Agilent

Software Status Bulletin


OpenLAB CDS EZChrom Edition     M82xxAA

Known Problem Report as of Mar 30 2026 4:00AM

Preface
-------------------------
This Software Status Bulletin (SSB) documents all known problems in the software
product designated above. The SSB is derived from Known Problem Reports (KPR)
which result from user problems that have been classified as documentation
problems or software defects. When a KPR is written, an identifying number is
assigned to it, and the KPR is added to the next edition of the SSB.

User inputs that have been classified as Enhancement Requests are not documented
in the SSB. User problems that have been submitted, but that have not been
classified by the time the SSB is generated are not included in the SSB.

How to use the SSB
-------------------------
When you experience a problem with a product, first check this SSB to see if the
problem has been reported already, and if there is a temporary workaround
available for the problem, or if the problem has already been fixed by a new
revision. If the problem is not listed in this SSB then you may wish to report
it to the Response Center or to your field support representative.


To determine if your problem is documented in this SSB, first look in the
Keyword Index section of the SSB. Under each keyword is a listing of one-line
descriptions of related KPRs. If any of these sound like yours, locate the KPR
# in the Known Problem Reports section of the SSB, and read the full KPR. The
KPRs in the Known Problem Reports section are sorted by KPR #.


There are two sections in the SSB:
Keyword Index: This index is categorized by keyword. For each KPR there is a
brief description and a KPR #. A KPR may be associated with more than one
keyword.


Known Problem Reports: This section contains KPRs, with all the available
information relevant to the problem. KPRs in this section are sorted by KPR #.


Keyword Glossary



35900 Control
3rd Party Drivers
68xx Driver
7890A Control
7890A Control,Acquisition
78xx Driver
88xx GC
Acquisition
Administration
Agilent LC Driver
Audit Trail
Automation
Autosampler
CDS
Client Report Rendering
Configuration
Configuration,Instrument Control
Control Panel
Custom Reports
DAD/MWD Driver
Data Analysis
Data Analysis,Installation
Data Store Integration
Data Store Integration,ECM Integration
Documentation
ECM Integration
External Component
Headspace
Help
Installation
Installation , 35900 Control
Instrument Control
Integration
Localization
Method
Method Usage
Output
Performance
Project Configuration
Pump Driver
Report Template Design
Report Templates
Reports
Reports - IR
Roles and Privileges
Sampler Driver
Sequencing
User Interface

Keyword Index


Keyword: 35900 Control

KPR#:213248 Update to A.04.09 can break 35900E driver installation with CTC
KPR#:307590 Preview run not working in A.04.09 & new RC.Net Driver functions differ
KPR#:381770 Install manual(M8201-90038) references installing Classic Drivers for 35900 and is not correct
KPR#:2121 35900E ADC-instrument cannot be configured with RC.Net driver after reinstalling the OpenLAB CDS application

Keyword: 3rd Party Drivers

KPR#:138882 The running sequence can not be updated
KPR#:389878 Hitachi LCE and CM: Cannot Use Submit Button in Run Queue to Submit Runs
KPR#:394168 Hitachi LCE and CM: Cannot Use Submit Button in Run Queue to Submit Runs

Keyword: 68xx Driver

KPR#:16 Vial Missing Error Message
KPR#:17 How to Change Column Dimensions
KPR#:18 If One Signal Needs to be Saved, it Must be Assigned as Signal 1
KPR#:22 In 68XX driver revision 5.02 - 6.11, it is not possible to Modify a Method During a Run to Extend the Run Time or Change the Oven Temp
KPR#:23 Info in Edit instrument parameters\Columns is incomplete
KPR#:31 Ignore missing vial option is not present in GC
KPR#:33 Instrument Setup graph has incorrect column legend
KPR#:38 ALS errors abort sequence; no Retry/Abort dialog box option offered with 68xx RC.NET Driver
KPR#:59 In 68XX Driver Version 5.02, it is not Possible to Start the 6890 RC.NET from the Front Panel Start Key
KPR#:66 CIS4 inlet is not recognized with 6890 GC RC.NET
KPR#:83 Only Fast Plunger Speed is available for PTV Solvent Vent multiple injections
KPR#:89 Incorrect injection volume shown in method report if PTV solvent vent is used
KPR#:94 68xx Method Download Times Out
KPR#:95 "Method modified after loading" occurs when running a sequence (6890 RC.NET)
KPR#:99 Sequence continues while GC temperature is not ready
KPR#:100 A Method Can be Created Which Will Not Download to the Instrument and the Sequence Does Not Run
KPR#:101 GC Back Injector Not Present in Edit Method Screen
KPR#:102 Validation failed shown on left corner of Resolve Method screen
KPR#:103 Injector Panels in GC Edit Method Show as Uninstalled When an Injector is Installed.
KPR#:105 Plunger Speed is Not Saved with 6850 GC Methods
KPR#:106 Opening the GC method parameters to only look (no change made) and selecting OK to close stops the sequence or single run
KPR#:109 Method Contents Report Displays the Incorrect Configuration of Front and Back Detectors for 6890 GC
KPR#:113 Cannot Run a Packed GC Column with a 6890
KPR#:116 Pulsed Split/Splitless Injection Does Not Wait For Ready on a 6890.
KPR#:117 In 68XX GC Driver Version 6.01, 2 Injections are Made When Using the PTV
KPR#:119 Vial left in gripper; no message shown to alert user about the problem
KPR#:120 6850 vial numbers are not consistent with standard numbering system.
KPR#:122 Method Resolution Changes Oven Temperature Program
KPR#:123 Column info is missing with 6890
KPR#:124 Method Modification Flag is Set after Edit Method: Occurs on first connection or after upload from GC.
KPR#:128 Method Ddownload Failure Occurs Due to Column Calibration Limits
KPR#:129 Instrument actual value for COC inlet does not display correctly
KPR#:130 Instrument actual value for anode purge flow on an ECD does not display correctly
KPR#:132 6850 Support of Column Compensation
KPR#:136 Data Files do not Contain the Correct Signals if the Front Inlet is Connected to Back Detector
KPR#:140 The Oven Ramp table does not display correctly in IR with Acquisition method selected.
KPR#:141 6890 PTV Inlet not recognizable in OpenLAB EZChrom
KPR#:142 Multiple errors when the 6890 doesn't have any inlets
KPR#:143 Sequence table Inj Vol Override goes to wrong stops for 6890/7683: C.01.05, A.04.05, C.01.06, A.04.06
KPR#:145 Often get two or more ChemMain.exe's running following shutdown of Control Panel or online session.
KPR#:147 6890N with 7693 barcode scanning checksum mismatch does not return vial to tray and skips a vial.
KPR#:148 Customer can't install colum information from catalog if column description is blank
KPR#:149 EZChrom: Installation of GC Drivers A.03.02 breaks a 7697 configuration if installed
KPR#:150 Setting a Hold time of 0.0 does not work correctly with GC6890
KPR#:154 Elite: G1888 status only shows connecting after GC driver installation
KPR#:155 Instrument session continues to run in background after session is closed.
KPR#:159 6890 online help has the bad strings in Injector page.

