                 HP 4500 ChemStation Software
                   Software Status on Rev.B.01.01
                                         (supplemental disk Rev.B.01.03)
*************************************************************************

This file contains information about installation, new features in this
version of system disk: Rev.B.01.01 and current software status of 
this version of supplemental disk: Rev.B.01.03.
After installation of the supplemental disk Rev.B.01.03, we recommend 
reading this file before using the HP4500.

[Information about installation]

1. About installation of the supplemental disk
  The supplemental disk can be installed automatically after installation
  of the system disk for Rev.B.01.01. You can also install it separately as 
  follows:

     1-1. Insert the disk into drive A.
     1-2. Click the [Start] button on your Windows task bar.
          and select [Settings] - [Control Panel].
     1-3. Double click the [Add/Remove Programs] icon in the Control Panel.
     1-4. Click the [Install] button on the [Add/Remove Program Properties] window.
     1-5. Click the [Next] button, the installer "a:\setup.exe" will be detected.
     1-6. Click the [Finish] button to launch the installer.

2. For the Vectra XM Series 4 5/166, we recommend using the display driver
   S3(256 colors) or S3 Trio 64 PnP PCI (256 colors).

3. About ASX500
  If the ROM version of the ASX500 software is ver.1.4 or earlier, the
  ASX500 may hang up when you set 0 sec for the "Optional Rinse". This is set 
  in the dialog box shown by selecting the "PeriPump"-"Set PeriPump Program" 
  or when you stop a sequence that has already started.
    3-1. Select "ASX500" in the autosampler combobox on Configuration Task.
    3-2. Click "Cmdline on" in the system menu of Acquisition Task.
    3-3. Enter "ALSCOMMAND "VER"(ENTER-KEY)" in the command line.

4. For HP Vectra XM series 4 and XA 5/166 Users
  HP-IB select code must be set to 6 if your computer is Vectra XM series 4 
  and Vectra XA 5/166. This is set in the configuration task. The default 
  value is usually 7, but we set it to 6 when shipping XM series 4
  and XA 5/166.

5. You cannot use long file names(more than 8 characters) for files and 
   directories that the ChemStation refers to.

6. Optional Software
  Following optional software products can be installed and operated on the ChemStation 
  system disk Rev.B.01.01.
          EnviroQuant Software Rev.B.01.00
          ETV Operating Software Rev.B.01.00
          Laser Operating Software Rev.B.01.00
          FI Operating Software Rev.A.01.00
  Note for the FI Operating Software Rev.A.01.00:
  The FI Operating Software Rev.A.01.00 overwrites some parts of the ICP-MS 
  ChemStation software. Please install the supplemental disk Rev.B.01.03 for 
  the ICP-MS ChemStation Software again. It is necessary to keep your 
  ChemStation updated to the latest version.


[New features in this version]
1. Year 2000 Compliance
  This version of the ChemStation Software (combines the system disk  
  Rev.B.01.01 and the supplemental disk Rev.B.01.02 or later) is Year 2000 Compliant. 
  Accessary software for Windows 95 is also Year 2000 Compliant when 
  they are installed on the ChemStation Rev.B.01.01 or later.

2. Bug fix for the default data path for TOP Task
  There was a problem that when the latest data path opened on offline data 
  analysis, it set as the initial data path for TOP task. Consequently,
  the path is displayed as the default data path in Start Sequence 
  dialog box. The problem was fixed in this version.

3. Bug fix for the output value of Count in reports
  CPS was output to Quantitative Reports, Custom Reports and QC Reports, instead of 
  Counts. The problem occurred when Data Analysis did not quantitate some 
  elements; for example, the measured masses are set as "Excluded" in the 
  calibration table of the used method. The bug was fixed in this version.

4. Bug fix for the sensitivity auto tuning.    
  It is fixed that auto tuning macro stops with the error "Expected a 
  variable or function name", when sensitivity auto tuning was executed with
  the option, "Reduce Doubly Charged Ions", "Reduce Oxide Ions" 
  or "Reduce Background".   

5. Bug fix for calibration table.
  It is fixed for mass numbers that sometimes displayed incorrectly in 
  a calibration table, when they are deleted and added again.

 Feature 2 to 5 requires the combination of the system disk Rev.B.01.01 
 and the supplemental disk Rev.B.01.03 or later.
   
[Current software status]
1. Possible error message displayed when ChemStation is started.  
   When you start the top level of ChemStation, a message that says 
  "APG_TOP is already running" is sometimes displayed.
  In this case, please close Windows and reboot your computer.

2. Displaying the Notepad window
   If you select Screen as a reports destination, the Notepad window will      
   sometimes be hidden behind the ChemStation window. This occurs when you      
   generate reports using the "DoList" function. If this happens, click on 
   the Notepad icon displayed on the Windows task bar, and the Notepad 
   window will be brought to the front.

3. About deleting the current method  
   Please do not delete the method file that is currently loaded on 
   ICP-MS Top or ICP-MS Data Analysis. If you do, the drive in which 
   the method file is located will be disabled in the Selecting Method 
   dialog box. 
   To solve this problem, please close Windows and reboot your computer.    
   
4. About Autotune
  When you would like to abort "Set EM" or "Autotune", please click "Abort" on 
  the MENU BAR. DO NOT CLICK "Abort" on the SYSTEM MENU. If you click "Abort" on 
  the system menu, the program may not finish normally. In addition, the tuning 
  file name will be changed to "temp.u$" while the "Set EM" program is running. 
  The file name will change to "atune.u" once you close the tuning task and 
  then open it.

5. About 'Edit Average File'
  If you start Edit Average File when another instance of 
  Edit Average File is already running, the data file is loaded 
  not on the second instance of Edit Average File but on the first one.
  Please operate as following, when you want to open two or more
  data files on Edit Average File:

   (1) Load the first data file by Data Analysis task.
   (2) Start [Edit Average File] from the menu command
      [Files]  >>  [Edit Average File...]
   (3) Load the second data file, using the menu command
      [ICP-MS Data] that is displayed in the menu bar
      of Edit Average File task. 