Agilent

Software Status Bulletin


UV VIS ChemStation     G1118AA

Known Problem Report as of Mar 14 2017 10:27AM

Preface
-------------------------
This Software Status Bulletin (SSB) documents all known problems in the software
product designated above. The SSB is derived from Known Problem Reports (KPR)
which result from user problems that have been classified as documentation
problems or software defects. When a KPR is written, an identifying number is
assigned to it, and the KPR is added to the next edition of the SSB.

User inputs that have been classified as Enhancement Requests are not documented
in the SSB. User problems that have been submitted, but that have not been
classified by the time the SSB is generated are not included in the SSB.

How to use the SSB
-------------------------
When you experience a problem with a product, first check this SSB to see if the
problem has been reported already, and if there is a temporary workaround
available for the problem, or if the problem has already been fixed by a new
revision. If the problem is not listed in this SSB then you may wish to report
it to the Response Center or to your field support representative.


To determine if your problem is documented in this SSB, first look in the
Keyword Index section of the SSB. Under each keyword is a listing of one-line
descriptions of related KPRs. If any of these sound like yours, locate the KPR
# in the Known Problem Reports section of the SSB, and read the full KPR. The
KPRs in the Known Problem Reports section are sorted by KPR #.


There are two sections in the SSB:
Keyword Index: This index is categorized by keyword. For each KPR there is a
brief description and a KPR #. A KPR may be associated with more than one
keyword.


Known Problem Reports: This section contains KPRs, with all the available
information relevant to the problem. KPRs in this section are sorted by KPR #.


Keyword Glossary



Autosampler
Communications
Data Acquisition
Data Analysis
Edit
Graphics/Views
Import/Export
Load/Save
Method
OQ/PV
Print
Saving/Uploading

Keyword Index


Keyword: Autosampler

KPR#:105 Pump times incorrect
KPR#:296 XY autosampler stops during offline dissolution testing run.

Keyword: Communications

KPR#:270 Configuration of new instruments provides empty data folders.

Keyword: Data Acquisition

KPR#:17 GPF during dissolution run on Windows 2000 SP2
KPR#:104 Blank might become invalid, when changing the sampling system
KPR#:257 Dissolution run is aborted after clicking on the 'Close' button
KPR#:264 GUI updates are slowed down when running MCA in the Multibath mode.

Keyword: Data Analysis

KPR#:48 'Used Vessels' is not applied to 'Absorbance' and 'Match Factor' results
KPR#:49 Different types of Blank spectra are shown in the Blank Spectrum History
KPR#:96 Changes in Calculation Parameter Variables dialog box are not executed
KPR#:139 Measurement cycle control change does not initiate results update
KPR#:276 ChemStation accepts negative values for Tablet Weight.
KPR#:278 Uncorrect calculation of the Match Factor.
KPR#:747 No access to processing parameters, misleading message box displayed
KPR#:753 Dissolution results not calculated

Keyword: Edit

KPR#:277 The Paste Append item is not available via the Edit menu.

Keyword: Graphics/Views

KPR#:153 Dialog window hidden behind application

Keyword: Import/Export

KPR#:99 Decimals are cutt off in user defined profiles

Keyword: Load/Save

KPR#:47 Blank Spectrum History is not saved in Multibath Dissolution Mode

Keyword: Method

KPR#:106 Blank has to be measured twice after switching to a method with only one lamp.
KPR#:121 Help link for editing post-run sequence lost
KPR#:208 Overwrite Method in Multibath Operator Mode possible
KPR#:751 Red line warning displayed when setting up a control as pre/post run action

Keyword: OQ/PV

KPR#:39 Crosscontamination test fails in the multibath dissolution mode
KPR#:103 Valve Channel not printed on Flow Rate Test Report
KPR#:291 Filename of the dummy bath driver for OQ/PV DDE Interface

Keyword: Print

KPR#:256 System hangs up when a printer is installed during a dissolution run
KPR#:269 Method Information is truncated on printed reports.

Keyword: Saving/Uploading

KPR#:744 Post run action Save Result to ECM offered in Security Pack


KPR#:766  Product:UV VIS ChemStation  G1118AA  TBD

Keyword: 

One-line Description:

Result data corrupted in Multibath dissolution testing

Problem:

When saving dissolution result data in a multibath configuration of concurrent dissolution runs, the data on file are corrupted.