Keyword: 7890A Control

KPR#:272474 Instrument Type does not include 8860 and 8890 GC

Keyword: 7890A Control,Acquisition

KPR#:183609 Dual channel issue online plot

Keyword: 78xx Driver

KPR#:268 Behavior of the Maintenance Screen When Adding Modules to the GC
KPR#:269 ALS Behavior When Aborting a Sequence
KPR#:270 7890B GC Column Maintenance Counter Behavior
KPR#:271 Update GC Column Catalog dialog is limited to 10 items; no way to display the rest (...)
KPR#:272 Method Differences Viewer Behavior with Scanned Columns
KPR#:273 7890B GC Method Compatibility
KPR#:274 Logbook Messages for Scanned Consumables Items
KPR#:275 The Syringe and Liner are Not Shown in the Method
KPR#:276 The PC and GC Clocks Need to be Synchronized to Account for Daylight Savings Time
KPR#:277 Logging of Injection Tower Location
KPR#:279 Explanation for How Default Early Maintenance Feedback Counters Work
KPR#:281 The 7890 GC Goes Offline When Non-English Characters are Entered into the Column Information
KPR#:282 Instrument Actuals Show "GC is Offline" Quickly at the Method or Sequence Start, then Returns Back Online and the Method Does Not Run
KPR#:283 78xx GC configuration fails; Failed to Create Instance
KPR#:284 Data not collected, sequence stopped when vial removed from turret after injection
KPR#:285 C.01.06: Import eMethod or Export eMethod freezes system
KPR#:287 0-volume blank run in overlapped sequence causes injection
KPR#:294 Standard and Easy Sample Prep programs run incorrectly in a calibration sequence (run for unknown samples only) when overlap is involved.
KPR#:295 7890B configuration, model number serial number and IP address change after restart
KPR#:296 Makeup Gas mode terminology should be consistent with the GC front panel
KPR#:297 Agilent Solvent Elimination Wizard issue
KPR#:298 H2 Sensor button crashes EZChrom
KPR#:299 Clicking OK after making changes to Events does not save the changes made to the events.
KPR#:301 Can't select uECD makeup mode to "Does not affect Makeup Flow"
KPR#:302 Can't save "Included in Makeup Flow" on uECD setting screen.
KPR#:304 The IQT is failing for GC Drivers version A.02.03.074(5.03) for Chinese and Japanese and passing for English in EZChrom A.04.05 with update 5 is installed.
KPR#:305 7820 GC: Method modification flag set after starting a sequence or method
KPR#:306 EZChrom: Method returns to the "untitled.met" after running single or sequence.
KPR#:307 EZChrom: Method Parameters incorrect after loading a new method. OpenLAB CDS EZChrom
KPR#:308 Method printout missing column parameters: flow or pressure, setpoint On, (Initial), and post run
KPR#:309 There is an incorrect character when displaying temperature information in the method report in Japanese
KPR#:310 MMI Cryo Fast Cooldown not saved
KPR#:311 Method is always seen as modified, nothing can be added to the queue.
KPR#:312 Oven Slow fan will be off after method is downloaded the slow oven mode on to 7890
KPR#:313 During a pulsed splitless run, the purge time shorter than the pulse time cannot be entered.
KPR#:314 Method print out with 7890 configured does not show total GCc run time.
KPR#:315 There is no "Inlet Septum Purge" event in Runtime Events table - EZChrom
KPR#:316 External Device "Not Ready" ignored by OpenLAB CDS ChemSataion and EZChrom
KPR#:318 Aux Flow Limited to 150 mL/min even when pressure is less than Maximum
KPR#:319 Post Run Flow Rate has default that out of range and value seems not to be stored
KPR#:320 Method User Interface does not correctly load the content of a method.
KPR#:324 Does not delete previous open GC method event table when load method. Wrong injections can be created.
KPR#:325 Method download to GC while in sequence causes it to disconnect and sequence stops.
KPR#:326 7890 method print doesn't include post run time
KPR#:335 Crash of Software after check "Show Threshold Value as remaining time" in EMF Options
KPR#:339 Not allowed to set proper GC column max temp because it is less than 150degrees.
KPR#:340 FID signal subtraction setting changes spontaneously from "Nothing" to "Column Compensation" during run.
KPR#:341 "Extend Run" in GC settings does not work correctly
KPR#:342 Cannot configure a 7890B on an existing client server system
KPR#:343 Copy to Clipboard Function in Instrument Actuals throws exception error
KPR#:344 The column flow calculator calculation is off by almost 3x the flow rate for the system
KPR#:345 Sample Overlap settings can not be edited after driver update, but are still part of the method.
KPR#:346 Error Unhandled Exception when changing columns and a Gerstel CIS4 in installed in the 7890.
KPR#:347 Incomplete Method Parameters for xCD in Acquisition Method Report
KPR#:349 Actual flows displayed by CDS aren't the same as the setpoint flows
KPR#:350 Print method does not include all MMI setpoints: Purge time, Post Run Temp, Time, Flow
KPR#:351 Extend run does not extend the acquisition of the Intuvo 9000 GC
KPR#:352 Cannot Add User-Defined EMFs on Intuvo 9000 GC
KPR#:353 EZChrom: Some settings not accepted after method migration
KPR#:355 ChemStation Instrument actuals showing D2 Bus Temperature with no D2 bus installed
KPR#:356 Method print is showing Solvent wash mode as A-A2, B-B2 when it should be A,B since it is a 16 vial turret
KPR#:357 Inlet pressure setpoint displayed assuming all thermal zones are turned ON, results in confusing mismatch when heated zones are Off.
KPR#:359 IQT report failed on EZChrom Elite 3.3.2 SP2 with GC Driver B.01.01.
KPR#:360 Method Editor for PP inlet & Packed Column incorrectly allows both flow and pressure setpoints to be entered
KPR#:361 ChemStation can lock up when changing Intuvo Bus Flow Chip types with ChemStation session running.
KPR#:364 Selection box remains after selection made on Events page
KPR#:365 Gray bars hide setpoints after context menu download of method
