Agilent

Software Status Bulletin


UV VIS ChemStation     G1115AA

Known Problem Report as of Mar 14 2017 10:27AM

Preface
-------------------------
This Software Status Bulletin (SSB) documents all known problems in the software
product designated above. The SSB is derived from Known Problem Reports (KPR)
which result from user problems that have been classified as documentation
problems or software defects. When a KPR is written, an identifying number is
assigned to it, and the KPR is added to the next edition of the SSB.

User inputs that have been classified as Enhancement Requests are not documented
in the SSB. User problems that have been submitted, but that have not been
classified by the time the SSB is generated are not included in the SSB.

How to use the SSB
-------------------------
When you experience a problem with a product, first check this SSB to see if the
problem has been reported already, and if there is a temporary workaround
available for the problem, or if the problem has already been fixed by a new
revision. If the problem is not listed in this SSB then you may wish to report
it to the Response Center or to your field support representative.


To determine if your problem is documented in this SSB, first look in the
Keyword Index section of the SSB. Under each keyword is a listing of one-line
descriptions of related KPRs. If any of these sound like yours, locate the KPR
# in the Known Problem Reports section of the SSB, and read the full KPR. The
KPRs in the Known Problem Reports section are sorted by KPR #.


There are two sections in the SSB:
Keyword Index: This index is categorized by keyword. For each KPR there is a
brief description and a KPR #. A KPR may be associated with more than one
keyword.


Known Problem Reports: This section contains KPRs, with all the available
information relevant to the problem. KPRs in this section are sorted by KPR #.


Keyword Glossary



8453 Control
Autosampler
Communications
Configuration
Data Analysis
G1115A
Graphics/Views
Import/Export
Installation
Instrument Setup
Load/Save
Measure
Method
OQ/PV
Print
Programming
Reports

Keyword Index


Keyword: 8453 Control

KPR#:746 Lamp switching via lamp time table generates error on blank measurement

Keyword: Autosampler

KPR#:143 Occasionally MCTs are loosing their correct stepper motor positions
KPR#:214 XY-Autosampler x vial number increment does not work with all rack codes
KPR#:308 Gilson may be initialized when changing unit id in offline version
KPR#:722 No XY autosampler control in Windows VISTA

Keyword: Communications

KPR#:255 Loss of communication when using Win95 and HP 82341C card on a Kayak PC
KPR#:582 Selftest UV-Vis 8452A hangs up

Keyword: Configuration

KPR#:128 The ChemStation s/w does not recognize that the lamps are switched on.
KPR#:283 Page Header Text is truncated in the report
KPR#:311 The Selection Box for the GPIB address is greyed out during Configuration.
KPR#:500 instruments runs into error

Keyword: Data Analysis

KPR#:124 Labelling of Peaks and Valleys is incomplete
KPR#:141 Data Analysis: Equation: The units of the analyte value are different
KPR#:144 No dilution correction in the standard mode (Ratio/Equation)
KPR#:212 Change of a method is not indicated after deleting standards
KPR#:215 Annotation of Peaks and Valleys not always complete
KPR#:280 No calibration possible if parentheses () are used in analyte name
KPR#:281 X-axis scale of math results does not match x-scale of sample spectra.
KPR#:288 Results are displayed with an accuracy that is too high
KPR#:289 Wrong Initial Rate calculation if start time is other than zero
KPR#:549 Paste of blank spectrum to math result window does not work
KPR#:755 No transmittance data calculated
KPR#:767 Update of calibration view does not work always automatically

Keyword: G1115A

KPR#:499 Lamp counter data is reversed in Verification reports

Keyword: Graphics/Views

KPR#:137 Date field truncated in Verification and Diagnostics, Information, Histories
KPR#:164 Wrong enumeration in OQ/PV summary table
KPR#:287 Wrong title for overlaid spectra in fixed standard mode reports
KPR#:307 Wavelength range settings in Spectrum/Peaks Task are not updated

Keyword: Import/Export

KPR#:250 No warning when DIF Export function stops saving data due to full disk

Keyword: Installation

KPR#:749 Dialog text partially in German
KPR#:758 Misleading ChemStation.ini information

Keyword: Instrument Setup

KPR#:720 Wrong Dialog Description in LAN interfaced instrument setup

Keyword: Load/Save

KPR#:50 Loading sample leads to error message: The object is not a UV-Vis Spectrum
KPR#:258 Annotations not saved.
KPR#:263 sw truncate to 8 characters
KPR#:719 Task no adjusted when results file loaded

Keyword: Measure

KPR#:286 The Automation Run aborts if a read-only file is specified.

Keyword: Method

KPR#:146 Spectra Display axis is not stored with the method.
KPR#:184 Line breaks in multiple line comments are ignored
KPR#:195 Blank has to be measured twice after switching to a method with only one lamp.
KPR#:196 Wrong lamp status is displayed if the lamp time table is used.

Keyword: OQ/PV

KPR#:142 No acceptance criterion defined for OQ/PV of multicell transports
KPR#:165 Error Message during Validation of 8452A
KPR#:221 Intensity test in diagnosics indicates error

Keyword: Print

KPR#:27 Overlaid printed spectra are not visible
KPR#:95 Entered sample comments are not printed in results report
KPR#:114 Set up wavelength analysis with list of wavelengths. Overlay spectra on
KPR#:123 The print out of Lamp Counters are truncated
KPR#:279 Missing characters in report header.
KPR#:598 Printing overlaid spectra in color does not work

Keyword: Programming

KPR#:285 The loading of a macro fails.

Keyword: Reports

KPR#:754 Method information missing in various different reports


KPR#:4  Product:UV VIS ChemStation  G1115AA  A.02.05

Keyword: 

One-line Description:

If a run starts at 1:00PM, the first data point

Problem:

In the G1115A UV-Vis Software is no kinetics acquisition possible. The BUG B400-106740 belongs to the old revision 2.04 and is fixed with Rev. 2.05.