Temporary Solution:

None

Fix Information:

This solution is based on the fact, that a dissolution testing result (in mode multibath dissolution testing) is saved based on the currently selected bath. 
Fixed in code.


KPR#:774  Product:UV VIS ChemStation  G1118AA  B.02.01 SP1

Keyword: 

One-line Description:

Long file names not completely printed in reports

Problem:

In many reports long filenames are truncated. All report templates provided must be updated for at least 32 characters.
For example the dissolution results report displays method and result filenames truncated.

Temporary Solution:

n/a

Fix Information:

Fixed report generator code.


KPR#:787  Product:UV VIS ChemStation  G1118AA  B.02.01 SP1

Keyword: 

One-line Description:

Dissolution average profile not printed

Problem:

When setting up a report style using the Dissolution Average Profile graphical element, an error is reported"Dissolution Average Profile not available".

Temporary Solution:

n/a

Fix Information:

Fixed.


KPR#:805  Product:UV VIS ChemStation  G1118AA  B.04.02

Keyword: 

One-line Description:

Problem loading Dissolution method from Navigation Bar

Problem:

When loading a Dissolution method in Multibath Dissolution Mode over the Navigation Bar the mode is changed to Dissolution Mode. It is not possible to load Samples over the Navigation Bar in Multibath Dissolution Mode. There is no lost data. Loading the method by the File menu is a good workaround.

Temporary Solution:

n/a

Fix Information:

Fixed in code B.04.02


KPR#:105  Product:UV VIS ChemStation  G1118AA  A.06.04

Keyword: Autosampler

One-line Description:

Pump times incorrect

Problem:

A slight deviation between the configured pump time and the actual pump time has been measured. The measured times are 2-3% longer than the pump times set in the software.

Temporary Solution:

n/a

Fix Information:

Fixed in revision A.10.01.


KPR#:296  Product:UV VIS ChemStation  G1118AA  A.08.03

Keyword: Autosampler

One-line Description:

XY autosampler stops during offline dissolution testing run.

Problem:

If the autosampler is configured with rack code 22 and " First increment x
vial number" is selected, the sequence stops at autosampler position
60 during the automated run.

Temporary Solution:

Choose the default setting "First Increment y vial number" or choose
the standard rack code 21.

Fix Information:

No fix planned.


KPR#:270  Product:UV VIS ChemStation  G1118AA  A.05.02

Keyword: Communications

One-line Description:

Configuration of new instruments provides empty data folders.

Problem:

Configuration of new instruments by means of the HP UV-Vis Configuration Editor provides the complete data directory structure (e.g. Hpchem\2\Automat,Data, ...), but the folders contain no data files.

Temporary Solution:

On principle, it is not possible for the Configuration Editor to install or copy files. Hence, you have to add further instruments by means of CD Setup to install all data files on your hard disk.

Fix Information:

No fix planned.


KPR#:17  Product:UV VIS ChemStation  G1118AA  A.09.01

Keyword: Data Acquisition

One-line Description:

GPF during dissolution run on Windows 2000 SP2

Problem:

On systems running Windows 2000 SP2 certain dissolution runs might
abort with a General Protection Fault (GPF) in module KRNL.exe at 0001:6382.
In particular this GPF occurs during long dissolution runs, in the
multibath mode and during dissolution runs with short cycle times
(< 5 minutes).

Temporary Solution:

In order to avoid this problem on Windows 2000 SP2 the following temporary solution is available:

1. Start the regedit and remove/rename the following entry:
   HKEY_LOCAL_MACHINE\SYSTEM\CurrentControlSet\Control\Session Manager\AppCompatibility\ntvdm

2. Reboot the machine in order to make the above setting effective

Fix Information:

Fix under investigation.


KPR#:104  Product:UV VIS ChemStation  G1118AA  A.06.04

Keyword: Data Acquisition

One-line Description:

Blank might become invalid, when changing the sampling system

Problem:

After changing from multicell to any other sampling system the blank becomes invalid, if it hasn't been measured on position 1.

Temporary Solution:

Always measure a blank after switching to the multicell sampling system
before proceeding with any sample or standard measurement.

Fix Information:

Fix under investigation.


