How are timepoints programmed in the Apparatus 3 and 7?
 
Apparatus 3 and Apparatus 7 timepoints are programmed differently than Apparatus 1/2/5/6.  Since Apparatus 3 and Apparatus 7 utilize different rows, timepoints are programmed independently for each row rather than a continuous running time as with the Apparatus 1/2/5/6.  If Apparatus 3 and Apparatus 7 timepoints are desired for 1, 2, 3, and 4 hours for example you would program 1 hour, 1 hour, 1 hour, 1 hour for the timepoints.
 
Should you need assistance with this or any other Agilent Dissolution Systems product, contact Agilent’s Technical Support team at +1-800-227-9770, select option 3, and then option 2, to be connected to the correct assistance team member. You may also contact our Dissolution Hotline at dissolution.hotline@agilent.com