KPR#:366 Multiple clicks necessary for combo box to list choices.
KPR#:367 Right Clicking on Method Editor makes UI hide behind Instrument Screen
KPR#:368 Incorrect Values in 9000 Pressure/Flow Calculator
KPR#:369 Make changes to guard chip on Intuvo 9000 GC, audit trail just says temperature changed
KPR#:370 Backflush Wizard illogically sets trial backflush times after the elution time of last peak of interest
KPR#:371 Should not be able to download a method while a in a GC run or sequence.
KPR#:375 No indication that the AutoSampler tray is in park mode.
KPR#:376 The Local User Interface(LUI) on the Intuvo 9000 GC shows long name for connected data system.
KPR#:377 For both the 7890 and Intuvo 9000 there is no change in the UI to show the user that a rack is not present.
KPR#:378 Post Run Backflush. The Guard Chip temperature was not set to the Post Run Oven Temperature.
KPR#:379 Location of Intuvo 9000 GC System Health Report should be moved
KPR#:380 Link to Intuvo 9000 GC help does not work.
KPR#:381 Event drop downs are black when creating a new method
KPR#:383 PCM will not control in constant flow.
KPR#:388 RTL for OpenLAB CDS 2.X does not calculate the pressure correctly when locking or re-locking the method when setting a Target retention time.
KPR#:422 HS Sleep/Wake mode should sync with GC Sleep/Wake mode
KPR#:431 Overlap injections on the 9000 uses only the injection volume from the first injection for subsequent injection.
KPR#:439 Downloading method parameters (like oven temperature) on a 6890A does not work with OpenLab CDS 2.4
KPR#:472 Chromeleon: Intuvo trigger two instrument notification for configuration changed
KPR#:473 Chromeleon: GC: Change of column length (and related parameters like flow etc) via calibration in online method sometimes reported in audit trail sometimes not
KPR#:474 Chromeleon: GC: Column length can only be calibrated once in Online Method Editor
KPR#:475 Chromeleon: Setting Liner on Intuvo triggers two config change messages in audit trail
KPR#:476 8860 Open Lab CDS EZChrom compact license is invalid.
KPR#:477 Selecting Partial Data for the 6890 causes the retention times and areas to be wrong.
KPR#:478 6890 - cannot set post run flow for packed column
KPR#:479 Chromeleon: GC 6850A and 6890A: GC online method is not editable
KPR#:480 Chromeleon: GC 6890A: Wrong temperature shown in chromatogram for signal GC_OvTempe_Act
KPR#:481 Chromeleon: GC 6850A: Cannot create instrument method
KPR#:482 Chromeleon: GC8890: Current Signal Value in 'Blank evaluation' shows incorrect value in Method editor
KPR#:483 Chromeleon: GC8890: Blank evaluation setpoints missing in method report
KPR#:484 Signal Scaling is incorrect for the 6890
KPR#:485 Chromeleon: Intuvo: Help button in window 'Configure Agilent Intuvo 9000 GC' opens 78xx Series GC Help
KPR#:486 Chromeleon: GC8890: Issues during sequence execution
KPR#:487 Chromeleon: Changes made in 68xx Edit Online Method not applied or reported
KPR#:488 Error when loading GC method without Blank Run Analysis setpoint
KPR#:489 Chromeleon: Corrected column length will not save after manual method resolution
KPR#:490 Chromeleon: Signals from previous method persist in new method
KPR#:491 Chromeleon: ECD Help page cannot be accessed
KPR#:492 Chromeleon: All 8860 signals not displayed or available in report
KPR#:493 Chromeleon: Error occurred at several points of operation
KPR#:494 Chromeleon: GC6890: Front Injector parameter change in Online method not applied
KPR#:495 Chromeleon: GC6890A: Method download fails
KPR#:496 8860 on Chromeleon -Unhandled exception in Chromeleon Driver Host (CmDDKHost): AgilentTechnologies.DdkIcfAdapter|Agilent
KPR#:497 Chromeleon does not recognize the tray with Back Injector installed.
KPR#:498 Chromeleon: Automatic de-activation of not supported feature blank evaluation cannot be saved
KPR#:499 Getting multiple Errors in the PC Event Log for AgGC7890 after disconnecting and Reconnecting LAN cable.
KPR#:500 GC6890: Valid vial ranges are not correct for 6890 in some cases
KPR#:501 Chromeleon: GC6890A: Error 'The communication FIFO of the DDK driver is corrupted.' after connecting and disconnecting several times.
KPR#:502 When click the "Apply the Option" in GC Chromatographic Column Directory under Column Configuration, an exception occur on Openlab 2.4
KPR#:503 Upgrade installation from GC B.01.03 to 3.0 SR2 fails to install Ag7697WS.dll (and .pdb)
KPR#:504 Chromeleon: GC6850A: Changing method parameter creates method inconsistent to configuration
KPR#:505 Chromeleon: GC6850: Partial Data option causes issues during sequence execution
KPR#:506 Chromeleon: GC6890: Unexpected audit trail entries during endurance run
KPR#:507 SR1 for the GC3.0 drivers breaks the connection between an 8890 and 7000D transferline temperature reading in MH ACQ 10.
KPR#:508 Not able to download flow setpoints to 6890 VI inlet.
KPR#:509 Chromeleon: Incorrect conversion and missing unit when using pressure setting bar
KPR#:510 68xx: Chromeleon: System with G4513 won't make 1.1 uL injection
KPR#:511 Chromeleon: System crash: system slow and then crashing
KPR#:512 8860 with Headspace/PAL does not consume correct VL license for OpenLAB 2.4
KPR#:513 B,04.03 SP2 will not connect to an 8890 GC.
KPR#:514 Chromeleon: 6890GC Oven temperature disabled in method - Run does not start.
KPR#:515 3.0 SR2 driver fails SVT on OL CS C.01.09.144
KPR#:516 OLCS C.01.09 with Driver 3.0.650 will not launch any instrument.
KPR#:517 Chromeleon: SVT Chameleon 2.1.30 Fails
KPR#:518 Function_build number not display in SVT report in EzChrom and OpenLab CDS 2.4
KPR#:519 8890: the MS transfer line temp displayed is 0 C but its actual temp is 250 C.
KPR#:599 Instrument Status shows HS Comm as 'OFF' when it is not.
KPR#:1196 7890A: Shutdown events are not published to the software GC Status screen

