DNF-471-0500
Please login to add product to your favorites.

Fragment Analyzer RNA Kits

Part Number:
DNF-471-0500
RUO

RNA Kit (15NT), 500. Fragment Analyzer system kit for the quantitative and qualitative analysis of total RNA from 200 to 6000 nt. Includes reagents necessary for running 500 samples.

For Research Use Only. Not for use in diagnostic procedures.
Item successfully added to your list

Subscribe to this item in cart or checkout

More lab efficiency with your auto delivery schedule, modify and cancel it at any time. Simply select subscription delivery frequency in the cart or checkout, and submit your order. How does it work?

List Price:
$1,686.00 /1 Kit

Please enter valid quantity

Specifications
Analysis Time
  • Short array Total RNA: 40 min; Short array mRNA: 45 min
Assay Sensitivity
  • Standard
Detection Limit
  • 5 ng/µL
Guaranteed Shelf Life
  • Minimum 4 months
Qualitative Range
  • Total RNA: 5 - 500 ng/µL; mRNA: 1 - 100 ng/µL
Quantitative Accuracy
  • ± 20 %
Quantitative Precision
  • 10 % CV
Quantitative Range
  • Total RNA: 25 - 500 ng/µL; mRNA: 20 - 100 ng/µL
Sample Volume
  • 2 µL
Sizing Accuracy
  • Total RNA: ± 5 %; mRNA: ± 10 %
Sizing Precision
  • 5 % CV
Uses Per Kit
  • 500
Kit Contents
Part Number Product Description Quantity
DNF-265-0240 RNA Separation Gel, 240mL 1
DNF-300-0008 BF-25 Blank Solution, 8mL 1
DNF-355-0125 5x 930 dsDNA Inlet Buffer, 125mL 1
DNF-369-0004 RNA Diluent Marker (15 nt), 4 mL 3
DNF-382-U020 RNA Ladder, 20 µL 5
DNF-475-0050 5x Capillary Conditioning Solution, 50mL 1
DNF-497-0125 0.25x TE Rinse Buffer, 125mL 1
DNF-600-U030 Intercalating Dye, 30 µL 1
Support
User Manuals

Agilent DNF-471 RNA Kit (15 nt) Quick Guide for Fragment Analyzer Systems

This Quick Guide is intended for use with the Agilent 5200, 5300, and 5400 Fragment Analyzer systems only. This kit is designed to detect Total RNA within the range of 5 ng/μL to 500 ng/μL input sample concentration and IVT RNA in the range of 1 ng/ μL to 100ng/ μL