Software Status Bulletin
MassHunter Quantitative Analysis GQuantAA
Known Problem Report as of Mar 14 2017 10:27AM
Preface
-------------------------
This Software Status Bulletin (SSB) documents all known problems in the software
product designated above. The SSB is derived from Known Problem Reports (KPR)
which result from user problems that have been classified as documentation
problems or software defects. When a KPR is written, an identifying number is
assigned to it, and the KPR is added to the next edition of the SSB.
User inputs that have been classified as Enhancement Requests are not documented
in the SSB. User problems that have been submitted, but that have not been
classified by the time the SSB is generated are not included in the SSB.
How to use the SSB
-------------------------
When you experience a problem with a product, first check this SSB to see if the
problem has been reported already, and if there is a temporary workaround
available for the problem, or if the problem has already been fixed by a new
revision. If the problem is not listed in this SSB then you may wish to report
it to the Response Center or to your field support representative.
To determine if your problem is documented in this SSB, first look in the
Keyword Index section of the SSB. Under each keyword is a listing of one-line
descriptions of related KPRs. If any of these sound like yours, locate the KPR
# in the Known Problem Reports section of the SSB, and read the full KPR. The
KPRs in the Known Problem Reports section are sorted by KPR #.
There are two sections in the SSB:
Keyword Index: This index is categorized by keyword. For each KPR there is a
brief description and a KPR #. A KPR may be associated with more than one
keyword.
Known Problem Reports: This section contains KPRs, with all the available
information relevant to the problem. KPRs in this section are sorted by KPR #.
Keyword Glossary
Keyword Index
Keyword:
One-line Description:
Extra levels added when reloading a method after an existing batch update
Problem:
Going from batch view and then going to open method from existing batch you will get the QC's added as Calibration levels in the table and they are not identified as QCs. 1. Created a new batch 2. Created the method by using acquired MRM data 3. Manually entered the concentration and dilution pattern 4. Now we have 5 calibration levels for every compound (5 lines in concentration setup) 5. He saved the method 6. He applied this method to other batch, then the calibration level has 7 lines (5 Cals + 2 QCs). this is referred to as the extra lines in calibration levels.
Temporary Solution:
Quant has been this way always, it is not a bug. The QC are considered to be Quality Control levels and are added to the calibration.
Fix Information:
n/a
Keyword:
One-line Description:
MolecularFormula visibility in Library Editor in Quant
Problem:
MolecularFormula column in dataset is missing from the database option.
Temporary Solution:
n/a
Fix Information:
N/A
Keyword:
One-line Description:
Changing RT (RI) parameter in Library Search, it has no effect on search results
Problem:
the search results should change according to the RT or RI window in Library search. But it has no effect at all: 1. use the new batch file PesticidesCompounds in my D:\Data\Data_FromHarry\QuantResults folder 2. Create New Method using Scan data with Library Search 3. in Library Search Setting, change the RT window to 2min and 100 min 4. Search the library, but no p,p-DDD compound returned in the results.
Temporary Solution:
The issue boils down to using the wrong RT window parameter in the Advanced Dialog. The TargetRT match tolerance window on the Compound Identification tab is only used during UA to match unknown compound candidates with target compounds. This setting does NOT affect the library match score. In order to make the RT matching more tolerant during library search is to relax the RT range in the Library Search tab. When this is increased to 100 seconds, the desired effect is achieved.
Fix Information:
n/a
Keyword:
One-line Description:
The chromatogram image in the report should be the same as "Batch as A Glance"
Problem:
Customers want to get the same chromatogram (scale) image in the report as in the Batch at a Glance screen. If customers change the scale of the chromatogram the updated chromatogram in scale should appear in the report.
Temporary Solution:
The issue is that the chromatogram used in the report is not updated when changing the scale of the chromatogram. But most cases, the customers are interested in the scale of the lower limit. If the chromatogram that is fit y axis to lower limit of calibration is used in the report, it will satisfy the requirement, then we can close this request.
Fix Information:
n/a
Keyword:
One-line Description:
When creating New Method using Manual setup, 2D signal type has only MS
Problem:
When creating New Method using Manual setup, 2D signal type has only MS to choose, no other channel like DAD to choose from. Tested both MRM and GCMS flavor, all are the same.
Temporary Solution:
When doing a manual method setup, Quant does not generally have access to a sample. Therefore, there is no way to get a list of available signals from the sample, so the list is populated with the MS as the only default signal. However, this is not an issue, since the user can type in any signal type value like GC, and the application will take it and interpret it correctly when the method is applied to a batch (assuming the signal name is present in the sample).
Fix Information:
Resolved in Build B.02.01
Keyword:
One-line Description:
Copy to Clipboard of Spectra in Qual (318) causes exception in Lib. Ed.3.1.166.0
Problem:
Copy to Clipboard of Spectra in Qual (318) causes exception in Lib. Ed.3.1.166.0
Temporary Solution:
None
Fix Information:
Fixed in version B.03.02
Keyword:
One-line Description:
Cannot load new data files into new batch if data files were used in previous batch
Problem:
creating a new batch in Quant, adding new samples to this batch, but the data file has the same name as the previous batch you will get an error message that file name already exist and brings you back to the original directory path.
Temporary Solution:
Files are being processed as a batch. Batch means all of the data files will exist under the existing directory so the file names need to be unique.
Fix Information:
n/a
Keyword:
One-line Description:
Save as *.library.xml to the *.library format could abort the library editor
Problem:
“Save as” library.xml format to library binary file format could cause the library editor to abort.
Temporary Solution:
n/a
Fix Information:
Fixed in B.03.02
Keyword:
One-line Description:
SIM data was labeled as SCAN after translation using the Translator
Problem:
AIA format cannot distinguish between SIM and full-scan data. So, there is no short way to determine the data in the AIA file to whether it is full-scan data or SIM data. So the software will continue to default the output file data type as Scan data.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
RT is "0.00" for every compound when creating new method in Quant
Problem:
When creating a quantitative analysis method using the Method-->New-->New Method from Acquired MRM Data, one or more Time Segments may have compounds that do not have Retention Times (i.e. 0.00). All other method parameters (e.g. Compound Name, Scan Type, Transition, etc.) will be listed but not the Retention Time. Other Time Segments may have all parameters including the Retention Time listed.
Temporary Solution:
The problem is caused by there not being at least 64 data points in the chromatograms extracted for the transition in those Time Segments. The default integrator for quantitative analysis method setup, the MS-MS Parameterless Integrator, requires at least 64 data points in the chromatogram in order to integrate a peak. Since no peak is found, no Retention Time is recorded for that compound. This problem has also been seen in MassHunter Quantitative Analysis rev. B.01.04 and the workaround described in the Solution section is the same. Solution/Action: The workarounds for this problem involve editing a file that sets quantitative analysis method setup and user interface defaults. To edit this file: 1) Close MassHunter Quantitative Analysis. 2) Make a copy of the file QuantAnalysis.exe.config in the C:\Program Files\Agilent\MassHunter\Workstation\Quant\bin folder. 3) Open C:\Program Files\Agilent\MassHunter\Workstation\Quant\bin\QuantAnalysis.exe.config using the Notepad editor or an editor specific for XML files. Search for the lineChange the "AutoMagic" to "RTE" i.e. The General (RTE) Integrator will now be the default integrator for creating the quantitative method using the Method-->New-->New Method from Acquired MRM Data menu item; it does not require 64 data points to integrate a peak. However, a superior solution from a data quality standpoint is to examine and adjust the Acquisition conditions. Often, data files that have exhibited this problem have Time Segments that are too short with too many MRM transitions. Due to the short Time Segments, some compound peaks are on the boundary and cut off, such that no integrator could successful integrate them. 4) Save and Close the file QuantAnalysis.exe.config 5) Restart MassHunter Quantitative Analysis.
Fix Information:
n/a
Keyword:
One-line Description:
Manually integrating a qualifier does not update quantifier area with Area Sum
Problem:
If the user creates a Quantitative Analysis method for a compound with Area Sum turned on for its qualifiers, when a qualifier is manually integrated it does not update the quantifier area. The correct behavior would be to update the quantifier area so that it reflects the new area of the qualifier.
Temporary Solution:
n/a
Fix Information:
For this localization issue, since we can't change any UI behavior, so the fix is in the next version of Quant (FAST). Fixed in version B.04.00
Keyword:
One-line Description:
Timed events Area Sum On/Off in Genie Integrator fixed
Problem:
Timed events Area Sum On/Off in Genie Integrator was not working in B.03.01 unless patched
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Data is acquired for the same precursor and product ions (transition) but at different collision energies (CID).
Problem:
A Quant method is set up which describes each transition with the associated CID. When doing analysis Quant B.03.02 and earlier do NOT reference the CID during extraction and integration and thus reports each transition using the sum of all matching transitions.
Temporary Solution:
None
Fix Information:
Quant 04.00 has an option to extract and integrate based on CID. User must enter a value in the new Collision Energy Delta field to activate the new feature. Fixed in version B.04.00
Keyword:
One-line Description:
Instrument Name and Injection Volume now in Quant
Problem:
Instrument Name and Injection Volume were not appearing in Quant in B.3.01
Temporary Solution:
none
Fix Information:
na
Keyword:
One-line Description:
Displaying UV 2-D data in Quant requires the UV signal to have a "Time Segment"
Problem:
When creating a 2-D UV or ADC compound, the chromatogram will not show up in the Compound Information pane until you enter something in the TS column of the Quantifier field for the new UV compound. Until you enter something arbitrary in the TS field, the Compound Information pane will say "No Data Points".
Temporary Solution:
Enter a value of 1 in the TS
Fix Information:
n/a
Keyword:
One-line Description:
Quant shows TIC and 2D chromatogram in same windows
Problem:
for 2D GC data, Quant should separate TIC and 2D chromatogram in 2 windows, currently they overlay in one window, user has to normalize it in order to view 2 signals together.
Temporary Solution:
none
Fix Information:
n/a
Keyword:
One-line Description:
Library editor crashes when undo paste function to existing library
Problem:
Library editor crashes when undo pasted library entries into an existing library in
Temporary Solution:
none
Fix Information:
Fixed in B.03.02
Keyword:
One-line Description:
2D GC compound displayed poorly in Compound Information window
Problem:
1) RT in chromatogram window is not limited to RT window but shows entire run time 2) data shows up in the qualifier window (no qualifiers) 2) Spectra window shows some sort of data (no mass data)
Temporary Solution:
none
Fix Information:
n/a
Keyword:
One-line Description:
Add units to peak labels for Calc. Conc.
Problem:
Although one can asks to see peak labels for Calc Conc. in the Compound Information window you do not have the option to display its units. Would like to see this in the form 1.234 ug/ml where 1.234 is the Calc. Conc. and not just add another column to the stack "Units = ug/ml"
Temporary Solution:
none
Fix Information:
Fixed in B.03.02
Keyword:
One-line Description:
.NET Framework message appear even though it's already installed.
Problem:
1. Install B317 Quant(released) to JPN Windows Vista SP1. 2. .NET Framework 2.0 (or higher) is already installed (Qual is already installed), but the message appears.
Temporary Solution:
none
Fix Information:
fixed in B.03.02
Keyword:
One-line Description:
GCMS translator may have problem when translating the SIM/Scan Data
Problem:
Sine LCMS QQQ format cannot support multiple MS signals in the same data file with DIFFERENT number of scan records. The GCMS file format does support same data files with different number of scan record and as a result the translator will fail in this situation by design.
Temporary Solution:
None
Fix Information:
Fixed in B.03.02
Keyword:
One-line Description:
4 shortcuts are created when only "GCQQQ" is selected on Installation
Problem:
1. Install Build 170 Quant(released) to JPN Windows Vista SP1 2. Select only "GCQQQ" 3. 4 shortcuts are created. they are named "ShurtcutX". in English shortcuts are removed after install
Temporary Solution:
none
Fix Information:
Fixed B.03.02
Keyword:
One-line Description:
Calibration curve window, the peak being viewed should show conc/response
Problem:
In the Calibration curve window, the peak that is being viewed should have its relative concentration and relative response shown and the intersection with the calibration curve shown. I have tried to draw in the way that it should look. when the user scrolls up and down of the sample, it would be nice to have it reflected in the calibration curve so that the user can see where does the sample reside in comparison to the levels which were taken. This has been entered into the LCMS Quantitation Database earlier and users have been requesting this quick visualization tool.
Temporary Solution:
none
Fix Information:
Fixed in B.03.02
Keyword:
One-line Description:
Quant will not load *.xml library while it is selected
Problem:
User just created an xml library, no problem. Go into Quant and the software sees the file, no problem User then tries to load the file and it will not accept the file name that was selected, it will stay open. 1 - problem is two parts, the user has a new PC and the file extensions state hide known types. The system knows the file is an xml format because it will be displayed but the error is the open command, it will not open the file because the file program is looking for the full string to be passed and it was not. 2 - If hide file extensions is declared to have hide known types then the user is forced to type in ".XML" to make the library work.
Temporary Solution:
Configure Windows to show file extensions
Fix Information:
Fixed in B.03.02
Keyword:
One-line Description:
Unable to export more than 256 Batch Table columns to MS Excel.
Problem:
Cannot export to Excel when there are more than 256 columns in the batch table. When you try to export to an .xls file it tells you "Excel worksheet is limited to 256 columns, and the batch table exceeds this limit.
Temporary Solution:
None
Fix Information:
Fixed in B.04.00
Keyword:
One-line Description:
Time Reference Flag has no effect when checked in Quant
Problem:
When Time Reference Flag is checked for an ISTD compound (eg. AMP-D5 ), According to Time Reference Flag definition, it would adjust the RT of AMP but doesn’t.
Temporary Solution:
Time Reference should not cause any significant shift in the reported retention times, and certainly not in the RTs specified in the method. As the Quant documentation states, Time Reference simply helps the Quantitation Engine find and identify peak groups which would otherwise not be detected due to a large shift in RTs. The correct way to test the Time Reference feature is to use it on a data set where it actually makes a difference, namely on a sample where all RTs (for target compounds and ISTDs) have shifted relative to the values specified in the method.
Fix Information:
Fixed in B.04.00
Keyword:
One-line Description:
Manual integration of ISTD disappears after manually integrating Target peak
Problem:
Seen in B.03.01 Build 170.0 and 170.18. The steps to reproduce are: 1) Manually integrate the sample MN2 and compound NAD's target chromatogram. 2) Manually integrate its ISTD PIPES. 3) Go back and manually integrate the target chromatogram. The ISTD chromatogram's manual integration disappears. In addition, you can no longer make any changes to the manual integration of the target chromatogram. If you try, nothing changes.
Temporary Solution:
none
Fix Information:
Fixed in B.03.02
Keyword:
One-line Description:
"Manual integration failed" error while manually integrating in Quant.
Problem:
Found in Q-TOF data Quant batches with 150 data files and 50-100 compounds. They inspect every chromatogram and manually integrate those that they judge need to be better integrated. About every hour they have been seeing a message: Manual integration failed Failed to get compound chromatogram for OpenDataFile is not called prior to calling this method OpenDataFile is not called prior to calling this method after attempting to manually integrate a peak. Then they are unable to save their batch unless they close Quant and reopen Quant. As a result, they lose their results from the last save.
Temporary Solution:
none
Fix Information:
Fixed in B.03.02
Keyword:
One-line Description:
UserDefined1 through 9 are not transferred from worklist to Quant.
Problem:
A customer wanted to use the UserDefined1 through 9 parameters that they saw in the B.03.01 Quantitation Dataset. However, they were unable to get any of these to show up in their report. I reproduced this and found that I could transfer UserDefined from a worklist but not UserDefined1 through 9.
Temporary Solution:
none
Fix Information:
Fixed in B.03.02
Keyword:
One-line Description:
Qualifier Ratio is 0 when creating New method from Acquired MRM data
Problem:
Qualifier Ratio is not zero in the Qualifier Graph, but in method, it is set 0 when creating new method from Acquired MRM data.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
scan type for TOF data is MS2Scan
Problem:
While creating a method using TOF data. The scan type ("Scan" column) is listed as "MS2Scan".
This is TOF data, there is no MS2Scan.
Temporary Solution:
n/a
Fix Information:
Fixed in B.04.00
Keyword:
One-line Description:
Index Converter tries to convert TOF or Q-TOF data, takes a very long time.
Problem:
Seen on Q-TOF LC/MSs. When they try to create a batch, it starts converting the Q-TOF data and takes a very long time. It does this every time they create a batch with the data.
Temporary Solution:
None
Fix Information:
Fixed in B.04.00
Keyword:
One-line Description:
No cal point is displayed for zero concentration level with no response.
Problem:
A user would like to use a zero concentration standard in the Cal Curve. If one puts it in and the standard (actually a blank) has no response, it does not display a cal point in the Calibration Curve of Quantitative Analysis.
Temporary Solution:
Can manually integrate a peak (or a background peak), then use Zero Peak. Then a Level 0 Cal point is displayed on the curve. The Blank Offset calibration curve fit uses the Blank Offset to calculate an offset. That is often an acceptable answer to those who are looking to push the lower limit of the quantitation. But, using 'Blank offset' for the calibration curve Origin does not display a zero concentration point on the curve.
Fix Information:
n/a
Keyword:
One-line Description:
Need a way of thresholding TOF and Q-TOF data when creating Quant method
Problem:
Right now, when you attempt to create a TOF or Q-TOF method from Acquired Scan Data, it will "deconvolute' the data. The data is extremely rich and it can take a very long time to deconvolute it. I have even rebooted my PC after starting because it took so long it appeared to be locked up. It would be helpful to have a way to "threshold" this so it only looks for the 20 top peaks or peaks above 500 abundance values. This way the search for compounds could go much faster.
Temporary Solution:
workaround is the following procedure: 1) Open the scan data batch in Quant 2) Go to the Method Editor 3) Right-click anywhere in the Sample information window and select Perceive Compound Settings. You will get the Parameters dialog. 4) Set the Spectrum peak threshold to a high-enough value to reduce the number of peaks. Click OK to apply and cache the value. 5) Run "New Method from Acquired Scan Data". The new threshold value will be used.
Fix Information:
Fixed in B.04.00
Keyword:
One-line Description:
Extraction window by Retention Time Flag
Problem:
Version B.03.02 and below of the software uses the retention time and left and right retention times deltas as specified in the method to extract data for peak detection rather than the adjusted retention time, as currently calculated and then applied the left and right deltas from the method. This can result in a peak outside of the method RT window but in the adjusted RT window not being found.
Temporary Solution:
none
Fix Information:
B.04.00 uses the adjusted retention time window to extract signal and detect peaks. Fixed in version B.04.00
Keyword:
One-line Description:
MRM transitions with same precursor and product ions signed MS1SIM and 0 retention time
Problem:
While running "Create MRM method from MRM data' the MRM transitions, which have same precursor and same product ion, are signed as MS1 SIM. They are coming with zero retention time. You have to change the Scan type to MRM and manually enter right ret times and then peaks are integrated. data file added
Temporary Solution:
None
Fix Information:
Fixed in B.03.01 (except RT = 0 when too few data points to detect peak)
Keyword:
One-line Description:
MassHunter Qual using the shortcut from Quant MassHunter encounters an error
Problem:
When triggering Tools/Go to Qualitative Analysis... menu item, the user see an error message, but the user can continue
Temporary Solution:
Qual opens the data file, but does not correctly extract the ions (transitions). Use Qual to manually extract the appropriate ions for that compound.
Fix Information:
Fixed in B.03.02
Keyword:
One-line Description:
Performance issue for library editor when importing large JCAMP file
Problem:
B.03.01 Library editor was very slow importing JCAMP files.
Temporary Solution:
None
Fix Information:
fixed in B.03.02
Keyword:
One-line Description:
Missing information in MS-MS copy for spectral library entry
Problem:
In B.03.01, a copy/paste into the Library Editor does not transfer the InstrumentType and acquisition mode, i.e. MS/MS and precursor ion does not get entered automatically
Temporary Solution:
None
Fix Information:
Fixed B.03.02
Keyword:
One-line Description:
Arrow points to Cal levels in Calibration Curve when on Calibration
Problem:
Feature was present in version before B.03.01 release where it was lost.
Temporary Solution:
none
Fix Information:
Fixed in B.03.02
Keyword:
One-line Description:
Error when using Timed events in Universal (Genie) Integrator
Problem:
B.03.01 version - When setting up or changing existing Timed Events (with the Universal/Genie integrator), clicking 'Apply to All' displays an error message. Quant could hang after the message is displayed.
Temporary Solution:
none
Fix Information:
Fixed in B.03.02
Keyword:
One-line Description:
Accurate mass spectra from TOF rounded when copied to Library Editor
Problem:
When accurate mass spectra are copied from Qual or Quant into the Library editor the masses are rounded to a tenth AMU.
Temporary Solution:
none
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Library Editor does not read Retention Index data when importing JCAMP files
Problem:
Library Editor does not read Retention Index data when importing JCAMP files
Temporary Solution:
none
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Curve fit type, origin, weight for ISTDs are modified when saving batch
Problem:
Seen in version B.03.02 Curve fit type, origin, weight for ISTDs are modified when saving batch. Steps to reproduce: 1. Open Quant batch 2. Apply a method. Make sure Curve fit, origin, weight are 'NULL' for ISTD compounds 3. Analyze batch and Save batch (not method) 4. Close batch and open this batch again 5. Curve fit, origin, weight for ISTDs are filled by default values while those value should be NULL.
Temporary Solution:
None
Fix Information:
Issue does not impact results as ISTD do not have calibration curves. Fixed in version B.04.00
Keyword:
One-line Description:
Injection Volume is blank
Problem:
The Injection Volume field can be blank in Quant and thus Quant reports when a valid Injection Volume is in the sample_info.xml field. Seen in B.03.01
Temporary Solution:
None
Fix Information:
Fixed in B.03.02
Keyword:
One-line Description:
Creating Method from MRM data fails when multiple acquisition methods are found
Problem:
When creating a Quant method from MRM data using the “New method from Acquired MRM Data” feature Quant looks in the AcqData/*.m directory of the selected data file and take the first file that fits the pattern *qqqacqmethod*.xml as the acquisition method file. It then uses this file to determine which transitions to load into the method not the data in the data file selected by the user. When multiple files fit the pattern, as would potentially occur when the acquisition method is edited and saved under a slightly different name, the possibility exists that the software will select an acquisition method file other than the one used to collect the data. If this happens the method may not reflect the transitions in the data file.
Temporary Solution:
Delete or rename (to avoid the *qqqacqmethod*.xml format) all undesired method files that are not the desired one.
Fix Information:
Fixed in B.03.02
Keyword:
One-line Description:
Curve Fit assistant give no calibration curves (rows)
Problem:
When asked to display the Curve Fit assistant on some data sets the software displays no suggested calibration curves (rows) in the Curve Fit table. Problem seen in B.03.02
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Universal integration events are altered unexpectedly
Problem:
The “Use same parameters as the parent compound” in the Universal integration dialog box associated with Qualifiers does not properly copy or track the parent compounds events to the Qualifiers. Problem could be seen after editing the method, applying to batch and returning to the method to find expected changes to events. Problem found in version B.03.01 and B.03.02.
Temporary Solution:
None
Fix Information:
Resolved in B.04.00 Fixed in version B.04.00
Keyword:
One-line Description:
Bad Integration crashes Quant during reporting
Problem:
At times the MS-MS or MS-MS(GC) integrators will integrate a peak erroneously such that the peak variables have undefined or illogical values. An example of this is when the integrated area appears to be completely above the peak and stretch upward forever. When this happens to any peak in a graphics window the program encounters an error and crashes producing no report. Usually it is not the peak that corresponds to the compound of interest but a poorly defined peak near the edge of the graphical window. Problem seen in B.03.01 and B.03.02.
Temporary Solution:
Review the data looking for bad peak integrations and manually integrate the peak. Then generate the desired reports.
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Manual Integration fails under Universal integrator
Problem:
When attempting to manually integrate a peak under the Universal integrator the integration fails and gives the error dialog that reads “ Manual Integration Failed The method or operation is not implemented.” Seen in B.03.02.
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Manual Integration of Batch with Disabled Cal Curve Points Changes Calc. Conc.
Problem:
In Quantitative Analysis rev. B.03.02 Build 170.21, manual integration of a peak in a batch that has a Cal Curve with disabled points can result in an incorrect Calculated Concentration. The problem does not occur with every batch. The problem also does not occur with the older rev. B.03.01 Build 170.0 software. To reproduce the problem: 1) Load or create a batch. 2) Disable one or more points in the Calibration Curve. 3) Manually integrate a chromatogram peak. The change in value of the Calculated Concentration will be different than it should be if the points were not disabled. A fix is currently under investigation.
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Unable to Change Mass Extraction Window for TOF/Q-TOF Scan Data
Problem:
With Quantitative Analysis rev. B.03.02 Build 170.21, it is not possible to change the extraction window for TOF or Q-TOF Scan data. For example, if the user were to load a batch with TOF Scan data, go to edit the method (Method--Edit Method menu item or F10 button), then make a change to the Extract Left m/z or Extract Right m/z in Mass Extraction Setup, it would not be reflected in the method. The mass extraction will remain as the default value 20 ppm. The problem does not occur in version B.01.04 Build 126.0 nor B.03.01 Build 170.0 of Quantitative Analysis. The problem also does not occur with other types of data (i.e. LC/QQQ, GC/QQQ, translated single quad data).
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Qualifier Ion Area Sum function does not work in rev. B.03.02
Problem:
This issue involves the Area Sum function for quantitating with the addition of the qualifier ions. This function does not work at all in the shipping Quantitative Analysis, rev. B.03.02 Build 170.21.
Temporary Solution:
Install point patch. The implementation in this patch is different than previous MassHunter implementations. The previous implementation involved summing the chromatograms of quantifier and all qualifiers, then integrating and using the area. The implementation in the patch and future MassHunter Quantitative Analysis versions involves integrating each individual quantifier and qualifiers, then summing the areas. This patch implementation matches the way the area sum function has worked in the MSD ChemStation G1034xx and G1701xx since its first release.
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Manual integration gives very large area change for small adjustments
Problem:
Seen in B.03.01 and B.03.02. When one attempts multiple manual integrations by making fairly small changes in the start and stop points such that the area appears visually to fluctuate by 50% to 100% results in much larger variance in the area up to 10 fold. The problem seems to happen when the start integration is placed near the start of the peak and the stop integration point is placed in a region from peak half height to and past the peak apex.
Temporary Solution:
Install Patch
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Quant error when trying to create new method with TOF data
Problem:
Issue reported in B.03.01. While creating a method in Quant from an acquired TOF data file Quant gives the following error: "Failed to create method from scan data Deconvolution failed: FindPeakMS: Missing required input parameter set of type PSetTofPeakFinder" The data was acquired on a 6230A instrument with AJS, FPS disabled.
Temporary Solution:
None
Fix Information:
Resolved in B.03.02
Keyword:
One-line Description:
DA method and DA method path are missing in Quant
Problem:
When a Quant method is saved and applied to a batch, the DA method and DA method Path columns are empty when viewed in Batch-at-a-Glance. Correct operation would be to display the method name and path where the method was previously saved. Reported in B.03.01.
Temporary Solution:
None
Fix Information:
Resolved in B.03.02
Keyword:
One-line Description:
Quant can crash when right-clicking the mouse to add columns in Batch-at-a-Glance
Problem:
Occasionally Quant will crash when the user attempts to add a column via the add columns feature. Problem seen in Quant B.03.01.
Temporary Solution:
None
Fix Information:
Resolved in B.03.02
Keyword:
One-line Description:
Scan spectrum shows small spurious peaks at each m/z in the sample file
Problem:
Problem can be seen when viewing a spectrum. If one zooms into a region of the spectrum where there appears to be no masses, they will observe spurious peaks at every m/z with a step of 0.1 AMU.
Temporary Solution:
None
Fix Information:
Resolved in B.03.02
Keyword:
One-line Description:
CC Relative Response Limit Low (High) table column names changed to CC Response Ratio Low(High)
Problem:
In outlier setup, limits for CC Response Ratio are "CC Relative Response Limit Low" and "CC Relative Response Limit High" are not accurate and have been changed to "CC Response Ratio Low " and "CC Response Ratio High".
