Software Status Bulletin
OpenLAB CDS ChemStation Edition M83xxAA
Known Problem Report as of Sep 25 2023 4:00AM
Preface
-------------------------
This Software Status Bulletin (SSB) documents all known problems in the software
product designated above. The SSB is derived from Known Problem Reports (KPR)
which result from user problems that have been classified as documentation
problems or software defects. When a KPR is written, an identifying number is
assigned to it, and the KPR is added to the next edition of the SSB.
User inputs that have been classified as Enhancement Requests are not documented
in the SSB. User problems that have been submitted, but that have not been
classified by the time the SSB is generated are not included in the SSB.
How to use the SSB
-------------------------
When you experience a problem with a product, first check this SSB to see if the
problem has been reported already, and if there is a temporary workaround
available for the problem, or if the problem has already been fixed by a new
revision. If the problem is not listed in this SSB then you may wish to report
it to the Response Center or to your field support representative.
To determine if your problem is documented in this SSB, first look in the
Keyword Index section of the SSB. Under each keyword is a listing of one-line
descriptions of related KPRs. If any of these sound like yours, locate the KPR
# in the Known Problem Reports section of the SSB, and read the full KPR. The
KPRs in the Known Problem Reports section are sorted by KPR #.
There are two sections in the SSB:
Keyword Index: This index is categorized by keyword. For each KPR there is a
brief description and a KPR #. A KPR may be associated with more than one
keyword.
Known Problem Reports: This section contains KPRs, with all the available
information relevant to the problem. KPRs in this section are sorted by KPR #.
Keyword Glossary
Keyword Index
KPR#:2022
Signals from different LC modules show a shift in the Online Plot
KPR#:2133
Online spectra not available for some instruments
KPR#:1840
Instrument communication error may cause incorrect injection time stamp on subsequent run
Keyword: 1100 Control , 1200 Control
KPR#:2133
Online spectra not available for some instruments
Keyword: 1100 Control,1200 Control
KPR#:89210
Online spectra not available for some instruments
KPR#:2133
Online spectra not available for some instruments
KPR#:1892
Missing error handling for errors during fraction collection
KPR#:1894
Manual Fraction Collection: Delay time not applied
KPR#:1926
Multidraw injection volume >1000ml cannot be entered from the context menu.
KPR#:1928
Stopping a run is sometimes executed as Abort
KPR#:1968
LC Column information is not transferred to the data file
KPR#:100941
In ChemStation "CP Macro" errors are added as logbook entry for Fraction Collectors, when a run has been completed
KPR#:339647
The MSD goes not ready (NRDY) and the sequence and queued sequences are aborted
KPR#:354548
After an Upgrade to C.01.10, no Fraction Tickmarks shown in the Online Plot
KPR#:1720
Cluster information can be overwritten from the offline session
KPR#:1840
Instrument communication error may cause incorrect injection time stamp on subsequent run
Keyword: 1200 Control , Method Scouting Wizard
KPR#:1720
Cluster information can be overwritten from the offline session
KPR#:1623
35900E - Error The Current Event Name is the same for a special event
KPR#:1677
Cannot start two instruments using the two channels of one ADC at the same time
KPR#:1702
35900E A/D converter cannot be split into two instruments from a ChemStation client
KPR#:1721
View ACQ method from Data Analysis does not list acquisition parameters for the 35900 AD converter.
KPR#:2073
Activity log for 35900E will truncate last leter in errors or status messages
KPR#:92291
Sequence stops when pushing the 35900 stop button during sequence
KPR#:2121
35900E ADC-instrument cannot be configured with RC.Net driver after reinstalling the OpenLAB CDS application
KPR#:1412
6890 RC.NET driver: instrument diagram not updated to show configuration
KPR#:1464
Addition of new columns to the column catalog (database) is not possible after upgrade
KPR#:1933
Data file of sequence with manual inject of 6890 may be incomplete
KPR#:1993
Only one 6890 instrument with GPIB connenction can be configured on one controller
KPR#:2003
Migrating a classic 6890 instrument to RC.Net does not work smoothly after upgrade
KPR#:16
Vial Missing Error Message
KPR#:17
How to Change Column Dimensions
KPR#:18
If One Signal Needs to be Saved, it Must be Assigned as Signal 1
KPR#:20
Instrument Diagram Does Not Match Installed Column in Method
KPR#:22
In 68XX driver revision 5.02 - 6.11, it is not possible to Modify a Method During a Run to Extend the Run Time or Change the Oven Temp
KPR#:23
Info in Edit instrument parameters\Columns is incomplete
KPR#:31
Ignore missing vial option is not present in GC
KPR#:33
Instrument Setup graph has incorrect column legend
KPR#:38
ALS errors abort sequence; no Retry/Abort dialog box option offered with 68xx RC.NET Driver
KPR#:55
Opening 68xx RC.NET GC types result in 'FetchCPVar cannot talk to application'
KPR#:59
In 68XX Driver Version 5.02, it is not Possible to Start the 6890 RC.NET from the Front Panel Start Key
KPR#:66
CIS4 inlet is not recognized with 6890 GC RC.NET
KPR#:83
Only Fast Plunger Speed is available for PTV Solvent Vent multiple injections
KPR#:87
IQ1.CFG file of 6890/6850 instrument corrupted after regional setting change
KPR#:89
Incorrect injection volume shown in method report if PTV solvent vent is used
KPR#:92
Instrument Diagram Does Not Match Column 2's Inlet Configuration
KPR#:94
68xx Method Download Times Out
KPR#:95
"Method modified after loading" occurs when running a sequence (6890 RC.NET)
KPR#:98
GC Column Catalog File Cannot be Opened When Using the 68xx Driver Version 6.10
KPR#:99
Sequence continues while GC temperature is not ready
KPR#:100
A Method Can be Created Which Will Not Download to the Instrument and the Sequence Does Not Run
KPR#:101
GC Back Injector Not Present in Edit Method Screen
KPR#:102
Validation failed shown on left corner of Resolve Method screen
KPR#:103
Injector Panels in GC Edit Method Show as Uninstalled When an Injector is Installed.
KPR#:105
Plunger Speed is Not Saved with 6850 GC Methods
KPR#:106
Opening the GC method parameters to only look (no change made) and selecting OK to close stops the sequence or single run
KPR#:109
Method Contents Report Displays the Incorrect Configuration of Front and Back Detectors for 6890 GC
KPR#:113
Cannot Run a Packed GC Column with a 6890
KPR#:116
Pulsed Split/Splitless Injection Does Not Wait For Ready on a 6890.
KPR#:117
In 68XX GC Driver Version 6.01, 2 Injections are Made When Using the PTV
KPR#:119
Vial left in gripper; no message shown to alert user about the problem
KPR#:120
6850 vial numbers are not consistent with standard numbering system.
KPR#:122
Method Resolution Changes Oven Temperature Program
KPR#:123
Column info is missing with 6890
KPR#:124
Method Modification Flag is Set after Edit Method: Occurs on first connection or after upload from GC.
KPR#:128
Method Ddownload Failure Occurs Due to Column Calibration Limits
KPR#:129
Instrument actual value for COC inlet does not display correctly
KPR#:130
Instrument actual value for anode purge flow on an ECD does not display correctly
KPR#:132
6850 Support of Column Compensation
KPR#:134
Easy SamplePrep reset of one or any group of resources creates log entry: 'User reset all resources'
KPR#:136
Data Files do not Contain the Correct Signals if the Front Inlet is Connected to Back Detector
KPR#:137
Slow download speed from GC Datafiles stored on network drive
KPR#:138
C.01.06: 6850 classic driver will not accept valid vial locations in Run Method, Sequence, or Easy Sequence
KPR#:140
The Oven Ramp table does not display correctly in IR with Acquisition method selected.
KPR#:141
6890 PTV Inlet not recognizable in OpenLAB EZChrom
KPR#:142
Multiple errors when the 6890 doesn't have any inlets
KPR#:143
Sequence table Inj Vol Override goes to wrong stops for 6890/7683: C.01.05, A.04.05, C.01.06, A.04.06
KPR#:145
Often get two or more ChemMain.exe's running following shutdown of Control Panel or online session.
KPR#:146
PDID Valco Detector isn't recogonized in the RC.net driver in OLCS EZChrom
KPR#:147
6890N with 7693 barcode scanning checksum mismatch does not return vial to tray and skips a vial.
KPR#:148
Customer can't install colum information from catalog if column description is blank
KPR#:149
EZChrom: Installation of GC Drivers A.03.02 breaks a 7697 configuration if installed
KPR#:150
Setting a Hold time of 0.0 does not work correctly with GC6890
KPR#:155
Instrument session continues to run in background after session is closed.
KPR#:159
6890 online help has the bad strings in Injector page.
KPR#:163
6890, 6850 session start up hangs or is very slow to start up with Windows 10 version 1511 as the operating system
KPR#:336
GC instrument Help page cannot be displayed
KPR#:1239
7890A Runs cannot be started until the GC and injector are idle, view GC Status
KPR#:2126
GC Backflush Wizard does not work properly in C.01.07 SR1 HotFix01 [113]
KPR#:1840
Instrument communication error may cause incorrect injection time stamp on subsequent run
KPR#:2041
Unprivileged users can change instrument setpoints using the "GC Instrument Diagram"
KPR#:5
Error Message when certain methodsare loaded onto an offline configured 7890.
KPR#:15
Shutdown events are not displayed on the software Status screen
KPR#:17
Barcode mismatch error
KPR#:26
Sample Prep Program with only a 7693A Back Injector does not save
KPR#:37
ALS Errors Abort Sequence; no Retry/Abort dialog box option offered
KPR#:54
Method edit: Must click twice with the mouse on each data field to change number
KPR#:67
ChemStation does not show "Hydrogen Safety Shutdown" in 7890A status
KPR#:75
7890A valve names added in the config screen but are not available in the method
KPR#:76
7820 method printout includes parameters and setpoints unavailable for PP inlet
KPR#:77
7820 method printout inserts negative number for Septum Purge Flow, invalid parameter
KPR#:80
If 7890 gets into an error status (flame off) this is NOT reported within the 7890 status window
KPR#:84
Read barcode just before injection and use barcode as sample name in report
KPR#:85
7890A: Column part number is not included in the method report
KPR#:90
Partial data collection is not available on the signals screen
KPR#:91
Need ability to do Digital switching for detector signals in Runtime events
KPR#:93
Read barcode just before injection and use barcode as the sample name in the report
KPR#:98
Park during sequence with Overlap causes unresponsiveness
KPR#:100
7890GC: inlet pressure unit inconsistency - method audit trail vs instrument setup
KPR#:103
RTL: New calculated pressure doesn’t update column 2 - 6 on current 7890A methods
KPR#:107
New Abort Behavior
KPR#:115
7890A: Shutdown events are not published to the software GC Status screen
KPR#:117
Upgrading B.01.03 to B.04.02 GC ChemStation: Addition of new columns to the column catalog (database) is not possible; error occurs
KPR#:118
Electrometer on/off not needed for FID or NPD
KPR#:124
Freeze Signal Value and Resume Signal Value missing from runtime events
KPR#:126
The Override Column Max Temp Checkbox Behavior
KPR#:127
Cryo Trap Run Time Event Support
KPR#:132
Error in flow setpoint upload when method has more than 100 events
KPR#:134
Running a sequence requires the arm from the tray, G4514A to be in the UN-PARKED position
KPR#:137
Extended Parameters does not disable the BCR when no hardware is present
KPR#:139
Switched and standard mode for septum purge not available on MMI
KPR#:140
Negative Split Splitless inlet Pressure values during Post Run Backflush
KPR#:154
Column compensation menu is available during the sequence
KPR#:166
Column: need to include Ramped and Constant pressure and flow modes as selection
KPR#:167
GC goes offline momentarily because of SOAP Exception
KPR#:170
No indication of inlet pressure shutdown in software status with 7890A
KPR#:174
GC Keypad Lock Behavior
KPR#:182
GC Instrument Status for Detectors
KPR#:189
It is not possible to set a positive pressure when column outlet is set to "other"
KPR#:190
FID Flame out; no feedback in software
KPR#:195
uECD and TCD report printout for makeup flows is wrong
KPR#:200
GC continues when STOP on 7820 is pressed
KPR#:203
Open GC edit parameters and close it without changes; method modification flag is set
KPR#:204
7890A logbook shows 7693A turret, plunger errors as GCI1 did not start a run
KPR#:209
Unable to set equilibration time for LTM columns from Edit Parameters
KPR#:211
Changes made to 7890 method from GC instrument diagram result do not load.
KPR#:223
Backflush icon is not displayed when the outlet of the column configuration is set as "Other".
KPR#:227
When running Backflush from ChemStation with segmented columns only the in-segment is taken into account in the calculation.
KPR#:228
The display of the parameter setting screen of the method setup is small.
KPR#:229
After editing the GC method and selecting ok - validation error message is displayed.
KPR#:230
7820 Back Injector Only Resets Injection Source to Manual
KPR#:238
PCM Inlet Heater Shutdown
KPR#:247
Multiposition valve resets to position # 1 between injections
KPR#:252
MMI in Split or Pulsed Split Mode does not allow editing of Total Flow value
KPR#:266
The GC Maintenance View May Not Change Until the Software is Closed and Re-Opened
KPR#:267
Description of How to Delete Columns from the Inventory
KPR#:268
Behavior of the Maintenance Screen When Adding Modules to the GC
KPR#:269
ALS Behavior When Aborting a Sequence
KPR#:270
7890B GC Column Maintenance Counter Behavior
KPR#:271
Update GC Column Catalog dialog is limited to 10 items; no way to display the rest (...)
KPR#:272
Method Differences Viewer Behavior with Scanned Columns
KPR#:273
7890B GC Method Compatibility
KPR#:274
Logbook Messages for Scanned Consumables Items
KPR#:275
The Syringe and Liner are Not Shown in the Method
KPR#:276
The PC and GC Clocks Need to be Synchronized to Account for Daylight Savings Time
KPR#:277
Logging of Injection Tower Location
KPR#:279
Explanation for How Default Early Maintenance Feedback Counters Work
KPR#:280
7890 back injector only sequence is missing Acq.txt and acq.macaml in data directory
KPR#:281
The 7890 GC Goes Offline When Non-English Characters are Entered into the Column Information
KPR#:282
Instrument Actuals Show "GC is Offline" Quickly at the Method or Sequence Start, then Returns Back Online and the Method Does Not Run
KPR#:283
78xx GC configuration fails; Failed to Create Instance
KPR#:284
Data not collected, sequence stopped when vial removed from turret after injection
KPR#:285
C.01.06: Import eMethod or Export eMethod freezes system
KPR#:287
0-volume blank run in overlapped sequence causes injection
KPR#:291
Backflush Wizard never advances past Ready to add backflush screen
KPR#:293
7890A ALS Recoverable Error Actions cannot function after upgrading to latest GC driver build on OpenLAB CDS before A.02.01
KPR#:294
Standard and Easy Sample Prep programs run incorrectly in a calibration sequence (run for unknown samples only) when overlap is involved.
KPR#:295
7890B configuration, model number serial number and IP address change after restart
KPR#:296
Makeup Gas mode terminology should be consistent with the GC front panel
KPR#:297
Agilent Solvent Elimination Wizard issue
KPR#:299
Clicking OK after making changes to Events does not save the changes made to the events.
KPR#:300
OL CDS ChemStation C.01.06 does not download the run time events to the 7890 correctly.
KPR#:301
Can't select uECD makeup mode to "Does not affect Makeup Flow"
KPR#:302
Can't save "Included in Makeup Flow" on uECD setting screen.
KPR#:303
GC7890B IP address changes when loading default method
KPR#:305
7820 GC: Method modification flag set after starting a sequence or method
KPR#:308
Method printout missing column parameters: flow or pressure, setpoint On, (Initial), and post run
KPR#:309
There is an incorrect character when displaying temperature information in the method report in Japanese
KPR#:310
MMI Cryo Fast Cooldown not saved
KPR#:311
Method is always seen as modified, nothing can be added to the queue.
KPR#:312
Oven Slow fan will be off after method is downloaded the slow oven mode on to 7890
KPR#:313
During a pulsed splitless run, the purge time shorter than the pulse time cannot be entered.
KPR#:314
Method print out with 7890 configured does not show total GCc run time.
KPR#:315
There is no "Inlet Septum Purge" event in Runtime Events table - EZChrom
KPR#:316
External Device "Not Ready" ignored by OpenLAB CDS ChemSataion and EZChrom
KPR#:318
Aux Flow Limited to 150 mL/min even when pressure is less than Maximum
KPR#:319
Post Run Flow Rate has default that out of range and value seems not to be stored
KPR#:320
Method User Interface does not correctly load the content of a method.
KPR#:321
Sample Prep Audit Trail is not logging changes.
KPR#:323
Sample Prep steps of previous method are included in printout of methods saved from SP methods, if checked but no steps are specifed.
KPR#:324
Does not delete previous open GC method event table when load method. Wrong injections can be created.
KPR#:325
Method download to GC while in sequence causes it to disconnect and sequence stops.
KPR#:326
7890 method print doesn't include post run time
KPR#:333
Elite: Overlap injection no longer working on EZChrom Elite with GC Drivers A.03.02
KPR#:335
Crash of Software after check "Show Threshold Value as remaining time" in EMF Options
KPR#:339
Not allowed to set proper GC column max temp because it is less than 150degrees.
KPR#:340
FID signal subtraction setting changes spontaneously from "Nothing" to "Column Compensation" during run.
KPR#:341
"Extend Run" in GC settings does not work correctly
KPR#:342
Cannot configure a 7890B on an existing client server system
KPR#:343
Copy to Clipboard Function in Instrument Actuals throws exception error
KPR#:344
The column flow calculator calculation is off by almost 3x the flow rate for the system
KPR#:345
Sample Overlap settings can not be edited after driver update, but are still part of the method.
KPR#:347
Incomplete Method Parameters for xCD in Acquisition Method Report
KPR#:349
Actual flows displayed by CDS aren't the same as the setpoint flows
KPR#:350
Print method does not include all MMI setpoints: Purge time, Post Run Temp, Time, Flow
KPR#:351
Extend run does not extend the acquisition of the Intuvo 9000 GC
KPR#:352
Cannot Add User-Defined EMFs on Intuvo 9000 GC
KPR#:355
ChemStation Instrument actuals showing D2 Bus Temperature with no D2 bus installed
KPR#:356
Method print is showing Solvent wash mode as A-A2, B-B2 when it should be A,B since it is a 16 vial turret
KPR#:357
Inlet pressure setpoint displayed assuming all thermal zones are turned ON, results in confusing mismatch when heated zones are Off.
KPR#:360
Method Editor for PP inlet & Packed Column incorrectly allows both flow and pressure setpoints to be entered
KPR#:361
ChemStation can lock up when changing Intuvo Bus Flow Chip types with ChemStation session running.
KPR#:367
Right Clicking on Method Editor makes UI hide behind Instrument Screen
KPR#:368
Incorrect Values in 9000 Pressure/Flow Calculator
KPR#:369
Make changes to guard chip on Intuvo 9000 GC, audit trail just says temperature changed
KPR#:370
Backflush Wizard illogically sets trial backflush times after the elution time of last peak of interest
KPR#:371
Should not be able to download a method while a in a GC run or sequence.
KPR#:372
GC Dashboard does not get populated for ChemStation
KPR#:374
Cannot download GC set points to 7820 configured with a Nickel Catalyst heater and no inlets.
KPR#:375
No indication that the AutoSampler tray is in park mode.
KPR#:376
The Local User Interface(LUI) on the Intuvo 9000 GC shows long name for connected data system.
KPR#:377
For both the 7890 and Intuvo 9000 there is no change in the UI to show the user that a rack is not present.
KPR#:378
Post Run Backflush. The Guard Chip temperature was not set to the Post Run Oven Temperature.
KPR#:379
Location of Intuvo 9000 GC System Health Report should be moved
KPR#:380
Link to Intuvo 9000 GC help does not work.
KPR#:381
Event drop downs are black when creating a new method
KPR#:382
Cannot install B.01.03 on ChemStation C.01.07 SR4 by Welcome.html
KPR#:383
PCM will not control in constant flow.
KPR#:388
RTL for OpenLAB CDS 2.X does not calculate the pressure correctly when locking or re-locking the method when setting a Target retention time.
KPR#:422
HS Sleep/Wake mode should sync with GC Sleep/Wake mode
KPR#:425
Script error when starting help for "Runtime Events"
KPR#:431
Overlap injections on the 9000 uses only the injection volume from the first injection for subsequent injection.
KPR#:435
GC Method Resolution results in “Object Reference” error message and occasional crash
KPR#:439
Downloading method parameters (like oven temperature) on a 6890A does not work with OpenLab CDS 2.4
KPR#:472
Chromeleon: Intuvo trigger two instrument notification for configuration changed
KPR#:473
Chromeleon: GC: Change of column length (and related parameters like flow etc) via calibration in online method sometimes reported in audit trail sometimes not
KPR#:474
Chromeleon: GC: Column length can only be calibrated once in Online Method Editor
KPR#:475
Chromeleon: Setting Liner on Intuvo triggers two config change messages in audit trail
KPR#:476
8860 Open Lab CDS EZChrom compact license is invalid.
KPR#:477
Selecting Partial Data for the 6890 causes the retention times and areas to be wrong.
KPR#:478
6890 - cannot set post run flow for packed column
KPR#:479
Chromeleon: GC 6850A and 6890A: GC online method is not editable
KPR#:480
Chromeleon: GC 6890A: Wrong temperature shown in chromatogram for signal GC_OvTempe_Act
KPR#:481
Chromeleon: GC 6850A: Cannot create instrument method
KPR#:482
Chromeleon: GC8890: Current Signal Value in 'Blank evaluation' shows incorrect value in Method editor
KPR#:483
Chromeleon: GC8890: Blank evaluation setpoints missing in method report
KPR#:484
Signal Scaling is incorrect for the 6890
KPR#:485
Chromeleon: Intuvo: Help button in window 'Configure Agilent Intuvo 9000 GC' opens 78xx Series GC Help
KPR#:486
Chromeleon: GC8890: Issues during sequence execution
KPR#:487
Chromeleon: Changes made in 68xx Edit Online Method not applied or reported
KPR#:488
Error when loading GC method without Blank Run Analysis setpoint
KPR#:489
Chromeleon: Corrected column length will not save after manual method resolution
KPR#:490
Chromeleon: Signals from previous method persist in new method
KPR#:491
Chromeleon: ECD Help page cannot be accessed
KPR#:492
Chromeleon: All 8860 signals not displayed or available in report
KPR#:493
Chromeleon: Error occurred at several points of operation
KPR#:494
Chromeleon: GC6890: Front Injector parameter change in Online method not applied
KPR#:495
Chromeleon: GC6890A: Method download fails
KPR#:496
8860 on Chromeleon -Unhandled exception in Chromeleon Driver Host (CmDDKHost): AgilentTechnologies.DdkIcfAdapter|Agilent
KPR#:497
Chromeleon does not recognize the tray with Back Injector installed.
KPR#:498
Chromeleon: Automatic de-activation of not supported feature blank evaluation cannot be saved
KPR#:499
Getting multiple Errors in the PC Event Log for AgGC7890 after disconnecting and Reconnecting LAN cable.
KPR#:500
GC6890: Valid vial ranges are not correct for 6890 in some cases
KPR#:501
Chromeleon: GC6890A: Error 'The communication FIFO of the DDK driver is corrupted.' after connecting and disconnecting several times.
KPR#:502
When click the "Apply the Option" in GC Chromatographic Column Directory under Column Configuration, an exception occur on Openlab 2.4
KPR#:503
Upgrade installation from GC B.01.03 to 3.0 SR2 fails to install Ag7697WS.dll (and .pdb)
KPR#:504
Chromeleon: GC6850A: Changing method parameter creates method inconsistent to configuration
KPR#:505
Chromeleon: GC6850: Partial Data option causes issues during sequence execution
KPR#:506
Chromeleon: GC6890: Unexpected audit trail entries during endurance run
KPR#:507
SR1 for the GC3.0 drivers breaks the connection between an 8890 and 7000D transferline temperature reading in MH ACQ 10.
KPR#:508
Not able to download flow setpoints to 6890 VI inlet.
KPR#:509
Chromeleon: Incorrect conversion and missing unit when using pressure setting bar
KPR#:510
68xx: Chromeleon: System with G4513 won't make 1.1 uL injection
KPR#:511
Chromeleon: System crash: system slow and then crashing
KPR#:512
8860 with Headspace/PAL does not consume correct VL license for OpenLAB 2.4
KPR#:513
B,04.03 SP2 will not connect to an 8890 GC.
KPR#:514
Chromeleon: 6890GC Oven temperature disabled in method - Run does not start.
KPR#:515
3.0 SR2 driver fails SVT on OL CS C.01.09.144
KPR#:516
OLCS C.01.09 with Driver 3.0.650 will not launch any instrument.
KPR#:517
Chromeleon: SVT Chameleon 2.1.30 Fails
KPR#:518
Function_build number not display in SVT report in EzChrom and OpenLab CDS 2.4
KPR#:519
8890: the MS transfer line temp displayed is 0 C but its actual temp is 250 C.
KPR#:564
Calculated Pressure for Relock is not updated for the 6890 in C.01.10.
KPR#:1196
7890A: Shutdown events are not published to the software GC Status screen
KPR#:336
GC instrument Help page cannot be displayed
Keyword: 78xx Driver , 68xx Driver
KPR#:336
GC instrument Help page cannot be displayed
KPR#:1660
No special characters should be used for names of methods, data files or sequences
KPR#:1701
Terminal Service can close under high workload
KPR#:1747
Queue Method vial range reflects range of loaded method, not the method browsed to
KPR#:1811
7890 back injector only data file is missing acq.txt and acq.macaml in the data directory
KPR#:1841
Run time for unsynchronized modules is shortened
KPR#:1842
Memory leak in RDPShell.exe may cause significant performance decrease on OpenLAB CDS ChemStation AICs
KPR#:1853
GC7890 acquisiton methods cannot be reported with Intelligent Reporting of OpenLAB CDS ChemStation
KPR#:1960
Chromatograms of manually stopped runs are shorter than expected
KPR#:1962
LCDIAG.REG content incomplete
KPR#:1964
Data system uses the currently loaded method to validate the vial range in the Sequence table
KPR#:1969
Using FIle> Snapshot causes the software to become unresponsive on systems with a Varian Prep Star module
KPR#:2067
Missing data points in chromatogram can be perceived as retention time shift.
KPR#:2072
Incorrect XML import of MultiSampler vial positions
KPR#:2077
Duplicate or corrupted files or unresponsive software after stop/abort
KPR#:2102
Order of data points in instrument signals is inverted
KPR#:2108
ChemStation Scheduler hpsched.exe fails to start on Windows 2008 Server R2 SP1
KPR#:2110
Error 17923 occurred for both LC and GC
KPR#:2136
Order of data points in instrument signals is inverted
KPR#:2147
Sequence Start button is enabled when sequence is paused.
KPR#:2159
G7100A Wait time after loading new method is applied to every run
KPR#:2186
"Error 41092 occurred" is shown in the system activity logs
KPR#:89225
Calibration run was canceled during injection, File/Path error message – could not save method with data shown
KPR#:105304
ChemStation does not set the Device Master-Runtime correctly to the ELSD RC.Net driver
KPR#:123728
Cannot start the FIA Method in the Single Sample Information down menu with G7129A Vial sampler
KPR#:124268
Intermittent signal lost of HPLC with G7115A
KPR#:132730
Fractions missing in Data File after Instrument Error
KPR#:149080
Instrument driver error crashes the ChemStation
KPR#:168241
Intermittently the signal is missing or saved in wrong data folder for runs not started from the ChemStation
KPR#:172406
Intermittently the signal is missing or saved in wrong data folder for runs started at GC
KPR#:218783
Counter used for single data file naming does not increment correctly with multiple runs in C.01.08 [216]
KPR#:232469
Counter used for single data file naming does not increment correctly with multiple runs in C.01.08 [216]
KPR#:239891
Data saved in reverse order for multiple single runs specified in Sample Info
KPR#:399554
Missing error method in sequence parameters
KPR#:451930
Snapshot does not work: Error "Data file does not exist"
KPR#:474709
When doing Fraction Collection without sampler it can happen that the fraction start locations is wrong
KPR#:519791
Performance problems with OpenLab ChemStation C.01.07 and C.01.08 with 8890 and 8860 GC instruments
KPR#:523782
Column table continues to grow each time a column is changed in GC and LC
KPR#:524638
Column table grows over time when changing column settings
KPR#:603801
Sample naming counter issue
KPR#:694563
Commands setup in the Command Scheduler window also used in the Offline session
KPR#:770071
Cannot remove custom wellplate definitions
KPR#:1996
A missing vial error in an LC system with two samplers can be irrecoverable
KPR#:1763
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:36
Windows Domain tab is grayed out in OpenLAB Server Utility
KPR#:39
Cannot stop Shared Serviceshost service on computer after changing the Domain Account password
KPR#:119981
Users with privilege of view and run instrument can take over instrument used by another user
KPR#:128072
Microsoft Windows 10 update (build 1703) may result in bluescreen
KPR#:129420
Software Verification Report does not include Hotfix04 in the base revision number
KPR#:149886
EMF help information not available using the Instrument Control Framework
KPR#:149890
Do not install msi files by double-clicking the file
KPR#:165543
Administrative Break Session lock functionality from ChemStation Admin tool is not working correctly
KPR#:217785
Installation Qualification Test fails when a newer RapidControlIF.dll is installed than has been expected by the LC Driver
KPR#:274555
Installation Qualification Test fails when A.02.19 SR2 is installed on OpenLab Chemstation Edition C.01.09
KPR#:431252
Driver release note states wrong firmware revision for G7104C
KPR#:764090
Secure File IO cannot be set
Keyword: Administration,User Interface
KPR#:267323
Activity Log in OLCP does not record the Application name when login from the Web UI
KPR#:683
Unable to download temperature setpoint to Gasifier II
KPR#:1749
Method Audit Trail does not include change of Injection source when HS/PAL is no longer configured
KPR#:1819
UIB II - set a timetable and control "Method Audit Trail"
KPR#:1876
Automatically saving of method is missing in audit trail
KPR#:2053
Selection of storage of instrument curves is not reported in method audit trail
KPR#:2059
Sequence Audit Trail is not written for 2D LC instruments
KPR#:2092
Update Master Method with DA settings does not copy the full list of changes to the audit trail
KPR#:2099
Pre-run changes to sequences are not captured in the sequence audit trail
KPR#:2116
Sequence audit trail reports outdated sequence line number
KPR#:123232
Sequence Method audit trail is missing entries for some runs
KPR#:203019
Changes of integration events in combination with preview results may not be tracked correctly in method audit trail
KPR#:291500
Severity levels of activity log did change and need explanation
KPR#:292364
Snapshots are not recorded in the audit trail
KPR#:293333
Snapshot not disabled when “After any data modification” is checked
KPR#:296074
Time stamps of method audit trails are printed in ISO format
KPR#:317916
Instrument error stopped sequence, but log also states “stopped by user”
KPR#:431170
Sequence audit trail can be closed without entering a reason for the update
KPR#:488786
Acquisition method last change timestamps do not match between Intelligent and Classic reports
KPR#:58
Activity Log not displayed in web browser when using Chrome 5.1
KPR#:338
EZChrom: PAL XT module not available after installation
KPR#:1860
Cannot open batch - specified range parameter is bad
KPR#:1903
Getting errors when selecting any Classic Performance report templates in Batch Review
KPR#:360620
Batch Reporting does not allow the same compound name to be used for both signals
KPR#:694004
Acquisition Method changed by user without required privileges
KPR#:149884
CE Method not working for CEC Mode and/or MSInstalled soft configuration option
KPR#:150050
Using timetable setting Internal Pressure in CE/p and CEC mode is not applied
KPR#:1089
Run a sequence with recalibration, only the first ISTD is updated
KPR#:1817
Sequence run for bracketing standard which is started with blank completed without creating data files
KPR#:1827
C.01.04 Hot Fix03 : incorrect calibration curves legend in calibration curves printout
KPR#:1923
Compounds without ISTD correction are not reported
KPR#:2118
0.000 amount in Sequence summary report using ISTD% report
KPR#:2178
Wrong vial number in intelligent report with bracketing calibration
KPR#:153540
Chromatograms extracted from UV iso-absorbance plot cannot be directly used for quanitfication
KPR#:171163
in Signal Details and Calibration tables has an old functionality of "Clear" instead of "Copy"
KPR#:253208
Include/ignore origin in default calibration settings shows a warning
KPR#:49
LC/MS data acquired on Chinese or Japanese OpenLAB CDS ChemStation C.01.07 cannot be reported with Intelligent Reporter.
KPR#:52
LC/MS data acquired on Chinese or Japanese OpenLAB CDS ChemStation C.01.07 cannot be reported with Intelligent Reporter.
KPR#:53
LC/MS ChemStation - Intelligent Reporting would report a red cross when using Analytical Studio Reviewer
KPR#:55
Intelligent Reporting template migration from OpenLAB CDS ChemStation Edition C.01.04 or C.01.05 to C.01.06 or higher
Keyword: Client Report Rendering
KPR#:41
Data Peak without Apex Spectrum cannot render IR report with Apex Spectrum selected
KPR#:46
Intelligent report of English openlab chemstation can not show chinese name in PDF report
KPR#:628
Large matrix reports may produce additional empty pages
KPR#:653
Calculating the standard deviation of identical numbers may return NaN
KPR#:657
Changes in report templates are not immediately shown on report printouts
KPR#:659
Rendering error when using a template containing a large result matrix
KPR#:149860
TCC CLuster remains offline with message "Initialization of ColumnCompCluster failed
KPR#:149912
Position number after run is reported wrongly for Multicolumn Thermostat G7116B in the report and audit trail
KPR#:149913
Field sizes for column description and comment have been increased from 16 to 32 characters for G7116B Multicolumn Thermostat
KPR#:149914
Method transfer from G1316A/B/C to G7116B failed if option "next position after run" was chosen
KPR#:149917
Wrong display for 2pos/10port and 2pos/6port valves in method UI for G1316A/B/C modules
KPR#:149918
No tooltip displayed if linked detector is missing for G7116B Multicolumn Thermostat
KPR#:149921
Incompatible change in behavior of 2-position/6-port and 2-position/10-port valves in G1170A, G7116B and G4227A compared to Thermostatted Column Compartments G1316A/B/C
KPR#:149925
Error dialog and error logbook entry for TCCs G1316A/B/C
KPR#:149929
Control UI for G1316C does not work when using the Instrument Control Framework ICF in client/server setups.
KPR#:149930
Changing control settings like switching on/off the Column Compartment Thermostat or Leak Detection in the instrument control UI does not work
KPR#:149943
Import/export functionality in Column Assignment dialog of MCT has been removed
KPR#:149962
No comments can be entered for virtual tags
KPR#:149976
Column information may get lost within VTCs during auto-configuration in some cases
KPR#:149977
Last injection date and injection counters are not updated in VTC in some cases
KPR#:149978
Valve-Thermostat-Cluster time table can be edited in read-only mode
KPR#:149985
Maximum column void volume is 50 mL
KPR#:149998
Auto configuration fails when changing the valve type in an MCT for an existing valve thermostat cluster
KPR#:272678
G1316C does not recognize 800 bar Quick Change Valve Heads configured in LabAdvisor
KPR#:1609
Companion fails to run blank run for valved/ALS injections
KPR#:117721
Exception when login in OpenLab Control Panel. ECM does not display all the subfolders when browsing for Project Path during project creation; exception displayed
KPR#:217841
Unable to configure multiple instruments with the same name in different locations (on the same AIC)
KPR#:283789
Create new instrument causes error "System.Net.Sockets.SocketException.."
KPR#:1905
Default value for a custom field is not cleared
KPR#:2106
Custom Fields can disappear after acquisition
KPR#:105305
Ability to store results in sample or compound custom fields
KPR#:1837
Empty Custom Fields are not printed in Intelligent Report
KPR#:2098
Custom Field values can be accessed during acquisition of a sequence.
KPR#:980
Relative Retention Time solution does not work for uncalibrated peaks
KPR#:1995
FIltering syntax has changed
KPR#:2012
Calibration Levels and the corresponding amounts might not match in Intelligent Reporting
KPR#:471016
Sample order number change in C.01.10
KPR#:602877
Topology report considers real instruments only
KPR#:149820
In localized CDS for ELSD detector G4218A no signal data is shown or stored
KPR#:149830
High load on CAN may cause "Spectra runbuffer overflow" entries and a loss of spectra
KPR#:149837
Autobalace executed although in Sequence table set to "OFF"
KPR#:149851
Occasional shift in retention times for G1315A/B or G1365A/B
KPR#:150000
Retention times may be slightly too high for DAD data acquisition rates below 1 Hz
KPR#:149949
Module stays offline after a configuration change using the "Modify" function
KPR#:1578
Reprocessing copy starting as a background window.
KPR#:1639
Data System will not "Autoscale" a TCD data file to the correct peak
KPR#:1652
Y-axis labeled incorrectly when loading GC signals acquired with MSD ChemStation
KPR#:1654
ChemStation B.04.03 [54] will not "Autoscale" the TCD data file
KPR#:1785
C.01.05: no longer possible to print reports directly from the DA.M in Recalculate Last Result mode
KPR#:1796
Report template not found when reprocessing with a master method
KPR#:1843
Snapshot runs cannot be integrated manually
KPR#:1863
ChemStation CDS client appears to freeze after hiding a save as dialog
KPR#:1879
Sample parameters for sequence runs are not copied to a self-assembled result set
KPR#:1884
Relative Retention Time calculation differs from the EP definition.
KPR#:1897
Japanese ChemStation: Intelligent report template settings are not getting saved
KPR#:1975
AIA export/import for MS-data does not work
KPR#:1980
Selection of signals does not work in Japanese ChemStation
KPR#:1999
Negative spike can appear in some zoom range
KPR#:2000
Plate number and Resolution of negative peaks is not calculated
KPR#:2010
IR Sequence Summary Report is always created
KPR#:2049
Calibration time stamps may differ by a few seconds
KPR#:2101
Bracketing calibrations are not performed when runs other than samples are used in a sequence
KPR#:2151
6890 GC data acquired with ChemStation B.04.03 cannot be loaded with ChemStation C.01.07
KPR#:2161
Display of fraction annotation in chromatogram depends on the signal options
KPR#:2171
Fractions in MSD traces are only annotated for the first signal
KPR#:89199
Cannot read Spectra (*UV) files from OpenLab CDS 2.x
KPR#:89262
Reprocessing in online session automatically paused after third (3rd) line
KPR#:117028
Data Analysis remains busy when trying to generate a report
KPR#:119120
ChemStation stopped working after deleting UIB signal in Data Analysis
KPR#:124571
Data Analysis remains busy when trying to generate a report
KPR#:156495
Pressure signal is not shown when "high pressure" is selected in Data Analysis
KPR#:176783
Macro error prevents 2DLC DA plugin from loading and EIC extraction fails when Instrument Curves loaded
KPR#:242519
Cannot view Isoabsorbance Plot for DAD2, DAD3 detector
KPR#:262772
G7117 detector data does support calculated purity threshold function with rev. C.01.07
KPR#:274472
Loading data takes 20 times longer
KPR#:292505
Split peak on MS data generates error message stating "NO peak detected"
KPR#:295747
Fraction Data from classic cluster configurations cannot be loaded
KPR#:300364
ChemMain crashes when reprocessing with method containing huge number of events in audit trail
KPR#:318693
RX files missing in the resultsets acquired with MS instrument
KPR#:359503
ChemStation crash during delay calibration if instrument name is too long
KPR#:385216
ADFExport for ChemStation result may contain invalid instrument description
KPR#:408053
Character missing in signal label when loading data imported from OpenLab CDS 2.x
KPR#:472767
Macro Verftest.mcx for Verftest will not run on AIC instrument 10
KPR#:599531
Column length value in the report has the wrong value
KPR#:679265
System suitability not calculated for some peaks when using tail tangent skim timed event.
KPR#:715384
Un-zooming chromatogram does not go back to the defined scale settings
KPR#:732976
CP Parsing error when using Recalibrate from Quantitation icon in the System Diagram
KPR#:804715
LTS 01.11 Chinese version: "Not a string" error shown during DAD1A signal overlay
KPR#:1845
Garbage characters can appear in English LCMS reports
KPR#:1688
Preferences set in offline overwrite online Preferences and vice versa
Keyword: Data Analysis , Programming
KPR#:1845
Garbage characters can appear in English LCMS reports
KPR#:1575
HTML Help requires Internet Explorer revision 6.0.28.00.1106CO or higher
KPR#:2018
M8301-96025 Japanese references manual has wrong formula for PeakValleyRatio
KPR#:2190
Insufficient description of regression coefficient in the OpenLAB CDS ChemStation reference manual
KPR#:101356
Peak Model Retention Time used but not saved or documented
KPR#:139546
Wrong description about “Correlation coefficient” in Reference Manual
KPR#:194088
ChemMain.exe crash and the LCMS instrument window closes automatically
KPR#:311116
OpenLab Chemstation 1.10 Release notes sets incorrect Allotrope expectations
KPR#:410634
ChemMain stopped working and other instabilities due to Adobe Acrobat Reader and updates pushed by Adobe
KPR#:434199
After Any Data Modification is selected – not all restrictions are listed in Help
KPR#:721701
Automation Interface documentation updates were not included for LTS 01.11
KPR#:1611
Path definition in OpenLab Control Panel is not used in ChemStation
KPR#:1659
The storage path for files in ECM is user-specific
KPR#:1667
Cannot switch current ECM authentication to a different ECM server
KPR#:1672
it is not possible to remove AIC from networked workstation
KPR#:1794
ECM user licenses not released when user is changed in OLCDS ChemStation
KPR#:1814
In file information of a datafile the Operator name shows as "Disconnected"
KPR#:1818
Submitting multiple sequences to queue, the User token in upload path is resolved to current user - not to submit user
KPR#:1833
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:1844
Single runs in self-assembled data sets cannot be integrated
KPR#:1856
Instrument access rights are not enforced after user change
KPR#:1867
Incomplete Result Set package (sc.ssizip) was sent to ECM: versions do not contain the data files
KPR#:1873
Instrument access rights are not enforced after user change
KPR#:1893
OLCDS Chemstation - DynaZip Zip error in ECM Queue management
KPR#:1901
Result Sets with a % in the folder name cannot be uploaded to backend storage
KPR#:1912
ECM upload may fail with warning: Could not read VCS insformation for 'C:\Chem32\1\VcsCache\
KPR#:1925
incomplete versions of result sets are uploaded to the back end storage.
KPR#:1939
File permissions set incorrectly after backup on OpenLAB Secure Workstation
KPR#:1987
ECM data path status in ChemStation C.01.05 (41) HF6 is not always updated correctly
KPR#:1992
ChemStation with no access to any menu or icon.
KPR#:1997
ECM path not recognized in preferences after upgrade
KPR#:2004
Sequence name using special characters does not upload to ECM
KPR#:2019
Wrong take over message can be displayed in a Distributed System when a user had online and offline session
KPR#:2050
The Upload path can edited for a running sequence
KPR#:2058
Sequence not uploaded to the correct location in backend after user change
KPR#:2076
Queue maintenance after shutdown could conflict with short inactivity lockout
KPR#:2094
Runs in a sequence with bracketing can be accessed too early
KPR#:2113
Sequence/single runs manually uploaded to ECM still shows the asterisk
KPR#:2115
Number of function arguments is invalid when loading data from backend
KPR#:2127
Files names containing special characters are scrambled after downloading SSIZIP from Data Store / ECM
KPR#:2135
Shortutl.mac needs modification to show the ECM path
KPR#:2155
Notification for data not uploaded to database backend not clear
KPR#:2162
Path Preference issue with C.01.07 SR2 and ECM LCDF-applied role
KPR#:2174
Sequence upload to ECM leaves an incomplete progress bar, ECM Uploading in Process
KPR#:89186
It is not possible to remove AIC from networked workstation
KPR#:123515
"Error 41092 occurred" in the system activity logs
KPR#:135793
LCDF applied role prevents download
KPR#:161869
ECM Uploading in Process dialog box does not close in offline session after Resultset is uploaded to ECM
KPR#:166923
Error occurred after clicking on Save Report Template on ECM tab
KPR#:193594
No “Connection Error” message displayed if "save data in ECM" function is performed while ECM is disconnected
KPR#:263762
Not able to configure ECM data path in ChemStation C.01.09 with ECM 3.6 as backend
KPR#:321577
If data exceeds the maximum size it cannot be uploaded to ECM
KPR#:527213
ECM Chemstation XML Filterpack is failing to filter ssizips when instrument includes fraction collector
KPR#:541911
Logon to OpenLab Control Panel and unlock ChemStation is slow with ECM 3.6 storage
KPR#:570626
Enhanced SSIZIP file modification detection
KPR#:716634
Exception of type 'System Out of Memory Exception' when data > 550Mb uploaded to ECM at SSIZip checksum generation
KPR#:2065
Report Upload from Review View fails with error
KPR#:2112
Error 16420 occurred while printing a report
KPR#:1957
Local files are not stored securely with ChemStation Edition
KPR#:1763
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:149823
ELSD G4218A can not handle change of Sample Time during run
KPR#:149825
The ELSD method cannot be called by a right-click from the context menu
KPR#:149856
Runtime of a G4218A ELSD cannot be changed during a run
KPR#:1917
Easy Sequence: Running back table headspace seq, select front table, add sample name, click back table and select cell causes seq to hang
KPR#:2014
Error message when Blank with Vial is entered in Easy Sequence Setup
KPR#:2082
Vial values not accepted for 96 well plate tray in Easy Sequence Setup
KPR#:516688
Easy Sequence is not added to the Run Queue when resources are insufficient
KPR#:149862
FLD G1321 Scan ranges and threshold under certain conditions not available
KPR#:1608
FetchCPVar Cannot Talk to Application Error
KPR#:1628
FetchCPVar Cannot Talk to Application Error
KPR#:1630
Column compensation cannot be started from the Data System with 68xx RC.NET
KPR#:1683
OpenLab CDS, C.01.02, 7820A run continues when STOP on 7820A is pressed
KPR#:1868
After switching the injection source from front to back it is no longer possible to change the counter part of the Prefix/Counter naming scheme
KPR#:1887
After switching the injection source from front to back it is no longer possible to change the counter part of the Prefix/Counter naming scheme
KPR#:1986
GC Back injector only used, the sequence specified volume is ignored and the method setting is used
KPR#:2066
Wrong Error Message Used When GC Method Download Fails
KPR#:2103
Instrument session continues to run in background after session is closed
KPR#:205926
ChemMain stopped working during running sequence; 'Out of Memory' in acquisition logs
KPR#:260731
Unexpected abort of a sequence with GC dual injection
KPR#:368024
Event table not in chronological order, not all events executed
KPR#:380477
Reports generated based on snapshot data do not contain column information
KPR#:453813
Overlapped injection enabled for a dual sequence run is not handled correctly for back injection
KPR#:502630
GC sample diagram shows red vials for successfully completed injections
KPR#:514410
Headspace not ready condition hold up non-HS injections invalidly
KPR#:534214
Failed to create instance! error message when launching GC
KPR#:540016
Run queue will show Front injector on Pending runs when method actually uses back
KPR#:583491
Detector signal name change causes recalibrate
KPR#:593325
Misleading warning is displayed if the barcode reader writes the barcode in sample name field
KPR#:677663
Resource Library dialog is not working correctly
KPR#:711328
Method Resolution Wizard does not work when loading 6890 or 6850 methods to other GC Types
KPR#:736317
'Parsing Error' on ChemStation taskbar after resolving method
KPR#:1966
Sequence stops for no apparent reason
KPR#:1966
Sequence stops for no apparent reason
KPR#:375261
Pause/Retry after ALS error re-numbers locations can abort sequence with HSS
KPR#:150035
Content of localized Online Help texts may not be complete (Japanese, Chinese and Portuguese version)
KPR#:150072
Flow Modulator requires regional settings in English
Keyword: Globalization,Pump Driver
KPR#:149973
Pressure range settings may seem to change when using unit psi
KPR#:90
Methods Must be Stored on the C Drive
KPR#:164
Injection Volume is reported as 0 uL or blank in report for headspace
KPR#:171
ChemStation crashes on start-up if the headspace is in an error state
KPR#:179
7697A temperature(Oven,loop and Tr line) loaded with the method parameters of line 1 after the end of the sequence.
