When you turn on the Cary power, it makes a screeching noise or initialization has failed errors appear. The errors may include:
-
Error code :9:9305EQ: E9305 unable to initialize instrument Fault reported by the instrument while calibrating peak not detected or too small
-
Error code :9:9301SEQ:E9301 Internal system error
seq_cc_invalid_dic
See INITIALIZATION FAILURE (E9300 below)
-
Error code :9:9302SEQ:E9302 Unable to initialize instrument
Fault reported by instrument while calibrating, shutter open
See INITIALIZATION FAILURE (E9300 below)
-
Error code :9:9303SEQ:E9303 Unable to initialize instrument
Fault reported by instrument while calibrating, peak transmission too small.
See INITIALIZATION FAILURE (E9300 below)
-
Error code :9:9304SEQ:E9304 Unable to initialize instrument
Fault reported by instrument while calibrating, peak transmission too large
See INITIALIZATION FAILURE (E9300 below)
-
Error code :9:9619SEQ:E9619 Internal system error.
Fault reported by the Instrument nmp_cc_datum_not_found.
Monochromator intialization failed
See INITIALIZATION FAILURE (E9300 below)
-
Error code :9:9328SEQ:E9328 fatal initialization
Fault reported by the instrument software while calibrating.
See INITIALIZATION FAILURE (E9300 below)
-
E9300 Internal system error
INITIALIZATION FAILURE
The instrument has failed to initialize properly. During the initialization process the instrument has failed to locate the deuterium 656.1 nm emission line. The following actions are recommended:
-
Make sure that the rear beam is empty.
-
Make sure that the sample compartment lid is closed.
-
Re-power the instrument
-
If the initialization still fails, please check the following:
-
Check the lifetime of the lamp.
-
Older lamps with weak signals can sometimes cause failures.
-
Check the alignment of the deuterium lamp.
-
A misaligned lamp can also cause failure.
-
Optimize the lamp for maximum transmission (%t).
Note: If the instrument fails to find the deuterium emission line, wavelength calibration will be performed using the visible lamp at Zero order. The instrument can still be used. However, the wavelength accuracy will be reduced.
|