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Manual Integration Sometimes Incorrect or Not Updated
Problem:
When manually integrating, the retention time is sometimes moved also indicating the area response is incorrect.
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Quant and Qual has different MRM extraction peak appearance for same data
Problem:
When Quant has UseIndexDataAccess active (the default) some data files can produce very jagged peak shapes that do not match the same peak when viewed in Qual. This problem was seen in B.03.02.
Temporary Solution:
Disable IndexDataAccess
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
After changing sample type from 'Cal' to 'CC', the sample type cannot be changed back to 'Cal'
Problem:
This defect is seen in B.03.02 when a Calibration sample is changed to Continuing Calibration and then the batch is reanalyzed. After this, the sample type cannot be set back to Calibration and successfully reanalyzed. If this is attempted an error message displays: “Batch analysis failed Sample not validated : Continuing Calibration level (some level name) does not match any existing calibration levels”.
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Unable to manually integrate quantifier or qualifier after turning on Area Sum
Problem:
After turning on Area Sum for qualifiers, if a qualifier is manually integrated the quantifier's area is not updated. Seen in B.03.01 and B.03.02.
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Outlier Custom Calculation doesn't work
Problem:
After installing .Net Framework 3.5 custom calculation feature does not work. The on the fly compiler fails because a new parameter required in .net 3.5 is not supplied.
Temporary Solution:
Uninstall .net 3.5 framework
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Left and right sides of peak give different areas
Problem:
When one manually integrates a symmetrical peak measuring the area on the right half (drop from apex) vs. the left half of the area, the ratio is approximately 2:1 when it should be about 1:1. The sum of the left and right areas are essentially equal to the peak when it is manually integrated as a whole.
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Blank sample name stops batch analysis
Problem:
If a sample without Sample Name information is added to a batch, the batch analysis fails since Sample Name is a required field and edits are not allowed.
Temporary Solution:
None
Fix Information:
Fixed in 4.00
Keyword:
One-line Description:
Sample type "ConCal" in worklist shows up as "Sample" in Quant
Problem:
The sample type "ConCal" in worklist should be "CC" in Quant but it shows up as "Sample". Seen in B.03.01 and B.03.02.
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
New Method from Acquired MRM Data loses some information from acquisition method
Problem:
The problem can be seen when running “New Method from Acquired MRM Data” in B.03.02 and there are mixed polarities. The method is created but not all transitions get the name assigned that has been used in the acquisition method. Instead they get names like “Compound_##” where ## is a number. Another symptom is for some qualifiers ions to become separated from the target compound and form new compounds
Temporary Solution:
none
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Quantitative Analysis crashes when adding sample to batch created in B.01.04
Problem:
If a batch, created in B.01.04, is opened in B.03.02 and a new sample is added to the batch Quant immediately crashes.
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Quant assigns wrong polarity to negative MRM data
Problem:
While running “Create new method from acquired MRM data “ in B.03.02, a limited number of cases have created methods where negative MRM data was assigned a positive polarity.
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Replicate calibration samples with disabled levels in method does not work
Problem:
Two sets of standards with replicate injections are in the same quant method. The Enable check box is used to disable some compounds in some levels (where the compound is not present in that standard) and enabled in some levels (where the compound is present in that standard). Quant can handle this case when no replicates are present but not when replicates are present. Method becomes corrupted during the Analyze Batch step.
Temporary Solution:
Split the method in two, one for each standard set.
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Data files convert each time that batch is loaded.
Problem:
Each time the batch is opened Quant will indicating it is converting the data files when in fact it has already done this in previous batch use. Seen in B.03.02 Build 170.21.
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
ISTD Conc. Ratio and ISTD Resp. Ratio set to zero when Analyzing a Batch.
Problem:
When the user loads a batch and makes either the ISTD Response Ratio or the ISTD Concentration Ratio visible, then presses Analyze Batch, these ratios are set to 0.0, regardless of what the true values should be. Seen in B.03.02.
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Manual integration of peak does not give same area as automatic integration
Problem:
The problem is seen B.03.02 and prior versions when a peak is found and integrated automatically and then a manual integration is made which very slightly changes the baseline. The change in area from the automated integration will change more than would be expected and may in fact change in the wrong direction (reported area increases when peak area is actually being slightly reduced). Attempting to recreate the automated peak area via manually integration by returning the baseline to the original position will not give the same area. The difference seen is area cannot be explained in terms of slight differences in cursor position.
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00 when using Agile integrator. Very small variance still seen in other integrators.
Keyword:
One-line Description:
MRM chromatographic peak with too short RT Delta shows as jagged
Problem:
In a MRM data file where the right RT Delta is too short to display the entire chromatographic peak the peak gives a jagged appearance. If the right RT Delta is increased the jagged appearance goes away.
Temporary Solution:
Disable index file use
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Universal Integrator Shows 10x More Area than Other Integrator in B.03.02
Problem:
MassHunter Quant B.03.02 and prior shows 10 times higher results for the areas in the Universal Integrator than with other integrators. When manual integration is enabled and a manual integration is performed the factor 10 times goes away.
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Copy Calibration Level does not copy response information
Problem:
In version B.03.02 and prior the copy calibration level function duplicates the level identifier and concentration but does not copy response information from the current batch.
Temporary Solution:
None.
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Japanese Localization: Export Batch Table to CSV file creates garbled file.
Problem:
Export of the batch table in CSV file creates garbled file. Problem not seen in other localizations.
Temporary Solution:
Export in .XLS format then use Excel to save file as CSV.
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Unable to Open Existing Batch File From Previous Versions
Problem:
Quant B.03.02 attempts to open a batch file created under a previous version and an error containing "[ERROR] AppCommand.BatchFile Agilent.MassSpectrometry.DataAnalysis.Quantitative.ApplicationCommandException: Failed to read binary batch file ---> Agilent.MassSpectrometry.DataAnalysis.Quantitative.ApplicationCommandException: Hashcode does not match value stored in batch" appears.
Temporary Solution:
None
Fix Information:
Fixed in version B.04.00
Keyword:
One-line Description:
Open Method from existing batch fails when batch missing sample files
Problem:
When attempting to open a method from an existing batch and all of the samples files associated with the existing batch are no longer in the batch folder then Quant will fail to load the method and report an error “Failed to import method from batch. Failed to load compounds from batch sample Data Access failed for sample path {file path and name} : File not found {file path and name}”. The error may also been seen if the only some of the sample files are missing and the software selects one of the missing files to open for information.
Temporary Solution:
Restore missing sample files.
Fix Information:
This defect is not planned to be fixed.
Keyword:
One-line Description:
Quant and Qual report different Acquisition times
Problem:
Quant and Qual are reporting different Acquisition times. Both Qual and Quant are consistent in terms of screen and reports within its own software but they differ between the packages by 1 or more minutes. Quant and Qual are reading different event times and thus reporting different values.
Temporary Solution:
None
Fix Information:
Future release of Qual with become consistent with Quant.
Keyword:
One-line Description:
MI is disabled in Universal Integrator
Problem:
When using Universal Integrator, MI is disabled and Quant gives error message saying "method or operation is not implemented" when manual integration is attempted.
Temporary Solution:
Disable Indexed Data Access. This is done by editing the “QuantAnalysis.exe.config” file in the in the \Program Files\Agilent\MassHunter\Workstation\Quant\bin directory and changingto . Note that this will greatly slow data processing in the application.
Fix Information:
Fixed in Quant B.04.00 SP1
Keyword:
One-line Description:
Isotopic dilution give incorrect calculated concentrations
Problem:
When Isotopic dilution is activated (Rx and Ry supplied) the calibration curve is corrected but the calculated concentrations appear to be off as indicated by significantly larger accuracy figures than is justified by the data. The problem appears to be the order in which the correction and quant are applied. The accuracy figures are changed to the correct values if, after you analyze the batch, you perform a manual integration (zero peak) and restore the original area with a “clear manual integration”. It appears that the quant is being calculated before the curve is corrected.
Temporary Solution:
None
Fix Information:
Fixed in B.04.00 SP1
Keyword:
One-line Description:
Area correction feature does not work
Problem:
When the area correction m/z and area correction factor are entered in the single quad quant method there is no effect on the target response even when a peak at the m/z is found. The area of the m/z peak multiplied by the area correction factor should have been subtracted from the target area thus reducing the target response.
Temporary Solution:
None.
Fix Information:
Fixed in B.04.00 SP1
Keyword:
One-line Description:
Create Quant Method from Acquired Scan Data for Scan causes error message
Problem:
When one attempts to create a new method from scan data Quant reports an error : Failed to create method from scan data Deconvolution failed : Index was outside the bounds of the array.
Temporary Solution:
Disable Indexed Data Access. This is done by editing the “QuantAnalysis.exe.config” file in the in the \Program Files\Agilent\MassHunter\Workstation\Quant\bin directory and changingto . Note that this will greatly slow data processing in the application.
Fix Information:
Fixed in B.04.00 SP1
Keyword:
One-line Description:
Corrupt MStree2.bin file is created if the file is being acquired and gives NO DATA POINTS message
Problem:
When a batch is opened or created, Quant scans the data files in the associated directory and forms the Mstree2.bin files if they are not already present. If the directory contains a data file into which data is being acquired, Quant attempts to create an MSTree2.bin file which is not correctly formatted. If the file with the corrupt MSTree2.bin file is later added to a batch, the chromatogram graphic is blank and labeled “***NO DATA POINTS***”. Deleting the corrupt MSTree2.bin file and reopening the batch corrects the problem. The core issue is that Quant does not check the status of the file to see if it is being collected and skip the conversion if it is.
Temporary Solution:
Delete the corrupt MSTree2.bin file then reopen the batch to correct the problem.
Fix Information:
Fixed in B.04.00 SP1
Keyword:
One-line Description:
Searching MPW2007.L library crashes Quant
Problem:
An attempt to search the MPW2007.L library will cause Quant to crash. This can be seen when a spectra is extracted using “Extract Spectrum” and the MPW2007.L library is searched. The MPW2007.L library contains only condensed spectra (no full).
Temporary Solution:
None
Fix Information:
Fixed in B.04.00 SP1
Keyword:
One-line Description:
New method from Acquired MRM data not working correctly when high number of compounds, transitions and time segments are acquired
Problem:
When attempting to create new method from MRM data, from the example data file, a large number of compounds are created with names like “Compound_##”. The problem disappears and all compounds are correctly created when the "UseIndexedDataAccess" value="true" is changed to false. An error like this was seen B.03.01 and fixed in B.03.02 but appears again in B.04.00 for some but not all files.
Temporary Solution:
The problem disappears and all compounds are correctly created when you disable Indexed Data Access. This is done by editing the “QuantAnalysis.exe.config” file in the in the \Program Files\Agilent\MassHunter\Workstation\Quant\bin directory and changingto . Note that this will greatly slow data processing in the application.
Fix Information:
Fixed in B.04.00 SP1
Keyword:
One-line Description:
Advanced reporting crashes when a sample with a blank Name field is selected
Problem:
When samples that have a blank name field are selected using the Advanced reporting option, Quant crashes.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
CurveFitStatus and CurveFitR2 are blank on samples where no peak found in target
Problem:
When no peak is found for the target compound on a given sample the CurveFitStatus and CurveFitR2 are left blank. The curve data should be present and the same on a given compound in all samples. This defect can leave some fields in reports blank if that sample did not find the target. Seen in B.03.02, B.04.00 and B.04.00 SP1
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Batch Table export to Excel gives corrupt data file when > 676 columns are displayed
Problem:
When exporting a table, if the number of columns exceed Excel 2007 column “XZ” the result file will be corrupted and after the recovery the Excel column from “ZA” to “ZZ” are without values, data restart correctly from Excel column “AAA”.
Temporary Solution:
Export in CSV format then open in Excel. The data will transfer although some the Excel label formatting information is lost.
Fix Information:
n/a
Keyword:
One-line Description:
Old calibration data on compounds with no peaks not cleared resulting in errors
Problem:
Create a new batch and add calibration standard and some samples. Copy a method from a previous batch or method which contains calibration curve information. Method validates without errors. Exit and apply. Analyze batch (no errors) Compounds where no peaks were found in the current calibration standard but were found in the old calibration retain the original calibration curve and give quantitation on samples. Analyze batch (fails with error) Edit method and validate ( errors appear – Calibration STD Path mismatch between target and ISTD)
Temporary Solution:
Clear the old file names from the old batch. Method should validate without error.
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Qualifier signal is not displayed with MS-MS integrator when normalization on
Problem:
"Not Found" is displayed in the qualifier signal window when using MS-MS integrator and the Normalize option in the Qualifiers section of the properties window is checked. Seen in B.04.00.
Temporary Solution:
Three workarounds: 1) Use the Agile integrator. 2) Turn off ‘normalization’ in the qualifier window. (Right-click in the Compound Information window, select ‘Properties’. Under Qualifiers, uncheck the box for ‘Normalize’.) 3) Manually integrate a peak in the target window.
Fix Information:
Moving forward, the Agile integrator is the integrator of choice. There are no plans to change the MS-MS integrator.
Keyword:
One-line Description:
Method information is lost when batch contain no samples
Problem:
Method information is lost when doing the following. 1) Create a new batch 2) Do NOT add any samples 3) Open a method from existing method or batch 4) Exit the method editor and apply 5) Now add samples 6) Go to method editor – no method info
Temporary Solution:
Add samples before adding method
Fix Information:
n/a
Keyword:
One-line Description:
Quant Console can't find a method file if the file is saved in a root directory
Problem:
If a method file is saved in the root of a directory, Quant is not able to find the file even if the correct location or path is specified in the sequence editor. No batch is created during the sequence run.
Temporary Solution:
Do not save method files that are to be used by Quant Console in a root directory.
Fix Information:
n/a
Keyword:
One-line Description:
New Method from Acquired SIM Data displays the incorrect scan type in the method
Problem:
The method displays Ms2SIM instead of Ms1SIM for scan type.
Temporary Solution:
In Method Edit, use the pulldown menu under "Scan" to change the type.
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Surrogate outlier does not change when changing Final conc via dilution , amt or total amt
Problem:
The surrogate outlier is based on Final Concentration. Global setup should have “Apply Multiplier to Surrogate” checked to see this issue. When the final concentration is changed through manual integration the surrogate indicates high or low recovery properly. But when the multiplier is changed via dilution, amount or total amount the final concentration will change but the surrogate outlier is not updated.
Temporary Solution:
Perform an Analyze Batch after changes to dilution, amount or total amount to update surrogate outlier status.
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
If Integration Parameters Modified (IPM) is done, the integrator cannot be changed for that compound in the method editor.
Problem:
If Integration Parameters Modified (IPM) is done, then the integrator for the compound cannot be changed in Method Edit even though it appears that it is being changed.
Temporary Solution:
The workaround is to Restore Integration Parameters first, then change the integrator for the compound in the method.
Fix Information:
n/a
Keyword:
One-line Description:
Sample Information Window does not affect reporting graphics
Problem:
Sample Information Window's properties don't affect reporting graphics
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Product Ion and Precursor Ion are switched when manually picking the transition from the spectrum in Precursor Scan data
Problem:
While in Method Edit
Temporary Solution:
Go to method editor on a file containing Precursor Scan data and extract a spectrum. In the extracted spectrum, select an ion and right click the mouse to select New Compound. A new compound entry is created but the product Ion contains the m/z value of the precursor ion and precursor ion contains the m/z value of the product ion.
Fix Information:
Resolved in B.05.00
Keyword:
One-line Description:
"Relative Response" determined by “create method from data file” can give wrong value
Problem:
When creating a new method from MRM data under certain circumstances the Relative Response in the method does not match the value shown in the compound information panel. The issue is caused by the largest peak in the qualifier EIC not being the same as the peak that is ultimately selected as the qualifier.
Temporary Solution:
Update qualifiers for all compounds after creating a new method from MRM data.
Fix Information:
n/a
Keyword:
One-line Description:
Quant TIC window display ***NO DATA POINTS*** for GC only signal
Problem:
When a file containing only GC data is displayed in the Sample Information window a trace of the GC data can be seen but the display of TIC data is annotated with ***NO DATA POINTS***.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Exporting nested table with no visible columns crashes
Problem:
In the nested Horizontal view if and there are no visible columns in a table, such as the Qualifier Method or ISTD Compound method, the program crashes.
Temporary Solution:
Have at least one visible column for each table displayed.
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Samples being acquired may be added to a batch
Problem:
If the user attempts to add a sample to a batch using the “Add Samples” feature and there is a file that data acquisition is collecting into that file appears in the list of files and can be added to the batch. Once added any operation in quant will result in incorrect operation with regard to the sample and possibly the batch as a whole (the sample might be a calibrator).
Temporary Solution:
Avoid adding samples that are being acquired by checking the status of the file in the data acquisition software.
Fix Information:
n/a
Keyword:
One-line Description:
Surrogate % Recovery is not displayed nor available for QC and Cal Types
Problem:
In Quant rev. B.03.02 and B.04.00, when the batch has a surrogate, the Surrogate % Recovery is not reported in the Batch Table if the Type is set to QC or Cal. It will be reported for Samples. This function worked for QC and Cal Types in B.01.04
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Creating Quant Batches with DAD UV Spectral Data Fails
Problem:
If user try to create a batch with DAD UV spectral data, it may constantly convert. The error message in the conversion log is: ERROR: Failed to parse spectrum file D:\MassHunter\Data\1290DADTest\test008.d\AcqData/DAD1.sd 04/05/2010 10:40:40.125 Stack trace: at Agilent.MassSpectrometry.DataAnalysis.Quantitative.SpectrumDirFileReader.Read() at Agilent.MassSpectrometry.DataAnalysis.Quantitative.SignalConverter.ConvertNonMSSpectrumData(DeviceInfo deviceInfo) at Agilent.MassSpectrometry.DataAnalysis.Quantitative.QuantDataAccess.ConvertToIndexedData(Boolean overwrite) at Agilent.MassSpectrometry.DataAnalysis.Quantitative.IndexedData.IndexedDataConverter.ConvertSample(String path) 04/05/2010 10:40:40.125 Inner Exception: Invalid number of values per data point: expected 2, got 1
Temporary Solution:
n/a
Fix Information:
Fixed in B.05.00.
Keyword:
One-line Description:
Deleting sample from Standard Addition Batch Table causes error message
Problem:
When a user deletes a sample injection from a batch that is set as Standard Addition and then presses Analyze Batch, an error message appears that says Batch Analysis failed: the value for column 'CompoundName' in table 'TargetCompound' is DBNull. Unable to cast object of the type 'System.DBNull' to type 'System.String'
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Standard Deviation Bars do not appear until after the Curve Fit Assistant is run
Problem:
The Standard Deviation bars in the Cal Curve will not appear until after running the Curve Fit Assistant.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Create Quant Method from Acquired Scan Data for TOF causes error message
Problem:
When the user tries to create a “New Method from Acquired Scan Data” with an LC/MS TOF data set, they will get an error “Deconvolution failed: Deconvolution requires sample data acquired in Scan, Product Ion scan, or Neutral Loss scan mode; found Unspecified”.
Temporary Solution:
n/a
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
MRM Transition/Collision Energy gives good chromatogram in Qual, none in Quant
Problem:
An MRM chromatograms extracts with no problem in Qual. However, when in Quant same MRM signal does not give me a chromatogram. Instead, it says “Failed to access sample data” and “Object reference not set to an instance of an object”.
Temporary Solution:
None.
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Method Global parameters don't display correctly when loaded from quantmethod.xml file
Problem:
Load a method from quantmethod.xml file, the parameters in file don't show correctly on UI. However, when method is applied to the batch then return to the method editor the correct values are shown.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
ADC data type not appearing in menus, cannot be set up in Quant Method
Problem:
When setting up a method the ADC data does not appear in the menus for signal type so the method cannot be configured to use ADC
Temporary Solution:
None
Fix Information:
Fixed in B.05
Keyword:
One-line Description:
MRM Graphic with no peak shows ~40 count baseline until scaled
Problem:
When there is no variation in baseline (flat line) in the MRM, the graphic in Quant initially shows a baseline at ~40 counts. If the scale is changed, example to the lowest Cal level, it will show the baseline at zero counts. Once the scale is corrected cannot change the scale to produce a baseline of ~40 again, until the batch is re-opened.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Zero peak does not work when the quantitation is based on height
Problem:
Zero peak does not work when the quantitation is based on height. Height is not set to zero.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Incorrect label in Compounds at a Glance while overlaying sample groups
Problem:
In Compounds-at-a-Glance there is the possibility to overlay sample groups. Feature is broken in Quant B.04.00 Build 4.0.225.0 and, instead of the compound name itself, the label in the top right corner of the chromatographic windows becomes “$(CompoundName)”:
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
If 'Acq Date-Time' is null, cannot Add Calibration, Average Calibration, or Replace Calibration
Problem:
An error message is obtained if 'Acq Date-Time' is null and Add Calibration, Average Calibration, or Replace Calibration is attempted.
Temporary Solution:
Be sure the 'Acq Date-Time' field has an appropriate value.
Fix Information:
All data files should have a value in the 'Acq Date-Time' field unless the data files are old or unusual.
Keyword:
One-line Description:
Sample Group and Sample/Compound Group do not default to "All" when a new batch is opened
Problem:
The Sample Group pulldown menu in the Filtering toolbar, and the Sample/Compound Group pulldown menu in the Group Navigation toolbar do not default to "All" when a new batch is opened. Sample Group retains the last value, and Sample/Compound Group is blank. (If Quant is closed, then opened, Filtering and Group Navigation pulldown menus default to “All”.) This defect is observed in B.04.00 SP2.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
MatrixSpike compound type should calculate and display Accuracy for calibrators
Problem:
Software does not calculate and display Accuracy for calibrators when compound type is MatrixSpike.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Cannot modify "Integration Parameters" with Quant Compliance loaded
Problem:
If an attempt is made to modify "Integration Parameters" from the Compound Information window, a message is displayed saying that you don't have permission to execute that command.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Quant Audit Trail not reporting or incorrectly reporting three commands
Problem:
1) The “Action” for “CmdSetMethodTargetCompoundAttribute” is blank when assigning ISTD compounds to targets and assigning ISTD concentrations. 2) The “Action” for “CmdSetQualifierIntegrationParameters” is blank. 3) When the integrator is changed from General to Universal and one Universal parameter is changed, the integration parameters listed don’t reflect the parameters entered.
Temporary Solution:
Apply SP2
Fix Information:
Fixed in Quant B.04.00 Sp2
Keyword:
One-line Description:
Calibration point which appears above the regression line calculates out to a low value
Problem:
Under some conditions the points displayed in the calibration curve and the calculated conc in the BAG table do not agree. The values in the calibration curve are correct while the BAG table is wrong.
Temporary Solution:
None
Fix Information:
Apply SP2 to Quant B.04, fixed in B.05.00
Keyword:
One-line Description:
Unable to 'Append Method from Acquired Scan Data with Library Search'
Problem:
Method created using Method --> New --> New Method from Acquired Scan Data with Library Search. Attempt to use 'Method --> Append --> Append Method from Acquired Scan Data with Library Search' results in error message.
Temporary Solution:
None
Fix Information:
Apply SP2 to Quant B.04.00
Keyword:
One-line Description:
Non-zero concentration calculated for a sample where ‘Zero Peak’ was performed after re-analysis
Problem:
Analyze Batch after Zero Peak performed shows non zero calculated concentration for the zero-peaked sample.
Temporary Solution:
none
Fix Information:
Apply SP2 to Quant B.04.00
Keyword:
One-line Description:
Change in EIC's Exp RT in Quant method makes significant change to chromatogram
Problem:
When editing a method, the customer notices that if he changed an expected retention time (e.g. 14.482 to 14.3) it makes significant changes to the chromatogram, where new peaks emerge in the middle of the chromatogram. It appears that there is also a change to either the accurate mass extracted for the EIC (this is Q-TOF data) or the extraction window. Issue on B.04.00 (Build 225.0) and B.04.00 SP1 Build 225.2.
Temporary Solution:
None
Fix Information:
Apply SP2 to Quant B.04, Fixed in Quant B.05.00
Keyword:
One-line Description:
Compound name displayed in the Compound pulldown menu is incorrect
Problem:
Compound name displayed in the Compound pulldown menu in the Batch Table toolbar is from the previously opened batch when changing between MRM batches.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Perceive compounds setting for Sample Information window of batch mode in Quant Q-TOF is using LC defaults
Problem:
Perceive compounds setting for Sample Information window of batch mode is using LC defaults and user changes on deconvolution parameters in the Scan Analysis Parameters dialog box, 'perceive compounds' is always using LC default parameters for deconvolution.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Average Qualifier Ratio ignores manual integrations and gives incorrect answer
Problem:
When running Update -> Average Qualifier Ratio in the method editor the calculated ratio is always the same (and wrong) with or without manual integration. The same value is reported with or without manual integrations even when this significantly alters the ratios seen in the cal samples. It also does not appear to be the correct average ratio.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Update Qualifier Ratio when the selected sample has no Ratio gives error
Problem:
When "Update Qualifier Ratio" is attempted using "Select All", error message is displayed, "QuantAnalysis has encountered a problem and needs to close ...".
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Audit trail does not display previous response information after Manual Integration
Problem:
Previous and current information for response and x, y coordinates are not displayed in the 'Action' field of the audit trail after manual integration is performed.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Cannot launch Qual Application from tool menu when only Qual B.04 is installed
Problem:
From Quant, use the Tool > Start Qual Application menu item. Error message appears "Cannot find Qualitative application file './AgtQual.exe' Would you like to specify?".
Temporary Solution:
Manually navigate to the file system to the Qual B.04.00 application AqtQual.exe. or Launch Qual from desktop or start menu and then navigate to desired file
Fix Information:
Resolved in B.05.01. Use Tools --> Actions --> Go To Qualitative Application.
Keyword:
One-line Description:
Chromatograms with missing MRM data not centered on expected RT and missing data not clearly identified
Problem:
Chromatogram is not centered around the expected RT. The problem is due to Quant method having a RT window that extends beyond the data acquisition window for the MRM being plotted. When data collection stops before the Quant method expects (RT + Right RT delta) the software takes the latest data point found in the method window and uses this to assign the right side. The same process is used on the left side of the chromatogram (using RT – left RT delta) and the earliest data point in the window to set the left side graphics limit.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00 When RT Reference display is ON, software will center the chromatogram by Reference RT
Keyword:
One-line Description:
Conversion of Quant B.01.03 batch to B.04.00 gives "NO DATA POINTS"
Problem:
When B.01.03 batches are opened with B.04.00 all of the compounds had "NO DATA POINTS" in the Compound Information pane. B.04.00 was creating an MSTree2.bin file of only 45 bytes which is far too small and indicates a failed conversion.
Temporary Solution:
Disable Indexed Data Access. This is done by editing the “QuantAnalysis.exe.config” file in the in the \Program Files\Agilent\MassHunter\Workstation\Quant\bin directory and changingto . Note that this will greatly slow data processing in the application.
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Quant B.04.00 SP3 Crashes When Trying to Change Quantifier
Problem:
Seen on LC/QQQ with pos / neg switching and give a Quant crash when trying to change the quantifier in the method. Go to qualifier setup in the method and click on the product ion and Quant crashes.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Previous integrator parameters is null in audit trail when should be default Universal integrator parameters
Problem:
For an instance of 'CmdSetQualifierIntegrationParameters', the “previous value’ is null in the audit trail when the previous values should be the default Universal integrator parameters. Observed in B225.19 (Quant SP2 build)
Temporary Solution:
None
Fix Information:
Since the most common setting is for the qualifier to use the same integration parameters as parent target ion, and the qualifier IPM is not as important as the target ion IPM, this issue will not be fixed.
Keyword:
One-line Description:
“New Method from SCAN” or “New Method from SCAN with Library Search” will fail with the error message
Problem:
This problem is seen when SIM & SCAN are acquired in the same run. Quant failed to create method from scan data and gives an error message "Deconvolution failed: Source array was not long enough. Check srcIndes and length, and the array's lower bounds."