Temporary Solution:

n/a

Fix Information:

Replace software with current revision 2.05


KPR#:225  Product:UV VIS ChemStation  G1115AA  A.02.04

Keyword: 

One-line Description:

"EdTab 100: item not found..." error message after aborting automation

Problem:

If the an automated run using the multicell transport is aborted then the error message "EdTab 100: item not found in data table" sometimes occurs. If automation is started again the same message appears after the first "Preparing automation" message.

Temporary Solution:

None. The error message is harmless.

Fix Information:

Fix under investigation


KPR#:251  Product:UV VIS ChemStation  G1115AA  A.02.04

Keyword: 

One-line Description:

Software does not run properly using italian Windows version

Problem:

Software does not run properly using italian version of windows. Problems occur in the Verification and Diagnostics MODE of the software.

Temporary Solution:

Use US english version of windows and set country and language options to "United States" and "English".

Fix Information:

No fix planned


KPR#:252  Product:UV VIS ChemStation  G1115AA  A.02.04

Keyword: 

One-line Description:

No automatic annotation in hardcopies.

Problem:

In "Print Selected Window" and "Print View" menu tasks no automatic annotation of graphs is done.

Temporary Solution:

Use "Annotate..." menu task of "Edit" menu to create annotation text.

Fix Information:

No fix planned


KPR#:752  Product:UV VIS ChemStation  G1115AA  B.02.01 SP1

Keyword: 

One-line Description:

Wrong revision display in B.02.01 installation

Problem:

For the qualification of the software the respective revision needs to be checked after the installation.
On the target system the revision was displayed as Rev. B.04.01 [0].

Temporary Solution:

n/a

Fix Information:

Not a defect.  This version was the version number of the IQ software. This now contains the correct UV-Visible version, e.g. B.02.01


KPR#:756  Product:UV VIS ChemStation  G1115AA  B.02.01

Keyword: 

One-line Description:

Wrong date information displayed in instrument logbooks

Problem:

The date column is to small to display and print the date information properly.
The data displayed is e.g. 1/11/2008 rather than 11/11/2008.

Temporary Solution:

n/a

Fix Information:

Modified diagnosis register and adjusted column width from 9 to 11.


KPR#:764  Product:UV VIS ChemStation  G1115AA  B.02.01

Keyword: 

One-line Description:

UV-visible ChemStation fails to start and locks up

Problem:

Eventually a UV-visible ChemStation installation of revision B.02.01 in the Windows XP environment fails to start. During the initialization the system displays the error message "C:\chem32\UvExe\agtuvvus.dlc failed. (in load)" and locks up.

Temporary Solution:

n/a

Fix Information:

Fixed with code and setup changes .


KPR#:768  Product:UV VIS ChemStation  G1115AA  B.02.01 SP1

Keyword: 

One-line Description:

Tracking of deleted standards does not work

Problem:

In the security pack environment no raw data should be permanently deleted form a record. In standard mode this does not work for standard data. If the delete operation is performed, no dialog specifying a reason and the tracking in with the deleted spectra view is available. Standard data are permanently removed.

Temporary Solution:

When loading more than one results files in a sequence, the clear standards function is still offered. In addition the ability to overwrite the existing record is offered.
See also TeamTrack ID 796 of the same project.

Fix Information:

Fixed in code.


KPR#:769  Product:UV VIS ChemStation  G1115AA  B.02.01

Keyword: 

One-line Description:

Wrong date in instrument status displayed

Problem:

When using the instrument status window, the "Power on time" is displayed completely wrong. The date is in the future and unrelated to the PC/instrument time

Temporary Solution:

n/a

Fix Information:

The behavior of localtime() and localtime_s() are a little bit different, making some code changes necessary.


KPR#:771  Product:UV VIS ChemStation  G1115AA  B.02.01

Keyword: 

One-line Description:

Invalid data not printed correctly

Problem:

In the Spectrum/Peak Report invalid or not available values should be indicated using "---" as with revisions prior to B.02.01. However, now various different values like zeros or -1.7977E308 are printed.

Temporary Solution:

n/a

Fix Information:

The core problem is the invalid value for displaying ‘***’ in tables.  The default value can’t be ‘***’ (= invalid). The value can be set, but is rounded to the smallest valid value. 

The original UV-Visible code is setting the default value for the absorbance column to ‘***’ (= invalid). After setting the default value all values are set with there absorbance. 

The new code is now setting the absorbance column to ‘***’ (= invalid) for all peak rows, and after this setting the values to the absorbance.


KPR#:773  Product:UV VIS ChemStation  G1115AA  B.02.01

Keyword: 

One-line Description:

WordPad dialog partially in German language

Problem:

After the installation of the UV-visible ChemStation software B.02.01 on the supported operating systems, the Print Preview dialog of WordPad is displayed with German command buttons.

Temporary Solution:

n/a

Fix Information:

Removed inappropriate language files in Windows.


KPR#:776  Product:UV VIS ChemStation  G1115AA  B.02.01 SP1

Keyword: 

One-line Description:

Results not saved in automated run

Problem:

When checking "Save results to file:" with a long filename no data is saved and no warning is displayed.

Temporary Solution:

n/a

Fix Information:

Fixed in code.


KPR#:781  Product:UV VIS ChemStation  G1115AA  B.02.01 SP1

Keyword: 

One-line Description:

Update of calibration view does not work always automatically

Problem:

When manually switching to the standard view in the UV-visible ChemStation's Quantification task, this view is not properly updated. Rather than displaying the standards and a calibration curve, sample data may appear with the "Processed Standard Spectra" window and the "Calibration Curve" window.
When a Standard spectrum is acquired the two windows might be empty.

Temporary Solution:

n/a

Fix Information:

Fixed in code.