KPR#:257  Product:UV VIS ChemStation  G1118AA  A.06.01

Keyword: Data Acquisition

One-line Description:

Dissolution run is aborted after clicking on the 'Close' button

Problem:

Clicking on the Close button of the HP ChemStation window ('x' in the upper right corner) during a dissolution run generates the warning "Automation still in progress. Cannot proceed closing Application. OK". Nevertheless after clicking on OK the current dissolution run is aborted (CP_Autom_2:Aborted by operator).

Temporary Solution:

Do not click on the Close button during a dissolution run.

Fix Information:

No fix planned.


KPR#:264  Product:UV VIS ChemStation  G1118AA  A.05.02

Keyword: Data Acquisition

One-line Description:

GUI updates are slowed down when running MCA in the Multibath mode.

Problem:

When running a MCA method in the Multibath Dissolution Testing mode it may occur that the graphical user interface is not updated correctly. This has no influence on the run or the measurement results.

Temporary Solution:

Reduce the evaluation range of the spectra.

Fix Information:

No Fix planned.


KPR#:48  Product:UV VIS ChemStation  G1118AA  A.06.04

Keyword: Data Analysis

One-line Description:

'Used Vessels' is not applied to 'Absorbance' and 'Match Factor' results

Problem:

Even for 'Not Used' Vessels (Dissolution/Setup Vessel Usage) the disssolution result tables 'Absorbance' and 'Match Factor' are calculated (select View/Dissolution Results -> Absorbance/Match Factor) and printed on the 'Dissolution Results' Report.

Temporary Solution:

n/a

Fix Information:

Fix planned.


KPR#:49  Product:UV VIS ChemStation  G1118AA  A.06.04

Keyword: Data Analysis

One-line Description:

Different types of Blank spectra are shown in the Blank Spectrum History

Problem:

In the 'Blank Spectrum History' window of the Multibath Dissolution Testing mode the blank spectra are displayed in different ways :
A) 
Blanks performed in the 'PreRun' are displayed as absorbance spectra. This is correct, because due to time restrictions a Blank measurement in the Multibath Dissolution mode comprises only one measurement. 
Note: In all other Chemstation modes two measurements are performed during the Blank process. These two spectra are subtracted from each other and the resulting 'Baseline Spectrum' is shown in the Blank Spectrum window. Hence, the Blank Spectrum window shows not the absorbance of the blank but a baseline spectrum

B)
Blanks performed during the 'Dissolution Run' as part of 'Measure Vessels' are displayed as Baseline spectra although only one measurement is performed. This is not correct, since two spectra need to be acquired to calculate a Baseline spectrum.

Temporary Solution:

n/a

Fix Information:

Fix planned.


KPR#:96  Product:UV VIS ChemStation  G1118AA  A.06.01

Keyword: Data Analysis

One-line Description:

Changes in Calculation Parameter Variables dialog box are not executed

Problem:

If 'Based on Equation' is selected in the 'Calculation' section of the 'Evaluation Method Parameters' dialog box and any of the user defined variables in the 'Calculation Parameter Variables' dialog box is changed (adding or removing checkmarks for 'Use' or 'Prompt' or changing the 'Value' of the variable) without changing the specified equation, the new variable settings are not taken into account for the calculation of the dissolution result.

Temporary Solution:

Whenever the variable settings are changed, change also the specified equation in a simple way that does not effect the calculated result. E.g. add zero or multiply the complete equation by a factor of one.

Fix Information:

Fix planned.


KPR#:139  Product:UV VIS ChemStation  G1118AA  A.08.01

Keyword: Data Analysis

One-line Description:

Measurement cycle control change does not initiate results update

Problem:

If an existing dissolution results file is loaded and the 'Define Control for Measurement Cycle' is changed, no recalculation is forced. If a report is printed, the results for the controls during the run can be wrong.

Temporary Solution:

n/a

Fix Information:

Fixed in code B.04.02


KPR#:276  Product:UV VIS ChemStation  G1118AA  A.05.02

Keyword: Data Analysis

One-line Description:

ChemStation accepts negative values for Tablet Weight.

Problem:

The HP ChemStation accepts negative values for Tablet Weight in the Dissolution

Temporary Solution:

Enter only positive values for Tablet Weight.

Fix Information:

No fix planned.


KPR#:278  Product:UV VIS ChemStation  G1118AA  A.05.02

Keyword: Data Analysis

One-line Description:

Uncorrect calculation of the Match Factor.

