HHF100019 for Data Handling MSWS 6.8 (.exe, 1.3MB)
Changes/Corrections:
Issues resolved in this hotfix 100019 for MS Workstation 6.8:
- Fix for Incorrect S/N values reported for Compound peaks that cross segment boundaries - fix for S/N values were reported for Target compounds in the following situations when the integrated peak spanned a data file segment boundary:
- Quantitation used Peak-to-Peak Noise
- Quantitation used either Peak-to-Peak or RMS Noise and the Peak Retention Time was in a different data file segment than the Method Expected Retention Time.
- Fix for Zooming chromatogram segment with more than 40000 points - if a chromatogram segment contains more than 40,000 points and is displayed in single-segment mode, MSDR would crash if the chromatogram was zoomed untile the time range is so small that individual points should be displayed.
This Hotfix can only be installed on MSWS 6.8