KPR#:792  Product:UV VIS ChemStation  G1115AA  B.04.01

Keyword: 

One-line Description:

Remove "untitled" from reports where information is unknown

Problem:

Change the Uv-Vis code to eliminate the string "untitled" from any report where the data, method, or any other file has not been saved or is unknown.  This is a customer request - In a regulated environment it never looks good to have "untitled" in a report.  This generates a false impression that the data was lost or not acquired correctly.

Also,  when saving data to ECM add ECM location and filename to report.

Temporary Solution:

n/a

Fix Information:

Changed code to say "Results not yet saved" unstead of "untitled" on reports.


KPR#:793  Product:UV VIS ChemStation  G1115AA  B.04.01

Keyword: 

One-line Description:

Advanced Analysis - Wavelength step will not accept small steps less than .2

Problem:

Any wavelength step under .2nm causes report generation to fail. Problem is located in the ChemStation Core command "DataGetMinMax"

Temporary Solution:

n/a

Fix Information:

Fixed to allow small incremental steps.


KPR#:795  Product:UV VIS ChemStation  G1115AA  B.04.01

Keyword: 

One-line Description:

Keyboard Cursor Arrows not working to traverse graph

Problem:

In B.04.01 Uv-Vis the keyboard arrows do not work to move the pointer in small steps to different locations on the spectrum graph.  This worked in previous versions and is a useful feature.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:796  Product:UV VIS ChemStation  G1115AA  B.02.01 SP1.1

Keyword: 

One-line Description:

ECM not working properly with BPM client 3.4.0

Problem:

A certain version of the BPM client that does not work correctly.  ECM and BPM client 3.3.1 installed on Windows 7 system seem to work fine.  However there seems to be a problem on a ECM 3.3.2 client along with a BPM 3.4.0 client.

Temporary Solution:

n/a

Fix Information:

Fixed in code B.04.02


KPR#:798  Product:UV VIS ChemStation  G1115AA  B.02.01 SP1

Keyword: 

One-line Description:

Password longer than 9 characters crashed ChemStation Session

Problem:

The access to the manager mode of the UV-visible ChemStation can be secured by providing a password. In the entry dialog for the password no length limit is given nor is the entry length limited to less than 10 characters. If a password exceeding 9 characters is specified, a dialog box pops up indicating a Chemstation execution error. Even a new launch of the ChemStation fails.

Temporary Solution:

n/a

Fix Information:

Fixed in SI62.
Fix done in file password.cpp, part of agtuvvis.dlc.


KPR#:799  Product:UV VIS ChemStation  G1115AA  B.04.01

Keyword: 

One-line Description:

Kinteics Single Cell Measurements Can Lose Data Over Long Tests

Problem:

In Kinetics mode, a single cell setup can "lose" data over long tests.  This is a firmware problem where the interval time counter apparently overflows and causes random times to be associated with the data.  This jumbles the data and causes an empty gap when looking at the results.  Note that the interval timer is reset with each measurement in Multicell sample setups, so the problem does not exist in Multicell.  (Workaround for a single cell is to run in Multicell mode with one cell in position one).

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:800  Product:UV VIS ChemStation  G1115AA  B.04.01

Keyword: 

One-line Description:

Need to be able to export Sample and Standard Data from Tables

Problem:

Currently the only way to export sample and standard attenuation data is with the focus on the graph window.  If a spectra is selected from a table without focus on the graph window, only the hidden table data (not from spectrum) is exported.  Need to be able to highlight a sample or standard in a table and export same as if highlighted in the graph.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:806  Product:UV VIS ChemStation  G1115AA  B.04.02

Keyword: 

One-line Description:

Selecting XY Autosampler, or Gilson Autosampler in Windows 7 hangs ChemStation

Problem:

Selecting XY Autosampler, or Gilson Autosampler in Windows 7 hangs ChemStation after getting the message "This application is for Windows NT, 2000, or XP".    Selection on an XP computer works ok.

Temporary Solution:

n/a

Fix Information:

Update to Gilson GSIO library files version 2.0


KPR#:808  Product:UV VIS ChemStation  G1115AA  B.02.01 SP1

Keyword: 

One-line Description:

Change of integration time does not invalidate last blank

Problem:

When changing the integration time parameter, a new reference measurement (blank) must be forced. Usually this is achieved by clearing the last blank measurement. This works for changes of the lamp status, but not when changing the integration time parameter.

Temporary Solution:

n/a

Fix Information:

Fixed in code B.04.02


KPR#:809  Product:UV VIS ChemStation  G1115AA  B.04.01

Keyword: 

One-line Description:

Incorrect measurement counter of standard results during automation

Problem:

1. Start UV-ChemStation and switch to Standard Mode.
2. Setup a simple Method and Automation.
3. Open Automation Setup dialog and check “Save results to file”. 
4. Enter a file name which is shorter than 40 characters (“StdMode_Automation”)
5. Run Automation 2 times.
6. Open windows explorer and check filenames of the 2 automations.

Two files were created. Both starts with the entered filename:  “StdMode_Automation”. Attached is a counter which fills up the filename with 0s until the filename contains 40 characters starting with 0. Here: “StdMode_Automation0000000000000000000000.SR” and “StdMode_Automation0000000000000000000001.SR”.

7. Open Automation Setup dialog again and change filename to “StdMode_Automation_HHHHHHHHHHHHHHHHHHHHH” (40 characters, the maximum).
8. Run Automation 2 times again.
9. Open windows explorer again and check filenames of the 2 automation.

Now only one file was created with name  “StdMode_Automation_HHHHHHHHHHHHHHHHHHHHH.SR”. No counter was attached to the filename. The second automation overwrites the results of the first automation.

Temporary Solution:

n/a

Fix Information:

Fixed in code B.04.02


KPR#:810  Product:UV VIS ChemStation  G1115AA  B.02.01 SP1

Keyword: 

One-line Description:

Automation Results printing command not active in Standard mode

Problem:

When a results file generated by an automated sequence is loaded into standard mode, the printing of Automation Results is grayed out. Therefore reprinting of the statistical results is not possible.