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Help is incorrect for Average Response Factor
Problem:
Help says that Average Response Factor is calculated for "enabled calibration points". But the Avg Resp Factor is being calculated for calibration and QC points.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
RRT outlier fails when calibration curve is taken from another batch
Problem:
The RRT outlier only works correctly when the current batch contains calibrators. If Calibration curve is loaded from a previous batch the RRT outlier fails. This problem may be seen with the DOA batch. 1) Create new batch 2) Add only Samples and blanks 3) Open method from existing batch (with full cal curve) 4) Analyze batch 5) RRT fails 6) Add QCs 7) Analyze batch 8) RRT fails (but should work as Help says it uses QC or Cal) 9) Add Calibrators 10) Analyze batch 11) RRT works The calibration table stores only the response not the RT of the calibration standards so the calculation must fail (see image) when there are no Calibrators (or QC) samples in the batch which can supply the RT.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Quadratic calibration curves give no or Infinity conc. on samples > highest level
Problem:
If a sample’s response is higher than the highest cal level when using a Quadratic calibration curve, Quant could give the following 3 conditions: 1) Conc is calculated with Red Background—Flag conc. is outside of calibration range. 2) Conc = Infinity with Red Background--- Flag conc. is outside of calibration range. 3) Conc is not calculated (empty in Calc Conc column) with pink background over all cells --- Quantitation message displays “Response is beyond the range of calibration curve” Case 1 is the expected situation where the response of the sample falls somewhere on the calibration curve, but Case 2 and 3 occur when the response of the sample is above the top of the parabolic curve fit. Case 2 occurs when the peak is found by automatic integration (no manual integration). Case 3 occurs when manual integration is performed . Seen in B.04.00 including service packs (SP1 and SP2)
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Quant Help for 'Ignore Peaks Not Found' is incorrect
Problem:
B.04.00 version Help says 'ignore non-target peaks', but this feature is actually for target peaks.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Help for CC Relative Response Factor outlier is incorrect
Problem:
Should read RRFcc < MinimumCCRelativeResponseFactor
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
"Cannot find RT calibration file" displayed and no calculated results displayed if RT cal file path is invalid
Problem:
If the RT calibration file path referenced in the Library Search tab is invalid and a reference library is setup in the method, applying the method and analyzing the batch causes a Quantitation method ("Cannot find RT calibration file " to be displayed for all samples.
Temporary Solution:
Remove the RT calibration file path: Select Method --> New --> New Method from Acquired Scan Data with Library Search. Select any library and sample for the Library and Sample paths, respectively. Click 'Advanced'. Select the Library Search tab. Delete the RT calibration file path reference. Click OK. Click Cancel. Press F5 or click Analyze Batch. Results are displayed and Quantitation message about not finding RT Cal file disappears.
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Error message obtained when using 'New Method from Acquired Scan Data with Library Search'
Problem:
Attempting 'New Method from Acquired Scan Data with Library Search' sometimes results in error with message “Failed to create method from scan data Deconvolution failed: Arithmetic operation resulted in an overflow.” Problem seen in B.04.00
Temporary Solution:
None
Fix Information:
Fixed in B.04.00 SP2
Keyword:
One-line Description:
Fixed graphic for X in Quant report
Problem:
When setting up the fixed graphic range for X, compound extracted window doesn't re-plot using the new fixed range, it just re-scaled in the new range. This can result in missing (blank) signal when the scale extends beyond the RT window of the method.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
"DA. Date-Time" in Batch Sample Information is not set to a value
Problem:
The "DA. Date-Time" field/column in the batch Sample Information is always blank. 'DA. Date-Time' field should store the date and time that the sample’s data was last processed.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Zoomed graphics copied to PowerPoint expand when Ungrouped
Problem:
Display a TIC or EIC. Zoom in to show small peaks, letting large peaks go off scale on the top. Copy image to clipboard. Paste into PowerPoint as metafile or enhanced. Ungroup a few levels. Zoomed TIC or EIC goes off the top of the slide rendering it useless for presentation. The copy to clipboard is copying all the data in the y-scale, not just the zoomed data. The x-scale seems to be okay.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Universal integrator parameters cannot be changed when using "Integration Parameters" feature
Problem:
Unable to change the Universal "Threshold" parameter after initial change using "Integration Parameters" feature.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Library Editor JCAMP Import crashes
Problem:
Some JCAMP files will cause the Library Editor to crash during import.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Quant Help for Average Retention Time feature is incorrect
Problem:
The 'Average Retention Time' feature does not involve sample-averaging. Also, the setup dialog that displays when the feature is selected does not contain checkboxes that allow the selection of Cal or QC samples.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
In compliance mode report generation does not appear in Audit Trail
Problem:
Use ATM Configuration to require User Validation and reason for generating reports. No reason is asked, no password required and the task does not appear in the Audit Trail.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Apply same integration parameter to qualifiers as parent ions, smoothing doesn't look same
Problem:
In Integration Parameters Setup, select integration parameters for qualifier ion, check the box "Use same parameters as Parent compound". Smoothing is performed but "Smoothing" checkbox is not checked as in target ion parameters.
Temporary Solution:
none
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Quant crashes when ‘Average Calibration replicates’ is performed
Problem:
If the Response is null for any Cal samples and "Average Calibration replicates" is performed, Quant crashes. Seen in B.04.00
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Universal integrator integrates peak with far too large area compared to other peaks
Problem:
The use of Tangent Skim and Baseline Hold events in Universal integrator have resulted in inaccurate peak area calculations.
Temporary Solution:
Avoid use of Tangent Skim and Baseline Hold events.
Fix Information:
Will not be fixed
Keyword:
One-line Description:
Warning - Qualifiers have the same M/Z value = x in QQQ method editor unnecessary
Problem:
When a QQQ method has the same product ion but different precursor ions an unnecessary warning message is given. The message would be reasonable on a MS method and indicate a duplicate signal but in QQQ it is reasonable to select the same product M/Z if the precursor ions are different. This logic could be extended to allow the same transition to be used in the qualifier if they have different collision energy and collection energy delta is such that they are different signals.
Temporary Solution:
Ignore warning.
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Replace Calibration for ONE level adds multiple levels into the method
Problem:
Performing a “Replace Calibration” on a single level causes the method to become corrupt. The problem is that when the replace calibration is performed on a compound (A) that shares an internal standard (IS1) with some other compound (B) then the software will add an additional level to IS1. At that point you cannot enter the method editor and edit the method without having validation problems due the mismatch in levels between compound (B) and the ISTD.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00 The new behavior block updates unless the update is made to ALL compounds that shares an internal standard. This avoids the corruption of the method but limits the functionality to replacing sets of compounds rather than to single compound or subset of compounds.
Keyword:
One-line Description:
Manual integrations are lost after same method is applied to data
Problem:
Problem occurs when the method editor is entered while the sample and compound in the batch table has no peaks integrated. This is frequently true of blanks and samples that do not contain the chemical in question. Once this is done if the method is applied to the batch, even if no changes to the method are made, manual integrations will be lost.
Temporary Solution:
User must insure that there is an integrated peak.
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Quant Help for Average RT and Average RRT is incorrect
Problem:
Relative Retention Time and Average Relative Retention Time is incorrect as the Help says “enabled calibration points”. Calculations indicate that it uses QCs and calibration points.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
New methods created in B.05.00 cannot find compounds in older Drugs of Abuse data set
Problem:
If a new method is created in B.05.00 using the pre B.05.00 Drugs of Abuse data files from previous Quant media some compounds will not be found. Methods created by versions of Quant before B.05.00 give results are as expected. This issue is due to a change in the default mass extraction window to plus and minus 0.05 for MRM data. Drugs of Abuse data files on the B.05.00 media were manually-edited so peaks are found.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Concentration units not displayed in Compounds-at-a-Glance
Problem:
No Concentration units are displayed when "Display units for Conc. and RT" is checked in Peak annotations for Compounds-at-a-Glance. Seen in B.04.00 and all service packs
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Hash Code and Audit Trail boxes are not checked in Check Batch Files
Problem:
After checking batch in Check Batch Files (Quant SP3), the Hash Code and Audit Trail boxes are not checked for batch files. Audit Trail box is not checked for sample files.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Quant does not launch Qual B.04 from menu
Problem:
If the user has installed Qual B.04 and B.03.01 and they use the main menu (Tool > Start Qual Application) to launch Qual and open the current file Quant launches Qual B.03.01 rather than the current release. If Qual B.04 is installed but B.03.01 is not installed the Quant brings up a message "Cannot find Qualitative application file: './AqtQual.exe' Would you like to specify? If the user knows where the application is loaded they may select "Yes" and navigate through multiple directory levels to complete the launch.
Temporary Solution:
Navigate file system to complete the launch or launch Qual B.04.00 from desktop and then load desired data file.
Fix Information:
Resolved in B.05.01. Use Tools --> Actions --> Go To Qualitative Application.
Keyword:
One-line Description:
Compound at a glance crash when link Y axis is been used with wrap row function
Problem:
View, Compounds at a glance, Layout, Setup, Organize tab ………then use the following step : Organize Rows by; Compound Overlay; None – Target only Review Mode; Compound by Compound Display options; Baselines, Uncertainty, Fill peaks Using the above - Compounds at a Glance will be OK. But it only fills the top row of the grid. When Wrap rows feature is checked, applied then select "Link Y axis" the system crashes. This happens with different batches of data. Seen in B.04.00 SP2
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
When the user selects "Qualifier Outliers" in online Help, they get Help for the Queue Viewer
Problem:
If the user goes to the Index and selects "Qualifier Outliers" in online Help, they get Help for the Queue Viewer. Seen in B.04.00 and all B.04.00 service packs
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Peaks not integrated when non reference window is set to minutes
Problem:
Agile integrator finds peaks in the data file when non reference window type is percent but does not when set to minutes. This issue is observed when the window is set very wide by either minutes or percent so that the peak should never be rejected by being outside of the window. Seen in B.04.00 SP2
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Display of Qualifier range is the same when Uncertainty settings of Relative vs. Absolute
Problem:
When Qualifier Uncertainty is Relative, the graphic representation of the band in the Compound Information is correct. But when Qualifier Uncertainty is Absolute, it is unchanged from Relative.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Manual integration of batches using Standard Addition fails
Problem:
If a batch has the Global Standard Addition checked then the manual integration feature fail to operate properly. 1) Attempt a manual integration of one of the calibrator target compounds. The peak will get an * placed by it and the MI in the batch table will be checked, but the graphic for the area will snap back to its original position and the area will not change. 2) Uncheck Standard Addition , then manually integrate a target peak, manual integration works and the concentration is calculated. Check Standard Addition, apply the method to the batch and analyze the batch. No concentration is calculated. Observed with MS/MS and Agile as integrators. Observed in B.04.00 SP2
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
'Fit to Highest Cal Level' truncates top of peak in Compound Information
Problem:
When using Fit to Highest Calibration Level, the top of the peak can become truncated or clipped in the Compound Information area.
Temporary Solution:
None
Fix Information:
Fixed in B.05.00
Keyword:
One-line Description:
Quantitative Analysis B.04.00 SP2 is unable to open a batch
Problem:
MassHunter Quantitative Analysis B.04.00 SP2 is unable to open a batch or create a new batch for specific users. For some users, when they try to either open a batch or create a new batch in MassHunter Quantitative Analysis B.04.00 Service Pack 2 (SP2), they will see an error message: 'QuantAnalysis has encountered a problem and needs to close. We are sorry for the inconvenience.' However, when other Windows users log onto the PC, they are able to successfully open a batch or create a new batch.
Temporary Solution:
The problem is caused by corrupt user-specific files that control the UI display in MassHunter Quantitative Analysis rev. B.04.00 SP2. Reinstalling MassHunter Qualitative Analysis will not fix the problem. To fix the problem: 1) Select the "Restore Default Layout" button in the Batch Table view which appears when MassHunter Quantitative Analysis starts up. Then try to open a batch or create a new batch. If 1) above does not fix the problem 2) Delete the corrupt files so they can be recreated with default values a) Shutdown MassHunter Quantitative Analysis b) Delete the files in the folder C:\Documents and Settings\\Application Data\Agilent Technologies_Inc\QuantAnalysis.exe\4.0.225.19 (where "user name" is the name of the user currently logged into Windows, e.g. Administrator.) c) Restart MassHunter Quantitative Analysis. The files will have been recreated and the user should now be able to open batches and create new batches.
Fix Information:
Resolved in B.05.00
Keyword:
One-line Description:
Cannot open Quant Familiarization Guide from the Help with Internet Explorer 9
Problem:
In Quant, select Help --> Contents. Under the 'Getting Started' section (purple), click 'Read the Familiarization Guide'. Quant closes and an error message is displayed.
Temporary Solution:
Go to the 'Manuals' folder on the Quant DVD and open the Familiarization Guide file from there. Alternatively, if Internet Explorer 9 is installed, change to Internet Explorer 8.
Fix Information:
Resolved in B.05.01
Keyword:
One-line Description:
Crash occurs if levels exist in the Calibration table and the Quant/Actions/DummyCalibrationSetup script is run
Problem:
Crash occurs with these steps: 1) Enter Method Edit 2) Have levels exist in the Calibration table for any target compound 3) Select Tools-->Actions-->Custom Action 4) Select the DummyCalibrationSetup script from MassHunter/Scripts/Quant/Actions 5) Click Open 6) Application crashes
Temporary Solution:
Delete any existing calibration levels from the Calibration table in Concentration Setup prior to running the DummyCalibrationSetup script.
Fix Information:
This script is designed to be run when no levels exist in the Calibration table.
Keyword:
One-line Description:
When using Compliance, a command is missing from the Command Group in ATM Configuration
Problem:
Message displayed saying that a Quant command could not be performed because the right permissions were not in place, even though ATM Configuration was in default (should be able to do perform all commands). Turns out the command is missing from the Command Group in ATM Configuration.
Temporary Solution:
Re-enter permission settings in ATM Configuration after installing the new Quant version.
Fix Information:
Resolved in Quant version B.05.01. 1) Copy the ComplianceConfiguration.xml file from the Program Files/Agilent/MassHunter/Workstation/Quant/bin folder, then copy it back after installing the new Quant version. 2) ‘Export’ the compliance file from ATM Configuration, then ‘Import’ the saved file from the MassHunter/log/quant folder.
Keyword:
One-line Description:
Retention Time method validation errors are not flagged the first time the quantitation method is applied for scan data
Problem:
Retention Time method validation errors are not flagged the first time that the method is applied after 'Append Method from Acquired Scan Data' or 'Append Method from Acquired Scan Data with Library Search' is used. Upon entering Method Edit and applying the method a second time, the Retention Time errors are flagged. After appending a method from a scan data file, data access is switched to the scan data that was used to do deconvolution/library search. In order to catch the error, need to switch back to the original data.
Temporary Solution:
n/a
Fix Information:
The workaround is to exit Method Edit applying the method, then re-enter Method Edit and apply the method again.
Keyword:
One-line Description:
S/N and Noise Calculations give Infinity with the Universal Integrator
Problem:
The signal-to-noise calculation gives infinity for the S/N calculation and the Noise calculation when the Universal integrator is selected.
Temporary Solution:
None.
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Cannot 'View demonstration videos' from Quant Help if Adobe Flash Player is not installed
Problem:
In Quant, select Help --> Contents. Under the 'Getting Started' section (purple), click 'View demonstration videos'. Click on any link to a demonstration video. The message "'Loading multimedia control' is displayed and no video plays.
Temporary Solution:
n/a
Fix Information:
Install Adobe Flash Player.
Keyword:
One-line Description:
Blank Response outlier not flagged for sample of type 'DoubleBlank' if no ISTD found
Problem:
The target peak is not integrated if ISTD peak is not found. Attempting to manually integrate the target peak results in the message "Manual integration failed. Previous integration results restored"
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Average RF RSD outlier is not visible in Compound Table view
Problem:
Average RF RSD outlier is flagged in Flat Table view but not Compound Table view. RF RSD field is also blank Compound Table view.
Temporary Solution:
Use Flat Table layout to check RF RSD status.
Fix Information:
n/a
Keyword:
One-line Description:
Reference Library Match Score is not shown in the Spectrum in the Report
Problem:
When generating a report for samples with Reference Library Match Score, the Match score is not shown in the spectrum on the report. Observed in B.05.00
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Mixed SIM and MRM batch crashes when generating report – bad SIM spectrum
Problem:
Quant fails to extract a SIM spectra. This results in a blank spectrum in the Sample Information section of Quant and a crash in reporting that results in no report being generated.
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
QUANT B0500, does not have IQT reference file support for the COMPLIANCE module
Problem:
The IQT reference file for the COMPLIANCE module is not available in the QUANT B.05.00 installation
Temporary Solution:
None
Fix Information:
Resolved in B.05.00 Service Pack 01.
Keyword:
One-line Description:
Names of the integrators in the audit trail are not the same as the ones in the drop down menu
Problem:
The audit trail displays Universal as 'Genie' and General as 'RTE'. The method editor displays 'Universal' and 'General'.
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Unknowns Analysis, selected library path gets overwritten with C:\MassHumter\Library
Problem:
After altering the library information in the Advanced version of the Unknowns Analysis method editor the new path of gets overwritten with C:\MassHumter\Library
Temporary Solution:
Create the directories C:\MassHunter\Library and copy the library into C:\MassHunter\Library
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Unknown Analysis will crash if the SAME library is selected two times
Problem:
If the same library is specified more than once in the list of libraries then Unknowns Analysis will crash during the library search.
Temporary Solution:
Enter the library only once.
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Audit Trail missing 'Action' information for some commands
Problem:
Some blank lines written in the audit trail, on occasion when clicking on the “apply” button after a method change (example - changing the integrator type).
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Unknown Analysis – crashed when 176 or more data files are loaded and user scrolls down the sample list
Problem:
If a large number of files are loaded into Unknowns Analysis and the user scrolls down the list the application will crash.
Temporary Solution:
Limit number of data files loaded.
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Quant Method Integrator Peak Filter – maximum number of peaks nonfunctional
Problem:
The maximum number of peaks cannot be activated or retain values. 1) Open Method for edit 2) Go to Integration Parameter Setup 3) Go to Peak Filter 4) Check “limit to the largest” in the “Maximum number of peaks” section. 5) Edit the default number in “peaks” from 100 top 5 6) Click Apply or Apply to All 7) Click OK 8) Reopen Integration dialog and check Peak filters’ 9) Values entered are gone (set back to unchecked and 100)Limit number of data files loaded.
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Audit trail - timed events for integration codes need to be spelled not coded
Problem:
Timed events for integration that are being logged to the Compliance Audit Trail, are not understandable. Customer wants to know what :MECEND-ECEND-MECALLVALLEY-ECALLVALLEY-MECBASELINE-ECBASELINE and also EC , MEC”, mean. The list on the right should appear in the audit trail instead: ECALLVALLEYS Baseline All Valleys ON ECAREAS Area Sum ON ECBACKWARD Baseline Back ECBASELINE Baseline Next Valley ECDISABLE Solvent Peak OFF ECEND Integrator OFF ECHOLD Baseline Hold ON ECIMMEDIATE Baseline Now ECNEGATIVE Negative Peak ON ECREJECT Area Reject ECTANGENT Tangent Skim ECTHRESHOLD Threshold ECWIDTH Peak Width MECALLVALLEYS Baseline All Valleys OFF MECAREAS Area Sum OFF MECDISABLE Solvent Peak ON MECEND Integrator ON MECHOLD Baseline Hold OFF MECNEGATIVE Negative Peak OFF
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Library editor: Running the mol script will crash the library editor
Problem:
Running the scrip, DownloadMolfiles.txt, will crash the library editor. Seen in B.05.00
Temporary Solution:
Fixed for B.05.01, but the workaround in B.05.00 is to have the following lines at the top of the script file: //!
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
No "BL Draw" values or baseline calculation points displayed for 'Sulfa_QTOF_TargetedMSMS.batch.bin'
Problem:
BL Draw column in the batch table is blank. If Baseline calculation points is checked in Compound Information - Chromatogram properties, no rectangles or lines are displayed in the target chromatogram window. Data is from the Quant shipping CD: QTOF/QTOF_4GHz_SulfaQuantTargetedMSMS. Seen in B.05.00
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
.xds files are not installed in B.05.00
Problem:
Schema files (.xsd) are missing from the Quant template directories. They were installed in the B.04 version of the software and are required for some modifications of the templates.
Temporary Solution:
Request .xds files
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Printing “\” in path results in improper PDF output in Japanese Quant
Problem:
Released B.05.00 (build 291) Japanese , Win 7 SP1 64bit, Excel 2010 1. Generate report 2. select "PDF" as format 3. PDF report is created. but "\" on batch data path is misprinted. Excel has no problem, so direct printed report has no problem. 4. when I print out "PDF" report, can see misprinted "\", too.
Temporary Solution:
Print from Excel Use CuteWriter as PDF driver
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
"Fit to Lowest/Highest Calibration Level" function is inactive in compounds at a glance
Problem:
When selecting Setup/Organize/Overlay None - target Only, the "Fit to Lowest Calibration Level" and "Fit to Highest Calibration Level" functions are not active. When selecting other Organize options, such as target and qualifier, both functions are ON.
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Noise not calculated when no peaks found
Problem:
The noise field is left blank if the integrator fails to detect a peak. The noise is not even calculated when the method specifies a noise region.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01 The "Noise" field in the Compound Results table is still blank when no peaks are integrated but new field "Noise of Raw Signal" found in the Compound Method table contains the noise value.
Keyword:
One-line Description:
When noise regions are manually specified, the graphic representation does not change
Problem:
When noise regions are manually designated in the method, the baseline calculation points that designate noise regions for the parameterless integrators remain the same. In this case, the graphic in the chromatogram window is not correct.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
B.05.00 DrugQuant Report fails to display Qualifier graphics when no target peak found
Problem:
Example of this issue can be seen in the DOA data set in the Blank and Sample1 for Amp. When no Qualifier peak is found for a compound the report fails to properly VLOOKUP the required graphic because the SampleID, CompoundID, and PeakID fields are blank and thus the key is malformed. See attached modified report template for one approach that works. Seems that the key fields from PeakQualifer were used rather than TargetQualifier where the graphic is stored.
Temporary Solution:
Template available on request.
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Crash when attempting 'Restore Integration Parameters'
Problem:
Steps to reproduce: 1) Open DrugsOfAbuse batch 2) Delete all but 3 samples 3) Perform IPM on each of the three samples changing Agile to Universal 4) ‘Restore Int Params’ on one sample displays the message --> "Failed to find a sample with unmodified integration method for compound Amp" 5) Restore Int Params on the 2nd or 3rd sample --> crash Observed in B.05.01 Build 307.1 and B.05.00.
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Quant cannot generate report for batch containing only 2D data
Problem:
For 2D data only batch, Quant crashes when generating report. Not dependent on Template and can't be avoided by generating report without graphic files.
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Schema file (batch.results.xsd) for Unknown Analysis is missing from the default template directory
Problem:
batch.results.xsd is required to add new data fields to an existing template and customize the template.
Temporary Solution:
Request file
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Target Peak doesn't show the center of the windows in compound information
Problem:
In method editor, select "Show Reference R.T" in compound information property. Move to next compound at method table. Target Peak and Qualifier peak don't show in the center of the window. Uncheck "Show Reference R.T". Target Peak and Qualifier peak show in the center of the window. Seen in B.05.00
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Changing Retention Time Window CC units for CC Retention Time outlier does not change the outlier range
Problem:
Changing the 'Retention Time Window CC' units for 'CC Retention Time' outlier, from Percent to Minutes or vice-versa, does not change the outlier range and result after the method is applied and Analyze Batch or Quantitate Batch is done. Also, when the 'Retention Time Window CC' outlier is clicked in the method, the 'RT Units' column should display along with the 'Retention Time Window CC' column. Seen in B.05.00
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
A Zero Peak still flag as being below the calibration range
Problem:
When the quantifier is set to a zero area with the Zero Peak function it is not treated as “not found” but as a real value of zero and the below calibration range outlier is activated.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
User Annotation field is not protected and is corrupted when you edit the method and apply it to the batch.
Problem:
Users wish to use the User annotation field to record comments during data review. However the data is not protected during a method edit session (edit method, apply) and the comments are lost. Normal data review practice can easily require a method modification after some comments have been entered. Seen in B.04.00 and B.05.00.
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Concentration is incorrect with Standard Addition and no ISTD in the batch
Problem:
Re-analyzing a batch after manual integration of a sample does not update the calibration curve. After clearing calibration then analyzing the batch, the calibration curve is updated. But the calculated concentration does not agree with the calibration curve equation. Seen in B.05.00
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Target compound response is not updated in the Calibration Curve window after manual integration when using Standard Addition
Problem:
Manually integrate a sample in a batch that is using standard addition, then Analyze Batch. After the first manual integration, the response in the Calibration Curve pop-up window is updated. After a second manual integration, the target compound final conc is updated in the batch table, but the response in the Calibration Curve pop-up window remains unchanged.
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Mass accuracy value displayed is not the expected value when profile data is used
Problem:
For profile data, the manual mass accuracy calculation does not match the mass accuracy displayed in MassHunter Quantitative Analysis.
Temporary Solution:
MassHunter Quantitative Analysis does not support profile data. Centroid data must be used.
Fix Information:
Will not be fixed
Keyword:
One-line Description:
Calculation of Library Match Score changed
Problem:
Revised algorithm results in improved Library Match scores. Library Match scores in Quant B.05.01 may not be the exact same as in previous versions of Quant.
Temporary Solution:
n/a
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Method validation error that GC-FID peak is outside the retention time window
Problem:
If a non-MS signal is selected, such as a GC-FID, and the retention time (RT) of the signal is outside the RT range of the MS signal, the method editor validation displays an error message indicating that the retention time is out of the window. Issue observed in Quant B.04.00 and B.05.00.
Temporary Solution:
Ensure that the MS signal is collected over the RT range of other signals.
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Uninstallation of MassHunter Quantitative Analysis Reporting will uninstall Report Templates
Problem:
Prior to MassHunter Quantitative Analysis B.05.01, Report Templates would remain after uninstallation of the Quantitative Analysis Reporting application. Beginning with MassHunter Quantitative Analysis B.05.01, Report Templates will be removed with uninstallation of the Quantitative Analysis Reporting application. Customized report templates with file names that are different from the installed templates will remain after uninstallation.
Temporary Solution:
n/a
Fix Information:
Report templates from previous versions of MassHunter Quantitative Analysis (prior to B.05.01) will not uninstall.
Keyword:
One-line Description:
Data files are displayed in random order in 'Add Samples' dialog box when files are on some server locations
Problem:
When 'Add Samples' is done from a particular server location, the samples are not in any particular order. If the data files are moved to a local drive, 'Add Samples' displays the data files in the order that the files were collected. Issue seen in B.04.00 SP2
Temporary Solution:
Move files to local drive.
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Translated Chemstation file can't be added to Quant batch
Problem:
ChemStation file was translated using the GCMS Translator. It can be opened in Qual B.04.00 SP2 without error and all signals can be extracted (FID and SIM). However if it is added to a Quant batch it generates the error "There were problems with adding one or more samples: Failed to import sample from worklist" The issue was found to be an illegal character string in the barcode field that was present in the ChemStation file. No Barcode reader was installed and the field contained random data.
Temporary Solution:
None.
Fix Information:
The B.05.02 Build 1032 of the GCMS Translator will remove illegal characters and fixes this problem.
Keyword:
One-line Description:
Average RF in method editor does not match graphics and appears to be based on calibration and QC/CC values
Problem:
The Average RF value shown in the method editor does not match the calibration curve graphics (slope) when Average of Response Factors is the selected curve type. The graphics value is correct and based only on the calibrators. The value in the method editor includes QCs or CCs.
Temporary Solution:
None.
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Batch with SIM data files hangs or aborts during reporting
Problem:
SIM data files are translated to MassHunter format and included in a Quant batch. There are two clear problems. 1) The Spectrum area of the Sample Information is blank and the title area reads “Spectrum: Failed to access sample data” 2) Attempts to generate reports with graphics on fail with by either hanging or give error message. No Graphics report is generated. Problem is associated with spectrum graphics. Problem seen only in B.05.00. Problem does not exist in B.04.00 SP2.
