When required
If an application needs a different type of flow cell or the flow cell needs repair. 
Tools required
Two 1/4 inch wrenches for capillary connections
Parts required
Standard flow cell, 10mm, 13µl, 120bar, G1315-60012
Semi-micro flow cell, 6mm, 5µl, 120bar, G1315-60011
High pressure flow cell, 6mm, 1.7µl, 400bar, G1315-60015
500 nl flow cell, 10 mm, refer to "500 nl Flow Cell - Replacing or Cleaning"


  1. Press the release buttons and remove the front cover to have access to the flow cell area.
  2. Press the release button and open the flow cell door.
  3. Disconnect the cell inlet and the cell outlet capillary and the waste capillary from the capillary holder.
  4. Remove the flow cell by pressing the flow cell holder.
  5. The label attached to the flow cell provides information on part number, path length and maximum pressure.If you want to replace flow cell parts, see Repairing the Standard or Semi-Micro Flow Cell (Video Clip) or Repairing the Standard or Semi-Micro Flow Cell (Video Clip).
  6. Insert the flow cell into the cell compartment and close the cell door.
  7. Reconnect the waste capillary and the cell inlet and the cell outlet tubing to the capillary holder.
  8. To check for leaks, establish a flow and observe the flow cell (outside of the cell compartment) and all capillary connections.