KPR#:699742
OpenLab Control Panel help launches in English
KPR#:717784
ChemStation Online Help needs updated to include the changes in USP 2022
KPR#:1618
Sporadically launch of an online session will not proceed
KPR#:1869
ChemStation does not initialize when system paths are not accessible
KPR#:1898
Software does not initialize properly when the last method used, was located on an unaccessible drive
KPR#:1959
Running processes can prevent instruments from being deleted
KPR#:2023
Misssing GC license message even though system has enough GC licenses
KPR#:2032
Misleading message about missing license during session takeover
KPR#:2100
“Autostart Macro failed, error #999 occurred”
KPR#:2158
A configuration change to one of the DA-only instruments causes configuration panel to pop-up every session start
KPR#:89201
"Autostart Macro failed, error #999 occurred"
KPR#:131104
AutoStart Macro error on start up when using min/max dimensions within the SampleContainerDevice
KPR#:147982
Offline MS ChemStation hangs during startup
KPR#:393124
Corrupt Sample Containers cannot be deleted from the Sample Container database
KPR#:2045
Error during Secure Workstation installation
KPR#:22
FTP protocol is enabled by default on OpenLAB Data Store A.01.02 systems
KPR#:42
OpenLAB Control Panel does not open after upgrading Windows Server 2008 R2 to Windows Server 2012 R2
KPR#:45
Not possible to uninstal DataStore in a Secure WorkStation after doing a repair
KPR#:1627
AIC ports not opened for 7890A acquisition
KPR#:1885
Software upgrades from a share will not work
KPR#:1938
An error occured after clicking the 'Configure Instrument' Button twice
KPR#:2130
Secure Workstation: User is not prompted to change password on first log in
KPR#:120427
Wrong OpenLAB Shared Services database type installed when selecting PostgreSQL with C.01.07 SR3
KPR#:173564
Installer wrongly detects the presence of OpenLAB Data Analysis
KPR#:176707
Warning message “The installed OpenLAB software is incomplete” or “Repair is not possible as no complete topology”
KPR#:193962
Error occurred for OpenLAB CDS ChemStation Drivers shows red in status during upgrade via DVD
KPR#:224734
Cannot access GC configure instrument in Control panel after upgrading from C.01.08
KPR#:243280
Error occurred for OpenLAB CDS ChemStation Drivers shows red in status during upgrade via DVD
KPR#:298255
AIC Installation from DVD is not supported
KPR#:318797
Japanese Installation of ChemStation after the OpenLab CDS 2.4 product fails
KPR#:366738
Fail to Register Instrument Controller
KPR#:372511
With C.01.10 Update 1, it is no longer possible to disconnect from a running instrument session and a launching failed error occurs
KPR#:384350
Install will fail if user name contains a #
KPR#:414556
Installaton of AIC fails due to security policies with internal default CSI user password length of 12 characters
KPR#:416759
Use of pain text passwords for CSI users unclear
KPR#:470764
For disabling TLS 1.0 on AIC's with Server 2016, SQL with high availability DB is needed
KPR#:674125
System preparation sometimes hangs at the Tcp Port Sharing and Activation step
KPR#:694858
System Preparation sometimes hangs at the TCP Port Sharing and Activation step
KPR#:698383
System Preparation Tool: Check Pending Reboot step fails even after rebooting computer
KPR#:758423
C.01.10 Site Preparation Tool fails the Microsoft Hyper-V check
KPR#:785114
AIC registration is failing
KPR#:2121
35900E ADC-instrument cannot be configured with RC.Net driver after reinstalling the OpenLAB CDS application
KPR#:2045
Error during Secure Workstation installation
Keyword: Installation , 35900 Control
KPR#:2121
35900E ADC-instrument cannot be configured with RC.Net driver after reinstalling the OpenLAB CDS application
Keyword: Installation , Initialization
KPR#:2045
Error during Secure Workstation installation
KPR#:1534
User is not asked to save manual integration events when manual events are stored in the method
KPR#:1744
Misleading display of tangent peak width
KPR#:1838
GC compound peaks might be labelled as solvent peaks
KPR#:1854
Background subtraction does not work on chromatograms with manual events
KPR#:1921
Manual integration changes need actively to be saved, when the method contains stored events.
KPR#:1932
Manual integration may become slow when Results Audit Trail is switched on
KPR#:2095
Baseline not shown in IR report when baseline end too near signal end
KPR#:2120
Foley-Dorsey Plate calculation formula uses incorrect asymmetry factor
KPR#:2128
Adjacent featureless peaks might be reported with the same retention time.
KPR#:2150
User privileges are not properly transferred to the offline session
KPR#:182387
Unexpected baseline and peak integration when using GC demo data file
KPR#:193315
Integrator timed event: Split peak can cause integrate to show a misleading error message box
KPR#:283472
Manual integration events not available with ECM connection and µ in data file name
KPR#:454723
Number format of Area% column in integration results is set back to default when selecting a row in the table
KPR#:385410
Some Japanese or Chinese character are replaced by "_" in the data file name
KPR#:398646
ChemStation Admin Tool settings have been changed when no modification was done; Translation issue
KPR#:2040
Peak Purity Report in Chinese Version Wrongly use Peak Number as Peak's Retention time
Keyword: Localization (Japanese)
KPR#:1668
Japanese characters from saved pdf report are garbled when copied and pasted to another application
KPR#:2164
Data file name has Japanese character if data system create data file
Keyword: Localization,Report Template Design
KPR#:277791
Composite group saved with special character in the name
KPR#:803
Only 20000 lines are kept in the sequence log
KPR#:1637
'Method not downloaded to device' message when PAL is present
KPR#:1889
User log off is not recorded in the logbook
KPR#:2039
Unresolved controllers interrupt status updates and system activity logs
KPR#:2140
OpenLAB Control Panel activity log and current log book are missing entries for GC Back injector
KPR#:2146
System activity log shows a failed user change as a successful login (private session lock)
KPR#:2179
Differences between Offline Instrument Logbook and OLCP Activity log after connectivity loss to ECM
KPR#:241458
Incorrect run logbook information about unauthorized integration parameter change
KPR#:256761
ChemMain aborted when printing instrument logbook
KPR#:281314
Internal server error after long query
KPR#:300470
Entry in logbook appears multiple times for one parameter change
KPR#:300963
"stop by user" in sequence log will not affect data integrity
KPR#:303382
User selection "Stop" for ALS error, it is not logged in sequence log
KPR#:370213
Unwanted log entries: Method not found: 'Void Agilent.Windows.Forms.SampleContainerControls
KPR#:432950
Valve switch information missing in logbook
KPR#:570794
Logbook viewer does not display correctly if logbook entries contain the '>' character
KPR#:726331
File accessed at same time by both online and offline sessions result in a Logbook File error
KPR#:1958
"AgilentAddOnMSDeconvolution" license is requested although it is not configured
KPR#:142695
No ability do fast polarity switching on G6125B LC/MSD
KPR#:156401
LC/MS Peptide tools incompatibility
KPR#:156987
Value for Polarity Switch Delay was not updated if user presses "Enter" key
KPR#:157119
LC/MSD ChemStation terminates during run
KPR#:158433
Printout from LCMS method does not include the MSD model
KPR#:162667
C.01.08: NIST Library searches are not automatically loaded; NIST Library search is not possible
KPR#:162907
NIST Library search stopped working
KPR#:168054
NIST library searches are not automatically loaded
KPR#:170285
C.01.08 ChemStation crashes when loading a method with 10 or more SIM ions
KPR#:189566
Delay Sensor Correction in Delay Evaluation is incorrect
KPR#:218911
RMS Signal-to-Noise specification should be 100:1 on the report for ES-API source
KPR#:234463
List Content of MS Spectra Window produces Parameter 1 Invalid in type or name
KPR#:258401
Extracting ion chromatograms from TIC can lead to an unresponsive ChemStation
KPR#:276189
Walkup rejects any injection volume as invalid with G2258A dual-loop autosampler
KPR#:297893
Sample Purity Options dialog becomes read-only when ASR File Generator is installed
KPR#:298455
Chemstation hangs when calling classic LCMSQual report preview from the Sample Table's context menu
KPR#:370173
Crash occurs when DAD or MWD is configured and WalkUp user submits sample with wavelength at lower or upper end
KPR#:375369
Vial location column missing on classic report with FIA
KPR#:480544
Error extracting MS spectra: "There are not enough points in the object to do the operation"
KPR#:499793
If the tune name is too long, it is not correctly presented on classic report
KPR#:638529
LCMS spectrum not visible in 'write data to CSV" box
KPR#:654884
LCMS Chemstation hangs when noise intervals are outside of the chromatogram range
KPR#:722644
Target masses greater than 18,000 can't be entered in the Sample Entry User Interface
KPR#:724794
Permission missing to control UV Spectra Library deletion
KPR#:740710
EIC extracted signal of Abundance greater than 999.9 cannot be calibrated
KPR#:755236
Walkup: Intermittently, an unexpected exception is thrown when submitting a sample
KPR#:2125
Peak Performance calculations are off when data points are not equidistant
KPR#:1616
Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group
KPR#:1675
Automated Delay Calibration of the MS signal for Fraction Collection is not supported in the RC.Net environment
KPR#:1686
MSD Tune View will awlays load ATUNES.TUN if tune file loading is cancelled
KPR#:1730
LC/MS may remain in a run state after Method/Sequence termination due to an HPLC hardware error
KPR#:1910
A missing vial in the sampler can cause the MSD to be non responsive
KPR#:1916
Automated LC/MS Delay Time Calibration does not work with Preparative Fraction Collector (G1364B) using RC.NET drivers
KPR#:2001
LC/MS selection valve delay time set at 4.0 minutes cause intermittently no signals in the MS
KPR#:120188
"Instrument Error occurred": FIA run does not complete when “Edit FIA Series” Dialog box is open
KPR#:213171
Logbook error when saving LC/MS method
KPR#:280527
Sequence hangs from Method Scouting Wizard A.02.09 generated campaigns/sequences
KPR#:286295
LC-MS system goes not ready and aborts the sequence and queued sequences with method Instrument nRdy timeout occurred
KPR#:339648
ChemStation MSD driver remains in NRDY forever and aborts the currently running sequence
KPR#:348537
Not applicable MSD warning “NozzleVolt/VCharge is not active” in MSD Dashboard
KPR#:357697
Internal error occurs when switching MS divert valve in MSD user interface
KPR#:1353
GC Offline Data Analyisis checks for ALS location when loading method
KPR#:1473
CTC portion of method printout does not include overlap settings
KPR#:1624
Undefined symbol: PrDARep error displayed when selecting Export e-Method
KPR#:1643
Open GC edit parameters and close it without changes; method modification flag is set
KPR#:1650
Headspace or PAL Sampler sequence with methods saved to a location other than C:\ does not run
KPR#:1663
No information around handling for special characters
KPR#:1710
Cannot transfer an English method to a Japanese or Chinese ChemStation and vice versa
KPR#:1748
Add/Removal of 7697 HS from 7890 does not show a change occurred; not displayed in Method Load options; method not shown as modified
KPR#:1846
[DEL] key in time tables does not work
KPR#:2021
ACQ.txt file contains Column(s) information that does not match actual column used and leads to confusion
KPR#:2131
Secure File IO enabled blocks Export eMethod... and Import eMethod: 'Invalid File Path'
KPR#:2138
G7167B dual needle: injector program set from context menu is not visible in Instrument method setup
KPR#:2168
Method modification date in classic reports is the sequence acquisition date
KPR#:2177
Method listing from [View Method] contains parameters from method loaded in Method Run Control
KPR#:2181
G4218A method parameters cannot saved when SecureFileIO is enabled
KPR#:123270
Method listing from [View Method] contains parameters from method loaded in Method Run Control
KPR#:141882
6850, 6890 with PAL or headsapce: method not saved and logbook shows 'Checksum error: uncontrolled method modification'
KPR#:197219
C.01.09 method loaded on C.01.08 session gives many error messages (due to Custom Commands feature added in C.01.09)
KPR#:297466
report incorrectly shows method modified after generating an LCMS classic report with an old method
KPR#:408346
Method resolution unclear for DAD after reconfiguration
KPR#:489525
Changing state of Enable SamplePrep Program does not set dirty method flag
KPR#:717061
Graphical range option for signal display of the method is not loaded correctly and defaults to autoscale at method reload
KPR#:725145
No Negative Nozzle voltage with AJS Source
Keyword: Method Scouting Wizard
KPR#:1720
Cluster information can be overwritten from the offline session
KPR#:1911
Loading result sets or methods with long names causes errors
KPR#:1752
Changing the sample type in a sequence with cyclic calibration corrupts the sequence structure
KPR#:1836
Date Time entries can be incorrect for aborted fraction collection runs
KPR#:1865
For back injection samples, the Acq. Method column of the Navigation Table incorrectly displays the analysis method
KPR#:1878
Autofilter function is not implemented
KPR#:2143
Reprocessing of old (C.01.05 or earlier) result sets might skip runs from the Navigation Table
KPR#:1612
Printing fails if printer name exceeds 100 characters
KPR#:1855
Unable to perform "XML Export" with the tenth instrument on an Agilent Instrument Controller (AIC)
KPR#:506719
Chemstation spectra CSV extraction not properly rounding the decimal places
KPR#:516136
Format for XML export of location changed
KPR#:946408
Print out of CM search results has the wrong message stating that there is no search result. But there are results found and printed.
KPR#:392755
Usage of "Next location" does not work as expected when using multiple Recovery Collectors in a cluster
KPR#:1845
Garbage characters can appear in English LCMS reports
KPR#:1617
AIC can experience communication loss if Application Experience service is running
KPR#:1761
The command "macro " cannot be used to select *.mac files
KPR#:2025
The csv export of tables results in an error
KPR#:89227
Acq-svc.log shows: permission denied in VC_CopyOneDataFiel line 1678 code 41084 at copy
KPR#:91024
Instrument driver error crashes ChemStation
KPR#:370835
Long annotation text cannot be read using .Net libraries for register access
KPR#:371551
"CoreAddCmd redefining Alert!" error displayed during instrument launch
KPR#:371555
"Empty list" problem in Show dialog for Chinese and Japan Systems
KPR#:371556
Inconsistent handling of illegal characters in file names
KPR#:149819
Pump overpressure: impossible to turn off pump via UI
KPR#:149838
Misrepresentation of Bottlefillings in Dashboad of PumpValveCluster
KPR#:149840
Solvent compositions during run do sometimes not sum up to 100%
KPR#:149845
PumpValveCluster with ISET - Solvent composition: resulting %A is wrong (at 1 line in timetable)
KPR#:149868
Blend Assist : Timetable not persisted properly
KPR#:149879
Unused solvents are handled differently for 1100/1200 Series/1260 Infinity vs. 1290 Infinity/Infinity II pumps
KPR#:149881
Prep-pump cluster stays not ready after restarting OL ChemStation (reconnect)
KPR#:149885
The 1290 Quaternary Pump G4204A is specified for a maximum flow rate of 5 ml/min
KPR#:149895
Purge function has been improved for Preparative Pump Clusters
KPR#:149909
Solvent type may change during a method transfer from 1290 Infinity/II Quaternary Pumps G4204A or G7104A to 1290 Infinity/II Binary Pumps G4220A/B or G7120A
KPR#:149928
Recorded Acquisition Method has Incorrect Pump Parameters
KPR#:149938
Implementation of seal wash behavior has been improved for G4220A/B and G4204A
KPR#:149950
Warning "No matching ISET solvent combination" for G4220A/B pump as emulated pump
KPR#:149951
Using the OpenLAB CDS EZChrom edition feature "Extend run" will disable ISET features in the 1290 pumps
KPR#:149986
Seal wash for Preparative Pump G1361A or Preparative Pump Cluster is displayed as "off" even when it's on
KPR#:149989
Wrong purge flow and duration shown in prepare pump dialog depending on regional settings
KPR#:149992
Shutdown recovery does not work for Preparative Pump Clusters in a Multi-Controller environment
KPR#:397976
Reporting matrix not correctly exported to csv report output
Keyword: Report Template Design
KPR#:25
Intelligent Reporting : Last displayed digits might be cut off, in cases where the number is zero
KPR#:40
Missing named peaks are reported in Intelligent Reporting for all the channels or signals
KPR#:42
The Fixed Text field in IR Templates does not accept Chinese characters
KPR#:45
After installation of LC Dissolution AddOn the Report Template Editor is no longer functional
KPR#:50
Intelligent Reporting - Spaces are stripped from filters when they are saved
KPR#:51
.NET Framework 4.6.1 update not recommended for OpenLAB CDS ChemStation Edition C.01.07 SR2
KPR#:54
Mean summary calculation with report parameters in Intelligent Reporting on OpenLAB CDS A.01.04
KPR#:56
Error seen in Intelligent Report header (#error) if any related data field is used in an expression
KPR#:58
Intelligent Reporting - Stacked Chromatograms do not display the time axis in reports
KPR#:59
Intelligent Reporting - Square brackets cannot be used in filter expressions
KPR#:60
Cannot calculate summary statistics on columns containing variable or aggregator
KPR#:63
The sorting option of MS spectra report item in Intelligent Reporting is not functional
KPR#:620
Error message when adding a chromatogram to a report template
KPR#:621
Internal grouping of tables and usage of calculation variables cannot be applied at the same time
KPR#:622
Summary Calculations cannot be applied to table columns containing calculation variables
KPR#:623
Adding a template description to a previously erased and saved description field causes a KeyNotFoundException
KPR#:627
Customized composite groups should have a unique name
KPR#:662
Filter definition does not always allow usage of OR operators
KPR#:668
Reporting time format does not offer an option of (dd/mm/yyyy and time)
KPR#:385844
Reporting method information snippets are duplicated in summary report outputs
KPR#:36
Min and Max column summary calculations report GMT and not local time for date fields
KPR#:37
Only Main peaks for a compound are shown in Intelligent Report.
KPR#:39
The report date is printed incorrectly in the Agilent supplied intelligent reports
KPR#:43
The template fields "SignalToNoise_EP" and "SignalToNoise6Sigma" do not work in Japanese version
KPR#:44
The Chinese and Japanese templates for the Extended Performance Report need updating
KPR#:47
Center of triangular calibration point labels do not represent the measured value
KPR#:48
XLS export of intelligent reporter templates generate hidden rows
KPR#:57
The subtract blank feature used in OpenLAB CDS ChemStation does not match with the Intelligent Reporting report generated
KPR#:658
Customized chromatogram or axis lables may not display correclty
KPR#:117645
Migration of templates from C.01.04 to C.01.07 SR2 - Aggregator with category corruption -
KPR#:362808
Using a variable expression in conditional formatting
KPR#:383434
Export to excel does not export the report header logo
KPR#:401975
Report template containing multi-line expressions fail to generate printout
KPR#:1305
Group sum results are not exported to the result.xml
KPR#:1572
Batch mode does not allow usage of intelligent reports
KPR#:1579
Report Templates newly added to a Result Set do not appear in the list of available templates
KPR#:1582
Mismatch in drawing of peak performance between Classic Report and Intelligent Report
KPR#:1619
Intelligent Reports do not contain ECM specific information in Review mode
KPR#:1634
B.04.03 DSP1 [54] Data Analysis, Graphics, response range does not work
KPR#:1673
CSV, DIF, and EMF files are not generated for some Classic report styles
KPR#:1703
ChemMain.exe can crash when a long file path is used
KPR#:1705
Chinese/Japanese characters cannot be searched in the Report Viewer
KPR#:1722
Tangent peak width not correctly evaluated.
KPR#:1751
eSignature data file privilege is missing from the ChemStation lab manager, analyst, and operator roles
KPR#:1754
The structure of the reports to file has changed
KPR#:1756
CP3: Number format error in: %d when printing report to screen in Data Analysis
KPR#:1778
'Number Format error' in Sequence log file after printing report to screen
KPR#:1779
Peaks are calculated by area only or height only in a report
KPR#:1786
A Sequence Summary Report after Partial Sequence reports all the runs of the full sequence
KPR#:1821
Classic detailed report does not show calibration curves after encountering a compound with invalid calib curve
KPR#:1823
Default column information report shows incorrect GC column information and labels
KPR#:1839
File corruption may cause sequence runs appearing as single runs
KPR#:1850
Calibration curve plot shows only the order of the calibration level but not it's label.
KPR#:1875
"ISTD and Reference Not Found " and "Reference compound(s) not found" printed on ESTD report when peak summing activated
KPR#:1881
Multiplier in method cannot be smaller than 0.0001
KPR#:1883
Peak Summing, Use Reference does not work for Norm% report
KPR#:1895
Some reports end without the *** End of Report ***
KPR#:1900
"Result Set Templates" group window is unavailable while selecting the IR report template
KPR#:1907
Some method parameters are reported with a limited number of digits
KPR#:1909
Peak width cannot be reported in Intelligent Reports (Chinese ChemStation only)
KPR#:1913
Injection volume and instrument name are not reported in case of aborted data acquisition
KPR#:1922
Manipulation of chromatograms has no effect on Intelligent Reports
KPR#:1930
Width of the integration line is not an element in IR template editor
KPR#:1950
Intelligent Report does not print the 6890 oven ramp
KPR#:1961
Datastore/ECM path under sequence parameters is not reported in sequence template printout
KPR#:1965
Norm% calculation normalizes per channel
KPR#:1972
The y axis for ISTD calibration curves is always displayed as height
KPR#:1974
Cannot change the Report Name for the Signal in the Classic Sequence Summary Report
KPR#:1983
Summed Peak reports adds ISTD amounts to the Total Amount
KPR#:1985
Missing parameters on reports, if files were acquired with a packed column
KPR#:2017
Barcode option 'Write Barcode to Sample Name' does not write if back injection
KPR#:2044
Data files are named odd or even when using the same data file Prefix in Sequence Parameters
KPR#:2051
Intelligent Reporting: The fields for Sample_OrderNumber cannot be used to group replicates
KPR#:2052
LCMS Qualitative Classic Report. EIC signals missing in some Data files
KPR#:2062
Using GC back injector, ISTD 2 - 8 are not included in report when Dual Simultaneous checkbox is unchecked
KPR#:2063
Sample and compound specific custom fields are listed in the sequence but not in reports
KPR#:2064
ChemStation shuts down when Report viewer cleans up temp files
KPR#:2080
The theoretical plates per meter are reported as plates per mm
KPR#:2085
C.01.06 HF03: Classic Extended Performance report is missing information
KPR#:2086
Incorrect method modification date shown on IR reports
KPR#:2088
Data files are named odd or even when using the same data file Prefix in Sequence Parameters
KPR#:2089
The sample information that is input in three lines is displayed in one line in C.01.06
KPR#:2093
Sample Type "Blank" is reported as "2"
KPR#:2111
Session freezes generating Intelligent Report During Sequence Acquisition
KPR#:2119
Single Peak's Peak area is different from Total area for classical built in report
KPR#:2132
490 micro GC column information is not available for Plates per Meter or Peak Performance calculations
KPR#:2134
Intelligent Report snippet for instrument module: 35900 reports incorrectly 'Auto Sampler'
KPR#:2152
Classic Report area% calculation can provide incorrect results
KPR#:2167
Column information missing for G7116B Multi Column Thermostat
KPR#:2169
Column information is not printed, if the Column Description is empty
KPR#:2170
Operator name is reported as disconnected or blank in both classic and IR reports
KPR#:2173
Sequence reprocessing updates and rounds the numeric sample parameters
KPR#:2184
Preprocessing Results stuck in the upload queue will not match what is in Content Management
KPR#:2185
Sequence Method audit trail is missing entries for some runs
KPR#:89195
C.01.06: Sim by Target masses show wrong targeted masses on first injection only
KPR#:102753
Preprocessing Results stuck in the upload queue will not match what is in Content Management
KPR#:114422
Significant figure limit for compound group amount values in calibration table
KPR#:115779
G1364E cluster: Fraction report is mixed; the first fraction is from recovery
KPR#:117643
Report "print to file" of subtracted chromatogram fails
KPR#:123241
Use another sequence method to preview report won't generate new Central storage version
KPR#:126672
Self-assembled Result set, incorrectly lists (F) when the injection was originally done with the Back injector
KPR#:130917
Printing 2 report files causes flickering of screen and no access to ChemStation
KPR#:131124
Blank Subtraction changes the sample name in the Classic report
KPR#:149972
Columns are not reported correctly when switching valves
KPR#:150046
Not acquired signal A (VWD) leads to a "Manual Control" entry in the Result Set Audit Trail
KPR#:151390
Labeled peaks of a mass spectrum are shown with red peak integration lines when printed
KPR#:160573
ASTM -Long-Term Noise Calculation is not done
KPR#:165636
FCC: When switching Fraction Collectors during a run in an asymmetrical setup, unexpected fractions might be collected and reported
KPR#:180068
In generating report the injection number of a seqLine replicate can be wrong or missing
KPR#:187755
3D stack overlay does not report the correct time
KPR#:187890
Error message appears when adding Report Styles or removing Report Styles
KPR#:188182
Interactive PeakPerformance efficiency result is too low
KPR#:202614
Waters Alliance working directory no longer ChemStation’s CORE folder causes issues
KPR#:203100
Waters Alliance: error when loading IR template
KPR#:267383
Operator name is "Disconnected" during unlocking ChemStation
KPR#:277843
Unnecessary calibration warnings appear if report is generated a second time
KPR#:291121
Printing time is 1 second different in the exported report from in the printed report
KPR#:293854
Missing Signal Descriptions for LCMS Signals on Intelligent Reports
KPR#:300787
Calibration Curve last modified date may appear with ambiguous date format
KPR#:301635
Separator date may change in OpenLAb CDS ChemStation Edition C.01.10 for customized OS date formatting
KPR#:302286
Recalibration date is incorrect on IR using ExtendedPerformance.rdl
KPR#:342383
Analytical Studio Reviewer show duplicate peaks on report
KPR#:359409
Intelligent Report extended sequence report missing units in acq method listing
KPR#:361138
Intelligent Report lacks data in the Advanced Run Information
KPR#:371553
Garbage: Error message for File name length of Classic Report template
KPR#:376839
PDF extraction failing
KPR#:419459
Fraction location label is not flexible in C.01.09 and C.01.10
KPR#:421132
Blank subtraction: chromatogram issue in IR
KPR#:467325
Custom report created in C.01.04 is loaded on C.01.10 has all audit trail entries with current date and time
KPR#:476514
Unable to print sample description and Lims ID for IR when Load Signal
KPR#:495250
Reported injection time follows time zone and daylight saving setting and reports wrong time
KPR#:499383
Report item becomes uneditable in Intelligent Reporting editor
KPR#:508000
UV spectra data from CDS2 does not work in IR reporting of spectra when loaded in ChemStation
KPR#:538317
IR report error when loading template that includes composite group
KPR#:568535
XLSX sequence summary file creation crashes for IR template when large batch is processed
KPR#:610844
Sample amount value in IR-report not updated after change in quantitation settings
KPR#:695912
Sequence ACAML has fraction result from a previous run
KPR#:728830
"Print Sequence Template" does not work when the barcode option "On barcode mismatch don’t inject" is selected in Sequence Parameters
KPR#:747935
The xml result file is not correctly formatted if the "Add fraction table and ticks" option is enabled in "Specify report"
KPR#:751862
Incorrect Title of Fraction List in classic report
KPR#:752405
Incorrect Title of Fraction List in classic report
KPR#:788216
LIMSID fields are missing in XML exported result for back injection
KPR#:794750
LC slave module reports the host firmware revision
KPR#:1840
Instrument communication error may cause incorrect injection time stamp on subsequent run
KPR#:1837
Empty Custom Fields are not printed in Intelligent Report
KPR#:2065
Report Upload from Review View fails with error
KPR#:2112
Error 16420 occurred while printing a report
KPR#:2040
Peak Purity Report in Chinese Version Wrongly use Peak Number as Peak's Retention time
KPR#:2125
Peak Performance calculations are off when data points are not equidistant
KPR#:1715
Sequence Summary Report does not do statistics when the first compound is not present
KPR#:2145
Incorrect Sample location reported for bracketing
Keyword: Reports , 1100 Control , 1200 Control , 7890A Control
KPR#:1840
Instrument communication error may cause incorrect injection time stamp on subsequent run
Keyword: Reports , Custom Fields
KPR#:1837
Empty Custom Fields are not printed in Intelligent Report
Keyword: Reports , ECM Integration
KPR#:2065
Report Upload from Review View fails with error
KPR#:2112
Error 16420 occurred while printing a report
Keyword: Reports , Localization
KPR#:2040
Peak Purity Report in Chinese Version Wrongly use Peak Number as Peak's Retention time
Keyword: Reports , MSD Analysis
KPR#:2125
Peak Performance calculations are off when data points are not equidistant
KPR#:1715
Sequence Summary Report does not do statistics when the first compound is not present
KPR#:2145
Incorrect Sample location reported for bracketing
KPR#:228201
Spectra cannot be reported in Intelligent Reporting with MS profile data
KPR#:245690
Report generation hanging when the report template has a custom scaling
Keyword: Retention Time Locking
KPR#:131756
Update Master Method from Data Analysis view does not update RTL calibration
KPR#:345843
Calculated Pressure for Relock is not updated automatically for the 6890 in C.01.10
KPR#:207007
Users can delete their own data without deletion permission via web client
KPR#:1851
Unexpected shifts in the Online Plot
KPR#:1857
In Chinese and Japanese versions the counter function in sample info dialog is not incrementing when the start button is used.
KPR#:1918
Instrument not ready time out message box may block other dialog boxes
KPR#:1919
Single runs do not turn to 'Not Ready time out' automatically and OpenLAB Control panel shows 'running' status
KPR#:1920
ChemStation Shutdown may not update instrument status in OpenLAB Control Panel
KPR#:1927
Systems under heavy load can miss raw data inside the data folder
KPR#:2024
Submission of single runs to the queue blocks the queue
KPR#:2047
Queuing a modified sequence cannot be undone
KPR#:2056
Fraction preview does not work for UIB II as peak detector
KPR#:2075
Japanese ChemStation: The Sequence Table cannot be evoked from the sequence context menu
KPR#:139793
Using double quotes in the sample name for signal runs will not run with no error
KPR#:1936
Only sequences from the default sequence path can be used in a queue plan
KPR#:1936
Only sequences from the default sequence path can be used in a queue plan
KPR#:1966
Sequence stops for no apparent reason
KPR#:1715
Sequence Summary Report does not do statistics when the first compound is not present
KPR#:2145
Incorrect Sample location reported for bracketing
KPR#:1936
Only sequences from the default sequence path can be used in a queue plan
KPR#:778
Change of sample type during sequence acquisition is ignored
KPR#:1514
The path to a data file has a limited length
KPR#:1604
HeadSpace sequence method will not load when it follows a Valve sequence
KPR#:1648
Sequence parameter subdirectory overwrites sample info parameters during run
KPR#:1670
Queue Planner will not run PAL Autosampler sequences that follow a valve
KPR#:1725
Sample statistics are not done, when the first compond in the calibration table is not detected.
KPR#:1787
78xx: Sequence stops when pushing the GC STOP key during sequence
KPR#:1822
Sequence parameters: Changes to Fraction start location ignored
KPR#:1824
After upgrade, Sequence modified flag is set even if no changes are made before reprocessing
KPR#:1847
C.01.03: Opening the back sequence table, click on a cell, make no changes, and select OK causes headspace sequence to stop
KPR#:1849
Opening the back sequence table, click on a cell, make no changes, and select OK causes headspace sequence to stop
KPR#:1858
Queue Sequence will not run sequences with different injection sources
KPR#:1859
No new lines can be added to running sequence when Unique Folder is set to Off
KPR#:1862
C.01.04 Custom Field Values cleared for back sequence table
KPR#:1864
Method of front injector incorrectly used for Data Analysis of the back injections when vial conflict occurs
KPR#:1870
Well Plate locations are not accepted in offline sequences
KPR#:1871
Opening the back sequence table, click on a cell, make no changes, and select OK causes headspace sequence to hang
KPR#:1872
No new lines can be added to running sequence when Unique Folder is set to Off
KPR#:1880
Editing of a running online sequence can lock the table cells
KPR#:1882
With UFC off, sequences will halt the queue
KPR#:1888
Selected data path in sequence template is not remembered.
KPR#:1915
Sequence subdirectory not working with unique folder creation OFF for GC
KPR#:1931
Data files from different sequences can end up in one result set
KPR#:1934
CSV sequence import does not validate vial positions
KPR#:1937
The icon for barcode reader does not get active after a sequence is finished
KPR#:1940
Changing the sample type in a running sequence does not activate the Apply button
KPR#:1951
Queue Sequence will not run sequences with different injection sources
KPR#:1954
GC/G1888 headspace data is missing or partially missing with ChemStation C.01.05 with HotFix installed (Hotfix 01 – 08)
KPR#:1955
Paste function in Sequence Table does not support Datafile names
KPR#:1976
Methods from different paths are not accepted in Sequences
KPR#:1978
Fill Down wizard increment logic is inverse for 100 vial plates
KPR#:1979
Sequences do not execute all runs when they contain Calibration runs for 10 levels
KPR#:1982
Fraction collection location in the sequence table vanishes
KPR#:1988
Client session frozen if user clicks "start" on paused sequence instead of "resume" for a paused sequence
KPR#:1991
Printing Sequence table with old layout shows front injector but actual injector is on back
KPR#:1998
Sequence has a Blank Option in Part of the Methods to Run in the Sequence Parameter Dialog
KPR#:2002
Entry of Numeric Sample Parameters is Not Allowed for Injections of Sample Type Calibration
KPR#:2006
Message The sequence is already running unbalanced Front/Back injector runs, therefore all remaining runs must be on the ‘Front’ Injector Location. appears
KPR#:2007
C.01.06 and higher: Sequence template changes are not saved to the \Sequence directory when ‘Unique Folder Creation Off’ is used
KPR#:2008
Cannot append a sequence line using copy, paste or append to a sequence running back injector only samples for GC
KPR#:2015
C.01.07 Vial range shown as invalid (even though they are valid) when appending sequence lines (no tray used)
KPR#:2016
C.01.07: GC back turret location (no Tray) flagged as invalid in the Filldown Wizard
KPR#:2026
The Sample Entry Properties do not retain a selected data path
KPR#:2028
“Write barcode to Sample name” does not update the Sample Name in the Sequence table during editing
KPR#:2030
C.01.07: Data file name overwritten and changed to OnlineEdited.D when updating sequence table
KPR#:2031
Incorrect message "Sequence is in use by online Session"
KPR#:2034
Sequence table locks if a dialog box is about to open while current run is completing.
KPR#:2035
Headspace sequence will not stop or pause from the ChemStation menu
KPR#:2036
C,01.06: Samples added to running sequence are named OnlineEdited<counter>.D
KPR#:2037
Number of samples "Done" count is incorrect under "Hide/show sample information" table
KPR#:2048
The maximum calibration level of 99 is not enforced
KPR#:2054
GC Data File naming is limited by the currently loaded method
KPR#:2055
Sequence rollout no longer possible, after closing first instance
KPR#:2057
Editing a sequence table during acquisition can cause the software to become unresponsive
KPR#:2070
C.01.07: GC back turret location (no Tray) flagged as invalid in the Filldown Wizard
KPR#:2071
Data file name overwritten and changed to OnlineEdited.D when updating sequence table
KPR#:2074
Cannot append a sequence line using copy, paste or append to a sequence running back injector only samples for GC
KPR#:2079
Resume button for a paused sequence is not active
KPR#:2081
Start position for fraction collection in sequence table is ignored
KPR#:2083
The Undo function does not undo changes made in dropdown cells
KPR#:2084
ChemStation hangs at the end of a run
KPR#:2087
ChemStation hangs at the end of a run
KPR#:2090
Adding lines to sequence table for Dual GC injection are in the wrong order
KPR#:2091
Duplicating the currently running sequence line can lead to overwriting of data files
KPR#:2096
Sequence stopped for unknown reason; Error 41300 occurred: invalid path shown in logfile
KPR#:2097
Incrementing locations does not work well for some samplers
KPR#:2105
Sequence hangs
KPR#:2107
Copy/paste of rows in sequence table fails, when sample information contains line breaks
KPR#:2109
Print sequence template does not show correct parameter for Data file naming
KPR#:2114
Result Set name pattern with trailing or leading blank causes issues; sequence does not start
KPR#:2117
Unable to edit the sequence table while the sequence is running in localized versions.
KPR#:2129
Sequence hangs, instrument remains in postrun
KPR#:2141
Sample Overlapped injections not possible with a GC back injector
KPR#:2144
Intermittently, sequence aborts after multiple successful runs without errors
KPR#:2154
Option to Clear Calibration Curve in Sequences does not work for multiple Calibration runs
KPR#:2165
Partial sequence does not support run-to-run relations
KPR#:2166
Nested Data folders in Preferences conflict with sequence subdirectories
KPR#:2176
Dual injection GC bracketing sequence uses incorrect method for quantitation and reports
KPR#:2180
Copy paste operations do not work as expected for locked sequence lines
KPR#:2189
Chemmain.exe crashed unexpectedly during sequence
KPR#:100307
ChemMain has stopped working, crashed when trying to read barcodes for > 100 samples
KPR#:121987
Data is not saved for PAL3 Barcoded vials; Barcodes are not written to sample name when specified
KPR#:127744
Write Barcode to Sample Name fails and the sequence does not continue
KPR#:135530
Partial Sequence fails to initialize if the default data folder is not selected
KPR#:141775
Sample type specified in the sequence table as 'control' but it is reported as 'sample'
KPR#:145403
Performance becomes very slow after deleting sequence lines in which an injection volume value was added
KPR#:148739
Modified sequence needs to be saved
KPR#:150620
Partial Sequence fails to initialize if the default data folder is not selected
KPR#:160318
Not ready timeout within sequence parameters no longer available
KPR#:161847
sequence template editing using copy and paste of large sequence tables takes too long
KPR#:163945
Opening large sequence table takes too long
KPR#:176251
User can delete pending samples in run queue when multi selection is done
KPR#:182210
Run Queue will abort if sequence uses different method with custom fields
KPR#:187826
Sequence Template (ChemStation\1\Sequence) does not match sequence that was saved with data set
KPR#:200061
Reprocessing very large result set fails to upload to Content Management (CM) with error message
KPR#:202056
Renaming a sample name in Sequence Table does not update the sample name in File Information
KPR#:206905
Running sequence aborts while scheduling another sequence
KPR#:222874
Pipe character "|" is removed from sequence table fields
KPR#:250899
Not able to print the report to the screen and to a file at the same time during a sequence run
KPR#:295239
Sequence Table cannot be closed after clearing errors
KPR#:341079
Sequence won't abort with the mismatch barcode vial containing calibration samples
KPR#:341084
No overlap observed between sequential front injection and back injection
KPR#:351385
Reprocessing very large result set fails to upload to Content Management (CM) with error message
KPR#:360927
Run queue injection location does not match sequence table of pending runs when dual blank run is the first run
KPR#:370254
Manual integration done on the Front data are showing up in the RUN.log for the back data file after reprocessing
KPR#:371040
Deadlock when closing Sequence Table user interface
KPR#:371069
Post-Sequence command was active, not executed
KPR#:375252
Sequence table that contains a Manual injection with blank vial location aborts sequence 'due to an error'
KPR#:388820
Fraction collector start field is not populated using a CSV imported sequence
KPR#:388823
Fraction collector start field is not populated using a CSV imported sequence
KPR#:484053
Screen refresh issue in the sequence table, far right columns show dark gray wipe over when edited
KPR#:665860
Unable to 'View Original Sequence' using data acquired from an earlier software revision
KPR#:789186
Start Sequence Button is not disabled when the sequence is paused
KPR#:1996
A missing vial error in an LC system with two samplers can be irrecoverable
KPR#:2098
Custom Field values can be accessed during acquisition of a sequence.
KPR#:1763
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:1996
A missing vial error in an LC system with two samplers can be irrecoverable
KPR#:2098
Custom Field values can be accessed during acquisition of a sequence.
Keyword: Sampler Configuration
KPR#:444022
Issues when re-configuring capillaries in Dual Loop and High Performance Autosamplers
KPR#:149821
Control of Tray illumination missing for G1329B & G4303A
KPR#:149822
Parameter mismatch for custom-defined Well Plates
KPR#:149824
Control of Tray illumination missing for new G1329A
KPR#:149826
Control of tray illumination missing for G1329A/B & G4303A in Japanese and Chinese ChemStation
KPR#:149827
Larger injection volume with 90 ul head may be incorrectly displayed after injection
KPR#:149831
LC driver starts run with ALS Chiller turned off
KPR#:149841
ICF A.01.04 - No method download, if method contains pretreatment/injector program steps.
KPR#:149843
Injection specific injection volumes may get propagated to following injections
KPR#:149844
Dua Loop Sampler: ACQ method shows wrong loop with overlap injection
KPR#:149847
Bouncing menu
KPR#:149850
Wrong Inj Volume with German settings
KPR#:149852
the "edit Wellplate type" could not work in OpenLab ChemStation Edition Japanese .
KPR#:149855
Wrong Injection Volume using overlapped injection and setting InjVol in sequence table
KPR#:149863
Carry over with dual Loop Sampler: Injection cleaning: Rinse volume not set in Firmware
KPR#:149880
For overlapped injections in vial samplers G1313A and G1329A/B, injection volumes and sample locations are not updated synchronously in the status UI
KPR#:149940
In OpenLAB CDS ChemStation Edition C.01.07 and below, vial positions larger than 50 are not allowed for Vialsamplers G7129A/B
KPR#:149941
Description in user interface has been improved for Smart Overlap and Overlapped injections
KPR#:149948
How to handle configuration changes in EZChrom and OpenLAB 2.0
KPR#:149954
Older versions of ChemStation cannot handle trays in Vialsamplers G7129A/B with cartesian coordinates
KPR#:149990
G1329A Autosampler tray types are not updated in Thermo Chromeleon
KPR#:149999
Sampling diagram has been removed for Infinity II samplers
KPR#:150004
Do not use instructions Inject, Valve and Remote in injector programs in case of overlapped injections
KPR#:150005
Do not change injection method parameters during overlapped injection
KPR#:150009
Under special circumstances, wrong injection volumes may be reported for ChemStation sequences of overlapped injections by the Multisampler
KPR#:150014
When a method specifies a wash position, that is physically not available, the method will not be sent to the instrument
KPR#:350975
Sampler goes offline in rare cases during run
KPR#:187313
Method download fails when cooler is controlled in method
KPR#:1973
Overlaid spectra are not properly displayed
KPR#:2069
Incorrect message: No spectra available in data file
KPR#:2156
Data points of UV spectra can carry inaccurate wavelength values
KPR#:343647
Load Signal dialog fails to load spectra because of case sensitive name comparison
KPR#:406883
No Spectra within peak available when OpenLab CDS 2.x exported UV spectra is loaded into ChemStation
KPR#:1616
Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group
KPR#:117797
No sample name for imported AIA
KPR#:158465
RegisterCS.bat is not working with C.01.08 default path
KPR#:731
Error Fetching CP Variable observed sometimes when starting the chemstation
KPR#:1574
OpenLAB Control Panel does not prevent connecting to an instrument already in use
KPR#:1577
Slow performances observed while loading datafiles when a networked printer is not accessible
KPR#:1583
Additional information about the [ChemStation:Sequence][Save Sequence] privilege
KPR#:1644
PAL Sampling diagram always shows 'Vial Status: Normal' for every vial
KPR#:1674
Graphic artifacts/software crashes caused by graphic settings
KPR#:1682
7890A, 7820A missing sampling diagram when only a rear injector is installed
KPR#:1745
Finish Queue Sequence Help is hidden behind dialog
KPR#:1746
Sampling diagram for 68xx Classic driver appears, but if selected and reselected, disappears; also when switching betw Meth&Run and DA
KPR#:1764
Run-time error ‘282’: No foreign application responded to a DDE initiate
KPR#:1815
OLCDS Chemstation - No warning before taking over session for DA (data analysis only) instruments.
KPR#:1852
Display of connected user is not updated after user switch
KPR#:1874
Takeover without privilege can result in a locked session
KPR#:1877
OpenLAB CDS client: Scrambled user interface after unlocking locked Windows session
KPR#:1886
Several dialogs are not usable with screen resolutions higher than 100%
KPR#:1947
Zoom range of chromatogram is not used when a new file is selected.
KPR#:1981
Pasting selected cells into a sequence table omits empty cells
KPR#:1984
Chinese/Japanese ChemStations: Scrollbars in Partial Sequence Dialog not fully functional
KPR#:2005
Browse method button unavailable in offline sequence table
KPR#:2033
With Unique Folder OFF a running sequence cannot be loaded in Data Analysis view
KPR#:2043
For single runs, the vial is not displayed in white in the sampling diagram
KPR#:2060
Sequence Parameters grayed out in LC sequence context menu
KPR#:2153
Zooming out of the active window can set the focus to the negative time range
KPR#:2182
Start and abort buttons on the override GC start dialog box are available when session is locked
KPR#:105359
Sample Entry User Interface does not update MultiSampler Layout
KPR#:106507
Create instrument shortcuts missing Chinese and Japanese systems
KPR#:128630
Injection info not available and Sample "Location" or "Sampling Diagram" are not updated when running an FIA Method
KPR#:130832
After time based lock, ChemStation hangs when trying to unlock
KPR#:132001
After time based lock, ChemStation hangs when trying to unlock
KPR#:135644
Sample diagram for G1329B is blank when system is configured as type LCMS
KPR#:140355
“The target path is invalid” error message is displayed when no privilege to upload files to ECM
KPR#:146183
Locked sessions appear unresponsive when minimized and cannot be accessed, must end processes
KPR#:148467
ChemStation crashes when locking or unlocking session with ECM authentication in C.01.07 SR3, SR4, C.01.08
KPR#:149082
Frequent ChemMain.exe crash or hang when unlocking sessions that were minimized in Windows Taskbar
KPR#:149911
Time table graph of UV Detector ends before run time of system
KPR#:150246
Sample Entry User Interface: Injection volume shows warning after updating location
KPR#:160280
ChemStation crashes when locking or unlocking session with ECM authentication in C.01.07 SR3, SR4
KPR#:168139
Locked sessions appear unresponsive when minimized and cannot be accessed, must end processes
KPR#:168842
PDFs Which Include Localized Characters Can Appear Garbled in Web UI Previews
KPR#:178249
Method used during reprocessing changed for no reason
KPR#:180338
First time login to the Content Management Web UI can take up to 60s
KPR#:196708
Deadlock occurs when dialog box is open while sequence is finishing
KPR#:298860
Sample Name is not displayed on the 8890 display
KPR#:303876
Data displayed in sequence table is not consistent with navigation table
KPR#:307134
Pagination disappears when search returns more than 250 results
KPR#:348531
Sample Operator shown on reports doesn't show up in user interface
KPR#:413147
Graphical controls which visualize the sampler and fraction collector bed-layout in an LC instrument environment need update
KPR#:445628
GC Sampling Diagram does not display on load
KPR#:487484
Cosmetic changes in the edit integration events UI
KPR#:490394
Resizing chromatogram window does continuous resize jumps even after let go of the mouse
KPR#:554961
Adobe Reader 2020 "sign in" dialog causes ChemStation to become unresponsive when left open
KPR#:740714
Instrument status dashboard missing important information about the module
KPR#:799726
LC module reports the host firmware revision
KPR#:2041
Unprivileged users can change instrument setpoints using the "GC Instrument Diagram"
KPR#:58
Activity Log not displayed in web browser when using Chrome 5.1
KPR#:1688
Preferences set in offline overwrite online Preferences and vice versa
KPR#:1957
Local files are not stored securely with ChemStation Edition
KPR#:1763
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
Keyword: User Interface , 7890A Control
KPR#:2041
Unprivileged users can change instrument setpoints using the "GC Instrument Diagram"
Keyword: User Interface , Audit Trail
KPR#:58
Activity Log not displayed in web browser when using Chrome 5.1
Keyword: User Interface , Data Analysis
KPR#:1688
Preferences set in offline overwrite online Preferences and vice versa
Keyword: User Interface , ECM Integration
KPR#:1957
Local files are not stored securely with ChemStation Edition
Keyword: User Interface , SEQ , ACQ , ECM Integration
KPR#:1763
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:2157
Manual integration tool is still accessible after a user change
KPR#:2163
Removal of "Edit integration events" permission will not disable access to the "Specific Event For Signal"
KPR#:2172
"User"s and Group's Role Assignment Report" will not display users with group but no roles assignment
KPR#:2183
Missing ChemStation privileges when connected to OpenLAB Server 2.1 with OpenLAB CDS 2.1
KPR#:115034
Users with only the view and run instrument privilege can take over the online/offline used by other user
KPR#:116731
User can switch to DA-view when method change flag is set and run queue is paused
KPR#:120340
C.01.07 SR1 User can delete pending samples in run queue that is paused when multi selection is done
KPR#:123269
Missing privileges for NWS chemstation connected to OpenLAB 2.1 server with OpenLAB CDS 2.1
KPR#:130664
Lock and Unlock during running sequence may lead to deadlock / user permission loss
KPR#:146287
User without save method changes privilege can save changes to the method
KPR#:146402
User without save method changes privilege can save changes to the method
KPR#:165817
User with more privileges cannot preview report without saving method than a User with less privilege
KPR#:187099
Exception to "Edit Calibration Table" privileges
KPR#:381733
Orphan Property ID leading to sync failure
KPR#:801721
User without specific Project Group access can see the name of the Project Group
KPR#:1902
The "Import worklist" feature does not work when Chinese characters are present
KPR#:1990
Error message when selecting "Print Report" for an AIA imported file, report not shown
KPR#:2046
Zoom level is lost when overlaid UV and MSD data are copied
KPR#:2124
No sample name for imported AIA
KPR#:1616
Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group
Keyword: Utilities , Spectra Analysis , MSD Analysis
KPR#:1616
Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group
Keyword:
One-line Description:
Custom DPI Setting
Problem:
When increasing the dpi settings, the purification software cannot be used since many controls overlay each other and error icons are not visible.
Temporary Solution:
N/A
Fix Information:
The supported DPI Setting is the default one, 100%. No other DPI settings is currently managed by the Purification software.
Keyword:
One-line Description:
Common & channel# tab order mixed up when a channel is deactivated
Problem:
[OpenLAB driver] common & channel# tab order mixed up when a channel is deactivated
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Pump settings not set to initial value after run and/or after loading a Method
Problem:
During run a stop is requested to the Software. Opening the User Interface where the system is shown the Initial pump parameters were not loaded. It was still A: 99.84% B: 0.16% instead of A:100% and B:0%. Loading the Default method still the gradient is wrong displayed in UI.
Temporary Solution:
n/a
Fix Information:
there's any fix for this issue. The simptoms was very seldom and random. Restarting the PC and the Modules will erase the wrong information displayed on the User Interface.
Keyword:
One-line Description:
4 Channel 490 Micro GC is not configured properly
Problem:
Tried to configure a four channel 490 micro GC. After clicking on Activate Communication button, the Configure screen indicates that communication has been established, but the information for the 4 channels has not been uploaded. Information is provided when connecting with PROstation.
Temporary Solution:
Config 490GC remotely, but need to wait about 60 seconds for driver to get all configuration from remote 490GC
Fix Information:
This defect is fixed in build 1.11.1.1003 .
Keyword:
One-line Description:
LC-3D, CE, and MSD technique-specific columns missing
Problem:
The columns that are missing include: LC-3D and MSD: Target Masses (number, range, list) and AutoBalance (blank or ‘OFF’) CE: Voltage (number), StopTime (number), and User1-10 (1-48 or blank)
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
6890: Add\Advanced Tower Selection Syring Volume does not update
Problem:
6890: Add\Advanced Tower Selection Syring Volume does not update until session is closed/reopened. For any Add step, select Advanced. The volume determines which tower will be used if automatically select by volume is selected (default). The Syringe Volume of that injector is displayed read-only below it. If you go to the Configuration and change the syringe size for any tower, then add a new Add step, save and return, this syringe Volume change is not detected or displayed. It is only refreshed by closing and reopening the online session of ChemStation.
