When required
When poor injection-volume reproducibility
When leaking injection valveĀ 
Tools required
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
9/64 inch 15 cm long, T-handle hex key 8710-2394 (supplied in accessory kit)
Parts required
0100-1853 Vespel Rotor Seal for 0101-0921 injection valve (G1367A/68A) or
0100-1849 Tefzel Rotor Seal for 0101-0921 injection valve (G1367A/68A)
0100-2088 Vespel Rotor Seal for 0101-1050 injection valve (G1377A/78A)

  1. When opening capillary or tube fittings, solvents may leak out. Please observe appropriate safety procedures (for example, goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the solvent vendor, especially when toxic or hazardous solvents are used.
  2. The micro injection valve, 0101-1050 has no stator face.
  3. Open the front door.
  4. Remove all capillaries from the injection-valve ports with the 1/4 inch wrench.
  5. Unscrew and remove the three stator screws from the stator head with the 9/64 inch wrench.
  6. Ensure the stator face does not fall out of the valve.
  7. Remove the stator head, the stator face and the stator ring.
  8. Remove the rotor seal (and isolation seal if required).
  9. Install the new rotor seal and isolation seal (if required). The metal spring inside the isolation seal must face toward the valve body. In other words, the metal spring should not be visible when the isolation seal is installed.
  10. Reinstall the stator ring.
  11. Place the stator face on the stator head. The pins on the stator face must engage in the holes on the stator head.
  12. Install this stator head/face assembly on the injection valve. Tighten the screws alternately with the 9/64 inch wrench until the stator head is secure.
  13. Reconnect all the capillaries to the injection valve ports with the 1/4 inch wrench.
  14. Close the front cover.