Keyword: 88xx GC

KPR#:345674 8860 GC System: there is no core license available

Keyword: Acquisition

KPR#:138296 On-line signal for Hitachi Chromaster is only displayed during acquisition
KPR#:146968 Certain auxiliary traces of Accela pump are not supported
KPR#:147007 The default "Use Method" setting is not working as expected.
KPR#:147010 Mismatched channel/peak data for Thermo Trace 1310GC
KPR#:147797 LC does not abort sequence after pressure exceeds upper limit
KPR#:200819 Delay observed when comparing the Run Time and Analysis Time of the data.
KPR#:220292 Shimadzy 2010 GC and HS-20 synchronization issue
KPR#:316749 Delay observed when comparing the Run Time and Analysis Time of the data.
KPR#:356008 Hitachi LCE and CM: Cannot Use Submit Button in Run Queue to Submit Runs
KPR#:380636 Bruker GC: Changing the current time of the computer makes the driver go offline sporadically
KPR#:401753 No error message for missing vial
KPR#:536644 Hitachi Chromaster LC: No Injection Happens if "No injection/Instrument blank" Used as the Injection Source
KPR#:695883 Hitachi ChromasterUltra RS: Displayed Runtime Doubles for Detectors Online Plot During a Run
KPR#:949857 Data is not being saved when executing a sequence using the PAL 3 RTC with Tool changes
KPR#:1192808 Shimadzu LC: Manual Injection Starts Without Waiting for Trigger
KPR#:1382228 CTC PAL: Activity Log: “Could not read barcode” message does not identify the vial/tray position

Keyword: Administration

KPR#:36 Windows Domain tab is grayed out in OpenLAB Server Utility
KPR#:39 Cannot stop Shared Serviceshost service on computer after changing the Domain Account password

Keyword: Agilent LC Driver

KPR#:179143 Error in RCNet_RCPluginProvider::SetConfigSettings...-Exception
KPR#:400080 Warning message “The installed OpenLAB software is incomplete” or “Repair is not possible as no complete

Keyword: Audit Trail

KPR#:146969 Timetable changes audit traces on general level only
KPR#:146973 PDA 2998/ UV 2489 : Audit trail messages do not contain unit information when values are reported
KPR#:191802 Error message for e-signing twice can be unclear on Citrix Clients
KPR#:210882 Level changes in Peak Table are not captured in the Method Audit Trail
KPR#:251903 Method audit trail will be lost once loading method in different instruments
KPR#:292456 Units for carrier A timetable
KPR#:297766 Method Audit Trail Incorrectly References Detector A Gas 2 as Gas 3
KPR#:589099 Hitachi Chromaster Ultra: No System Activity Log Entries Upon Vial Settings Change

Keyword: Automation

KPR#:256374 “Unable to Create UILess IR control” error message in Instrument activity after launching instrument from Automation tool.

Keyword: Autosampler

KPR#:338 EZChrom: PAL XT module not available after installation

Keyword: CDS

KPR#:146999 EZChrom Edition only: Increased memory consumption after a sequence.
KPR#:516793 PE LC300: Incorrect Status of Non-Detector Modules While Running Shutdown Method
KPR#:517638 Bruker GC456: Stepwise Online Signal
KPR#:548087 Hitachi Chromaster and LaChrom Elite LC: Differences in DAD channel naming in Data Analysis
KPR#:554454 CTC PAL3 for LC Cannot be Launched with OpenLab CDS 2.6
KPR#:1378523 Shimadzu LC: A Task with Shutdown Functions Not Executed

Keyword: Client Report Rendering

KPR#:46 Intelligent report of English openlab chemstation can not show chinese name in PDF report
KPR#:628 Large matrix reports may produce additional empty pages
KPR#:653 Calculating the standard deviation of identical numbers may return NaN
KPR#:657 Changes in report templates are not immediately shown on report printouts
KPR#:659 Rendering error when using a template containing a large result matrix

