This can be a sample preparation or system cleanliness problem. Try new solvents, both the ones for sample and blank preparation, as well as the solvents in the sample wash bottles. Try a new syringe and septum. Remove and clean the split vent line. Make a run without any injection, to see if the peaks show up just by heating the column.