Temporary Solution:

n/a

Fix Information:

Fixed in code B.04.02


KPR#:812  Product:UV VIS ChemStation  G1115AA  B.04.01

Keyword: 

One-line Description:

No information to user if saving results to ECM failed

Problem:

UV-ChemStation installed with ECM.

1. Start online ChemStation and switch to Dissolution Testing Mode.
2. Setup a Dissolution Run with 'Save Dissolution Result to ECM' Action in PostRun Sequence.
3. Start Dissolution Run and select result file which is checked out by a different user.

After Dissolution Run upload of result file to ECM failed. No MessageBox will be shown. Only a logbook entry was created which says that the result file was saved to disk.
In Advanced Mode a MessageBox appears

Temporary Solution:

n/a

Fix Information:

This error is now handled and a message box appears.


KPR#:813  Product:UV VIS ChemStation  G1115AA  B.02.01

Keyword: 

One-line Description:

Run logbook and Deleted Spectra missing in results report

Problem:

In the Security Pack environment the run logbook and the deleted spectra (if any) should be printed as part of the results report.

Temporary Solution:

n/a

Fix Information:

Fixed in code B.04.02


KPR#:816  Product:UV VIS ChemStation  G1115AA  B.04.01

Keyword: 

One-line Description:

Error Printing Results in Dissolution Mode

Problem:

1. Start offline UV-ChemStation and switch to Dissolution Testing Mode.
2. Load existing Dissolution Result file.
3. Print result report to screen.

When you want to print the result, than you get a printer error.
This is caused by the default reportstyle. If you create your own reportstyle you will be able to print the dissolution result without any problems.

Temporary Solution:

n/a

Fix Information:

Fixed in code B.04.02


KPR#:817  Product:UV VIS ChemStation  G1115AA  B.02.01 SP1

Keyword: 

One-line Description:

Lamp settings may change when switching mode

Problem:

When switching from Verification&Diagnostics mode to a different application, the lamps might be switched on/off without any warning.

Temporary Solution:

n/a

Fix Information:

Fixed in code B.04.02


KPR#:818  Product:UV VIS ChemStation  G1115AA  B.04.01

Keyword: 

One-line Description:

Possible to save methods to ECM with different file extensions.

Problem:

UV-ChemStation installed with ECM-Plugin.

1. Start ChemStation offline.
2. Switch to Standard Mode and create new method (File -> New Method).
3. Change method settings (Setup-Button from GUI).
4. Save Method to ECM (ECM -> Save Method As...).
5. Enter file name with file extension different from '.M' (e.g.: test.x).

ChemStation saves the method to ECM with entered file extension and shows filename in GUI. It is not possible to load this method again from ECM with ChemStation because 'ECM Desktop Open'-dialog only shows files with extension '.M'.

Temporary Solution:

Workaround: Download method file via Internet Explorer, change file extension to '.M' and upload file via Internet Explorer again.

Fix Information:

Fixed in code B.04.02


KPR#:819  Product:UV VIS ChemStation  G1115AA  B.02.01 SP1

Keyword: 

One-line Description:

ReadPort and WritePort commands not available

Problem:

A customer macro does not run anymore due to non existing functions/commands. The help text describes the command syntax, but an error message is displayed when using the command.

Temporary Solution:

n/a

Fix Information:

Not available in ChemStation core.  Removed from help file


KPR#:746  Product:UV VIS ChemStation  G1115AA  B.02.01

Keyword: 8453 Control

One-line Description:

Lamp switching via lamp time table generates error on blank measurement

Problem:

When using the lamp timetable to switch on one e.g. the Tungsten lamp, the GUI and the Instrument Status are displaying the correct status, however when trying to measure a blank using the blank button an error message "Tungsten lamp not ready!" is displayed.

Temporary Solution:

n/a

Fix Information:

The lamp state of the method register was updated to handle lamp switch events from the 8453. 

The same problem occured when switching the UV lamp. This is fixed too.


KPR#:143  Product:UV VIS ChemStation  G1115AA  A.06.03

Keyword: Autosampler

One-line Description:

Occasionally MCTs are loosing their correct stepper motor positions

Problem:

The 7- and 8-position multicell transports are loosing sporadically their correct stepper motor positions. 
When trying to fix the problem by performing the MCT recalibration procedure of the Verification and Diagnostics mode (Task: Maintenance -> MCT Test/Recalibration -> 'Start Test') a white bar shows up for the related MCT position indicating low intensity. In this case it might not be possible to realign the MCT by clicking the 'Recalibrate' button.
Pressing the 'Default Positions' button to set the MCT stepper motor positions back to the factory values might also fail under these conditions

Temporary Solution:

To retrieve the factory default settings for the multicell transport a forced cold start has to be performed (refer to the service manual G1103-90003, page 147.)
Note: It is recommended to perform a wavelength recalibration after the forced cold start is done.

Fix Information:

Fix under investigation.


KPR#:214  Product:UV VIS ChemStation  G1115AA  A.06.02

Keyword: Autosampler

One-line Description:

XY-Autosampler x vial number increment does not work with all rack codes

Problem:

The XY-Autosampler setup offers the choice between incrementing the y vial number or the x vial number first. The option to start incrementing the x vial number first does not work with rack code 21. Using the Autosampler Control dialog, negative vial numbers are displayed. To leave this dialog, a valid vial number must be entered first.

Temporary Solution:

Use y vial number increment only for racks of type 21 or use racks of type 22.

Fix Information:

Fix planned.


KPR#:308  Product:UV VIS ChemStation  G1115AA  A.02.03

Keyword: Autosampler

One-line Description:

Gilson may be initialized when changing unit id in offline version

Problem:

If an offline version of the software is started with a Gilson autosampler connected, changes to the Gilson id will cause the Gilson to initialize. This could be a problem when running an off-line version simultaneously with an on-line version that is currently executing and automation on the Gilson.

