- Cool all temperature zones and power off the GC. Turn off gas flows.
- Remove the autoinjector tray and bracket.
- Remove the left side panel.
- Remove the 1/8'' brass swagelock nut from the injection port. Trace the split vent line to the solenoid valve and remove it.
- Using the solvent(s) the majority of your samples are dissolved in, using vacuum or pressure, force the solvent through the tube until the solvent runs clear. Use a solvent with the highest vapor pressure as the last rinse.
- Draw or force air through the tube to dry it, then reassemble the GC.
- Do not heat the injection port or oven until carrier flow has been established for 10 minutes.
|