Temporary Solution:
Close and reopen the session to update the syringe volume.
Fix Information:
n/a
Keyword:
One-line Description:
Methods can not be stored in Micro GC after rebooting
Problem:
After a power shutdown, the uGC does not load the last active method, but remains in the default settings (temp 30C). A method download is required.
Temporary Solution:
n/a
Fix Information:
This defect is fixed in build 1.11.1.1003 .
Keyword:
One-line Description:
Calibration Blanks from Easy Sequence using an Easy SamplePrep program with 'Use for unknowns only' checked reserves resources
Problem:
No Injection Calibration Blanks from Easy Sequence using an Easy SamplePrep program with 'Use for unknowns only' checked reserves resources, even though the Easy SamplePrep program does not run in this case.
Temporary Solution:
Supply an excess of resources to allow the sequence to be submitted to the Queue.
Fix Information:
n/a
Keyword:
One-line Description:
Multiple fractions collected by MSD within one peak due to ionic species
Problem:
Multiple fractions could be collected within one peak due to different ion species of similar intensity for the same target.
Temporary Solution:
Check results of analytical results evaluation, what adducts apply to what compounds / samples. Create separate tasks for each ionic species. Uncheck "Use same ion species as in analytical run" in preparative system ion species of the task. Within one task you select only one adduct in preparative system and select all samples, which compounds have a good response for this adduct.
Fix Information:
n/a
Keyword:
One-line Description:
Sequence location directory selection is not correct
Problem:
When creating an Easy Sequence template and select the sequence location funcion, the software displays the normal windows function for selecting a directory / sub-directlry. This is wrong. The software should show the selection of sequence locations as specified in the sequence preferences.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
ChemStation: Vials are validated against the currently loaded method
Problem:
In the Easy Sequence Setup, a method is selected and then the vial numbers are entered. These vial numbers are passed to the data system to be validated. The validation occurs against the currently loaded method, not the method used in the template that will actually use the vials.
Temporary Solution:
n/a
Fix Information:
Fixed in Easy Sequence 2.4[002].
Keyword:
One-line Description:
Logbook file: no string 4234
Problem:
Empty Bottles will cause the window with the "no string 4234" to popup during a Task execution.
Temporary Solution:
Verify the level of solvents prior to execute the task.
Fix Information:
Fixed in ChemStation C.01.06 requires Purification Software A.01.02.
Keyword:
One-line Description:
Calibration Blanks with injection run Sample Prep when 'Use for unknowns only' is checked
Problem:
Calibration Blanks with injection run Sample Prep when 'Use for unknowns only' is checked
Temporary Solution:
Use a different methods for blanks that has 'Use Sample Prep program' unchecked
Fix Information:
n/a
Keyword:
One-line Description:
Custom Fields are missing in Extended Parameters dialog
Problem:
Sample Custom Fields and Compound Custom Fields are missing for Extended Parameters.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Stop running tasks in purification Software
Problem:
When a Task is running the "Stop" will lead to a Task contents to be not complete. The ACAML file created at the end of the acquisition is not present. The sample present during the acquisition is not listed.
Temporary Solution:
Analytical runs by purification software, will not display analytical results in case of STOP. Workaround: Create a new task and import the analytical run as result set: Stopped sample and later samples will be missing, but first samples will be displayed. Preparative runs by purification software will not display stopped sample and beyond.
Fix Information:
Fixed in ChemStation C.01.06 requires Purification Software A.01.02.
Keyword:
One-line Description:
Bracketing with unique folders off
Problem:
When bracketing is used with Easy Sequence with Unique Folder Creation OFF, data file names can not be generated for the calibration runs. The potential exists to overwrite files when using Bracketing with Unique Folder Creation OFF with Easy Sequence.
Temporary Solution:
When using Bracketing with Easy Sequence, use Unique Folder Creation ON from Preferences.
Fix Information:
n/a
Keyword:
One-line Description:
6890 Easy SamplePrep Add step fails for both by use and by volume resources
Problem:
If there is an ‘Add’ step in Easy SamplePrep program with just the ‘Solvent Wash’; even though the Back injector is to be used, it may not use the Back injector.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Resource layout on 7890A GC instrument dialog is not translated
Problem:
Resource layout on 7890A GC instrument dialog is not translated into Chinese and Japanese.
Temporary Solution:
n/a
Fix Information:
Fixed in revision A.02.03.
Keyword:
One-line Description:
PAL3 method fails at runtime when an inject amount is out of range for the installed syringe
Problem:
If you have a 10 ul syringe installed and specify a 10 ul injection volume, the system will not indicate any problem until runtime. This is because the driver cannot look into the script to determine total volume (ie sample volume and air gap). This calculation is done at run and the error shows then.
Temporary Solution:
Use has to calculate total volume required when setting up the method.
Fix Information:
n/a
Keyword:
One-line Description:
Easy SamplePrep Method printout is not translated
Problem:
Easy SamplePrep Method printout is not translated.
Temporary Solution:
n/a
Fix Information:
Fixed in Revision A.02.03.
Keyword:
One-line Description:
Manual/Valved/HS templates cannot be loaded if inj source does not match loaded method
Problem:
6890 with no ALS or tray. Configured to a HeadSpace. Create a front valve method, a back manual method, and a HeadSpace (front) method. Depending on the method loaded, load an .est and this fails with: '--------------------------- Inconsistent Easy Sequence Setup --------------------------- Sequence requires a back injector not present on the instrument. --------------------------- OK ---------------------------' The message really does not advise the user of what should be done to correct this situation.
Temporary Solution:
This is current ChemStation behavior. The template created in Easy Sequence must use the currently loaded method.
Fix Information:
n/a
Keyword:
One-line Description:
Resources replenished on 7890A GC instrument dialog not translated
Problem:
Resources replenished on 7890A GC instrument dialog not translated
Temporary Solution:
n/a
Fix Information:
Fixed in revision A.02.03.
Keyword:
One-line Description:
Certain combination of setpoints cause setpoint error from tower
Problem:
The software allows the user to enter invalid parameters for the draw and dispense speeds for the syringe. Using an "Add" command it generates a "Command to XXX Tower failed: XXX Tower Prep Failure Code: 142: Autosampler setpoint error..." The error message is limited and it takes effort to find the "invalid" parameter.
Temporary Solution:
n/a
Fix Information:
Fixed in A.03.01.
Keyword:
One-line Description:
Easy Sequence Setup does not display dual sources on method load; must Clear All to activate
Problem:
Open online session to a 6890 with dual ALS (Plasticman 6890A/7693). DEF_GC.M is configured by default to front injector. Set injection source to Dual and save as a new method, now loaded. Select Easy Sequence Setup. No injection sources are displayed, (presumably because of the injection source selection of the previously loaded method had one injector.) Load the Dual method. Still no injection sources.
Temporary Solution:
The user must select 'Clear All' to refresh the screen in order to display the Front/Back injection source checkboxes after loading the dual injector method.
Fix Information:
n/a
Keyword:
One-line Description:
No Fraction Results Table if sample name + data path length exceeds 103 characters
Problem:
[EDTAB60: could not set window title because register was locked] With or without Purification Software having a combination of sample name and result set path, which exceed the ChemStation limit of 103 characters. In that case no fraction results table is displayed in ChemStation Data Analysis View: Purify and error message in the ChemStation status line: "EDTAB60: could not set window title because register was locked"
Temporary Solution:
Workaround: Use short sample and folder names. Use purification software to review fraction results instead of ChemStation.
Fix Information:
N/A
Keyword:
One-line Description:
No peak detector - Fraction Collection tab inaccessible
Problem:
When a task evaluation results in No peak detector error (e.g., MSD was on and UV off but no mass was entered) the appropriate error occurs but Fraction Collection tab is Task Configuration is inaccessible
Temporary Solution:
Change any parameter on systems or scale up task configuration pages. This will reset the run state of the task and the fraction collection task page will be accessible, again.
Fix Information:
Fixed in Purification Software A.01.02.
Keyword:
One-line Description:
Need ability to 'shake' with Mix command
Problem:
The Mix command needs some more detailed control. The mixer to is used to shake a vial (Mix 2500 rpm 1 sec, stop 1 sec , and change direction mix 2500 1 sec, repeat), this cannot be done with the current mix command.
Temporary Solution:
n/a
Fix Information:
Fixed in revision 2.3.
Keyword:
One-line Description:
Clear fraction collector delay volumes for unused peak detectors
Problem:
Using the "Time table" > "Trigger time windows" for fraction triggering and a [Peak Detector] is setup for fraction collection, which is not used for triggering but has a non-zero delay volume setup in the fraction collector configuration could happen that fractions are not collected. This happen, if you were using the fraction collector in a previous instrument configuration with a different peak detector or with another detector connected to UIB. If the previous [Peak Detector] is no longer in the current instrument configuration, but its delay volume has not been reset to zero in the fraction collector by LabAdvisor, then the delay volume still contributes in correcting time windows. Time windows are shifted (corrected) by the maximum delay time of all peak detectors, which are setup in the fraction collector.
Temporary Solution:
If you remove a peak detector from your instrument configuration, make sure that any delay volume that is set to it in fraction collector is reset to zero using LabAdvisor.
Fix Information:
Fixed in Purification Software A.01.02.
Keyword:
One-line Description:
Driver 1.12.3.1008 installation may cause error in RCNet
Problem:
The direct cause of this issue is the absence of the VC++ runtime library.
Temporary Solution:
A quick solution is to manually install the missing runtime lib. Please download the Visual C++ Redistributable for Visual Studio 2015 (x86) installer and install it. https://download.microsoft.com/download/9/3/F/93FCF1E7-E6A4-478B-96E7-D4B285925B00/vc_redist.x86.exe
Fix Information:
Update gc_dll to 1.40.0.9, downgrade the dependency runtime to VC100 to avoid some runtime missing issue.
Keyword:
One-line Description:
Instrument ChemStation Logbook entry for DAD Internal error 10010
Problem:
Randomly the ChemStation logbook is reporting the "DAD internal error 10010". The issue is due to few spetra lost during the acquisistion. The Automated Purification Software is capable to manage this small lack of data. Calculations and "Target Compound" identification are working as expected.
Temporary Solution:
N/A
Fix Information:
No fix planned.
Keyword:
One-line Description:
Easy Sequence Setup does not validate HeadSpace vial ranges
Problem:
HeadSpace Method loaded in ChemStation. The Injection Source is set to HeadSpace Sampler Easy Sequence Setup Sample Information Starting Vial Location Any number entered into this location will error with the following error message: The vial position could not be validated against the current configuration
Temporary Solution:
n/a
Fix Information:
Fixed in Easy Sequence 2.4[012].
Keyword:
One-line Description:
Subsequent cuts require a short gap
Problem:
If multiple cuts are scheduled adjacently (i.e. start time + sampling time = next start time), some cuts may got lost randomly because the valve needs a short time for switching. The gradient preview does not show a warning in such cases.
Temporary Solution:
Leave a gap of at least 0.01 minutes (0.6 s) between each two adjacent cuts. This can be done by either slightly decreasing the sampling time or increasing the start time for the next cut.
Fix Information:
This has been fixed in firmware A/B/C/D.07.01
Keyword:
One-line Description:
Export selected fractions with long ChemStation sequence file names cause ChemStation crash, at end of execution
Problem:
Export a fraction reanalysis file as ChemStation sequence file for selected fractions. Have the specified file name longer than 40 characters, or at least long enough that together with the maximum automatic appendix "_collectorX_Y", where X is the collector number, Y is the file number in case of 4 plates, the complete sequence file name exceeds 40 characters in length. Execution of such sequence file will run, but at completion time it will cause ChemStation to crash.
Temporary Solution:
Do not specify sequence file names longer than 27 characters in the export fraction reanalysis dialog box.
Fix Information:
Fixed in Purification Software A.01.02.
Keyword:
One-line Description:
Add step fails to execute due to mismatch in # injections/volume/syringe size/solvent allowed by firmware
Problem:
Method with a single add step of 1000 uL. Syringe size is 10 uL. Vials are moved to turret and then error message displays: method not running. This is a mismatch in # injections/volume/syringe size/solvent allowed by firmware.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
A method using Easy Sample Prep cannot be used to create a method not using Easy Sample Prep
Problem:
Create a method which has the "Use Sample Prep Program" checkbox enabled. Use this method to create a method which does not have the "Use Sample Prep Program" checkbox enabled. Load the method which does not have the checkbox enabled. Error message displays: Configuration mismatch: Default parameters loaded for 'Agilent 7890'.
Temporary Solution:
n/a
Fix Information:
Fixed in 3.1[21]. This revision will only fix with Openlab Chemstation C.01.03.
Keyword:
One-line Description:
Export selected fractions injection in CS sequence file is set to 10 uL ignoring the re-analysis method injection volume
Problem:
When exporting fraction reanalysis sequence file in ChemStation format, the injection volume is set to 10 ul as default.
Temporary Solution:
Open the sequence table of the exported sequence in ChemStation and fill down the desired injection volume.
Fix Information:
Fixed in Purification Software A.01.02.
Keyword:
One-line Description:
Unable to unlock ChemStation / Purification Software, if purification software dialog is displayed
Problem:
If an inactivity time to lock application has been set in OpenLAB Control Panel, administration, then Chemstation usage will be protected by a lock dialog after the set time of inactivity. Purification software dialogs hide the lock dialog, but are locked themselves. Therefore it is not possible to unlock the application (ChemStation and purification software).
Temporary Solution:
Set the “Inactivity time before locking the application” to zero in OpenLAB Control Panel > administration > security policy; if this complies to your security policy. If ChemStation and purification software got locked, you can access the unlock dialog with help of the Windows Task Manager. Start the task manager, go to the applications screen and select the Purification Software task window, which is called Task XY - Set Up and Run Purification Tasks. Right-click this application and select "Minimize". Then select the Agilent OpenLab Login window from the application list of the task manager, right-click and select "Bring to Front". Then log in as normal for getting back access to ChemStation and purification software.
Fix Information:
Fixed with Purification Software A.01.03. It is possible that the ChemStation lock dialog appears behind a Purification Software (system or task) dialog. In such case, you can move or minimize the Purification dialog, or select the ChemStation window from the Windows Taskbar in order to access the ChemStation lock dialog to unlock.
Keyword:
One-line Description:
Delay volume setting can cause delay in method download for preparative runs
Problem:
Purification Software sets the fraction collector peak detector delay volume in fraction collector firmware, if the current volumes in firmware do not match the preparative system settings in the purification task. This causes a configuration change of the fraction collector, which could lead to a delayed download of the initial method of the preparative run and temporary offline status of the fraction collector. This could take up to a few minutes. During this time, the pump will run with the set preparative flow rate. The likelihood of this situation increases with the number of delay volumes to be set, i.e. are different from the actual settings in the fraction collector; and the number of peak detectors or fraction collectors which are configured. At the end of the standby sample run, all delay volumes that had been changed before the preparative run are restored to their previous values - as part of the post-run actions. This could lead to a timeout and thus failed reporting of the standby sample run. The run queue is paused during post-run. Due to a delay volume restore, it is possible that the queue is still paused while another scheduled task is trying to start. Task runs are rejected with a message box, when the queue is paused.
Temporary Solution:
The delay volume setting by purification software is needed for changes in delay volumes by different flow rates. If your flow rate is fix or does the delay volume does not depend on the flow rate (no usage of active splitter), set all peak detector delay volumes by LabAdvisor for your configured fraction collectors. If a UIB is configured as well, but not used as auxiliary peak detector for fraction triggering, then set its fraction collector delay volume to zero in LabAdvisor.
Fix Information:
With Purification Software A.01.03 there is no need to run LabAdvisor to perform the delay volume calibration or set the peak detector to fraction collector delay volume. The delay time/volume calibration can be executed from the "Preparative Systems" dialog. If you want to apply the delay times/volumes from your preparative system parameters to the next purification runs, just click the "Write delay times" button from the "Preparative Systems" dialog. This avoids setting for the values before the purification sequence run and restoring the previous values afterwards. If you want the values to be set and restored for only one purification run, purification software is waiting for the configuration change to finish displaying "Waiting for configuration update in ChemStation" in its status line, before submitting the purification run.
Keyword:
One-line Description:
Results from stopped analytical runs are not used for preparative runs
Problem:
If an analytical run has been stopped, the result may be incomplete. The purification task logbook tracks the run states. Analytical data source is set to "read analytical data" on task configuration page "Analytical run" after the (stopped) analytical run to indicate where data had been produced. If you proceed with the purification task run and "read analytical data" for the stopped analytical run, the analytical run will be restarted.
Temporary Solution:
If you like to restart the analytical run anyway, then purification software is behaving as expected. If you would like to use the available data from the stopped analytical run, please set up a new task from the same template as your original task and set the analytical data path of your stopped analytical run in the "read analytical data" field on the on purification task configuration page "Analytical run".
Fix Information:
With Purification Software A.01.03, results from stopped or failed (analytical) runs are kept with the task. Therefore, you can proceed with a preparative run for the available samples from a failed or stopped analytical run. If you wish to (re-)run the missed analytical samples of a failed or stopped analytical run: Create a new task as a clone from the task with the failed or stopped analytical run. In the analytical run setup of the cloned task: Click on "Revert to sequence" and select the samples to re-run by editing the sequence.
Keyword:
One-line Description:
HeadSpace Injection Volume is accepting negative numbers; no checking; no meaning
Problem:
Create an Easy Sequence Template for a G1888/6890. There is no error checking for Injection volume for this injection source in either the Setup or Sample List. All values are accepted, including negative numbers. There is no place to setup an 'injection volume' for a HeadSpace. There is only a time and pressure.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Easy Seq and Setup Starting Vial Location and Number of Samples do not check for limits
Problem:
Starting Vial Location and Number of Samples do not do any range checking against any method or configuration.
Temporary Solution:
n/a
Fix Information:
Fixed in revision 2.3
Keyword:
One-line Description:
Easy Seq Setup: Sample Name text entry should check for a Counter if used as Data File name
Problem:
Easy Sequence Setup: Entry of text in the Sample Name combo box does not check for and advise that a counter token is needed if the user also uses the token 'Sample Name' for the Data File Name. The user will find out when sending to the queue 'Error: Sample Name xxxx occurs more than once.'
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Easy Sequence Setup not refreshing the Injection Location
Problem:
7890/7693: Create methods for front, back, and dual ALS injection sources. Easy Seqeunce Setup does not refresh to show the correct injection location of the loaded method until user loads the corresponding method in ChemStation. Even loading a different method, the previous injection source is kept.
Temporary Solution:
If the injection source has been changed, use the "Clear All" button and reload the method.
Fix Information:
n/a
Keyword:
One-line Description:
No error checking for vial ranges in Easy Sequence Setup; fails at runtime
Problem:
There is no error checking within Easy Sequence Setup to determine if the vial range of the loaded method's injection source is accurate. The method can be saved and submitted to the queue. It fails at this time with: '--------------------------- Error --------------------------- Invalid location found in FRONT sequence, line 1 There may be others --------------------------- OK ---------------------------' This defect occurs because there is a difference between the loaded method's injection source and the method that is selected in the Easy Sequence Template.
Temporary Solution:
Load the method in ChemStation from Method > Load Method menu that will be used to create the Easy Sequence Setup Template.
Fix Information:
n/a
Keyword:
One-line Description:
Pause/Resume button is in the wrong state after a sequence is edited and then pauses
Problem:
Send a sequence to the Queue and while it is running, edit it. This modified/changed sequence pauses the Queue, but the first blue button is still in a 'Pause||' state. Click the button and change it to Resume, but it does not do anything as it is waiting for the current sequence to be saved before starting the next sequence. In this case, because the sequence has been modified, the ChemStation is not accepting sequences. If you click pause || button, you pause the queue and it does not do any thing. The message on the active queue is red and the message says the sequence has been modified.
Temporary Solution:
Save the sequence that has been modified. Once the sequence has been saved, the queue will resume running the next sequence.
Fix Information:
n/a
Keyword:
One-line Description:
7890A with no ALS shows Read Barcode button in Easy Sequence Sample List
Problem:
Load an Easy Sequence.es for a valve or HeadSpace Injection Source and the Sample Name cell in the Sample List shows a Read barcode button.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
HeadSpace Simple Calibration Blank type should only allow 'Blank with injection'
Problem:
HeadSpace software will not run a sequence without a vial location. Therefore, the Easy Sequence Setup\Calibration\Simple Calibration options should only allow Blanks with Injection and not allow blanks without injection.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Calibration Blank with no injection loads with Inj Vol as 0
Problem:
Create a setup with a blank with injection and a blank with no injection. Save and load to Easy Sequence. The Injection volume for the Blank with no injection is entered as 0. Native sequence table also shows this as zero. This is not consistent with the volume for the blank with injection
Temporary Solution:
n/a
Fix Information:
Fixed in revision 2.2
Keyword:
One-line Description:
Calibration table clears on method reload from Samples tab
Problem:
Create a Setup with Calibration. Return to Sample tab and load the same or a different method with the same injection location. Return to Calibration tab and all has been cleared. Reloading a method does not clear entries made on the Samples tab.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Queue displays running vials incorrectly when initial HeadSpace vial not found
Problem:
Start a 3 line headspace sequence from the queue, but remove the FIRST vial in the sequence. Queue does not update correctly, showing the first vial as running, though the HeadSpace has moved on to vial 2.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Aborted HeadSpace run displays Status as 'Complete' in Queue History
Problem:
HeadSpace aborts, but ChemStation does not. Select the Abort Run button from the Waiting for ready dialog and ChemStation aborts the sequence. Queue History shows this run as 'Complete' instead of 'Aborted'.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Easy Sequence does not advise of invalid vial locations until sequence is fully loaded into Queue
Problem:
There is no advisory about the invalid vial locations until the sequence has been added to the queue.
Temporary Solution:
n/a
Fix Information:
Fixed in revision 2.3
Keyword:
One-line Description:
Sequence submitted to queue but will not run and can not be removed/deleted
Problem:
Sequence submitted to the queue but will not start and cannot be deleted.
Temporary Solution:
As a workaround, delete the following file: SequenceQueue.que from the C:\Chem32\4\TEMP directory.
Fix Information:
Fixed in revision 2.4.
Keyword:
One-line Description:
Injection Volume on ES Setup should show units (uL); valved inj are in mL and user must convert
Problem:
The Easy Sequence Setup dialog should show Injection volume units, which in ChemStation always seem to be uL. When the user uses EasySequence with a Valved injection, the volume, which for valves is mL, must be multiplied by 1000 to get a value that will be accepted.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Easy Sequence Setup should check for Sample Name length and not allow invalid entries
Problem:
Enter a long Sample Name in Easy Sequence Setup and add a counter. Save and load to Easy Sequence. Fill Samples. Not until then is the user warned (with a red X) in the Sample list that there is a 40 character limit. It is possible to save this but it will not be possible to add to the Queue.
Temporary Solution:
n/a
Fix Information:
Fixed in Easy Sequence 2.4[014].
Keyword:
One-line Description:
Easy Sequence Setup: Change Inj/vial to a value >1; all labels and menu bar disappear
Problem:
Select Easy Sequence Setup. Enter a number greater than 1 in the Inj/vial cell on the right side and then tab. All labels and menu items disappear.
Temporary Solution:
Restart the session.
Fix Information:
n/a
Keyword:
One-line Description:
Vial Status shows first run as completed for Easy Sequence following Run Method
Problem:
Run a single vial using Run Method. Send sequence to queue vials 2-4. When the run method completes and the sequence loads and then the first vial in the sequence starts, Vial status shows the first line (vial 2) as immediately 'completed'. The system stalls at Sequence running. The GC is in prerun state and the system cannot start another run until it is shutdown and restarted.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Easy sequence actions are not included in the audit trail and no separate audit log exists
Problem:
Any changes done on Easy Sequence are not included in the audit trail; there is no separate audit log present for Easy Sequence.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Vial not found reported in software Status not consistent with instrument status
Problem:
Run an Easy Sequence using a HeadSpace injection source. When a Vial not Found event happens, the final Status for this run, which did not start a run, is reported as 'Completed'. The status being reported by the HS device is 'Run Aborted'.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
PAL autosampler plate/vial locations not accepted in Setup location box
Problem:
Easy Sequence 2.01.41 on Chemstation B.04.03 DSP1 will not accept plate designations of PAL injection source in the Setup location combo box. (P1A1) The user should be able to choose the location designation.
Temporary Solution:
n/a
Fix Information:
Resolved in revision 2.4 [12].
Keyword:
One-line Description:
Print preview screen does not appear when blank is added to calibration program
Problem:
1. Method and Run control screen> easy sequence settings tab > calibration tab 2. Select "simple calibration" for calibration mode, add Blank to the sequence. click print button to print preview easy sequence settings. Print preview screen does not appear.
Temporary Solution:
n/a
Fix Information:
Fixed in Easy Sequence 2.4[30].
Keyword:
One-line Description:
Headspace continues after stop sequence is clicked from the ChemStation menu or Sampling Diagram
Problem:
Select Stop Run/Inject/Sequence from the ChemStation menu or from the Sampling Diagram. This shows yellow 'Sequence Stopping//Data Acquisition' but the headspace sequence continues running. Headspace continued to process vials and generate data until end of sequence.
Temporary Solution:
n/a
Fix Information:
No fix planned. The headspace (G1888 or 7697) cannot be stopped or paused from the ChemStation menu or Sampling Diagram. It can/must be paused from the headspace (HS) Status bar ‘Pause Sequence’.
Keyword:
One-line Description:
Injection Volume is reported as 0 uL or blank in report for headspace
Problem:
The injection volume is reported as 0 uL or blank in the data system report. This works by design since headspace loops are not calibrated and the sample injected is not known by the data system.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
6890 method does not allow SAVE DATA to be checked
Problem:
It is not possible to save data in a 68xx method.
Temporary Solution:
A workaround is to edit the column nominal dimensions (length and diameter) to different values and download them. The method editor will change the internal values stored in the methd to valid values. The column nominal dimensions can then be edited to the correct values and downloaded to the GC. The problem should not recur if that is done once and all parameter values are edited through the data system with driver version 6.21 from then on.
Fix Information:
Resolved in 68xx Driver 6.23 [009].
Keyword:
One-line Description:
C.01.06 ChemStation with GC/G1888: last run completes but the ChemStation remains in the run state
Problem:
OpenLAB CDS ChemStation Edition C.01.06 with a G1888 sequence never finishes. The last vial completes but the sequence continues to be in the run state and must be stopped manually. Once stop has been selected, allow time for the system to process and cleanup. Try to avoid using abort. Use of ABORT almost immediately terminates all ChemStation activities and processes. This has the potential to leave any currently running sequence in a strange state. Also any consecutive Data Analysis is aborted. When the ChemStation is connected to central storage (OpenLAB ECM or OpenLAB Data Store), no files will be uploaded to the repository after aborting the sequence. Loss of data may occur. After using the Abort function, close and restart the ChemStation Application. Always try STOP first and allow enough time for consecutive processing and cleanup.
Temporary Solution:
n/a
Fix Information:
This problem is addressed with Integrated Headspace Software revision B.01.06.
Keyword:
One-line Description:
Headspace system acquires all data into a single data file
Problem:
Intermittently, all of the headspace runs are collected in a SINGLE GC data file. The ChemStation never closes the raw data file for the first run and all further runs are acquired into the first data file. The first run never completes in the ChemStation, but the Headspace continues with further injections.
Temporary Solution:
Increase the GC cycle time.
Fix Information:
n/a
Keyword:
One-line Description:
User is not able to save 'Column Parameter' changes at the same time when change is made
Problem:
User is not able to save 'Column Parameter' changes at the same time when change is made unless they click on another field on the screen
Temporary Solution:
n/a
Fix Information:
Resolved in B.01.02
Keyword:
One-line Description:
Invalid Resource Item implementation for oven program
Problem:
Invalid Resource Item implementation for oven program
Temporary Solution:
n/a
Fix Information:
Resolved in B.01.02
Keyword:
One-line Description:
The peak inflection point position does not match classical ChemStation reporting.
Problem:
The peak inflection point position is drawn different compared to the classic reporting. This is because there are multiple valid positions where the point can be drawn. The effect can be observed for peaks that show only very small changes of the slope over a "longer" time range and the 2nd derivative is close to 0 over that range.
Temporary Solution:
The reported peak performance values are the same in classic and Intelligent Reporting. Users who prefer the old drawing need to use classic reporting which is still available in ChemStation C.01.01.
Fix Information:
There is currently no fix planned
Keyword:
One-line Description:
Exponential calibration curve plot isn't shown in Calibration Report when using Intelligent Reporting.
Problem:
Exponential calibration curve plots cannot be reported with Intelligent Reporting. An empty plot is shown instead.
Temporary Solution:
If exponential calibration curves plots need to be reported, the (old) classic reporting needs to be used in ChemStation. EZChrom users need to use the EZChrom standard reporting.
Fix Information:
A potentially fix is currently under investigation.
Keyword:
One-line Description:
LC/MS spray chamber temperatures are no longer be applied to the spray chamber heated zones following a spray chamber open/close action.
Problem:
After an LC/MS spray chamber has been opened and closed, temperature will no longer be applied to the Spray Chamber heated zones until the spray chamber parameters are updated (by clicking "ok" in the Spray Chamber Parameters dialog box or loading a method) or the system is turned "off" or to "standby" and then back to on. This occurs with the APCI, APPI and Agilent Jet Stream spray chambers. The MultiMode source will set the system to "Standby" if opened & closed, however.
Temporary Solution:
Turn the instrument to "off" or to "standby" in ChemStation, then back to "on" if this condition is encountered.
Fix Information:
n/a
Keyword:
One-line Description:
UNC paths are not supported as report file storage loaction
Problem:
The "copy to" function of the intelligent report method setup does not support usage of UNC paths.
Temporary Solution:
Map the network path as a drive letter and select the new drive letter as a report destination.
Fix Information:
There is no fix planned.
Keyword:
One-line Description:
Multiplier is applied on compound group ISTDs
Problem:
Create a compound group, select it as ISTD. In dialog "Specify Report...", tab "Quantitation Settings", uncheck "Use multiplier and dilution factor with ISTDs" Select "Use sample data From Sample Defaults below Enter multiplier different from 1.0. Multiplier is still applied on compound group. It should not be applied, as for normal ISTD compounds
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB CDS C.01.06.
Keyword:
One-line Description:
Intelligent Report Method Information does not correctly display calibration units when units changed in Calibration Table
Problem:
When adding the Method Information section to an Intelligent Report, the calibration amounts units are not correctly displayed if the value was changed in the method.
Temporary Solution:
This issue is in the method information section of the Intelligent Report template. There is no way to modify this value in the report template.
Fix Information:
Fixed in OpenLAB CDS C.01.06.
Keyword:
One-line Description:
78xx Stop button pressed no longer continues to the next run in the sequence
Problem:
Pushing the stop key at the GC keyboard will stop the current run and generate a report but does not continue to the next line in the sequence. This occurs beginning on ChemStation version B.04.01 SP1. In revisions B.03.01 - B.04.01, pushing the stop key at the GC Keyboard would stop the current run and generate a report and continue to the next line in the sequence.
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB CDS C.01.06. Setup conditional System Behaviors for the 78xx in Manage Rules & Alerts found under the Instrument menu -Continue -Abort
Keyword:
One-line Description:
Signal options "Each in full scale" does not work correctly
Problem:
Under Data Analysis, select Graphics Range :select time range Multichromatogram:Saparated Scale:Each in Full Scale Each in Full Scale does not get used
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB CDS C.01.06.
Keyword:
One-line Description:
acq.macaml for 7890B is displayed as 7890A
Problem:
In the acq.macaml of an 7890B, there is always the > ModuleDisplayName 7890A
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB CDS C.01.06.
Keyword:
One-line Description:
Sequence is aborted soon after the start and must be restarted
Problem:
Started a sequence (55 injections), but the sequence was aborted soon after the start. Chemstation showed status "ready" and instrument showed status "idle". The sequence could be restarted and ran without error after the restart.
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB CDS C.01.06.
Keyword:
One-line Description:
Baseline not displayed properly when using ChemStation AIC’s
Problem:
A nearly horizontal baseline is not displayed or printed completely when doing a Remote Desktop connection to a ChemStation WorkStation or to a ChemStation AIC.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
User without Permission to Change Transfer Location Can Change Transfer Location in Sequence Parameters
Problem:
In OpenLAB Shared Services, the user does not have any [ChemStation Security] privileges, specifically they do not have the [Modify Storage Transfer Settings]. But in [Sequence Parameters] the user can still uncheck [Remote Data Path] [Use Preferences] and then select the [Browse] function to select a new LCDF location for the instrument data transfers.
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB C.01.06.
Keyword:
One-line Description:
The Active Splitter does not always stop at the end of a sequence when "Run Sync" is enabled in the Active Splitter Software
Problem:
The Active Splitter is not always stopping at the end of a sequence when Run Sync is enabled . With "Run Sync" enabled, the operation of the Active Splitter should be controlled through ChemStation/Active Splitter software. Intermittently, the software control of the Active Splitter does not stop its operation at the conclusion of a sequence/run, even with "Run Sync" enabled.
Temporary Solution:
An external Contact Closure card, G1351A, can be used to Start/Stop the Active Splitter independently of the Active Splitter software.
Fix Information:
n/a
Keyword:
One-line Description:
Users can't log in with domain authentication - Message User must change Password
Problem:
When using Windows Domain Authentication, the message "User must change the password" is displayed even when this is not correct.
Temporary Solution:
When using Domain Authentication, the application needs to read user attributes like userAccountControl (https://msdn.microsoft.com/en-us/library/ms680832(v=vs.85).aspx) and pwdLastSet (https://msdn.microsoft.com/en-us/library/ms679430(v=vs.85).aspx) In order to get information whether or not a user must change password. When read permission is not granted to the user, the application will assume that the password of the user expired and will refuse the access Steps to enable the permissions to read attributes: 1 Open Active Directory Users and Computers 2 In the View menu, click Advanced Features 3 Under Users, select a user that is not able to log in to OLCP 4 Right click and go to properties 5 In the Security tab, click select 'Authenticated Users' 6 Check 'Read' checkbox below and click OK
Fix Information:
n/a
Keyword:
One-line Description:
Software Verification report of Agilent Data store Fail While upgrading Server of A.04.07 SR1 to A.04.07 SR2
Problem:
Running the Software Verification Tool (SVT) with a user other than the user who did the upgrade will result in a failed report.
Temporary Solution:
The user performing the upgrade should be the user that runs the SVT tool.
Fix Information:
n/a
Keyword:
One-line Description:
Copy/Paste of text fields or character failing from reporting output to another text editor
Problem:
Copying, and pasting text fields from the intelligent report printout onto notepad or any other text editor may not give the expected results. This report generation uses a Microsoft component which does not allow all character or text field to be reportable from the output to another text editor.
Temporary Solution:
N/A
Fix Information:
This is fixed in another product release: OpenLAB CDS v2.1 or higher. A fix in OpenLAB ChemStation Edition is for now not possible
Keyword:
One-line Description:
Report export to doc or txt file fails when using data fields in the report header and footer
Problem:
C.01.08 and higher allows users to add data fields in the report headers and footers. However this functionality will not be functional when the report output chosen is a DOC or TXT file extension. In this case an error message will appear in the report file.
Temporary Solution:
The temporary fix would be to modify the existing report template, and delete the data fields from the report header and footer.
Fix Information:
No fix is for now possible.
Keyword:
One-line Description:
Injection order lost for replicates when reprocessing with a different analysis method
Problem:
The report does not show the correct injection order number for replicates if the processing method is changed for a replicate and the sequence is reprocessed.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS ChemStation C.01.08.
Keyword:
One-line Description:
OLSS Upgrade error with MS SQL 2012
Problem:
Trying to upgrade an Shared Services using SQL server windows authentication with a user who is not the owner of the already installed OLSS DB will lead to an error. Microsoft does not support such an update with a different user.
Temporary Solution:
n/a
Fix Information:
For installs or upgrades using SQL server windows authentication, please make sure that the SYSTEM user is the owner of the OLSS Database.
Keyword:
One-line Description:
EIC generation fails in Intelligent Reporting report printout (intermittent)
Problem:
In some circumstances which will be described below, the EIC generation in the report printout may fail. In the cases this extraction are failing, a blank space will be displayed at the area where the EIC should appear. This can only occur when using one decimal digit (such as m/z = 721.3) in the "Extract Ion..." in the "File" menu, or by modifying the signal description in the signal details in the ChemStation Data Analysis view. However, not all EIC with one decimal digit would fail. This is only intermittent in the case that the ion chromatogram description required is not matching to what the report template algorithm would expect.
Temporary Solution:
There are several ways to temporary fix the misleading ion chromatogram extraction: 1. Move the one decimal value with a m/z shift of 0.1 (higher or lower value) 2. Change the m/z extraction range in "signal options" in the "graphics" menu. Default mass defect values are: [-0.3; +0.7]. This can be changed to another value to allow fine EIC extraction
Fix Information:
This will be fixed in C.01.08 as part as other new EIC extraction functionalities.
Keyword:
One-line Description:
Installing a lamp with tag may invalidate "alway use lamp" setting in detector's control
Problem:
Installing a lamp with tag may invalidate the detector's control setting "always use lamp" and silently modify it to "use lamp only if lamp tag is available".
Temporary Solution:
Check detector's Control setting after installing a new lamp.
Fix Information:
Fixed in A.02.04
Keyword:
One-line Description:
Flexible Cube G4227A does not work with HiP ALS G1367E
Problem:
On Chinese or Japanese CDS only the 1290 Infinity Flexible Cube G4227A does not work with High Performance Autosampler G1367E. On English CDS the operation of 1290 Infinity Flexible Cube G4227A with High Performance Autosampler G1367E is not affected.
Temporary Solution:
n/a
Fix Information:
Resolved in revision A.02.02
Keyword:
One-line Description:
Lamps and cells of G1314F show incorrect dates in the RFID tags
Problem:
Lamps and cells of G1314F show incorrect dates in the RFID tags
Temporary Solution:
n/a
Fix Information:
Fixed with the driver version A.02.06
Keyword:
One-line Description:
Cannot Check "Pump and solvent selection valve installed" Box for Flex Cube if Once Clear it
Problem:
LC Driver version: A.02.06 [363] (Test application/Install on Japanese ChemStation) 1. Open Instrument configuration dialog 2. Select Flexible Cube 3. Open Flexible Cube Configuration dialog 4. Clear the "Pump and Solvent selection valve installed" check box, then press OK 5. Re-open Flexible Cube Configuration dialog 6. Thick that check box and press OK 7. Open Flexible Cube Configuration dialog again -> check mark is gone
Temporary Solution:
Workaround: use autoconfig to set up the FlexCube
Fix Information:
Fix in the next rev RC.Net A.02.07.
Keyword:
One-line Description:
AFC & UIB II: AFC does not fractioning as expected
Problem:
System configuration: 1 DAD 1 UIB II 2 AFCs -> 1 AFC Cluster Analog out of DAD connected to Analog In of UIB II Simulated signal of DAD was used Method settings for modules (expect threshold for UIB II signal): see attached method 1. Created a data file with the simulated signal recorded by the UIB II 2. Loaded data file into Fraction Preview 3. Used Fraction Preview to collect peakbased fractions with threshold > 0.12 V 4. The value 0.12 V is written into the table on the left side without conversion (the table shows mV as unit) Using this method will cause that the AFC will fractioning all the time. Cause: The firmware interface expects the threshold in mV but the driver writes 0.12 (V) to the AFC This causes that the AFC is fractioning all the time.
Temporary Solution:
n/a
Fix Information:
Fixed in UIB II signal is delivered as mV signal similar to UIB I since LC Driver Release A.02.07.
Keyword:
One-line Description:
Use Vial as Fraction collections
Problem:
Started a run with fraction collection. The fraction collector config does not show 10 vials but only trays.
Temporary Solution:
If you have installed a 2 WP-Tray, vials are allways forbidden to use for fractionation, regardless if there are wellplates inserted or not.
Fix Information:
Issue fixed with LC and CE Driver revision A.02.07
Keyword:
One-line Description:
Recovery volume and fraction volume: behaviour for reset volume
Problem:
I worked with one AFC. Configured 2 wps +10 vials and 15 vial tray for recovery. Started multiple runs. For each run a new Recovery location was used. Then removed the recovery tray and entered it again. Next run started with first recovery location --> this was already used.
Temporary Solution:
If you want to reset the Funnel Trays only, use "Reset Funnel Tray Volume" If you want to reset Wellplates and vials (all except funnels), use "Reset Fraction Volume"
Fix Information:
Issue fixed with the LC and CE driver revision A.02.07
Keyword:
One-line Description:
Wrong datarate for Vwd Aux signals for 'Board Temperature' and 'Optical Unit Temperature'
Problem:
Wrong lookup values in VwdDeviceCapabilities.xml.
Temporary Solution:
n/a
Fix Information:
OpenLAB CDS ChemStation Edition - SP2 for LC RC.Net Driver A.02.05 and A.02.06
Keyword:
One-line Description:
Fraction collector stop times do not end runs
Problem:
For previous firmware revisions, fraction collectors could be stop time masters same as pumps and injectors. If a stop time is set for a stop time master, this will end the run after the time set. In case of multiple possible stop time masters, the shortest time will apply. Other modules can refer to the stop time master (option as pump/injector) or set their own stop times. For other modules, the stop time will not end the run but the module activity, e.g. if the detector has a shorter stop time than the pump, signal acquisition will end before the run. Since firmware A.06.54, fraction collectors can no longer be stop time masters, i.e. their stop times will no longer end a run.
Temporary Solution:
If a stop time is set for the fraction collector, this change may cause a changed method behavior. If a run stop time is needed, please set it for the pump or injector.
Fix Information:
n/a
Keyword:
One-line Description:
Method parameters are applied by module order, not by type, serial number or name
Problem:
If there are several modules of the same type (e.g. two pumps, detectors or column thermostats), method parameters are applied by their order in the configuration. If order changes e.g. because of newly configuring a system, parameters may be applied to the wrong modules.
Temporary Solution:
Ensure to keep the same order of modules in a system.
Fix Information:
Resolved in OpenLab ChemStation LTS 01.11.
Keyword:
One-line Description:
SFC remains in "Ready" (Idle) state while pump is in Error.
Problem:
Have set up a long sequence (600 injections). Pump run out of solvent in channel B und stopped with Error as it was defined in "Bottle fillings" UI. SFC remained in "Ready" (Idle) state. Should go to "Standby". System was in "SFC mode".
Temporary Solution:
Workaround could be to not use bottle fillings with SFC
Fix Information:
Solved with LC & CE Drivers A.02.08 SP1
Keyword:
One-line Description:
Turning off the SFC device G4301A while connected to an online OL ChemStation will lead to an application crash
Problem:
If you turn off your SFC device while the OL ChemStation is connected to this instrument this will crash this application.
Temporary Solution:
Close the OL ChemStation before switching off the SFC module.
Fix Information:
n/a
Keyword:
One-line Description:
Changing the backpressure method parameter of the SFC pump during a run does not work properly
Problem:
If you try to change the back pressure of the SFC pump during a run, this change will be applied correctly to the module, i.e. the instrument control changes as requested. However the method parameter will not be updated correctly and if the method is saved, the original setting will be used.
Temporary Solution:
Change the method parameter outside a run.
Fix Information:
Since driver version A.02.16 backpressure can no longer be changed during a run.
Keyword:
One-line Description:
Analytical SFC System (G4301A) delivers still CO2 even if the SFC binary pump has flow 0
Problem:
The 1260 Infinity Analytical SFC System (G4301A) still delivers CO2 even if the SFC binary pump has flow 0.
Temporary Solution:
n/a
Fix Information:
Upgrade to the most recent firmware version (minimum A.03.09) of the 1260 Infinity Analytical SFC System.
Keyword:
One-line Description:
Driver does not start the run / CS stays in Run in Progress / Data Acquisition
Problem:
The ChemStation has prepared the run as expected but the driver didn't start the run from CS / User perspective. It can be reproduced by opening the door of the sampler after the run preparation was done (need some tries to have the correct timing)
Temporary Solution:
Solved with LC & CE Drivers A.02.09
Fix Information:
n/a
Keyword:
One-line Description:
Fraction start location in a sequence parameter cannot accept the number "1" when using 2P+10Funnel tray at RC.net driver.
Problem:
The customers want to use 10 Funnels at the tray 2P+10Funnel without plates.
Temporary Solution:
Solved with LC & CE Drivers A.02.09
Fix Information:
N/A
Keyword:
One-line Description:
Analytical SFC System (G4301A) looses method settings when the module is powered off and on again while the application is connected
Problem:
The Analytical SFC System (G4301A) looses method settings when the module is powered off and on again while the application is connected. This may lead to a changed method depending on the used CDS.
Temporary Solution:
Do not turn off the Analytical SFC System (G4301A) while connected to a CDS.
Fix Information:
No fix available
Keyword:
One-line Description:
Data acquisition rate changes when importing old methods to new detectors
Problem:
The definition of data acquisition rates/peak widths changed between detector types. This is because the data rate is saved as an index number (0 = 20 Hz, 1 = 40 Hz, ...) rather than the actual data acquisition rate and the index definition was changed. DADs G1315A, B and D use the same indexing scheme. DADs G1315C, G4212A/B use a different indexing scheme. MWDs G1365A, B and D use the same indexing scheme. MWD G1365C uses a different indexing scheme. VWDs use several different indexing schemes.
Temporary Solution:
n/a
Fix Information:
Verify data acquisition rates/peak width settings when migrating to a different detector type.
Keyword:
One-line Description:
HDR Cluster does not allow using additional detectors
Problem:
Using several detectors creates a large amount of data, which is communicated at the same time. In a HDR cluster setup, spectral data is communicated via LAN but single wavelength data is communicated via the internal CAN bus. If detectors are used in addition to the HDR cluster, the very high data volume may impair data integrity, e.g missing data points.
Temporary Solution:
n/a
Fix Information:
Do not connect additional detectors to HDR clusters via CAN.
Keyword:
One-line Description:
Incorrect lamp type settings for VWDs
Problem:
The G1314A/B/C VWDs allow the use of either VWD or DAD lamps, which can be configured in the module driver. In driver revisions A.02.09 and below the selection of DAD or VWD lamp is reversed. The different lamps use a different lamp ignition process. The DAD filament heating is turned off after ignition while the VWD lamp filament heating is kept on a lower level. RC.NET LC Driver (<= A.02.09) CLASSIC driver G1314A/B/C REVERSED OK G1314D/E/F OK OK
Temporary Solution:
In case of noise and ignition issues please verify the setting. In ChemStation or LabAdvisor, you may use the command line command to check the actual setting in the VWD. ChemStation: print sendmodule$(lvwd,"ltyp?") LabAdvisor: send ltyp? to VWD module Following instrument answers are possible: 0 = DAD lamp 1 = VWD lamp
Fix Information:
This issue has been fixed in LC Driver version A.02.10. If the correct setting has been used before (i.e. VWD lamp present and VWD setting has been chosen), the behavior will now be correct and has previously been wrong. If a wrong setting has been used before (i.e. VWD lamp is present and DAD setting has been chosen) as a workaround, the behavior will then be wrong for A.02.10 and following and needs to be changed.
Keyword:
One-line Description:
Tube heights and volumes are overwritten for inserted tube trays for G1364A/B/C/D, G5664A fraction collectors and their clusters
Problem:
Module firmware and driver store settings for tube heights and volumes for each tube tray type. If tray of any type is replaced by a tube tray, settings for this tube tray may be overwritten by a previous tray setting. These new and potentially wrong settings are stored in firmware and driver.
Temporary Solution:
After inserting a new tube tray, please verify settings for height and volume in the fraction collector configuration and correct them if needed.
Fix Information:
This has been fixed with driver release A.02.11 SP1. Additionally, height and volume settings need to be verified for each fraction collector and tube tray type.
Keyword:
One-line Description:
HDR stays offline if one DAD of the HDR cluster has been switched off and on again
Problem:
HDR stays offline if one DAD of the HDR cluster has been switched off and on again
Temporary Solution:
Restart the CDS after switching on both HDR modules
Fix Information:
n/a
Keyword:
One-line Description:
Status display for wavelengths and absorption values are not updated in dual wavelength mode for the VWD G7114B.
Problem:
For the dual wavelength mode of the VWD G7114B, both set wavelengths and measured absorptions are not updated in the extended status UI after switching between single and dual wavelength mode.
Temporary Solution:
not available
Fix Information:
Fixed with driver release A.02.11 SP1.
Keyword:
One-line Description:
1290 Infinity II VWD G7114B shows only "Signal" instead of "Signal A"
Problem:
When using the dual-wavelength mode for the 1290 Infinity II VWD G7114B, the user interface for instrument curves shows signal A as "Signal without reference" and "Reference only" for the reference wavelength of signal A. Signal B is shown correctly as "Signal B without reference" respectively "Reference B only" for the reference wavelength signal.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Method screens may not be displayed correctly if the Windows text size is increased
Problem:
Windows allows adjusting the text size. If this size is increased beyond 100%, the display may distorted and become partially invisible or inaccessible.
Temporary Solution:
Please use the default Windows text size.
Fix Information:
n/a
Keyword:
One-line Description:
Resetting tray volumes does not reset the fraction start location
Problem:
Resetting tray volumes does not automatically reset the fraction start location unless a dedicated start location is set. The fraction collector will continue from its current location.
Temporary Solution:
Set a fraction start location.
Fix Information:
Will be fixed in G1364A/B/C firmware A.07.01
Keyword:
One-line Description:
Changed stoptime behavior for G1364x devices
Problem:
Due to different FW implementation (introduced with FW A.06.54 or later) of the stop time behavior, it can happen that, in case of defining the stop time ONLY in the fraction collector method, a run will not stop after the specified time.
Temporary Solution:
Define the stoptime in addition in the method of another device e.g. the pump or the sampler.