Temporary Solution:
If only non-graphical reports are required then turn off Graphics generation (uncheck "Generate graphics files" at the top of the "Report graphics files" page in Advanced reporting options). If graphical reports are required consider down grading to Quant B.04.00 SP2.
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Unknowns Analysis - Cannot specify an RT Cal file for second library
Problem:
If using more than one library and using different RT Cal file for different libraries, only one RT Cal file can be chosen in the Library Search window of Method Edit.
Temporary Solution:
None.
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
User Annotation is not displayed in Compound Table view of the batch table
Problem:
The User Annotation field is not displayed In Compound Table view. The User Annotation field is only displayed in Flat Table view.
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
VolatileOrganics.batch.bin shipping with Quant B.05.00 gives warning due to improper method
Problem:
When method validation is performed on the VolatileOrganics.batch.bin file from the VOA data set, warnings are displayed because some of the MZ extraction window units for the quantifiers are different from the MZ extraction window units for the qualifiers.
Temporary Solution:
n/a
Fix Information:
Fixed in B.05.01. MZ Extraction window units set to Thomsons for all quantifiers and qualifiers.
Keyword:
One-line Description:
Results from Genie (Universal) integrator are not consistent between MassHunter Qual-Quant and MSD Chemstation
Problem:
Results from Genie (Universal) integrator are not consistent between MassHunter Qual-Quant and MSD Chemstation
Temporary Solution:
n/a
Fix Information:
Difference between Mustang-genie and either Qual-Genie or Quant-Genie: Genie was changed to obtain the same area as all the other MassHunter integrators so manual integration was consistent across integrators (Agile, MS-MS, …). Difference between Qual-Genie and Qual-Genie: Peak areas are in the ballpark of each other. Qual and Quant do not necessarily share the same time or mass windows (RT windows, mass extraction windows). If those settings are different, they can easily explain the different in areas. The units on the X axis are different for different integrators in Mustang. Different units easily explains different values. In MassHunter the same X-axis units is used for all integrators.
Keyword:
One-line Description:
Unable to add 2D compounds after loading a previously saved Quant method
Problem:
The issue is observed when the data file contains signals other than MS such as GC or UV. Normally, if a sample is selected from the batch before entering the method editor, the method editor displays the sample information, such as name and data file name, in the Sample row. The Quant method editor uses this information to determine what signals are present in the data file. If an attempt is made to define a GC or UV signal using the Compound 2D Setup menu item found in Advanced Tasks, Quant populates the drop down list under Scan Type with the GC or UV signals. However if an existing method is loaded into the method editor, either from a batch or method file, the data file name information in the Sample row is cleared. This breaks the link to the data file and the drop down list under Scan Type loses the GC or UV signals. Issue observed in B.05.00.
Temporary Solution:
Exit the method editor, apply changes, then re-enter the method editor. The selected sample will be restored as will the drop down list.
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Entering Method Edit while a TuneCheck sample is selected causes crash
Problem:
If the selected sample in the batch table is a TuneCheck and an attempt to enter the method editor is made the program will crash.
Temporary Solution:
Avoid entering the method editor on a TuneCheck sample.
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
Options in custom to Fit to Lowest/Highest Calibration level lost when switching samples or compounds
Problem:
Setting options to display peaks scaled to lowest or highest calibrator level will be lost when switching samples or compounds in a layout which is doing review by sample or compound. The setting is a temporary zoom that is not automatically applied each time the grid is reloaded with the next sample or compounds data. This issue is also seen when the pre-defined or user defined layouts are used and it attempts to set the scale. It works when initially applied but fails when the next sample or compound is displayed.
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
When no peaks are detected in any calibration sample, the calibration curve is not cleared and rebuilt from the batch data.
Problem:
If no target peaks are found in all calibration samples, performing Analyze Batch does not rebuild the calibration curve. The calibration curve that previously existed (if any) is retained. In this situation, the status bar (lower right) displays “Processed” but the batch table does not reflect results using the current calibration curve.
Temporary Solution:
None
Fix Information:
Do Not Fix
Keyword:
One-line Description:
CC Relative Response Factor and CC Average RF outlier do not display the expected values
Problem:
With two different CC samples in the same batch, CC Relative Response Factor and CC Average RF outliers (1) do not display the expected values, and (2) the different CC samples display the same outlier results
Temporary Solution:
n/a
Fix Information:
Resolved in Quant B.05.01: Different CC samples in the same batch display the expected values for CC RRF and CC Average RF
Keyword:
One-line Description:
Qualifier ratio outlier does not work for samples of type 'Blank'
Problem:
Qualifier ratio outlier does not work for samples of type 'Blank'
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
When a pattern reference library is modified, the accurate mass scores and spectrum are not automatically updated in the Quant batch
Problem:
1) Quant displays a batch with accurate mass scores. 2) The pattern reference library that is used in the Quant batch is modified in Library Editor. 3) 'Analyze Batch' in Quant still displays the original accurate mass scores and spectrum. 4) Re-selecting the library in method edit, applying the method, then analyzing batch still displays the original accurate mass scores and spectrum.
Temporary Solution:
1) Save the batch 2) Close the batch 3) Open the batch 4) Analyze Batch The accurate mass scores and spectrum based on the changed library are displayed.
Fix Information:
The library file is cached, so it's a trade-off for performance..
Keyword:
One-line Description:
'Analyze Batch' does not calculate concentrations for one or more Calibration samples in the batch
Problem:
Samples deleted from the original, analyzed batch so only four samples remain. Sample type changed such that two of the four samples are 'Cal' type and the other two are 'QC' type. Appropriate level entered for each Cal or QC. Concentration of one or more Cal samples is not displayed when Analyze Batch is done.
Temporary Solution:
Enter method edit. Exit method edit applying the method. Then Analyze Batch displays concentrations for all of the samples.
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Sample field too small to hold values
Problem:
The sample field in the header of the batch table does not display long sample names fully even when there is space in the tool bar.
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Names of the integrators in the audit trail are not the same as in the method when using Integration Parameters Modified
Problem:
Audit trail shows integrator’s development name instead of displayed names when Integration Parameters Modified (IPM) is used: 'Universal' displays in the method but 'Genie' displays in the audit trail. 'General' displays in the method but 'RTE' displays in the audit trail.
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Some audit trail fields are not correct for commands implemented in B.05.01
Problem:
Audit trail needs the following corrections: 1) Action for 'CmdZeroOutQualifierPeak' says 'Zero out primary peak ...' instead of 'Zero out qualifier peak ...' 2) Name for Append method from CEF file says 'CmdCreateMethodFromCefFile' instead of 'CmdAppendMethodFromCefFile' 3) Name for Append method from library file says 'CmdCreateMethodFromLibrary' instead of 'CmdAppendMethodFromLibrary' 4) Action for Append method from CEF file says 'Create method from CEF file ...' instead of 'Append method from CEF file ...' 5) Action for Append method from library file says 'Create method from library file ...' instead of 'Append method from library file ...'
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Incorrect Average Response Factor RSD
Problem:
Average Response Factor Relative Standard Deviation is not calculated correctly if calibration samples of type 'Method' are in the method calibration table.
Temporary Solution:
n/a
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
CAS number displays in date format when exported .csv file is opened in Excel
Problem:
After exporting Components from Unknowns Analysis to Excel (.csv), some CAS numbers are displayed in date format. Changing the format of the cell in Excel does not get the original CAS number back.
Temporary Solution:
n/a
Fix Information:
Import the file in Excel instead of opening: 1) Open the .csv file in Excel. 2) Click the ‘Data’ tab 3) Under the ‘Get External Data’ section, click ‘From Text’ 4) Select the .csv file and click Import. 5) Select ‘Delimited’. Click Next 6) Select ‘Comma’. Click Next 7) Select ‘Text’ and select the CAS# column. Click Finish 8) Select ‘New worksheet’. Click OK
Keyword:
One-line Description:
Number of scans is not displayed in the spectrum of the batch table for DMRM data
Problem:
Number of scans is not displayed in the spectrum of the batch table for DMRM data when a qualifier does not have the same precursor ion as the quantifier in Quant B.05.01
Temporary Solution:
None
Fix Information:
Will not be fixed
Keyword:
One-line Description:
No method created for 6490 dynamic MRM pos/neg data
Problem:
'Create New Method from Acquired MRM Data' does not create a method for dynamic MRM data. Data is from a 6490 with pos/neg switching. This issue is observed in MassHunter Quantitative Analysis but is actually caused by LCMS Acquisition.
Temporary Solution:
Method can be created with generic compound names by removing or renaming the .m file in each data folder.
Fix Information:
Final resolution will be in MassHunter Acquisition.
Keyword:
One-line Description:
Setting up a pattern reference library in method edit is not recorded in the audit trail
Problem:
Setting up a pattern reference library in method edit is not recorded in the audit trail
Temporary Solution:
None
Fix Information:
Since the pattern reference library is generated by Library Editor (Library Editor is launched from Method Editor), no method edit command exists.
Keyword:
One-line Description:
ISTD data points don't show in the calibration report graphics
Problem:
When ISTD is checked in the Calibration Curve pane, the ISTD data points are displayed in the Calibration Curve pane. However, they are not present in the graphic files generated by reporting and missing from the reports.
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Add support for accuracy value and its outlier for CC sample type
Problem:
The Accuracy outlier feature was not working in B.05.00 on CC sample type.
Temporary Solution:
None
Fix Information:
Fixed in B.05.01
Keyword:
One-line Description:
MassHunter Quant crash when producing graphical reports on UV DAD data batches
Problem:
MassHunter Quant B.05.00 crashes each time it tries to run a report made with UV DAD data files. Will successfully print if "generate graphics files" is unchecked.
Temporary Solution:
Downgrade to Quant B.04.00 SP2
Fix Information:
Fixed in Quant B.05.00 SP2 and Quant B.05.01
Keyword:
One-line Description:
'Update Retention Times' does not update when peaks are manually integrated
Problem:
Performing 'Update Retention Times' on peaks that are manually integrated does not change the method retention times
Temporary Solution:
A workaround is to do the manual integration in Method Edit (instead of in the batch table), then the method retention times are automatically changed.
Fix Information:
Fixed in B.06.00 SP1
Keyword:
One-line Description:
Batch status is incorrectly displayed as 'Processed'
Problem:
If a calibration sample is manually-integrated, then Quantitate Batch is done, the status bar displays 'Processed' when it should display 'Modified'.
Temporary Solution:
Use 'Analyze Batch' instead of 'Quantitate Batch' to update the batch table results.
Fix Information:
n/a
Keyword:
One-line Description:
Integration differences between Quant B.05.01 and Quant B.05.02 cause differences in spectrum display and accurate mass scores
Problem:
Opening a batch that was created in Quant B.05.01 in Quant B.05.02 and performing Analyze Batch may result in different integration responses and, therefore, different spectrum display and accurate mass scores.
Temporary Solution:
None
Fix Information:
Will not be fixed
Keyword:
One-line Description:
Response is not displayed after clearing manual integration in a batch with Standard Addition
Problem:
In a batch with Standard Addition, performing manual integration then clearing manual integration results displays an empty Response in the batch table.
Temporary Solution:
After clearing manual integration, analyze the batch.
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Purity calculation revised in B.05.02
Problem:
Quant B.05.02 contains a new, completely reworked purity calculation. Changes in purity results between B.05.02 and previous versions should be expected.
Temporary Solution:
N/A
Fix Information:
Changed in B.05.02
Keyword:
One-line Description:
DrugQuant Reports indicates template fails and gives Quant crash.
Problem:
The following DrugQuant reports in Quant B.05.00 and B.05.01 will not generate reports without additional method setup steps (see Temporary Fix) and will generate an error message when run from Quant. "DrugQuant_withOneQualifier_B_05_00" "DrugQuant_withTwoQualifier_B_05_00" "DrugQuant_withOneQualifier_B_05_01" "DrugQuant_withTwoQualifier_B_05_01"
Temporary Solution:
In order to use the Drug Quant report templates to generate Drug Quant report, users have to setup their Quant method by grouping the Target and its ISTD as the same group. This is done by setting the Compound Group field so that each Target compound and its ISTD have a common value. Once users setting up the compound group, then they can generate report using the 2 Drug Quant report templates with one or two Qualifiers. Without setup compound group, it will fail the report. DrugQuant report sorts the target compounds by its Compound Group name, so when you pick the group name, please bearing this in the mind.
Fix Information:
n/a
Keyword:
One-line Description:
Area Reject in Universal integrator parameters is off by 10X
Problem:
When the Area Reject parameter is used, the value must be increased 10 times for the filter to work. This cutoff is being applied to the internal uncorrected area before the 10-fold reduction in area is applied and shown in the batch table.
Temporary Solution:
Area Reject parameter can be increased 10 times
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Signal-to-Noise (S/N) is Infinity when manual integration is done for all integrators except Universal
Problem:
Manual integration causes Noise to be zero and, therefore, Signal-to-Noise to be Infinity. With the Agile integrator, all noise algorithms result in S/N of Infinity except Auto-RMS.
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Baseline calculation points display differently when using Agile integrator vs. MS-MS integrator
Problem:
When you show the baseline calculation region on the MRM chromatogram, the Agile integrator shows regions down by zero counts. MSMS integrator shows the regions on the actual baseline.
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
When sample name is null, observe unexpected display in the Current Sample pull-down menu of the batch table toolbar
Problem:
When creating a new batch, "Agilent.MassSpectrometry.DataAnalysis.Quantitative.UIScriptFImpls.ScriptDataRow" is displayed in the Current Sample pull-down menu of the batch table toolbar. See attached. After the batch is saved then re-opened, the Current Sample pulldown menu displays as expected. Seen in B.05.01
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Spectrum display in method edit and the batch table are different for DMRM data
Problem:
Spectrum display is method edit and the batch table are different for DMRM data when the quantifier and qualifier have different precursor ions.
Temporary Solution:
None
Fix Information:
Fixed in B.05.02. Target compound's spectrum is displayed if target is selected, and qualifier compound's spectrum is displayed if qualifier is selected. Spectrum view in method editor can be different from spectrum view in Batch-at-a-Glance.
Keyword:
One-line Description:
Auto Review Compounds with sorted Compound Table view does not display compounds in the desired order
Problem:
In Compound Table view with the Mass Match Score column sorted in decreasing order, selecting View --> Auto Review Compounds does not display the compounds in the sorted order.
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Load layout in Unknowns Analysis can display columns in different order than seen when saved
Problem:
Load layout in Unknowns Analysis can display columns in different order than seen when saved. An example was reported where the File Name (Samples window) appears in a different place from when Save Layout was done Seen in B.05.01
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Treatment of curve settings - Origin set to 'Include' with 'Weight' set to 1/x^2,1/x, 1/y or 1/y^2 actually ignores the origin
Problem:
When choosing a curve fit setting where Origin = Include and 'Weight' = 1/x^2,1/x, 1/y or 1/y^2, the origin gets a weight of zero. So the origin is not included in the curve.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Method Editor does not automatically update/refresh when copying calibration levels to other targets
Problem:
Some of the data fields in the Calibration table do not update on screen automatically after a calibration level is copied to other levels. The data has been copied correctly but the software does not display the data correctly until the user scrolls down or click on the field.
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Unknowns Analysis shows unit mass on GC-QTOF acquisition software
Problem:
Data files acquired in GC-QTOF version of acquisition are displayed in unit mass while data file acquired under the old LCMS acquisition software show accurate mass. When exported to a Library the library is initially created as unit mass but when manually switched to accurate mass, the accurate mass Info is shown (not rounded to tenths place).
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Samples missing from report when using QuantReport_ISTD_ResultsComplete_B_05_01.xltx as template
Problem:
Problem is seen in Quant B.05.01 when running QuantReport_ISTD_ResultsComplete_B_05_01.xltx as the template where some samples being lost from the summary page in the section on each chemical. The lost samples are the ones next to the page break header.
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Trying to create a new method from a .cef file generated by Qual will display a value of "0" under the precursor ion column
Problem:
Problem in B.05.01 and may be seen when the following steps are taken: In Qual: -Run "Find by Targeted MS/MS" algorithm -Export as a .cef file In Quant: -Create a new batch with data files used in Targeted MS/MS -Method -> New -> New Method from File -Select the just exported .cef file -In the method editor, all values in the "Precursor Ion" column read "0". Product ion column values DO read correctly.
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Odd behavior when ISTD Resp Min % Dev is blank while ISTD Resp Max % Dev contains value
Problem:
A method with a maximum ISTD response percent deviation but no minimum ISTD response percent deviation caused a number of incorrect actions to take place while working with the batch - Peak integration failed - Manual Integration failed - Manually integrated peaks were lost
Temporary Solution:
Setting minimum ISTD response percent deviation = 0 resolves the issue
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Concentrations are only calculated for the first sample in a batch after a sample type is changed
Problem:
If the sample type is changed for a sample (frequently the first sample in a batch), the concentration is only calculated for the first sample when the batch is analyzed.
Temporary Solution:
Enter method edit. Exit method edit applying the method (no changes to the method). Analyze Batch
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
Method does not validate: Target and ISTD calibration tables have different number of rows
Problem:
After a batch is analyzed, entering method edit and attempting to validate the method gives 'Target and ISTD calibration tables have different number of rows' error.
Temporary Solution:
Edit the method making sure that the target/ISTD levels match.
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Cannot manually add calibration levels without entering 'Dil. High Conc.' and 'Dil. Pattern'
Problem:
In method edit, click Concentration Setup. Display the Serial Dilution toolbar. Enter # of Levels, then click Create Levels--> Please enter 'Dil. High Conc.' message displays
Temporary Solution:
Click 'New Calibration Level' under Manual Setup Tasks for each level desired. For example, to add five levels, click 'New Calibration Level' five times.
Fix Information:
n/a
Keyword:
One-line Description:
'Add Calibration’ cannot be performed if no target peak is found
Problem:
Can only add calibration when a response is found for the target compound. If no peak is found, a message is displayed when Analyze Batch or Quantitate Batch is done: "Batch calibration failed. Compound calibration failed. Calibration response not found for compound."
Temporary Solution:
Only add calibration for those compounds where a peak is found
Fix Information:
Fixed in Quant B.05.02 Note: When adding a level for all compounds, the workflow now requires clicking "Select All", even when expected concentrations are entered for all compounds.
Keyword:
One-line Description:
Audit trail report is missing part of the title and has some Action items that are not fully displayed
Problem:
The Audit Trail report (QuantReport_AuditTrail_B_05_01.xlsx) shows the title missing the first letter and some Action items with the last line not fully displayed
Temporary Solution:
The Title and Action items are fully displayed when viewed on screen. Alternatively, the audit trail can be exported to Excel and printed from Excel, or the pdf report template can be used.
Fix Information:
n/a
Keyword:
One-line Description:
Unknowns Analysis - Method problems with mz limits, no effect on score, can’t be cleared
Problem:
Changing the upper mz limit has no effect on match score as the feature is not implemented. More over once a value is entered in the lower or upper mz limit, the value cannot be cleared
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
When using 'Import Calibration Levels from File', method validation flags target and ISTD calibration tables as having different number of rows
Problem:
Previous (Quant B.05.01 and earlier) method validation does not flag when target and ISTD have a different number of rows in the calibration table.
Temporary Solution:
n/a
Fix Information:
If the method has ISTD compounds, add the ISTD compounds to the file being used for import so the target and ISTD have the same number of rows in the calibration table.
Keyword:
One-line Description:
Profile data cannot be viewed and processed.
Problem:
Quant is unable to process profile data.
Temporary Solution:
n/a
Fix Information:
The following changes must be made in the QuantAnalysis.exe config file. In the example given below replace "[" with "<" and "]" with ">". [add key=”UseProfiledate” value=true”/] [add key="UseIndexedDataAccess" value="false"/] This file can be found under: C:\Program File(x86)\Agilent\MassHunter\Workstation\Quant\bin
Keyword:
One-line Description:
Compounds missing from reflibrary.xml that was created from .L library
Problem:
Creating a reference library (*.reflibrary.xml) from a compressed library (.L) results in the reflibrary.xml file missing compounds due to a Compound Name case-sensitivity issue
Temporary Solution:
Manually edit or create the library to ensure that all desired compounds are present
Fix Information:
Fixed in Quant B.05.02
Keyword:
One-line Description:
Blank Offset alters slope
Problem:
When “Blank Offset” is used as origin, the slope of the line is altered from what it would be when set to “Ignore”. This behavior is unexpected based on the information in the help as it claims that the y –intercept will be forced to match the blank level. Such an adjustment would not affect the slope of the line.
Temporary Solution:
None for the behavior of the “Blank Offset”. Blank Subtraction Add-In is delivered as a solution in B.06.00
Fix Information:
n/a
Keyword:
One-line Description:
Manual and automatic integration peak area disagree by 30% for the same peak shape
Problem:
The issue was seen when using the General integrator to find peaks. In some cases when a manual integration is made to adjust the peaks, a large difference in area is seen even when little or no change is made to start and stop integration points. Observed in Quant B.05.00, B.05.00 SP2 and B.05.01.
Temporary Solution:
Switch integration to Agile to ensure identical manual and automatic integration
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Cannot restore Qualifier integration parameters
Problem:
After performing 'Integration Parameters' on a qualifier peak, the 'Restore Integration Parameters' option is not active. Steps to reproduce: 1) Change integrator to 'General' in the batch 2) Perform 'Integration Parameters' on one qualifier peak (Uncheck the box for 'Use same parameters as parent compound' and change one parameter) 3) 'Restore Integration Parameters' is grayed out for the qualifier (and remains inactive after Analyze Batch is done) Can perform 'Integration Parameters' for the related target peak, then 'Restore Integration Parameters' for the target peak clears the integration parameters for the target peak and the qualifier peak. Observed in Quant B.05.01 and Quant B.05.02
Temporary Solution:
None
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Auto-RMS noise algorithm gives a Signal-to-Noise value of 'Infinity'
Problem:
Under certain conditions the Auto-RMS noise algorithm supplied a noise value of zero and thus a Signal-to-Noise value of 'Infinity'. Issue exists in B.05.01
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
‘Spectrum Peak Height % Threshold’ does not have an effect for MRM data
Problem:
Method settings should apply to the spectrum extraction no matter what scan type it is.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Calibration Curve Properties window checkboxes for ISTD/QC/CC points do not affect display
Problem:
‘Show ISTD responses’, ‘Show QC points’, and ‘Show CC points’ do not appear to be linked with the Calibration Curve toolbar icons (Show/Hide ISTD Responses, Show/Hide QC, Show/Hide CC). Unchecking these boxes does not change the calibration curve display as expected (i.e., remove the ISTD points, etc.)
Temporary Solution:
Use controls in the calibration curve display to show / hide data
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
‘Show indicator’ checkbox in the Calibration Curve Properties window has no effect
Problem:
Checking or unchecking the ‘Show indicator’ box does not affect the display of the downward-pointing vertical arrow in the Calibration Curve window if a calibration sample is selected in the batch table. The box always remains checked even after unchecking and clicking Apply or OK. Sample’s ‘Data Path’ column and calibration ‘Cal Path’ column must match in order for indicators to display. If batch was moved from its original location, the sample data path stays as the original path, so it no longer matches the Cal path.
Temporary Solution:
n/a
Fix Information:
Will not be fixed. No display of sample indicator is expected if the batch and data files were moved from the original location where the batch was created.
Keyword:
One-line Description:
Batch status is incorrectly displayed as 'Modified'
Problem:
If a 'Sample' sample type is manually integrated, the sample concentration is automatically updated. So the status bar displays 'Modified' when it should display 'Processed'. Issue exists in Quant B.05.01.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Analyze Batch required after changing sample type for Quantitation Message to update
Problem:
Steps to reproduce: 1) Open batch --> Blank sample displays Quantitation message 2) Change sample type from 'Blank' to 'DoubleBlank' --> Quantitation message still displayed 3) Analyze Batch --> Quantitation message is not displayed
Temporary Solution:
Perform 'Analyze Batch' after changing sample types
Fix Information:
n/a
Keyword:
One-line Description:
Audit trail for target compound UserDefined column displays a 'previous value' when 'previous value' was null
Problem:
When entering a value into the target compound UserDefined column when that column was previously null, the audit trail displays the previous value as the just entered value rather than null.
Temporary Solution:
None
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Audit trail for 'Add Calibration' level displays a 'previous value' when 'previous value' was null
Problem:
When Analyze > Add Calibration is used, the previous level name and previous concentration are identified as the level name and concentration that was entered for the new calibration level. The 'previous values' should be null.
Temporary Solution:
None
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
When the TIC is normalized in Sample Information, the normalized graphics are not displayed in the report
Problem:
When using the 'Normalize Each' button in Sample Information, the graphics are normalized in Batch-at-a-Glance. But when generating a report using the exported graphic settings file option, the normalized results are not displayed in the report.
Temporary Solution:
None
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Avg RF RSD value does NOT get updated when a calibration level is removed
Problem:
The Average Response Factor RSD value displayed in the calibration curve window is not updated when a calibration level is removed. Performing 'Analyze Batch' updates the Avg RF RSD value displayed in the batch table, but the value in the calibration curve window is not updated.
Temporary Solution:
Refer to the 'Avg. RF RSD' value in the compound method section of the batch table
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Default value has been changed for the deconvolution parameter 'SNR threshold'
Problem:
The default value has changed from 2 to 0 for the deconvolution parameter 'SNR threshold'.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
The default row arrangement in Compounds-at-a-Glance has changed
Problem:
The default for ‘Organize Rows by’ is now by ‘Samples’ instead of by 'Compounds'
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
The default for display of Gridlines has been changed to not display any lines
Problem:
This default applies to the Sample Information, Compound Information, Calibration Curve, and Compounds-at-a-Glance windows.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
The sample types that the Ignore Peaks Not Found global applies to has been changed
Problem:
The' Ignore Peaks Not Found' global for Quantitation Message is now applicable to samples of type Sample, Blank, MatrixBlank, and DoubleBlank.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Relative Response is off by 100 when using New Method from File with a library file
Problem:
When creating a new method from a library file, the Relative Response is one hundred times less than expected.
Temporary Solution:
Multiply the Relative Response value by 100 and enter this value in the method as the Relative Response
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
'CC Retention Time' and 'CC Time' show as outliers enabled by default when no outlier limits are set in the method
Problem:
Steps: 1) Open Outliers window 2) Click 'Select All' --> Boxes for 'CC Retention Time' and 'CC Time' are checked 3) Close the Outliers window. 4) Enter method edit 5) CC Maximum Elapsed Time In Hours is 0.00. 'Retention Time Window CC' for CC Retention Time outlier is null.
Temporary Solution:
None
Fix Information:
'CC Time' outlier fixed in B.06.00
Keyword:
One-line Description:
Method Export Table is not complete
Problem:
From the method Editor perform an "Export table" function. In some cases the export is incomplete, typically target and of qualifier information is missing.
Temporary Solution:
Click on the Sample row before performing the export.
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Compounds-at-a-Glance crashes when Fit To Peak option is selected after loading a saved Compounds-at-a-Glance configuration
Problem:
Compounds-at-a-Glance crashes when Fit To Peak option is selected after loading a saved Compounds-at-a-Glance configuration
Temporary Solution:
None
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
The Advanced Print option only retains one report template
Problem:
The Advanced Print option allows for more than one report template to be entered however only the top most report in the list is remembered by the software.
Temporary Solution:
None
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Qualifier peak labels are displayed for all integrated peaks in the qualifier window
Problem:
Only the peak for which results are displayed in the batch table should have the label displayed Seen in Quant B.05.00 SP2 , Quant B.05.01
Temporary Solution:
None
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
QC LCS outlier displayed in the batch table does not automatically update
Problem:
After changing a value in the 'Info' column (Sample section of the batch table), Analyze Batch or Quantitate Batch does not update the QC LCS outlier display in the batch table. The batch status remains 'Processed' even though the batch table display is not in sync with 'Info' column changes.
