Software Release Bulletin


ICP-MS MassHunter     G7200B

Known Problem Report as of Nov 18 2010 11:15PM


Keyword Glossary




Keyword Index



KPR#: 1225    Product: ICP-MS MassHunter     G7200B    B.01.01.03

Keyword: 

One-line Description:

Meter reading of Option gas line is not displayed after installing Patch 2

Problem:
After installing patch 2, the Option gas flow sets to the correct value as inputted in the tune screen, however the option gas flow meter reading always shows 0%.  Also an incorrect warning error "2416 Warning: Optional gas flow rate is low" appears.
Temporary Solution:
n/a
Fix Information:
The defect is fixed with B.01.01 Patch 003.

KPR#: 1226    Product: ICP-MS MassHunter     G7200B    B.01.01.03

Keyword: 

One-line Description:

High Li RSD after 7700s High Sensitivity Autotune

Problem:
After Ion Lens Autotune on 7700s Hot Plasma, sometimes Li RSD is high (like 6 - 8%).
Temporary Solution:
n/a
Fix Information:
The defect is fixed with B.01.01 Patch 003.

KPR#: 1227    Product: ICP-MS MassHunter     G7200B    B.01.01.03

Keyword: 

One-line Description:

"NaN" are shown in DA-Calib pane

Problem:
In the calibration pane, "NaN" are shown even it should show numbers.
And, some blank column should also show numbers. (e.g. Na)
Temporary Solution:
n/a
Fix Information:
The defect is fixed with B.01.01 Patch 003.

KPR#: 1228    Product: ICP-MS MassHunter     G7200B    B.01.01.03

Keyword: 

One-line Description:

Can't save and print replicate data in QC Tune Report

Problem:
Can't save and print replicate data in QC Tune report. The QC Tune report is also over-written everyrun. Can't regenerate QC report.
Temporary Solution:
n/a
Fix Information:
The defect is fixed with B.01.01 Patch 003.

KPR#: 1229    Product: ICP-MS MassHunter     G7200B    B.01.01.03

Keyword: 

One-line Description:

Auto RF Matching can't find correct matching

Problem:
When auto RF Matching is run, the algorithm can't find correct matching value
Temporary Solution:
n/a
Fix Information:
The defect is fixed with B.01.01 Patch 003.

KPR#: 1231    Product: ICP-MS MassHunter     G7200B    B.01.01.03

Keyword: 

One-line Description:

Number order of repetitions is not sequential when you have a large number of repetitions

Problem:
When you display a large number of replicates (e.g. >10) in the Batch-at-a-Glance data table, the numbering of the replicate columns is not sequential in the view.
Temporary Solution:
n/a
Fix Information:
This defect is fixed with B.01.01 Patch 003.

KPR#: 1232    Product: ICP-MS MassHunter     G7200B    B.01.01.03

Keyword: 

One-line Description:

Bkgrnd doesn't work in semi-quant when changing the order of the data for Background correction with Specify Process Order

Problem:
When one or more samples are set to Bkgnd type in the batch table, the counts of the samples after the Bkgnd sample will be recalculated (background corrected) after executing Process Batch.  However, if the sample list process order is changed using Specify Process Order, the files that follow a Bkgnd sample type are not recalculated (background corrected) when Process batch is executed.
Temporary Solution:
n/a
Fix Information:
The defect is fixed with B.01.01 Patch 003

KPR#: 1233    Product: ICP-MS MassHunter     G7200B    B.01.01.03

Keyword: 

One-line Description:

ISTD concentration settings are changed without the user input

Problem:
When no peak is recognized as analyte or ISTD peak in a calibration standard data, the conc level of an ISTD would be overwritten by the level conc of an analyte.
Temporary Solution:
n/a
Fix Information:
The defect is fixed with B.01.01 Patch 003.