Fix Information:
n/a
Keyword:
One-line Description:
ChemStation runs may not finish for SFC systems which are not operated in SFC mode
Problem:
In SFC systems, the SFC mode may be disabled for running a pump in normal HPLC mode (SFC off in instrument control). Such runs may not finish correctly
Temporary Solution:
Abort the run if the run has finished (by time).
Fix Information:
n/a
Keyword:
One-line Description:
Pump pressure is limited by lowest pressure of a valve in a valve thermostat cluster
Problem:
If a valve thermostat cluster is configured using one or multiple valves, the valve with the lowest possible maximum pressure determines the maximum pressure of the entire cluster, e.g. a 2-pos/6-port valve with 1300 bar is used for switching between to multicolumn thermostats with column selector valves of 600 bar and 800 bar. Then 600 bar would be the maximum pressure for that cluster and would limit the pump pressure accordingly. In case of two 8-pos/9-port valves using topology 5 (V-C-V), one valve is before the column in the high pressure flow path and the other one after the column in the low pressure flow path. In that case, the valve in the low pressure flow path should not limit the pump pressure but it does. In all other topologies, all valves are as well in the high pressure as low pressure flow path.
Temporary Solution:
n/a
Fix Information:
This has been fixed with driver version A.02.16.
Keyword:
One-line Description:
Changing multiple trigger settings for Infinity II Fraction Collectors at the same time did not work properly
Problem:
When loading a method from the instrument, trigger settings with the same time are discarded.
Temporary Solution:
Avoid duplicate times for trigger settings in the method.
Fix Information:
This has been fixed with driver version A.02.16.
Keyword:
One-line Description:
Unexpected run may start if the Analytical SFC System (G4301A) is switched on
Problem:
Unexpected run may start if the 1260 Infinity Analytical SFC System (G4301A) is switched on and the LC modules of the system connected via APG remote cable may also start the run.
Temporary Solution:
Ensure that the Analytical SFC System is the first module that is switched on in the entire system.
Fix Information:
Not available
Keyword:
One-line Description:
The delay calibration does not work for the G1364A with preparative setup in ChemStation CDS
Problem:
The delay calibration does not work for the G1364A Automatic Fraction Collector with preparative setup in ChemStation CDS. The proposed steps for delay calibration are not provided. This results in a failure of the calibration process.
Temporary Solution:
Agilent Lab Advisor Software can be used for a delay calibration of the G1364A with preparative setup.
Fix Information:
This issue has been fixed in LC driver version A.02.17.
Keyword:
One-line Description:
Pressure Limit Cluster is set to auto for any valve that is not used as switching valve in the VTC
Problem:
For any valve, which is not used as switching valve in the VTC, the pressure limit cluster is set to "auto", which is also displayed in the Control UI. This means, that the valves, which are not used in the VTC determine nevertheless the pressure limits for all pumps because of the "auto" setting.
Temporary Solution:
Make sure to configure all connected valves either in a Cluster or as stand alone device.
Fix Information:
n/a
Keyword:
One-line Description:
Tubing Kit cannot be modified from Dashboard for new Fraction Collectors
Problem:
The LC dashboard provides a Modify-dialog called "Needle and Tubing". It is not possible to modify the tubing kit of the G1364E/F or G5664B Fraction Collectors using this function.
Temporary Solution:
Modify the Tubing Kit with the ChemStation command line instead. The commands are as follows: For a single fraction collector: print sendmodule$(AFC,"AFC:CAPILLARYKIT 'G1364-68723' "), replace G1364-68723 with the desired Tubing Kit part number. For G1364E/F and G5664B clustered: print sendmodule$(AFC,"G1364E:AR42351409@AFC:CAPILLARYKIT 'G1364-68723' "), replace G1364E:AR42351409 with <Type>:<SerialNumber> of the Fraction Collector in a cluster for which the tubing kit should be set and G1364-68723 with the desired Tubing Kit part number.
Fix Information:
The behaviour will be changed with LC driver version A.02.18 so that the Tubing Kit can be set in the LC dashboard "Modify... Needle and Tubing" dialog.
Keyword:
One-line Description:
Overfill Behavior unclear
Problem:
The Fraction Collectors (G7159B, G1364E/F, G5664B and G7166A) provide a setting for "Overfill Behavior" with which it is possible to specify the action to be taken if the fill volume of a location reaches the max. fill volume per location. This can lead to samples / fractions being switched to waste. In case of using pooling the behavior is unclear to the user.
Temporary Solution:
N/A
Fix Information:
With driver version A.02.19 the parameter "Overfill Behavior" is removed, in order to prevent a loss of sample / fractions and to make the behavior more clear. The LC drivers now imply the following behavior: When Pooling: in case the fill volume of a location reaches the maximum fill volume, the run or sequence is stopped with an error. When not Pooling: in case the fill volume of a location reaches the maximum fill volume, the collection switches to the next available location.
Keyword:
One-line Description:
Fraction Collector Cluster: Vial 1-10 and right tray can now be used for fraction collection - consider behavior when working with "Next Location"
Problem:
With LC and CE driver A.02.18 and a Fraction Collection Cluster with 2 well plates + 10 Vials (or 2 well plates + 10 collecting funnels) configuration, the positions Vial 1-10 and the ones on the right tray can now also be used for fraction collection (in addition to P1 and P2 on the left tray). Please consider that when working with "Next Location" as Fraction Start Location, locations may be used for fractioning, that were ignored with previous driver versions (Vial 1-10 and the right tray). In order to prevent that these locations are used for fractioning, specify Vial 1-10 and Vial 201-240 in the Collection Settings as "Forbidden Location" for all fraction collectors in the cluster. The modules affected are: G1364A/B/C and G5664A.
Temporary Solution:
N/A
Fix Information:
In order to prevent that locations that were previously ignored are now used for fractioning, specify Vial 1-10 and Vial 201-240 in the Collection Settings as "Forbidden Location" for all fraction collectors in the cluster.
Keyword:
One-line Description:
Start location for fraction collection is reset when removing a tray and closing the fraction collector front door
Problem:
Start location for fraction collection is reset when removing a tray and closing the fraction collector front door. This behavior may not be wanted, if the same tray is inserted afterwards.
Temporary Solution:
If the same tray shall be inserted again, do not close the front door in order to retain the fraction collection start location.
Fix Information:
n/a
Keyword:
One-line Description:
Fraction Collector Method Timetable must be empty in order to perform a delay volume calibration
Problem:
In order to perform a delay volume calibration with the Fraction Collectors G1364A/B/C it must be made sure that the timetable within the Fraction Collector Method is empty. If this is not the case, the Fraction Collector will execute the timetable and collect fractions. Affected modules are the Agilent Fraction Collectors G1364A/B/C. Newer modules are not affected.
Temporary Solution:
n/a
Fix Information:
Make sure that the timetable in the Fraction Collector method is empty in order to perform a delay volume calibration.
Keyword:
One-line Description:
Upgrade from v2.1 fails when the application is installed on a folder name with short name
Problem:
Upgrade from v2.1 fails when the application is installed on a folder name with 2 or less characters
Temporary Solution:
Do not install the application on folder where the name is less than 3 characters
Fix Information:
n/a
Keyword:
One-line Description:
Integrator timed event: Split peak can cause integrate to show a misleading error message box
Problem:
When the Split Peak timed event happens to be at a very front edge of a peak start, it can cause an error message on the message line or logbook or show an error message box when the edit integration user interface is open. The error message is: "Could not integrate signal. Probably too many peaks found! Please adjust integration events for less sensitivity ". There are critical times when the Split Peak timed event can cause the integrator to return a misleading error message.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Snapshot file will be created but not displayed if long data paths are used
Problem:
Snapshot file will be created but not displayed if long data paths are used. The following error message is displayed: Display File not set
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab ChemStation C.01.09 Update 1 and C.01.10.
Keyword:
One-line Description:
MS Spectra not always reported in Intelligent Reporting when acquiring with high mass range
Problem:
When acquiring with high mass range (for example m/z: 150 - 3000), some peaks may not have reported spectra in the report printout. Because of the number of mass peaks to scale in the spectra plot is so high, the report generation may fail for few peaks.
Temporary Solution:
Modification of the report template would be necessary. As a temporary solution, increasing the spectra plot size in the report template will resolve the issue for non-reported peaks.
Fix Information:
This issue will be fixed in OpenLab CDS ChemStation Editon C.01.10.
Keyword:
One-line Description:
Purification - When the same recovery start location is used multiple times in a sequence, the instruments uses "Next Location" instead
Problem:
When a sequence is set up with the same absolute Recovery Start Location value in multiple sequence lines, then the instrument ignores these absolute settings and instead uses the next empty recovery location for each run. This is done to protect existing recovery locations from used multiple times accidentally.
Temporary Solution:
None, works as designed currently.
Fix Information:
None. Works as designed currently. A logbook entry has been created to show that a specified Recovery Start Location has been skipped by the instrument. An additional behavioral change to enhance usability will be implemented in a future revision.
Keyword:
One-line Description:
Purification - G1364A-C - Using the Fraction or Recovery Start Location "Next Plate" in the first line of a sequence leads a sequence abort before injection
Problem:
When a sequence is created that has the Fraction Start Location or Recovery Start location value "Next Plate" and all fill volumes of the Purification instrument have been reset right before executing that sequence line, then the sequence execution will abort. No loss of sample is to be expected.
Temporary Solution:
Do not use the Fraction Start or Recovery Start Location value "Next Plate" in a sequence line.
Fix Information:
The issue has been fixed.
Keyword:
One-line Description:
Purification - Specific Recovery Start Locations are currently ignored by the instrument when using OpenLab CDS Chemstation Ed. C.01.10 Update 1
Problem:
When using LC Drivers within Chemstation C.01.10 Update 1, specific Recovery Start Locations are ignored by the instrument. The driver instead defaults to "Next Location" and proceeds to the next free position.
Temporary Solution:
Either do not upgrade to OpenLab CDS ChemStation Ed. C.01.10 Update 1 or do not use specific Recovery Start Locations for the time being when upgrading to LC Drivers 3.1 at the same time. If your workflows do require OpenLab CDS ChemStation Ed. C.01.10 Update 1, then a workaround is to use Update 1 with LC Drivers 3.0 instead.
Fix Information:
None currently. Will be fixed in a future release of OpenLab CDS ChemStation Edition.
Keyword:
One-line Description:
Missing Signal Descriptions for LCMS Signals on Intelligent Reports
Problem:
In the chromatogram displayed in Data Analysis for a MS signal, the signal description is displayed as: "MSD1 TIC, MS File, Pos Scan, MM-APCI, Frag: 70, "Scan" ". The same signal displayed on an Intelligent Report will only display as: "MSD1 TIC, MS File".
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Purification in OpenLab CDS 2.5 - Error when using legacy Fraction Collector Cluster with added Recovery collector
Problem:
When using OpenLab CDS 2.5 together with the legacy Fraction Collector Cluster (supporting G1364A/B/C and G5664A), an error can occur during sequence setup. The error is caused by using the graphical "Sample Trays" UI when one of the collectors in the cluster is used as Recovery collector.
Temporary Solution:
As a temporary workaround, users are asked to not use the graphical "Sample Trays" UI when setting up sequences for the Legacy Fraction Collector Cluster.
Fix Information:
There is no fix for this issue at the moment.
Keyword:
One-line Description:
Missing Signal Descriptions for LCMS Signals on Intelligent Reports
Problem:
In the chromatogram displayed in Data Analysis for a MS signal, the signal description is displayed as: "MSD1 TIC, MS File, Pos Scan, MM-APCI, Frag: 70, "Scan" ". The same signal displayed on an Intelligent Report will only display as: "MSD1 TIC, MS File".
Temporary Solution:
n/a
Fix Information:
n/a
Keyword:
One-line Description:
Peak based fraction collection does not work
Problem:
If a pump time table starts with 0 flow and peak based fraction collection starts with RT=0 no fractions are collected.
Temporary Solution:
Either start the peak based faction collection with RT>0 or start the pump time table with a flow > 0
Fix Information:
n/a
Keyword:
One-line Description:
Nozzle Volt is not active warning message
Problem:
Nozzle Volt is not active warning message on G6120B and G6130 Single Quad.
Temporary Solution:
n/a
Fix Information:
ChemStation C.01.10 Update 4
Keyword:
One-line Description:
Instrument configuration change is not registered in Activity Log
Problem:
If the instrument configuration is changed from the Online session (from Method and Run Control --> Instrument --> Instrument Configuration --> Autoconfiguration ), nothing is registered in the Activity Log.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab ChemStation LTS 01.11.
Keyword:
One-line Description:
Import XML worklist does not include target mass
Problem:
Import XML worklist does not include target mass
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: 1100 Control
One-line Description:
Signals from different LC modules show a shift in the Online Plot
Problem:
LC modules running on a firmware A.xx.xx will interrupt the data transfer between runs. LC modules running with a firmware B.xx.xx will send data without interruption between runs. When traces from modules of different firmware generations are overlaid in the onliine plot, there will be a visible shift of signals, because they are synchronized after every injection.
Temporary Solution:
N/A
Fix Information:
No fix planned
Keyword: 1100 Control , 1200 Control
One-line Description:
Online spectra not available for some instruments
Problem:
In some cases the online spectra windows do not display the spectra for some detectors. In systems with more than one or two DADs spectra have been missing for one detector and not for the other. The reason was that the plot has only been designed for ten modules. Any detector with an internal number >9 will not be able to deliver the spectra.
Temporary Solution:
Change the order of modules in the Instrument Configuration. Use the Up arrows to move the detectors to the top of the list.
Fix Information:
Resolved in OpenLAB CDS ChemStation C.01.08.
Keyword: 1100 Control,1200 Control
One-line Description:
Online spectra not available for some instruments
Problem:
In some cases the online spectra windows do not display the spectra for some detectors. In systems with more than one or two DADs spectra have been missing for one detector and not for the other. The reason was that the plot has only been designed for ten modules. Any detector with an internal number >9 will not be able to deliver the spectra.
Temporary Solution:
Change the order of modules in the Instrument Configuration. Use the Up arrows to move the detectors to the top of the list.
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.09
Keyword: 1200 Control
One-line Description:
Missing error handling for errors during fraction collection
Problem:
Logical errors are not handled well, when locations in fraction collection cannot be accessed. An example would be a next location command, when the arm is already at the last position. Various symptoms can occur, like macro errors (Error in Hook RepPreChrom ) or duplicate injection times.
Temporary Solution:
Plan ahead to avoid these situations. They will cause loss of sample even with better software handling.
Fix Information:
Fixed in OpenLAB CDS C.01.06.
Keyword: 1200 Control
One-line Description:
Manual Fraction Collection: Delay time not applied
Problem:
The delay time of a signal is not applied to the fraction table. The fraction annotation in the chromatogram is correct.
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB CDS ChemStation C.01.06.
Keyword: 1200 Control
One-line Description:
Multidraw injection volume >1000ml cannot be entered from the context menu.
Problem:
The module context for a sampler does not allow entering an injection volume higher than 999.99 ml. Use the topic: Set up Instrument Method... from the Instrument menu to enter volumes up to the allowed maximum.
Temporary Solution:
N/A
Fix Information:
N/A
Keyword: 1200 Control
One-line Description:
Stopping a run is sometimes executed as Abort
Problem:
Systems with RC.Net drivers sometimes interpret a Stop command as an Abort command. The symptom is a system with strange status combination, like all modules IDLE (yellow), but the overall status is still "RUN" (blue).
Temporary Solution:
Restart the online session and the instrument hardware. Fixed with Hotfix08 for C.01.05.
Fix Information:
Fixed in C.01.06.
Keyword: 1200 Control
One-line Description:
LC Column information is not transferred to the data file
Problem:
The column information from tagged LC columns is sometimes not written to the data file. Without column information no capacity factors or plates per meter can be calculated in reports.
Temporary Solution:
Enter the column information manually each time a column changes. Capacity factors or plates per meter can be calculated with the interactive Peak Performance tool, if the column dimensions are known.
Fix Information:
Fixed in C.01.06
Keyword: 1200 Control
One-line Description:
In ChemStation "CP Macro" errors are added as logbook entry for Fraction Collectors, when a run has been completed
Problem:
When a run has been completed, "CP Macro" error logs will be added to the logbook in ChemStation C.01.07 SR3, in case a Fraction Collector is used. The error logs do not indicate an actual problem. These error logs can safely be ignored by the user. Besides these logbook entries, no other side effects are observed. The error logs can safely be ignored by the user.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS ChemStation C.01.08.
Keyword: 1200 Control
One-line Description:
The MSD goes not ready (NRDY) and the sequence and queued sequences are aborted
Problem:
In an 1200 LC-MS system, the MSD goes not ready (NRDY) and aborts the sequence and any queued sequences.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.10 Update 1.
Keyword: 1200 Control
One-line Description:
After an Upgrade to C.01.10, no Fraction Tickmarks shown in the Online Plot
Problem:
For better performance and memory usage of ChemStation, the ChemMain.exe was split into two processes, ChemMain.exe and a ChemMainAcq.exe in C.01.10. After an upgrade to C.01.10, the ChemMain.exe process split is not active, due to a missing ChemStation.ini key. The ChemStation is running in a single process instead of two. Due to the missing key, the online instrument does not start the ChemMainAcq.exe process. Possible symptoms: - No fraction tickmarks shown in on-line plot of LC prep - Large LC configurations with 10 or more modules may terminate after several days of usage The problem does not occur on a new install of ChemStation C.01.10.
Temporary Solution:
Adding the new line "AcqRemoteServer=1" at the end the PCS section of the Chemstation.ini file will enable the second ChemStation process. [PCS] UIFont1+... … AcqRemoteServer=1
Fix Information:
Fixed in C.01.10 Update 2
Keyword: 1200 Control , Method Scouting Wizard
One-line Description:
Cluster information can be overwritten from the offline session
Problem:
Changes in a TCC or pump cluster configuration (like e.g. changing a solvent name) usually cause method resolution. The changes will also be updated, even when the configuration dialog is closed by pressing Cancel. This is true for online and offline. This will consequently modify methods, although nothing has actually been changed.
Temporary Solution:
None
Fix Information:
Fixed in revision B.04.03 SP1.
Keyword: 35900 Control
One-line Description:
35900E - Error The Current Event Name is the same for a special event
Problem:
B.04.03 GC ChemStation with a 35900E configured. Select the CHANGE, DEFINE EVENTS tab in device configuration and you get an error "The Current Event Name is the same for a special event. Please change the current event name." This error is reported on a clean install of B.04.03 [16]. ChemStation B.03.02 works correctly.
Temporary Solution:
n/a
Fix Information:
Fixed in revision B.04.03 SP1.
Keyword: 35900 Control
One-line Description:
Cannot start two instruments using the two channels of one ADC at the same time
Problem:
When two instruments are configured that use channel A and channel B of the same ADC converter, respectively, the two instruments cannot be running at the same time. When the second instrument is started, both ChemStation instruments will freeze and can no longer be operated.
Temporary Solution:
The two instruments will work when using only one of the instruments at a time.
Fix Information:
Resolved in ChemStation Edition C.01.07 SR1 Hotfix01 [113]. Use the 35900E RC.NET driver if you need to split the channels between 2 instruments on an AIC distributed system.
Keyword: 35900 Control
One-line Description:
35900E A/D converter cannot be split into two instruments from a ChemStation client
Problem:
The two channels of a 35900 ADC cannot be part of two different instruments when controlled from a ChemStation client.
Temporary Solution:
When the 35900 is connected to an AIC and its two channels need to be part of two different instruments, they need to be setup from a networked workstation. The two instruments will work when using only one of the instruments at a time.
Fix Information:
Resolved in ChemStation Edition C.01.07 SR1 Hotfix01 [113]. The latest 35900E RC.NET driver must be used to split the channels between 2 instruments on an AIC distributed system. The latest 35900E firmware must be used.
Keyword: 35900 Control
One-line Description:
View ACQ method from Data Analysis does not list acquisition parameters for the 35900 AD converter.
Problem:
The acquisition parameters of an AD converter are not listed in the method printout from Data Analysis view.
Temporary Solution:
Use the Print Method... option from Method&Run Control view.
Fix Information:
Resolved in ChemStation Edition C.01.07 SR1 Hotfix01 [113] and with the Agilent 35900E RC.Net Add-On driver.
Keyword: 35900 Control
One-line Description:
Activity log for 35900E will truncate last leter in errors or status messages
Problem:
When you view the Activity log from OpenLAB Control Panel for a given instrument with a 35900 attached, some of the status messages and error messages will truncate the last letter of the message. For example, Power Failure will be displayed as Power Failur, or Power On will be displayed as Power O. The actual run.log and other logbooks will display the complete error message correctly.
Temporary Solution:
n/a
Fix Information:
Resolved in ChemStation Edition C.01.07 SR1 Hotfix01 [113] and with the Agilent 35900E RC.Net Add-On driver.
Keyword: 35900 Control
One-line Description:
Sequence stops when pushing the 35900 stop button during sequence
Problem:
With the RC.net driver, pushing the 35900 stop button will stop the current run and generate a report but does not continue to the next line in the sequence.
Temporary Solution:
n/a
Fix Information:
There is no fix planned.
Keyword: 6890 Control
One-line Description:
6890 RC.NET driver: instrument diagram not updated to show configuration
Problem:
In the 6890 RC.NET driver, changed the GC column configuration to have the back inlet go to the back column to the back detector. This change was not reflected in the instrument diagram as it is still showing the back inlet going to the front column to the back detector.
Temporary Solution:
After the configuration change is made to the column connections, the workaround is to shut down the ChemStation software and restart to reflect the new configuration changes. Beginning with B.04.03 DSP1 [54], after updating the column configuration, save and reload the method will update the instrument diagram. Classic 68xx driver works correctly.
Fix Information:
No fix Planned.
Keyword: 6890 Control
One-line Description:
Addition of new columns to the column catalog (database) is not possible after upgrade
Problem:
Upgrading from B.01.03 to B.04.02, it is no longer possible to add new columns to the GC column catalog (database) in Rev. B.04.02. The following error is shown: The Catalog File cannot be opened. If needed, you may obtain an appropriate replacement file from Agilent. C:\Chem32\CORE\...\DRIVERS\catalog.mdb
Temporary Solution:
The problem is that there are several columns that have the same manufacturer and model number. Our new database software doesn’t handle this. The newer database software (in B.04.02 for example) no longer allows duplicates of manufacturer + model number. i.e., there can only be one HP HP-5 in the database. Note: the new software will inform the user that the model number “xxx” is already in use and request that a different value be chosen. (The user is returned to the original entry screen with all the info filled in.) In theory, customers should not be creating more than one catalog entry with the same model number. It’s like having two items in the LLBean catalog with the same order number! But, of course, the older software allowed the duplicates.
Fix Information:
Before upgrading, fix the problem with the column database duplicates while running B.01.03 for example. Review the column catalogs and make sure there are no duplicates and fix any that exist before the upgrade.
Keyword: 6890 Control
One-line Description:
Data file of sequence with manual inject of 6890 may be incomplete
Problem:
Data file created by sequence with manual inject , with 6890 only, may be incomplete and not useable.
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB CDS C.01.06.
Keyword: 6890 Control
One-line Description:
Only one 6890 instrument with GPIB connenction can be configured on one controller
Problem:
It is not possible to configure more than one 6890 with GPIB in OpenLAB CDS.
Temporary Solution:
n/a
Fix Information:
No fix planned. 6890s controlled via the GPIB interface must be updated to LAN. The LAN kit to order is G2335A. Re-configure the 6890 to use the RC.NET driver beginning with OpenLAB CDS ChemStation C.01.06. Reference service note: OpenLAB CDS ChemStation Edition-047
Keyword: 6890 Control
One-line Description:
Migrating a classic 6890 instrument to RC.Net does not work smoothly after upgrade
Problem:
The C.01.07 ChemStation no longer supports the 6890 GC with classic drivers. Customers upgrading from earlier revisions of OpenLAB CDS ChemStation may want to switch their classic instruments to the new drivers. Removing the Tickbox "Use classic drivers" does not offer a configurable instrument.
Temporary Solution:
Toggle the check box in front of Enable Intelligent Reporting, meaning: uncheck, then check it again. Uncheck Use classic drivers. Leave the Configuration dialog by clicking OK. Call the Configuration dialog again. The 6890 instrument will now be available for configuration.
Fix Information:
N/A
Keyword: 68xx Driver
One-line Description:
Vial Missing Error Message
Problem:
When the 6890 GC encounters a missing vial, the IALog displays Stop key pressed error #1 instead of "Missing Vial."
Temporary Solution:
None
Fix Information:
This is fixed in 68XX Driver version 6.21 which is included in C.01.05 OpenLAB ChemStation and A.04.05 OpenLAB EZChrom.
Keyword: 68xx Driver
One-line Description:
How to Change Column Dimensions
Problem:
Changing Column Dimensions requires being logged in as an administrator.
Temporary Solution:
Column dimensions are part of the GC Configuration and the end user must be logged on as an Administrator to change these parameters.
Fix Information:
This is fixed in GC Driver version 6.01/6.02 which is included in OpenLAB ChemStation C.01.01 and OpenLAB EZChrom A.04.01 and later.
Keyword: 68xx Driver
One-line Description:
If One Signal Needs to be Saved, it Must be Assigned as Signal 1
Problem:
If only one signal needs to be saved, you must assign it as signal 1.
Temporary Solution:
If only one signal needs to be saved, you must assign it as signal 1.
Fix Information:
Fixed in 68xx driver revision 6.11[008]. This driver is included in B.04.03 SP1 ChemStation and C.01.03 OpenLAB CDS.
Keyword: 68xx Driver
One-line Description:
Instrument Diagram Does Not Match Installed Column in Method
Problem:
The Instrument diagram of the GC instrumentation indicates a different column configuration than is set in the method. The GC keyboard will reflect the new changes, only the software instrument diagram is affected.
Temporary Solution:
Beginning with B.04.03 DSP1 [54], after the configuration change is made to the columns, the workaround is to save the method and then load it. This will update the displayed column configuration. Another workaround is to shut down the ChemStation software and restart to reflect the new configuration changes. These workarounds may not work for all configurations. Classic 68xx driver works correctly.
Fix Information:
None
Keyword: 68xx Driver
One-line Description:
In 68XX driver revision 5.02 - 6.11, it is not possible to Modify a Method During a Run to Extend the Run Time or Change the Oven Temp
Problem:
It is possible to the software edit parameters screen and change the oven hold time or oven temp and select OK. The GC does not accept the updates. The method in the software is changed to the new values; the GC ignores all.
Temporary Solution:
n/a
Fix Information:
68XX Driver 6.11 is included in B.04.03 SP1, OpenLAB ChemStation C.01.04 and OpenLAB EZChrom A.04.04.
Keyword: 68xx Driver
One-line Description:
Info in Edit instrument parameters\Columns is incomplete
Problem:
Edit instrument parameters; columns Applies to GC’s: 7890A, 6890N, 6850A The window displayed is incomplete; it shows the column type and flow/pressure parameters, but does not give the length, ID or film thickness. These are critical parameters, and should be displayed just as they were in the previous ChemStation revisions. To find these parameters, the user now has to look on the configuration window.
Temporary Solution:
n/a
Fix Information:
Fixed in 68xx driver revision 6.11[008]. This driver is included with OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.
Keyword: 68xx Driver
One-line Description:
Ignore missing vial option is not present in GC
Problem:
Run a sequence with 11 runs and the 11th vial is missing. The sequence aborts and SSIZip file was not uploaded to ECM as the 11th vial was missing. The ignore missing vial option is not present in GC.
Temporary Solution:
n/a
Fix Information:
This is fixed in GC Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.
Keyword: 68xx Driver
One-line Description:
Instrument Setup graph has incorrect column legend
Problem:
In the 6890, the legend for the Instrument Setup graph says Column 2, irrespective of which column is highlighted. The graph is correct, it is just the legend that is wrong.
Temporary Solution:
n/a
Fix Information:
This has been fixed in 68XX Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.
Keyword: 68xx Driver
One-line Description:
ALS errors abort sequence; no Retry/Abort dialog box option offered with 68xx RC.NET Driver
Problem:
6890A RC.NET driver with ALS. If an ALS error occurs such as a Vial not found event, there is no dialog box option presented to Retry/Abort. The sequence simply aborts.
Temporary Solution:
n/a
Fix Information:
This is fixed in GC Driver version 6.21, which is available with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.
Keyword: 68xx Driver
One-line Description:
Opening 68xx RC.NET GC types result in 'FetchCPVar cannot talk to application'
Problem:
Configure 6850 RC.NET or 6890N RC.NET GC. Open an online session, initial or any subsequent try. This results in a benign pop-up error: 'FetchCPVar cannot talk to application'.
Temporary Solution:
Go to view menu and turn off the sampling and instrument diagrams. This will keep hpgcui00.exe from starting up.
Fix Information:
Fixed in driver revision 6.02. Driver included in OpenLab C.01.01.
Keyword: 68xx Driver
One-line Description:
In 68XX Driver Version 5.02, it is not Possible to Start the 6890 RC.NET from the Front Panel Start Key
Problem:
In 68XX Driver Version 5.02, it is not possible to start the 6890 RC.NET from the front panel Start key. Software configuration is Injection source = manual. 6890 is READY, press start and get error message "ERROR Start Key disabled".
Temporary Solution:
The workaround is for the user to start the run from the "run control" then "run Method" in the Chemstation software and then press start at the GC.
Fix Information:
Fixed in driver revision 6.10. This driver is included with B.04.03 DSP1 ChemStation. It is also included with OpenLAB ChemStation C.01.02 and OpenLAB EZChrom A.04.02.
Keyword: 68xx Driver
One-line Description:
CIS4 inlet is not recognized with 6890 GC RC.NET
Problem:
CIS4 inlet is not recognized with 6890 GC RC.NET
Temporary Solution:
n/a
Fix Information:
This is fixed in GC Driver version 6.21, which is available with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.
Keyword: 68xx Driver
One-line Description:
Only Fast Plunger Speed is available for PTV Solvent Vent multiple injections
Problem:
The fast plunger speed is only available when using multiple PTV Solvent Vent injections. Most users of multiple PTV Solvent Vent injections need to use Slow or Variable plunger speed, not Fast.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 6.10. This driver is included with B.04.03 DSP1 ChemStation and OpenLAB C.01.02 ChemStation and OpenLAB A.04.02 EZChrom.
Keyword: 68xx Driver
One-line Description:
IQ1.CFG file of 6890/6850 instrument corrupted after regional setting change
Problem:
From Control panel\Regional settings: change the decimal separator from a period to a comma. It is not possible to open a 6890/6850 ChemStation.
Temporary Solution:
1. Backup data, methods, sequences, column databases (found in C:\Chem32\DRIVERS), etc Record GC IP addresses 2. Delete IQ1.CFG located under the Chem32\1 directory for example 3. Uninstall Chemstation 4. Reinstall Chemstation
Fix Information:
Period is required in Regional Settings.
Keyword: 68xx Driver
One-line Description:
Incorrect injection volume shown in method report if PTV solvent vent is used
Problem:
6850 with a PTV inlet The method specifies PTV solvent vent mode with 3 injections, ten seconds apart. The method volume is 5 µL. Specify the method to use method volume for three injections. Each injection is 5 µL. However the report shows the injection volume as 1 µL. This is incorrect. The report should show the injection volume as either 5 µL or 5 µL x 3, or 15 µL.
Temporary Solution:
n/a
Fix Information:
Fixed in software driver 6.10 [023]. This fix is included in B.04.03 DSP1 and later versions of ChemStation. It is included in OpenLAB ChemStation C.01.02 and OpenLAB EZChrom A.04.02 and later.
Keyword: 68xx Driver
One-line Description:
Instrument Diagram Does Not Match Column 2's Inlet Configuration
Problem:
Column 2 inlet configuration will never be correct if column 2 is connected to anything other than inlet 2.
Temporary Solution:
n/a
Fix Information:
Fixed in 68xx driver revision 6.10[044]. This fix is included in these minimum versions:B.04.03 SP1 ChemStation, OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.
Keyword: 68xx Driver
One-line Description:
68xx Method Download Times Out
Problem:
Connect to 6890N GC. Within seconds of successfully opening an online session, logbook logs the following simultaneously with the loading of the method. 'Logbook Welcome to ChemStation 12:17:12 06/28/10 0 1da6 4c28cb1f 0 6890 1 Timed out waiting for method download to fin> 12:17:35 06/28/10 0 1da6 4c28cb1f 0 6890 1 ish. 12:17:35 06/28/10 7da 41e0 4c28cb1f 0 Method Loading Method LCV_GCCONTROL.M 12:17:35 06/28/10'
Temporary Solution:
n/a
Fix Information:
Fixed in GC Driver revision 6.10 [027]. This driver version is included in B.04.03 DSP1 ChemStation and C.01.02 OpenLAB ChemStation and A.04.02 OpenLAB EZChrom.
Keyword: 68xx Driver
One-line Description:
"Method modified after loading" occurs when running a sequence (6890 RC.NET)
Problem:
The "Method modified after loading" flag occurs after lauching the sequence. Started the sequence from the ChemStation software. The method modification status indicates that the method has changed when it was not changed and the report will indicate 'modified after loading'. The GC had a calibrated capillary column that was changed to a packed column. When using this method that now has a packed column, the GC reports an error in the length parameter, and that is causing the method download not to reset the 'modified after loading' flag.
Temporary Solution:
There is a workaround: bring Method into MethodEdit UI, and "reconfigure" column (Packed to Packed would be OK) to get new structure.
Fix Information:
Fixed in GC Driver revision 6.10 [027]. This driver is included in B.04.03 DSP1 ChemStation and OpenLAB C.01.02 ChemStation and OpenLAB A.04.02 EZChrom. Note, this fix will only apply to new methods, not to current methods.
Keyword: 68xx Driver
One-line Description:
GC Column Catalog File Cannot be Opened When Using the 68xx Driver Version 6.10
Problem:
It is not possible to add new columns to the GC column catalog (database) using the RC.NET driver. It is possible to add columns to the same column database using the Classic driver. User switched back to classic on same system and could add columns without error. The following error is shown: The Catalog File cannot be opened. If needed, you may obtain an appropriate replacement file from Agilent. C:\Chem32\CORE\...\DRIVERS\catalog.mdb The RC.NET driver is prevented from opening the db because you can’t have two column entries with the same model number!
Temporary Solution:
Start with a new, “stock” db column catalog file from another system.
Fix Information:
This has been fixed in 68XX Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.
Keyword: 68xx Driver
One-line Description:
Sequence continues while GC temperature is not ready
Problem:
When running a sequence where consecutive runs use the same method using the ALS on a data system containing 6890 driver version 6.10.027 or earlier, the data system may proceed to start a new run before the GC has equilibrated after the conclusion of the previous run. This is more likely if the method contains a post run period of sufficient length that the GC equilibrates to the post-run conditions and the data system has completed data analysis well before the end of the post-run period. This could result in a (retention time) RT shift.
Temporary Solution:
n/a
Fix Information:
Patch available for RC.NET GC driver revision 6.10 [027] for B.04.03 DSP1 [54]. Fixed beginning in: B.04.03 SP1 [87] and OpenLAB C.01.03 [37].
Keyword: 68xx Driver
One-line Description:
A Method Can be Created Which Will Not Download to the Instrument and the Sequence Does Not Run
Problem:
Method download fails, sequence aborts after start; no injection made Error reported in logbook: Attempt to download method to GC failed. The Data System will not allow the run to continue when the method download has failed.
Temporary Solution:
n/a
Fix Information:
This fix is in the 68xx driver version 6.10.032 and 6.11.001. This fix is included in B.04.03 SP1 and OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.
Keyword: 68xx Driver
One-line Description:
GC Back Injector Not Present in Edit Method Screen
Problem:
After successfully running a sequence with a back method, another back method is loaded. The back injector is not present in the edit method screen.
Temporary Solution:
Change a parameter at the GC keyboard and click "upload from instrument" in the edit method screen. The back injector will show in the edit method screen.
Fix Information:
This has been fixed in 68XX Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.
Keyword: 68xx Driver
One-line Description:
Validation failed shown on left corner of Resolve Method screen
Problem:
Load a method which was developed for both detectors and has the front detector signal chosen under signals tab. Load this method on a GC with no front detector - method resolution. Changes made to the detector signal tab are not detected, tab remains yellow (invalid).
Temporary Solution:
Click on a different tab and modify the value of any field to a valid value or click on the configuration tab and then click away. Change the signals tab to a valid value or turn it off to resolve the method.
Fix Information:
This has been fixed in 68XX Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.
Keyword: 68xx Driver
One-line Description:
Injector Panels in GC Edit Method Show as Uninstalled When an Injector is Installed.
Problem:
The front injector was unplugged. Close software, plug injector in, restart software. Method Configuration/ALS panel and Injector panel both say: Front Injector Not Installed. The GC shows that it is installed. Software inlet and injection parameters panel show that it is installed as well.
Temporary Solution:
n/a
Fix Information:
This has been fixed in 68XX Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.
Keyword: 68xx Driver
One-line Description:
Plunger Speed is Not Saved with 6850 GC Methods
Problem:
Plunger speed parameters are not saved in methods.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 6.11[002]. This fix is included in B.04.03 SP1 ChemStation, C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom.
Keyword: 68xx Driver
One-line Description:
Opening the GC method parameters to only look (no change made) and selecting OK to close stops the sequence or single run
Problem:
During a run, the GC method parameters are opened via the instrument diagram icon. Select OK after looking at the oven parameter to close the screen. The sequence or single run will just stop and the method is dirty. See Log File for error message. Log File: 67 41e2 4e69f8b3 0 Method Method started 07:29:55 09/09/11 67 41e2 4e69f8b4 0 Method Instrument running sample Vial 1 (front) 07:29:56 09/09/11 67 41e2 4e69f8b4 0 Method Instrument running sample Vial 2 (back) 07:29:56 09/09/11 0 1da5 4e69f91a 0 6890 1 Attempt to download method to GC failed. 07:31:38 09/09/11 67 41e2 4e69f91a 0 Method Instrument Error - Method/Sequence stopped 07:31:38 09/09/11 67 41e2 4e69f91a 0 Method Method aborted 07:31:38 09/09/11
Temporary Solution:
Click the Cancel button rather than the OK button if the intent was to check the values of some of the instrument setpoints. This would not interfere with any run which is in progress.
Fix Information:
Fixed in 68xx driver revision 6.11[008]. This is included in B.04.03 SP1 and OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.
Keyword: 68xx Driver
One-line Description:
Method Contents Report Displays the Incorrect Configuration of Front and Back Detectors for 6890 GC
Problem:
The report displays the incorrect configuration. With signal-1 selected as the back detector and integration events for back detector. The method contents report displays the integration events for the front detector and the remaining paramaters are for the back detector are configured to the front detector.
Temporary Solution:
n/a
Fix Information:
Fixed in 68xx driver revision 6.11[003]. This fix is included in B.04.03 SP1 ChemStation, C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom and later.
Keyword: 68xx Driver
One-line Description:
Cannot Run a Packed GC Column with a 6890
Problem:
Method download fails, sequence aborts after start when trying to run a 6890 with a packed column. Error reported in logbook: Attempt to download method to GC failed. The Data System will not allow the run to continue when the method download has failed.
Temporary Solution:
This can be run by setting the column as a capillary column but that changes the chromatography.
Fix Information:
Fixed in driver revision 6.10[032] and 6.11[001]. This fix is included with B.04.03 SP1 GC ChemStation, C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom
Keyword: 68xx Driver
One-line Description:
Pulsed Split/Splitless Injection Does Not Wait For Ready on a 6890.
Problem:
The GC does not wait for the inlet pressure to increase to and stabilize at the pulsed pressure before injection and data acquisition. The GC starts and injects sample at the same time, then starts data acquisition and increase of inlet pressure to pulsed pressure.
Temporary Solution:
If the 6890 is operated from the GC keyboard, the GC has the expected result.
Fix Information:
Fixed in 68xx driver revision 6.11[007]. This fix is included with B.04.03 SP1 ChemStation, OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03
Keyword: 68xx Driver
One-line Description:
In 68XX GC Driver Version 6.01, 2 Injections are Made When Using the PTV
Problem:
Set up the injector to make 1 ul injection x 1. The number of injections will revert to 2 even if the user enters 1.
Temporary Solution:
n/a
Fix Information:
This is fixed in driver revision 6.10 [27] which is included in B.04.03 DSP1 [54] GC ChemStation, C.01.02 OpenLAB ChemStation and A.04.02 OpenLAB EZChrom.
Keyword: 68xx Driver
One-line Description:
Vial left in gripper; no message shown to alert user about the problem
Problem:
After a stop, if there is a vial in the gripper it is not returned to its original location, it is kept in the gripper until the start of a new run. In the interim, if a new vial is placed in the open location, during the next run the system runs into an error because the gripper wants to return the vial to its original location which is now occupied.
Temporary Solution:
n/a
Fix Information:
This is fixed in GC Driver version 6.21, which is available with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.
Keyword: 68xx Driver
One-line Description:
6850 vial numbers are not consistent with standard numbering system.
Problem:
A 6850 with a G2613A tower and a three vial turret will accept only single digit vial locations, ie., 1, 2, 3 instead of 101, 102, 103. Attempts to enter the correct vial number generates an error message "The vial 101 is not allowed with the current device configuration." The also occurs when entering correct vial locations in a sequence table.
Temporary Solution:
n/a
Fix Information:
Fixedi n 68XX driver revision 6.11. This driver is included with B.04.03 SP1 ChemStation, OpenLAB C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom.
Keyword: 68xx Driver
One-line Description:
Method Resolution Changes Oven Temperature Program
Problem:
Load a method from a 6890 GC with a fast oven on a 6890 GC with different hardware and a normal oven. Method resolution occurs. The oven temperature program from the fast oven GC has not been loaded on the normal oven GC.
Temporary Solution:
n/a
Fix Information:
Fixed in 68xx driver revision 6.11[002]. Fix is included in B.04.03 SP1 ChemStation, OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.
Keyword: 68xx Driver
One-line Description:
Column info is missing with 6890
Problem:
The column specified as being used is missing information: description, Manufacturer and Model Number.
Temporary Solution:
n/a
Fix Information:
Fixed in 68xx driver revision 6.11[005]. This fix is included in B.04.03 SP1 GC ChemStation, C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom.
Keyword: 68xx Driver
One-line Description:
Method Modification Flag is Set after Edit Method: Occurs on first connection or after upload from GC.
Problem:
1. First Connection to an online GC and save the method OR connect to an instrument and save the method, open method editor, click on ‘Upload from instrument’. Click ‘Ok’. 3. Click ‘Edit entire Method’, go on clicking ‘Ok’ without making any change. 4. Method is dirty at the end. This only occurs on 6850 GCs with G4513A injectors.
Temporary Solution:
Save the method as a new name after initial connection or subsequent method upload.
Fix Information:
Fixed in 68xx driver revision 6.11[008]. This is included in B.04.03 SP1 and OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.
Keyword: 68xx Driver
One-line Description:
Method Ddownload Failure Occurs Due to Column Calibration Limits
Problem:
A valid column calibrated using calculate diameter function will result in a method download failure if the change from the original diameter is more than 20 microns. This occurs after OKing method changes that are downloaded, and will occur even if the column values are valid dimensions.
Temporary Solution:
n/a
Fix Information:
This is fixed in 68XX GC Driver Version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.
Keyword: 68xx Driver
One-line Description:
Instrument actual value for COC inlet does not display correctly
Problem:
6890N GC with COC inlet connected to FID. The COC inlet pressure setpoint reflects the correct pressure set for the column, however the instrument actual value (blue text) in the setup method screen is displayed as 0.0.
Temporary Solution:
Enable setup actuals (inlet) in the setup method screen, the inlet pressure is properly displayed.
Fix Information:
Fixed in 68xx driver revision 6.11[018]. This driver is included in B.04.03 SP1 ChemStation and OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.
Keyword: 68xx Driver
One-line Description:
Instrument actual value for anode purge flow on an ECD does not display correctly
Problem:
The ECD anode purge flow setpoint reflects the correct value, however the instrument actual value (blue text) in the setup method screen is displayed as 0.0.
Temporary Solution:
Enable setup actuals in the setup method screen, the actual anode purge flow displayed properly.
Fix Information:
Fixed in 68xx driver revision 6.11[018]. This driver is included in B.04.03 SP1 ChemStation and OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.
Keyword: 68xx Driver
One-line Description:
6850 Support of Column Compensation
Problem:
The 6850 GC does not support column compensation from the software. This is only able to be done with the G2629A handheld controller which is now obsolete.
Temporary Solution:
n/a
Fix Information:
This is fixed in 68XX GC Driver Version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.
Keyword: 68xx Driver
One-line Description:
Easy SamplePrep reset of one or any group of resources creates log entry: 'User reset all resources'
Problem:
Create a resource layout. Select menu item Reset Resources... From the list, select one or several resources and OK. The logbook always logs 'User reset all resources' no matter how many resources are actually reset.
Temporary Solution:
n/a
Fix Information:
There is no fix available.
Keyword: 68xx Driver
One-line Description:
Data Files do not Contain the Correct Signals if the Front Inlet is Connected to Back Detector
Problem:
Data files do not contain the correct signals if front inlet is connected to back detector
Temporary Solution:
n/a
Fix Information:
This is fixed in GC Driver version 6.21, which is available with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.
Keyword: 68xx Driver
One-line Description:
Slow download speed from GC Datafiles stored on network drive
Problem:
Loading GC6890 data files stored on a network drive to a ChemStation C.xx version is very slow when the operating system is Windows 7. The data is loaded in small chunks instead of reading one big block. Only 6890 GC files are showing this issue. Agilent does not support the usage of a mapped network drive during data acquisition. Accessing and loading a data file from a mapped network drive will take few seconds longer than the same file stored locally in the ChemStation computer. Active file scanning done by an antivirus software may also impact the loading speed.
Temporary Solution:
n/a
Fix Information:
No fix planned.
Keyword: 68xx Driver
One-line Description:
C.01.06: 6850 classic driver will not accept valid vial locations in Run Method, Sequence, or Easy Sequence
Problem:
It is not possible to change the default '1' Vial location in Run method. Changing the vial entry in the sequence table results in red text 'Invalid Injection Location: Front'. It is not possible to enter vial locations in Easy Sequence Setup which results in either 'Injection Location not available' or 'Obj Reference not set to an instance of an object'.
Temporary Solution:
n/a
Fix Information:
Customers upgrading to C.01.06 must re-configure their 6890/6850 to use the RC.NET driver.
Keyword: 68xx Driver
One-line Description:
The Oven Ramp table does not display correctly in IR with Acquisition method selected.
Problem:
When you specify an Intellignet Report with Method Info report added to the Sample Summary report to get the Acquisition method it is formatted incorrectly..
Temporary Solution:
n/a
Fix Information:
The rows in tables now have the same names and IDs so that the data system can organize them into a table. Fixed in revision 6.25 (GC Driver revision A.02.05).
Keyword: 68xx Driver
One-line Description:
6890 PTV Inlet not recognizable in OpenLAB EZChrom
Problem:
The Extracted configuration shows PTV back inlet and AIB Back Detector. Launching the instrument offline shows the column connected to the Back Inlet and the Back Detector.The Module Configuration Lists the PTV Back Inlet with Blank Back Detector, however the Back Inlet is seen as Unrecognized inlet type.
Temporary Solution:
n/a
Fix Information:
The code wasn't checking for the right inlet type code for a PTV mounted in the back. Fixed in revision 6.25
Keyword: 68xx Driver
One-line Description:
Multiple errors when the 6890 doesn't have any inlets
Problem:
The OpenLAB CDS ChemStation gives errors when the 6890 doesn't have inlets installed and both inlets are configured as none on the 6890 GC.
Temporary Solution:
n/a
Fix Information:
The error was occurring in the code which set up the possible inlet connections for columns. Fixed in revision 6.25
Keyword: 68xx Driver
One-line Description:
Sequence table Inj Vol Override goes to wrong stops for 6890/7683: C.01.05, A.04.05, C.01.06, A.04.06
Problem:
If an Injection volume is entered into the Sequence which overrides the method injection volume setting the injector goes to the wrong stops for a 6890 with a 7683 ALS. The injection volume will be incorrect for versions C.01.05, A.04.05, C.01.06, A.04.06 of the software
Temporary Solution:
n/a
Fix Information:
Fixed in 68xx Driver Version 6.23 [011].
Keyword: 68xx Driver
One-line Description:
Often get two or more ChemMain.exe's running following shutdown of Control Panel or online session.
Problem:
Run Task manager always on top. With no ChemMain.exe's running, open a 6890 online session. Close session and the ChemMain.exe does NOT go away.
Temporary Solution:
n/a
Fix Information:
Fixed in A.02.05
Keyword: 68xx Driver
One-line Description:
PDID Valco Detector isn't recogonized in the RC.net driver in OLCS EZChrom
Problem:
Customer running a a 6890 with OLCDS EZChrom A.04.05. The 6890 that they have has a 3rd party detector, PDID from Valco. The old EZChrom driver recognized the detector, but the new RC.net driver does not recognize it.
Temporary Solution:
n/a
Fix Information:
Fixed in A.04.06
Keyword: 68xx Driver
One-line Description:
6890N with 7693 barcode scanning checksum mismatch does not return vial to tray and skips a vial.
Problem:
There was a timeout in the driver; if the reading of the bar code was not complete by the time that the timer timed out, the driver would declare that the reading had terminated. This was misleading the data system into advancing to the next vial. The time required to finish reading the bar code or declare that the attempt had finished is indeterminate. If the tray is parked, the system will sit there waiting to move until the tray is unparked. The timeout was removed from the code.
Temporary Solution:
n/a
Fix Information:
Fixed in A.03.01
Keyword: 68xx Driver
One-line Description:
Customer can't install colum information from catalog if column description is blank
Problem:
OpenLAB EZChrom A.04.06 and 6890N Step: 1, Customer selects 122-10BE column. this column doesn't have description. 2, Error message appears when you try to install this column.