Temporary Solution:
After changing a value in the 'Info' column: 1) Enter method edit 2) Exit method edit applying the method 3) Analyze or quantitate batch The QC LCS outlier displayed in the batch table is updated.
Fix Information:
n/a
Keyword:
One-line Description:
Batch status is incorrectly displayed as 'Processed' after dilution is changed when using Add-In
Problem:
When using the 'TPH ISTD/Surrogate Subtraction' Add-In, the modified concentration values displayed in the Custom Calc column of the batch table do not automatically update when dilution values are changed and the batch status does not change from 'Processed' to 'Modified'.
Temporary Solution:
n/a
Fix Information:
Add-Ins do not allow automatic updating of batch table values. If columns are changed, 'TPH ISTD/Surrogate Subtraction' needs to be clicked again.
Keyword:
One-line Description:
New Method using Manual Setup has no sample when you reach method edit
Problem:
When 'New Method using Manual Setup' is selected from the batch table view, no TIC is displayed in method edit. In this case, a sample cannot be used as a guide when creating a method manually.
Temporary Solution:
Select 'New Method using Manual Setup' in method edit: 1) Select sample in the batch table 2) Enter method edit 3) Select Method --> New --> New Method using Manual Setup Sample Information window displays trace
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Creating a reference library from a sample results in Library Match Scores <100 for compounds in the sample used to create the library
Problem:
Creating a reference library from a sample results in Library Match Scores <100 for compounds in the sample used to create the library. Issue observed in B.05.01 and B.05.02
Temporary Solution:
None
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Audio is not understandable in three Quant Method Assistant videos
Problem:
For the ‘Excel walkthru1 – approach’ and 'Excel walkthru2' videos, the sound is static (no voice). ‘Dynamic Quant method creation in Excel 2007-Vista - Wizard’ has some audio problems too.
Temporary Solution:
None
Fix Information:
Fixed in B.06.00 Dynamic Quant method creation in Excel 2007-Vista - Wizard’ won't be changed. Audio fades out at end of video but otherwise is understandable.
Keyword:
One-line Description:
The default has changed from 'false' to 'true' for the "noPenaltyForLowPeakCount” parameter in “RefLibraryMatchParams.xml”
Problem:
Certain workflows (e.g., tMRM) want to avoid the penalty for missing peaks. When set to 'false', the matching algorithm applies penalties when less than five peaks are present: 75% (1 peak) 88% (2 peaks) 94% (3 peaks) 97% (4 peaks) This parameter is found in the configuration file at Program Files\Agilent\MassHunter\Workstation\Quant\bin\RefLibraryMatchParams.xml.
Temporary Solution:
n/a
Fix Information:
Changed in B.06.00
Keyword:
One-line Description:
When 'second order ln' is used for calibration curve type, lower calibration curve points give unexpected concentrations
Problem:
Concentrations are much greater than expected.
Temporary Solution:
None
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Change the default for Time Segment boundaries to 'Not Displayed'
Problem:
Turning off the time segment boundaries by default makes the MRM or SIM acquisition method less complicated
Temporary Solution:
n/a
Fix Information:
Changed in B.06.00
Keyword:
One-line Description:
Abundance column is not sorted as expected when the column header is clicked in the Library Editor Spectrum window
Problem:
Also, no triangles are displayed in the column header to indicate that the column is sorted and in what order the sorting is being done.
Temporary Solution:
None
Fix Information:
Fixed In B.06.00
Keyword:
One-line Description:
Chromatogram or spectrum labels overlap so displayed information can be unreadable
Problem:
Observed in B.05.02 1) Chromatogram window: Qualifier ratio overlaps retention time. 2) Spectrum window: Lib Match Score or Mass Match Score overlaps mass
Temporary Solution:
Current workaround is to unzoom the vertical axis or widen the window.
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Custom Calc values remain in the batch table after Analyze Batch is done when using Add-In
Problem:
After using the 'Quantitate with Blank Subtraction' Add-In, clicking 'Analyze Batch' still displays the blank-subtracted results in the Custom Calc column. Need to either 1) Clear Results before doing Analyze Batch, or 2) Enter method edit, apply the method, then Analyze Batch to clear the blank-subtracted values from the Custom Calc column
Temporary Solution:
n/a
Fix Information:
When using the Add-In, need to select 'Analyze' -->' Clear Results' to clear batch table results before analyzing or quantitating the batch without the Add-In action.
Keyword:
One-line Description:
Qualifier response ratio acceptable range in the Compound Information window is always displayed as 'relative' even when 'absolute' is selected for Uncertainty
Problem:
When 'Absolute' is selected for Uncertainty, the 'Relative' qualifier ratio acceptable range is displayed in the Compound Information window
Temporary Solution:
Display the 'Qual Min' and 'Qual Max' columns in the batch table (Qualifier Method section) to see the correct qualifier ratio response range when 'Absolute' is selected
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Cannot uninstall MassHunter Quantitative Analysis Reporting B.05.00
Problem:
Message displayed when attempting to uninstall reporting:
Error 1721. There is a problem with this Windows Installer package. A program required for this install to complete could not be run. Contact your support personnel or package vendor. Action: UninstallIQTReference, location, command rm -p:"MassHunter Quantitative Analysis B.05.00 Reporting"
-f:"C:\Users\admin\AppData\Local\Temp\{E07D6730-08E7-4977-BA85-ACEBC76580C7B\MassHunter_Quant_Reporting_B.05.00.xml"
Reporting does not uninstall.
Temporary Solution:
Steps: 1) Uninstall Agilent Installation Qualification Tool A.04.xx.xxx 2) Install Agilent Installation Qualification Tool A.03.03.009. If this IQT version is not available, re-install MassHunter Quantitative Analysis Reporting B.05.00 3) Uninstall MassHunter Quantitative Analysis Reporting B.05.00
Fix Information:
Fixed in B.06.00
Keyword:
One-line Description:
Warning messages displayed during uninstallation of MassHunter Quantitative Analysis Software
Problem:
'Warning 1910' messages say that shortcuts cannot be removed. After clicking 'OK' to the messages, the desktop shortcuts appear to be removed. Right-clicking in the desktop and selecting 'Refresh' displays the shortcuts that do not work.
Temporary Solution:
Click 'OK' when the messages are observed to close the messages. Right-click on the desktop shortcuts and select 'Delete'.
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Unknowns query results not updated after clicking 'Refresh' or re-running the query
Problem:
After making changes to the Components table, clicking 'Refresh' in the Query Results window does not update the results displayed. Re-running the query still does not update the results displayed. Seen in 'Batch Non-Target Hits containing Chlorine in Formula.sql.qry'
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Library Editor - ChemSpider mol files display with jagged lines
Problem:
Many customers are using chemical structure (mol) files from the ChemSpider server however they display with jagged bond lines in MassHunter software (Library Editor, Qual) while displaying normally in applications such as ChemDraw.
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
All outlier flags are not cleared from the batch table when 'Clear Results' is done
Problem:
'Clear Results' is done but outlier flags (e.g., blue background) are still displayed in the batch table
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
The batch data path is truncated in multiple report templates
Problem:
This issue has been observed in the following templates but may also exist in other templates that are not listed: Path: D:\MassHunter\Report Templates\Quant\ja\A4\ISTD\Layout_Graphics QuantReport_ISTD_Condensed_B_06_00_Ja.xltx QuantReport_ISTD_CondensedQualifier_B_06_00_Ja.xltx Path: D:\MassHunter\Report Templates\Quant\ja\A4\ISTD\Parts_Graphics QuantReport_ISTD_Complete_B_06_00_Ja.xltx QuantReport_ISTD_Samples_B_06_00_Ja.xltx Path: D:\MassHunter\Report Templates\Quant\ja\A4\ESTD\Results_NoGraphics QuantReport_ESTD_ResultsByCompound_B_06_00_Ja.xltx QuantReport_ESTD_ResultsComplete_B_06_00_Ja.xltx QuantReport_ESTD_ResultsSummary_B_06_00_Ja.xltx Path: D:\MassHunter\Report Templates\Quant\ja\A4\Outliers\Calibration QuantReport_Outlier_AverageResponseFactor_B_06_00_Ja.xltx QuantReport_Outlier_AvgResponseFactorRSD_B_06_00_Ja.xltx QuantReport_Outlier_CalibrationRange_B_06_00_Ja.xltx QuantReport_Outlier_CurveFit_B_06_00_Ja.xltx QuantReport_Outlier_RelativeResponseFactor_B_06_00_Ja.xltx QuantReport_Outlier_ResponseFactor_B_06_00_Ja.xltx
Temporary Solution:
Possible workarounds: 1. Change the data location so the batch path is shorter (Share drive locations usually result in longer path names) 2. Use a smaller font for the batch path so the path name fits within the page width (i.e., modify the template) 3. Truncate the path and only use the batch file name in the report (Customers need to contact Agilent for this workaround)
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
DrugQuant flavor: Outlier dialogs show hidden outlier and standard outlier names
Problem:
The Outliers window (displayed when 'Select Outliers' is clicked) includes ALL outliers that exist in MassHunter Quant. But in the Drug Analysis configuration, many of these outliers are not visible in the Outliers section of the Method Edit view and thus aren't accessible by the user when running Drug mode.
Temporary Solution:
n/a
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
When Excel is used while generating a report, an error message is displayed and the report will not be generated
Problem:
Using two instances of Excel, one for Quant reporting and another for something else, is not supported
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Quant Reporting does not automatically enable the 'MassHunter Reporting Quant' Add-In in Excel as part of the installation
Problem:
After installing MH Quant Reporting, the 'MassHunter Reporting Quant' Add-In box in Excel is not checked. Also, the box does not stay checked (e.g., Check the 'MassHunter Reporting Quant' Add-In box, click 'OK' to close the Add-In window, close Excel, then open Excel results in an unchecked 'MassHunter Reporting Quant' Add-In box).
Temporary Solution:
Reports can be generated successfully when the 'MassHunter Reporting Quant' Add-In box is not checked
Fix Information:
n/a
Keyword:
One-line Description:
Quant_B.04.00_Reporting_SP01.exe displays wrong version in the Windows Add or Remove Program display
Problem:
When Quant_B.04.00_Reporting_SP01.exe is installed on a Windows XP or Windows Vista system it displays the wrong version in the Windows “Add or Remove Program” display. It should display a version of 4.0.255.1 but instead displays a version of 5.0.225.1. The version displayed in the ACE compliance report of B.04.01 is correct. Quant_B.04.00_Reporting_SP01.exe was distributed on media with PN G3336-60049 to enable Windows XP and Windows Vista based systems with Excel 2010 to generate reports from MassHunter software.
Temporary Solution:
None.
Fix Information:
n/a
Keyword:
One-line Description:
Stepping through compounds in the batch table when the method has >400 compounds is slow towards the end
Problem:
When a quantitation method has many compounds ( approximately 450), stepping through the compounds gets slower towards the end.
Temporary Solution:
Closing then re-opening MassHunter Quantitative Analysis in the middle makes it faster again
Fix Information:
Fixed in B.06.00 SP1
Keyword:
One-line Description:
For samples on which Zero Peak has been done, values greater than zero display for concentrations when Analyze Batch is done
Problem:
Initially, performing Zero Peak causes the concentration for the sample to display as zero But after performing Analyze Batch, small non-zero amounts are displayed in the concentration fields. Zero peaked compounds should always report zero concentration as it is meant to be an indication of the peak not being present.
Temporary Solution:
None
Fix Information:
Fixed in B.06.00 SP1
Keyword:
One-line Description:
When data is on a share drive and replicate calibration levels exist in the batch, disabling a level in method edit does not give the expected results in the calibration curve window
Problem:
When one level is disabled, all replicates of that level are disabled in the calibration curve window. Also, accessing the data from a different computer will show all levels enabled or disabled when only one level was disabled.
Temporary Solution:
None
Fix Information:
Fixed in B.06.00 SP1
Keyword:
One-line Description:
Accuracy outlier flag is not displayed when a batch is opened on a different computer
Problem:
Accuracy outlier flags are displayed when the batch is saved then closed on PC1. When the batch is opened on PC2 (PC1 drive is mapped on PC2), the Accuracy outlier flags are not displayed unless 'Analyze Batch' is done. Appears to be fixed in Quant B.05.02 and Quant B.06.00 (Quant B.05.01 not tried)
Temporary Solution:
Perform 'Analyze Batch' to display the Accuracy outlier flags
Fix Information:
Fixed in B.05.02
Keyword:
One-line Description:
When using ‘Update Retention Times’, only “Greatest Response” or “Closest RT” criteria are supported. "Close RT" and "Close RT with Qualifiers" will instead have the closest RT being used
Problem:
In the method Editor, the function Method -> Update Retention Times only supports "Greatest Response" or "Closest RT". If “Close RT” or “Close RT with Qualifiers” is selected for Criteria, the closest RT will be used.
Temporary Solution:
None
Fix Information:
Fixed in B.06.00 SP1
Keyword:
One-line Description:
Video in Outlier -> Peak Results -> Purity (Advanced) lacks critical steps in B.06.00
Problem:
The video "Outliers - Peak Purity - Advanced.wmv" is now obsolete and incomplete due to code changes in B.06.00 change involving targeted deconvolution. It does not include all of the necessary details to setup the outlier correctly. The video give only two setup steps and leaves out adding a reference library and selecting "deconvoluted scan" in Spectrum Extraction Override.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Setup_Supplemental.exe on the B.06.00 SP1 DVD does not work
Problem:
Messages are displayed during installation: "Error 1308: Source file not found ..." The SP1 installation is looking for supplemental files that are on the B.06.00 DVD but got changed on the B.06.00 SP1 DVD.
Temporary Solution:
There are two possible workarounds: (1) Use the Setup_Supplemental.exe on the B.06.00 DVD, or (2) Manually copy the desired Data, Manuals, Videos, and Report Templates folders from the Supplemental folder of the B.06.00 SP1 DVD to the appropriate location. Default locations are as follows: MassHunter\Data\QuantExamples MassHunter\Manuals\Quant MassHunter\Videos\Quant MassHunter\Report Templates\Quant\Index
Fix Information:
n/a
Keyword:
One-line Description:
Library Match Score and Purity values may differ between MassHunter Quantitative Analysis B.06.00 and B.06.00 Service Pack 1
Problem:
The difference is due to a difference in the window size factor values used for deconvolution. In B.06.00 SP1, low-resolution window size factors of 200, 100, 50, and 25 are used every time and not just in the case of duplicate peaks, whereas in build B.06.00, low-resolution window size factors of 100, 50, and 25 are used.
Temporary Solution:
None
Fix Information:
Possibly a partial solution to the saturation problem in Targeted Deconvolution
Keyword:
One-line Description:
Editing a report method (EditReportMethod) is not recorded in the audit trail
Problem:
Steps to reproduce: 1) Select Report --> Generate 2) Click 'Edit' 3) Remove the existing template. Add a different template. Save the method 4) Close the Report Method Edit window 5) Click 'OK' to generate the report Only 'Generates report' is recorded in the audit trail
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Crash when 'Create Levels from Calibration Samples' and a null level exists
Problem:
Problem seen when: 1) Create new batch 2) Add samples 3) Change one of the 'Sample' types to 'Cal' 4) Type a value in the Level cell for the sample that was changed to a 'Cal' but do NOT press 'Return' or 'Enter' instead go directly to Select Method --> New --> New method from acquired scan data. 5) Select a sample path, then click 'OK' 6) go to the menu and select Method --> Create Levels from Calibration Samples 7) Crash Problem is the null (empty) level caused by grid not saving data until pressing 'Return' or 'Enter' or clicking on another cell. Crash is not specific to the way the method is created, New method from acquired scan data was only given as an example.
Temporary Solution:
Make sure data is entered for all Samples of type "Cal' and that the date is accepted by either pressing 'Return' or 'Enter' or clicking on another cell.
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
On a domain user, unable to generate an audit trail report in Quantitative Analysis B.06.00 SP1 with compliance enabled
Problem:
Unable to generate an Audit trail report from Quantitative Analysis( b06.00 sp1) When the user logins as the local administrator on the workstation, a Quant reports is generated with no issues. When logged in as a domain user, however, the AUDIT Quant report fails to generate, ( all other work) and the process just hangs. No error is thrown.
Temporary Solution:
Workaround: create a new batch ( add the same samples and use the same method) and generate the audit trail report.
Fix Information:
n/a
Keyword:
One-line Description:
MRM data of zero intensity displayed when MRM signal of same precursor exists
Problem:
Quant mistakenly extracts a zero intensity signal when there is a MRM signal with matching precursor but a different that specified product ion. This has no effect in the time segment or time region in which the desired MRM signal is found because the zero intensity is added to the correct signal and the signal is extracted correctly. The problem is seen when the MRM signal with matching precursor but a different that specified product ion appears in a region without the desired MRM. Here Quant draws a flat baseline with zero intensity. This in turn can cause the integrators to make mistakes. This behavior was seen B.05.01 thru B.07.00 only.
Temporary Solution:
Take extra care to set RT left and RT right window so only the desired signals are extracted.
Fix Information:
Fixed in B.07.01 There is the possibility to see different integration behavior, including but not limited to peak area, noise, and signal to noise, between software versions which do or do not display and integrate the zero intensity signal. The versions that do not display and integrate the zero intensity signals are considered to be the correct interpretation. .
Keyword:
One-line Description:
Error message displayed when attempting to uninstall Quant
Problem:
The error message “Error opening installation log file. Verify that the specified location exists and is writable.” is displayed when attempting to uninstall Quant. Microsoft has confirmed this to be a problem in the operating systems including Windows 7 Professional. This can occur if Windows Installer Logging is enabled, but the Windows Installer engine cannot properly write the uninstallation log file. This condition can occur if the Windows Installer's application heap becomes freed and thus loses the information on where to store the log file. When this occurs, Windows Installer attempts to write to the location 'C:\Windows\System32' and addresses it as a file. Proper behavior would be to write to the following location and file name: 'C:\Users\\AppData\Local\Temp\MSIxxxxxx.log'.
Temporary Solution:
To work around this issue, stop and restart the Explorer.exe process using Task Manager.
Fix Information:
Dependent on Microsoft
Keyword:
One-line Description:
DrugQuant Sample types incorrect for pdf reporting
Problem:
The non DrugQuant sample types are being used: Blank = double blank negative = blank etc
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Quant method in .quantmethod.xml as well as the method file in the .M folder lacks checksum
Problem:
Quant method in .quantmethod.xml as well as the method file in the .M folder lacks checksum and can be edited by notepad or other applications without detection when Quant is run in compliance mode. These changes are not picked up by the “Check Batch Files” tool, either.
Temporary Solution:
Place method on a write protected disk.
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Limited number of calibration points in large batch
Problem:
Customer reports that in B.06.00 an error takes place indicating a limit of 200 calibration points. Customer indicated that this was not the case in previous versions.
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Option to “Delete graphics files after report generation” is missing
Problem:
The feature to delete graphic files after reports are generated was present in Quant releases before B.06.00 but missing in B.06.00.
Temporary Solution:
The workaround in this is using a bat file (see text below) as post process. This bat file deletes *.emf files and graphics folders from report output directory. Please modify the report method to use the bat file as Post Process --- Bat file --- : DeleteGraphicsFile.baterase /s /q %1\*.emf rmdir %1\PeakChromatogram rmdir %1\PeakQualifierChromatogram rmdir %1\PeakQualifiers rmdir %1\PeakSpectrum rmdir %1\SampleChromatogram rmdir %1\TargetCompoundCalibration t
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Change to LCS outlier recovery calculation
Problem:
Prior to B.07.00 the recovery was calculated against the Final Concentration. Starting in B.07.00 the recovery is based on Calculated Concentration.
Temporary Solution:
None
Fix Information:
Changed in B.07.00
Keyword:
One-line Description:
Manual vs. Automatic integration gives extreme change in noise and S/N
Problem:
When a very small change in integration is made via manual integration the noise and S/N values change far too much. When manual vs. automatic peaks with almost identical peak height and start/end points are compared noise increases by 3 to 5 fold, and thus a related drop in S/N, is seen. Seen in B.06.00 and B.06.00 SP1. .
Temporary Solution:
For manual integration, the noise region is derived from peak boundaries and chromatogram. When using automatic integration, the noise region is derived from baseline information. If manual noise regions are designated in the method then the noise values are more similar.
Fix Information:
Help updated to include explanation of behavior.
Keyword:
One-line Description:
Changing the "Titles" in the Peak Chromatogram Graphics section of the report generator does not affect the final print out
Problem:
-Generate report -Edit method -Graphics Setting tab -Peak Chromatograms: Graphics Settings -Titles -Custom title -> then clear everything. -Start printing The default titles are still in the final report.
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Selecting the first row in Compound Table does not change the compound displayed in Current Compound
Problem:
Sometimes selecting the first row in Compound Table does not change the compound displayed in Current Compound. Observed in B.06.00
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Cannot open Quant after user configuration/layout file is corrupted
Problem:
Cause of corrupt file is not known. Quant crashes when opened. Uninstalling then re-installing Quant does not stop the crash. Seen in B.05.00
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Deconvolution fails on "New Method from Acquired Scan Data with Library Search" on a set of GC TQ data files collected in profile product ion mode
Problem:
Quant B.06.00 does not create compounds when running "New Method from Acquired Scan Data with Library Search" on a set of GC TQ data files collected in profile product ion mode. All version of Quant prior to B.06.00 (B.05.02 was used to confirm previous operation) can process the data and create compounds. This workflow is used by the TQ community during method development.
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Deconvolution Parameters displays library score tab
Problem:
When running “New method from acquired scan data” click on the “Deconvulsion Parameters” button. The “Library RT Match” Tab is visible but shouldn’t be. Seen in B.06.00.
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Report folder path does not automatically advance – reporting fails
Problem:
Load a batch Select Report/Generate select a report folder path and a report method and click OK Repeat the last two steps. The second time you cue reports, the path should automatically increment to create a new folder but it does not.
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Report generation is not recorded in the audit trail unless the batch is saved after report generation is performed
Problem:
Printing report could be done many times without record of the printout in the audit trail. Printouts are used in a hybrid system. Seen in B.06.00
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
'Specified cast is not valid' message is displayed when using Tools --> Actions --> Go to Qualitative Application on Accurate Mass data
Problem:
Happens only when Tools --> Actions --> Go to Qualitative Application is done in Method Edit using some examples of Accurate Mass data. Observed with Quant B.06.00 SP1 and Qual build 6.0.633.0
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Zoom in Spectrum extracted from TIC can expose region below baseline
Problem:
Steps to reproduce 1) got to method editor 2) in the Sample Information window Extract a spectrum 3) In the spectrum window draw a zoom box to the baseline 4) the zoom works but exposes region below baseline
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Avg RF RSD is 'Infinity' or blank when standard addition is checked
Problem:
Avg RF RSD in the Batch table is displayed as 'Infinity' or blank when standard addition is checked. Seen In B.06.00 SP1
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Display Extract left(right) mz with two digits to the right of decimal point - SIM and MRM window showing as 0.1 when it is in fact 0.05
Problem:
See Advanced Tasks -> Mass Extraction Setup. This request is driven by confusion about the SIM and MRM window showing as 0.1 when it is in fact 0.05. Asking only that the default numeric format be altered to F2 for Extract left mz and Extract left(right) mz.
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
.L library from ChemStation gives errors -
Problem:
Customer created the attached library in.L format from ChemStation software. When searched in Qual it gives an error about XY pairs. When opened in Library editor the application has a hard crash. The library contains spectrum MZ values with descending sorted (MZ=150,140,130,120… instead of 80,90,100,120…).
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Cannot print out log-log calibration curve plots in Quant B.06.00
Problem:
Prior versions of Quant could save the graphics settings to for a log-log plot of the cal curve and then use those settings to make the Report display a log-log plot as well. However, with Quant B.06.00 this was no longer possible as this option was not in the Report Method file.
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Mass Match value is displayed when target is zero peaked
Problem:
When peak is zero peaked the mass of the peak that was zeroed is used to calculate a Mass Match value.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Signal-to-Noise (S/N) value is displayed when target is zero peaked
Problem:
When peak is zero peaked the peak that was zeroed is used to calculate a Signal-to-Noise (S/N) value.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Sample with Zero Peaked target compound is included as Response = 0 when Average RF RSD outlier is calculated
Problem:
Sample with Zero Peaked target compound is included as Response = 0 when Average RF RSD outlier is calculated.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Sample with Zero Peaked ISTD is included as Response=0 when ISTD Response Percent Deviation outlier is calculated
Problem:
Sample with Zero Peaked ISTD is included as Response=0 when ISTD Response Percent Deviation outlier is calculated
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Sample with Zero Peaked target compound is included as RF=0 when Average RF outlier is calculated
Problem:
Sample with Zero Peaked target compound is included as RF=0 when Average RF outlier is calculated
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Sample with Zero Peaked target compound is included when Curve Fit R2 outlier is calculated
Problem:
Sample with Zero Peaked target compound is included when Curve Fit R2 outlier is calculated
Temporary Solution:
Exclude data point from curve
Fix Information:
n/a
Keyword:
One-line Description:
Zero peak and integration metric/metric flag
Problem:
After zero peaking, the integration metric outlier is no longer flagged but (1) Qualifier Int Metric and Int Metric Flag values remain the same as before the zero peak was done, and (2) Int Metric and Int Metric Flag values change to ‘Accepted’ and ‘NoProblem’.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Unidentified option displayed in the Column Format, Format pattern pulldown menu
Problem:
Right-clicking then selecting 'Format Column' in a numerical batch table column shows a blank option in the pulldown menu. A crash results if the blank option is selected then Format Column is selected by right-clicking. Seen in B.06.00 SP1
Temporary Solution:
None
Fix Information:
Fixed in B.07.00.
Keyword:
One-line Description:
Unknowns Analysis: using the ‘closest ISTD’ to estimate RF results in ‘infinity’ for the Estimated Concentration
Problem:
Unknowns Analysis: using the ‘closest ISTD’ to estimate RF results in ‘infinity’ for the Estimated Concentration
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Avg RF RSD for the ISTD for the first sample in the batch is incorrect
Problem:
Analyze Batch, Quantitate Batch, or enter method edit/apply method/analyze batch do not change the value Seen in B.06.00 SP1.
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
After manual integration, analyzing the batch does NOT always update concentrations
Problem:
Problem seen when all samples of a given compound are manually integrated. Issue seen in Quant B.06.00 SP1 (388.1), Quant B.05.00 (291.0).
Temporary Solution:
(1) Enter method edit, apply the method, and analyze the batch, or (2) Clear results then analyze the batch. Both methods will update the concentrations.
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Library search with screening on misses compounds with very abundant ions at low masses
Problem:
When a library search with screening checked is performed the screening algorithm can miss compounds which have abundant ions below the scan range. This is mostly seen in compounds of low molecular weight.
Temporary Solution:
Turn off screening.
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Zero peak of ISTD causes curve calculation error
Problem:
To reproduce 1) load DOA method 2) go to a calibrator sample (L1 was used) 2) zero peak the amp ISTD target peak for AMP 3) Analyze batch Note calibration curve has infinite slope and NaN for intercept (and no line).
Temporary Solution:
Disable data point in Calibration curve
Fix Information:
n/a
Keyword:
One-line Description:
Selected manual integration are not accepted until user configuration files are deleted
Problem:
When attempting manual integration Quant would not accept what was selected and it would snap to a strange integration Issue reproducible until the user configuration files were deleted, then results are as expected.
Temporary Solution:
1) Close Quant 2) Delete user configuration files located in the folder located at \Users\\AppData\Roaming\Agilent_Technologies,_Inc\QuantAnalysis.exe 3) Restart Quant 4) Perform manual integration Seen in B.06.00
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Target compounds using Isotopic Dilution give incorrect results with and without manual integration.