Keyword: Configuration

KPR#:22100 Auto-configuration for instruments is not supported
KPR#:146924 Starting Acquity instrument for first time after an AIC reboot has errors
KPR#:146964 Some Autosampler parameters are not uploaded e.g. loop size.
KPR#:146966 Manual tray type configuration required
KPR#:146970 Auto-configuration for instruments is not supported
KPR#:146988 It is not possible to configure a Shimadzu SPD-M20A detector in the same system configuration as a Shimadzu Nexera integrated system.
KPR#:147000 200uL Sample loop not configurable in Alliance instrument configuration
KPR#:147014 For the GC column properties, there are missing units for column length and internal diameter.
KPR#:285964 Hitachi LaChrom Elite auto configuration fails on EZChrom with Win 10
KPR#:288817 Instrument status shows temperature value for non-configured injector
KPR#:397013 Two different pressure units displayed in data analysis
KPR#:456679 Hitachi LCE and Chromaster LC: Comma as a Digital Separator Limitations
KPR#:486957 PE AutoSystem XL GC: cannot add VICI valves
KPR#:551474 PE TurboMatrix HS with Agilent GC Cannot be Configured More Than Once
KPR#:715153 PE Flexar LC: Configuration of Flexar Column Oven with Peltier Fails
KPR#:1216158 CTC PAL: Imported/removed scripts are not reflected in Method editor unless acquisition is relaunched

Keyword: Configuration,Instrument Control

KPR#:732721 PE TurboMatrix HS: High Pressure Injection Not Usable Without Vial Pressure Sensor
KPR#:800415 PE TurboMatrix HS: High Pressure Injection Not Usable Without Vial Pressure Sensor

Keyword: Control Panel

KPR#:117721 Exception when login in OpenLab Control Panel. ECM does not display all the subfolders when browsing for Project Path during project creation; exception displayed

Keyword: Custom Reports

KPR#:1995 FIltering syntax has changed

Keyword: DAD/MWD Driver

KPR#:149830 High load on CAN may cause "Spectra runbuffer overflow" entries and a loss of spectra
KPR#:149837 Autobalace executed although in Sequence table set to "OFF"
KPR#:149898 Lowest possible DAD wavelength is 190.5 nm

Keyword: Data Analysis

KPR#:139618 Incorrect value entered at peak group table if GC method parameter window is open using Agilent GC driver B.01.02
KPR#:153192 Chromatogram Annotation for Purity does not reflect actual value less than 1
KPR#:379394 Sequence audit trail create time differs from report

Keyword: Data Analysis,Installation

KPR#:408720 Analyses turns into error state "HRESULT: 0x80010105 (RPC_E_SERVERFAULT)"

Keyword: Data Store Integration

KPR#:191223 The Users and Groups report generated on a A.02.02 client from a 2.0 Datastore Server is incorrect

Keyword: Data Store Integration,ECM Integration

KPR#:339790 Mismatch between result set modification time and result set creation time on CM and ECM3.5 backend

Keyword: Documentation

KPR#:31 The module type of the Alliance instrument is reported as a 'Detector' in OpenLAB CDS ChemStation Edition reports
KPR#:146997 Instrument Configuration report can not be printed when offline.
KPR#:147024 The firmware version required by the driver is incorrectly stated as v1.41.6164 for the Waters Alliance 2489 detector in the User Guide
KPR#:177259 No help for reason type in Method Audit Trail
KPR#:260862 PE Clarus GC: Some Method parameters are missing in the method report
KPR#:391787 Hitachi's Sub CPU Firmware can Only be Visible from Hitachi Maintenance Tool
KPR#:601921 Hitachi Chromaster Ultra Rs: Document for the Hitachi Maintenance Software is not the latest
KPR#:634039 Shimadzu LC: No Information about Wavelength for UV Detector in Chromatogram in DA
KPR#:651995 PE Clarus GC 500: Min Firmware for Mainboard Missing in Training
KPR#:701847 Hitachi Chromaster Ultra LC: Inconsistent driver version 3.20 reported in Software Verification Tool

Keyword: ECM Integration

KPR#:251997 ECM LCDF name ending with a period "." prevents resultset download
KPR#:277494 Report Template Long File Name Not Uploaded and Causes InstClient crash
KPR#:278036 Long File Name Prevents Single Level Calibration

Keyword: External Component

KPR#:350455 Hitachi Chromaster can't Initialize after AIC reboot with "IFC cannot be found" error
KPR#:658322 Hitachi LC: Incorrect Display of Parameters in Maintenance Tool with Win 11
KPR#:708791 Hitachi LC: USB Drivers incompatible with OS Memory Integrity Option
KPR#:762177 Hitachi LC: USB Drivers incompatible with OS Memory Integrity Option
KPR#:788048 Shimadzu LC: Current Value of Wavelength Accuracy 656nm (NAR) Not Displayed in Shimadzu LC System Check Report
KPR#:1036196 Shimadzu LC: SPD-M20A Excluded from the System Check Report if Upgraded to SPD-M40
KPR#:1077556 Hitachi LC: Incorrect Display of Parameters in Maintenance Tool with Win 11
KPR#:1092976 CTC PAL 3 for LC: Importing Pre-installed LC Injection Script Leads to Exception

Keyword: Headspace

KPR#:180 EZChrom: Original method not restored when sequence run from EZChrom

Keyword: Help

KPR#:22140 Error with detectors help
KPR#:90310 Incorrect units for autosampler method setup shown in Help
KPR#:592464 Hitachi Chromaster Ultra: Help Clarification for the Purge Options of the Pump
KPR#:592467 Hitachi Chromaster Ultra: Help Clarification to Rack Settings of Autosampler
KPR#:595483 Hitachi LaChrom Elite: Incorrect Error Message for Analog Input Device Issue