Temporary Solution:
n/a
Fix Information:
Fixed in A.03.01
Keyword: 68xx Driver
One-line Description:
EZChrom: Installation of GC Drivers A.03.02 breaks a 7697 configuration if installed
Problem:
Installation of GC Drivers A.03.01 breaks a 7697 configuration if installed.
Temporary Solution:
Uninstall the 7697 HS driver and install the latest HS driver.
Fix Information:
None
Keyword: 68xx Driver
One-line Description:
Setting a Hold time of 0.0 does not work correctly with GC6890
Problem:
Setting up Hold Time of 0.0 does not work correctly with GC6890.
Temporary Solution:
No fix planned.
Fix Information:
None
Keyword: 68xx Driver
One-line Description:
Instrument session continues to run in background after session is closed.
Problem:
When an instrument session is closed the ChemMain.exe process continues to run in the background when an instrument connection isn't made or lost. This can lead to multiple ChemMain.exe processes running at the same time causing performance issues.
Temporary Solution:
Close the ChemMain.exe process in Task Manager.
Fix Information:
Fixed in GC Drivers version A.02.05.024
Keyword: 68xx Driver
One-line Description:
6890 online help has the bad strings in Injector page.
Problem:
6890 online help has the bad strings in Injector page.
Temporary Solution:
n/a
Fix Information:
Resolved in B.01.02
Keyword: 68xx Driver
One-line Description:
6890, 6850 session start up hangs or is very slow to start up with Windows 10 version 1511 as the operating system
Problem:
Online 6890, 6850 ChemStation session startup hangs on "Loading Toplevel Menu and view" , initializing start up Macros when using Windows 10 version 1511 as the operating system. The session hangs up when initializing the sampling and instrument diagrams. The Online session start up could take more than 30 minutes to initialize.
Temporary Solution:
n/a
Fix Information:
Upgrade to Windows 10 version 1703.
Keyword: 7890A Control
One-line Description:
7890A Runs cannot be started until the GC and injector are idle, view GC Status
Problem:
Occassionly, when trying to start a run or sequence on a GC whose front display and ALS are Ready, fails to start. Logbook shows: 'Runs cannot be started until the GC and injector are idle'. This can occur after an abort or stop of sequence, or connecting to a GC that is controlled by another PC.
Temporary Solution:
Pressing "stop" on the GC front panel enabled the GC to makes runs again. Power cycling the GC may be necessary.
Fix Information:
Fixed in revision B.04.03.
Keyword: 7890A Control
One-line Description:
GC Backflush Wizard does not work properly in C.01.07 SR1 HotFix01 [113]
Problem:
The Backflush Wizard starts and runs the first sample but at the end of the run does not bring up the box to select the last peak of interest. It then goes into a loop where it only runs the sample and never completes the backflush set up.
Temporary Solution:
Backflush Wizard works in C.01.07 [27] and C.01.07 SR1 [110].
Fix Information:
Resolved in revision OpenLAB CDS ChemStation C.01.07 SR2.
Keyword: 78xx Driver
One-line Description:
Error Message when certain methodsare loaded onto an offline configured 7890.
Problem:
If methods ISTD_EX.m or DEF_GC.M are loaded onto a configured 7890 offline session an error message (Exception in command "translateMethod$" occured). These 2 methods are 6890 specific and should not be loaded onto a 7890.
Temporary Solution:
n/a
Fix Information:
Fixed in GC driver 4.01 [047]. This driver was included in the following datasystem packages: GC ChemStation B.04.03 SP1 [87] and OpenLab CDS C.01.02 GCMSD ChemStation E.02.02 OpenLab CDS EzChrom A.04.02
Keyword: 78xx Driver
One-line Description:
Shutdown events are not displayed on the software Status screen
Problem:
Shutdown events, such as a thermal or pneumatic shutdown, displayed on the GC Keyboard are not reported to the software and therefore not displayed on the “Agilent GC Status” screen.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.01. 7890A requires a minimum Firmware revision of A.01.12. This driver is included with the following datasystems: GC ChemStationB.04.03 DSP1 [52] and OpenLAB C.01.02 GCMSD ChemStation E.02.02 OpenLab CDS EzChrom A.04.02
Keyword: 78xx Driver
One-line Description:
Barcode mismatch error
Problem:
When a barcode mismatch occurs with overlapped injections (option of do not inject on barcode mismatch), the mismatched vial is taken to the turret and chemstation stalls with the sequence running and waiting. An abort is required to recover from this error.
Temporary Solution:
n/a
Fix Information:
Fixed in 78xx driver revision 4.02[023]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Keyword: 78xx Driver
One-line Description:
Sample Prep Program with only a 7693A Back Injector does not save
Problem:
The problem occurs when a 7693A Tray is installed and there is an injector in the Back and no injector in the Front. If a Sample Prep Program is created, the default values for a Move command are incorrect, and cause erratic behavior.
Temporary Solution:
See Service Notes 7693A-002 and 7693A-002A for available patch which is used with the 2.12.040 driver.
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
ALS Errors Abort Sequence; no Retry/Abort dialog box option offered
Problem:
The setting is selected by opening the Tray/Overlap tab in Method Editor. The Instrument Status must be opened to view the dialog box. Pause is the default setting for ChemStation and MassHunter. Skip is the default setting for EZChrom.
Temporary Solution:
n/a
Fix Information:
Fixed in the 5.01 Driver. This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
Keyword: 78xx Driver
One-line Description:
Method edit: Must click twice with the mouse on each data field to change number
Problem:
In Method Edit of GC setpoints, the user has to click twice with the mouse on each data field to change a numerical entry. The first click highlights the field, the second click highlights just the number, and then the number can be changed.
Temporary Solution:
n/a
Fix Information:
This problem is fixed in Software Driver Version: 3.01. This driver is included in the following datasystem packages: GC ChemStation B.04.02 SP1 [208] Ezchrom Elite 332 SP2 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
ChemStation does not show "Hydrogen Safety Shutdown" in 7890A status
Problem:
When the 7890A goes into hydrogen safety shutdown the ChemStation shows "Front inlet flow shutdown" and other messages, but not "Hydrogen Safety Shutdown". The chemstation does not report the hydrogen shutdown but it does list the other causes. In addition, the heated zones can be turned on from the chemstation. Since a hydrogen shutdown should require operator intervention AT THE GC remote control from the (any) chemstation should be disabled.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.01. 7890A requires a minimum Firmware revision of A.01.12. The 4.01 driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
7890A valve names added in the config screen but are not available in the method
Problem:
The 7890A valve names can be added in the configure screen but the valve names are not available in the method.
Temporary Solution:
n/a
Fix Information:
This is fixed in GC driver version 3.01. This driver is included in the following datasystem packages: GC ChemStation B.04.02 SP1 [208] Ezchrom Elite 332 SP2 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
7820 method printout includes parameters and setpoints unavailable for PP inlet
Problem:
Purged Packed Inlet includes parameters: 'Pressure On 20.261 psi' 'Total Flow, On, 5.4077 mL/min' 'Septum Purge Flow On -0.0089673 mL/min' Pressure, Total Flow, and Septum Purge Flow are parameters that are not available for this GC.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.01. This driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
7820 method printout inserts negative number for Septum Purge Flow, invalid parameter
Problem:
7820 method printout for a Purged Packed (PP) inlet includes not only the invalid parameter 'Setptum Purge Flow', but inserts a negative number as its setpoint.
Temporary Solution:
n/a
Fix Information:
Fixed in Driver revision 3.02. This driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.01 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
If 7890 gets into an error status (flame off) this is NOT reported within the 7890 status window
Problem:
If the 7890 gets into an error status (such as flame off) this is NOT reported within the 7890 status window The flame failed to ignite at the FID but the status window ONLY displays that the GC is Not Ready.
Temporary Solution:
n/a
Fix Information:
Fixed in 78xx driver revision 4.02[009]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Keyword: 78xx Driver
One-line Description:
Read barcode just before injection and use barcode as sample name in report
Problem:
Users want to add vials to the tray as they go and have the vial bar code read just before injection and recorded as the name of the sample. Reading the vial ahead of time into the sequence table for each vial is time consuming. Read the barcode just before injection since users are adding samples as they go. This is working with the 6890 Classic driver.
Temporary Solution:
n/a
Fix Information:
This problem is fixed in Software Driver Version: 3.01. The driver is included in the following datasystem packages: GC ChemStation B.04.02 SP1 [208] Ezchrom Elite 332 SP2 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
7890A: Column part number is not included in the method report
Problem:
The 7890A column part number is not printed in the method report (English, Japanese, Chinese). Revision B.03.01 SR1.1 does include the column part number in the method report. B.03.02, B.04.01, B.04.01 SP1, and B.04.02 do not include the part number of the column in the method report.
Temporary Solution:
n/a
Fix Information:
This problem is fixed in Software Driver Version: 3.01. The driver is included in the following datasystem packages: GC ChemStation B.04.02 SP1 [208] Ezchrom Elite 332 SP2 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
Partial data collection is not available on the signals screen
Problem:
On the signals screen, need the ability to collect partial data for specified times.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.01 [31]. This driver version is included with the following data systems: GC ChemStation B.04.03 DSP1 [54] GCMSD ChemStation E.02.02 OpenLab CDS EzChrom A.04.02
Keyword: 78xx Driver
One-line Description:
Need ability to do Digital switching for detector signals in Runtime events
Problem:
The ability to do digital switching of detector signals in the Runtime events table is needed. It is only possible to do analog switching.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.01. 7890A requires a minimum Firmware revision of A.01.12. Driver version 4.01 is included with the following data systems: GC ChemStation B.04.03 DSP1 [52] GCMSD ChemStation E.02.02 OpenLab CDS EzChrom A.04.02
Keyword: 78xx Driver
One-line Description:
Read barcode just before injection and use barcode as the sample name in the report
Problem:
It is not possible to add vials to the tray and have the vial barcode read just before injection and recorded as the name of the sample. Reading the vial ahead of time into the sequence table for each vial is time consuming. It is not possible to read the barcode just before injection and have the barcode used as the sample name in the report for 78xx.
Temporary Solution:
n/a
Fix Information:
Fixed in 78xx driver 3.01. The driver is included in the following datasystem packages: GC ChemStation B.04.02 SP1 [208] Ezchrom Elite 332 SP2 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
Park during sequence with Overlap causes unresponsiveness
Problem:
Running ChemStation B.04.02.96. Start a sequence with overlapped injection set to immediately after injection. While the sample is being delivered to the turret, the Park button is pressed on the 7693A Tray. After delivering the vial, the tray is put in the Park position and the Injector completes the injection. At this point the ChemStation is unresponsive. (For example, when the ChemStation is in this state, any report printed to the screen can not be closed.) The ChemStation responds after the Park button is pushed and vial is returned.
Temporary Solution:
The ChemStation responds after the Park button is pushed and vial is returned.
Fix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
7890GC: inlet pressure unit inconsistency - method audit trail vs instrument setup
Problem:
Problem with the Agilent 7890GC inlet pressure unit: With method audit trail turned on, make changes to the inlet pressure (configured in psi in this case), save the method, check method audit trail log, it recorded the pressure in pascal.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 3.02. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] Ezchrom Elite 332 SP2 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
RTL: New calculated pressure doesn’t update column 2 - 6 on current 7890A methods
Problem:
RTL: New calculated pressure doesn’t update column 2 - 6 on current 7890A methods. Only Column 1 is updated
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.02[034]. This driver is included with the following datasystems: GC OpenLab CDS C.01.01 GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.04
Keyword: 78xx Driver
One-line Description:
New Abort Behavior
Problem:
Once it detected a vial not found, everything would abort, including samples that were in process (that have already been injected and are being analyzed). The new behavior, is that the vial that is currently being processed will be finished. Everything else will abort, if the abort option is selected.
Temporary Solution:
n/a
Fix Information:
Fixed in the 5.01 Driver. This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
Keyword: 78xx Driver
One-line Description:
7890A: Shutdown events are not published to the software GC Status screen
Problem:
The thermal inlet shutdown is not reported in the software “Agilent GC Status” screen. Only the first event, which is 'not ready', is reported to the Agilent GC Status screen. The second event, which is the shutdown, is not displayed.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.01. 7890A requires a minimum Firmware revision of A.01.12. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
Upgrading B.01.03 to B.04.02 GC ChemStation: Addition of new columns to the column catalog (database) is not possible; error occurs
Problem:
Upon upgrading from B.01.03 to B.04.02 it is no longer possible to add new columns to the GC column catalog (database). The following error is shown: The Catalog File cannot be opened. If needed, you may obtain an appropriate replacement file from Agilent. C:\Chem32\CORE\...\DRIVERS\catalog.mdb
Temporary Solution:
The problem is that there are several columns that have the same manufacturer and model number. Our new database software doesn’t handle this. The newer database software (in B.04.02 for example) no longer allows duplicates of manufacturer + model number. i.e., there can only be one HP, HP-5 in the database. Note: the new software will inform the user that the model number “xxx” is already in use and request that a different value be chosen. (The user is returned to the original entry screen with all the info filled in.) Temporary Fix: Before upgrading, fix the problem with the column database duplicates while running B.01.03 for example. Review the column catalogs and make sure there are no duplicates and fix any that exist before the upgrade.
Fix Information:
This is fixed in GC Driver version 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
Electrometer on/off not needed for FID or NPD
Problem:
The software shows an on/off checkbox for the electrometer on both the FID and NPD. This setting is not available in the normal local user screens in 7890A. The checkbox should be removed from the edit parameters screen.
Temporary Solution:
n/a
Fix Information:
This is fixed in driver version 3.01 which started to be included in the following data system packages: GC ChemStation B.04.02 Driver Service Pack 3 (DSP3) [256] Ezchrom Elite 332 SP2 and OpenLab EZChrom GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
Freeze Signal Value and Resume Signal Value missing from runtime events
Problem:
Freeze Signal Value and Resume Signal Value missing from runtime events
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.01. 7890A requires a minimum Firmware revision of A.01.12. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
The Override Column Max Temp Checkbox Behavior
Problem:
The override column max temp checkbox in the oven parameter screen is automaticaly checked by default in versions of the driver prior to 5.01. In driver version 5.01, it will not be checked and the box will by highlighted in yellow if the column maximum temperature exceeds the value that is recommended for the column that is installed. To make the box not yellow, either check the "override column max temp" checkbox, or enter a value that is lower than the recommended maximum temperature for the installed column.
Temporary Solution:
n/a
Fix Information:
Change of Behavior in Driver Version 5.01 This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
Keyword: 78xx Driver
One-line Description:
Cryo Trap Run Time Event Support
Problem:
When a thermal aux zone is configured as a Cryo Trap, the cooling of the trap is accomplished via a run time event entitled Cryo trap cooling with an ON an OFF state.
Temporary Solution:
n/a
Fix Information:
Fixed in the 5.01 Driver. This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
Keyword: 78xx Driver
One-line Description:
Error in flow setpoint upload when method has more than 100 events
Problem:
When a method with more than 100 events is uploaded from the GC, all the steps above the 100th event that are related to flow (ml/min) are multiplied by a factor 1000.
Temporary Solution:
n/a
Fix Information:
Fixed in 7890A firmware A.01.11. No software update is necessary.
Keyword: 78xx Driver
One-line Description:
Running a sequence requires the arm from the tray, G4514A to be in the UN-PARKED position
Problem:
The tray "Park" status overrides the start of a sequence.
Temporary Solution:
n/a
Fix Information:
The G4514A arm from the tray must be UN-PARKED in order for a sequence to proceed. User must press Park button to UN-PARK arm.
Keyword: 78xx Driver
One-line Description:
Extended Parameters does not disable the BCR when no hardware is present
Problem:
Extended Parameters for both the native sequence Parameters and for Easy Sequence Setup does not disable the Bar Code Reader section when no such hardware exists.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
Switched and standard mode for septum purge not available on MMI
Problem:
For the split/splitless inlet, it is possible to run the septum purge in either standard mode or "switched" mode. In the switched mode, the septum purge is turned off during the injection cycle so that the only flow out of the inlet is through the column. The MMI would benefit from having this mode as well, particularly for splitless and solvent vent modes. In both modes on the MMI, the septum purge does not remain off until the inlet switches back to the purge state.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
Negative Split Splitless inlet Pressure values during Post Run Backflush
Problem:
When using a 7890A GC with Split Splitless inlet to perform backflush the Inlet Setpoint and Actual Pressure assume negative values on the GC and Chemstation display during backflush. This occurs with a splitter (>2 columns) or with a purged union (Primary column plus restrictor). The columns must be running in "Flow" mode. A similar symptom is the inlet pressure gets set to "0.00" PSI during backflush rather than the wizard recommended value of "1.0" PSI.
Temporary Solution:
The workaround is to set the columns to "pressure" mode then back to flow and re-save the method.
Fix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
Column compensation menu is available during the sequence
Problem:
The Column compensation menu is available during the sequence and if selected will stop the sequence.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
Column: need to include Ramped and Constant pressure and flow modes as selection
Problem:
User interface needs Constant Flow, Constant Pressure, Ramped Pressure, and Constant Flow for the Control Mode selections.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.01. This driver was included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] GCMSD ChemStation E.02.02 OpenLab CDS EzChrom A.04.02
Keyword: 78xx Driver
One-line Description:
GC goes offline momentarily because of SOAP Exception
Problem:
Setup 7693 injector and Tray. Put Tray in Park and then tried to start a run. The ChemStation showed that the GC went offline momentariily. The Event Log showed the following events : Disconnecting because a System.Web.Services.Protocols.SoapException was caught when sending a command; 29: GC refused to do an injection The connection state of the GC at 130.30.245.60 is now Offline Attempting to connect to 130.30.245.60 In DataChannel[1].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000 In DataChannel[2].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000 In DataChannel[3].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000 In DataChannel[3].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000 The connection state of the GC at IP address is now Online Connected to 7890 at 130.30.245.60, using license key Cannot start runs until GC and injector are idle.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Keyword: 78xx Driver
One-line Description:
No indication of inlet pressure shutdown in software status with 7890A
Problem:
Inlet pressure shutdown advisories indicating a problem with the instrument not shown in software status user interface.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.01. 7890A requires a minimum Firmware revision of A.01.12. This driver is included with the following datasystems: GC ChemStation B.04.03 DSP1 [52] and OpenLab CDS C.01.02 GCMSD ChemStation E.02.02 OpenLab CDS EzChrom A.04.02
Keyword: 78xx Driver
One-line Description:
GC Keypad Lock Behavior
Problem:
In configuration we now offer the option to configure how the keypad lock will behave.
Temporary Solution:
n/a
Fix Information:
The driver with this new behavior is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
Keyword: 78xx Driver
One-line Description:
GC Instrument Status for Detectors
Problem:
In previous versions of software, only the temperature of the detectors was shown in instrument status. In driver version 5.01, we show all GC detector status in instrument status.
Temporary Solution:
n/a
Fix Information:
Fixed in the 5.01 Driver. This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
Keyword: 78xx Driver
One-line Description:
It is not possible to set a positive pressure when column outlet is set to "other"
Problem:
If the colum is configured from one inlet to other (in case for a non-purged CFT or other applications) it's not possible to define the pressure of the outlet. It's possble to set it from the GC keyboard.
Temporary Solution:
n/a
Fix Information:
Fixed in the 5.01 Driver. This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
Keyword: 78xx Driver
One-line Description:
FID Flame out; no feedback in software
Problem:
The GC status and instrument actuals list the signal value as 0.0 but does not provide feedback if the flame is out.
Temporary Solution:
n/a
Fix Information:
Fixed in 78xx driver revision 4.02[001] This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Keyword: 78xx Driver
One-line Description:
uECD and TCD report printout for makeup flows is wrong
Problem:
7890A has a uECD detector with flows configured. With the following method settings... Makeup Flow(N2) : 30 mL/min Constant Column + Makeup : 31 mL/min ... the method printout reads: Makeup Flow(N2) : 31 mL/min Constant Column + Makeup : 0 mL/min
Temporary Solution:
n/a
Fix Information:
Fixed in 7890A SW Driver 4.01 [048]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Keyword: 78xx Driver
One-line Description:
GC continues when STOP on 7820 is pressed
Problem:
User makes a manual injection and waits until the peaks of interest elute, then hits the STOP on the 7820 front panel the GC stops (and resets the initial GC settings) but the chemstation continues and reports the system is running data analysis but does not go into data analysis. System appears to hang up waiting on the expected run time and report is generated at that point rather than directly after STOP is pressed.
Temporary Solution:
Use STOP run from the software.
Fix Information:
Fixed in driver revision 4.02[11]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Keyword: 78xx Driver
One-line Description:
Open GC edit parameters and close it without changes; method modification flag is set
Problem:
Saved a method. Open the GC edit parameters to review a setpoint and made no changes or changed the view. Selected OK and the method modification flag indicates a change was made. No changes were made.
Temporary Solution:
n/a
Fix Information:
Fixed in 78xx driver revision 4.02[019]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Keyword: 78xx Driver
One-line Description:
7890A logbook shows 7693A turret, plunger errors as GCI1 did not start a run
Problem:
7693A ALS errors, e.g. turret, plunger, door open, etc. show a non-specific logbook entry: GCI1 did not start a run.
Temporary Solution:
n/a
Fix Information:
Fixed in 78xx driver revision 4.02[023]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Keyword: 78xx Driver
One-line Description:
Unable to set equilibration time for LTM columns from Edit Parameters
Problem:
The LTM column's equilibration time is shown as 0. Change the equilibration time by entering 0.5 and downloading the method. This does not change the value at the GC.
Temporary Solution:
n/a
Fix Information:
Fixed in 78xx driver revision 4.02[005] and GC Firmware revision A.01.13. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Keyword: 78xx Driver
One-line Description:
Changes made to 7890 method from GC instrument diagram result do not load.
Problem:
Open the instrument parameter from GC Instrument Diagram and confirm sample prep is off. Set inlet heater - save changes. Open the instrument parameter from the GC Instrument Diagram and confirm sample prep is off. Change the inlet heater temp and save as a new method. Load the new method - a configuration mismatch message is displayed. Requires Easy Sample Prep
Temporary Solution:
n/a
Fix Information:
Fixed in 78xx driver revision 4.02[004]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Keyword: 78xx Driver
One-line Description:
Backflush icon is not displayed when the outlet of the column configuration is set as "Other".
Problem:
When the outlet of the column configuration is set as "Other", there is no backflush icon displayed in the 7890 Edit Parameter Screen.
Temporary Solution:
n/a
Fix Information:
Fixed in 78xx driver revision 4.02[009]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03 Please note, when the configuration is changed from backflush not available to backflush available, the user must exit the method editor and begin the edit a second time in order for the backflush icon to be shown.
Keyword: 78xx Driver
One-line Description:
When running Backflush from ChemStation with segmented columns only the in-segment is taken into account in the calculation.
Problem:
If you run Backflush from Chemstation with segmented columns (usually associated with LTM), only the in-segment portion of the configuration is used in the backflush flow rate and void volume calculations. This leaves the user unable to calculate the backflush time.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.02[023]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Keyword: 78xx Driver
One-line Description:
The display of the parameter setting screen of the method setup is small.
Problem:
In the Edit Parameter screen for the oven and the inlets, the display of this screen is small. Adjusting the method setup screen will not resize this screen.
Temporary Solution:
Right click on parameter setting screen, and go to "show graphical plots". The display will resize to a normal size.
Fix Information:
Fixed in 78xx driver revision 4.02[006]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Keyword: 78xx Driver
One-line Description:
After editing the GC method and selecting ok - validation error message is displayed.
Problem:
In driver revision 4.01, after editing the GC method and selecting ok, a "Validation Error" error message flashes in the left hand corner of the screen. User must select cancel to exit the GC parameter screen. No method resolution occurs when loading method Software loads fine and can make runs. Methods cannot be edited in online or offline mode. The configuration includes 4 packed columns that flow to the Back TCD Detector. Column 1's inlet is PCM C-2 (@ const pressure); 4's inlet is Aux EPC 5 (@ const pressure); and both 2's and 3's inlet is PCM C-1. Column 2's mode is constant flow; and 3, since it is also connected to PCM C-1, is reported as being controlled by #2.
Temporary Solution:
Workaround for customers with driver revision 4.01: replace columns #2 and #3 with a single undefined (e.g., packed) column, suitably described as being a parallel combination. The user can then set either a flow or a pressure setpoint to the new combined column #2.
Fix Information:
Fixed in driver revision 4.02. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03 OpenLab CDS ChemStation C.01.04
Keyword: 78xx Driver
One-line Description:
7820 Back Injector Only Resets Injection Source to Manual
Problem:
On a 7820, with only a Back Injector Installed the following message is displayed when launching the 7820: Upon launching the 7820A, the following message appears: DEF_GC.M: 7820 Resetting to Manual injection. For method 'DEF_GC.M, injection source 'GC Injector' conflicts with injection location. This does not happen when there is only a Front Injector installed.
Temporary Solution:
Open the injection source and change it to GC Injector.
Fix Information:
Fixed in the 5.01 Driver. This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
Keyword: 78xx Driver
One-line Description:
PCM Inlet Heater Shutdown
Problem:
7890 GC with PCM controlled inlet heater. Open Edit Parameters, enable PCM inlet heater checkbox and make a change to the heater settings. Download the method to the GC or hit apply and the inlet heater is turned OFF at the GC. If the user closes edit parameters, and then re-opens edit parameters the PCM inlet heater checkbox is disabled.
Temporary Solution:
Create your method, save, and download to the GC (the PCM inlet heater will shutdown). At the GC, set the PCM inlet heater to the desired setting. Upload the method from the GC (Instrument, Upload Method), and save your method as a new method (Save method as …). This method can now be used and loaded successfully in the future. The workaround will need to be performed any time changes are made to the method.
Fix Information:
This driver is included with the following datasystems: OpenLab GC ChemStation C.01.04 GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.04
Keyword: 78xx Driver
One-line Description:
Multiposition valve resets to position # 1 between injections
Problem:
The set position of the gas streams used for a multiposition valve is controlled in the sequence. The user can run from different gas streams by entering the different numbers in the sequence table. The position is not stored as part of the method. The MPV set position is always reset to start from position #1 between runs.
Temporary Solution:
n/a
Fix Information:
This is fixed in 7890A firmware revision A.01.14. No software update is required.
Keyword: 78xx Driver
One-line Description:
MMI in Split or Pulsed Split Mode does not allow editing of Total Flow value
Problem:
MMI in Split or Pulsed Split Mode does not allow editing of Total Flow value. The Total Flow combo box is active, but any entry made there reverts to its original entry.
Temporary Solution:
Total Flow is editable in this mode from the GC keypad in driver versions prior to 5.01.
Fix Information:
Fixed in the 5.01 Driver. This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
Keyword: 78xx Driver
One-line Description:
The GC Maintenance View May Not Change Until the Software is Closed and Re-Opened
Problem:
The Maintenance View May Not Change Until Software is Closed and Re-Opened
Temporary Solution:
In the Maintenance menu, the user has the option to change the view of the menu by changing it under the Options button. Sometimes the change may not go into effect until the software is closed and re-opened.
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
Description of How to Delete Columns from the Inventory
Problem:
Editing can only take place in the catalog file and you can edit both the inventory data and the catalog data. For an inventory item, you can edit the actual length and the comments. For a catalog item, you can edit the keywords, comments, and check/uncheck favorites. You can only delete from the catalog, this is where you delete your items from inventory. You can remove a column or an inventory item by clicking on the minus sign in the navigation bar.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
Behavior of the Maintenance Screen When Adding Modules to the GC
Problem:
You will not see the module ID that is installed, but the module location will be identified If you add a back injector, the back injector tab will come up and all relevant counters will show. The module ID will not be shown. To make it show up, you have to close the data system and reopen the datasystem.
Temporary Solution:
Add hardware modules when the hardware is turned off and the data system is disconnected.
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
ALS Behavior When Aborting a Sequence
Problem:
If a sequence is aborted in the middle of a sequence, the vial will be left in the turret until the next run is started. At that time, it will be put back in the tray.
Temporary Solution:
This is expected behavior.
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
7890B GC Column Maintenance Counter Behavior
Problem:
If you move a column from one GC to another that is using a different database, you will not see the current Maintenace Counters for the column. If you return the column to the original GC, it will have the old values.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
Update GC Column Catalog dialog is limited to 10 items; no way to display the rest (...)
Problem:
Update Column Catalog dialog is limited to 10 items; no way to display the rest (...) When you view the catalog, all items will appear.
Temporary Solution:
Description of behavior in driver version 5.01: when you view the catalog, all items will appear.
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
Method Differences Viewer Behavior with Scanned Columns
Problem:
When you scan a column in to the GC when you are not connected to the datasystem, and you connect with the data system, you will see differences in your method viewer upon opening the data system. The change will not be shown in highlight differences. No matter what is selected, the column that was scanned into the GC will be the column that is configured.
Temporary Solution:
Scan the column in with the data system connected.
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
7890B GC Method Compatibility
Problem:
The7890B GC driver included in the new data systems (A.01.05 OpenLAB CDS and B.07.00 MassHunter) supports up to 12 valves. The GC driver in previous data systems only reported 8 valves. Because of this difference, you cannot develop a method on a new data system and run it on an old data sytem. You can take a method loaded on an old data system and run it on the new data system. If you try to load a new data system method on an old data system, the behavior that you will see is that after starting a run, the software and the GC just sit there waiting for prerun to start, but nothing happens.
Temporary Solution:
n/a
Fix Information:
Documented Behavior for GC Methods Developed Using A.01.05 OpenLAB CDS and B.07.00 MassHunter.
Keyword: 78xx Driver
One-line Description:
Logbook Messages for Scanned Consumables Items
Problem:
There is no logbook message about scanned consumables or reset of Maintenance Counter values if the consumable is scanned when the Data System is Offline.
Temporary Solution:
In order for the logbook to capture messages about the scanned consumables or reset of Maintenance Counter values, make sure the Data System is connected when scanning the barcode of the consumable.
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
The Syringe and Liner are Not Shown in the Method
Problem:
The syringe and liner not shown in method. This is expected behavior for the 5.01 GC driver that is included in the C.01.05 OpenLAB CDS and B.07.00 MassHunter.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
The PC and GC Clocks Need to be Synchronized to Account for Daylight Savings Time
Problem:
For the resource conservation feature, the GC and PC clocks need to be set to the same time(synchronized). After daylight savings time, make sure the GC and PC clocks are synchronized by pressing the synchronize clocks button, which is available from the configuration screen and from the maintenance screen. The times will not be synchronized automatically.
Temporary Solution:
Synchronize the clocks by opening the Maintenance Window, select the Maintenance Actions tab and synchronize the clocks.
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
Logging of Injection Tower Location
Problem:
When the injector tower is moved from front to back position (or the opposite way) the method and sequence run correctly. However, on the report there is still the original (back) position printed, the change of the position is not tracked.
Temporary Solution:
Recommended steps: Move the injector to the new location, click "Don't resolve" (DO NOT RESOLVE THE METHOD) and then load the method for that injector position.
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
Explanation for How Default Early Maintenance Feedback Counters Work
Problem:
When "set default recommended values for limits" is selected for the first time, and if those values are not changed, the Maintenance window needs to be closed and re-opened for the values to take effect.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
7890 back injector only sequence is missing Acq.txt and acq.macaml in data directory
Problem:
Version 4.02.23 of the 78xx driver did not support the XML interfaces for reporting the method. These interfaces are used to populate acq.macaml. There is no acq.txt nor acq.macml in the data directory.
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB CDS C.01.04, (78xx Driver 4.02 [043]).
Keyword: 78xx Driver
One-line Description:
The 7890 GC Goes Offline When Non-English Characters are Entered into the Column Information
Problem:
The 7890 GC goes offline when non-english characters are entered into the column information screen. The issue has to do with the number of characters that the column description accepts.
Temporary Solution:
If non-english text is entered in the column information field and the GC can not connect to the data system, reboot the GC by holding the clear button down. Do not enter the column information in non-english text.
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
Instrument Actuals Show "GC is Offline" Quickly at the Method or Sequence Start, then Returns Back Online and the Method Does Not Run
Problem:
Instrument actuals show that the "GC is Offline" quickly at the start of the method and/or sequence and then returns back online. The method and/or sequence does not run. This is likely caused by a difference in the method parameters and what is loaded on the GC.
Temporary Solution:
n/a
Fix Information:
Open a brand new Def.M and create a new method and/or sequence. Download the method to the GC from the software.
Keyword: 78xx Driver
One-line Description:
78xx GC configuration fails; Failed to Create Instance
Problem:
There is an issue with the GC Drivers Install Package that is included on the "July 2013" Disk 3 and Disk 5 of OpenLAB CDS. When Configuring the 78xx GC, failed to create instance error message appears.
Temporary Solution:
n/a
Fix Information:
Use the August 2013 Disk 3 and Disk 5 when installing OpenLAB CDS with a GC. Disk3_M8301-60035 OpenLAB ChemStation Drivers C.01.05 (Aug 2013) Disk5_M8201-60029 OpenLAB CDS EZChrom Edition Drivers A.04.05 (Aug 2013)
Keyword: 78xx Driver
One-line Description:
Data not collected, sequence stopped when vial removed from turret after injection
Problem:
After the sample was injected, the vial was removed from the turrett. The tray showed an error and the log file reported that there was no vial in gripper. The Sequence was stopped when the vial was removed and the data was not collected.
Temporary Solution:
n/a
Fix Information:
Fixed in driver 5.03.
Keyword: 78xx Driver
One-line Description:
C.01.06: Import eMethod or Export eMethod freezes system
Problem:
In GC ChemStation C.01.06, the system will freeze when selecting either "Export eMethod" or "Import eMethod". The status line in the ChemStation will show PrintMethod macro completed, but the system will be frozen and task manager must be used to close ChemStation.
Temporary Solution:
n/a
Fix Information:
Fixed in revision 5.05
Keyword: 78xx Driver
One-line Description:
0-volume blank run in overlapped sequence causes injection
Problem:
In EZChrom, a blank run can be specified by specifying 0 as the injection volume. If a blank run is so specified following an ALS non-blank run in a sequence and the sequence is run with overlap enabled, the injector makes a zero-volume injection for that run. This means the sample will be moved from its location in the tray and subjected to any heating, mixing, bar code reading, sample washes, and sample pumps specified in the method, as well as having the vial's septum pierced and the syringe needle stuck into the sample.
Temporary Solution:
n/a
Fix Information:
Fixed in revision 5.05 Fixed in revision 5.05
Keyword: 78xx Driver
One-line Description:
Backflush Wizard never advances past Ready to add backflush screen
Problem:
For a GC with a column configuration that should allow backflush (though this is not necessary), select to launch Backflush Wizard. This displays the ‘Welcome to the Backflush Wizard’ dialog. Select the ‘Ready to add backflush to the current method?’ button and it closes, never advancing to the wizard itself.
Temporary Solution:
n/a
Fix Information:
Existence of Headspace methods were causing the problem. Fixed in revision 5.05
Keyword: 78xx Driver
One-line Description:
7890A ALS Recoverable Error Actions cannot function after upgrading to latest GC driver build on OpenLAB CDS before A.02.01
Problem:
OpenLAB CDS EZChrom A.04.03, ALS Recoverable Error Actions setting is under ALS Tab from Instrument Parameters, but from OpenLAB CDS A.02.01, the setting has been changed to EZChrom Sequence Action tab. This leads to the problem that after upgrading to latest GC Driver build, since the platform EZC A.04.03 doesn't support ALS Recoverable Error Actions setting in Sequence Action tab, therefore 7890A ALS Recoverable Error Actions cannot function since you cannot find where to set up.
Temporary Solution:
n/a
Fix Information:
The DVD will make a config available to address this issue. Fixed in revision 5.05
Keyword: 78xx Driver
One-line Description:
Standard and Easy Sample Prep programs run incorrectly in a calibration sequence (run for unknown samples only) when overlap is involved.
Problem:
Standard and Easy Sample Prep programs run incorrectly in a calibration sequence (run for unknown samples only) when overlap is involved. Sample Prep is correctly run for samples only when overlap is NOT involved. Also works fine with overlap inv lved and NO calibration. Also happens the same way in Easy Sample Prep. Calibrated sequences using Sample Prep/Run for Unknown Samples only/Overlap method do not execute correctly.
Temporary Solution:
None. Do not use 'don't prep calibration vials' and 'overlapped prep' together, as they don't work.
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
7890B configuration, model number serial number and IP address change after restart
Problem:
During installation the 7890B is configured with an IP address, the instrument configured in ChemStation and the details are confirmed. The ChemStation is restarted and the GC rebooted. The 7890 IP address changes and the Model and serial numbers change in the configuration to 7890A and CNxxxxxxxx..
Temporary Solution:
n/a
Fix Information:
Fixed in revision 5.05 [019]
Keyword: 78xx Driver
One-line Description:
Makeup Gas mode terminology should be consistent with the GC front panel
Problem:
Makeup Gas mode terminology should be consistent with the GC front panel display. Causes confusion when setting up the system.
Temporary Solution:
n/a
Fix Information:
Also changed wording for ECD and TCD to be consistent with the GC display Fixed in revision 5.05
Keyword: 78xx Driver
One-line Description:
Agilent Solvent Elimination Wizard issue
Problem:
When select Solvent Vent mode in MMI, select 10uL syringe and Injection Volume 5uL in Solvent Elimination Calculator on first window. In second window, injected volume 5uL is out of the range. Because the range is 0.01-0.25uL. This is solvent elimination calculator issue.
Temporary Solution:
n/a
Fix Information:
Fixed in revision 5.05
Keyword: 78xx Driver
One-line Description:
Clicking OK after making changes to Events does not save the changes made to the events.
Problem:
Clicking OK after making changes to Events does not save changes made to the events. If you click Apply first the events get saved.
Temporary Solution:
n/a
Fix Information:
Fixed in revision 5.05
Keyword: 78xx Driver
One-line Description:
OL CDS ChemStation C.01.06 does not download the run time events to the 7890 correctly.
Problem:
OL CDS ChemStation C.01.06 connected to a 7890A with firmware A.01.16. Select the Runtime Events screen in the software. FIRST DEFECT. Enter multiple valid events and APPLY, does not download events to the 7890. SECOND DEFECT. Exit screen and re-enter the Runtime Events screen. Enter multiple valid events and OK, does not download events to the 7890. THIRD DEFECT Exit screen and re-enter the Runtime Events screen. Enter multiple valid events and OK. Again, Select the Runtime Events screen in the software, after the OK, the Events are deleted from the table.
Temporary Solution:
n/a
Fix Information:
Fixed in revision 5.05
Keyword: 78xx Driver
One-line Description:
Can't select uECD makeup mode to "Does not affect Makeup Flow"
Problem:
Can't select uECD makeup mode to "Does not affect Makeup Flow" If you select "Included in Makeup Flow" on uECD setting screen can't select "Does not affect Makeup Flow". If you remove column configuration from uECD both selection appear but, these are grayed out.
Temporary Solution:
n/a
Fix Information:
Fixed in revision 5.05
Keyword: 78xx Driver
One-line Description:
Can't save "Included in Makeup Flow" on uECD setting screen.
Problem:
Can't save "Included in Makeup Flow" on uECD setting screen.
Temporary Solution:
n/a
Fix Information:
Fixed in revision 5.05
Keyword: 78xx Driver
One-line Description:
GC7890B IP address changes when loading default method
Problem:
GC7890B IP address was configured to an IP address on instrument configuration. Whenever you open the ChemStation and load the default method the IP address of the GC in instrument configuration will change to a different address every time. Need to manually change the IP address back to the original address via instrument configuration was required every time the ChemStation was restarted and a hinder to daily routine work.
Temporary Solution:
n/a
Fix Information:
Fixed in revision 5.05 [019]
Keyword: 78xx Driver
One-line Description:
7820 GC: Method modification flag set after starting a sequence or method
Problem:
Method is saved. Start the 7820 sequence. The method modification flag is set indicating the method was changed, however, no changes were made. The report will indicate the method was modified when it was not.
Temporary Solution:
n/a
Fix Information:
Fixed in 5.05.019.
Keyword: 78xx Driver
One-line Description:
Method printout missing column parameters: flow or pressure, setpoint On, (Initial), and post run
Problem:
Using Print Method from the ChemStation Software C.01.05, the method printout is missing column parameters: flow or pressure, setpoint On, (Initial), and post run. Column Column #1 Flow or Pressure Setpoint On (Initial) 2.3922 mL/min Post Run 2.1209 mL/min Agilent 19091J-413: 496.46579 HP-5 5% Phenyl Methyl Siloxan
Temporary Solution:
n/a
Fix Information:
Resolved in A.02.05.024.
Keyword: 78xx Driver
One-line Description:
There is an incorrect character when displaying temperature information in the method report in Japanese
Problem:
There is an incorrect character when displaying temperature information in the method report in Japanese. There is a question mark instead of a dash.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
MMI Cryo Fast Cooldown not saved
Problem:
The MMI has a parameter called Cryo fast cool down that turns the coolant on as soon as the run is finished to cool the inlet down quicker. When this parameter is checked in the Method set up screen and then applied and OK it is not downloaded to the GC and when Method edit is launched the check box is unchecked again. There is a workaround which is to manually turn it on from the GC Keyboard, check the fast cool down box and Press Upload from Instrument. This is the only way that the parameter stays in both GC and ChemStation.
Temporary Solution:
n/a
Fix Information:
Fixed in a.03.01
Keyword: 78xx Driver
One-line Description:
Method is always seen as modified, nothing can be added to the queue.
Problem:
The septum purge on/off state and flow value were being considered as part of the method, although the 7820 does not allow these items to be changed by software. The flow value for the septum purge (and total flow, which has the purge as a component), obtained by uploading the method is thus an actual and not a setpoint.
Temporary Solution:
n/a
Fix Information:
Fixed in A.02.05
Keyword: 78xx Driver
One-line Description:
Oven Slow fan will be off after method is downloaded the slow oven mode on to 7890
Problem:
OL CDS EZChrom A,04.05 and 7890B 02.03.1 step1. turn on Oven Slow fan, step2. click "Download Method to GC". step3. 7890B turned Oven Slow fan on correctly. then "Method Download Complete" message appeared on OL EZChrom. step4. instrument set up screen refreshed. step5. Oven Slow fan check box turned off.
Temporary Solution:
n/a
Fix Information:
Fixed in A.03.01
Keyword: 78xx Driver
One-line Description:
During a pulsed splitless run, the purge time shorter than the pulse time cannot be entered.
Problem:
On a 7890B one of the modes of operation is pulsed splitless. In some cases the purge time needs to be shorter than the pulse time. Recent changes in GC driver has invalidated this parameter change. Many customers need to turn the purge time shorter than the pulse time.
Temporary Solution:
n/a
Fix Information:
Fixed in A.03.01
Keyword: 78xx Driver
One-line Description:
Method print out with 7890 configured does not show total GCc run time.
Problem:
With 7890 configured; printing the method does not show total GC run time, however if 6890 configured it does show total run time.
Temporary Solution:
n/a
Fix Information:
Fixed in A.03.01
Keyword: 78xx Driver
One-line Description:
There is no "Inlet Septum Purge" event in Runtime Events table - EZChrom
Problem:
Method setting; On OpenLAB EZChrom Online session, Click dropdown list on Runtime events table, there is no "Inlet Septum Purge" event in this list.
Temporary Solution:
n/a
Fix Information:
Fixed in A.03.01
Keyword: 78xx Driver
One-line Description:
External Device "Not Ready" ignored by OpenLAB CDS ChemSataion and EZChrom
Problem:
When doing a valve injection with a Multi-Position valve and an external device the system ignores the external device readiness. When the external device is "Not Ready" the sequence will continue to make the injection and ignore the APG readiness. If a sequence is initiated locally at the front panel of the 7890 with the external device being "Not Ready" the sequence will wait for the external device to become ready prior to making the injection and starting the run. (This is the correct behavior.)
Temporary Solution:
n/a
Fix Information:
Fixed in A.03.01
Keyword: 78xx Driver
One-line Description:
Aux Flow Limited to 150 mL/min even when pressure is less than Maximum
Problem:
Aux Flow Limited to 150 mL/min even when pressure is less than Maximum
Temporary Solution:
n/a
Fix Information:
Fixed in A.02.05
Keyword: 78xx Driver
One-line Description:
Post Run Flow Rate has default that out of range and value seems not to be stored
Problem:
Post Run gas flow defaults to 250 mL min even though the highest allowed value is 150 mL min. Even if you change it to save the analysis method as soon as you re-open the method it defaults back to 250 mL min.
Temporary Solution:
n/a
Fix Information:
Fixed in A.03.01
Keyword: 78xx Driver
One-line Description:
Method User Interface does not correctly load the content of a method.
Problem:
The content of a loaded method is only partly reflected in the method user interface. I seems that only the visible tab is updates. Steps to reproduce: 1.) Create a new method 2.) Change Oven Ramp, Events table or Signals 3.) Go back to the ALS Tab and save the method 4.) Create new method or load a different one 5.) Load the method saved in step 3 Verify the content of Oven Ramp, Events table or Signals. It does not display the correct content. See also attached document. This issue was found during the integration in Waters UNIFI and could also be reproduced in ICF TestApp and Shimadzu LabSolutions.
Temporary Solution:
n/a
Fix Information:
Fixed in A.03.01
Keyword: 78xx Driver
One-line Description:
Sample Prep Audit Trail is not logging changes.
Problem:
Create and save a Easy Sample prep method, reopen and make same changes and save it. Go to Audit Trail and the Description says: Saved method - no changes. Expected behavior: Changes made in Easy Sample prep method should be recorded in the Audit Trail.
Temporary Solution:
Easy Sample Prep does not currently support the audit trail.
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
Sample Prep steps of previous method are included in printout of methods saved from SP methods, if checked but no steps are specifed.
Problem:
Create a resource layout and simple method with Add Sample Prep steps. Run this method. Remove the steps but leave the 'Use Sample Prep Program' checkbox checked. Save As a new method. Print method. The Sample Prep steps of the original method are included.
Temporary Solution:
n/a
Fix Information:
Fixed in A.03.02
Keyword: 78xx Driver
One-line Description:
Does not delete previous open GC method event table when load method. Wrong injections can be created.
Problem:
Steps: Open instrument setup panel Create method with some lines into the event table. Save method Create a second method with empty event table. Click detectors tab to see detectors info. Save this method. Open the first method, detectors tab is shown. Go to events table and it is blank.
Temporary Solution:
n/a
Fix Information:
Fixed in A.03.02
Keyword: 78xx Driver
One-line Description:
Method download to GC while in sequence causes it to disconnect and sequence stops.
Problem:
When a method is downloaded to the GC while running a sequence the GC disconnects and the sequence stops.
Temporary Solution:
n/a
Fix Information:
Fixed in A.03.02
Keyword: 78xx Driver
One-line Description:
7890 method print doesn't include post run time
Problem:
When printing a method the GC setpoints section does not include the post run time as expected.
Temporary Solution:
n/a
Fix Information:
Fixed in A.03.02
Keyword: 78xx Driver
One-line Description:
Elite: Overlap injection no longer working on EZChrom Elite with GC Drivers A.03.02
Problem:
Overlap injection no longer working on EZChrom Elite with GC Drivers A.03.02. EZChrom Elite with the latest GC Drivers does not support this feature.
Temporary Solution:
No fix planned.
Fix Information:
None
Keyword: 78xx Driver
One-line Description:
Crash of Software after check "Show Threshold Value as remaining time" in EMF Options
Problem:
Once you check the "show threshold value as remaining time" in EMF options, and close the EMF dialog. Once you re-enter the dialog, software freezes and has to be closed by task manager.
Temporary Solution:
n/a
Fix Information:
Fixed in A.03.02
Keyword: 78xx Driver
One-line Description:
Not allowed to set proper GC column max temp because it is less than 150degrees.
Problem:
Problem when a column has the following restrictions: The maximum allowable operating temperature Isothermal: 130 °C The maximum allowable operating temperature Programmed: 150 °C When they want to enter this number in this configuration the field becomes yellow. This value can can't be entered in the field due to value limitations of related field. Maximum value should be in the range of 150 ≤ °C ≤ 450.
Temporary Solution:
n/a
Fix Information:
Fixed in B.01.01
Keyword: 78xx Driver
One-line Description:
FID signal subtraction setting changes spontaneously from "Nothing" to "Column Compensation" during run.
Problem:
When customer set FID - > Subtraction from signal value to "Nothing". Save method and upload method to instrument. Function is automatically changed to option "Column compensation run. During measure, there is activity log settings that the settings for method was changed in Subtract from signal function.
Temporary Solution:
n/a
Fix Information:
Fixed in B.01.01
Keyword: 78xx Driver
One-line Description:
"Extend Run" in GC settings does not work correctly
Problem:
Setting the run time to extend the run for a length of time is not handled correctly. The length of time is longer than the setpoint.
Temporary Solution:
n/a
Fix Information:
Fixed in B.01.01
Keyword: 78xx Driver
One-line Description:
Cannot configure a 7890B on an existing client server system
Problem:
Error in configuration panel "There are unconfigured modules"
Temporary Solution:
n/a
Fix Information:
Fixed in B.01.01
Keyword: 78xx Driver
One-line Description:
Copy to Clipboard Function in Instrument Actuals throws exception error
Problem:
Customers have reported encountering an exception when using the "Copy to Clipboard" function. Intermittent errors when clicking the ‘copy to clipboard’ button in the Status UI on the 7890 – which gives an exception error.
Temporary Solution:
n/a
Fix Information:
Fixed in B.01.01
Keyword: 78xx Driver
One-line Description:
The column flow calculator calculation is off by almost 3x the flow rate for the system
Problem:
The values which are setup for the column and the flow rates which were entered. The problem is when you look at the GC for the values they don't match the column flow calculator.
Temporary Solution:
n/a
Fix Information:
Fixed in B.01.01
Keyword: 78xx Driver
One-line Description:
Sample Overlap settings can not be edited after driver update, but are still part of the method.
Problem:
With OpenLAB EZChrom A.04.07 SR1 Update2 and GC driver version A.02.05.021 (from OpenLAB installation media) on a 7890A+ with 7693 sampler under ALS/ [Tray/Other] Sample Overlap parameters can be configured and saved to the method. Those settings are still active after upgrading to GC Drivers A.03.02.019 in the method and can cause that 2 methods behave in terms of sample overlap totally different. But they can no more been edited with the new driver.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
Incomplete Method Parameters for xCD in Acquisition Method Report
Problem:
When we do an Acquisition report, not all the xCD parameters are stored. The Reaction Cell Pressure and the Flow through the Ozone generator should also be reported. Can be confusing to customer.