Problem:
If a target compound has Isotopic Dilution parameters Rx and Ry the resulting calculation of the are incorrect. The following results were not correct, the “ISTD concentration Ratio” as well the concentrations (calculated and final). If user executing a manual integration the calculation difference appeared to be more obvious. Seen in B.06.00
Temporary Solution:
1) Remove the Isotopic Dilution values of Rx and Ry from target compounds 2) Analyze the batch and review data and perform all required manual integrations - remember areas and other values are uncorrected by Isotopic Dilution 3) Return to the method editor and restore Rx and Ry setting for all target compounds 4) Analyze the batch - Results are corrected by Isotopic Dilution
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Navigation not always synchronized between BAG and CAG
Problem:
When 'Synchronize Navigation' and a Review Mode other than 'None' is selected in CAG, navigation is not always synchronized between BAG and CAG
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
'Create Levels from Cal Samples' does not define the correct level values for one compound
Problem:
In certain cases the 'Create Levels from Cal Samples' does not define the correct level values.
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
Warning messages are displayed during installation of Quant B.07.00 when Quant B.06.00 SP1 is not uninstalled first
Problem:
When installing B.07.00, B.06.00 SP1 uninstallation is launched. B.06.00 SP1 uninstallation has a warning message issue: "Warning 1910. Could not remove shortcut Unknowns Analysis.lnk. Verify that the shortcut file exists and that you can access it" "Warning 1910. Could not remove shortcut Library Editor.lnk. Verify that the shortcut file exists and that you can access it"
Temporary Solution:
Uninstall Quant B.06.00 SP1 before installing Quant B.07.00
Fix Information:
(None)
Keyword:
One-line Description:
'Quantitate Sample' does not calculate Matrix Spike Percent Recovery
Problem:
'Quantitate Sample' (Hotkey F7) does not calculate Matrix Spike Percent Recovery.
Temporary Solution:
Use "Quantitate Batch" (Hotkey F6)
Fix Information:
n/a
Keyword:
One-line Description:
Create New Method from Acquired Scan Data using an All Ions data file does not display collision energy
Problem:
When an All Ions data file is used with the "Create New Method from Acquired Scan Data" feature, the collision energy field is blank rather than displaying the collision energy in the data file.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Crash when running three UA queries
Problem:
Crash obtained when attempting to run the following UA queries: Batch Non-Target Hits containing Chlorine in Formula.sql.qry Batch Non-Target Hits custom filter.sql.qry Batch Quant Target Compound List.sql.qry Seen in B.07.00
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
In method edit, Edit --> Find does not give expected results when using ‘Does not contain’
Problem:
In method edit, Edit --> Find does not give expected results when using ‘Does not contain’
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Tooltips for 'Baseline Start' and 'Baseline End' display incorrect units
Problem:
Baseline End and Baseline Start tooltips show units as minutes (min.), but they are Y-values (counts) Seen in B.06.00 SP1 & B.07.00
Temporary Solution:
n/a
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Unknown Analysis - Crash after sorting one column then selecting 'Refresh' in queries
Problem:
Steps to reproduce: 1) Open 'Target and Non-Target Analysis.uaf' (MassHunter\Data\QuantExamples\MS\RI-PEST-MATRIX from Supplemental installation) 2) Select Report > Query. Select any of the first three queries (Batch Non-Target ..., Batch Quant Target ...). Click Open 3) In the Query window, sort any one column by clicking the column header 4) Right-click in the query window and select 'Refresh' --> Crash
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Unknowns Analysis - Print Preview of Ion Peaks window is cut off on the right side when selecting 'Fit to Sheet Width' and 'Fit to Full Sheet'
Problem:
Unknowns Analysis - Print Preview of Ion Peaks window is cut off on the right side when selecting 'Fit to Sheet Width' and 'Fit to Full Sheet'
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Fewer than the specified number of qualifier ions created during method creation
Problem:
Some automated method creation steps have an option to create a specified number of qualifier ions. However new logic introduced in B.07.00 to select optimal ions based on mass, S/N, etc. may not find the specified number of ions with an acceptable quality score. In such a case fewer than the specified number of ions may be created including the possibility that no qualifiers are created.
Temporary Solution:
There is an option to force the software to create at least one qualifier regardless of the ion quality scores found. Locate the file IonSelectionParameters.xml found in the \Program Files\Agilent\MassHunter\Quant\bin folder and edit it. Change the value of theparameter from “false” to “true”. Restart Quant so that the new parameter will be used.
Fix Information:
n/a
Keyword:
One-line Description:
Quant displays incorrect Reference Library source after overwriting existing Reference Library file
Problem:
Steps to reproduce 1) Open VOA batch 2) select a cal file 3) enter method editor, Create Reference library based on a cal file 3) Exit, apply, analyze 4) name of Reference file displayed in spectra 5) enter method editor, Create Reference library based on NIST11 by overwriting existing library (i.e., Leave 'Create reference library at' as it was for the first Step 3) 6) Exit, apply, analyze 7) name in spectra is still cal file, not NIST11
Temporary Solution:
Use different file name rather than overwrite existing file.
Fix Information:
n/a
Keyword:
One-line Description:
Universal Integrator Timed Event clears Value field when a Time is entered
Problem:
To reproduce this issue: 1- go to integrator dialog box. 2 -in integrator tab, select Universal for integrator. 3 -go to Universal tab, in time event, select peak width or threshold which can make value box active in event droplist. 4 -input one number such as 5 for value, then input time. The value box becomes blank when a time is entered.
Temporary Solution:
To avoid this issue, enter the time first, then the value, then click Enter.
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Analyze Batch is slow
Problem:
Batch for LC-QQQ data takes four to five minutes when analyzed
Temporary Solution:
n/a
Fix Information:
Fixed in Quant B.07.00. The batch that took four to five minutes for analysis in Quant B.06.00 SP1 takes less than one minute for analysis in Quant B.07.00.
Keyword:
One-line Description:
Exported Table to Excel Prints in letter even when A4 format has been selected
Problem:
To see the issue 1) Use the menu item Select File > Page Setup, and set or confirm A4 size is selected 2) Export the Batch table in excel format by using File > Export > Export Table, then generate export table in Excel format. 3) The generated excel file has letter format not A4.
Temporary Solution:
None
Fix Information:
Fixed in B.07.00
Keyword:
One-line Description:
'Create Levels from Calibration Samples' creates levels for QC and CC types
Problem:
'Create Levels from Calibration Samples' creates levels for QC and CC types This behavior is different from B.06.00 SP1 where levels are only created for 'Cal' types.
Temporary Solution:
Enter the required information (Conc.) for the rows created even if it is a duplicate of the level seen in Calibrators.
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Value remains displayed for 'Avg. RF RSD' after 'Clear Results' is done
Problem:
If the 'Avg. RF RSD' is displayed in the Batch table and a 'Clear Results' command is done the 'Avg. RF RSD' field is not cleared, the value is still displayed.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
B.07.00 TPH Add-In results are different from B.07.01 implementation of TPH Subtraction in Compound Math
Problem:
The difference is due to the TPH Add-In using the original Responses for the calibration curve, whereas TPH Subtraction of Compound Math uses the adjusted responses. The Add-In had the limitation of not being able to update the calibration curve. Using 'Analyze Batch' updates the calibration curve. Using 'Quantitate Batch' does not update the calibration curve. 'Quantitate batch' should give the same results as the TPH Add-In.
Temporary Solution:
Not Applicable
Fix Information:
n/a
Keyword:
One-line Description:
Peak response using General integrator is different for GC-FID data
Problem:
Observe different peak response when comparing Quant B.07.00 and Quant B.07.01
Temporary Solution:
n/a
Fix Information:
Resolved with improved fidelity for borderline cases involving peak height rounding
Keyword:
One-line Description:
Quantitate Compound only works on the last line selected
Problem:
In Quant, the "F8" function to quantitate a single compound is great because it lets a user update a calibration table and re-calculate the concentration of the one compound that was updated. This is particularly helpful when there are many compounds. However, there is no way to process that one compound for the entire batch.
Temporary Solution:
To process every sample in the batch, you have to press the down key -> F8 -> down key -> F8, etc, etc.
Fix Information:
n/a
Keyword:
One-line Description:
Tools -> Go To Qualitative Application function no longer sends multiple selected data files to Qual
Problem:
Under the Tools -> Go To Qualitative Application function will only send the most recently selected data file from the batch table to Qual. In prior revisions, all selected data files were sent.
Temporary Solution:
Send each data file one at a time.
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Chromatograms extracted on QTOF give zigzag appearance due to ignoring CE
Problem:
Set up Quant method with QTOF target MS/MS data. In this target MS/MS data, the same precursor ion and same product ion with different two collision energy( example 10eV, 20eV). When select single collision energy( 10eV ) and CE delta 1V, EIC turns to zigzags. The extraction is not using CE as a filter thus both signals are extracted.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
TPH add-in does not subtract the surrogate that falls in the RT range
Problem:
In some cases the TPH add-in does not subtract the surrogate that falls in the RT range of certain compounds
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Unable to display TIC chromatogram for Fast Polarity Switching on LCMS QTOF Data
Problem:
For data acquired in fast polarity switching (FPS) mode on a LCMS QTOF, the TIC in Quant displays "No Data Points". If you open the same data file in Qual, it displays without problem, verifying the data is present.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
DrugAnalysis_DopingScreening.report.xml report has error when run on certain data sets
Problem:
DrugAnalysis_DopingScreening.report.xml report gives error "Failed to compare two elements in the array".
Temporary Solution:
An updated version of DrugAnalysis_DopingScreening.report.xml is available.
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Last entry in customer's .L library created by ChemStation export not found
Problem:
If a search is made of a .L library in Quant B.07.00 (or Qual B.06.00 SP1) the last entry is not found although it should. The problem is in the MassHunter library search code not the Library format.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Report Method - Peak Chromatogram Graphics - Peak fill not retained in method
Problem:
To see this defect 1) Report-> Generate, create a Reporting Method, go to the Peak Chromatogram Graphics, note peak fill = 75% 2) set peak fill to "No Fill" 3) apply, exit and save as test_fill.m 4) generate report now , check report, peak graphics are not filled 5) Report-> Generate, Chose, select test_fill.m, go to the Peak Chromatogram Graphics check peak fill Peak fill back to 75% also seen as 1) go to the Peak Chromatogram Graphics, note peak fill = 75% 2) set peak fill to "No Fill" 3) go to the Peak Chromatogram Graphics, note peak fill = 75% (flipped back!) seen in B.07.00.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Validation on AreaCorrectionMZ is incorrect in MRM mode where AreaCorrectionSelectedMZ is different than target precursor
Problem:
Example seen using the MRM to perform area correction. Example data has target compound with transition 200 -> 121 and correction compound 190 -> 121. Method will not validate and gives an error "Area correction MZ must be different than target m/z". This would be a valid issue in MS mode as this would be the same signal but is invalid in MRM when target and correction have different precursor ions thus different signals are extracted.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
General Integrator gives logically inconsistent and incorrect peak heights
Problem:
General integrator is giving inconsistent peak heights in a series of manual integrations. In many cases the peak’s baseline is raised yet peak height goes down.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
The Script UpdateQualifierRatioOutlierLimitsFollowingPointSystemRules.quant is still using the old tolerances.
Problem:
The Sanco Document SANCO/12571/2013 Method Validation and Quality Control Procedures for Pesticide Residues Analysis in Food and Feed have been updated. The Script UpdateQualifierRatioOutlierLimitsFollowingPointSystemRules.quant is still using the old tolerances.
Temporary Solution:
Request new script.
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Signal extracted is not valid near edge of RT window when RT window shifted by Time Reference compound feature
Problem:
When a compound's RT window is shifted by using the Time Reference compound feature the visual display of data near the edge of the chromatogram is inaccurate. This inaccurate region of the EIC can lead to integration error especially for peaks near the edge of the RT windows.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Report: "Display Label Name" option does not work
Problem:
1. Report > Generate 2. Select "Edit" button to Report method 3. "Graphics Setting" tab > Select "Graphics setting" for Peak Chromatogram 4. Select "Peak Label > check "Display Label Name". 5. Save and create a report 6. Label name is not shown in Graphics in Report.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
IQT fails for Quant B.07.00 Reporting SP1
Problem:
After installing the MassHunter Quantitative Analysis Reporting B.07.00 SP1 the IQT (Installation Qualification Tool) will generate a report showing “Fail” when run on the Product “MassHunter Quantitative Analysis B.07.00 Reporting”. See SN "G6845AA-070B:Release of Service Pack 1 (SP1) for MassHunter Quantitative Analysis Reporting B.07.00" to see which specific files and registry entries are invalid. The IQT report for the Product “MassHunter Quantitative Analysis B.07.00” is not affected and will pass.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Report output in TEXT doesn't work using Excel template.
Problem:
If the option to create a text version of a report created by an Excel report template is selected it will fail. The issue is in the Report Add-In not the Quant application.
Temporary Solution:
Install Quant Reporting B.07.00 SP1
Fix Information:
Fixed in Quant Reporting B.07.00 SP1
Keyword:
One-line Description:
PDF QuantDataSet schema is for Unknowns Analysis instead of Quant
Problem:
The PDF version of the QuantDataSet schema, QuantitationDataSet_B_07_00.pdf located in the \Additional\Schema folder on the Quant B.07.00 media, contains information on Unknowns Analysis Data Set not the Quant Data Set. The Word version of the document, QuantitationDataSet_B_07_00.docx located in the same folder is correct and should be used.
Temporary Solution:
Recreate PDF document from the Word document.
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Spectrum window displays unfamiliar path for reference library source
Problem:
This issue will be observed when a batch, containing a reference library, is initially created outside of the ECM storage and then that batch is transferred to ECM. The path to the reference library will not match the expected value. Recreating the reference library after it is migrated to ECM will update the path to the expected value.
Temporary Solution:
Recreate reference library after it is migrated to ECM.
Fix Information:
n/a
Keyword:
One-line Description:
Fixed graphics not working for compound Chromatograms
Problem:
Go to the report method editor, Graphics settings, Fixed Range graphics settings, enter a compound name and some values for min & max for X and Y. Print an excel report and note that the range do not change from the default.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Extracted spectrum disappears unexpectedly while creating new compound and qualifiers
Problem:
When manually creating a quant method using the following steps gives error 1. Double LMC (left mouse click) on TIC to get spectra 2. LMC on target ion to add blue triangle then RMC to get context menu 3. Select ‘New Compound’ then enter compound name in Method table. So far all good 4. Go back to spectra and LMC the qualifier ion. Blue triangle appears 5. RMC. Context menu does not appear. The spectra disappears, leaving a zoomed in TIC with an EIC of the target ion overlaid If you skip entering the compound name and go straight from adding the target to adding the qualifier this does not happen. It does not seem to happen every time but enough to be reproducible.
Temporary Solution:
at the end of step #6 press 'Enter' or 'Tab' or click another cell.
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
QuantReportPostProcess-CopyCSVFile.bat file contains error
Problem:
There is wrong information regarding what argument to use within file QuantReportPostProcess-CopyCSVFile.bat. In this file it says that argument $(ExcelDestinationFilePath) should be used which is wrong. Correct argument is $(DestinationFilePath).
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Create New Method from SIM data gives wrong results
Problem:
When running "New Method from SIM data" on certain data files the resulting method is not correct. Typical issues include compounds being placed at a RT outside of the time segment in the acquisition method. Others errors exist.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Instrument Name is not read into Quant if data was acquired on GCMS Acquisition B.07.01
Problem:
Instrument Name is not read into Quant if data was acquired on GCMS Acquisition B.07.01.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
PDF Reporting: "Injection Volume" label is missing in "Gen_BySample" and "Gen_BySample_withSN" reports.
Problem:
The PDF reports "Gen_BySample" and "Gen_BySample_withSN" display the value for the Injection Volume in the Header region of the report but the label "Injection Volume" is not printed.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Unknowns Analysis - Switch Head-to-Tail on/off in Spectrum Window fails to re-scale Library spectrum
Problem:
In Unknowns Analysis go to the Spectrum window. Switch Header-to-tail to checked. Display switches and looks fine. Switch Header-to-tail to un-checked. Display changes but the ions in the Library spectrum are not re-scaled (just nubs at <1% of display). Auto Scale fails to correct problem. Clicking to alternate compound then back to initial compound required to re-scale.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Library searches does not recognize updates from Library Editor
Problem:
The issue can be seen in the steps below as 1) Add library X to Qual (same seen with Unknowns Analysis) library list 2) Perform search 3) Open Library Editor and add a new compound, then save library X 4) search again, new compound not found (bad) 5) edit library to use Y and search 6) change back to Library X, search, new compound not found (bad) 7) close application, restart application, search library X and compound is found.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Accurate mass score results are different in B.07.01 vs. previous versions
Problem:
The accurate mass score results (Mass Abundance Score, Mass Accuracy Score, Mass Match Score) in B.07.01 are different than in versions prior to B.07.01. This change is caused by an improvement to the underlying algorithm that determines which ions are considered to be part of an isotopic cluster. This can cause a different set of ions to be selected and thus the different score is a reflection in ion set rather than a change in the way the scores are calculated.
Temporary Solution:
None
Fix Information:
Not a defect
Keyword:
One-line Description:
Batch Status in PDF Reports is not translated
Problem:
Batch Status is not translated in Reports. The string is found in the Reports below: Gen_ByCompound Gen_Calibration Gen_Complete Gen_ResultsQualifierRatio Gen_ResultsSumary Gen_Samples
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Different target peak selected in Quant B.07.01 as compared to Quant B.07.00
Problem:
Peak at different retention time integrated by automatic peak selection
Temporary Solution:
None
Fix Information:
For multiple targets, targets are now assigned within the acquisition-specified Time Segments instead of first come-first serve logic
Keyword:
One-line Description:
Chromatogram Information window: Page Setup does not work properly
Problem:
The printed chromatogram is truncated under certain conditions: Upper margin of page is approximately doubled and the bottom of the page is truncated about one inch
Temporary Solution:
n/a
Fix Information:
Using 'Fit to Sheet' in Page Setup prints the chromatogram(s) without truncation
Keyword:
One-line Description:
Update Qualifier feature in the method edit can crash Quant if there is no peak found
Problem:
Problem can be seen when an attempt to update the qualifier ratios using Update -> Update Qualifier Ratios within the method editor. If one of the selected compounds has a target or qualifier where the integrator has failed to find a peak then a crash will happen. Normally this operation would be done on a standard where all compound target and qualifiers would be integrated and have peaks but if one is missed resulting in a "no peaks found" then the application has a hard crash.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Unknowns Analysis: Library search using accurate mass gives low library match scores
Problem:
Accurate mass library search does not work
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
MRM/Scan data does not display correct spectra when "Scan" or "Deconvoluted scan" selected in "Spectrum Extraction Override"
Problem:
In Quant B.07.00 it is possible to override the normal spectrum type, default is signal that is used to extract the target compounds EIC, to either "Scan" or "Deconvoluted scan". This is done by going to the Method Editor, then Advanced Tasks -> Spectrum Extraction Setup -> Spectrum Extraction Override. For MRM/Scan data this feature does not work for either "Scan" or "Deconvoluted scan" choices and MRM spectra are still shown. The "Scan" mode is fixed in B.07.01, the "Deconvoluted scan" is still non-functional in B.07.01
Temporary Solution:
None
Fix Information:
Fixed in B.07.01 (for Scan spectra only)
Keyword:
One-line Description:
Accurate mass reference library gives less that 100 fit on spectra used to create it
Problem:
This issue can be seen when an accurate mass spectra of a compound is used to create a reference spectra for the compound. The resulting library match is less than 100% when comparing the library spectra to the sample spectra. The expectation would be a 100% match. However when an accurate mass library entry is created from sample data, there is a 1% threshold applied to remove noises from the library spectrum. So when the cleaned up library spectrum is used to do the Library match score against a non-cleaned up spectrum generated from extraction, the score is not 100.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
In Quant Report, the fixed range graphics settings is not applied to the qualifier chromatogram.
Problem:
Set up the fixed range graphics to fix the Y scale in peak chromatogram report. When the quant report is generated the target peak chromatogram is displayed in the fixed scale as defined, but the qualifier peak chromatogram is not set to the fixed scale. Seen in B.07.00.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Tune Evaluation displays 'Object reference' message on some data files
Problem:
Tune Evaluation displays 'Object reference' message on some data files.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Unknown Analysis - Chromatogram Information window: If anchored chromatogram is not displayed, chromatogram order is not the original order
Problem:
If the anchored chromatogram is not displayed (checked) then the order of the chromatogram graphics will not match the order in the list.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
QC Relative Standard Deviation outlier is not calculated for separate streams when using Sample Group Bracketing
Problem:
QC RSD is the same value for each individual stream and the same as when all streams are calculated together.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Targeted Deconvolution Workflow - Quantitative Analysis criteria failures as a screening tool with selecting incorrect peak (or non-peak) as compound due to "out" qualifiers
Problem:
The Close RT with Qualifiers parameter was chosen in method to act as a screening tool. Compounds, known to be in sample either due to cross-check with AMDIS (and/or Unknowns) or based on standard/known sample, had low or no LMS values with bad deconvoluted mass spectra matches, but typically good target/qualifier ratios. The "true retention time," which was typically closer to (or same as) the library RT than the selected incorrect peak, but at least one of the integrated qualifiers were "out" (either too high or low) and the system then looked for another peak with possibly better qualifiers. These compounds were misidentified due to criterion that matches target/qualifiers prior to MS match. When the "correct" RT peak is manually integrated, the deconvoluted MS has a better match and the LMS value increases to typically 80 or higher. These failures were caught because the samples were either known (standards) and/or were also run in AMDIS/Unknowns. The main issue with this criteria failure exists for unknown samples because the customer does not know what is in the sample, so these failures would lead to false negatives, unless the customer also ran AMDIS or Unknowns Analysis. The criteria failures is seen in the Targeted Deconvolution Workflow in B.06.00 and B.07.00.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Unknown Analysis - Deconvolution peak filtering not functional - Component area is less than Area Filter Absolute
Problem:
Unknown Analysis - Deconvolution peak filtering not functional - Component area is less than Area Filter Absolute
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Some Excel Report templates show acquisition time blank when Compound group is set
Problem:
A problem has been seen in an Excel report template in which the acquisition time becomes blank once Compound Group values are entered. Prior to entering the Compound Group values into the batch the same report gave the correct acquisition time.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Initiate Quantitation: If Quant opens in Compound Table view, the sample is not automatically selected (sample is blank) in the batch table so results are not displayed
Problem:
Steps to reproduce: Open Quant. Select 'Compound Table' view. Close Quant 1) Open Qual 2) Open a MRM data file 3) Under 'Find Compounds', click 'Find by MRM'. Click 'Find Compounds by MRM' 4) Left-click on 'Compounds' in the Navigation Tree 5) Right-click and select 'Initiate Quantitation'. Click 'Launch' --> Quant opens and displays method edit 6) Exit method editor applying the method and analyzing the batch --> No sample is selected so no results are displayed (see attached screenshots) Note: Sample is automatically selected and results are displayed when the Quant batch table is in Flat Table view
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Unknowns Analysis - Cannot import the quant method after adding samples to the
Problem:
If you need to make any changes to the imported quant method after you’ve imported it (added samples first) you have to create a new UA batch. It does not allow you to import the quant method again. Steps to see this: 1) Create new analysis adding some but not all samples in the data folder 2) Select File > Import Target Method (Quantitative Analysis method). Select the quant method. Click Open 3) Add more samples to the analysis 4) Select File > Import Target Method (Quantitative Analysis method). Select the quant method. Click Open --> "Please import method after adding samples" message displays
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Rounding of data values to desired numeric format can make outliers appear to be incorrect
Problem:
In some cases the number of digits used to express the outlier limit is greater than the number of digits used to display the value in the batch table on which it acts. When this is the case rounding takes place for display purposes but the outlier limit is checked with unrounded, full precision numbers. As a result the outlier state may indicate there is an high or low limit issue yet show rounded data that contradicts this.
Temporary Solution:
Make sure that the number of digits displayed in the batch table is as great or greater than the precision of the outlier limits.
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Save a Tune Eval method as the same file fails
Problem:
Modify tune eval method and then "Save As" to the same file, it fails with the message "The process cannot access the file.... because it is being used by another process."
Temporary Solution:
Use "Save" rather than "Save As"
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Chinese and Japanese systems show error when generating reports based on PDF templates and requesting Single sample mode
Problem:
Chinese and Japanese systems show error when generating reports based on PDF templates and requesting Single sample mode. Not seen in English systems. Seen in B.07.00
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
PDF reporting fails with Compliance
Problem:
If an attempt is made to generate a report using a PDF report template using the "Generate report now" option the process will abort with a message "Failed to generate report". Using the Queue report fails also.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01 (for "Generate report now" option only, will not be addressed in Queue)
Keyword:
One-line Description:
'SaveBatchTableAs' and 'SaveBatch' commands are recorded twice in the audit trail
Problem:
'SaveBatchTableAs' and 'SaveBatch' commands are recorded twice in the audit trail
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Start time is not recorded in the batch audit trail
Problem:
Start time is not recorded in the batch audit trail when New Method from Acquired Chromatographic Data is done
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Excel Report Template - Filter on Peak_id fails on last row of table when using
Problem:
This issue can be seen in some customized reports where the last row in a table is using the Excel based filter to select a Peak_id (equal to zero or blank) and that table also contains an aggregate function such as Max, Min, StdDev. etc. There is an issue in Excel which causes it to mistakenly think that the last row is a "Totals Row" and not apply the filter to it. This results the last row being displayed in the report when it should not.
Temporary Solution:
To avoid it add a sort on the peak_id such that the last row will contain a value that would normally be includes (zero or blank).
Fix Information:
This is an Excel issue outside of Agilent's control. May or may not be resolved in a future Excel release or patch.
Keyword:
One-line Description:
“Calibration Range” inappropriately applied to Surrogate when sample type is changed
Problem:
The “Calibration Range” outlier which is displayed in the Calc conc” column should not be active for Compound Type of Surrogate however when the sample type is changed to a Cal, QC or CC the code calculates and if it is out of limits displays the red or blue outlier indicator.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
“Accuracy” and “Calc conc” outliers highlighted on Surrogate compound when changing sample type but not on Analyze batch command
Problem:
Problem appears when a compound of type Surrogate is Analyzed and the "Accuracy" and "Calc conc" outliers are outside of limits set in the method. Changing the Sample type from Cal, QC or CC to some other value and then back to the original Cal, QC or CC causes the "Accuracy" and "Calc conc" outliers to display. The core issue is that the outliers should not have worked at all as in B.07.00 they only apply to CompoundType of Target, not Surrogate (as is the case in the example) or MartixSpike. The "defect" is that the changing of sample type code didn't check the Compound Type and thus inconstant behavior. The code will be changed in B.07.01 to calculate “Accuracy” outlier on CompoundType equal to Target, Surrogate and MatrixSpike including the Analyze Batch command. The “Calibration Range”, which highlights the “Calc conc” column, will not be active on CompoundType of Surrogate or MartixSpike. Usually there is only one concentration in the calibration table for Surrogate and MatrixSpike compounds so any number other than the one in the calibration table will be flagged.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00 The code was changed in B.07.01 to calculate “Accuracy” outlier on CompoundType equal to Target, Surrogate and MatrixSpike including the Analyze Batch command. The “Calibration Range”, which highlights the “Calc conc” column, will not be active on CompoundType of Surrogate or MartixSpike. Usually there is only one concentration in the calibration table for Surrogate and MatrixSpike compounds so any number other than the one in the calibration table will be flagged.
Keyword:
One-line Description:
Matrix Spike outlier not flagged unless 'MatrixA' entered in 'Info' column
Problem:
In Quant B.07.00 if the "Info" column in the sample table is left blank then the software defaults to the Matrix Spike outlier using MatrixA values. In Quant B.07.01 if the "Info" is blank the outlier is not enabled and the '"Info" column must be set to "MartixA".
Temporary Solution:
Update older sample's 'Info' field to "MarixA"
Fix Information:
n/a
Keyword:
One-line Description:
RRT outliers not always highlighting
Problem:
In certain workflows the RRT outlier has failed to display in the batch table until the same unaltered method is reapplied to the batch.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Tune Evaluation is not combining mass fragments in the graphics but does sum the correct in the text portion of the Report Mass Hunter Quant Env.