Keyword: Installation

KPR#:22 FTP protocol is enabled by default on OpenLAB Data Store A.01.02 systems
KPR#:38 Project activity logs not displayed after upgrading to A.02.01
KPR#:42 OpenLAB Control Panel does not open after upgrading Windows Server 2008 R2 to Windows Server 2012 R2
KPR#:94355 Installer & Uninstaller wizards still show incorrect references to the driver name as an "Addon" in the installer screens
KPR#:130809 If a privilege is removed from the default roles in OLSS the application registration fails at installation
KPR#:175952 SVT report does not work when installing the driver in non-default location
KPR#:175999 SVT Report Does Not Check 34 Files for Thermo Fisher Trace GC Driver
KPR#:203907 Not all SVT reports get displayed on the screen when manually run SVT on Server
KPR#:222652 Missing file in the reference file in the MSI Package for Shimadzu LC in EZChrom Edition
KPR#:281960 SVT report has 41 files not found in reference
KPR#:290491 No warning when installing later PE GC driver version with the preinstalled older version might lead to malfunction
KPR#:297111 SVT for OpenLab CDS disappears upon uninstalation of PE Clarus GC Drivers
KPR#:298681 There is no Warning When Another Version of PE Driver is Installed Over The Previous One
KPR#:303159 SVT report does not show the installed driver for non-default path CDS installation
KPR#:605033 Mixed Mode OpenLab EZChrom A.04.10 and CDS 2.5 Warning
KPR#:606448 Error During Driver Installation
KPR#:2121 35900E ADC-instrument cannot be configured with RC.Net driver after reinstalling the OpenLAB CDS application

Keyword: Installation , 35900 Control

KPR#:2121 35900E ADC-instrument cannot be configured with RC.Net driver after reinstalling the OpenLAB CDS application

Keyword: Instrument Control

KPR#:32 OpenLAB CDS ChemStation Edition sequence shutdown macro PUMP ALL OFF is not executed.
KPR#:146967 Thermo Accela pump communication LED remains active
KPR#:146992 During acquistion of aWaters Alliance HPLC connected to OpenLab CDS EZChrom Edition, the software allows access/change instrument parameters even if roles/privileges for manual control are disabled.
KPR#:146993 Unexpected change of Alliance pump flow rate and solvent composition after a run observed on a French operating system
KPR#:147009 Shimadzu HPLC shows "Ready" even if Pump or Lamp is set to Off
KPR#:211507 Inlet time program does not work properly
KPR#:212506 Only one valve functions
KPR#:212515 GSV and Split valves cannot be controlled from the events table
KPR#:286720 Status of Oven and Autosampler is "Ready" when the GC is switched off
KPR#:288479 PE Clarus GC: at the end of a run ""The current method has changed" message
KPR#:394374 Shimadzu 2010 Plus GC: instrument connection not closing gracefully
KPR#:401780 "Switch Valve to Load" not available after "Switch Valve to Inject"; "Home arm" action needed first
KPR#:492287 Hitachi LC: Long Launching Time if Instrument Setup Window Left Open Prior To Exiting Instrument Window
KPR#:516796 PE LC300: Uploading Method to the Instrument with Turned off Pump Switches Pump On
KPR#:536651 Hitachi LaChrom Elite: Intermittent Waiting For Trigger Does Not Start the Run
KPR#:543340 VICI Valves: Acquisition Crash on Pressing Change Button in VICI Dashboard
KPR#:645291 Bruker 456 GC: FID Flame status inconsistent
KPR#:723437 PE Clarus 480 GC Cannot Be Connected to the AIC Due to Communication Error
KPR#:939675 PerkinElmer Flexar LC: Communication to Instrument Lost During Long Sequence
KPR#:947158 VICI +Agilent 7890: Changing Position of VICI Valve during a Sequence Can Cause Its Abortion
KPR#:962546 Shimadzu PDA Cannot Be Configured with Agilent LC
KPR#:1129855 Hitachi Chromaster: Script Error Appears Under Certain Conditions
KPR#:1185772 Shimadzu LC: Manual Injection Starts Without Waiting for Trigger

Keyword: Integration

KPR#:138745 Application crash if back data files resultset is reprocessed with the clear calibration flag
KPR#:263508 Reprocess or analyze of data file causes instclient.exe crash

Keyword: Localization

KPR#:287503 PE Clarus GC: Decimal symbol for syringe volume
KPR#:738391 Hitachi CM and LCE: Acquisition Window Crashes if Non-English OS Used for AIC/Client

Keyword: Method

KPR#:146963 Pretreatment methods are saving invalid parameter
KPR#:146971 Method upload functionality is not supported for both OpenLAB CDS Chemstation and EZChrom Editions
KPR#:146977 Run Time Mismatch after running a Gradient Method
KPR#:146995 The 450GC oven does not cool down as expected when the run finishes.
KPR#:146996 Not Ready in Run status/warning appears during run of 450GC and stops run
KPR#:147008 Missing instrument configuration information in the method reports.
KPR#:147013 Instrument Configuration information is not contained within method report.
KPR#:147017 Instrument Configuration information is not contained within method report.
KPR#:279725 Shimadzu LC Vial message: "Object reference not set to an instance of an object"
KPR#:288808 Split flow in instrument status is always X mL/min higher than in instrument setup
KPR#:290303 Instrument Setup Window: Time constant for FIDW is displayed without a unit
KPR#:297200 Incorrect Units (°C) for carrier gas settings in Method report
KPR#:354452 Shimadzu LC 2030C+: invalid range(0.1-50.0) error for syringe >50 uL
KPR#:370796 Method resolution is possible without privilege "Instrument Setup"
KPR#:530381 PE TurboMatrix HS: New Method Throws an Error Message Upon Saving
KPR#:542480 PE AutoSystem XL: Method Downloading to the Instrument Not Logged in the Activity Log
KPR#:695899 Hitachi Chromaster Ultra LC: Direct Control "Turn on at" Functions Properly Only at UTC Time Zone Entry
KPR#:751372 Hitachi CMU: Method Resolution Error When Upgrading from CMU1.0 to CMU1.1
KPR#:761728 PE TurboMatrix and Clarus GC: System Sometimes Prompts to Save Unmodified Method
KPR#:781493 Hitachi Chromaster LC: Actual Run Time Can Be 10x Longer Than Set in the Method Due to Certain System Settings
KPR#:800420 PE TurboMatrix and Clarus GC: System Sometimes Prompts to Save Unmodified Method
KPR#:1063041 PE Clarus GC Default Method Setting Combination for Carrier Not Allowed with EZChrom
KPR#:1155105 Clarus GC: Method UI Might Appear Hanging For Carrier Tab Under Certain Condition