Problem:

It is possible to calculate the Match Factor in a wavelength range where one or both of the compared spectra have no data points. Executing 'Match Factor' results in the Match Factor Result '*** Passed ***' even with a Match Limit of 1000.

Temporary Solution:

Make sure that the selected wavelength range is covered by both spectra.

Fix Information:

No fix planned.


KPR#:747  Product:UV VIS ChemStation  G1118AA  B.02.01

Keyword: Data Analysis

One-line Description:

No access to processing parameters, misleading message box displayed

Problem:

In the Advanced analytical setup in spectral processing the parameter dialog boxes are not accessible. Rather than displaying the parameter dialog, a message box stating "No Parameters necessary" is displayed when pressing the Setup button.

Temporary Solution:

n/a

Fix Information:

Fixed the 'more parameters given...' message


KPR#:753  Product:UV VIS ChemStation  G1118AA  B.02.01

Keyword: Data Analysis

One-line Description:

Dissolution results not calculated

Problem:

If a dissolution run is executed, invalid results may be displayed during the run. No results are available with the dissolution table and the profile is not updated.

Temporary Solution:

Downgrade to UV-visible ChemStation revision B.01.01.

Fix Information:

Fix in code.


KPR#:277  Product:UV VIS ChemStation  G1118AA  A.05.02

Keyword: Edit

One-line Description:

The Paste Append item is not available via the Edit menu.

Problem:

The 'Paste Append' item in Edit menu is always deactivated (grey).

Temporary Solution:

To execute 'Paste Append' press Shift+Ctrl+V.

Fix Information:

No fix planned.


KPR#:153  Product:UV VIS ChemStation  G1118AA  A.09.01

Keyword: Graphics/Views

One-line Description:

Dialog window hidden behind application

Problem:

If the "Advanced ..." button of the "Bath" tab in the "Product, Bath and Info Method Parameters" dialog is pressed the first time no dialog is visible.

Temporary Solution:

Switch to the bath driver application by means of the operating system's task bar to display the dialog box on top of the open applications.

Fix Information:

Fix under investigation


KPR#:99  Product:UV VIS ChemStation  G1118AA  A.06.03

Keyword: Import/Export

One-line Description:

Decimals are cutt off in user defined profiles

Problem:

If user defined data are send to the ChemStation dissolution application, decimals are cut off. These data are required.

Temporary Solution:

Multiply the questionable data by an appropriate factor to get all significant digits transferred.

Fix Information:

Fixed in Rev. A.07.01


KPR#:47  Product:UV VIS ChemStation  G1118AA  A.06.04

Keyword: Load/Save

One-line Description:

Blank Spectrum History is not saved in Multibath Dissolution Mode

Problem:

In the Multibath Dissolution Mode the Blank Spectrum History is not saved with the result file (*.dr).

Temporary Solution:

Print and/or save the Blank Spectrum History manually after the dissolution run has finished.

Fix Information:

Fix planned.


KPR#:106  Product:UV VIS ChemStation  G1118AA  A.06.04

Keyword: Method

One-line Description:

Blank has to be measured twice after switching to a method with only one lamp.

Problem:

When the user switches from a method with both lamps on to a method with only one lamp on, a new blank spectrum has to measured. After the
first measurement the Last Blank Spectrum window is still empty. The blank has to be measured again, before a sample or standard spectrum can be acquired.

Temporary Solution:

Measure the blank twice after changing from a method using two lamps
to a method using only one lamp.

Fix Information:

Fix under investigation.


KPR#:121  Product:UV VIS ChemStation  G1118AA  A.05.02

Keyword: Method

One-line Description:

Help link for editing post-run sequence lost

Problem:

The problem occurs, when the dissolution run method parameters are edited in the method menu. If one or more actions (apart from 'operator message') have just been set up, the help button cannot be selected for the pre run tab. Instead of help topics for pre run setup the 'Reference Contents' page is displayed.
This error does not occur, if the diisolution run tab or the post run tab is selected.

Temporary Solution:

If the help topics for the pre run tab are needed:

- Click on OK to exit dissolution run method parameters dialog box.
- Open the dissolution run method paramter dialog box again.
- Select Help for the pre run tab.
- If the correct link is available, do not close the help window, before setting up another pre run action.