Problem:
This typically occurs with BFB. In the some files mass 176 is fragmented as 175.7 175.8 and 176. The text combines these as one mass 176 and the rel abundance and raw abundance is correct however the spectrum in the graphics shows them as individual masses. ChemStation could handle this and combine them as 1 mass, however Quant does not and this could cause questions.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Unknown Analysis – Tentatively Identified Compounds Query does not respect best hits, shows same compound multiple times
Problem:
The Sample Tentatively Identified Compounds filter.ling.qry the software returns the same compound (with identical RT and Area) multiple times. This is due to the multiple RT window feature and the query not taking this into account.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Quant Reporting: Audit trail report using Excel template has empty rows where recorded actions should be
Problem:
An empty row is sometimes displayed for the last row on a page.
Temporary Solution:
Use the Quant PDF audit report template (AuditTrail.report.xml) instead of the Excel audit trail report template. The Excel audit trail report has been removed.
Fix Information:
Fix Not Planned - Use the Quant PDF audit report template (AuditTrail.report.xml) instead of the Excel audit trail report template. The Excel audit trail report has been removed.
Keyword:
One-line Description:
MRM/Scan spectra fails to give Library Match score
Problem:
If a scan is created from MRM/Scan data using the "Scan" option the spectra is seen but fails to give Library Match score
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Quant ECM: Choosing an older revision of a report method does not display the older revision
Problem:
Steps to reproduce: 1) Open batch 2) Select Report > Generate. Click 'Choose' 3) Right-click a report method with multiple versions and select 'Open Older Revision'. Select a previous revision. Click 'OK' 4) Click 'Edit'. Most recent revision of this report method is displayed instead of previous revision
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Drug Analysis Mass Hunter Quant report pointing to Sample Info not Sample name. B.07.00 Druganalyis.cs
Problem:
Drug Analysis Mass Hunter Quant report prints Sample Info not Sample name. Problem seen in B.07.00 and B.07.01 copy of Druganalyis.cs
Temporary Solution:
Modified report files (Druganalysis.resx and Druganalysis.cs) that print Sample name are available upon request.
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Unknowns Analysis with ECM: If the report method cannot be located, message displayed that '... Revision number "...' does not exist
Problem:
Unknowns Analysis with ECM: If the report method cannot be located, message displayed that '... Revision number "...' does not exist
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Unknowns Analysis: A blank option from pulldown menu for Integrator and Algorithm in Deconvolution Advanced parameters
Problem:
There is a blank entry in the pulldown menu for both the Integrator and Algorithm in Deconvolution Advanced parameters
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Unknowns Analysis: Audit trail window shows 'Succeeded'' as not checked but no error message is displayed
Problem:
The audit trail window in Unknowns Analysis can indicate that an operation was not successful, the 'Succeeded' as not checked, but no error message is displayed.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
IQT Report will fail with bad file if Excel is open when running IQT Report
Problem:
IQT fails for Quant Reporting, one invalid file is reported in generated IQT Report, when IQT is run while Excel is open.
Temporary Solution:
Close Excel and rerun IQT report, report should be good.
Fix Information:
Not considered a defect
Keyword:
One-line Description:
Japanese version of Help lacks Glossary section
Problem:
Japanese version of Help lacks Glossary section
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Online Help - Links to Familiarization Guides are not working
Problem:
Online Help - Links to Familiarization Guides are not working
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Agile2 integrator fails to integrate the qualifier peak
Problem:
What appears to be simple integration of a qualifier peak is missed using the Agile2 integrator. The General integrator finds this qualifier peak.
Temporary Solution:
n/a
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Previous/Next compound works erratically when Quant is in Compound Table mode
Problem:
When Quant is in Compound Table mode the order of visiting compounds is erratic and never visits some compounds when using the Previous/Next compound buttons. In one observed case the software repeatedly cycled back and forth between two compounds as if stuck in a loop.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
MRM Spectra not normalized when library shown
Problem:
When MRM spectrum from a sample and Reference Library are displayed the Reference Library spectra is not properly normalized to match the sample spectra. Problem seen in B.07.01 but not B.07.00.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Library Match Score and/or Purity are different
Problem:
Library Match Score and/or Purity are different in Quant B.08.00 when compared to Quant B.07.01
Temporary Solution:
Since the integrator changed from B.07.01, deconvolution results may be different so LMS, purity, etc are different
Fix Information:
Not applicable
Keyword:
One-line Description:
Reference Library Spectrum is being scaled differently than the source spectrum in Compound Information
Problem:
When running a library search in Quantitative Analysis, users have observed that the library spectrum (below the line) is being scaled to a small normalized axis that does not match the sample spectrum. As such, it's very difficult to see and compare.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Unable to generate report using 'Unknowns_BestHits_B_07_00.xltx' template
Problem:
Message displays: "Failed to generate the report ... Please specify one batch ID"
Temporary Solution:
n/a
Fix Information:
Must select 'Selected sample(s)' when generating the report as this report only applies to one sample
Keyword:
One-line Description:
TIC Action script and entering m/z=0 in the quant method do not give the same results
Problem:
The TIC Action script uses the m/z range in the sample selected when entering method edit. This range may not cover the m/z range in all samples.
Temporary Solution:
N/A
Fix Information:
N/A
Keyword:
One-line Description:
Integration difference between Quant B.07.01 and Quant B.08.00 using MS-MS (GC) integrator
Problem:
The change is due to a fix for MRM data and time segment boundaries, so results are different but more accurate
Temporary Solution:
None
Fix Information:
Change in B.08.00
Keyword:
One-line Description:
Some PDF reports show "ND" on some compounds when the compound is found
Problem:
This problem is seen in some PDF reports which produce a summary section that creates a list of concentrations of a compound across all samples. If the 1st sample in the batch, based on acquisition date, the column “Final Conc” has an “ND” (Compound not found) then all samples in the batch will show “ND” even though it was in fact found. Reports affected are Gen_ResultsSummary.report.xml and Gen_Compplete.report.xml. This behavior is also observed if MS data is not present in the data file. For example, if one is quantitative a UV only data file.
Temporary Solution:
Request updated report code.
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
The previous shipping script gives incorrect results
Problem:
The script does not widen the extraction window to include all MZs.
Temporary Solution:
None
Fix Information:
Changed in B.08.00 Enter m/z = 0 in the method for TIC compounds instead of using the action script
Keyword:
One-line Description:
Quant Actions script "Send To OpenLAB ECM..." used to send a Quant batch to ECM does not properly handle Quant batch files with SPACES in the name
Problem:
Quant Actions script "Send To OpenLAB ECM..." used to send a Quant batch to ECM does not properly handle Quant batch files with SPACES in the name. Issue can be seen in all versions of Quant from B.05.01 to B.07.01 SP1.
Temporary Solution:
An updated script is available on request.
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Cannot create a quantitation method from a PCDL
Problem:
Message displayed that the library is too large when attempting to create a new quantitation method from a PCDL
Temporary Solution:
If the library has more than 32766 compounds, need to create a subset library and use that smaller library to create the quantitation method
Fix Information:
Not a Defect - Info only
Keyword:
One-line Description:
TDA files created on all files in folder not only files in Add Samples
Problem:
When user brings in a subset of the TOF data files into a Quant batch via "add samples" all the files in the folder get the TDA files created rather than only the samples added. This can delay the user and consume extra disk space. Issue was seen in Quant B.06.00 and B.07.00.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01
Keyword:
One-line Description:
Samples visited in odd order or skipped when Quant in Compound Table view mode and using Next Sample button in Batch Table
Problem:
Problem is seen only when user creates a batch and then removes a sample and add it or another sample back to the batch.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
ECM Quant - User Validation
Problem:
Quant will accept a user validation by someone other the logged in user. 1 - The user that supplies the credentials doesn't have to have permission to do the action (only the logged in user). 2 - The audit trail shows the logged in user not the one supplying credentials.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Correct change in ATM Configuration logon behavior between B.07.00 and B.07.01
Problem:
Quant B.07.00: Required to be in QuantAdministrators group to login to ATM Configuration Quant B.07.01: Not required to be in QuantAdministrators group to login to ATM Configuration Quant B.08.00 will return to B.07.00 behavior.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Partially or completely missing deconvoluted mass spectrum in TD Quantitative and Unknowns Analysis with no or low LMS values after target and qualifier ion integrations
Problem:
When analyzing a sample, TD Quant (or Unknowns) will integrate the target ion and qualifier ions (in the peak extraction window) and generate qualifier ratios, but no deconvoluted mass spectrum or only a partial deconvoluted MS will be shown in the batch table (batch at a glance view). The LMS value in the batch table will be either low (for partial MS) or none will be found (for no MS), even though there is a peak in the TIC and a matching raw spectrum to the library spectrum. This issue is related to both data rate and Agile2 integrator.
Temporary Solution:
Increase scans / sec to degree possible may help reduce the frequency of occurrence.
Fix Information:
Info for B.08.00
Keyword:
One-line Description:
MRM/Scan data does not display spectra when deconvoluted scan selected
Problem:
MRM/Scan data does not display spectra when deconvoluted scan selected.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
When selecting compounds in the report method, the CC_mid point report fails
Problem:
When printing the "CC" reports if the user selects a subset of compounds rather than all compounds the report fails. An error message of "Error: empty sequence" is seen.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Quant crashes with left mouse drag in extracted spectrum
Problem:
Steps that cause crash. 1) go to method editor 2) Extract spectra 3) position mouse above the spectra x axis 4) hold down left mouse 5) drag mouse below spectra x axis 6) Quant has crash
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
ECM Quant : Adding Quant Reporting method to existing method "save as" not possible
Problem:
Only seen in ECM quant. This issue can be seen if the user uses the "New" button to create a Quant Reporting Method and then tries to use the "Save" option to place it into an existing >m method folder. When this is done Quant gives an error message saying method exists when in fact there is no Quant Reporting Method in the .M folder. This is a common workflow that works in non-ECM Quant where it works.
Temporary Solution:
Workaround is to "Edit" the existing method and then "Save".
Fix Information:
n/a
Keyword:
One-line Description:
ECM : Sample type is always SAMPLE even when set to Cal, QC, etc
Problem:
When samples are added to a Quant batch all samples are imported as SAMPLE regardless of the sample type specified in the acquisition sequence. Seen only in ECM mode.
Temporary Solution:
None
Fix Information:
Fixed in B.07.01 SP2
Keyword:
One-line Description:
Cannot log into ATM Configuration when user is in Power Users group
Problem:
Steps to reproduce: 1) Logon to PC as a user in the Power Users group (Reproduced with local user in Power Users group) 2) Attempt to open ATM Configuration (enter appropriate password) --> ATM Configuration never opens (cursor keeps spinning) 3) Open Task Manager and check Processes Not observed in Quant B.07.00
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Quant B.07.00 with Roles and Responsibilities enabled creates inconsistent audit trail items
Problem:
The user is comparing the audit trail visible from Tools -> Audit Trail against the contents of the audit trail report generated by the Audit trail PDF report template. Essentially the problem is that every time user generates an audit trail report, steps required to generate a report, such as a "CmdOpenBatchTable" entry, is added to the audittrail.pdf. However, this entry is NOT visible in the Tools -> Audit Trail table. Also seen in B.07.01
Temporary Solution:
None
Fix Information:
Fixed in B.07.01 SP2
Keyword:
One-line Description:
Excel report does not support 'All Samples' option in Unknowns Analysis
Problem:
If one of the Excel report templates is run with the "All Samples" option the report will not be generated and a "Failed to generate the report" error message is displayed. This is a limitation with the software present in all versions and it is not planned to be addressed in the future.
Temporary Solution:
None
Fix Information:
No Fix Planned
Keyword:
One-line Description:
Crash in Compounds-at-a-Glance when there is an empty chromatogram (“NO DATA POINT”) and showing reference RT
Problem:
Problem is seen when there is no data extracted and there is one or more empty chromatograms (“NO DATA POINT”) and the option to display the RT reference is active. If RT reference is not displayed then there is no crash.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Creating compound with Normalize Qualifier enabled causes the qualifier to flatten to 0 response
Problem:
When you adding a new compound (or creating compounds initially), there is a period where the normalized qualifier , as seen in the Qualifier pane of the Compound Information window, flattens to 0 response. The issue only occurs during method creation, then after first analyzing the batch, it seems to disappear until a new compound is added.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Unknowns Analysis - CEF file export shows deleted compounds
Problem:
The script "ExportCEFFile-TargetHits.Unknowns.script" exports compounds that have been deleted. To see this create a export of a "complete" list of hits, export, delete all but one or two hit compounds, export. Exported CEF files are the same size and 2nd file contains the delete compounds. Reported in B.07.00 and reproduced in B.07.01
Temporary Solution:
Updated Export files are available if requested
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Report folder does not increment when report queue processing is slow
Problem:
The problem occurs when a user is in background print queue mode, they ask for multiple reports on a batch and the Queue is not processing fast enough to keep up. To reproduce: 1 - Stop the Queue so no reports are processed (or run a really slow Excel based report) 2 - Generate a report using "Queue Report task" (don't start to simulate a backlog in processing) and note report folder 3 - Repeat step 2 and note that report folder doesn't increment. 4 - Start Queue running, note 1st report (step#2) runs OK but report #2 (step#3) fails as it uses the same report folder as step#2 Also seen in B.07.01.
Temporary Solution:
Workaround is to avoid the use the report queue and use the Generate Reports now option.
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Compounds-At-a-Glance :Zero peak runs slowly when layout by Compound and multiple cells zeroed
Problem:
When Compounds-At-a-Glance Layout is configured Organize Rows by Compound and Review Mode a Compound by Compound and more than 3 cells are selected and zeroed performance is poor when compared to other configurations. If a large number of compounds (400) are selected the delay can reach 3 hours. Also observed in B.07.01.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Compounds-At-a-Glance will crash if displaying of Reference RT and there is no Chromatogram (No Data)
Problem:
If there is a compound for which there is no data displaying in the chromatogram window and the user launches Compounds-At-a-Glance with the Reference RT displayed there will be a hard crash.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
In standard (non-ECM) Quant, when creating levels from calibration samples, the levels display as "1, 2, 3, ..." and need to be changed manually to the correct concentrations
Problem:
Behavior is different in Quant ECM: When creating levels from calibration samples, the concentrations in the method are correct.
Temporary Solution:
None
Fix Information:
No Fix Planned
Keyword:
One-line Description:
Default correlation window value changed from 2 minutes to 0.5 minute
Problem:
Default correlation window value changed from 2 minutes to 0.5 minute
Temporary Solution:
None
Fix Information:
Changed in B.08.00
Keyword:
One-line Description:
Internal script examples removed from the shipping product installation
Problem:
CompoundMath_Script, Batch_Script, CustomExpressions_Script, and OutlierAccuracy_Script were removed
Temporary Solution:
None
Fix Information:
Changed in B.08.00
Keyword:
One-line Description:
ECM Quant: Gen_LIMsIntegration.report.xml abort with error
Problem:
When running the report Gen_LIMsIntegration.report.xml in ECM mode the report generation fails with error "Failed to generate report". Problem is the code does not close the .csv file which causes an error when ECM upload is attempted.
Temporary Solution:
Contact service for revised report
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
PDF Report Builder - SpaceBefore parameter has no effect for the Batch table
Problem:
The SpaceBefore for the Batch table has no effect in the report output. This parameter works for other table, only has no effort in this batch table.
Temporary Solution:
None
Fix Information:
Limitation in iText, Not to be fixed
Keyword:
One-line Description:
Deconvolution only works in the first time segment for SIM/Scan data
Problem:
Deconvolution only works in the first time segment for SIM/Scan data
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Matrix Type Override outlier: Can add the same matrix type in the Matrix Override table with a different limit
Problem:
Matrix Type Override outlier: Can add the same matrix type in the Matrix Override table with a different limit
Temporary Solution:
User needs to check that there are not multiple rows with the same matrix type
Fix Information:
n/a
Keyword:
One-line Description:
LC-only data (no mass spec) is not automatically converted when adding samples to the batch table
Problem:
LC-only data (no mass spec) is not automatically converted when adding samples to the batch table
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Matrix Type Override outlier: Entering a negative value for outlier limit (e.g., Sample RSD) is not flagged in method validation
Problem:
Method error is flagged at batch analysis time in the batch table
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
The R2 and average response factor RSD outliers show in Flat Table (Single Compound) but not Compound Table (Single sample)
Problem:
The R2 and average response factor RSD outliers show in Flat Table (Single Compound) but not Compound Table (Single sample)
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Method validation fails to flag surrogate recovery limits for non-surrogate compounds
Problem:
n/a
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Defect in reporting API prevents proper scaling of qualifier graphic
Problem:
There is a defect in MassHunter Quant DA B.07.01 SP1 which prevents scaling of the Qualifier graphics - specifically all scaling options set within Quant are ignored.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Next Sample and Next Compound work differently in forward and reverse
Problem:
The next compound and next sample buttons work differently when progressing 'forward' and when progressing 'backwards'.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Library match score (LMS) is different in B.08.00 compared to B.07.01 SP2
Problem:
Due to change in accurate mass data library match score calculation
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Matrix Type override outlier: Does not use the expected outlier limits
Problem:
No method validation occurs in the Matrix Type Override window in Quant B.08.00, so the existing matrix type outlier (before Quant B.08.00) can be used in addition to the new matrix type outlier (override). If an outlier is set up with both the old (previously-existing) way AND the new way, outliers may not display or flag correctly.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Unknowns Analysis - checking Open published file does not work
Problem:
In Unknowns Analysis Report method if the Open published file is checked the document does not open automatically as seen in Quant.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Arrow key navigation inconsistent when in BAG in Compound Table-Multi-Sample mode
Problem:
To see the issue 1) Place BAG in Compound table mode 2) place BAG in Multiple Sample/Compound mode 3) click mouse into RT field of left most sample 4) use right arrow to move to next field, right arrow is in field by field mode 5) keep pressing right arrow until you leave the sample 5) behavior changes and right arrow now goes down to next compound and no longer in filed by field mode
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
"New Method from Acquired MRM data" algorithm picks peaks outside the RT range as specified in the dMRM acquisition method
Problem:
Starting with Quant B.08.00 the algorithm that locates the largest peak, and thus determines what RT will be used in the method, has been changed to respect the RT range specified in dMRM data files. All prior releases of Quant would locate the largest peak in the entire data file for the signal and in some cases that peak would be outside of the specified RT range from acquisition and thus pick an incorrect peak and RT.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Signal smoothing on individual summed compounds is not seen in the summed chromatogram
Problem:
If you set up a summed compound A+B and then activate smoothing on both compound A and compound B you will see the A and the B chromatograms displayed with smoothing but the summed compound A+B is unsmoothed .
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Different qualifier results in Quant B.08.00 when compared to Quant B.07.01
Problem:
Differences observed: 1) Integration different - Qualifier peak response different 2) Chromatogram looks different - Straight/flat trace is no longer displayed 3) Qualifier peaks are found in Quant B.08.00 that are not found in Quant B.07.01
Temporary Solution:
n/a
Fix Information:
Changes in Quant B.08.00 data access for MRM and SIM data to resolve another issue
Keyword:
One-line Description:
Excel 2016 reports stop when option to print by sample and $(SampleName) option is used
Problem:
When one tries to create a set of Excel based reports and use single sample mode and also specify the $(SampleName) Excel launches but stops.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Translated Ion trap data does not display spectra when deconvoluted scan selected
Problem:
Customer has translated MS ion trap data. All is well if you create a reference library from averaged spectra and compare that to the averaged spectra. But if you try and change the spectra type to deconvoluted scan there is no deconvoluted spectra displayed. Also seen when Create New Method from Scan Data is used as there will be no compounds created.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
MI of peak outside of Reference or Non Reference Window creates inconsistent state
Problem:
This issue was reported in Compounds-at-a-Glance but also seen in Batch-at-a-Glance 1) Open DOA demo data set 2) Visit ME set Non reference window to 20%, apply and analyze batch 3) visit Amp on Cal L2 4) Manually integrate the small peak outside the reference window (see before MI attachment). 5) Note error message 6) Now note that the peak is "not found" (all fields blank) and MI is not checked and the restore Clear MI is grayed out (see after MI Attachment). 7) analyze batch and note that the peak is not reintegrated. The problem here is 1) that the user has altered the results of the integrator using MI and there is no indication of a "MI". This is a great way to hide a hit with no indication that it was there (unlike zero peak). 2) The "phantom MI" is somehow protecting from the analyze batch fixing it. I am aware that I could recover by manually integrating the original peak then clearing manual integration but that is not the issue it is item #1 and #2. Seems like the correct action is not set the MI to checked as MI did fail, but leave the original peak as the selected peak. Reported in B.07.01.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
PDF Report QuantCSV.report.xml fails or has missing data
Problem:
Problems seen in the QuantCSV.report.xml report 1. The ‘batch mode’ will only put one sample in the CSV file, not the entire batch. 2. In ‘single sample’ mode, you need to select a folder that already exists. The process will NOT create the incremented folder shown in the Report UI. Seen in B.07.01 and Quant B.08.00.570. This was reported by Ken W. for "Jeff" at a customer site so this is external. Tom B. observed in B.07.01.
Temporary Solution:
Updated report available on request.
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Compounds-At-a-Glance Layout setup does not retain Pane Dimension after displaying Reference RT
Problem:
Compounds-At-a-Glance Layout setup does not retain Pane Dimension after displaying Reference RT
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Unknowns Analysis with ECM: Report generation does not complete and 'Cancel' option is inactive
Problem:
Progress bar continually shows the green scrolling from left to right. 'Cancel' is greyed out. Closed process using Task Manager. ECM shows the Best Hits report and it can be opened. Uaf file remains 'checked out'. Issue related to using .L library.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Quant with ECM: Crash when report template is selected instead of a report method
Problem:
Quant with ECM: Crash when report template is selected instead of a report method Reproduced issue using an Excel template, Quant B.07.01 SP1, and ECM 3.4.2 SP1
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Initial Calibration Report includes CC sample type
Problem:
Initial Calibration Report includes CC sample type
Temporary Solution:
Corrected report available on request
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Compound-at-a-Glance: Link to Y AXIS will be deselected after moving two compounds.
Problem:
If you select the "Link to Y-Axis" within Compound-at-a-Glance and then move to the next compound the following will happen: 1) After the first move the scale will completely disappear 2) After the 2nd move the scale will appear at each graphical window and the "Link to Y AXIS" will be deselected.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
CAG crash when Reference RT is displayed in CAG, not displayed in BAG, and a "no data found" cell is displayed in CAG
Problem:
If you set the Reference RT to not display in BAG, have it set to display in CAG, and you try and display a cell with No Data Points, CAG will crash with a "red x".
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
MassHunter Quant B.07.01 fails with "Arithmetic operation resulted in an overflow" while B.07.00 works fine
Problem:
MassHunter Quant B.07.01 fails with "Arithmetic operation resulted in an overflow" while B.07.00 works fine
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Open Report option no longer available in Queue Viewer Action menu
Problem:
In B.07.00 - a user could open the queue viewer and use the Action -> Open 'reportname.pdf' to open the published file from both current and older reports. In B.07.01, that functionality is gone, and now we can only open the report directory - making the customer find the .pdf or .xlsx output file manually.
Temporary Solution:
None
Fix Information:
Fixed in B.00.00
Keyword:
One-line Description:
Metrics Plot for ISTD Response visually corrupts Batch Table
Problem:
Open a batch. Add column for ISTD Response if not present. Plot this column. In the metrics plot window, SELECT or highlight one of the points to move to that line in the Batch Table. Note that ALL the information about the ISTD is lost and the Responses for the compounds are blank. Switching compounds will restore the values. Also seen in B.07.01
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Quant B.07.01 with UMAT enabled or ECM enabled cannot generate reports if CmdOpenBatchTable has User Validation and/or Reason checked
Problem:
If I have User Management and Audit Trails enabled, Quant will not generate reports if CmdOpenBatchTable has User Validation and Reason checked.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Some Excel based report templates that worked correctly in Quant B.07.00 will generate extra blank pages in Quant B.07.01 or greater.
Problem:
Some Excel based report templates that worked correctly in Quant B.07.00 will generate extra blank pages in Quant B.07.01 or greater. The issue is related to an change in the Quant Add-in going from version 4.31 in Quant B.07.00 to 4.33 in Quant B.07.01 and greater.
Temporary Solution:
Downgrade the Quant Add-In to version 4.31.
Fix Information:
No Fix Planned
Keyword:
One-line Description:
Changing Library Search parameter blocked by access to file in /Quant/bin
Problem:
To see the issue try doing a library search from the spectra window in Compound Information Window. After giving a library name and seeing search results access the Library Parameters via menu File > Search Parameters. Try editing the number of compounds or min match score and hit OK. A file access message appears.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
In report-generated graphics, the full scale display of the overlaid target compounds TIC occurs before the No Display setting is applied, instead of after like in B.07.01
Problem:
There is a issue about the scaling of the overlay targets chromatogram when the TIC is set to "no display". The behavior has changed from B.07.00 build 7.0.457.0 to B.07.01 SP1. In B.07.00 - the TIC would be scaled to 'full scale' AFTER the TIC was removed from the graphic. In B.07.01 - the TIC scales to 'full scale' BEFORE the TIC is removed from the graphic. To be clear - this is specifically an issue only in the graphics produced in the printed PDF report - NOT in the UI of the quant method editor (which always scales correctly). You will only see the problem if you generate reports.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Behavior when opening a batch that contains data files converted to TDA format
Problem:
Quant B.08.00 has a change in behavior to automatically use Archer data access if no converted files exist for accurate mass data. Results using Archer and results using TDA may not be the exact same results in some cases.
Temporary Solution:
NA
Fix Information:
Changed in B.08.00
Keyword:
One-line Description:
Manually-designated noise regions are not displayed when Auto-RMS noise algorithm is selected
Problem:
Auto-RMS looks for noise range, so it does not make sense to choose this type then manually enter noise regions.
Temporary Solution:
NA
Fix Information:
Documented in B.08.00
Keyword:
One-line Description:
Response Ratio does not update if Quant by Height is enabled and analyte peak is zeroed
Problem:
If Quant by Height is enabled for both ISTD and analyte and I zero the analyte peak, Batch-at-a-Glance retains the original RR. If Quant by Height is not enabled, then the RR becomes zero.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Unknowns Analysis crashes when selecting non-hit component and switching between samples in All mode
Problem:
Problem is seen when the interface is set to display Non-hits and the following steps are taken. 1. Click one sample row on Samples pane 2. Click one component row on Components pane (that is an Non-Hit) 3. Click another sample row on Samples pane 4. Error occurs, and Forces exit.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Peak Labels not displayed when TIC is not displayed.
Problem:
Problem most noticeable in SIM/ Scan data set where it is desired to hide the TIC due to zigzag appearance and only show the overlaid SIM TIC & Scan TIC. When the TIC is set to "No Display" in the Properties window the peak labels also disappear.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
TPH validation report only reporting samples
Problem:
Env_TPH_Validation.report.xml only reports on samples when it should also report on blanks and matrix spike and matrix spike dups.
Temporary Solution:
Corrected report template is available on request.
Fix Information:
Fixed in B.08.00.
Keyword:
One-line Description:
Average Qualifier Ion Ratio does not work in Quant B.07.01
Problem:
When the option is run in Method Editor it updates the ISTD, but using the qualifier ion ratio of the cal that was brought into Method Editor.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Quant ECM: Unknowns Analysis audit trail does not record that the library in the method was changed
Problem:
Quant ECM: Unknowns Analysis audit trail does not record that the library in the method was changed
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Unknowns Analysis - Loads file that is still being collected then crashes when sample is accessed
Problem:
If a data file is being acquired and a user tries to add it to the Unknowns Analysis it will add it but it will not be able to create the MSTree2.bin file and Unknowns Analysis will crash.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Sample Information display in method editor displays centroid spectra even when profile data is set in the config file
Problem:
Configure Quant to extract profile data and delete the TDA files to confirm that they are constructed using profile data correctly. Next - create a new blank method. In the method editing section, extract a spectrum from the peak apex. Note how the sample information spectra is displayed in centroid where the compound information spectra is profile. Request is to display profile spectra in sample information pane.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Integration Parameters checkbox 'Use same parameters as parent compound' not swapped when 'Swap Qualifier with Quantifier' done
Problem:
When swapping the Target and Qualifier the " 'Use same parameters as parent compound' not swapped.
