Applications
Identification of Contaminants

This phenol product contained small amounts of two contaminant species. By using very high loading, sufficient quantities of fractions 1 and 3 were collected to permit identification by infra-red spectroscopy.
Key
1 & 3 Contaminants
2 Phenol
|
Column: |
PLgel 10 µm 100Å , 2 x 300 x 25.0 mm |
|
Eluent: |
Acetone |
|
Flow Rate: |
10.0 mL/min |
|
Loading: |
100 mg/mL (200 g on column) |
|
Detector: |
RI |
Product Deformulation

Loading in preparative GPC is molecular weight dependent and is much greater for low MW materials than for polymers. Even for polymers, a minimum of 10x scale up is routine but for very low molecular weight materials, the loading can be significantly increased.
|
Column: |
PLgel 10 µm 100Å , 2 x 300 x 25.0 mm |
|
Eluent: |
THF |
|
Flow Rate: |
9.0 mL/min |
|
Loading: |
1. 5 mg/mL, 1 mL (5 mg on column) 2. 100 mg/mL, 3 mL (300 mg on column) |
|
Detector: |
RI |
Fractionation

Each fraction of this phenanthrene-formaldehyde resin was collected using a 'heart-cut' technique to avoid cross-fraction contamination. Identification was carried out using mass spectroscopy and infra-red spectroscopy.
|
Column: |
PLgel 10 µm 500Å , 2 x 300 x 25.0 mm |
|
Eluent: |
Dichloromethane |
|
Flow Rate: |
9.0 mL/min |
|
Loading: |
100 mg/mL, 2 mL (200 g on column) |
|
Detector: |
UV, 254 nm |