Temporary Solution:

n/a

Fix Information:

No fix planned.


KPR#:722  Product:UV VIS ChemStation  G1115AA  B.02.01

Keyword: Autosampler

One-line Description:

No XY autosampler control in Windows VISTA

Problem:

The installed GSIOCServer does not run on VISTA

Temporary Solution:

n/a

Fix Information:

Fixed with new GSIOC Server (GSIOC Server 2.0). The new GSIOC Server is included in the UV-Visible setup.


KPR#:255  Product:UV VIS ChemStation  G1115AA  A.06.01

Keyword: Communications

One-line Description:

Loss of communication when using Win95 and HP 82341C card on a Kayak PC

Problem:

When using Windows 95 as operating system on a Kayak PC, the communication between the HP8452A/HP8453 spectrophotometer and the HP ChemStation software may be lost, if a HP 82341C HP-IB card is used. In case of the HP8453 spectrophotometer the error message "Power Fail" is shown.

Temporary Solution:

Replace the HP 82341C HP-IB card by a HP 82341D card.

Fix Information:

No fix planned.


KPR#:582  Product:UV VIS ChemStation  G1115AA  B.01.01

Keyword: Communications

One-line Description:

Selftest UV-Vis 8452A hangs up

Problem:

The communication with the HP8452A shows sporadic errors. These errors are occuring in particular when using continous selftests.

Temporary Solution:

Communication with the HP8452A should be established using the USG to GPIB converter interface 82357A.
The build-in PC GPIB interfaces 82350A/B are too fast for establishing proper communication to the HP8452A.
In case of the occurence of a communication problem, the HP8452A should be power cycled and the PC restarted.

Fix Information:

No fix planned.


KPR#:128  Product:UV VIS ChemStation  G1115AA  A.02.05

Keyword: Configuration

One-line Description:

The ChemStation s/w does not recognize that the lamps are switched on.

Problem:

If the ChemStation software is booted before the spectrophotometer is started, the lamp settings in the Lamp(s) Parameter dialog box (Instrument -> Lamp(s) ...) are automatically set to 'OFF'. After starting the spectrophotometer the software does not recognize when the lamps are automatically switched on, i.e. the lamp settings are may still be 'OFF' although the lamps are on. If the lamp settings are switched to 'ON' the current method will be flagged as modified which may lead to inconsistencies.

Temporary Solution:

Always start the spectrophotometer first and then boot-up the ChemStation software.

Fix Information:

Fix under investigation.


KPR#:283  Product:UV VIS ChemStation  G1115AA  A.05.02

Keyword: Configuration

One-line Description:

Page Header Text is truncated in the report

Problem:

In the Config Page Header Text, but only a limited number of characters are printed in the report.

Temporary Solution:

Choose a short Page Header Text.

Fix Information:

No fix planned.


KPR#:311  Product:UV VIS ChemStation  G1115AA  A.09.01

Keyword: Configuration

One-line Description:

The Selection Box for the GPIB address is greyed out during Configuration.

Problem:

If you want to define the GPIB address of an instrument in the Configuration Editor, the Selection Box is greyed out.

Temporary Solution:

Click on the GPIB radio button to activate the selection box for the GPIB address.

Fix Information:

No fix planned.


KPR#:500  Product:UV VIS ChemStation  G1115AA  B.01.01

Keyword: Configuration

One-line Description:

instruments runs into error

Problem:

The communication speed using GPIB cards with PCI slots integrated to the PC mainboard is too fast to properly communicate with the HP8452A. When using the UV-visible ChemStation on-line, sporadic errors may occur and the communication to the instrument does not terminate.

Temporary Solution:

The USB to GPIB converter interface 82357A can be used to control the HP8452A. Probably the interface conversion delay cures the communication speed problem.

Fix Information:

No fix planned


KPR#:124  Product:UV VIS ChemStation  G1115AA  A.02.05

Keyword: Data Analysis

One-line Description:

Labelling of Peaks and Valleys is incomplete

Problem:

The result of the Valley/Peak Math-function are completely shown in the Peaks/Valleys tables, but the spectrum shown is not labelled completely.

Temporary Solution:

n/a

Fix Information:

Fix under investigation.


KPR#:141  Product:UV VIS ChemStation  G1115AA  A.08.03

Keyword: Data Analysis

One-line Description:

Data Analysis: Equation: The units of the analyte value are different

Problem:

Samples are already loaded in the Standard Mode. If the units Weight (Wt) and volume (V) are changed in the Ratio/Equation task and
then another set of Samples are  loaded, an red line error message appears:
The values of the analyte value are different.

Temporary Solution:

Make sure that only data acquired with identical weight/volume or concentration units are combined. Do not change these settings during an analysis sequence.

Fix Information:

Fix under investigation.


KPR#:144  Product:UV VIS ChemStation  G1115AA  A.06.03

Keyword: Data Analysis

One-line Description:

No dilution correction in the standard mode (Ratio/Equation)

Problem:

After the dilution factor has been changed in the sample result table of the Ratio/Equation task and the 'Analyze' button is selected again, the result does not change.

Temporary Solution:

Edit the dilution factor in the equation field of the Ratio/Equation Parameters dialog box. Select Ratio/Equation from the Method menu to
display this dialog box.

Fix Information:

Fixed in Revision A.09.01.


KPR#:212  Product:UV VIS ChemStation  G1115AA  A.06.03

Keyword: Data Analysis

One-line Description:

Change of a  method is not indicated after deleting standards

Problem:

Standard spectra are part of the method. However, the method modified flag in the GUI (Mod) does not appear after standards have been deleted.

Temporary Solution:

Save the changed method manually after deleting standard spectra. Use 
item "save method as..." in the file menu.

Fix Information:

Fixed in Revision A.09.01.