Temporary Solution:
None
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
Actual flows displayed by CDS aren't the same as the setpoint flows
Problem:
Examples: Setpoint Actual 3.000 3 2.998 3 2.95 3 2.92 2.9 The display for the 7890 shows three digits to the right of the decimal for both the setpoint and the actual. As shown above, the data system sometimes shows one digit to the right of the decimal and sometimes nothing to the right of the decimal (e.g. just "3")
Temporary Solution:
none
Fix Information:
Resolved in revision B.01.02
Keyword: 78xx Driver
One-line Description:
Print method does not include all MMI setpoints: Purge time, Post Run Temp, Time, Flow
Problem:
Print method for the Intuvo 9000 GC doesn't have any MMI Purge time or Post Run temp or Post Run total Flow or time.
Temporary Solution:
n/a
Fix Information:
Resolved in revision B.01.02
Keyword: 78xx Driver
One-line Description:
Extend run does not extend the acquisition of the Intuvo 9000 GC
Problem:
Intuvo 9000 GC on C.01.07.255 SR2. Extend run does not extend the run. LUI shows Idle during the extended part of the run.
Temporary Solution:
none
Fix Information:
Resolved in revision B.01.02
Keyword: 78xx Driver
One-line Description:
Cannot Add User-Defined EMFs on Intuvo 9000 GC
Problem:
Unable to created a user defined EMF Intuvo 9000 GC. Enter a name and tracking type, but nothing appears on the User-defined panel or any other panel. There are five user defined (seemingly blank) entries on that panel.
Temporary Solution:
none
Fix Information:
To Be fixed in B.01.02
Keyword: 78xx Driver
One-line Description:
ChemStation Instrument actuals showing D2 Bus Temperature with no D2 bus installed
Problem:
ChemStation Instrument actuals showing Second Detector Bus Temperature with no second Detector.
Temporary Solution:
n/a
Fix Information:
Resolved in revision B.01.02
Keyword: 78xx Driver
One-line Description:
Method print is showing Solvent wash mode as A-A2, B-B2 when it should be A,B since it is a 16 vial turret
Problem:
Intuvo 9000 with a ALS and a 16 vial turret. MassHunter software, the configuration of the ALS is set to Solvent Wash Mode A,B. Saved and printed the method. Solvent Was Mode on the method print shows A-A2, B-B2.
Temporary Solution:
n/a
Fix Information:
Resolved in revision B.01.02
Keyword: 78xx Driver
One-line Description:
Inlet pressure setpoint displayed assuming all thermal zones are turned ON, results in confusing mismatch when heated zones are Off.
Problem:
CDS displays inlet pressure setpoint assuming all thermal zones are turned ON, results in confusing mismatch when heated zones are Off.
Temporary Solution:
n/a
Fix Information:
Resolved in revision B.01.02
Keyword: 78xx Driver
One-line Description:
Method Editor for PP inlet & Packed Column incorrectly allows both flow and pressure setpoints to be entered
Problem:
When in flow mode, pressure should not be allowed. Entering a value messes up the flow setpoint. When in pressure mode, flow should not be allowed. Entering a value messes up the pressure setpoint.
Temporary Solution:
n/a
Fix Information:
Resolved in revision B.01.02
Keyword: 78xx Driver
One-line Description:
ChemStation can lock up when changing Intuvo Bus Flow Chip types with ChemStation session running.
Problem:
ChemStation can lock up when changing Intuvo Bus Flow Chip types with ChemStation session running. If replacing the flow chip with the same type then the problem will not occur.
Temporary Solution:
Workaround for now is to disconnect ChemStation, then turn off GC, do work, turn on GC, then connect ChemStation.
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
Right Clicking on Method Editor makes UI hide behind Instrument Screen
Problem:
Right Clicking on Method Editor makes UI hide behind Instrument Screen
Temporary Solution:
n/a
Fix Information:
Resolved in revision B.01.02
Keyword: 78xx Driver
One-line Description:
Incorrect Values in 9000 Pressure/Flow Calculator
Problem:
The values within the pressure/flow calculator when initializing from "Inlet 1" still do not match the instrument parameters.
Temporary Solution:
n/a
Fix Information:
Resolved in revision B.01.02
Keyword: 78xx Driver
One-line Description:
Make changes to guard chip on Intuvo 9000 GC, audit trail just says temperature changed
Problem:
With Audit trail enabled, I make changes to the GC parameters panel of the Intuvo 9000 GC. Changed the guard chip temperature and hold time and saved the method. Looked at the audit trail and it showed events but it just said Temperature setpoint changed and Temperature hold time changed. There was no reference to the guard chip.
Temporary Solution:
n/a
Fix Information:
Resolved in revision B.01.02
Keyword: 78xx Driver
One-line Description:
Backflush Wizard illogically sets trial backflush times after the elution time of last peak of interest
Problem:
Backflush Wizard illogically sets trial backflush times after the elution time of last peak of interest
Temporary Solution:
n/a
Fix Information:
Resolved in revision B.01.02
Keyword: 78xx Driver
One-line Description:
Should not be able to download a method while a in a GC run or sequence.
Problem:
Downloading a method while a in a GC run or sequence can cause issues.
Temporary Solution:
n/a
Fix Information:
Resolved in revision B.01.02
Keyword: 78xx Driver
One-line Description:
GC Dashboard does not get populated for ChemStation
Problem:
GC Dashboard does not get populated for ChemStation
Temporary Solution:
n/a
Fix Information:
Resolved in B.01.02.059
Keyword: 78xx Driver
One-line Description:
Cannot download GC set points to 7820 configured with a Nickel Catalyst heater and no inlets.
Problem:
Cannot download GC set points to 7820 configured with a Nickel Catalyst heater and no inlets. Without the Ni catalyst configured everything works fine however once the Ni catalyst is configured the ChemStation fails to download parameters to the instrument.
Temporary Solution:
n/a
Fix Information:
Resolved in revision B.01.02.059
Keyword: 78xx Driver
One-line Description:
No indication that the AutoSampler tray is in park mode.
Problem:
From the data system there isn't a way to determine that the AutoSampler tray is in park mode.
Temporary Solution:
n/a
Fix Information:
Resolved in B.01.03
Keyword: 78xx Driver
One-line Description:
The Local User Interface(LUI) on the Intuvo 9000 GC shows long name for connected data system.
Problem:
The front panel LUI shows the PC name with other information appended to it. Customers would like to see just the PC name.
Temporary Solution:
n/a
Fix Information:
Fixed in B.01.03
Keyword: 78xx Driver
One-line Description:
For both the 7890 and Intuvo 9000 there is no change in the UI to show the user that a rack is not present.
Problem:
For both the 7890 and Intuvo 9000 there is no change in the UI to show the user that a rack is not present.
Temporary Solution:
n/a
Fix Information:
Resolved in B.01.03
Keyword: 78xx Driver
One-line Description:
Post Run Backflush. The Guard Chip temperature was not set to the Post Run Oven Temperature.
Problem:
Using an Intuvo 9000 with Post Column Restrictor. The Backflush Wizard to set the Post Run Backflush. The Guard Chip temperature was not set to the Post Run Oven Temperature.
Temporary Solution:
n/a
Fix Information:
Resolved in B.01.03
Keyword: 78xx Driver
One-line Description:
Location of Intuvo 9000 GC System Health Report should be moved
Problem:
Location of Intuvo 9000 GC System Health Report is not consistent with the location on the LUI of the GC.
Temporary Solution:
The System Health Report location has been moved to the Maintenance/Diagnostics screen.
Fix Information:
Resolved in B.01.03
Keyword: 78xx Driver
One-line Description:
Link to Intuvo 9000 GC help does not work.
Problem:
Click on options > Show 9000 Help and Information. This gives a message that the 9000 cannot be found.
Temporary Solution:
n/a
Fix Information:
Resolved in B.01.03
Keyword: 78xx Driver
One-line Description:
Event drop downs are black when creating a new method
Problem:
When creating a new GCMS method and adding events, drop downs are black. If the screen is switched and you go back to events, it works.
Temporary Solution:
n/a
Fix Information:
Resolved in B.01.03
Keyword: 78xx Driver
One-line Description:
Cannot install B.01.03 on ChemStation C.01.07 SR4 by Welcome.html
Problem:
Using Welcome.html, GC Driver B.01.03 is not installed on ChemStation C.01.07 SR4 Procedure: 0. Uninstall previous GC Drivers. 1. Open Welcome.html by IE11. 2. Select Language (Japanese) and move to INSTALL GC DRIVERS tab. 3. Click on the hyperlink "OpenLAB CDS ChemStation Edition (C.01.07, C.01.06, C.01.05" 4. Select "Run" Expected Result: InstallShield wizard should start Actual result: A error Message is displayed. "To install this software, ChemStation must be installed on the system in advance."
Temporary Solution:
Workaround: Install the driver from OpenLAB Additional Software and Drivers.
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
PCM will not control in constant flow.
Problem:
PCM will not control in constant flow. Configured two packed columns using PCM Module to a single detector. The PCM will control in pressure mode but not flow mode. This was reported by a customer using OpenLAB 2.2 and tested using OpenLAB CDS ChemStation. .
Temporary Solution:
Use pressure control mode and ramp the pressure.
Fix Information:
Resolved in B.01.04
Keyword: 78xx Driver
One-line Description:
RTL for OpenLAB CDS 2.X does not calculate the pressure correctly when locking or re-locking the method when setting a Target retention time.
Problem:
RTL for OpenLAB CDS 2.X does not calculate the pressure correctly when locking or re-locking the method. This only occurs when the Target Retention field is set to a different retention time than the peak in the chromatogram.
Temporary Solution:
The work-around is to not update target retention time when selecting the peak for RTL locking. Use the retention time of the peak selection and the calculations are correct.
Fix Information:
Resolved in B.01.05
Keyword: 78xx Driver
One-line Description:
HS Sleep/Wake mode should sync with GC Sleep/Wake mode
Problem:
When a 7697 is configured to a 7890B, it should communicate such that the Sleep or Wake state of either one should be mimicked on the other.
Temporary Solution:
n/a
Fix Information:
Resolved in B.01.05
Keyword: 78xx Driver
One-line Description:
Script error when starting help for "Runtime Events"
Problem:
Script error when starting help for "Runtime Events"
Temporary Solution:
n/a
Fix Information:
Resolved in B.01.05
Keyword: 78xx Driver
One-line Description:
Overlap injections on the 9000 uses only the injection volume from the first injection for subsequent injection.
Problem:
Overlap injections on the 9000 uses only the injection volume from the first injection for subsequent injection.
Temporary Solution:
n/a
Fix Information:
Resolved in B.01.05
Keyword: 78xx Driver
One-line Description:
GC Method Resolution results in “Object Reference” error message and occasional crash
Problem:
GC Method Resolution results in “Object Reference” error message and occasional crash
Temporary Solution:
n/a
Fix Information:
Resolved in 3.0
Keyword: 78xx Driver
One-line Description:
Downloading method parameters (like oven temperature) on a 6890A does not work with OpenLab CDS 2.4
Problem:
Downloading method parameters (like oven temperature) on a 6890 does not work with OpenLab CDS 2.4. The GC is connected (current values are correctly displayed) and setting these parameters from the keyboard does work. Settings are updated in the method when selecting "Upload method".
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: Intuvo trigger two instrument notification for configuration changed
Problem:
Chromeleon: Intuvo trigger two instrument notification for configuration changed. Configure an Intuvo with guard chip feature and open the online method in context menu of status panel and get two config changed events with a timely distance of 30 seconds.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: GC: Change of column length (and related parameters like flow etc) via calibration in online method sometimes reported in audit trail sometimes not
Problem:
Chromeleon: GC: Change of column length (and related parameters like flow etc) via calibration in online method sometimes reported in audit trail sometimes not
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: GC: Column length can only be calibrated once in Online Method Editor
Problem:
Chromeleon: GC: Column length can only be calibrated once in Online Method Editor
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: Setting Liner on Intuvo triggers two config change messages in audit trail
Problem:
Chromeleon: Setting Liner on Intuvo triggers two config change messages in audit trail
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
8860 Open Lab CDS EZChrom compact license is invalid.
Problem:
8860 Open Lab CDS EZChrom compact license is invalid.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Selecting Partial Data for the 6890 causes the retention times and areas to be wrong.
Problem:
Selecting Partial Data for the 6890 causes the retention times and areas to be wrong.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
6890 - cannot set post run flow for packed column
Problem:
6890 - cannot set post run flow for packed column
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: GC 6850A and 6890A: GC online method is not editable
Problem:
Chromeleon: GC 6850A and 6890A: GC online method is not editable
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: GC 6890A: Wrong temperature shown in chromatogram for signal GC_OvTempe_Act
Problem:
Chromeleon: GC 6890A: Wrong temperature shown in chromatogram for signal GC_OvTempe_Act
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: GC 6850A: Cannot create instrument method
Problem:
Chromeleon: GC 6850A: Cannot create instrument method
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: GC8890: Current Signal Value in 'Blank evaluation' shows incorrect value in Method editor
Problem:
Chromeleon: GC8890: Current Signal Value in 'Blank evaluation' shows incorrect value in Method editor
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: GC8890: Blank evaluation setpoints missing in method report
Problem:
Chromeleon: GC8890: Blank evaluation setpoints missing in method report
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Signal Scaling is incorrect for the 6890
Problem:
The real time plot scaling is incorrect. The offset shown in the GC Status is 10.2. The Real time plot shows the signal level to be about 7.8 x10^4. It appears that the signal is not being scaled correctly, but is using the signal directly.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: Intuvo: Help button in window 'Configure Agilent Intuvo 9000 GC' opens 78xx Series GC Help
Problem:
Chromeleon: Intuvo: Help button in window 'Configure Agilent Intuvo 9000 GC' opens 78xx Series GC Help
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: GC8890: Issues during sequence execution
Problem:
Chromeleon: GC8890: Issues during sequence execution. During the execution of a sequence (using the blank evaluation) I observed some issues: 1.no audit trail entry if blank evaluation was used or not ( was performed as blank run, blank evaluation option enabled) 2.during execution of sequence the GC8890 reported high CPU usage 3.The blank evaluation failure was reported during : 16.04.2019 17:20:29 +02:00 2,059 Agilent Blank Evaluation Failed, Failure Action is Continue
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: Changes made in 68xx Edit Online Method not applied or reported
Problem:
Chromeleon: Changes made in 68xx Edit Online Method not applied or reported. 1. With Chromeleon open, selected to open Instrument Actuals 2. Right click in window and select Edit Online Method 3. Attempted several changes to online method. 4. Clicked Set Method Button. 5. No change in instrument actuals reflecting changes to method 6. No audit trail entries for changes made.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Error when loading GC method without Blank Run Analysis setpoint
Problem:
Error when loading GC method without Blank Run Analysis setpoint
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: Corrected column length will not save after manual method resolution
Problem:
Chromeleon: Corrected column length will not save after manual method resolution
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: Signals from previous method persist in new method
Problem:
Chromeleon: Signals from previous method persist in new method
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: ECD Help page cannot be accessed
Problem:
Chromeleon: ECD Help page cannot be accessed
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: All 8860 signals not displayed or available in report
Problem:
Chromeleon: All 8860 signals not displayed or available in report.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: Error occurred at several points of operation
Problem:
Chromeleon: Error occurred at several points of operation. 1. Open instrument, select instrument (8890) and choose Create Method 2. Error pops up (see pic) 3. Click OK and proceed setting method parameters 4. Finish method; new instrument method opens. 5. Error appears again 6. Click OK, check method and save. Method saves 7. Right click on Status window and select Edit Online Method 8. Error appears again 9. Click OK and proceed with method edit 10. No time in this instance did the error impede actual saving method or running a sequence Expected behavior: Method creation and editing proceed without error
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: GC6890: Front Injector parameter change in Online method not applied
Problem:
Chromeleon: GC6890: Front Injector parameter change in Online method not applied.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: GC6890A: Method download fails
Problem:
Chromeleon: GC6890A: Method download fails
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
8860 on Chromeleon -Unhandled exception in Chromeleon Driver Host (CmDDKHost): AgilentTechnologies.DdkIcfAdapter|Agilent
Problem:
8860 on Chromeleon -Unhandled exception in Chromeleon Driver Host (CmDDKHost): AgilentTechnologies.DdkIcfAdapter|Agilent
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon does not recognize the tray with Back Injector installed.
Problem:
Connected to a Back inlet only 8890 with a Back Injector and Tray. Was not able to use a vial position on the tray as the vial number in the sequence for Chromeleon.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: Automatic de-activation of not supported feature blank evaluation cannot be saved
Problem:
Chromeleon: Automatic de-activation of not supported feature blank evaluation cannot be saved
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Getting multiple Errors in the PC Event Log for AgGC7890 after disconnecting and Reconnecting LAN cable.
Problem:
Getting multiple Errors in the PC Event Log for AgGC7890 after disconnecting and Reconnecting LAN cable.
Temporary Solution:
n/a
Fix Information:
Unable to duplicate the issue.
Keyword: 78xx Driver
One-line Description:
GC6890: Valid vial ranges are not correct for 6890 in some cases
Problem:
GC6890: Valid vial ranges are not correct for 6890 in some cases
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: GC6890A: Error 'The communication FIFO of the DDK driver is corrupted.' after connecting and disconnecting several times.
Problem:
Chromeleon: GC6890A: Error 'The communication FIFO of the DDK driver is corrupted.' after connecting and disconnecting several times.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
When click the "Apply the Option" in GC Chromatographic Column Directory under Column Configuration, an exception occur on Openlab 2.4
Problem:
When click the "Apply the Option" in GC Chromatographic Column Directory under Column Configuration, an exception occur on Openlab 2.4.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Upgrade installation from GC B.01.03 to 3.0 SR2 fails to install Ag7697WS.dll (and .pdb)
Problem:
Upgrade installation from GC B.01.03 to 3.0 SR2 fails to install Ag7697WS.dll (and .pdb)
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: GC6850A: Changing method parameter creates method inconsistent to configuration
Problem:
Chromeleon: GC6850A: Changing method parameter creates method inconsistent to configuration
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: GC6850: Partial Data option causes issues during sequence execution
Problem:
Chromeleon: GC6850: Partial Data option causes issues during sequence execution
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: GC6890: Unexpected audit trail entries during endurance run
Problem:
Chromeleon: GC6890: Unexpected audit trail entries during endurance run
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
SR1 for the GC3.0 drivers breaks the connection between an 8890 and 7000D transferline temperature reading in MH ACQ 10.
Problem:
SR1 for the GC3.0 drivers breaks the connection between an 8890 and 7000D transferline temperature reading in MH ACQ 10.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Not able to download flow setpoints to 6890 VI inlet.
Problem:
Not able to download flow setpoints to 6890 VI inlet. Set up a method at the 6890. The column flow as set to 3 ml/min and the split ratio to 20 giving a total flow of 66.26 ml/min. Changed the split ratio to 10 and downloaded the method. The total flow did not change.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: Incorrect conversion and missing unit when using pressure setting bar
Problem:
Chromeleon: Incorrect conversion and missing unit when using pressure setting bar
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
68xx: Chromeleon: System with G4513 won't make 1.1 uL injection
Problem:
68xx: Chromeleon: System with G4513 won't make 1.1 uL injection
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: System crash: system slow and then crashing
Problem:
Chromeleon: System crash: system slow and then crashing. Chromeleon was unreactive and then suddenly the instrument connection was lost. After restarting the instrument controller the instrument was online again. Audit trail showns an OutOfMemory exception.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
8860 with Headspace/PAL does not consume correct VL license for OpenLAB 2.4
Problem:
8860 with Headspace/PAL does not consume correct VL license for OpenLAB 2.4
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
B,04.03 SP2 will not connect to an 8890 GC.
Problem:
B,04.03 SP2 will not connect to an 8890 GC.There is no way to configure an 8890 and if you try to connect as a 7890 it will not start the instrument session.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: 6890GC Oven temperature disabled in method - Run does not start.
Problem:
Chromeleon: 6890GC Oven temperature disabled in method - Run does not start.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
3.0 SR2 driver fails SVT on OL CS C.01.09.144
Problem:
3.0 SR2 driver fails SVT on OL CS C.01.09.144
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
OLCS C.01.09 with Driver 3.0.650 will not launch any instrument.
Problem:
OLCS C.01.09 with Driver 3.0.650 will not launch any instrument.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Chromeleon: SVT Chameleon 2.1.30 Fails
Problem:
Chromeleon: SVT Chameleon 2.1.30 Fails
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Function_build number not display in SVT report in EzChrom and OpenLab CDS 2.4
Problem:
Function_build number not display in SVT report in EzChrom and OpenLab CDS 2.4
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
8890: the MS transfer line temp displayed is 0 C but its actual temp is 250 C.
Problem:
8890: the MS transfer line temp displayed is 0 C but its actual temp is 250 C.
Temporary Solution:
n/a
Fix Information:
Resolved in version 3.0.652
Keyword: 78xx Driver
One-line Description:
Calculated Pressure for Relock is not updated for the 6890 in C.01.10.
Problem:
Calculated Pressure for Relock is not updated for the 6890 in C.01.10.
Temporary Solution:
Need to update manually.
Fix Information:
n/a
Keyword: 78xx Driver
One-line Description:
7890A: Shutdown events are not published to the software GC Status screen
Problem:
The thermal inlet shutdown is not reported in the software “Agilent GC Status” screen. Only the first event, which is 'not ready', is reported to the Agilent GC Status screen. The second event, which is the shutdown, is not displayed.
Temporary Solution:
n/a
Fix Information:
Fixed in driver revision 4.01. 7890A requires a minimum Firmware revision of A.01.12. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Keyword: 78xx Driver , 68xx Driver
One-line Description:
GC instrument Help page cannot be displayed
Problem:
Access to GC instrument HELP topics only available from the GC user interface
Temporary Solution:
GC help topics can be reached directly from the GC user interface
Fix Information:
n/a
Keyword: ACQ
One-line Description:
No special characters should be used for names of methods, data files or sequences
Problem:
Names for data files, methods and sequences should only contain alphanumeric characters. Special characters like ",#,$, %, &, @, ? and others should not be used and can in general cause trouble. The same is true for trailing blanks. While this is outlined in the written documentation several times, it is not enforced for all dialogs inside the software. There are still the occasional loop hole where such a character can be entered or imported.
Temporary Solution:
The possiblity of entering special characters does by no means imply that it is safe to do it.
Fix Information:
N/A
Keyword: ACQ
One-line Description:
Terminal Service can close under high workload
Problem:
It has been observed that the terminal service on an AIC can terminate under high workload. This is a limitation of Windows 2008 server. Since this only affects the remote session control to the ChemStation Client, no data will be lost.
Temporary Solution:
No fix available.
Fix Information:
Will not be fixed for Windows 2008.
Keyword: ACQ
One-line Description:
Queue Method vial range reflects range of loaded method, not the method browsed to
Problem:
Load a method in ChemStation. Select Queue Method. Within that dialog, Browse... to load another method with a different vial range. The vial range (Sample Parameters: Vial/Location) is not updated to reflect the range of the method loaded in Queue Method, though the Signal (1 and/or 2) Data File info is correctly loaded from the browsed to method.
Temporary Solution:
n/a
Fix Information:
Resolved in C.01.07 SR2
Keyword: ACQ
One-line Description:
7890 back injector only data file is missing acq.txt and acq.macaml in the data directory
Problem:
There is no acq.txt and acq.macml in the data directory for the 7890 back injection only data file.
Temporary Solution:
n/a
Fix Information:
OpenLAB CDS intelligent reporting of GC 7890 acquisition methods was introdcued with GC driver version 4.02.043 and OpenLAB CDS ChemStation C.01.04. OpenLAB CDS ChemStation C.01.03 and earlier and GC 7890 driver 4.02.023 and earlier do not allow creating Intelligent Reporter templates including GC7890 settings. Fixed in OpenLAB CDS C.01.04 which includes 78xx Driver 4.02 [043]. The 78xx driver 4.02 [043] supports the XML interfaces for reporting the method.
Keyword: ACQ
One-line Description:
Run time for unsynchronized modules is shortened
Problem:
The runtimes of modules on a CAN bus (1260, 7100, integrated LCs) are handled internally to synchronize start and stop. The same applies to devices attached to the remote bus (e.g. 35900 or MSD). Devices that are not synchronized might be either stopped too early which leads to data loss, or too late which causes time-out errors. This affects ELSD detectors, SFC modules, modules with GCI drivers or 3rd party modules with RC.Net drivers, e.g Waters Alliance modules.
Temporary Solution:
Make sure the runtimes of all modules producing data are identical and no other module has a longer runtime.
Fix Information:
Fixed in C.01.05 and C.01.04 HF04 by introducing a master runtime for all modules.
Keyword: ACQ
One-line Description:
Memory leak in RDPShell.exe may cause significant performance decrease on OpenLAB CDS ChemStation AICs
Problem:
The following problem has been reported for OpenLAB CDS ChemStation C.01.0x Acquisition Instrument Controllers (AIC) operated on Windows 2008 R2 SP1 operating systems. Windows 7 or Windows XP operating systems are not affected. A memory leak in RDPShell.exe may cause significant performance decrease as the overall memory usage may exceed 80% after several weeks of operation. This unintended and permanent memory allocation is triggered by moving the ChemStation application window over the client desktop. This problem is acknowledged by Microsoft (http://support.microsoft.com/kb/2636613/en-us) and hot fix is available on request. NOTE: The hot fix is not deployed as part of the automated update procedure.
Temporary Solution:
Unpatched AIC machines should be rebooted every 2nd week to release the allocated memory. Note: ChemStation instances need to be shutdown (rather than disconnected) before rebooting the AIC. Use the remote application window in full screen mode only and avoid moving the application window on the client screen. Shutting down the ChemStation instances does not release the memory on the Windows 2008 server as the RDP sessions remain in an idle state untill they are re-used.
Fix Information:
Microsoft offers a hot fix that is not autmatically distributed by the update process. The patch needs to be explicetly requested from http://support.microsoft.com/kb/2636613/en-us. Please follow the back-up and installation recommendations from Microsoft. Agilent performed limited tests with the mentioned hot fix and no negative impact was found.
Keyword: ACQ
One-line Description:
GC7890 acquisiton methods cannot be reported with Intelligent Reporting of OpenLAB CDS ChemStation
Problem:
OpenLAB CDS intelligent reporting of GC 7890 acquisition methods got introdcued with GC driver version 4.02.043 and OpenLAB CDS ChemStation C.01.04. OpenLAB CDS ChemStation C.01.03 and earlier and GC 7890 driver 4.02.023 and earlier do not allow creating Intelligent Reporter templates including GC7890 settings.
Temporary Solution:
Use the classic reporting functionality of OpenLAB CDS ChemStation to print GC7890 methods.
Fix Information:
OpenLAB CDS ChemStation C.01.03 HF08 introduces functionality to allow the 7890 GC driver 4.02.043 to store the instrument parameters to be used with intelligent reporting. Fixed in OpenLAB CDS C.01.04 which includes 78xx Driver 4.02 [043]. The 78xx driver 4.02 [043] supports the XML interfaces for reporting the method.
Keyword: ACQ
One-line Description:
Chromatograms of manually stopped runs are shorter than expected
Problem:
User sometimes stop runs as soon as the expected peaks have eluted as seen from the online plot. The run on disk might not contain those last peaks. The online plot uses a different channel which is unfiltered. Stopping a run will finalize the chromatogram immediately. When the detector has a high filter width (response time), the last data points are not yet stored to the chromatogram.
Temporary Solution:
With low data rates (and long filter widths) the online plot might be up to 30 seconds ahead of the data collection in the data file. Consider adding this time span before stopping a run.
Fix Information:
N/A
Keyword: ACQ
One-line Description:
LCDIAG.REG content incomplete
Problem:
The run contains LCDIAG.REG files which are missing vital information, like the sampler information.
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB CDS ChemStation C.01.05 HF04 and OpenLAB CDS ChemStation C.01.06.
Keyword: ACQ
One-line Description:
Data system uses the currently loaded method to validate the vial range in the Sequence table
Problem:
Sequence does not run or will abort if the currently loaded method vial range is different than the vail range of the next method in the sequence table. Set up two methods: Method E with injection source set to headspace (valid vials 1-111) Method A with injection source set to ALS sampler with tray. (valid vials 1-150) Construct a sequence Vial Method 1 A 1 E 130 A 2 E If method E is the currently loaded active method then this sequence will not validate because the first method in the sequence (method A) has a vial range that is valid for an ALS. The vial numbers are checked against method E which has a vial range for the headspace. The sequence cannot be run. If method A is the currently loaded active method, then this sequence will validate and start running. When the datasystem attempts to move to the 3rd line it validates the vial against E with a headspace injectorID and fails, aborting the sequence.
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB CDS C.01.06.
Keyword: ACQ
One-line Description:
Using FIle> Snapshot causes the software to become unresponsive on systems with a Varian Prep Star module
Problem:
Creating a file snapshot on an instrument with a Varian Prep detector can result in an unresponsive software.
Temporary Solution:
Adressed in in hotfix-09 for C.01.05
Fix Information:
Resolved in OpenLAB CDS ChemStation Edition C.01.07.
Keyword: ACQ
One-line Description:
Missing data points in chromatogram can be perceived as retention time shift.
Problem:
Under certain circumstances data points can be missing in chromatograms. This happens when the instrument is sending data, but the software has not yet acknowledged the start of the run. As a result data points are not recorded from the beginning of the chromatogram. The data points are not time stamped by the hardware. Due to the design of the data protocol they are sorted into the chromatogram starting at 0 minutes. Therefore it appears that data points are missing at the end of the chromatogram. For the user it appears as a retention time shift. The missing part of the chromatogram can range from seconds to over 10 minutes. Factors that make the problem more likely are run start from the hardware (handheld controller, front panel, manual injector, remote start). Fast network and high load on the instrument controller also increase the risk. The problem has so far been observed on FID and VWD signals.
Temporary Solution:
Fixed in OpenLAB CDS ChemStation Revision C.01.05 SP1 HF02. Fixed in Hotfix 05 for ChemStation C.01.06.
Fix Information:
This issue is fixed in OpenLAB CDS ChemStation Revision C.01.07 by redesign of the acquisition engine.
Keyword: ACQ
One-line Description:
Incorrect XML import of MultiSampler vial positions
Problem:
The vial positions of the Infinity II MultiSampler like D1F-A1 (for Drawer 1, Front, position A1) cannot be imported correctly from an XML file. The XML Import Table will show the correct Vial positions but the Import report will state "invalid Location in line". Also Sample Location field in the Sequence Table will be blank.
Temporary Solution:
N/A
Fix Information:
Fixed in C.01.07 SR1 HF1
Keyword: ACQ
One-line Description:
Duplicate or corrupted files or unresponsive software after stop/abort
Problem:
An inconsistency in the communication between instrument and software has been found during sequence runs. The defect is triggered in situations when external interaction occurs at state transitions. External interaction can mean: Stopping or aborting a run by user interaction, hardware error (Power fail, leak, communication loss), missing vial, barcode mismatch, etc. State transition can mean: at the time of injection, between runs, in situations that should be covered by rules and alerts, etc. As a result of such a conflict the same file (acaml file, method folder, logbook...) can be accessed by two instances, leading to a corrupted file or to unresponsive software.
Temporary Solution:
N/A
Fix Information:
This defect is addressed with HF03 for C.01.06.
Keyword: ACQ
One-line Description:
Order of data points in instrument signals is inverted
Problem:
On systems under heavy load, it can occur that the data points coming from the instrument are not sorted in the chronological order. This occurs typically with the instrument traces. In that case, the data points are sorted correctly on screen, but the line that connects the points appears to loop backwards. There is one known case where the issue occurred on the detector signal. In that case the connecting line was straight forward, but the order of data points was not correct. The issue might be more frequent, since a wrong order of data points will not be obvious outside of a peak.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS ChemStation C.01.05 SP1 HF02.
Keyword: ACQ
One-line Description:
ChemStation Scheduler hpsched.exe fails to start on Windows 2008 Server R2 SP1
Problem:
Upon start of HPSHED.EXE (either form ChemStation's View menu or by double-click on C:\Chem32\Core\hpshed.exe) only a plain dialog box titled "HPSHED" with this error text appears: "Run-time error '9': Subscript out of range".
Temporary Solution:
n/a
Fix Information:
Resolved in revision OpenLAB CDS ChemStation C.01.07 SR2.
Keyword: ACQ
One-line Description:
Error 17923 occurred for both LC and GC
Problem:
During acquisition, error message "Error 17923 occurred: Parameter 3: A value is outside the legal range" are displayed in instr.log
Temporary Solution:
n/a
Fix Information:
Resolved in revision OpenLAB CDS ChemStation C.01.07 SR2.
Keyword: ACQ
One-line Description:
Order of data points in instrument signals is inverted
Problem:
On systems under heavy load, it can occur that the data points coming from the instrument are not sorted in the chronological order. This occurs typically with the instrument traces. In that case, the data points are sorted correctly on screen, but the line that connects the points appears to loop backwards. There is one known case where the issue occurred on the detector signal. In that case the connecting line was straight forward, but the order of data points was not correct. The issue might be more frequent, since a wrong order of data points will not be obvious outside of a peak.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS ChemStation C.01.06 HF05.
Keyword: ACQ
One-line Description:
Sequence Start button is enabled when sequence is paused.
Problem:
When a locked online session is unlocked and the currently running sequence was paused, the Start button is enabled. Pressing it would start the same sequence a second time. The resulting conflict could make the software unresponsive or even lead to a crash.
Temporary Solution:
Adressed in Hotfix 05 for ChemStation C.01.06.
Fix Information:
Fixed in C.01.07.
Keyword: ACQ
One-line Description:
G7100A Wait time after loading new method is applied to every run
Problem:
Using the Sequence Parameter " wait x minutes after loading a new method" has been applied to every run within a sequence (e.g. multiple injections within a sequence line using the same method). This added extra time to the execution of a sequence.
Temporary Solution:
None
Fix Information:
The issue has been fixed with OpenLAB ChemStation Edition C.01.07 SR3. The wait time behavior has been corrected to be executed the same way as for other techniques. The wait time in applied once when a different method is loaded. In case that users made an advantage of the additional time added between each run (e.g. additional equilibration), either the methods post time (applied after a run is finished) or a preconditioning step (wait xx seconds) as a post run command can be used to apply the additional equilibration time.
Keyword: ACQ
One-line Description:
"Error 41092 occurred" is shown in the system activity logs
Problem:
This error intermittently happens when the sequence upload is stuck and the system must be shutdown. This is a general error that indicates that the system failed to get the correct response echoed back to it.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: ACQ
One-line Description:
Calibration run was canceled during injection, File/Path error message – could not save method with data shown
Problem:
When a calibration run is cancelled during the injection, the calibration dialog is closed automatically. A File/Path error - could not save method was shown and the run was removed from the run queue. The modules become idle again, but chemstation remained in the state: Ready/Saving method with data. Had to restart chemstation to remove the “Saving with data” state.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: ACQ
One-line Description:
ChemStation does not set the Device Master-Runtime correctly to the ELSD RC.Net driver
Problem:
If the initially loaded method has a Master-Runtime of exactly 10.0 minutes, the ChemStation does not propagate the Master-Runtime to the instrument driver. Thus, the instrument driver uses a wrong runtime for acquisition. The problem was seen with an LC stack together with an ELSD.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS ChemStation C.01.07 SR4.
Keyword: ACQ
One-line Description:
Cannot start the FIA Method in the Single Sample Information down menu with G7129A Vial sampler
Problem:
In the Sample Information form, the Vial location always has a "vial 1 location" by default. It needs to blank to able the start FIA method since the method has a P1-A1 vial location on the G7129A Vial sampler.
Temporary Solution:
Before loading the FIA method, edit the vial location field to be ‘blank’ in the Single Sample Information down menu.
Fix Information:
Fixed in C.01.10 Update 1
Keyword: ACQ
One-line Description:
Intermittent signal lost of HPLC with G7115A
Problem:
Intermittent signal loss of G7115A in a networked CDS system.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: ACQ
One-line Description:
Fractions missing in Data File after Instrument Error
Problem:
An instrument error occurs due to insufficient available locations causes an 'Abort Run' which closes the data files immediately (without collecting the remaining rawdata/fractions).
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: ACQ
One-line Description:
Instrument driver error crashes the ChemStation
Problem:
A instrument driver error can cause a crash in the ChemStation. The ChemStation should handle this situation more reliable and report valid error messages and not crash.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.07 SR4 Hotfix 1.
Keyword: ACQ
One-line Description:
Intermittently the signal is missing or saved in wrong data folder for runs not started from the ChemStation
Problem:
The instrument start is triggered by an external device based on its ready state. The instrument is not started first via "Run Control, Run Method from the ChemStation. Intermittently the signal is missing or saved in a wrong data folder for single runs that are not started via the ChemStation.
Temporary Solution:
Optional workflow: - Select Injection Source “Manual” - Before each device-run, first start a ChemStation-Run via “Run Control -> Run Method” - …ChemStation is now waiting for the instrument to start a run… - Trigger the instrument run (e.g. Start key) or via remote start With this approach it should be ensured the data path/file variables are set correctly before the instrument run starts.
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.08 Update 2. Resolved in OpenLab CDS ChemStation C.01.09.
Keyword: ACQ
One-line Description:
Intermittently the signal is missing or saved in wrong data folder for runs started at GC
Problem:
The instrument start is triggered by an external device based on its ready state. The instrument is not started first via "Run Control, Run Method from the ChemStation. Intermittently the signal is missing or saved in a wrong data folder for single runs that are not started via the ChemStation.
Temporary Solution:
Optional workflow: - Select Injection Source “Manual” - Before each device-run, first start a ChemStation-Run via “Run Control -> Run Method” - …ChemStation is now waiting for the instrument to start a run… - Trigger the instrument run (e.g. Start key) or via remote start With this approach it should be ensured the data path/file variables are set correctly before the instrument run starts.
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.07 SR4 Hotfix 1.
Keyword: ACQ
One-line Description:
Counter used for single data file naming does not increment correctly with multiple runs in C.01.08 [216]
Problem:
When running single runs from RunControl > Sample Info, the counter used for the single data file naming does not increment correctly when multiple runs are selected. Steps to reproduce it: 1. Single Injections at RunControl->Sample Info. 2. Signal1 Field = Any Field + Counter (i.e. counter =1, Length=any, Increment=1) with Number of Runs = more than 1 3. Result: If Run number is set as 3 and length=4, first injection data name of counter become 0003.D and second injection data become 0001.D. Also, the third injection naming becomes 0001.D again with notice for data overwriting.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.08 Update 2 for LC only. Resolved in OpenLab CDS ChemStation C.01.08 Update 3 for LC and GC. Resolved in OpenLab CDS ChemStation C.01.09 Update 1 for LC and GC. Resolved in OpenLab CDS ChemStation C.01.10 for LC and GC.
Keyword: ACQ
One-line Description:
Counter used for single data file naming does not increment correctly with multiple runs in C.01.08 [216]
Problem:
When running single runs from RunControl > Sample Info, the counter used for the single data file naming does not increment correctly when multiple runs are selected. Steps to reproduce it: 1. Single Injections at RunControl->Sample Info. 2. Signal1 Field = Any Field + Counter (i.e. counter =1, Length=any, Increment=1) with Number of Runs = more than 1 3. Result: If Run number is set as 3 and length=4, first injection data name of counter become 0003.D and second injection data become 0001.D. Also, the third injection naming becomes 0001.D again with notice for data overwriting.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.08 Update 2 for LC only. Resolved in OpenLab CDS ChemStation C.01.08 Update 3 for LC and GC. Resolved in OpenLab CDS ChemStation C.01.09 Update 1 for LC and GC. Resolved in OpenLab CDS ChemStation C.01.10 for LC and GC.
Keyword: ACQ
One-line Description:
Data saved in reverse order for multiple single runs specified in Sample Info
Problem:
Data is saved in the following format when running multiple single sample runs: Zen-006.d Zen-005.d Zen-004.d Zen-003.d Zen-002.d Zen-001.d
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: ACQ
One-line Description:
Missing error method in sequence parameters
Problem:
When Fraction Collector runs in Error during a sequence, the error method is not loaded.
Temporary Solution:
n/a
Fix Information:
Fixed in C.01.10 Update 3 and documented in help.
Keyword: ACQ
One-line Description:
Snapshot does not work: Error "Data file does not exist"
Problem:
When a subdirectory is configured in the sequence parameters or the sample info dialog, the snapshot function does not work. The error message is / Data File "C:\Users\Public\Documents\ChemStation\1\Data\snaphot.d" does not exits. /
Temporary Solution:
n/a
Fix Information:
Fixed in C.01.10 Update 4
Keyword: ACQ
One-line Description:
When doing Fraction Collection without sampler it can happen that the fraction start locations is wrong
Problem:
When doing Fraction Collection without sampler it can happen under certain configuration and unlucky timing that the fraction start locations is wrong.
Temporary Solution:
Configure the G7166A or other AFC to the first module in the configuration
Fix Information:
Fixed in C.01.10 Update 4
Keyword: ACQ
One-line Description:
Performance problems with OpenLab ChemStation C.01.07 and C.01.08 with 8890 and 8860 GC instruments
Problem:
The 8890 and 8860 GCs were released after OpenLab ChemStation C.01.07 and C.01.08 and are not properly recognized as an Agilent instrument in some column related functions. When running an unrecognized GC, every time a run is performed 6 lines are added to the columns table which causes it to grow with each injection. Delays related to reading the column table become noticeable when the column table gets too large.
Temporary Solution:
There are 2 workarounds available: 1. Delete the column table in the instrument folder, DRIVE:\Chem32\n\ for OpenLab ChemStation C.01.07, or DRIVE:\Program Data\Agilent Technologies\ChemStation\n\ for OpenLab ChemStation C.01.08 where n is the instrument number for the instrument. The column table is COLUMNSDB.CINV.REG and copies of this information may be stored in the configuration files for the online and offline sessions of the instrument: Config.Reg and Conf_Off.Reg. Make sure all instrument sessions in the ChemStation are closed. Delete COLUMNSDB.CINV.REG, Config.Reg, and Conf_Off.Reg. When the instrument sessions are launched the required files will be rebuilt. 2. Enter the following command on the ChemStation command line: if TabHdrVal(_config[1], "Columns", "NumberOfRows") > 250 then ; DelTabRow _Config[1], "Columns", 251:-1; endif ; The frequency of using either of these workarounds depends on system usage and performance.
Fix Information:
Resolved in OpenLab ChemStation C.01.08 Update 3, C.01.07 SR4 PP02, C.01.07 SR3 PP04 This issue no longer occurs beginning with: OpenLab ChemStation C.01.09
Keyword: ACQ
One-line Description:
Column table continues to grow each time a column is changed in GC and LC
Problem:
Column table continues to grow each time a column is changed in GC and LC.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab ChemStation LTS 01.11.
Keyword: ACQ
One-line Description:
Column table grows over time when changing column settings
Problem:
Every time a column setting is changed (for example, changing the name), a new row is added to the column table causing the table to grow: a) when the column does not have a serial number, or b) when the column has a serial number but not in the column table
Temporary Solution:
n/a
Fix Information:
Resolved In OpenLab ChemStation C.01.10 Update 5.
Keyword: ACQ
One-line Description:
Sample naming counter issue
Problem:
When runs are submitted to the beginning of the queue, the counter token for the first entry will have the highest value.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab ChemStation LTS 01.11.
Keyword: ACQ
One-line Description:
Commands setup in the Command Scheduler window also used in the Offline session
Problem:
If a data set is loaded in an offline session, that data set will be reprocessed in addition to the new data being collected.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.10 Update 7. Resolved in OpenLab ChemStation LTS 01.11.
Keyword: ACQ
One-line Description:
Cannot remove custom wellplate definitions
Problem:
Deleted custom wellplate definitions will reappear upon relaunching ChemStation.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: Administration
One-line Description:
Windows Domain tab is grayed out in OpenLAB Server Utility
Problem:
When the authentication provider is set to domain authentication, the user account used to browse the active directory is stored in the OpenLAB Shared Services database and can be modified or updated using the OpenLAB Server Utility. The OpenLAB Server Utility will then update the databse with the new information. When starting the OpenLAB Server Utility as a domain user which is not a database administrator, the information in the windows domain tab will be greyed out because the tool can't connect to the database.
Temporary Solution:
Login to windows as an user account which is granted database administration privileges
Fix Information:
n/a
Keyword: Administration
One-line Description:
Cannot stop Shared Serviceshost service on computer after changing the Domain Account password
Problem:
After changing the Domain Account password on the OpenLAB Server Utility, you may occasionally get an error message, “Cannot stop Shared Serviceshost service on computer”.
Temporary Solution:
As a temporary solution, restart the Agilent OpenLAB Shared Services service or reboot the shared services server.
Fix Information:
n/a
Keyword: Administration
One-line Description:
Users with privilege of view and run instrument can take over instrument used by another user
Problem:
Users with the privilege of view and run instrument can take over the online/offline session used by another user in Chinese and Japanese Systems. English systems are not affected.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS ChemStation C.01.07 SR4.
Keyword: Administration
One-line Description:
Microsoft Windows 10 update (build 1703) may result in bluescreen
Problem:
Microsoft has published a so called "creators update", a patch for Windows 10. The internal build number for this build is 1703. To find out the currently installed patch click the start button and enter "winver". Customers who haven´t deactivated those updates and have patched their C.01.07 SR3 installation will get a bluescreen when navigating to the detectors screen while editing a 7890B method. More issues may exist that we haven´t found yet.
Temporary Solution:
n/a
Fix Information:
Microsoft has provided two fixes that can be found here (install both patches): https://answers.microsoft.com/en-us/windows/forum/windows_10-performance/windows-10-creators-update-unexpected-kernel-mode/33b44144-f35b-431d-a57b-296e30fc2914 https://answers.microsoft.com/en-us/windows/forum/windows_10-performance/windows-10-creators-update-bsod-kernel-crash/b8ec01a2-4086-4048-8557-417eb79ce813 Or find it here as a cumulative update: https://support.microsoft.com/en-in/help/4022716/windows-10-update-kb4022716
Keyword: Administration
One-line Description:
Software Verification Report does not include Hotfix04 in the base revision number
Problem:
The Software Verification Report (SVT) does not show the correct base revison number for OpenLab Shared Services Server when Hotfix04 is installed. Hotfix04 is not included in the base revision number; only 2.1 is shown.
Temporary Solution:
Go to Windows Add/Remove Programs to verify that the OpenLab Shared Services Service Hotfix 4 is installed.
Fix Information:
n/a
Keyword: Administration
One-line Description:
EMF help information not available using the Instrument Control Framework
Problem:
Currently the EMF help information not available using the Instrument Control Framework
Temporary Solution:
The EMF (Early Maintenance Feedback) can only be used in conjunction with Agilent Lab Advisor. Within ICF, the EMF (Early Maintenance Feedback) status of the module is displayed by its EMF-specific color-coding. The tooltip gives further information about the EMF status. To modify EMF limits or clear EMF values, use the Agilent Lab Advisor.
Fix Information:
n/a
Keyword: Administration
One-line Description:
Do not install msi files by double-clicking the file
Problem:
Drivers and add-on software products are provided in .msi files (Microsoft Windows Installer). Double-clicking such files seems to create a valid installation. While the software may work correctly, this may create problems: - the OpenLAB Additional Software and Drivers Deployment Wizard does not register and list this software product. This information is needed for getting support from Agilent Technologies - uninstallation of the software may fail - upgrade of the software may fail
Temporary Solution:
Do not use a double-click for installation. Use instead the OpenLAB Additional Software and Drivers Deployment Wizard, which can be started by going to Windows Start > Agilent Technologies > OpenLAB > OpenLAB Additional Software and Drivers. Follow steps described in the Wizard for installation or software upgrades. If a software cannot be upgraded or uninstalled because of an invalid installation, it may help to install the software again using the procedure recommended here. Then uninstall the software using the wizard and try an upgrade again.
Fix Information:
Since LC and CE driver version A.02.18 a warning message appears which says that manual installs are not supported and the installation aborts. This only affects OpenLab CDS ChemStation Edition and OpenLab CDS EZChrom Edition. With OpenLab 2.x manual installations are supported.
Keyword: Administration
One-line Description:
Administrative Break Session lock functionality from ChemStation Admin tool is not working correctly
Problem:
In the "ChemStation Administration Tool" select the check box "Break session lock" in the login section. It is not possible to unlock ChemStation with ‘Cancel’ if Break session lock was activated in the ChemStation Administration Tool.
Temporary Solution:
n/a
Fix Information:
Resolved in Openlab CDS ChemStation C.01.07 SR4 Hot Fix 1. Resolved in Openlab CDS ChemStation C.01.08. It is now possible to unlock ChemStation with 'Cancel' if Break session lock is activated in the ChemStation Administration Tool.
Keyword: Administration
One-line Description:
Installation Qualification Test fails when a newer RapidControlIF.dll is installed than has been expected by the LC Driver
Problem:
Due to a mismatch within the Installation Qualification Test (IQT) setup of the LC Driver Installation Package, IQT may fail, if the A.02.19SR2 LC Driver package is installed. The issue occurs, when a previous installation of either the data system or another software package has installed the shared component "RapidControlIF.dll" with a more recent file version than expected by the LC Driver (Expected file version is 3.1.1.0). IMPORTANT: This issue does not affect proper function of the LC driver. It is an issue within the logic of the IQT files as it is being processed after the installation of an LC Driver package.
Temporary Solution:
n/a
Fix Information:
This IQT issue can not be fixed for LC Driver versions below A.02.19 SR3, if it already has occurred once. That is, if IQT has failed once for the reasons mentioned in the description. The issue can be PREVENTED, by directly installing A.02.19 SR3 onto a fresh installation of Chemstation C.01.09.