Keyword: Method Usage

KPR#:147025 After data has been acquired on Shimadzu LC-20 systems, a warning "The current method has changed would you like to save it?" is displayed.
KPR#:297768 Incorrect Carrier Info in Method Report
KPR#:414461 Incorrect Absorption Value in the Online signal and Status of 6420 UV-VIS Detector When Non-Default Range Value Used.
KPR#:415474 Hitachi ChromasterUltra Rs: DAD Online Signal Wavelength Needs to be Selected in DAD Direct Control Option
KPR#:691353 Shimadzu GC: Method Resolution Error Does Not Provide Details

Keyword: Output

KPR#:246004 EZChrom GPC/SEC uses a time slice algorithm that gives different results from other GPC software

Keyword: Performance

KPR#:261090 Abort Current Run Sets Instrument in Error State
KPR#:281316 Sporadic internal error "timeout expired" for activity log queries
KPR#:377731 Thermo SII Driver: Corona Veo Not Supported in Agilent LC stack
KPR#:511004 Refresh period of the Online Plot Might Increase for Long Sequences
KPR#:511440 Thermo Fisher SII: Submitting Shutdown Method in Acquisition Leads to Error for Corona Veo Detector
KPR#:614017 Thermo Fisher Ultimate 3000: The Injection Collection Could Not Be Initialized
KPR#:614020 Thermo Fisher SII: "Blank" Type Injections Cause Sequence to Abort
KPR#:658297 Hitachi Chromaster LC: Long Acquisition Launch Time
KPR#:664341 Thermo Fisher Trace GC, Integrion, Aquion: Method Part of the UI Not Accessible Due to Missing Scrollbar
KPR#:667230 Thermo Fisher Aquion Remains Offline After Reset
KPR#:667234 Thermo Fisher Aquion: Error Due to "Non-Specified Flow Rate"
KPR#:667333 Thermo SII Integrion: Setpoint Column Temp Shown Instead of Actual Temp in OpenLab Instrument Status
KPR#:667424 Termo SII: Integrion Method Editing Issues
KPR#:683446 Thermo CAD with Agilent LC: The Sequence Aborts if it Encounters No Injection/Blank Run
KPR#:695395 Shimadzu GC: Online Signal Does Not Work
KPR#:801770 Sequence aborts after updating running sequence
KPR#:804696 PE Clarus 680: Attenuation Time Change Events Do Not Work Properly
KPR#:1029458 Thermo Fisher SII: Take/Release Functionality Leads to an Unhandled Exception
KPR#:1058303 Injector temp Error happened when method switch from high temp. to low
KPR#:1064776 Thermo SII 1.3 with CDS 2.8: Issue about log4net in Station IQ and some Thermo instruments cannot be configured
KPR#:1073349 Sequence stalling and the time in log back to 1601 year
KPR#:1076815 Run can't be aborted due to Auxiliary trace enabled
KPR#:1087709 Unable to configure instrument in ChemStation
KPR#:1378700 Thermo Fisher SII: SQL engine not starting after installation on Win 11 and Win Server 2022 OS

Keyword: Project Configuration

KPR#:227940 "Edit running method" privilege has no effect on user privileges

Keyword: Pump Driver

KPR#:149838 Misrepresentation of Bottlefillings in Dashboad of PumpValveCluster
KPR#:149845 PumpValveCluster with ISET - Solvent composition: resulting %A is wrong (at 1 line in timetable)

Keyword: Report Template Design

KPR#:25 Intelligent Reporting : Last displayed digits might be cut off, in cases where the number is zero
KPR#:40 Missing named peaks are reported in Intelligent Reporting for all the channels or signals
KPR#:42 The Fixed Text field in IR Templates does not accept Chinese characters
KPR#:50 Intelligent Reporting - Spaces are stripped from filters when they are saved
KPR#:56 Error seen in Intelligent Report header (#error) if any related data field is used in an expression
KPR#:58 Intelligent Reporting - Stacked Chromatograms do not display the time axis in reports
KPR#:59 Intelligent Reporting - Square brackets cannot be used in filter expressions
KPR#:60 Cannot calculate summary statistics on columns containing variable or aggregator
KPR#:620 Error message when adding a chromatogram to a report template
KPR#:621 Internal grouping of tables and usage of calculation variables cannot be applied at the same time
KPR#:622 Summary Calculations cannot be applied to table columns containing calculation variables
KPR#:623 Adding a template description to a previously erased and saved description field causes a KeyNotFoundException
KPR#:627 Customized composite groups should have a unique name
KPR#:662 Filter definition does not always allow usage of OR operators
KPR#:668 Reporting time format does not offer an option of (dd/mm/yyyy and time)

Keyword: Report Templates

KPR#:36 Min and Max column summary calculations report GMT and not local time for date fields
KPR#:39 The report date is printed incorrectly in the Agilent supplied intelligent reports
KPR#:43 The template fields "SignalToNoise_EP" and "SignalToNoise6Sigma" do not work in Japanese version
KPR#:47 Center of triangular calibration point labels do not represent the measured value
KPR#:48 XLS export of intelligent reporter templates generate hidden rows
KPR#:658 Customized chromatogram or axis lables may not display correclty
KPR#:138370 Esignature table does not remember the formatting in Standard report
KPR#:138858 Only 50 .srp active template are reflecting under Report -> View option
KPR#:241653 VICI Valves: No Serial Number and no Firmware in IR report
KPR#:288580 Carrier table in method report is incorrect
KPR#:301630 Peaks and Repeating property in IR template does not present Missing Peaks
KPR#:309865 No Module Information is Reported in the Sequence Summary Report

Keyword: Reports

KPR#:138810 Versioning in the Method Report (Classic Reports)
KPR#:138849 System hangs when trying to generate sequence summary reports with Primaide Modules
KPR#:224121 QC Check Rounding limitation
KPR#:372183 SII: Wrong Entry in Instrument Module Report
KPR#:414840 An Additional Warning on Station IQ report for OpenLab CDS 2.5
KPR#:1370829 Thermo SII Instruments: Missing Information in Module List and Reports for Some Modules