Temporary Solution:
None
Fix Information:
No Fix Planned
Keyword:
One-line Description:
ECM Mode: PDF report template uses Windows user for the 'reporter name' not the user logged into Quant. Excel works correctly
Problem:
The PDF report template is using the logged in Windows user as the 'reporter name' while the Excel based template is correctly using the user logged into Quant.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Scaling of individual qualifier chromatogram leaves gaps and not scaled as target and overlaid qualifers
Problem:
A report that shows the target, overlaid qualifiers and a single qualifier seems not to be scaled the same. The single qualifier has a gap, white space, above the peak .
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Quant is not calculating PPM correctly in the mass extraction window - accurate mass when using TDA
Problem:
TOF Data Access (TDA) which Quant uses by default for all accurate-mass data can induce a mass difference between Qual and Quant. To gain performance, TDA bins data points into m/z slices of a small but finite width. If more than 1/2 the m/z extent of the slice is included in the m/z extraction window, all the data points in the slice are extracted. Typically this is not noticeable, unless the m/z extraction range is very small such as 10 ppm. This can also give rise to different peak shapes between Qual and Quant.
Temporary Solution:
Turn off indexed data access so TDA is not used.
Fix Information:
Fix Not Planned
Keyword:
One-line Description:
Apex spectra blank in SIM/Scan data
Problem:
When the method calls for using the Apex scan, rather than the Average (the default) over the peak on Sim/Scan data the extracted Spectrum is blank.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
DrugQuant_withOneQualifier_B_06_00 and DrugQuant_withTwoQualifier_B_06_00 Excel templates in Chinese and Japanese tune file name and tune time are displayed incorrectly
Problem:
When either the either the DrugQuant_withOneQualifier_B_06_00 or DrugQuant_withTwoQualifier_B_06_00 Excel templates in Chinese and Japanese versions are run both the file name and tune time are displayed incorrectly as "######....#".
Temporary Solution:
Use the PDF version of the report.
Fix Information:
Will Not Be Fixed
Keyword:
One-line Description:
QuantECM: Crash when attempting to change to Compound Table view and library method path is Windows not ECM
Problem:
Problem can be seen when a batch created outside of ECM is uploaded to ECM and it contains a reference to a Library Method that is still in the Windows file space.
Temporary Solution:
The workaround in Quant (B.07.01) is: 1. open the batch 2. close Chromatogram Information window 3. correct library method path in method editor and apply the method 4. reanalyze
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Toggling calibration levels on/off with Standard Addition does not function as expected
Problem:
Problem can be seen in Standard Addition batch when a Calibrator in one sample's calibration curve is disabled and this unexpectedly disables a data point in the preceding sample's calibration curve.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Quant ECM - Tune Evaluation: Unhandled exception
Problem:
Quant ECM - Tune Evaluation: Unhandled exception Note that ECM combined with Environmental mode are not a supported configuration.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
ECM Quant: Add Samples doesn't show Sample Name
Problem:
When adding samples in ECM mode the Sample Name is not displayed. This information is seen in standard Quant and used by some labs to selectively load desired files.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Quant not displaying customer 'User Defined' fields if acquisition column header is changed from 'User Defined x'
Problem:
In the LCMS acquisition if a user wants to configure a ‘User Defined’ column to be used/displayed in either Qual or Quant, they expose that column in Acquisition and they can choose to rename it: User Defined -> “Weight”. However when this is done the data is not imported to Quant.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
B.07.01 .xsd file not updated to include new fields
Problem:
The shipping .xsd files in B.07.01 lack some of the new data fields, one being CompoundMath.
Temporary Solution:
Customers may edit B.07.01 .xsd files. Assistance is available on demand.
Fix Information:
n/a
Keyword:
One-line Description:
Qualifier Ratio outlier flagged for samples of TuneCheck and Blank type
Problem:
Outlier for qualifier peak not found flagged in B.07.01 for TuneCheck and Blank sample types but should not be.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Using Profile data causes mass accuracy calculation to break if mass extraction window is too narrow
Problem:
If you are using Profile data in Quant, and have the mass extraction window set to a small value (~>10 PPM) - it may cause the mass accuracy calculation to break.
Temporary Solution:
The fix is to expand the mass extraction window to >10 PPM.
Fix Information:
n/a
Keyword:
One-line Description:
Quant mis-identifies dMRM RT during Create Method from MRM Data, Qual Disagrees
Problem:
If two compounds have similar MRM transitions (within 0.1 AMU of either product or precursor) - Quant may pick the wrong retention time when creating compounds 'from MRM data'.
Temporary Solution:
Correct the RT manually after creating the method.
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
ATM will attempt to launch when logged into Windows as a 'user' - hangs opening infinite copies of the exe
Problem:
Problem can be seen: 1.) Log into Windows as a local 'power user'. 2.) Launch the ATM utility. 3.) open the task manager to observe an unending list of compliance executables opening until the system runs out of memory.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
MRM Data disappears when window crosses TS boundary
Problem:
In some cases data from some time segments disappear when the extraction windows crosses a time segment boundary. Reducing the extraction to exclude the neighboring time segment will restore the data.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Quant ECM: Error is flagged if Quant Report Method has 'XPS' or 'TEXT' as Publish format.
Problem:
If one selects Publish format as XPS Or TEXT for Quant report method and click on Generate report option from Quant console then error message "Unable to upload file exceeding maximum URL length of 245 characters (error -27)" is flagged.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
The values of Exp.Conc in the Quant reports Gen_ResultsSummary.report.xml and Gen_Complete.report.xml shown the same value taken from 1st sample reported
Problem:
When either the Gen_ResultsSummary.report.xml or Gen_Complete.report.xml templates are run the value of Expected Concentration is read from the 1st sample. If this is a Cal, CC or QC the same value will be show, if from some other sample type the value will be blank.
Temporary Solution:
Corrected templates are available on request.
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Cannot 'Quantitate Batch' when ‘Concentration Sum’ in Compound Math is selected
Problem:
If the method has a Compound Math compound with the Concentration Sum’ option selected and they perform a 'Quantitate Batch' then a message indicates the method isn’t calibrated.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
'Cannot initialize compliance module' error message is displayed when trying to open ATM Configuration
Problem:
In ATM B.07.01, it is allowed (unintentionally) to copy the command node in the tree view that leads to duplicate command definitions in ComplianceConfiguration.xml file. If this is the case then ATM will fail to launch with the error message 'Cannot initialize compliance module' .
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Dilution values are not ported correctly to Unknowns when loading a batch from Quant
Problem:
Dilution values are not ported correctly to Unknowns when loading a batch from Quant
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Report template Gen_ResultsSummary.report.xml pulling position instead of vial number
Problem:
Report template Gen_ResultsSummary.report.xml pulling position instead of vial number
Temporary Solution:
Updated report code available on request.
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Replace Calibration action available under the ‘Analyze’ menu will re-enable the level for a compound that has the level disabled
Problem:
Replace Calibration action available under the ‘Analyze’ menu will re-enable the level for a compound that has the level disabled
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Con Cal reports not using correct Accuracy Outlier in reports B.07.0X
Problem:
The family of Continuing Calibration report templates ( Env_CC_Avg.report.xml, Env_CC_MidPoint.report.xml & Env_CC_Previous.report.xml) are not using the correct outlier to flag data that is out of limits.
Temporary Solution:
Corrected code is available on request
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Adjacent cell formula
Problem:
It has been discovered that some versions of the Excel may not perform calculations of formula if they're placed adjacent to another cell containing a formula. In one example, the 'Cal Update Time' and 'Batch Data Path' were right next to each other (one below the other). Because of the proximity - the Cal Update Time would simply return 'text' instead of actually evaluating the formula contained in the cell.
Temporary Solution:
Fixed the issue by moving the Cal Update Time to below the sample table in the example template. It looks like when 2 formula are next to each other, they interfere.
Fix Information:
No Fix Planned
Keyword:
One-line Description:
Unknowns Analysis in ECM mode fails to retrieve library spectrum in best-hits report
Problem:
It has reported that when running Unknowns Analysis in ECM mode, the best-hits report fails to retrieve the library spectrum correctly.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Baseline calculation points are not displayed after manual integration
Problem:
If Baseline calculation points are displayed they will appear when the peaks have been integrated but disappear when a manual integration is performed. Baseline calculations is a function of the integrator and become meaningless when manual integration is performed
Temporary Solution:
None
Fix Information:
No Fix Required
Keyword:
One-line Description:
Large data file size and slow processing in Quant due to very large MSTree2.bin files
Problem:
Systems that have installed GCMS Acquisition B.07.04 are experiencing large data file size and slow processing in Quant due to very large MSTree2.bin index files being created by Quant. This was caused by a change in data format from 16 to 32 bit format which stores results in many more unique mass values and thus a large and slow MSTree2.bin file. The extra precision in the 32 bit format is meaningless for unit mass data but results in a larger MSTree2.bin file.
Temporary Solution:
It is possible to run the GCMS Translator over the files which will recreate the 16 bit data files and then recreate the usual smaller faster MSTree2.bin files.
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Level code order in ISTD causes validation error
Problem:
Randomly Quant had an error after analyzing the batch : “sample not validated :calibration level name mismatch between target & ISTD”. Same results after checking/correcting/recreating the quant method. Problem found to be for ISTD, calibration level were set in the following order : 1-10-2-3-4-5-6-7-8-9 . For target, cal level were 1-2-3-4-5-6-7-8-9-10. The issue was solved after putting the calibration level for ISTD & target in the same order in the quant method."
Temporary Solution:
Manually reorder the levels in the ISTDs.
Fix Information:
n/a
Keyword:
One-line Description:
Compound math - ISTD method not supported.
Problem:
When using an ISTD method the raw area/height is used not the ISTD adjusted response to create the cal curve and quantitate the sample.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
'Open and Apply from Existing Batch' does not record which batch the method is from in the audit trail
Problem:
'Open and Apply from Existing Batch' does not record which batch the method is from in the audit trail
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
User Maintenance and Audit Trail - EditReportMethod permission not applying - doesn't ask for username nor requires reason
Problem:
User Maintenance and Audit Trail - EditReportMethod permission not applying - doesn't ask for username nor requires reason
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
TIC of SIM/Scan data gives zig zag Chromatogram
Problem:
Problem is seen when SIM/Scan data files are used with a method that sets ScanType=Scan and MZ=0 . The correct chromatogram would be based only on scan data but the one shown is based on both SIM and scan data.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Quant used summed spectra but calls it average
Problem:
When in Apex mode the scale is about 1/10 the intensity of the Average spectra. What is happening is summing. Help dose not cover the fact that it says Average but is really summed. This is both confusing and can make users think that there are saturated ions (based on high intensity). What is desired is that the Average be done (the preferred option) or that the term be changed to Summed (bad but at least accurate)
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Library Editor: Limited JCMP format support - error indicates invalid JCAMP record
Problem:
The Library Editor's import JCAMP feature does not support all JCAMP formats and when a valid but unsupported type is selected there will be an error indicates invalid JCAMP record.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
ATM Configuration shows only Quant commands
Problem:
No place to specify command groups, User Roles, User Validation/Reason associated for Unknowns Analysis commands
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Tune File Last Time Stamp is blank in Reports and in Batch Table when data file acquired with GCMS Acquisition B.07.04
Problem:
When a data file is collected under GCMS Acquisition B.07.04 the Tune File Last Time Stamp is blank in all reports and in Batch Table. This is the result in a defect in GCMS Acquisition B.07.04 which uses an incorrect date format that Quant cannot read. This defect is likely seen in any version of Quant. The defect is only seen in files created under GCMS Acquisition B.07.04
Temporary Solution:
There is no fix for the Quant as the issue is in GCMS Acquisition B.07.04. A post acquisition macro is available for GCMS Acquisition B.07.04 that will reformat the date format so it can be read by Quant and the issue is not seen in GCMS Acquisition B.07.05 or B.07.03 and earlier versions.
Fix Information:
There is no fix for the Quant as the issue is in GCMS Acquisition B.07.04. A post acquisition macro is available for GCMS Acquisition B.07.04 that will reformat the date format so it can be read by Quant and the issue is not seen in GCMS Acquisition B.07.05 or B.07.03 and earlier versions.
Keyword:
One-line Description:
Unknown Analysis - Help file - Relative ISTD Estimation Equation is incorrect
Problem:
The help file in Unknown Analysis indicates that the Hit.basepeak.DeconvolutedArea (or Height) is used when in fact it is the ComponentArea or Height)
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Changing Integration Parameters does not quantitate
Problem:
Changing the integration parameters from the Compound Information window will update response but clears the concentration values . This is inconsistent with manual integration which automatically determines concentration.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Library Methods Max number of peaks ignored by Quant's Create Method from Scan data
Problem:
The Library Method, which is required for the Create Method from Acquired scan data, has a feature to limit number of peaks by Area or height. It is used in Unknowns Analysis but ignored in Quant.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
2 - 3 second delay when changing sample type or level
Problem:
Changes in sample type or level in a small batch (5 samples, ~20 compounds) takes 2 to 3 seconds delay in response even on high-end laptop (i7, 16GB, SSD).
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Undefined cal level gives misleading results
Problem:
User assigned a level code to a an existing calibrator and this new level was not defined in the method. When they analyze the batch there is no error message appears warning that the level is not defined and the table shows old expected concentration of the old calibrator level and gives an accuracy value.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Unknowns Analysis ECM: Actions not recorded in the audit trail
Problem:
The following actions are not recorded in the audit trail: 1) Report generation 2) Unknowns report template used 3) Library change in the method
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Unknowns Analysis ECM: Method History Report and Comparison Results do not show which library changed in library method
Problem:
When a Library is changed there is a delete library and insert library event placed in the audit trail but the name of the Library being removed and the name of the library being added are not shown.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Unknowns Analysis ECM: Commands displayed in the audit trail are not displayed in ATM Configuration
Problem:
Because they are not displayed, cannot configure user validation and/or reason for these commands and reasons cannot be displayed in Method History Report
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
In Quant B.08.00 TDA files are not automatically created on accurate mass data files
Problem:
Quant B.08.00 has a change in behavior to automatically use Archer data, rather automatically create TDA files as was the case in version before B.08.00, access if no converted files exist for accurate mass data. Results using Archer and results using TDA may not be the exact same results in some cases. If TDA files are create using the new interface
Temporary Solution:
None
Fix Information:
Changed in B.08.00
Keyword:
One-line Description:
QuantReport_System_Suitability_B_06_00_Ja.xlsx, ISTD Response % deviation outlier field is not filled with Blue/Red.
Problem:
QuantReport_System_Suitability_B_06_00_Ja.xlsx, ISTD Response % deviation outlier field is not filled with Blue/Red.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
LC ChemStation sample information fields are not mapped to Quant fields correctly
Problem:
Some values from the LC ChemStation sample information dialogue are not mapped to the correct corresponding fields in Quant.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Env_Results.xml & Env_Results_withGraphics.xml report templates do not have a way to show partial compound list and also show ISTDs
Problem:
When a subset of compounds are selected none of the ISTDs, even the one related to the selected compounds, appear in the report
Temporary Solution:
Updated report templates available on demand
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Library Match Scores are different in B.08.00
Problem:
Library Match Scores are different in B.08.00 vs. older versions as the NIST Compatible was changed from False to True for library search settings.
Temporary Solution:
None
Fix Information:
Changed in B.08.00
Keyword:
One-line Description:
Quant ECM: Failed to create method "Created method from acquired scan file" gives Deconvolution failed messages
Problem:
This issue was seen on Profile Scan data collected on a Triple Quad instrument.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Quant ECM: Crash when attempting to extract spectrum from QQQ-profile-scan data
Problem:
Quant ECM: Crash when attempting to extract spectrum from QQQ-profile-scan data
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Quant ECM: Convert Accurate Mass Samples is not active
Problem:
If opening a Q-TOF batch without any converted files, the menu item Tools > Convert Accurate Mass Samples is not active
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Append method from file: Globals no longer uses values in appended file
Problem:
Behavior has changed from Quant B.07.01 to Quant B.08.00 Quant B.07.01: Globals setup values are from the method that was appended Quant B.08.00 - build 590: Globals setup values are from the existing method
Temporary Solution:
None
Fix Information:
Changed in B.08.00
Keyword:
One-line Description:
Action not displayed in User Validation and Reason box
Problem:
Steps to reproduce: 1) Setup ATM Configuration a) Each Required User Role is assigned to its corresponding Command Group b) Batch Analysis requires command reason and Method Development requires command reason and user validation c) One user is in the following groups: QuantBatchAnalyzer, QuantBatchReviewer, and QuantMethodDeveloper. 2) Open an audit-trailed batch in QQQ flavor 3) Attempt to manually-integrate a peak > User Validation and Reason window does not display "Manually integrate compound ..." for Action *'CmdManuallyIntegratePeak' does display in the audit trail for the command name*
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Quant ECM: "Open as Checked Out" for older file revisions open current version
Problem:
Quant ECM: "Open as Checked Out" for older file revisions open current version. This menu item is misleading and should be removed.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Update Retention Time in batch table view: Time it takes to update RT is proportional to the number of samples in the batch - The more samples, the longer update takes
Problem:
Update Retention Time in batch table view: Time it takes to update RT is proportional to the number of samples in the batch - The more samples, the longer update takes
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Noise region is part of an integrated peak
Problem:
Some integrators, such as Agile & Agile2, may define data points that appear to be part of an integrated peak. Such data points usually are near the base of the peak and may extend up the sides of the peak slightly. This is not a defect in the display of the noise region but an issue with the integrator.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Synchronize Navigation from Chromatogram Information window does not work
Problem:
With both 'Synchronize Navigation's turned on (Compounds-at-a-Glance and Chromatogram Information), selecting in Chromatogram Information window does not automatically select the same sample in Compounds-at-a-Glance. Also, selecting in Chromatogram Information does not automatically select the same sample in Batch-at-a-Glance.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
When 'ISTD Dilution' is entered or changed, results (concentrations) are not automatically updated
Problem:
When 'ISTD Dilution' is entered or changed, results (concentrations) are not automatically updated
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Quant ECM: Message displays that batch needs to be saved before reporting
Problem:
Steps to reproduce: 1) Open an exist batch in QuantECM 2) Select Report > Generate --> Message displays Since batch is already saved in ECM and there are no changes, this message should not display.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
If data files are not converted, then 'Noise of Raw Signal' is null
Problem:
In addition: 1) Baseline Draw is null when not converted 2) Baseline Standard Deviation is "1" when converted and "0" when not converted This will be seen in LC data due to lack of automatic "conversion" i.e. creating the MSTree2.bin index file.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Library Editor; Import of Jcamp from chemstation leaves the User Index and Retention Time is blank.
Problem:
Library Editor; Import of Jcamp from chemstation leaves the User Index and Retention Time is blank.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Negative outlier limits/associated values are not flagged by method validation
Problem:
Plates, Capacity Factor - Negative values not flagged (new outliers in Quant B.08.00) Peak Resolution - Negative values flagged prior to B.08.00 but no longer flagged in Quant B.08.00
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Quant Excel template adds border and blank pages in Excel 2013 but not in Excel 2010
Problem:
Some customized Excel template are behaving differently in Excel 2010 vs Excel 2013. In Excel 2013 a border appears that is not seen in Excel 2010. This border extends out to the right by enough to cause a "blank" page to be printed. The border appears to be based on the space needed to show the hidden fields.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
'Save batch' is recorded twice in the audit trail
Problem:
'Save batch' is recorded twice in the audit trail even though the command was only executed once
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Quant Reporting Repair utility creates $CustomerHome$ directory next to the customer home directory
Problem:
Run the 'Repair Quant Reporting' utility installed by Quant in the Start Menu -> Quant -> Quant tools program group. Open Windows Explorer and navigate to the "\MassHunter\ directory and note the new $CustomerHome$ directory.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
QuantECM: Original batch remains 'Checked out' when 'Save as' is done
Problem:
To see the problem: 1) Create a new batch or open an existing batch as 'checked out' 2) Do a 'Save As' and save the batch using a different name 3) Close the saved-as batch (No batches are open in Quant) 4) In ECM, the original batch remains 'Checked out'
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
QuantECM: Method History does not show all report method changes
Problem:
1) In Quantitative Analysis Report Method, Method 1 Value and Method 2 Value for 'Formatting[0]'/'Formatting[2]' display blank instead of report template path when report template changed 2) In Unknowns Method report, Method 1 Value and Method 2 Value for 'DeconvolutionMethod' are blank when LeftMZDelta/RightMZDelta and MZDeltaUnits are changed
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
CC reports using wrong target area if a surrogate follows last internal standard with all the same concentrations are used.
Problem:
In some cases the software will use a surrogate to determine the concentration of the CC and us this to locate the appropriate Initial Calibration level. In this case the level of the surrogate may not match the level of the target compounds and thus the wrong level is used in the report.
Temporary Solution:
Corrected report code is available on request
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Unknowns Analysis ECM: Unable to import a batch when user validation and/or reason is required to Open Batch
Problem:
If Quant compliance is configured to require username/password and/or reason to Open Batch, 'Import Quantitative Analysis' cannot be done in Unknowns Analysis. A message is displayed that a valid user name and password are required, but there is no place to enter this information.
Temporary Solution:
Configure Quant compliance such that username/password and/or reason are not required to Open Batch
Fix Information:
n/a
Keyword:
One-line Description:
ECM Quant : Not all locations are visible when the server has > 100 locations.
Problem:
When there are greater than 100 locations defined on the ECM server the file open/create/Save/Save As dialog box will not display all locations.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
RetentionIndex values in .L are 1000x larger than those in .mslibrary.xml, when save as a .mslibrary.xml library to .L library,
Problem:
If a library in .mslibrary.xml format that contains retention index information is opened in Library Editor and then saved as a .L format the retention index values are multiplied by 1000.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
With User Management and Audit Trail installation, no prompt to enter Reason and User Validation when generating a report
Problem:
This issue happens when the batch is not setup with an audit trail
Temporary Solution:
Workaround is to create the batch with an audit trail
Fix Information:
n/a
Keyword:
One-line Description:
Quant crashes when Create Levels is clicked with no compound selected
Problem:
QQQ Quantitative Analysis crashes when the Create Levels button is pressed with no compounds selected. To see this: 1. Create new Batch and add sample 2. Create new method from MRM data 3. Enable Serial Dilution Toolbar 4. Click Create Levels Button in Method Table Pane with no Compound selected in the Grid
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
QuantECM: Message displayed 'Failed to open data access for sample' after ECM session left open for days.
Problem:
If an ECM Quant session is left open too long ECM will timeout and recover the ECM License. Once this happens Quant fails.
Temporary Solution:
Workaround is to re-open the batch
Fix Information:
n/a
Keyword:
One-line Description:
Quant Create new method from SIM data gives no compounds when NCI file
Problem:
Customer is collecting SIM data in NCI mode and goes to "Create a method from SIM data". No compounds are created.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
IronPython 2.7.6.3 or greater will crash Quant and Unknowns Analysis
Problem:
Quant and Unknowns Analysis requires a version of IronPython prior to 2.7.6.3 and installing 2.7.6.3 or newer versions will result in a crash of Quant and Unknowns Analysis.
Temporary Solution:
None
Fix Information:
Info for Quant B.08.00
Keyword:
One-line Description:
Swap Qualifier With Quantifier - Peak Filter Threshold and Peak filter threshold Value are not swapping
Problem:
When a Swap Qualifier With Quantifier is performed the Peak Filter Threshold and Peak filter threshold Value are not moved from the qualifier to the quantifier as it should.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
Unknowns Analysis Familiarization Guide in B.08.00 release same as B.07.01
Problem:
Unknowns Analysis Familiarization Guide in B.08.00 release same as B.07.01. Some screen shots are not accurate for B.08.00.
Temporary Solution:
None
Fix Information:
Info for B.08.00
Keyword:
One-line Description:
Compounds-at-a-Glance : zero-peaking does not zero peak samples that are not visible on screen
Problem:
Steps to reproduce: 1) Open a batch and view in Compounds-at-a-Glance 2) Select a small number of panes in the pulldown menu such that not all samples are visible. Select 'Organize rows by' Compounds in CAG layout setup 3) Press Ctrl+A to select all panes. Activate manual integration. Press Delete on keyboard or click the zero peak icon 4) Some samples are zero-peaked. The same samples that were not visible on screen are not zero-peaked.
Temporary Solution:
Be sure to scroll over all samples such that they become visible at some point before performing the zero-peak.
Fix Information:
n/a
Keyword:
One-line Description:
QuantECM: Batch status in ECM is no longer 'Checked out' after generating a report
Problem:
Steps to reproduce: 1) Open a batch as checked out > ECM status (in web browser) shows "Checked out" 2) Generate a Quant report for the batch 3) After report generation completes, refresh the web browser > ECM status is null (no longer "Checked out") even though the batch is still checked out
Temporary Solution:
After generating a report always check the batch out again.
Fix Information:
n/a
Keyword:
One-line Description:
Create New Method from MRM data does not respect the
Problem:
It is possible to disabled compounds within a DMRM acquisition method. A disabled compound will have no data collected. Quant does not respect this setting and creates compounds in the method with "*** No Data Point ***" which them need to be deleted manually.
Temporary Solution:
None
Fix Information:
n/a
Keyword:
One-line Description:
PDF Reports do not respect the Area Scale option in the Peak Chromatogram Graphics section of the Quant Report Method
Problem:
PDF Reports do not respect the Area Scale option in the Peak Chromatogram Graphics section of the Quant Report Method. The Excel templates do.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Invalid JJCAMP format in the file
Problem:
In Library Editor b.07.01 SP2 Build 7.1.524.2, if the user attempts to Export JCamp File then Import that JCamp into another session of Library Editor, It fails with the Invalid JCAMP format in the file Required JCAMP record title not found. Steps to reproduce the error: 1. Open demo.l in Library Editor 2. Select one or more entries in the left hand margin 3. File-->Export JCamp file 4. Close Library Editor. Open a new session of Library Editor and attempt to Import JCamp file. This is an important workflow when customers are generating libraries in GCMSD ChemStation, AMDIS and NIST Search utilities and want to import them for use in MassHunter Qualitative or Quantitative Analysis.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Average Qualifier Ratios to via GCMS Acquisition sequence does not work
Problem:
If the sequence has the "Update Qualifier Ratios" field set to "Average" the feature will clear all levels but the one which makes the method invalid. The problem is due to a defect in the script "UpdateRetentionTimeQualifierRatio.automation.quant.script" located in the \MassHunter\Scripts\Quant\Automation folder.
Temporary Solution:
Request updated script from support and install.
Fix Information:
n/a
Keyword:
One-line Description:
Response of peak changes when sample type is changed and RTE Integrator is selected
Problem:
A case was reported where the peak start/stop was moving slightly when the sample type was being changed. Related to an RTE Integrator feature that tries to snap manual integrations to exact scan times and above the signal. Code altered to avoid the snap.
Temporary Solution:
None
Fix Information:
Fixed in B.08.00
Keyword:
One-line Description:
Library Editor - Export to JCAMP does not export library entries with no spectrum
Problem:
Library Editor - Export to JCAMP does not export library entries with no spectrum. JCAMP standard does not defile behavior in such a case.
Temporary Solution:
Information only - not a defect
Fix Information:
Information only
Keyword:
One-line Description:
Values are different from those in previous versions
Problem:
Some values are different in Quant B.09.00 (e.g., Noise of Raw Signal, Signal to Noise, Response, Baseline Draw, Baseline Standard Deviation) compared to previous Quant versions because non-mass spectrometric data that was not previously converted is now being converted.
Temporary Solution:
NA
Fix Information:
NA
Keyword:
One-line Description:
MZ Additional does not work on SIM data
Problem:
Summing SIM signals fails to work. Signal remains the same after adding an additional m/z value and the annotation in the chromatogram fails to show it after the target m/z value.
Temporary Solution:
None
Fix Information:
Fixed in B.09.00