Low Peak Area or Height | If using an inlet in split mode, check the split ratio. |
| Check the inlet for contamination. |
| Check each column and verify that it was cut and installed properly at each end. |
| Verify that the column type is correct. |
| Perform column maintenance: Bake out contaminants, remove the contaminated length of column near the inlet, and reverse and bake out the column as needed. |
| Verify that the liner type is appropriate for the sample. |
| Verify that the detector flow settings are correct. |
| Check the supply gas purity. |
| Check all trap indicators and dates. |
| Verify that the method parameters are correct. |
| Check sample stability. |
| Check configured syringe size. Some syringe sizes are specified at half-capacity. If the maximum syringe volume is marked at half-height on the barrel, not at the top of the barrel, enter twice the labeled volume when configuring the syringe size. |
If using an FID: | Verify that the correct jet is installed. |
If using a uECD: | Replace the fused silica indented mixing liner. |
| Replace and reinstall column. |
| Clean the makeup gas adapter. |
If using an NPD: | Check the detector for contamination. |
| Replace ceramic insulators. |
If using an FPD: | Verify correct column installation. |
| Check that the correct filter is installed and is clean. |
| Check the makeup gas type. |
|