KPR#:215  Product:UV VIS ChemStation  G1115AA  A.08.03

Keyword: Data Analysis

One-line Description:

Annotation of Peaks and Valleys not always complete

Problem:

If several spectra are loaded in the Standard Mode (Spectrum/Peaks),
the peaks and valleys are correctly identified and listed in the Sample/Result Table. However the peaks and valleys are only
annotated for one spectrum in the graphical Spectrum window.

Temporary Solution:

Only load and display the spectrum of interest.

Fix Information:

No fix planned.


KPR#:280  Product:UV VIS ChemStation  G1115AA  A.05.02

Keyword: Data Analysis

One-line Description:

No calibration possible if parentheses () are used in analyte name

Problem:

If parentheses are used in an analyte name, the calibration fails. In addition, the units proposal in the Quantification setup dialog box is wrong. Even if corrected manually, no calibration is possible.

Temporary Solution:

Do not use parentheses () in analyte names.

Fix Information:

No fix planned.


KPR#:281  Product:UV VIS ChemStation  G1115AA  A.05.02

Keyword: Data Analysis

One-line Description:

X-axis scale of math results does not match x-scale of sample spectra.

Problem:

There is no means of adjusting the math results x-axis scale exactly to the scale of the samples.

Temporary Solution:

First apply the Math Function to your data, then save the generated spectra as a *.sd-file. Load the saved *.sd-file to overlay it to the sample spectra window.

Fix Information:

No fix planned.


KPR#:288  Product:UV VIS ChemStation  G1115AA  A.08.03

Keyword: Data Analysis

One-line Description:

Results are displayed with an accuracy that is too high

Problem:

The results in the sample results table are displayed with an
accuracy that is too high. The results are rounded and the droped
digits a filled with "0". For example, the rounded value in the
sample result table is shown as 1.92300, wheras the correct reading
is 1.922980.

Temporary Solution:

The standard deviation of the result is also displayed in the sample
results table, which clearly defines the accuracy of the measurement.

Fix Information:

No fix planned.


KPR#:289  Product:UV VIS ChemStation  G1115AA  A.06.05

Keyword: Data Analysis

One-line Description:

Wrong Initial Rate calculation if start time is other than zero

Problem:

The initial rate calculation does not adapt to a change of the calculation range. The calculation always starts at the first acquired data point.

Temporary Solution:

Do not use a start of the calculation other than at the first acquired data point.

Fix Information:

Fix under investigation


KPR#:549  Product:UV VIS ChemStation  G1115AA  B.01.01

Keyword: Data Analysis

One-line Description:

Paste of blank spectrum to math result window does not work

Problem:

The Copy, Cut and Paste commands from the edit menu do not work for spectra.

Temporary Solution:

The short-cut keys Ctrl-C (Copy), Ctrl-X (Delete) and Ctrl-V (Paste) can be used instead of the respective Edit menu commands.

Fix Information:

Fix under investigation.


KPR#:755  Product:UV VIS ChemStation  G1115AA  B.02.01

Keyword: Data Analysis

One-line Description:

No transmittance data calculated

Problem:

If data are acquired and the data type is set to transmittance an error "Transmittance, Spectrum 1, Factor is too large is displayed" may occur.

Temporary Solution:

Downgrade to UV-visible ChemStation revision B.01.01.

Fix Information:

Fix in code.


KPR#:767  Product:UV VIS ChemStation  G1115AA  B.02.01 SP1

Keyword: Data Analysis

One-line Description:

Update of calibration view does not work always automatically

Problem:

In the UV-visible ChemStation B.02.01 SP1 [50] the update of
the calibration view is not always done properly. In many
cases misleading information is displayed.

Temporary Solution:

n/a

Fix Information:

Fix under investigation


KPR#:499  Product:UV VIS ChemStation  G1115AA  B.01.01

Keyword: G1115A

One-line Description:

Lamp counter data is reversed in Verification reports

Problem:

There is no impact from a GLP side, the correct lamp burn times are reported, but they are reported in the wrong fields (switched) on the verification report in UV/VIS Chemstation
- The problem is being fixed and will be part of UV-Vis ChemStation revision B.04.01

Temporary Solution:

n/a

Fix Information:

Fixed.


KPR#:137  Product:UV VIS ChemStation  G1115AA  A.06.04

Keyword: Graphics/Views

One-line Description:

Date field truncated in Verification and Diagnostics, Information, Histories

Problem:

If the date formats MM/dd/yyyy or dd/MM/yyyy are selected using Windows NT/Control Panel/Regional Settings/Date, the entries are truncated in verification and diagnostics histories (e.g.  /09/1999).

Temporary Solution:

Use a date format with 8 characters only (e.g. mm/dd/yy)

Fix Information:

Fix planned.


KPR#:164  Product:UV VIS ChemStation  G1115AA  A.09.01

Keyword: Graphics/Views

One-line Description:

Wrong enumeration in OQ/PV summary table

Problem:

The validation summary table shows an enumeration different to the sequence used in the execution of the validation tests.

Temporary Solution:

Be aware that the wavelength accurracy and reproducibility test is referred to as test 2 in the summary and the photometric accurracy test is number 1.

Fix Information:

Fix under investigation


KPR#:287  Product:UV VIS ChemStation  G1115AA  A.08.03

Keyword: Graphics/Views

One-line Description:

Wrong title for overlaid spectra in fixed standard mode reports

Problem:

If you choose View Last Spectrum instead of Overlaid Spectra and
print a report in the standard mode, the title of the graph is not correct.
It still reads "Overlaid Spectra".

Temporary Solution:

n/a

Fix Information:

No fix planned.


KPR#:307  Product:UV VIS ChemStation  G1115AA  A.06.01

Keyword: Graphics/Views

One-line Description:

Wavelength range settings in Spectrum/Peaks Task are not updated

Problem:

After a doubleclick in the Sample Spectra window, the wavelength range of the spectrum display is changed to the full range (190 nm to 1100 nm).
It is not possible to retrieve the settings in the Spectrum/Peaks Parameter dialog box.

Temporary Solution:

Use the view menu's Reset Current View command to retrieve the method setup.