Keyword: Administration
One-line Description:
Installation Qualification Test fails when A.02.19 SR2 is installed on OpenLab Chemstation Edition C.01.09
Problem:
When trying to update the LC Drivers on an existing installation of OpenLab Chemstation Edition C01.09, Installation Qualification Test (IQT) fails with an unexpected file version of a RapidControl component.
Temporary Solution:
None available.
Fix Information:
n/a
Keyword: Administration
One-line Description:
Driver release note states wrong firmware revision for G7104C
Problem:
The LC and CE Driver release note wrongly mentions "D.07.20" as the minimum required firmware revision. The correct minimum required firmware revision is "B.07.20". The issue is not impairing correct function or firmware installation on G7104C. It is a documentation-only issue and will be fixed in a later release of the LC&CE Drivers.
Temporary Solution:
Should you require evidence for the correct firmware revision of G7104C for documentation purposes in the meantime, please consider adding a copy or reference for this bulletin entry to your documentation.
Fix Information:
The issue has been fixed.
Keyword: Administration
One-line Description:
Secure File IO cannot be set
Problem:
Secure File IO file protections cannot be set on folders outside of users\public. If the user does not have local admin privileges, the folders outside of the users\public will not be secured when enabling Secure File IO.
Temporary Solution:
The user needs to have local admin privileges in order to secure folders outside of users\public or start the ChemStation Administrator Tool as an administrator.
Fix Information:
n/a
Keyword: Administration,User Interface
One-line Description:
Activity Log in OLCP does not record the Application name when login from the Web UI
Problem:
The Activity Log in the OpenLab Control Panel does not record the Application Name (e.g. Content Management) when a user connect to the Content Management web interface, only the action of login in and the user name is recorded.
Temporary Solution:
n/a
Fix Information:
This issue is resolved in OpenLab Server v2.5 and Openlab ECM XT v2.5
Keyword: Agilent GC Driver
One-line Description:
Unable to download temperature setpoint to Gasifier II
Problem:
Unable to download temperature setpoint to Gasifier II. This is a driver issue only, the function works from the browser interface.
Temporary Solution:
n/a
Fix Information:
Resolved in 3.3.70
Keyword: Audit Trail
One-line Description:
Method Audit Trail does not include change of Injection source when HS/PAL is no longer configured
Problem:
GC with HS or PAL as the injection source. Close the session and remove the external device. Open the session and select Compare. No differences are seen in the method. Download the original method. The injection source has been changed to Manual. No change is logged to the Method Audit Trail.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: Audit Trail
One-line Description:
UIB II - set a timetable and control "Method Audit Trail"
Problem:
If a timetable of UIB II is set and stored via "Save Method as..", the time of the change should be stored in the "Audit Trail". But the time of the method change is not correct.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: Audit Trail
One-line Description:
Automatically saving of method is missing in audit trail
Problem:
Method changed during the sequence is automatically saved but not recorded in audit trail. This automatic saving should be visible in the audit trail of the method but it is missing.
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB CDS ChemStation C.01.06.
Keyword: Audit Trail
One-line Description:
Selection of storage of instrument curves is not reported in method audit trail
Problem:
Selection of storage of instrument curves is not logged in method audit trail.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS C.01.07 SR2 and in OpenLAB CDS C.01.07 SR1.
Keyword: Audit Trail
One-line Description:
Sequence Audit Trail is not written for 2D LC instruments
Problem:
Sequence Audit Trail is not written for instruments without 3DUV Addon license in C.01.07.
Temporary Solution:
Fixed in C.01.07 SR1.
Fix Information:
n/a
Keyword: Audit Trail
One-line Description:
Update Master Method with DA settings does not copy the full list of changes to the audit trail
Problem:
When method audit trail is activated the changes made in data analysis are listed in full detail. This is usefull to correlate result versions with method versions. For a master method it is enough to have the current status at the time of acquisition. When the Data Analysis parameters are changed, the full audit trail is stored with the sequence method. When the master method is updated from Data Analysis, the audit trail of the master method contains only a link to the sequence method. With a certain combination of privileges, a user can update the master method, but discard the changes to the sequence method.
Temporary Solution:
N/A
Fix Information:
Beginning with rev. C.01.08 only saved methods can be used to update the master method.
Keyword: Audit Trail
One-line Description:
Pre-run changes to sequences are not captured in the sequence audit trail
Problem:
The current design of the sequence audit trail introduced with OpenLAB CDS C.01.07 does not cover the modification of sequence templates. Sequence template development is not considered to be traced as it happens before the validation process. Templates can be protected by user privileges and storing the master templates in OpenLAB DataStore or ECM provides versioning to track the history of validated templates. For the sake of usability, the sequence template can be modified before starting the sequence. These changes are currently not tracked in the audit trail.
Temporary Solution:
The pre-run changes done by the sequence submitter can be detected by comparing the sequence template, the final state of the result sequence and the sequence audit trail.
Fix Information:
With revision C.01.07 SR2, the audit trial will keep track of the number of sequence lines so the net changes can be interpreted easier.
Keyword: Audit Trail
One-line Description:
Sequence audit trail reports outdated sequence line number
Problem:
The design of the sequence audit trail introduced with OpenLAB CDS C.01.07 traces all committed changes to the sequence table while the sequence is acquiring data or when data gets reprocessed. The changes are traced when the sequence table gets closed and the net changes are written to the audit trail. As multiple insert, move and delete operations changes the sequence line numbers, the audit trail needs to refer to the line number at the begin of the operation. This may lead to artefacts where e.g. the original line #x is reported as being deleted, while other entries refer to line #x as being added and modified.
Temporary Solution:
N/A
Fix Information:
Begining with revision C.01.07 SR2, the audit trial contains additional information about how many lines have been added, deleted or moved in total. These entries are grouped together for increased readability.
Keyword: Audit Trail
One-line Description:
Sequence Method audit trail is missing entries for some runs
Problem:
The sequence method audit trail will capture all the method changes during data acquisition. For some sequences, the sequence method audit trail is missing entries for some runs. For example, the sequence log shows re-calibration was done by calibration standard, but the sequence audit trail does not capture it.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: Audit Trail
One-line Description:
Changes of integration events in combination with preview results may not be tracked correctly in method audit trail
Problem:
In case a user with appropriate privileges is working on the integration events in data analysis view and click immediately on ‘identify peaks, calculate and preview results’ it can happen that the changes are not tracked correctly in the method audit trail.
Temporary Solution:
To ensure all changes done in the integration events are tracked correctly in the method audit trail, it is strongly recommended to close the integration event window before generating a report.
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.09 Update 1. Resolved in OpenLab CDS ChemStation C.01.10.
Keyword: Audit Trail
One-line Description:
Severity levels of activity log did change and need explanation
Problem:
There is a mapping of old severity levels to new severity levels which is not well explained in the manuals. The old activity log severity level "Normal" and "Info" are both corresponding to the new severity level "Info". Definitions of the activity log severity levels: Info - Used for all informational logs generated by the system. Most application and instrument activities fall into this severity level. Warning - Used for entries describing activities or conditions that could lead to failure. Useful for instruments that want to record a consumable reaching a set limit. Error - A failure has occurred. Some examples: The software did not behave as expected, an instrument has prematurely shut down, or a run was prematurely ended.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: Audit Trail
One-line Description:
Snapshots are not recorded in the audit trail
Problem:
There is no audit trail for snapshots.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: Audit Trail
One-line Description:
Snapshot not disabled when “After any data modification” is checked
Problem:
Users are able to take multiple snapshots overwriting the previous snapshot which is not captured in an audit trail.
Temporary Solution:
n/a
Fix Information:
The Snapshot mode is disabled when “After any data modification (full data tracking)” is selected in the ChemStation Administration Tool beginning in OpenLab CDS ChemStation C.01.10 Update 1.
Keyword: Audit Trail
One-line Description:
Time stamps of method audit trails are printed in ISO format
Problem:
Beginning with rev. C.01.10, the time stamps of printed method audit trails appear in ISO format (global time), whereas all other time stamps used in ChemStation reflect local time. This issue is visible if data gets reviewed in a time zone different to the time zone where or when the data was acquired. The ISO time format was introduced with a new reporting component.
Temporary Solution:
n/a
Fix Information:
There is currently no design change planned.
Keyword: Audit Trail
One-line Description:
Instrument error stopped sequence, but log also states “stopped by user”
Problem:
The sequence did not complete due to an instrument error but the log states the sequence was also 'stopped by user' which is incorrect.
Temporary Solution:
n/a
Fix Information:
Fixed in C.01.10 Update 1
Keyword: Audit Trail
One-line Description:
Sequence audit trail can be closed without entering a reason for the update
Problem:
The sequence audit trail is enabled. The sequence table is opened, and edits are made. The sequence audit trail can be closed without entering a comment by right-clicking on ChemStation in the taskbar. This action will close the audit trail without entering a reason for the change done in the sequence table.
Temporary Solution:
Implement or update a procedural control (SOP) describing the meaning and usage of the audit trail comments in the workflow.
Fix Information:
Resolved in OpenLab ChemStation LTS 01.11.
Keyword: Audit Trail
One-line Description:
Acquisition method last change timestamps do not match between Intelligent and Classic reports
Problem:
Made changes to the sequence method and then updated the master method, (using the menu item in DA view : "Method"->"Update Master Method"). Ran this method and checked the audit trail, classic report, and IR report. The audit trail and classic report show a matching time for the acquisition method "Last Modified Date" due to the update. The IR report shows a few seconds earlier than the other 2 for the acquisition method last modified date time. Last method update time stamps do not match between IR and classic reporting audit trails.
Temporary Solution:
Workaround: The problem can be avoided by directly saving the master method via the sequence parameter option "Master Method Updates"->"Update data analysis parameters only" to update the master method from the data set method.
Fix Information:
Resolved in OpenLab ChemStation C.01.10 Update 5.
Keyword: Autosampler
One-line Description:
EZChrom: PAL XT module not available after installation
Problem:
1. Uninstalled existing PALxt driver. Driver was removed from Programs and Features 2. Installed PALxt B.01.07 using Gold Master disk. (dated 10/1/15) 3. Driver appeared in Programs and Features. 4. Restarted because it is EZChrom. 5. Opened EZChrom Control Panel and selected to configure. 6. PAL XT (CTC Sampler) was not present in modules window. (see pic) 7. PAL3 and HS were already present.
Temporary Solution:
n/a
Fix Information:
Fixed in B.01.01
Keyword: Batch
One-line Description:
Cannot open batch - specified range parameter is bad
Problem:
An error occurs when loading newly modified and saved batch files. A batch file that is created in an earlier revision does not save the modifications from batch review correctly. When the modified file is loaded again, the error occurs: "specified range parameter is bad" The original batch file will load without error, only the newly modified and saved batch file will not load.
Temporary Solution:
Fixed in Hotfix 04 for C.01.05
Fix Information:
n/a
Keyword: Batch
One-line Description:
Getting errors when selecting any Classic Performance report templates in Batch Review
Problem:
Using Extended Performance Report (classic) in batch review does not work. An error message shows: "ERROR Processing Run: The name given was not found"
Temporary Solution:
Use recalculate in Data Analysis
Fix Information:
Fixed in OpenLAB CDS C.01.06.
Keyword: Batch
One-line Description:
Batch Reporting does not allow the same compound name to be used for both signals
Problem:
The same compound name is used for the front and back detector signals in the calibration table. When a batch report is generated, the front injection results are being reported for the back injection. The same compound name can no longer be used for both signals without causing a problem with the batch report.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: Calibration
One-line Description:
Run a sequence with recalibration, only the first ISTD is updated
Problem:
The ASTM method for biodiesel uses two internal standards in the calibration table, but when you run a sequence with recalibration only the first ISTD gets updated.
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB CDS C.01.06. Up to 8 ISTD compounds can be updated during one run.
Keyword: Calibration
One-line Description:
Sequence run for bracketing standard which is started with blank completed without creating data files
Problem:
When a sequence is started with blank data files, data files are not generated. When the sequence is started with standard as bracketing std. (line 1) and sample (line 2), sequence completed and data files are generated
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB CDS C.01.06.
Keyword: Calibration
One-line Description:
C.01.04 Hot Fix03 : incorrect calibration curves legend in calibration curves printout
Problem:
When using "ISTD" as "Calculation mode" and "Area" for "Based on", the calibration curves x-axis and y-axis shown on the calibration curves were : "x-axis=amount ratio" "y-axis=area ratio” but the report printout for calibration curves have the following incorrect legend in English version: "x=amount" "y=area" And incorrect legend in Japanese version: "x= amount" "y= height"
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB CDS C.01.06.
Keyword: Calibration
One-line Description:
Compounds without ISTD correction are not reported
Problem:
In reports with calculation type ISTD or ISTD% amounts for compounds without ISTD reference are 0. If an ISTD compound is not detected, all compounds that use this reference are reported as 0.
Temporary Solution:
None.
Fix Information:
None. This is the intended outcome of the calculation. A mix of corrected and uncorrected amounts must be avoided.
Keyword: Calibration
One-line Description:
0.000 amount in Sequence summary report using ISTD% report
Problem:
Intermittently, when generating the report after a sequence with 0.000 amount, reprocessing the sequence gives 0.000 amount in Sequence summary report for ISTD% report when the amount should not be zero.
Temporary Solution:
n/a
Fix Information:
Resolved in revision OpenLAB CDS ChemStation C.01.07 SR2.
Keyword: Calibration
One-line Description:
Wrong vial number in intelligent report with bracketing calibration
Problem:
The wrong vial number is reported when using intelligent reporting with bracketing calibration.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS ChemStation Edition C.01.07 SR2.
Keyword: Calibration
One-line Description:
Chromatograms extracted from UV iso-absorbance plot cannot be directly used for quanitfication
Problem:
Calibration tables and associated corresponding signal descriptions generated based on extracted UV chromatograms cannot be directly used to generate quantitative reports. The signal description in the signal detail and the chromatogram window do not necessarily match precisely.
Temporary Solution:
If you aim to use the calibration table for compound identification or quantification, you need to ensure the signal description shown in the signal details window matches the signal description in the header of the chromatogram window. For certain UV detectors you need to manually adjust the precision of the wavelength numbers. After all, quantification based on extracted signals need to evaluated very carefully due to the reduced signal/noise ratio and its impact on quantification precision.
Fix Information:
There is currently no fix planed.
Keyword: Calibration
One-line Description:
in Signal Details and Calibration tables has an old functionality of "Clear" instead of "Copy"
Problem:
In Method and Run Control view, through the method editing steps the Signal Details window is prompted. A Signal Table Editor is available and rows with new Signals could be inserted, appended and added. Once a row is selected and the Keyboard function Ctrl+C is used, the entire sub-window is no longer available. An error message is generated ("Unknown error for register TempClipReg!") The same behavior was noticed for the Calibration Table and integration results table in Data Analysis view, with the error message "EDTAB 24: Unknown error for Register TempClipreg!" (EDTAB 65 for integration results table)
Temporary Solution:
Attached are macros that add a simplified menu item to copy the integration results to the clipboard. Installation: Just make sure there is no User.Mac file in the \Chem32\Core folder or \Program Files (x86)\Agilent Technologies\ChemStation\Core. If there is, provide a copy so it can be reconciled to the content of the attached User.Mac. If there is no existing User.Mac copy both macro files to \Chem32\Core or \Program Files (x86)\Agilent Technologies\ChemStation\Core folder. Restart the ChemStation and when you go to [Data Analysis] there will be a new [Clip] menu. Operation: Under the [Clip] menu is an item for [Integration Results]. When you click this item the current integration results are copied to the ClipBoard using the default settings. Note: This item copies the integration results for the currently loaded data file. It is not designed to handle more than one signal.
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.09.
Keyword: Calibration
One-line Description:
Include/ignore origin in default calibration settings shows a warning
Problem:
A situation can happen when the value of how the origin should be handled is different in the default calibration settings and the value in the calibration table. E.g. in the calibration table the value is “include” and in default calibration settings is “ignore” then a warning comes up, that a point is missing to generate the calibration curve.
Temporary Solution:
Change the values manually to have consistency or uncheck the Def check-box in the calibration table, then the default settings will not apply.
Fix Information:
Resolved in OpenLab CDS ChemStation Edition C.01.10
Keyword: CE Control
One-line Description:
Acquisition Method changed by user without required privileges
Problem:
When an instrument becomes online while an online session is running, instrument parameters in the method might be changed. This can happen, even if the logged on user does not have the privilege to change acquisition parameters.
Temporary Solution:
In order to protect a acquisition method from implicit method changes: - Prohibit usage of handheld LC instrument controller by a procedural control - Protect GC front panel keyboard by locking it in the configuration settings - When performing an instrument cold start the instrument method parameters are rest to factory defaults. Upon start-up of ChemStation or the user gets the option to load the instrument method. This needs to be prevented by choosing the configuration setting "New method from instrument". It is advised to load a stand-by method when closing ChemStation and shutting down the instrument. Upon re-start the stand-by method is loaded again.
Fix Information:
No design change planned.
Keyword: CE Method
One-line Description:
CE Method not working for CEC Mode and/or MSInstalled soft configuration option
Problem:
The pressure settings are not shown and therefore not working in the CEC Mode and/or MSInstalled soft configuration options. Vialtable implementation is not reliable.
Temporary Solution:
none
Fix Information:
Fixed in driver A.02.09 SP1
Keyword: CE Method
One-line Description:
Using timetable setting Internal Pressure in CE/p and CEC mode is not applied
Problem:
In CE/p and CEC mode, the method timetable offers the possibility to set internal and external pressure values. This function cannot be applied currently. When using CE in CEC mode, the timetable parameter "Internal pressure" can be set, but is not applied during run. When using CE in CE/p mode the parameter can be set, but results in a "download method failed" error during method download.
Temporary Solution:
As a workaround, make sure not to use the timetable parameter "Internal pressure".
Fix Information:
There is no fix available currently, It is planned to remove the "Internal pressure" parameter from the CE timetable in a future driver release.
Keyword: ChemStation
One-line Description:
LC/MS data acquired on Chinese or Japanese OpenLAB CDS ChemStation C.01.07 cannot be reported with Intelligent Reporter.
Problem:
When LC/MS data acquired with a Chinese or Japanese revisions of the ChemStation C.01.07, any intelligent report that contains a chromatogram in the template will show an error at the report generation (red cross).
Temporary Solution:
n/a
Fix Information:
Wait for this Service Release: OpenLAB CDS ChemStation Edition A.02.02 SR2. Once available, install the software service release.
Keyword: ChemStation
One-line Description:
LC/MS data acquired on Chinese or Japanese OpenLAB CDS ChemStation C.01.07 cannot be reported with Intelligent Reporter.
Problem:
When LC/MS data acquired with a Chinese or Japanese revisions of the ChemStation C.01.07, any intelligent report that contains a chromatogram in the template will show an error at the report generation (red cross).
Temporary Solution:
n/a
Fix Information:
Wait for this Service Release: OpenLAB CDS ChemStation Edition A.02.02 SR2. Once available, install the software service release.
Keyword: ChemStation
One-line Description:
LC/MS ChemStation - Intelligent Reporting would report a red cross when using Analytical Studio Reviewer
Problem:
This affects all ChemStation release from C.01.05 to C.01.07. Analytical Studio reviewer is often used with the MassHunter Walkup Add-on Software for the ChemStation. After report generation, you will observe a red cross in the Intelligent Reporting chromatograms section in the case of simultaneous use of Analytical Studio Reviewer and Intelligent reporting.
Temporary Solution:
Switch off either the Analytical Studio Reviewer report generator, or the intelligent reporting (by switching to classic reporting)
Fix Information:
For ChemStation release C.01.07: Wait for this Service Release: OpenLAB CDS ChemStation Edition A.02.02 SR2. Once available, install the software service release. For ChemStation releases from C.01.05 to C.01.06: Switch off either the Analytical Studio Reviewer report generator, or the intelligent reporting (by switching to classic reporting)
Keyword: ChemStation
One-line Description:
Intelligent Reporting template migration from OpenLAB CDS ChemStation Edition C.01.04 or C.01.05 to C.01.06 or higher
Problem:
With the C.01.06 release of the ChemStation, the schema how vial positions are used changed. While with the C.01.04 or C.01.05, "Vial 104" was still a valid vial position, this is no longer true for C.01.06. In C.01.06 and above, the vial position that is recorded in the result set does no longer contain the text "Vial ". Therefore any report template which uses the vial position for example in a filter, needs to be modified and use the correct vial position scheme in the filter value(s).
Temporary Solution:
n/a
Fix Information:
Modification of the report template: In a filter or a group repeating which contains the expression "Vial 102" for example. Change the filter or the group repeating to "102" to be valid in the new version of the OpenLAB CDS ChemStation. see comment:
Keyword: Client Report Rendering
One-line Description:
Data Peak without Apex Spectrum cannot render IR report with Apex Spectrum selected
Problem:
Data without Apex Spectrum, used to generate a report with a fresh template hangs during report generation, with the spectrum objects below. Spectra Plot, Spectra Plot All Peaks, Spectra Plot Compounds and Spectra Plot per Peak, the reports hangs and chemstation is unresponsive. Changing the IR template from Apex to All Peaks renders the IR report. If there is no Apex Spectrum for an identified Peak, and Apex is selected in the IR template, the report will not render.
Temporary Solution:
There several options to avoid this issue: - change the detector settings to collect all peak in spectra, not just the apex spectra - lower the detector threshold for the apex determination - add a area or height filter to the integrator or template that matches the detector threshold
Fix Information:
This defect is planned to be fixed in OpenLAB CDS ChemStation C.01.07.
Keyword: Client Report Rendering
One-line Description:
Intelligent report of English openlab chemstation can not show chinese name in PDF report
Problem:
Intelligent report of English openlab chemstation can not show chinese name in PDF report,but priview and export report as word report is ok ,only PDF report use "?" instead of chinese characters.
Temporary Solution:
n/a
Fix Information:
Customer needs to install the SimSun font on the English windows system
Keyword: Client Report Rendering
One-line Description:
Large matrix reports may produce additional empty pages
Problem:
Large matrix reports may produce additional empty pages. The issue is in the Microsoft Report Viewer control and cannot be fixed in the current product.
Temporary Solution:
Matrix reports easily exceed A4 or Letter sized paper formats and printing to a larger PDF paper format could be used as alternative.
Fix Information:
There is currently no fix planned.
Keyword: Client Report Rendering
One-line Description:
Calculating the standard deviation of identical numbers may return NaN
Problem:
When calculating the standard deviation or relative standard deviation of multiple identical numbers, the calculation may return "NaN" rather than 0. The problem is caused by the small delta derived from the rounded result of the average calculation and the individual values. Summing up these small numbers may result in a floating point exception.
Temporary Solution:
Do not apply standard deviation calculations to series of identical values.
Fix Information:
There is currently no fix planned.
Keyword: Client Report Rendering
One-line Description:
Changes in report templates are not immediately shown on report printouts
Problem:
Changes in report templates are not immediately shown on the printout if the user does not preview the report and save the template first.
Temporary Solution:
Always preview the report printout after applying changes to the template. Or save and reload the template from disk, if a preview should be avoided.
Fix Information:
There is no fix planned.
Keyword: Client Report Rendering
One-line Description:
Rendering error when using a template containing a large result matrix
Problem:
A template having a matrix inside a list may show a rendering error - The value 'x' is invalid. Valid values are between '0' and 'x' This error occurs when there is large data used for rendering and due to this the matrix data spans across multiple pages.
Temporary Solution:
For large result matrices it is recommended to print the results to PDF files, rather than paper. The paper format should adjusted to display large matrices on a single page.
Fix Information:
There is currently no fix planned.
Keyword: ColumnComp Driver
One-line Description:
TCC CLuster remains offline with message "Initialization of ColumnCompCluster failed
Problem:
After starting the application \ connecting to the instrument, the TCC Cluster remains offline. HowTo reproduce: - Configure an instrument with the single TCCs - In all single TCC method UI select the Combined mode for the right temperature zone and download it to the instrument - close \ disconnect the instrument - Configure a 2nd instrument with a TCC cluster. Just one column (no waste or bypass) and put in the upper right corner of one TCC. Trying to connect to this TCC cluster. It remains offline.
Temporary Solution:
N/A
Fix Information:
LC & CE Drivers A.02.08 SP1
Keyword: ColumnComp Driver
One-line Description:
Position number after run is reported wrongly for Multicolumn Thermostat G7116B in the report and audit trail
Problem:
The Multicolumn Thermostat G7116B allows setting a valve position to which the valve is switched after a run. This switch has been done correctly but was not reported correctly as well in the report as in the method audit trail. Offset between reported and actual valve position is 1, i.e. position 2 has been reported as 1.
Temporary Solution:
n/a
Fix Information:
This issue has been fixed with driver release A.02.11 SP1.
Keyword: ColumnComp Driver
One-line Description:
Field sizes for column description and comment have been increased from 16 to 32 characters for G7116B Multicolumn Thermostat
Problem:
For driver release A.02.11 and the corresponding firmware revisions, the maximum length for the column description and comment has been limited to 16 characters, which may not be sufficient for some columns. 16 characters apply to Latin characters, symbolic languages like Chinese or Japanese may require more memory for internal representation, which further limits the length of possible descriptions.
Temporary Solution:
not available
Fix Information:
This has been fixed with driver release A.02.11 SP1. Please note that this fix requires a firmware update for all relevant modules, which are the G7116B MCT module itself, which requires firmware C.06.72 as a hosted module and B.06.72/D.06.72 for all potential hosts. Please update all modules to the latest available firmware set including this fix. For details about the hosted module concept, please refer to the Firmware Update Manual.
Keyword: ColumnComp Driver
One-line Description:
Method transfer from G1316A/B/C to G7116B failed if option "next position after run" was chosen
Problem:
If a method with option "next position after run" was chosen for G1316A/B/C was loaded for a G7116B module, such a method became invalid and was rejected by the CDS and a CDS specific error message may be displayed or written to the log book.
Temporary Solution:
not available
Fix Information:
This issue has been fixed with driver release A.02.11 SP1.
Keyword: ColumnComp Driver
One-line Description:
Wrong display for 2pos/10port and 2pos/6port valves in method UI for G1316A/B/C modules
Problem:
For G1316A/B/C modules using the 2pos/10port valve or 2pos/6port valve, the method UI shows different port connections than the written connections, e.g. 1->2 is written but 1-> 10 is depicted. The description is correct but the display is wrong.
Temporary Solution:
Rely on written description.
Fix Information:
Fixed with A.02.11 SP1
Keyword: ColumnComp Driver
One-line Description:
No tooltip displayed if linked detector is missing for G7116B Multicolumn Thermostat
Problem:
If a detector has been linked optionally to a G7116B Multicolumn Thermostat (MCT) for aligning temperature control of MCT and detector flow cell and this detector is not present, the G7116B does not become ready as indicated by a yellow status LED. No tooltip is shown which would explain that reason.
Temporary Solution:
n/a
Fix Information:
Fixed with driver release A.02.11 SP1.
Keyword: ColumnComp Driver
One-line Description:
Incompatible change in behavior of 2-position/6-port and 2-position/10-port valves in G1170A, G7116B and G4227A compared to Thermostatted Column Compartments G1316A/B/C
Problem:
Multicolumn Thermostats (MCTs) G7116B, External Valve Drives G1170A and Flexible Cubes G4227A are controlling 2-position/6-port valves and 2-position/10-port valves differently from similar Thermostatted Column Compartments (TCCs) like G1316A/B/C. This change of internal valve positions has been introduced for the 1290 Infinity valve drive G1170A and the Flexible Cube G4227A but has not been relevant because no method transfer was needed between rather different modules. Both valves use two positions: Modules Valves Position 1 Position 2 G1316A/B/C 2pos/6port 1->2 1->6 G7116B, G1170A, G4227A 2pos/6port 1->6 1->2 G1316A/B/C 2pos/10port 1->2 1->10 G7116B, G1170A, G4227A 2pos/10port 1->10 1->2 Depending on the module type, different ports are connected by the same method parameter (valve position). If such valves were used e.g. for column selection and columns are connected to ports with same numbers in a new module, a different column may be used by a transferred method. This is displayed correctly in the user interface but may not be expected.
Temporary Solution:
n/a
Fix Information:
If your methods rely on valve positions, please verify that they are using correct connections. When transferring methods between different module categories (e.g. G1316C vs. G7116B), please either change your method parameters or corresponding connections to ports.
Keyword: ColumnComp Driver
One-line Description:
Error dialog and error logbook entry for TCCs G1316A/B/C
Problem:
When loading a method in OpenLab CDS EZChrom edition A.04.06, which includes any Thermostatted Column Compartment G1316A/B/C using LC and CE drivers A.02.11, an error dialog may be shown along with a log book entry "Error in RCNet_RCMethodUI::ReturnRequestResponse : There is an error in XML document (0, 0).". This error message has no impact and can be ignored.
Temporary Solution:
Ignore this error message
Fix Information:
This issues has been fixed with driver A.02.12.
Keyword: ColumnComp Driver
One-line Description:
Control UI for G1316C does not work when using the Instrument Control Framework ICF in client/server setups.
Problem:
In a client/server scenario, where the server has driver A.02.10 and the Client has been upgraded to A.02.11 SP1, the control user interface of the module dashboard a window may appear with the message that the settings are being read and stays unchanged.
Temporary Solution:
Close this dialog using "Cancel". For switching on the thermostat, use command "Switch on/off" from the context menu of the dashboard.
Fix Information:
This issue has been fixed with driver release A.02.12.
Keyword: ColumnComp Driver
One-line Description:
Changing control settings like switching on/off the Column Compartment Thermostat or Leak Detection in the instrument control UI does not work
Problem:
Two possibilities are available for switching on or off the thermostat of a column compartment 1. The Instrument Control User Interface (click "Control" in the context menu of the TCC). Then select Thermostat On or Off. 2. "Switch On" or "Switch Off" function in context menu. Option 1 does not work with driver A.02.11. Switching between on and off does not switch the thermostat. The same applies also to the settings for "Leak Detection".
Temporary Solution:
For switching the thermostat of the column compartment on or off, use Switch On or Off from the module dashboard context menu (option 2 above). The leak detection settings can still be changed using the Instant Pilot.
Fix Information:
This issue has been fixed in driver release A.02.12.
Keyword: ColumnComp Driver
One-line Description:
Import/export functionality in Column Assignment dialog of MCT has been removed
Problem:
Importing of column information via Column Assignment dialog would overwrite existing tags. In most cases, this would not be the expected behavior. Expected behavior would be to import such information only for columns without or with empty tags.
Temporary Solution:
Column information import has been removed for driver A.02.12 in order to avoid issues. Do not use this functionality for driver A.02.11.
Fix Information:
n/a
Keyword: ColumnComp Driver
One-line Description:
No comments can be entered for virtual tags
Problem:
Column information can be either stored on Column tags (recommended), in a column database (if supported by a CDS, e.g. OpenLAB CDS ChemStation Edition C.01.07 SR2) or temporarily on the instrument mainboard (as a virtual tag). Not all possible information may be supported outside the driver. For example, column comments cannot be stored on the mainboard. Previous drivers allowed entering comments or importing them from the ChemStation database but this information was not persisted for virtual tags.
Temporary Solution:
Do not try using comments for virtual tags, as they will not be saved.
Fix Information:
Since driver A.02.14, the comment field is no longer editable if no physical tag is present.
Keyword: ColumnComp Driver
One-line Description:
Column information may get lost within VTCs during auto-configuration in some cases
Problem:
A VTC allows handling of column information within clustered column thermostats or columns outside the system. This information is stored on column tags, on the system (mainboards) and - if applicable - in the ChemStation column database. Information for columns, which are either installed in column thermostats G1316C or outside the system may be reset during an auto-configuration. This is because modules configured in the VTC have changed their order.
Temporary Solution:
Re-order modules in the VTC configuration to the previous one after doing an auto-configuration.
Fix Information:
Fixed in LC and CE driver A.02.14
Keyword: ColumnComp Driver
One-line Description:
Last injection date and injection counters are not updated in VTC in some cases
Problem:
A VTC allows handling of column information within clustered column thermostats or columns outside the system. This information is stored on column tags, on the system (mainboards) and - if applicable - in the ChemStation column database. Information for columns, which are either installed in column thermostats G1316C or outside the system may not be updated correctly.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: ColumnComp Driver
One-line Description:
Valve-Thermostat-Cluster time table can be edited in read-only mode
Problem:
Chromatographic Data Systems (CDS) can temporarily set access to driver functions to read-only. By mistake, the time table of valve-thermostat clusters (VTC) is editable. This may cause issues for the CDS.
Temporary Solution:
Do not edit the VTC time table in read-only mode.
Fix Information:
This has been fixed in driver A.02.14.
Keyword: ColumnComp Driver
One-line Description:
Maximum column void volume is 50 mL
Problem:
Import of column information with a void volume larger than 50 mL from the OpenLAB CDS ChemStation Edition column database may fail because of a numeric limit in the driver.
Temporary Solution:
Avoid values larger than 50 mL. As a workaround, try to write void volume information e.g. to the comment field.
Fix Information:
This shall be fixed with an upcoming driver release.
Keyword: ColumnComp Driver
One-line Description:
Auto configuration fails when changing the valve type in an MCT for an existing valve thermostat cluster
Problem:
Auto configuration fails when changing the valve type in an MCT (Multi Column Thermostat) G7116A/B for an existing VTC (valve thermostat cluster). When trying to configure the valve thermostat cluster an unhandled exception is displayed.
Temporary Solution:
Do an auto configuration without clustering the modules, start the CDS and close it again, then do a second auto configuration and set up the VTC cluster as desired.
Fix Information:
This has been fixed with driver version A.02.16.
Keyword: ColumnComp Driver
One-line Description:
G1316C does not recognize 800 bar Quick Change Valve Heads configured in LabAdvisor
Problem:
The driver of G1316C Agilent 1200 Series Thermostatted Column Compartment is unable to recognize 800 bar Quick Change Valve Heads that have been configured in LabAdvisor.
Temporary Solution:
When configuring a 800 bar Quick Change Valve Head in a G1316C Agilent 1200 Series Thermostatted Column Compartment, use the built-in driver function only to manually configure the Valve Head. With the Valve Head installed, perform an auto-configuration. Start up your data system and connect to the instrument. Right-click on the dashboard tile of the G1316C and access the "Modify" > "Valve type..." function. Choose the correct valve type from the list and follow the instructions to modify the tag information on the device manually. Power down your device and install the 800 bar Valve Head. Perform an auto-configuration, connect to the instrument with your data system and verify that the proper Valve Head is displayed in the G1316C Dashboard tile.
Fix Information:
n/a
Keyword: Companion
One-line Description:
Companion fails to run blank run for valved/ALS injections
Problem:
Create a run table that uses valid locations and some that use 'blank' as the injection location (a valid choice in Companion). The runs associated with this location fail to run with a pop-up: 'Vial range must be in range 1-32'. This also happened for an ALS method.
Temporary Solution:
n/a
Fix Information:
Fixed in Revision B.04.01.
Keyword: Control Panel
One-line Description:
Exception when login in OpenLab Control Panel. ECM does not display all the subfolders when browsing for Project Path during project creation; exception displayed
Problem:
When the user creates projects, the standard procedure is to browse to the folder Location when selecting the Project Path. However, there are more than 100 sub-folders, it is not possible to set the project path correctly, because when browsing, the ECM window displays only the first 100 folders. The following exception is displayed: "Agilent.OpenLab.SharedServices.SharedServicesInternalException: The incoming request has too many parameters. The server supports a maximum of 2100 parameters..." An exception is also displayed when logging in OpenLab Control Panel: "Agilent.OpenLab.SharedServices.SharedServicesInternalException: The incoming request has too many parameters..."
Temporary Solution:
n/a
Fix Information:
The fix is OpenLAB Shared Services (OLSS) 2.1 Hotfix 4 and is distributed with the CDS. OpenLAB Shared Services 2.1 Hotfix4 is included with C.01.07 SR4 (Disk1) and is installed. Resolved in C.01.08.
Keyword: Control Panel
One-line Description:
Unable to configure multiple instruments with the same name in different locations (on the same AIC)
Problem:
On the same AIC, created an instrument in 2 different locations with the same instrument name. The first instrument can be configured as expected. The second instrument starts to configure, but the module configuration box never shows up. No errors are shown in software or operating system.
Temporary Solution:
Do not configure 2 instruments on the same AIC with the same name on 2 different locations.
Fix Information:
n/a
Keyword: Control Panel
One-line Description:
Create new instrument causes error "System.Net.Sockets.SocketException.."
Problem:
When creating a new instrument, the following dialog box is displayed: "An error prevented creating the instrument". The error text is: System.Net.Sockets.SocketException (0x80004005): The requested name is valid, but no data of the requested type was found at System.Net.Dns.GetAddrInfo(String name) at System.Net.Dns.InternalGetHostByName(String hostName, Boolean includeIPv6) at System.Net.Dns.GetHostEntry(String hostNameOrAddress) at Agilent.OpenLab.Communication.NetworkLocation.get_HostEntry() at Agilent.OpenLab.Communication.NetworkLocation.get_IsLocal() at Agilent.OpenLab.UI.Instruments.ViewModel.InstrumentEditor.IsLocal(InstrumentController controller) at Agilent.OpenLab.UI.Instruments.ViewModel.InstrumentEditor.GetInstrumentControllers(Connection connection, InstrumentApplicationViewModel application) at Agilent.OpenLab.UI.Instruments.ViewModel.InstrumentEditor..ctor(InstrumentLocationCollection parentLocation) at Agilent.OpenLab.UI.Instruments.Commands.Actions.CreateInstrumentAction.ExecuteAction(InstrumentLocationCollection parameter) This is caused by an instrument controller that has a Network Status of "Not resolved".
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: Custom Fields
One-line Description:
Default value for a custom field is not cleared
Problem:
The custom field default text is still displayed, even after the text has been deleted from the method and the method is aved. This happens in the following situation: A sequence contains methods with custom fields and the fields have been viewed from the sequence table. Then the method is modified, by making a text field mandatory and removing the default value. VIewing the filed values again shows that the field is labeled as mandatory, but the default text is still listed.
Temporary Solution:
The method printout shows the correct setup. To clear the values without rewriting the sequence change the value type from text to numeric and back.
Fix Information:
No fix available.
Keyword: Custom Fields
One-line Description:
Custom Fields can disappear after acquisition
Problem:
When custom fields are viewed from the sequence table of a stored result set, the values for the custom fields might be lost. This can occur when the master method used to create this result set has been modified since acquisition resulting in a different set of custom fields.
Temporary Solution:
Viewing the current Custom Field values from the menu will not change the entries. Viewing the current Custom Field values from the sequence table leads to this issue. The original value of the custom fields can be retrieved from earlier revisions in backend storage. Fix planned as Hotfix 05 for revision C.01.06.
Fix Information:
Fixed in revision C.01.07.SR2
Keyword: Custom Fields
One-line Description:
Ability to store results in sample or compound custom fields
Problem:
Custom Fields functionality doesn’t allow reporting any dynamic values (e.g. spectral purity results). Custom Fields are required to be set up before acquisition/reprocessing and their values remain static during the whole workflow which is not useable.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.07 SR4 Hotfix 1. The Method’s Run Time Checklist will be extended by an additional entry to run a macro command to modify custom fields values during data analysis shortly before reporting.
Keyword: Custom Reports
One-line Description:
Relative Retention Time solution does not work for uncalibrated peaks
Problem:
The Relative Retention Time solution available on the Agilent web site in the FAQ section only works for calibrated peaks. So does the solution that has been implemented in C.01.04 and Intelligent Reporting.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: Custom Reports
One-line Description:
FIltering syntax has changed
Problem:
In OpenLAB CDS A.02.01 expressions can be used for filtering. As a result of this new feature filtering for values has to use expressions as well. If the values are entered as plain numbers, the report will print an error: An error occured during local report processing. An error has occured during report processing. The processing of FilterExpression for the table 'OLIR_Table_XX' cannot be performed. Cannot compare data of types System.Int32 and System.String. Please check the data type returned by the FilterExpression.
Temporary Solution:
The correct syntax for A.01.05 and older for filtering peaks was: =Peak_RetentionTime > 1 when peaks with a retention time above one minute should be displayed. In A.02.01and higher the syntax needs to be: =Peak_RetentionTime > =1
Fix Information:
Existing report templates written with the old syntax will show up correctly in new software. No fix required.
Keyword: Custom Reports
One-line Description:
Calibration Levels and the corresponding amounts might not match in Intelligent Reporting
Problem:
If a multi-level calibration is recalibrated at the beginning of a sequence and later, only some of the levels are recalibrated, the Intelligent Reporting calibration report may associate the values and the levels incorrectly.
Temporary Solution:
Classic Calibration reports report the correct values.
Fix Information:
Fixed in C.01.05 SP1. Fixed in C.01.06 Hotfix 03. Fixed in C.01.07
Keyword: Custom Reports
One-line Description:
Sample order number change in C.01.10
Problem:
The sample order number is a sample field which can be used in Intelligent Reporting to reflect the number of samples as submitted in the sequence table. The sample order value for each replicated injection of a sample would have the same value. This field can be used to save report variable with unique key value or report aggregate (list of values) by category. In C.01.10, this field value was unexpectedly changed and cannot be used for the purposes described above.
Temporary Solution:
Other fields such as the sample vial number, sample name or sample description can be used to flag and categorize the replicated injections. This requires a template modification.
Fix Information:
This issue is fixed in C.01.10 update 04.
Keyword: Custom Reports
One-line Description:
Topology report considers real instruments only
Problem:
Data Analysis only instruments or instruments that are operated in off-line mode only do not get reported in the topology report.
Temporary Solution:
n/a
Fix Information:
A design change is considered for a future release of OpenLab ChemStation.
Keyword: DAD/MWD Driver
One-line Description:
In localized CDS for ELSD detector G4218A no signal data is shown or stored
Problem:
On a localized chromatograpy data system the driver for ELSD detector throws an exception and fails to show or store the detector's signal data. On an English system the ELSD driver's operation is not affected.
Temporary Solution:
n/a
Fix Information:
For a localized (e.g. Chinese, Japanese...) system a patch for the ELSD driver is made available.
Keyword: DAD/MWD Driver
One-line Description:
High load on CAN may cause "Spectra runbuffer overflow" entries and a loss of spectra
Problem:
If LAN is connected not to LC module generating most data (i.e. to Pump in spite of DAD's presence) then interconnecting CAN connection between modules might get overloaded. Potential consecutive loss of spectra may be observed and is flagged as "Spectra runbuffer overflow" in logbook and on report.
Temporary Solution:
Installation documentation for LC HW states that the LAN access should be in the module with the highest workload, i.e. by DAD/MWD - FLD - VWD - RID - Pump - ALS (in this order) and independent whether it is a 1100 - 1200 - 1290 - 1260 system.
Fix Information:
n/a
Keyword: DAD/MWD Driver
One-line Description:
Autobalace executed although in Sequence table set to "OFF"
Problem:
1) A method XYZ.M is saved with following DAD settings: Autobalance -> Prerun 2) A sequence table is saved using the method XYZ.M and AutoBalace set to "OFF" 3) The sequence gets started and although AutoBalace set to "OFF" it is executed.
Temporary Solution:
still under investigation
Fix Information:
none
Keyword: DAD/MWD Driver
One-line Description:
Occasional shift in retention times for G1315A/B or G1365A/B
Problem:
Occasionally some early data points (with negative time stamp) may occur that have not been handled correctly. Some slight deviation / jitter of retention times compared to similar runs may happen.
Temporary Solution:
n/a
Fix Information:
A patch for RC.Net Driver A.02.05 or A.02.06 is available: A.02.05 [SP1], respectively A.02.06 [SP1]
Keyword: DAD/MWD Driver
One-line Description:
Retention times may be slightly too high for DAD data acquisition rates below 1 Hz
Problem:
If very low data acquisition rates are used for the DAD, these frequencies are rounded in the driver and chromatographic data system. Affected data rates are 0.3125 Hz (rounded to 0.31 Hz, corresponding to 16 seconds response time) and 0.625 Hz (rounded to 0.62 Hz, corresponding to 8 seconds response time). Data acquisition rates are used by the data analysis for calculating the time values from the data point counter (e.g. at 2 Hz, the next data point comes 0.5 seconds later). Because of the rounding, the real module frequency is higher by 0.8% than what is shown in the UI/report and retention times are increased by the same factor. The effect increases with the run time, e.g. a peak coming at 10 minutes would be displayed at 10:05 minutes. This issue has been introduced with LC driver A.02.11 in 2015. It affects all DADs regardless of the CDS. The issue only becomes visible when comparing measurements run with affected drivers and non-affected drivers, but not e.g. between calibration and quantitation runs with the same software.
Temporary Solution:
n/a
Fix Information:
This has been fixed with driver version A.02.16.
Keyword: Dashboard
One-line Description:
Module stays offline after a configuration change using the "Modify" function
Problem:
If method resolution occurs due to a configuration change, after applying a modify function and the user did not accept this resolution within one minute, the module goes offline.
Temporary Solution:
Either accept the resolution within one minute or restart the CDS (Chromatographic Data System).
Fix Information:
n/a
Keyword: Data Analysis
One-line Description:
Reprocessing copy starting as a background window.
Problem:
When using "Reprocessing Copy" from the view menu, it may happen that the reprocessing session is hidden behind the Online session.
Temporary Solution:
You can minimize the online session to access the Reprocessing Session
Fix Information:
n/a
Keyword: Data Analysis
One-line Description:
Data System will not "Autoscale" a TCD data file to the correct peak
Problem:
Data analysis will not "Autoscale" a TCD data file. AUTOSCALE autoscales to the first (smallest) peak. NOTE, there is no solvent, first peak at 0.798 minutes is very small, software identifies peak 4 at 2.158 minutes as the solvent.
Temporary Solution:
Full Scale works correctly. Use Ranges can be made to work correctly.
Fix Information:
n/a
Keyword: Data Analysis
One-line Description:
Y-axis labeled incorrectly when loading GC signals acquired with MSD ChemStation
Problem:
When loading GC-signals acquired with the MSD ChemStation, the Y-axis is labeled: mAu. This should be pA. The signal units are read directly from the rawdata file "FID1A.ch" which are written by the acquiring software. The acquiring software shows that mAU was written. In order to be correct this, it must be fixed by the acquiring SW.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: Data Analysis
One-line Description:
ChemStation B.04.03 [54] will not "Autoscale" the TCD data file
Problem:
There is no solvent peak and the first peak is very small so the software identifies the wrong peak as the solvent. The Chromatogram is not autoscaled correctly.
Temporary Solution:
Full Scale works correctly. Use Ranges can be made to work correctly.
Fix Information:
n/a
Keyword: Data Analysis
One-line Description:
C.01.05: no longer possible to print reports directly from the DA.M in Recalculate Last Result mode
Problem:
Beginning with C.01.05, the Last Result Mode was converted to the “read-only/review” mode which means it was no longer possible to generate a report after changing the DA.M. The C.01.05 release introduced the possibility of saving the DA.M changes to a master method.
Temporary Solution:
n/a
Fix Information:
This workflow has been added back in OpenLAB CDS ChemStation C.01.06.
Keyword: Data Analysis
One-line Description:
Report template not found when reprocessing with a master method
Problem:
When a result set is analyzed with a master method, i.e a method from the method path, it will use a report template from the result set. The template name is taken from the method. Since the template is not copied to the result set automatically, no report will be created.
Temporary Solution:
N/A
Fix Information:
That is the intended workflow. Users who intend to use master methods might want to use report templates that are not part of the result set. The results of the analysis with different methods can be compared in a common template in the Review mode.
Keyword: Data Analysis
One-line Description:
Snapshot runs cannot be integrated manually
Problem:
Under specific circumstances it is not possible to integrate a snapshot run manually. The manual integration buttons are greyed out and the menu is also greyed out. This issue occurs if a sequence container is loaded while the snapshot is taken in off-line ChemStation.
Temporary Solution:
The recommended workflow is: - Double-click the single run node - Take the snapshot If you are in a situation where you encounter this issue: - Double-click the single run node - Double-click the run with file name SNAPSHOT.D The manual integration events are available, then.
Fix Information:
Fixed in OpenLAB CDS ChemStation C.01.04.
Keyword: Data Analysis
One-line Description:
ChemStation CDS client appears to freeze after hiding a save as dialog
Problem:
The following problem affects OpenLAB CDS Client (ChemStation) rev. C.01.03 or higher. The following workflow lead to a situation where a ChemStation terminal server client session appears to be frozen, where it is not: Whenever downloading a result set from OpenLAB ECM to OpenLAB CDS ChemStation a warning message pops up in case where the same result set is already stored on disk. A message pops up informing the user to save result set with a different name, as it does already exist on the client. After pressing o.k., the SAVE AS dialog comes up and then it may jump behind the ChemStation window to become invisible. This happens when either clicking the ChemStation window while the Save As dialog is open or if the previous info box got committed by a double-click. The second click brings the ChemStation window to the foreground. While the 'save as ' dialog is waiting for a user's input the ChemStation appears to be frozen during the ECM download operation. Even pressing the ALT+TAB keys does not allow accessing the SAVE AS dialog.
Temporary Solution:
If the ChemStation windows was not maximized users could move the foreground window aside and regain access to the 'save as' dialog. If the ChemStation window was maximized: - open the Windows task manager on the CDS client - Select the current instrument under the applications tab - Press the SWITCH TO button This brings the SAVE as dialog back to the foreground.
Fix Information:
Fixed in OpenLAB CDS ChemStation C.01.06.
Keyword: Data Analysis
One-line Description:
Sample parameters for sequence runs are not copied to a self-assembled result set
Problem:
When a self -assembled result set is created, Multiplier and Dilution factor and sample amounts are not copied to the new result set. This is the intended behavior. Self-assembled result sets can contain also single runs. Therefore, the sequence table cells cannot be populated anyway. The creation of these virtual sequences is supposed to simulate a result set without a history.
Temporary Solution:
N/A
Fix Information:
No change planned.
Keyword: Data Analysis
One-line Description:
Relative Retention Time calculation differs from the EP definition.