Fix Information:

Fix under investigation


KPR#:208  Product:UV VIS ChemStation  G1118AA  A.08.03

Keyword: Method

One-line Description:

Overwrite Method in Multibath Operator Mode possible

Problem:

I f an operator made changes to the current loaded method and runs the commands 'New Method' or 'Load Method' a message box 'Current Method has changed! Do you want to save the changes?' is displayed. If the 'yes' button is pressed the current loaded method can be saved and any other method overwritten.

Temporary Solution:

n/a

Fix Information:

Fixed in code B.04.02


KPR#:751  Product:UV VIS ChemStation  G1118AA  B.02.01

Keyword: Method

One-line Description:

Red line warning displayed when setting up a control as pre/post run action

Problem:

When adding the target concentration to a control measure in pre or post run sections of a dissolution test an error is displayed on the message line:
CP3: WARNING: More parameters given than expected by macro zzdcParamDlgMeasControlUpdCellEvent
The same message is displayed, when the menu command Define Control for Measurement Cycle... is applied.

Temporary Solution:

n/a

Fix Information:

The error happened when firing an event by changing a table cell. The event firing mechanism was changed in code.


KPR#:39  Product:UV VIS ChemStation  G1118AA  A.08.03

Keyword: OQ/PV

One-line Description:

Crosscontamination test fails in the multibath dissolution mode

Problem:

The crosscontamination test in the multibath dissolution mode might fail,
if the the flow test has been performed right before the crosscontamination test.
Instead of 100% the result will be around 93%
and instead of 0% the result will be around -7% on all tested channels.

Temporary Solution:

Exit and restart the ChemStation Software in between the
flow test and the crosscontamination test.

Fix Information:

Fix under investigation.


KPR#:103  Product:UV VIS ChemStation  G1118AA  A.06.04

Keyword: OQ/PV

One-line Description:

Valve Channel not printed on Flow Rate Test Report

Problem:

After the OQ/PV the Valve Channel Number is not printed on Flow Rate Test Report.

Temporary Solution:

Note the valve channel on the report.

Fix Information:

Fix planned.


KPR#:291  Product:UV VIS ChemStation  G1118AA  A.06.03

Keyword: OQ/PV

One-line Description:

Filename of the dummy bath driver for OQ/PV DDE Interface

Problem:

The filename of the dummy bath driver 'dummy.exe' which is used for the OQ/PV of the dissolution system does not match with the filename 'dummy16.exe' which is given in the OQ/PV manual.

Temporary Solution:

Use dummy.exe instead of dummy16.exe.

Fix Information:

Since Software Revision A.07.01 includes a 32-bit driver the filename 'dummy16.exe' is a wrong association. Therefore, the filename will remain 'dummy.exe'. Instead the filename will be changed in the next issue of the OQ/PV manual.


KPR#:256  Product:UV VIS ChemStation  G1118AA  A.06.01

Keyword: Print

One-line Description:

System hangs up when a printer is installed during a dissolution run

Problem:

Using Start->Settings->Printers->Add Printer to install a new printer to the PC after a dissolution run has been started may cause the ChemStation software to hangup. In this case the system has to be restarted while all measured data are lost.

Temporary Solution:

Do not add a new printer during a dissolution run.

Fix Information:

No fix planned.


KPR#:269  Product:UV VIS ChemStation  G1118AA  A.05.02

Keyword: Print

One-line Description:

Method Information is truncated on printed reports.

Problem:

The entered text for 'Method Information' in the 'Product, Bath and Info Method Parameters' dialog box is truncated on the printed report, if the defined number of rows for 'Method Information' in the choosen 'Report Style' is not large enough.

Temporary Solution:

If the entered text is truncated, start 'Report Setup File menu and enlarge the text field in your Report Style to a number of rows that fits to the text you would like to enter as 'Method Information' text.

Fix Information:

No fix planned.


KPR#:744  Product:UV VIS ChemStation  G1118AA  B.02.01

Keyword: Saving/Uploading

One-line Description:

Post run action Save Result to ECM offered in Security Pack

Problem:

In the current implementation the Security Pack and the ECM access module can only be installed alternatively: either the Security Pack or the ECM module can be installed.
However, in Dissolution Testing with the Security Pack the action "Save Result to ECM" is offered.
In addition no Warning in the pre interpretation is given!

Temporary Solution:

n/a

Fix Information:

A condition was implemented that will discard the ‘Save Dissolution Result to ECM’ action item