Fix Information:

No fix panned.


KPR#:250  Product:UV VIS ChemStation  G1115AA  A.02.05

Keyword: Import/Export

One-line Description:

No warning when DIF Export function stops saving data due to full disk

Problem:

The 'Export As .. Dif Format' function gives no warning when the disk is getting full. When activating the Export function a data file is created and data are stored until the disk is full. This can result in incompletely stored data.

Temporary Solution:

Check free disk space and file size to see if the file can be stored on the disk.

Fix Information:

No fix planned.


KPR#:749  Product:UV VIS ChemStation  G1115AA  B.02.01

Keyword: Installation

One-line Description:

Dialog text partially in German

Problem:

If .Net 2.0 does not exist on the target system, it is automatically installed during the setup of the ChemStation software. Part of the dialog displaying the license agreement including the buttons is in German.

Temporary Solution:

n/a

Fix Information:

Fixed setup program.


KPR#:758  Product:UV VIS ChemStation  G1115AA  B.02.01

Keyword: Installation

One-line Description:

Misleading ChemStation.ini information

Problem:

The content of the ChemStation.ini file is usually used to diagnose setup and configuration problems.
However the current information is misleading.
The revision is reported as B.01.01 P1 however it is a B.02.01 Patch 1 [45] setup reported by the "About" dialog.
The executable is indicated as HPUV-VIS.EXE, however a program called "launchcs.exe" is used to run the ChemStation.
The link information and setup of the serial communication is included, also this has been dropped with revision B.02.01.

Temporary Solution:

n/a

Fix Information:

Fixed by changing the setup script file.


KPR#:720  Product:UV VIS ChemStation  G1115AA  B.02.01

Keyword: Instrument Setup

One-line Description:

Wrong Dialog Description in LAN interfaced instrument setup

Problem:

In the configuration editor when setting up an 8453 connected via LAN, the dialog indicates a "Single Quad MSD".

Temporary Solution:

n/a

Fix Information:

While opening the configuration dialog a check on instrument type (eq. ‘UVVis’) is required. In This case the dialog box title was changed.


KPR#:50  Product:UV VIS ChemStation  G1115AA  A.08.03

Keyword: Load/Save

One-line Description:

Loading sample leads to error message: The object is not a UV-Vis Spectrum

Problem:

If the user selects a particular spectrum in the Sample/Result Table and saves this sample spectrum, it cannot be loaded again. Instead the red line error message "The object is not a UV-Vis Spectrum" is displayed.

Temporary Solution:

Select the sample spectrum in the Overlaid Sample Spectra window and choose Save Selected Spectrum as.. in the file menu.

Fix Information:

Fixed in Rev. A.09.01.


KPR#:258  Product:UV VIS ChemStation  G1115AA  A.06.01

Keyword: Load/Save

One-line Description:

Annotations not saved.

Problem:

When annotation has been added to the displayed spectra on the screen, this information is not saved with the data set. The annotation is available only during the ChemStation session.

Temporary Solution:

Only close you ChemStation session with annotated graphics if you have printed the annotated graph or do not need the annotation information any more.

Fix Information:

No fix planned


KPR#:263  Product:UV VIS ChemStation  G1115AA  A.05.03

Keyword: Load/Save

One-line Description:

sw truncate to 8 characters

Problem:

CANNOT EXPORT DATA UNDER FILENAMES NOT OBEYING THE 8.3 RULE (8 CHARACTER S FOR FILENAME AND 3 CHARACTERS FOR FILE EXTENSION) AND CANNOT EXPORT DA TA TO DIRECTORIES HAVING DIRECTORYNAMES LONGER THAN 8 CHARACTERS.

Temporary Solution:

WHEN USING DATA EXPORT FUNCTIONS IN THE VARIOUS UV-VIS CHEMSTATION MODES ENSURE TO USE FILENAMES OBEYING THE 8.3 CHARACTER FILENAME RULE AND THE 8 CHARACTER DIRECTORY NAME RULE. THIS RULE MUST BE OBEYED FOR WIN NT 4.0 AND WIN 95 OPERATING SYSTEMS.

Fix Information:

Fixed in A.06.01.


KPR#:719  Product:UV VIS ChemStation  G1115AA  B.02.01

Keyword: Load/Save

One-line Description:

Task no adjusted when results file loaded

Problem:

If a results file in the fixed wavelengths task is saved, after loading this file when the current task is on equation, no change is performed.

Temporary Solution:

n/a

Fix Information:

The result file format is changed from file version 1.0 to 1.1. The new result file format contains information about the current mode and the selected task.

Both, the mode and the task are checked while loading the result file. In case the current task does not match the task stored inside the result file the user is asked “The current task does not match the file info. Do you want to switch the task now to…“. 

The check for current mode is now also implemented, nevertheless only files containing the correct result file extension can be loaded.


KPR#:286  Product:UV VIS ChemStation  G1115AA  A.05.02

Keyword: Measure

One-line Description:

The Automation Run aborts if a read-only file is specified.

Problem:

The Automation run is aborted without any error message, if 'Save spectra to file' in the 'Automation Setup' dialog box is selected, and the specified file is marked as a read-only file.

Temporary Solution:

Make sure that the specified file is not marked as read-only.

Fix Information:

No fix planned.


KPR#:146  Product:UV VIS ChemStation  G1115AA  A.06.04

Keyword: Method

One-line Description:

Spectra Display axis is not stored with the method.

Problem:

In the Fixed Wavelength(s) Parameters dialog box, which is displayed by
selecting fixed wavelength in the Method menu, the operator can enter the wavelength range, that should be displayed. These settings are not stored with the method, though they are method parameters.

Temporary Solution:

n/a

Fix Information:

Fix planned.


KPR#:184  Product:UV VIS ChemStation  G1115AA  A.07.01

Keyword: Method

One-line Description:

Line breaks in multiple line comments are ignored

Problem:

Entering line breaks (Ctrl+Enter) are ignored in the printout.