Problem:
ChemStation's Relative RetentionTime_EP calculation uses a formula different from the EP definition. The formula should be (RT-t0) / (RT(ref)-t0) ChemStation calculates it as: (RT-t0) / RT(ref)
Temporary Solution:
This issue is resolved with Hotfix 06 for C.01.04 and Hotfix 06 for C.01.05.
Fix Information:
Fixed in OpenLAB CDS ChemStation C.01.06.
Keyword: Data Analysis
One-line Description:
Japanese ChemStation: Intelligent report template settings are not getting saved
Problem:
In Japanese version the intelligent report setting selected from Specify Report option is not getting saved. It becomes blank after each single run, the intelligent report has to be selected again under "Specify Report" tab after each single run analysis.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS ChemStation Edition C.01.07.
Keyword: Data Analysis
One-line Description:
AIA export/import for MS-data does not work
Problem:
AIA-export of MS-Data with default parameter settings fails (with 'invalid paramter' error). Once default parameters of AIA-export for MS-data are modified, later AIA-import of these previously exported MS-data fails (with 'can't get the variables from the AIA file').
Temporary Solution:
n/a
Fix Information:
No fix planned.
Keyword: Data Analysis
One-line Description:
Selection of signals does not work in Japanese ChemStation
Problem:
The dropdown menu that allows selection of one signal in multi-signal data does not work in Japanese ChemStation. When a signal is selected, an error occurs (parameter3, the name given was not found) and the software becomes unresponsive.
Temporary Solution:
Hotfix 01 for revision C.01.06 addresses this problem.
Fix Information:
N/A
Keyword: Data Analysis
One-line Description:
Negative spike can appear in some zoom range
Problem:
The smoothing algorithm in the display can cause artifacts for some curves in certain zoom ranges. Chromatograms acquired with a low data rate are likely to be affected.
Temporary Solution:
Disable the smoothing algorithm: 1. Create a backup copy of the file C:\Windows\ChemStation.ini 2. Open the file C:\Windows\ChemStation.ini with notepad.exe 3. Add the following text "GGL_GDIP=0"(without the double quotes) in the section called [PCS] example: [PCS] GGL_GDIP=0 UIFont1=Arial
Fix Information:
No fix planned
Keyword: Data Analysis
One-line Description:
Plate number and Resolution of negative peaks is not calculated
Problem:
In OLCDS Chemstation plate numbers and resolution are not calculated for negative peaks . This is done intentionally.
Temporary Solution:
N/A
Fix Information:
Evaluation of negative peaks has been added functionality in ChemStation C.01.07 SR2 for the interactive tool and the extended performance report.
Keyword: Data Analysis
One-line Description:
IR Sequence Summary Report is always created
Problem:
A classic sequence summary report is omitted when the Sequence or the run time checklist of all methods has Data Analysis disabled. An IR sequence summary report will always be created, as soon as a sequence.acaml file is present in the sequence container. In this case the report will only contain acquisition information.
Temporary Solution:
If no report is required, do not select the option in the Sequence Outout tab.
Fix Information:
No change planned
Keyword: Data Analysis
One-line Description:
Calibration time stamps may differ by a few seconds
Problem:
Due to internal update calls there may be more than one internal entry for one calibration update event. This results in time stamps on reports that may differ by up to five seconds. This is problematic in custom reports where results are grouped by or referenced to one calibration.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS C.01.07 SR2. Resolved in OpenLAB CDS C.01.07 SR1.
Keyword: Data Analysis
One-line Description:
Bracketing calibrations are not performed when runs other than samples are used in a sequence
Problem:
The documented procedure to run a sequence with bracketing involves only the calibration runs in the first sequence line(s) and the samples that are bracketed in the defined intervals. Users that want to include blanks and control runs outside of the bracket have two undocumented options: 1) Separated methods After the line(s) that define the bracket there can be blanks, calibration runs and control samples with a different acquisition method. The running sequence will execute those runs first, followed by the run with the bracketing method. The update of the methods with the most current calibration method can be done from data analysis view. This option works as expected. 2) Dummy Calibration runs The other option was to have the same acquisition method all over the sequence. All the additional blanks and controls are disguised as calibration runs with RF set to No Update. Also those dummy runs are executed first. The disadvantage here is that a Control needs to be evaluated manually. This second option does not work as expected in C.01.06 and C.01.07. When a bracket calibration should be executed, the software switches to Replace calibration. It is documented in the run log. Note: Easy Sequence does not support this workflow.
Temporary Solution:
The second option can still be realized, but the sequence table must be written differently. It must be written in its rolled-out state. That is to say that every singe injection needs to be scheduled by its own line in the sequence table. The entries in the fields inj/location must be 1. The Interval column is not used. The cells for the RF Update must be set to Bracket.
Fix Information:
Resolved in revision OpenLAB CDS ChemStation C.01.06 Hotfix03. Resolved in revision OpenLAB CDS ChemStation C.01.07 SR2.
Keyword: Data Analysis
One-line Description:
6890 GC data acquired with ChemStation B.04.03 cannot be loaded with ChemStation C.01.07
Problem:
B.04.03 allowed bad target mass strings to be put in the sample.mac and migration to sample.xml throws exception: "Caught System.FormatException, mscorlib,Version=2.0.0.0, Culture=neutral, PublicKeyToken=b77a5c561934e089 inUpdateChromRegWithRefDataFilePaths= Input string was not in correct format". Strange character is shown in the vial location.
Temporary Solution:
Addressed with Hotfix 01 for ChemStation C.01.07 SR2.
Fix Information:
Fixed in revision C.01.07 SR3.
Keyword: Data Analysis
One-line Description:
Display of fraction annotation in chromatogram depends on the signal options
Problem:
The fraction annotations in the purify task are only displayed for the multi chromatogram settings Separate and Full Scale. The green frames are not displayed for the other combinations. This affects the chromatogram window in Data Analysis view and the function "Print Selected Window". The annotations in classic reports are printed into the chromatograms.
Temporary Solution:
Addressed with Hotfix 01 for ChemStation C.01.07 SR2
Fix Information:
Fixed in ChemStation C.01.07 SR3.
Keyword: Data Analysis
One-line Description:
Fractions in MSD traces are only annotated for the first signal
Problem:
Fraction annotations are only drawn in Data Analysis view for fractions that are triggered from the MSD1 signal. A fraction that is triggered from the other available traces is not annotated on the screen.
Temporary Solution:
N/A
Fix Information:
Resolved in revision C.01.07 SR3.
Keyword: Data Analysis
One-line Description:
Cannot read Spectra (*UV) files from OpenLab CDS 2.x
Problem:
Generated some DAD signals plus UV data in OpenLab CDS 2.3 and converted to ChemStation format. The UV data could not be read in the data analysis view.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.09 Update 1. Resolved in OpenLab CDS ChemStation C.01.10.
Keyword: Data Analysis
One-line Description:
Reprocessing in online session automatically paused after third (3rd) line
Problem:
A few online sequences were edited during acquisition. The edited sequences were reprocessed in the Data Analysis View and gets automatically paused after the third line. Seems every reprocessing in the online instance gets paused after the 3rd line. There is no data loss is involved; data can be reprocessed in the offline copy.
Temporary Solution:
Switch to a Reprocessing copy or restart on-line ChemStation.
Fix Information:
n/a
Keyword: Data Analysis
One-line Description:
Data Analysis remains busy when trying to generate a report
Problem:
A datafile was created with the G7161A prep pump. Loaded this datafile in Data Analysis and checked all pump specific instrument curves. When printing this data file to screen, the chemstation remains busy for about 40 minutes.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS ChemStation C.01.08.
Keyword: Data Analysis
One-line Description:
ChemStation stopped working after deleting UIB signal in Data Analysis
Problem:
In Data Analysis Purify window, click on the icon "Delete object from Chromatography display". From the signal dropdown box, select UIB1 B, Analog In and then click with the mouse cursor on the signal in the displayed chromatogram. At the bottom left line in ChemStation there was a red cross displayed with a text "Parameter 3: A value is outside the legal range". Since then, no signal in chromatogram of any other detector could be displayed, the chromatogram was simply empty. When clicking inside the empty chromatogram for UIB again, ChemStation crashed. *The issue is also reproducible in offline session. * Open the provided data file in offline ChemStation. * Go to Purify and click on icon "Delete object from Chromatography display". * From the signal dropdown box select UIB B, Analog In (last signal). * Click inside the chromatogram twice and the crash occurs.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.07 SR4 Hotfix 1. Resolved in OpenLab CDS ChemStation C.01.08.
Keyword: Data Analysis
One-line Description:
Data Analysis remains busy when trying to generate a report
Problem:
A datafile was created with the G7161A prep pump. Loaded this datafile in Data Analysis and checked all pump specific instrument curves. When printing this data file to screen, the chemstation remains busy for about 40 minutes.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.07 SR4 Hotfix 1.
Keyword: Data Analysis
One-line Description:
Pressure signal is not shown when "high pressure" is selected in Data Analysis
Problem:
From Data Analysis, open the acquired data and check "high pressure" in the instrument curve tab. The is no pressure signal shown. Also, in the method editing page, the QuatPump pressure signal shows "25ms (bar)", which is not correct; it should be "pressure (bar)".
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.08.
Keyword: Data Analysis
One-line Description:
Macro error prevents 2DLC DA plugin from loading and EIC extraction fails when Instrument Curves loaded
Problem:
In an LC/MS system, load a 2DLC data file in Data Analysis. A second data analysis tab named '2D-LC Viewer' should be available but is not shown. This prevents EIC extraction when Instruments Curves are also loaded. The is not reproducible with LC alone.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.09 and OpenLab CDS ChemStation Edition C.01.08 Update 2.
Keyword: Data Analysis
One-line Description:
Cannot view Isoabsorbance Plot for DAD2, DAD3 detector
Problem:
In case more than one DAD detector is used, it is not possible to select the Isoplot for the different DAD detectors.
Temporary Solution:
Workaround: 1.) Backup your DAD1.UV signal 2.) Rename DAD2.UV or higher to DAD1.UV 3.) Open the Isoabsorbance Plot in OpenLab CDS ChemStation Edition to view the signal
Fix Information:
Fixed in C.01.09 Update 1
Keyword: Data Analysis
One-line Description:
G7117 detector data does support calculated purity threshold function with rev. C.01.07
Problem:
Some detector drivers (e.g. G7117, G7115) provide data may not be properly handled by the peak purity function of ChemStation C.01.07 and earlier. The "calculated threshold" or "fixed threshold" may not be used. Instead the peak controlled spectral threshold will be used for calculation and shown in the user interface and the report..
Temporary Solution:
n/a
Fix Information:
Fixed with C.01.08
Keyword: Data Analysis
One-line Description:
Loading data takes 20 times longer
Problem:
Loading data from B.04.03 takes 20 times longer when compared to C.01.08.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.09 Update 2. Resolved in OpenLab CDS ChemStation C.01.10.
Keyword: Data Analysis
One-line Description:
Split peak on MS data generates error message stating "NO peak detected"
Problem:
No peak detected when using split peak and trying to generate a report using classic report with MS Data.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.09 Update 2. Resolved in OpenLab CDS ChemStation C.01.10.
Keyword: Data Analysis
One-line Description:
Fraction Data from classic cluster configurations cannot be loaded
Problem:
Fraction Data from classic cluster configurations cannot be loaded.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.09 Update 2. Resolved in OpenLab CDS ChemStation C.01.10.
Keyword: Data Analysis
One-line Description:
ChemMain crashes when reprocessing with method containing huge number of events in audit trail
Problem:
By design, the older integration events are not deleted from the method. When there are a huge number of manual integration events saved in the method by making run by run adjustments will cause the ChemMain crash during reprocessing.
Temporary Solution:
The recommended workflow is to store the manual integration events with the data.
Fix Information:
Method cleanup macro added to the UCL included in C.01.10 Update 3 for methods containing huge number of events in the method audit trail. The macro creates a backup of older manual integration method audit trail entries and deletes them to decrease the method history table size.
Keyword: Data Analysis
One-line Description:
RX files missing in the resultsets acquired with MS instrument
Problem:
Writing acaml files to "C:\Windows\system32\config\systemprofile\AppData\Local\Agilent Technologies\ can sporadically fail. Hence it is recommended to add this path to the Anti Virus exclusion list.
Temporary Solution:
Add "C:\Windows\system32\config\systemprofile\AppData\Local\Agilent Technologies\ to Anti Virus exclusion list.
Fix Information:
n/a
Keyword: Data Analysis
One-line Description:
ChemStation crash during delay calibration if instrument name is too long
Problem:
If the instrument module name is too long, then ChemStation may crash at startup. ChemStation could be running slowly and could crash during delay calibration.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab ChemStation LTS 01.11.
Keyword: Data Analysis
One-line Description:
ADFExport for ChemStation result may contain invalid instrument description
Problem:
An export of ChemStation sequence data in the Allotrope Data Format (ADFExport) for OpenLab CDS ChemStation may not correctly describe the original instrument, if the result set was manually assembled from individual data files that were acquired by different instruments. The original instruments of all individual data files except the first are not included in the resulting data description. As a workaround, avoid ADFExport of self-assembled results sets or make sure the first selected data file was acquired with the desired instrument. It is recommended to only assemble data files into a new result set that were acquired with the same instrument.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: Data Analysis
One-line Description:
Character missing in signal label when loading data imported from OpenLab CDS 2.x
Problem:
When OpenLab CDS 2.x raw data is loaded in Chemstation, the signal name is missing the first character. For example: "Wavelength" is shown as "avelength" "Sig" is shown as "ig"
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.10 Update 7. Resolved in OpenLab ChemStation LTS 01.11.
Keyword: Data Analysis
One-line Description:
Macro Verftest.mcx for Verftest will not run on AIC instrument 10
Problem:
The verftest for instrument 10 offline data analysis will not run the test. The other 9 instrument verftest completed without any problem.
Temporary Solution:
The purpose of the mentioned test it to verify the correct installation of the entire Data Analysis chain. The test does not need to be repeated on every instrument, but only once per instrument technique (GC, LC, CE, ..), if applicable.
Fix Information:
n/a
Keyword: Data Analysis
One-line Description:
Column length value in the report has the wrong value
Problem:
The column length displayed in the report printed from the data analysis integration interface is incorrect.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab ChemStation LTS 01.11.
Keyword: Data Analysis
One-line Description:
System suitability not calculated for some peaks when using tail tangent skim timed event.
Problem:
System suitability not calculated for some peaks when using tail tangent skim timed event.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.10 Update 7. Resolved in OpenLab ChemStation LTS 01.11.
Keyword: Data Analysis
One-line Description:
Un-zooming chromatogram does not go back to the defined scale settings
Problem:
If a fixed response range is set, un-zooming does not go back to the scale settings defined in the graphical options.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab ChemStation LTS 01.11 Update 01.
Keyword: Data Analysis
One-line Description:
CP Parsing error when using Recalibrate from Quantitation icon in the System Diagram
Problem:
In an LC ChemStation, recalibration in the Method and Run Control View does not work. Right-mouse click on the Quantitation icon in the system diagram opens a context menu that offers “Recalibrate…”. Choosing this menu item causes a CP parsing error on ChemStation’s message line.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab ChemStation LTS 01.11 Update 01.
Keyword: Data Analysis
One-line Description:
LTS 01.11 Chinese version: "Not a string" error shown during DAD1A signal overlay
Problem:
In LTS 01.11 Chinese version, "Not a string" error is shown during DAD1A signal overlay.
Temporary Solution:
n/a
Fix Information:
A point patch for rev. LTS 01.11 Update 01 is available on request,
Keyword: Data Analysis , Programming
One-line Description:
Garbage characters can appear in English LCMS reports
Problem:
Some sample parameters from LCMS data files (like e.g. Operator Name) are not printed properly. Instead unreadable text appears, like a mix of Chinese and Japanese characters. Those parameters are incorrectly extracted from the data.ms files. If the data file also contains DAD traces, the sample parameters are taken from there. There is still the risk that other applications access the incorrect values from the LCMS objects.
Temporary Solution:
Fix planned for C.01.03.
Fix Information:
Fixed in C.01.04.
Keyword: Documentation
One-line Description:
HTML Help requires Internet Explorer revision 6.0.28.00.1106CO or higher
Problem:
The HTML Help requires Internet Explorer Rev. 6.0.2.800.1106CO or higher. If a lower IE revision is used, it is possible that executing the 'Reference' button on the top of the Help screen gives an error message: '...a Script Error ...'
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: Documentation
One-line Description:
M8301-96025 Japanese references manual has wrong formula for PeakValleyRatio
Problem:
In the Japanese manual (M8301-96025) the formula for the Peak/Valley ratio on page 37 is wrong: H1 = H2, Peak valley ratio = H1/Hv (this is wrong. It should be H2/Hv) and H1 < H2, Peak valley ratio = H1/Hv This is a misprint in the manual. The formula is used correctly in the software.
Temporary Solution:
For the correct formula please consult the reference manual in the ENU folder. The formula und graphics can be found on page 37.
Fix Information:
Resolved in OpenLAB CDS ChemStation C.01.07 SR2.
Keyword: Documentation
One-line Description:
Insufficient description of regression coefficient in the OpenLAB CDS ChemStation reference manual
Problem:
The OpenLAB CDS ChemStation reference manual states: "The correlation coefficient is the square root of the regression coefficient and gives a measure of the fit of the calibration curve between the data points." The term regression coefficient is not correct in this context and should be replaced by coefficient of determination, known as r square.
Temporary Solution:
The calculation of the correlation coefficient r is formally described in chapter "Evaluating System Suitability" and shall be used instead of the verbal description.
Fix Information:
This will be corrected in a later issue of the reference manual.
Keyword: Documentation
One-line Description:
Peak Model Retention Time used but not saved or documented
Problem:
For system suitability calculations, the ChemStation uses the true peak width at various heights and a retention time evaluated by a peak modelling algorithm. The retention time that is visible in all reports is the integrator RT. This value is used for peak identification (chapter 3) and in recalibration. The peak model algorithm considers the peak shape and is more suitable for the calculation of peak performance parameters. The value is exclusively displayed in the Interactive Peak Performance tool. The Peak Model's Apex may vary slightly to the Integrator's RT and is displayed in the interactive peak performance tool user interface.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS ChemStation C.01.08.
Keyword: Documentation
One-line Description:
Wrong description about “Correlation coefficient” in Reference Manual
Problem:
Definition of “Correlation coefficient” should be "A correlation coefficient is displayed with the graphic of the calibration curve. The correlation coefficient is the square root of the coefficient of determination when liner fitting by least squares method, and gives a measure of the fit of the calibration curve between the data points." and not "A correlation coefficient is displayed with the graphic of the calibration curve. The correlation coefficient is the square root of the regression coefficient and gives a measure of the fit of the calibration curve between the data points." as indicated in the CDS_CS-references manual.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.09
Keyword: Documentation
One-line Description:
ChemMain.exe crash and the LCMS instrument window closes automatically
Problem:
Online instrument window is automatically closing after the first run of the sequence.
Temporary Solution:
When you see rendering artifacts such as ChemMain.exe crashing in an application, try turning off hardware acceleration. If the artifact disappears, the problem might be with your video driver. The disable hardware acceleration option is a DWORD value that is either 0 or 1. A value of 1 disables hardware acceleration. From the Registry Editor, go to Computer\HKEY_CURRENT_USER\Software\Microsoft\xxx.Graphics The DisableHWAcceleration needs a DWORD value of 1 to turn off the hardware acceleration. This information has been added to the user documentation.
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.10 User Documentation.
Keyword: Documentation
One-line Description:
OpenLab Chemstation 1.10 Release notes sets incorrect Allotrope expectations
Problem:
In the OpenLab CDS Chemstation Edition Release Notes under ADFExport Plug-in is a ADFExport revision, product numbers and an expected release date mentioned. This information is misleading and should be ignored.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab ChemStation LTS 01.11.
Keyword: Documentation
One-line Description:
ChemMain stopped working and other instabilities due to Adobe Acrobat Reader and updates pushed by Adobe
Problem:
Only the Adobe Acrobat Reader Classic track is supported by ChemStation and runs in a 'sandbox' protected mode with the option of no automatic updates. Adobe Acrobat Reader Continuous track is not supported as it pushes updates on a regular basis which will interfere with ChemStation operation and result in ChemMain failures. The following error: "There is a problem with Adobe Acrobat/Reader. If it is running, please exit and try again. (523:523)" or "...(507:507)" occurs if Adobe Acrobat Reader settings are not configured for ChemStation or by a virus checker. There are installations of Acrobat Reader versions that push automatic updates and need to be uninstalled. To prevent Adobe from reaching out to the Web, there are other notification channels and messages that need to be disabled. These settings are ON by default at installation and must be changed manually. Also, some AntiVirus applications have an issue with adobe reader in our setup and exclusions are needed. In Windows Application event log, the faulting module names for Adobe Acrobat Reader are: AcroPDFImpl.dll_unloaded and AcroPDFImpl.dll.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.10 Update 3. Adobe Acrobat Reader recommended updates to improve ChemStation stability included in documentation.
Keyword: Documentation
One-line Description:
After Any Data Modification is selected – not all restrictions are listed in Help
Problem:
The following functionally are disabled when “After any data modification” (full data tracking) is selected in the ChemStation Administration Tool: • Load Signal • Manual Integration when overlaid signals come from different result sets • Reprocessing Copy menu • Snapshot (present in Offline mode only) menu • Save report files to local disk The following are missing from Help: • Reprocessing Copy menu • Snapshot (present in Offline mode only) menu • Save report files to local disk
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab ChemStation LTS 01.11.
Keyword: Documentation
One-line Description:
Automation Interface documentation updates were not included for LTS 01.11
Problem:
The C.01.10 version of the Automation Interface documentation was released with LTS 01.11.239.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab ChemStation LTS 01.11 Update 01.
Keyword: Easy Sequence
One-line Description:
Easy Sequence: Running back table headspace seq, select front table, add sample name, click back table and select cell causes seq to hang
Problem:
Submitted a sequence via Easy Sequence. If the sequence table is opened during a running sequence and the user clicks on a cell and clicks ok, there is a lock placed on the selected sequence line. The system now waits when it reaches the line where the cell was selected. The headspace doesn't put a vial in the oven and remains ‘waiting’. The Chemstation never continues past the locked vial. The system must be aborted. The error is not cleared until the chemstation is shutdown and restarted and the GC and headspace hardware are power cycled.
Temporary Solution:
n/a
Fix Information:
Fixed in OpenLAB CDS C.01.06.
Keyword: Easy Sequence
One-line Description:
Error message when Blank with Vial is entered in Easy Sequence Setup
Problem:
When a blank is defined in the Calibration tab of the Easy Sequence Setup, it will result in an error, if a vial location is specified. Error Object reference not set to an instance of an object The template cannot be saved. Blank without injection can be used without an error. The same problem occurs for control samples.
Temporary Solution:
Leave the vial location empty. and save the template. In the Easy Sequence tab use Fill samples to populate the sequence. Enter a vial number in this screen.
Fix Information:
n/a
Keyword: Easy Sequence
One-line Description:
Vial values not accepted for 96 well plate tray in Easy Sequence Setup
Problem:
Valid vial values not accepted for a LC instrument with 96 well plate tray in Easy Sequence Setup.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS ChemStation C.01.07.
Keyword: Easy Sequence
One-line Description:
Easy Sequence is not added to the Run Queue when resources are insufficient
Problem:
Easy Sequence is not added to Run Queue when resources are insufficient. There is no message indicating what resources are insufficient.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab ChemStation LTS 01.11.
Keyword: ECM Integration
One-line Description:
Path definition in OpenLab Control Panel is not used in ChemStation
Problem:
The storage path that is entered in the control panel is not supposed to be the default storage location in ChemStation. The storage path is related to EZChrom CDS projects.
Temporary Solution:
None
Fix Information:
No fix planned.
Keyword: ECM Integration
One-line Description:
The storage path for files in ECM is user-specific
Problem:
The LCDF storage path for saving methods to ECM is user specific.A user that logs on to ECM from different instruments will always save to "his" path by default. While this is the intended behaviour, it can be confusing. Method developers would prefer a workflow that makes the path instrument specific.
Temporary Solution:
Check the folder path before uploading methods. Consider including the instrument name in the folder name.
Fix Information:
Fix considered for a future revision.
Keyword: ECM Integration
One-line Description:
Cannot switch current ECM authentication to a different ECM server
Problem:
On a stand alone installation, the ECM authentication mode has been set. Due to a hardware problem (or something else), the current ECM authentication mode must be switched to a different ECM server with a different account. When the user tries to change the ECM authentication to a different ECM server and enters the new hostname, the administrator cannot be selected. The user is back on the previously configured ECM system.
Temporary Solution:
A workaround is to set authentication mode to none and then go to the new ECM server.
Fix Information:
N/A
Keyword: ECM Integration
One-line Description:
it is not possible to remove AIC from networked workstation
Problem:
If a workstation is removed from a networked system, the AIC remains in OpenLAB Shared Services and cannot be removed.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: ECM Integration
One-line Description:
ECM user licenses not released when user is changed in OLCDS ChemStation
Problem:
A user change in ChemStation does not trigger the release of ECM user licenses. The system could run out of licenses.
Temporary Solution:
Fixed in C.01.05 Hotfix 02. Fixed in C.01.04 Hotfix 05.
Fix Information:
Fixed in OpenLAB CDS C.01.06.
Keyword: ECM Integration
One-line Description:
In file information of a datafile the Operator name shows as "Disconnected"
Problem:
The Acquisition Operator name in the file information of Datafile shows as " Disconnected" when a user change is performed right before an injection.
Temporary Solution:
In IR templates the Submission Operator can be reported.
Fix Information:
n/a
Keyword: ECM Integration
One-line Description:
Submitting multiple sequences to queue, the User token in upload path is resolved to current user - not to submit user
Problem:
When submitting multiple sequences to the sequence queue, each sequence data path containing the <user name> token will be resolved to the currently logged in user.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS C.01.07 SR2. When submitting multiple sequences to the sequence queue, each sequence data path containing the <user name> token will be resolved to the user that submitted the sequences.
Keyword: ECM Integration
One-line Description:
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
Problem:
By OpenLAB ChemStation Edition C.01.04 the requirement for ECM's remote data path (LCDF) became more strict than before. If for editing the Sequence Parameters dialog is opened, and a backslash at the LCDF's begin is missing, the builtin check of this dialog-box will demand a manual correction by the user. This can happen especially once loading a sequence template file from an earlier version of OpenLAB ChemStation Edition. The later release of OpenLAB ChemStation Edition (like e.g. C.01.05) will tolerate and correct the missing leading backslash at ECM's remote datat path (LCDF) in Sequence Parameters dialog automatically. Execution of sequences by OpenLAB ChemStaation Edition is not affected. I.e. sequence templates do NOT have to be corrected, neither for sequence execution nor for sequence reprocessing.
Temporary Solution:
Leaving the opened Sequence Parameters Dialog by OK is interdicted until all parameters are in valid ranges. This now for OpenLAB ChemStation C.01.04 requires the remote data path (LCDF) for ECM to start with a leading backslash. If at some sequence template (e.g. from an earlier version of OpenLab ChemStation Edition) this character was missing, it now has to be inserted. Only then the Sequnece Parameters dialog can be successfully closed and left with OK-button.
Fix Information:
n/a
Keyword: ECM Integration
One-line Description:
Single runs in self-assembled data sets cannot be integrated
Problem:
When runs from outside are added to a result set (self-assembled data set), these runs cannot be manually integrated in a compliant workflow. The ECM conditions are "Upload after every modification". As this additional run is not part of the original result set, it cannot be uploaded as a new version of the sequence. Therefore no modification is allowed.
Temporary Solution:
Migrate the self-assembled data set to a self-assembled result set. This will be uploaded as version 1 of a new result set.
Fix Information:
This is the intended behaviour. No change planned.
Keyword: ECM Integration
One-line Description:
Instrument access rights are not enforced after user change
Problem:
OpenLAB Shared Services allows setting access privileges per user and instrument. Users without access rights cannot start an instrument. During a user change it is, however, possible for a user without the privilege to log on to an open session.
Temporary Solution:
This issue is resolved in OpenLAB CDS ChemStation C.01.03 Hotfix 09, C.01.04 Hotfix 06 and C.01.05 Hotfix 04.
Fix Information:
Fixed in OpenLAB CDS ChemStation C.01.06.
Keyword: ECM Integration
One-line Description:
Incomplete Result Set package (sc.ssizip) was sent to ECM: versions do not contain the data files
Problem:
Incomplete Result Set package (sc.ssizip) was sent to ECM. The incomplete ssizip package is caused by a specific character in the method name (e.g. DLX-ENANTIOMERIC-33℃.M). The degree symbol (℃) in the method name causes the Dynazip compression tool to fail and stop, resulting in an incomplete package uploaded to OpenLAB ECM. As long as the ChemStation is not closed, the local copy was available and was reprocessed to create additional versions. Each of these versions also was missing data since the compression failed again on the special character. The option "Cleanup Data on Shutdown" was selected, so the local copy of the data was also deleted when shutting down the ChemStation. In such a case, the data was lost and can't be recovered. The following error was observed: DynaZip UnZIP Error: Error creating output file.
Temporary Solution:
Special character usage in file names must be avoided. Refer to the Concepts Manual.
Fix Information:
Fixed in OpenLAB CDS C.01.06.
Keyword: ECM Integration
One-line Description:
Instrument access rights are not enforced after user change
Problem:
OpenLAB Shared Services allows setting access privileges per user and instrument. Users without access rights cannot start an instrument. During a user change it is, however, possible for a user without the privilege to log on to an open session.
Temporary Solution:
n/a
Fix Information:
This issue is resolved in OpenLAB CDS ChemStation C.01.03 Hotfix 09 and C.01.05 Hotfix 04. The issue will be resolved in the next minor revision following C.01.05.
Keyword: ECM Integration
One-line Description:
OLCDS Chemstation - DynaZip Zip error in ECM Queue management
Problem:
ECM queue management shows an entry “Unable to upload <~~XXX.D.SSizip> The system cannot find the file specified.(error -27) and when clicked process queue getting DynaZip Zip error”.
Temporary Solution:
The Dynazip error is a consequence of something else that went wrong first. Typically trying to upload a result where the path either does not exist, or exceeds the maximum path length. Avoid these situations.
Fix Information:
Fixed in OpenLAB CDS ChemStation C.01.06.
Keyword: ECM Integration
One-line Description:
Result Sets with a % in the folder name cannot be uploaded to backend storage
Problem:
Unique folder names must not contain the % character. The upload to Data Store fails with a message "Connection to Remote Storage is broken"
Temporary Solution:
N/A
Fix Information:
N/A
Keyword: ECM Integration
One-line Description:
ECM upload may fail with warning: Could not read VCS insformation for 'C:\Chem32\1\VcsCache\
Problem:
Under some rare circumstances the upload of result sets to ECM may fail with error message "Could not read Vcs information for 'C:\Chem32\1\VcsCache\ShouldBeHidden\<file identifier> . This error can be observed in cases where the file got uploaded successfully, but the upload command file did not get deleted. The system generated deletion of a file may happen if the file is currently accessed by another process or the PC is shutdown unexpectedly. Real-time protection and on-access scanning anti-virus software is the most likely cause for this unexpected file access.
Temporary Solution:
The likelihood of this error can be reduced by introducing controlled shutdown procedures for Workstations and Acquisition Controllers and by excluding the \\chem32 installation directory from "on-access" virus scanning - as generally recommended for OpenLAB CDS ChemStation. Virus scanning should be done on a scheduled basis while the OpenLAB CDS ChemStation application is closed.
Fix Information:
There is currently no design change planned.
Keyword: ECM Integration
One-line Description:
incomplete versions of result sets are uploaded to the back end storage.
Problem:
Certain situations are possible where incomplete versions of result sets are uploaded to the backend storage. This affects all ChemStation versions with ECM or Data Store. These situations are triggered by the way that online and offline versions of the same instrument are aligned. A workflow is likely to provoke such a situation, when the same result set is loaded in both sessions at the same time.
Temporary Solution:
Avoid situations that trigger a forced upload. Save all methods, sequences and manual integrations, both in offline and online, before closing either session. Or in other words: Don't close one of the sessions, while the upload barrel in either of the sessions is still yellow. Hotfix 08 for CHemStation C.01.05 addresses this problem.
Fix Information:
Fixed in OpenLAB CDS C.01.06.
Keyword: ECM Integration
One-line Description:
File permissions set incorrectly after backup on OpenLAB Secure Workstation
Problem:
Under certain conditions a situation can occur where the file permissions are not applied correctly after a backup of a Secure WorkStation. The following scenario leads to a situation where non-admin users can delete files: After a disk failure the system was restored from an image of the operating system. The OpenLAB CDS Secure ChemStation Workstation has been reinstalled, the folder structure for Data Store is created manually. After that the backup from Data Store is restored. In that case the permissions for the files is not propagated properly.
Temporary Solution:
A script is available on disk 2 of the OpenLAB installation media. Script and documentation to apply the correct file permissions are located in the Support\Tools folder.
Fix Information:
n/a
Keyword: ECM Integration
One-line Description:
ECM data path status in ChemStation C.01.05 (41) HF6 is not always updated correctly
Problem:
Subject: ECM data path status in ChemStation C.01.05 (41) HF6 The workflow on the installed Networked Workstation system is the following: The data acquisition is running on the ChemStation workstation. This means that the displayed status of the ECM data path is irrelevant for this. Then once the sequence is ready and the data file is completed the connection to the ECM-server is re-established and the data transfer will start. As we know, the upload of the data to ECM is working without issue so it can be concluded from this that all is working correctly in the background. It is just the UI that isn’t being updated correctly.
Temporary Solution:
The validity of the measured data is not at risk by this update problem! The status update also does not cause any kind of data loss.
Fix Information:
No timeframe is currently available for any code change to address the status issue.
Keyword: ECM Integration
One-line Description:
ChemStation with no access to any menu or icon.
Problem:
Distributed System C.01.06 plus ECM 3.4.1 ChemStation with no access to any menu or icon. A different window remains open behind current ChemStation and user cannot access it.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS ChemStation Edition C.01.06 HF01. Resolved in OpenLAB CDS ChemStation Edition C.01.07.
Keyword: ECM Integration
One-line Description:
ECM path not recognized in preferences after upgrade
Problem:
After an upgrade of the OpenLAB CDS software it is recommended to check the server path in the preferences (transfer settings). There have been cases where the LCDF path was not recognized. An error tool tip stated: " is not a valid target path. although the path in the dialog was valid.
Temporary Solution:
Re-enter the path and OK the dialog.
Fix Information:
No fix planned.
Keyword: ECM Integration
One-line Description:
Sequence name using special characters does not upload to ECM
Problem:
In Preferences, special characters are used in the sequence name pattern, for example ECM-seq3#$%19-19-15. Run the sequence. The sequence completes and can be seen in tree view in chemstation. The sequence does not upload to ECM and remains in the queue.
Temporary Solution:
Special characters cannot be used in the sequence name.
Fix Information:
No Fix Planned.
Keyword: ECM Integration
One-line Description:
Wrong take over message can be displayed in a Distributed System when a user had online and offline session
Problem:
When in a Distributed System one user is launching an online and the corresponding offline session and a second user is taking over the online session, the takeover message for the first user might name the wrong user depending on the system usage rate
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: ECM Integration
One-line Description:
The Upload path can edited for a running sequence
Problem:
When a running sequence is paused, it is possible to access the Sequence Parameters and make changes to the upload path to Data Store or ECM. The changes are not accepted, but the user receives no feedback.
Temporary Solution:
Fixed in C.01.06 HF 03 Fixed in C.01.07 SR1. The sequence path of a paused sequence is greyed out in the Sequence Parameters.
Fix Information:
Resolved in OpenLAB CDS ChemStation C.01.07 SR2.
Keyword: ECM Integration
One-line Description:
Sequence not uploaded to the correct location in backend after user change
Problem:
When the user changes during a sequence run and the upload path contains the user name, the upload path will change as well. The complete result set will upload to the path corresponding to the last user. The paths in reports of earlier runs will not be correct.
Temporary Solution:
Fixed in C.01.07 SR1. The name of that user will be used that submitted the sequence.
Fix Information:
Fixed in C.01.07 SR1. The name of the user that submitted the sequence will be used in the path location.
Keyword: ECM Integration
One-line Description:
Queue maintenance after shutdown could conflict with short inactivity lockout
Problem:
The maintenance of the upload queue (delete after shutdown, waiting for uploads) could take more than 1 minute. A short inactivity lockout can lead to a conflict. The result can be an incomplete shutdown and/or remaining orphan files in the queue.
Temporary Solution:
Increase the inactivity lockout time to 10 min.
Fix Information:
Resolved with C.01.06 HF03. Resolved in OpenLAB CDS ChemStation C.01.07 SR2.
Keyword: ECM Integration
One-line Description:
Runs in a sequence with bracketing can be accessed too early
Problem:
While a sequence is still acquiring, the runs in progress are not yet accessible in data analysis view. The lock on the active run is lifted, when data analysis is finished. This is not true for runs in sequences with bracketing. Samples inside a calibration bracket can be loaded and e.g. manually integrated, before they are ready for processing. There is no mechanism in place to react on the delayed analysis that is used with bracketing.
Temporary Solution:
No change planned. Activating Results Audit Trail will create a record of the manual integration.
Fix Information:
No change planned.
Keyword: ECM Integration
One-line Description:
Sequence/single runs manually uploaded to ECM still shows the asterisk
Problem:
The asterisk will not be removed if the data is manually uploaded by using right-click save data. This action does not use the queue to upload data and is not recommended as it can easily create versions that are out of sync.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: ECM Integration
One-line Description:
Number of function arguments is invalid when loading data from backend
Problem:
After the installation of the C.01.05 SP1 HF2, the following error pops up when trying to download data from data back-end (DA View -> ECM/DS -> Load Data...): Data Store Error << Number of function arguments is invalid >> occurred in Function VC_IsChemStoreMigration Code 41328 Please check your Data Store connection.
Temporary Solution:
The message box can be closed and the loaded data can be processed as usual.
Fix Information:
Resolved in C.01.05 SP1 HF3.
Keyword: ECM Integration
One-line Description:
Files names containing special characters are scrambled after downloading SSIZIP from Data Store / ECM
Problem:
File names containing special characters are internaly renamed during creatiion of the SSIZIP archive. Examples are ü, ß, ñ, æ from the character set of certain languages. A method with such a name can be created, saved, used in acquisitions and in sequence reprocessing without problems. A result set containing the method will upload to the back end storage correctly. After a download it shows that the exotic characters in the file name have been replaced by another character form extended ASCII set.
Temporary Solution:
Use file names containing only characters from the 26 standard character set
Fix Information:
Fixed in OpenLAB CDS ChemStation C.01.07 SR2.
Keyword: ECM Integration
One-line Description:
Shortutl.mac needs modification to show the ECM path
Problem:
Shortutl.mac needs modification to show the ECM path
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.08.
Keyword: ECM Integration
One-line Description:
Notification for data not uploaded to database backend not clear
Problem:
Result sets and single runs get marked with a green icon once the upload to the database backend has been triggered. When the connection to the database backend is lost or broken, the user is notified that single runs and sequences are waiting in the upload queue by a small lightning bolt in right hand corner. Users easily missed this notification that there are runs waiting in the queue. Users would assume the transfer occurred because they see the green icon.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLAB CDS ChemStation C.01.07 SR2. An Asterisk is now added to the green icon and indicates a pending transfer of data (result sets and single runs). This feedback is shown in the Data Tree and the Navigation Table.
Keyword: ECM Integration
One-line Description:
Path Preference issue with C.01.07 SR2 and ECM LCDF-applied role
Problem:
For an ECM user assigned LCDF-based access, the following is noted from an OLCDS instrument session: unable to assign Path Preference in Sequence Parameters. Error: "Transfer settings are not correct. No permissions to write to ECM path" indicated with OLCDS/ECM integration for ECM -> Preferences The same user added to the system-wide ECM role does not exhibit this behavior. This issue has not been noted on a base C.01.07 installation and is specific to C.01.07 SR2.
Temporary Solution:
Fixed in C.01.07 SR2 HF01
Fix Information:
Resolved in OpenLAB CDS ChemStation C.01.07 SR3.
Keyword: ECM Integration
One-line Description:
Sequence upload to ECM leaves an incomplete progress bar, ECM Uploading in Process
Problem:
There have been cases where the progress bar for an upload to ECM did not disappear after the upload was complete. This holds up the progress of the online session. The root cause is a bottleneck in the processing of internal messages. The risk for this situation increases with the size of the sequence and the complexity of reports.
Temporary Solution:
n/a
Fix Information:
Fixed in revision C.01.07 SR3
Keyword: ECM Integration
One-line Description:
It is not possible to remove AIC from networked workstation
Problem:
If a workstation is removed from a networked system, the AIC remains in OpenLAB Shared Services and cannot be removed.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: ECM Integration
One-line Description:
"Error 41092 occurred" in the system activity logs
Problem:
"Error 41092 occurred" can be found in the activity log which usually happens when the sequence upload hangs and it is necessary to do a force shutdown.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: ECM Integration
One-line Description:
LCDF applied role prevents download
Problem:
Roles applied at the LCDF location level, as opposed to the system level, prevents selection of LCDF and files when using ECM -> Load Data from ChemStation Data Analysis.
Temporary Solution:
n/a
Fix Information:
n/a
Keyword: ECM Integration
One-line Description:
ECM Uploading in Process dialog box does not close in offline session after Resultset is uploaded to ECM
Problem:
From the Offline session, while the resultset is being uploaded to ECM, another resultset is loaded from the ChemStation tree view and the ChemStation hangs. The “ECM Uploading in Process” dialog box does not close after Resultset is uploaded to ECM. No other operations in ChemStation can be done and it must be shutdown and restarted. The resultset does upload correctly.
Temporary Solution:
While the updated resultset is being uploaded to ECM, do not load another resultset from the ChemStation tree view . Wait for the upload to ECM to complete before loading another resultset in the tree view.
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.07 SR4 Hotfix 1. Resolved in OpenLab CDS ChemStation C.01.08 Update 1.
Keyword: ECM Integration
One-line Description:
Error occurred after clicking on Save Report Template on ECM tab
Problem:
In the Report layout view, click on ECM Tab -> click on Save Report Template. Error message is displayed after clicking on Save Report Template button.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.08 Update 01.
Keyword: ECM Integration
One-line Description:
No “Connection Error” message displayed if "save data in ECM" function is performed while ECM is disconnected
Problem:
The "save data to ECM" functionality is selected. When the ECM connection is lost and data cannot be uploaded to ECM, there is no "Connection error" message provided to the user.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.09 and resolved in OpenLab CDS ChemStation C.01.08 Update 2
Keyword: ECM Integration
One-line Description:
Not able to configure ECM data path in ChemStation C.01.09 with ECM 3.6 as backend
Problem:
With the CDS ChemStation C.01.09 installer the ECMAPI Wizard of ECM 3.5 is included. Before installing CDS ChemStation C.01.09 with a ECM 3.6 backend, install the ECMAPI of ECM 3.6 as described in the ECM installation Guide of ECM 3.6. The correct ECMAPI Wizard should be on the ECM 3.6 media.
Temporary Solution:
n/a
Fix Information:
Run the ECM 3.6 ECMAPI Wizard before installation of C.01.09 against ECM 3.6. In the ECM Installation Guide of ECM 3.6 it is documented which ECM API is needed and how to install it.
Keyword: ECM Integration
One-line Description:
If data exceeds the maximum size it cannot be uploaded to ECM
Problem:
ChemStation has a 4GB file size limitation on the creation of the container file (ssizip). The SSIZIP functionality used by ECM-client does not have a 4GB limitation.
Temporary Solution:
break up the sequence into smaller parts
Fix Information:
Fixed in C.01.09 Update 2 and C.01.10 Update 2
Keyword: ECM Integration
One-line Description:
ECM Chemstation XML Filterpack is failing to filter ssizips when instrument includes fraction collector
Problem:
The ECM Chemstation XML Filterpack is failing to filter ssizips (result.xml) from a workstation configured with an instrument including a fraction collector.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab ChemStation C.01.10 Update 5.
Keyword: ECM Integration
One-line Description:
Logon to OpenLab Control Panel and unlock ChemStation is slow with ECM 3.6 storage
Problem:
The logon times to the OpenLab Control panel and the time to unlock a ChemStation instance can be very slow. This is the case when an ECM 3.6 storage backend is used.
Temporary Solution:
Manually add the required registry entries that can be provided by your Agilent support team.
Fix Information:
n/a
Keyword: ECM Integration
One-line Description:
Enhanced SSIZIP file modification detection
Problem:
If a SSIZIP file was modified outside of ChemStation it is not obviously detected as such.
Temporary Solution:
n/a
Fix Information:
The software releases C.01.10 Update 5 and LTS 01.11 introduce a new mechanism to notify users upon loading result sets that got modified outside of the OpenLab ChemStation application.
Keyword: ECM Integration
One-line Description:
Exception of type 'System Out of Memory Exception' when data > 550Mb uploaded to ECM at SSIZip checksum generation
Problem:
In C.01.10 Update 5, an exception of type 'System Out of Memory Exception' occurs when data > 550Mb is uploaded to ECM at SSIZip checksum generation. This issue occurs for large data files when the checksum is calculated, because the contents of entire file are read into memory. This causes System Out of Memory Exception.
Temporary Solution:
n/a
Fix Information:
Resolved in OpenLab CDS ChemStation C.01.10 Update 7. This issue is resolved in OpenLab ChemStation LTS 01.11.
Keyword: ELSD Driver
One-line Description:
ELSD G4218A can not handle change of Sample Time during run
Problem:
With Rc.Net driver internal filters of ELSD G4218A get confused by change of "Sample Time" during run.
Temporary Solution:
n/a
Fix Information:
fixed with LC Drivers A.02.03
Keyword: ELSD Driver
One-line Description:
The ELSD method cannot be called by a right-click from the context menu
Problem:
In the method and run control view the ELSD has its own dashboard item. Unlike other modules in an RC.Net configuration the instrument setup is not available by right-clicking on it.
Temporary Solution:
Use the instrument menu to have access to all modules, including the ELSD.
Fix Information:
No fix available.
Keyword: ELSD Driver
One-line Description:
Runtime of a G4218A ELSD cannot be changed during a run
Problem:
The runtime of a G4218A ELSD cannot be changed explicitly during a run. However there are software mechanisms which implicitly change a run time, e.g. the function "ignore missing vessel" of an autosampler normally stops the run after some seconds if no vessel is present. In case of an G4218A ELSD this does not work and the run uses the full specified run time.
Temporary Solution:
A single run may be stopped manually.
Fix Information:
n/a
Keyword: FLD Driver
One-line Description:
FLD G1321 Scan ranges and threshold under certain conditions not available
Problem:
Scan ranges and threshold under certain conditions not available.
Temporary Solution:
n/a
Fix Information:
Solved with LC & CE Drivers A.02.08 SP1
Keyword: GC ACQ
One-line Description:
FetchCPVar Cannot Talk to Application Error
Problem:
Start up Chemstation, but no instrument or sampling diagrams are shown. FetchCPVar Cannot Talk to Application Error occurs. Systems use 7683 ALS.
Temporary Solution:
n/a
Fix Information:
Fixed in revision B.04.03 DSP1.
Keyword: GC ACQ
One-line Description:
FetchCPVar Cannot Talk to Application Error
Problem:
Open an online session, initial or any subsequent try. This results in a benign pop-up error: 'FetchCPVar cannot talk to application'.
Temporary Solution:
n/a
Fix Information:
Fixed in B.04.03 DSP1 [54].
Keyword: GC ACQ
One-line Description:
Column compensation cannot be started from the Data System with 68xx RC.NET
Problem:
Column compensation (comp) cannot be started from the Data System with 68xx RC.NET
Temporary Solution:
Column compenstation must be started via the GC keyboard.
Fix Information:
No Fix Planned.
Keyword: GC ACQ
One-line Description:
OpenLab CDS, C.01.02, 7820A run continues when STOP on 7820A is pressed
Problem:
A manual injection is done and if the STOP on the 7820A front panel is selected, the GC stops (and resets the initial GC settings) but the chemstation continues. When the STOP button is pushed on the 7820A, the "current logbook" reports the end of run correctly, and reports the system is running data analysis but does not go into Data Analysis right away. If this is a 30 minute run, and the STOP button is pressed on the 7820A after 10 minutes, AND YOU WAIT 20 more minutes until the Chemstation gets to the end of the run, a report is generated. The system appears to hang up waiting on the expected run time.
Temporary Solution:
Workaround is to use STOP run from the software.
Fix Information:
n/a
Keyword: GC ACQ
One-line Description:
After switching the injection source from front to back it is no longer possible to change the counter part of the Prefix/Counter naming scheme
Problem:
After executing a number of runs and then switching the injection source from front to back or vice versa it is no longer possible to change the counter part of the Prefix/Counter naming scheme. Even if you type a new counter, the next injections will always start with counter 000001.
Temporary Solution:
n/a
Fix Information:
Resolved in revision C.01.05 Hotfix05.
Keyword: GC ACQ
One-line Description:
After switching the injection source from front to back it is no longer possible to change the counter part of the Prefix/Counter naming scheme
Problem:
After executing a number of runs and then switching the injection source from front to back or vice versa it is no longer possible to change the counter part of the Prefix/Counter naming scheme. Even if you type a new counter, the next injections will always start with counter 000001.
Temporary Solution:
Resolved in revision C.01.04 Hotfix06 and C.01.05 Hotfix05.
Fix Information:
Fixed in OpenLAB CDS ChemStation C.01.06.
Keyword: GC ACQ
One-line Description:
GC Back injector only used, the sequence specified volume is ignored and the method setting is used
Problem:
The injection volume specified in the sequence table is ignored when only the back injector is used. It is not possible to override the injection volume specified in method when using the back injector. The report for the back injector will include the injection volume that was specified in the sequence table which is not correct.
Temporary Solution:
In sequences with only back injections set the injection mode to dual simulaneous. The execution of the sequence will not change, but the volume override will be accepted.
Fix Information:
This issue has been resolved in OpenLAB CDS ChemStation Edition C.01.06 Hotfix 02.