Temporary Solution:

n/a

Fix Information:

No fix planned.


KPR#:195  Product:UV VIS ChemStation  G1115AA  A.08.01

Keyword: Method

One-line Description:

Blank has to be measured twice after switching to a method with only one lamp.

Problem:

When the user switches from a method with both lamps on to a method with only one lamp on, a new blank spectrum has to measured. After the
first measurement the Last Blank Spectrum window is still empty. The blank has to be measured again, before a sample or standard spectrum can be acquired.

Temporary Solution:

Measure the blank twice after changing from a method using two lamps
to a method using only one lamp.

Fix Information:

Fix under investigation.


KPR#:196  Product:UV VIS ChemStation  G1115AA  A.07.01

Keyword: Method

One-line Description:

Wrong lamp status is displayed if the lamp time table is used.

Problem:

If the lamp time table is used to switch on the lamps automatically and the ChemSation is not closed, the lamp status is wrongly displayed on the graphical user interface.

Temporary Solution:

Close the ChemStation, if you are using the lamp time table to switch on the the lamps. Then the  lamp status will be displayed correctly after the
ChemStation has been started.

Fix Information:

Fix under investigation.


KPR#:142  Product:UV VIS ChemStation  G1115AA  A.06.03

Keyword: OQ/PV

One-line Description:

No acceptance criterion defined for OQ/PV of multicell transports

Problem:

The 'MCT Adjustment' report does not show a result for the 'Multicell Transport Performance Verification'. The acceptance criterion is defined in the OQ/PV manual G1115-90006 but not verified in the 'MCT Adjustment' report (File/Print Preview/Print MCT Adjustment Results).

Temporary Solution:

Following limits can be calculated for the 'Multicell Transport Performance Verification' from the specifications of the multicell transports and the acceptance criterion defined in the OQ/PV manual G1115-90006 (0.2 turns = 0.1 mm) :

8-position MCT G1120A:    Delta = +- 1.0
7-position MCT 89075C/D: Delta = +- 6.0 

'Delta' is shown in the last column of the MCT Adjust report (File/Print Preview/Print MCT Adjustment Results).

Fix Information:

Fix planned.


KPR#:165  Product:UV VIS ChemStation  G1115AA  A.09.01

Keyword: OQ/PV

One-line Description:

Error Message during Validation of 8452A

Problem:

A red-line error message Undefined Symbol ValidtestFilterUncert$
is displayed  during the 8452A OQ/PV. The error message doesn't
have an influence on the results nor the test procedure.

Temporary Solution:

Ignore message

Fix Information:

Fixed in revision A.10.01.


KPR#:221  Product:UV VIS ChemStation  G1115AA  A.02.04

Keyword: OQ/PV

One-line Description:

Intensity test in diagnosics indicates error

Problem:

If local language options are changed,such that decimal separator is "," instead of "." the intensity test reports an error.

Temporary Solution:

Set decimal separator to "." .

Fix Information:

Fix under investigation


KPR#:27  Product:UV VIS ChemStation  G1115AA  A.09.01

Keyword: Print

One-line Description:

Overlaid printed spectra are not visible

Problem:

If several overlaid spectra are printed, the spectra #9 and #10 are printed in yellow and light gray respectively and are not visible on the printed report. In the configuration editor only the colors
up to #8 can be defined. The occurs e.g. on Deskjet 840C, 940C, 990Cxi printers.

Temporary Solution:

n/a

Fix Information:

Fix planned.


KPR#:95  Product:UV VIS ChemStation  G1115AA  A.02.04

Keyword: Print

One-line Description:

Entered sample comments are not printed in results report

Problem:

Comments entered by the user for sample measurements are not printed with the results report.

Temporary Solution:

n/a

Fix Information:

Fix under investigation.


KPR#:114  Product:UV VIS ChemStation  G1115AA  A.05.03

Keyword: Print

One-line Description:

Set up wavelength analysis with list of wavelengths.  Overlay spectra on

Problem:

Set up wavelength analysis with list of wavelengths, the wavelengths are indicated by vertical lines on the screen. Vertical lines do not show up on print out and there is no way to turn this feature on and off.

Temporary Solution:

n/a

Fix Information:

Fix under investigation.


KPR#:123  Product:UV VIS ChemStation  G1115AA  A.02.05

Keyword: Print

One-line Description:

The print out of Lamp Counters are truncated

Problem:

When attempting to print out lamp counters using "Print Selected Window" in the Verification and Diagnostics mode, the results are truncated and incomplete.

Temporary Solution:

None.

Fix Information:

Fixed in Rev. A.09.01


KPR#:279  Product:UV VIS ChemStation  G1115AA  A.05.02

Keyword: Print

One-line Description:

Missing characters in report header.

Problem:

If duplex printing is selected, the report header is too short. Some characters of the report name are missing.

Temporary Solution:

Switch printer to none duplex printing.

Fix Information:

No fix planned.


KPR#:598  Product:UV VIS ChemStation  G1115AA  B.01.01

Keyword: Print

One-line Description:

Printing overlaid spectra in color does not work

Problem:

Even if a color printer is attached and configured properly, overlaid spectra are printed in black and white only

Temporary Solution:

None

Fix Information:

Fix under investigation.


KPR#:285  Product:UV VIS ChemStation  G1115AA  A.05.02

Keyword: Programming

One-line Description:

The loading of a macro fails.

Problem:

In some cases the HP ChemStation does not load a macro at first attempt.

Temporary Solution:

Retry to load the macro several times.

Fix Information:

No fix planned.


KPR#:754  Product:UV VIS ChemStation  G1115AA  B.02.01

Keyword: Reports

One-line Description:

Method information missing in various different reports

Problem:

Numerical information is missing in various reports. The labels are printed without the respective value.

Temporary Solution:

Not available

Fix Information:

The problem was the missing check for an empty format string. Fixed in code.