Keyword: Reports - IR

KPR#:213496 "Not-found compound" table in IR report not working
KPR#:242758 Difference in compound group area reported on standard and intelligent reports
KPR#:290069 Difference in compound group area reported on standard and intelligent reports
KPR#:291693 Wrong value of Sample Type in IR - A.04.08
KPR#:316761 "Not-found compound" table in IR report not working
KPR#:547223 Zeroes-trimming for injection volumes lower than 1
KPR#:709423 Integration droplines may not be displayed on a report chromatogram when using Intelligent Report templates

Keyword: Roles and Privileges

KPR#:177092 Analyze tab not greyed out for those without the privilege
KPR#:508781 SII: Direct Control Context Menu Disabled During a Run for User with All Permissions
KPR#:508896 SII: Direct Control Context Menu Disabled for User with Manual Control (in Run) Permissions
KPR#:509382 SII: Instrument Control is Still Possible for User Without Manual Control Privileges
KPR#:933834 Thermo SII: Method "Approve" Function Does Not Allow Method Changes Anymore

Keyword: Sampler Driver

KPR#:149831 LC driver starts run with ALS Chiller turned off
KPR#:149841 ICF A.01.04 - No method download, if method contains pretreatment/injector program steps.
KPR#:149880 For overlapped injections in vial samplers G1313A and G1329A/B, injection volumes and sample locations are not updated synchronously in the status UI

Keyword: Sequencing

KPR#:226629 Error message: Unable to open method (2) when the sequence is started
KPR#:236585 Sequence Table Column width, reduced making column invisible cannot be restored
KPR#:266379 Saving Easy Sequence without change results in "Filed to upload..."

Keyword: User Interface

KPR#:22101 The alignment of the instrument module icons on the Instrument status display within OpenLAB CDS Chemstation Edition are not correctly aligned and in some cases, may be partially obscured.
KPR#:22136 Comment field is not available in the Instrument setup dialog for the 2998 PDA
KPR#:138857 In Windows 10 Data Appearance Properties Schemas cannot be saved.
KPR#:146965 Tray image for 'Custom 384 well settings' is not displayed.
KPR#:146976 The alignment of the instrument module icons on the Instrument status display within OpenLAB CDS Chemstation Edition are not correctly aligned and in some cases, may be partially obscured.
KPR#:146978 Waters Alliance: The elapsed run time dislayed on the instrument dashboard is not updated when a UV2489/PDA2998 detector is connected (ChemStation Edition only)
KPR#:146990 Improved handling of spectrum acquistion wavelength paremeters requried in user interface.
KPR#:146998 Comment field is not available in the Instrument setup dialog for the 2998 PDA
KPR#:257130 Context Menu Entries in Instrument Status Appear Active During Run
KPR#:257664 No Online Help for Thermo Fisher Instruments Available in Status Dashboard
KPR#:391501 Column Oven Pictogram Displays the Oven is ON in case of a door opened condition even if it is OFF
KPR#:415443 Entries in Method Audit Trail Are Unclear
KPR#:489974 SII: Configuring The First Instrument for SII Might Display Two Modules, One being Empty
KPR#:502441 Using Non Unique Method for Identical Instruments Prompts to Save Changes Even When No Changes Exist
KPR#:508118 SII: Smart Startup, Smart Shutdown, Smart Standby Not Supported
KPR#:508900 SII: Method Can Still Be Edited But Not Saved Without Create and Edit Acquisition Method Permission
KPR#:664364 Thermo Fisher Trace GC: Method Parameters Out of Range Warnings Not Properly Displayed
KPR#:667210 Thermo Fisher Trace GC: AI/AS 1600 Autosampler Missing Info in Module List
KPR#:667219 Thermo Fisher Aquion: Imperfect Method Parameters Input Validation
KPR#:667312 Thermo Fisher Integrion: Waveforms Cannot Be Edited at First Trial
KPR#:675175 Acquisition delay is reset to 0 after reconnecting
KPR#:682958 Thermo SII: ICS-6000 Acquisition Method Parameter Not Correctly Shown
KPR#:684251 Thermo SII: Not Showing Online Signals Outside a Run
KPR#:703570 Thermo Fisher Trace GC, Integrion, Aquion: Method Part of the UI Not Accessible Due to Missing Scrollbar
KPR#:708978 uGC Channel status disappear from OpenLab CDS Aquisition
KPR#:710021 Don't support Aux trace feature on Openlab CDS
KPR#:735528 Method_Acqusition delay makes all peaks move backward
KPR#:758653 VICI valve can't configure successfully
KPR#:760039 No injection delay in EZ chrom 04.10
KPR#:763799 Hitachi ChromasterUltra RS: Selection in Tray A Display Also Reflects in Tray B of Microplates
KPR#:763901 Hitachi ChromasterUltra RS: Selection Column by Column on Tray Display Not Supported
KPR#:801393 SII Module Status May Show As Not Ready in Idle State
KPR#:948202 SII module cannot be selected during configuration
KPR#:978882 The color of Actuals for Gasifier should be updated after plug the USB cable.
KPR#:1038023 Truncation in Pump Limit Field Makes Input Semi-Invisible
KPR#:1043038 Hitachi Chromaster: Name of 2nd Channel Changes to None Outside Run in Online Signal View
KPR#:1052659 Dashboard shows pump as idle when running out of rear seal wash solvent
KPR#:1074052 Hitachi ChromasterUltra RS: Selection in Tray A Display Also Reflects in Tray B of Microplates
KPR#:1074053 Hitachi ChromasterUltra RS: Tray Display Vial Format Requirements
KPR#:1074165 OpenLab CDS--Method save must be before download.
KPR#:1080182 Temperature of Gasifier is blank when Status is not Ready
KPR#:1082987 Version on "About" page not consistent with actual installed version
KPR#:1084188 Difference driver version between Configuration and Windows Control Panel (On ChemStation C.01.10 Update07)
KPR#:1091615 The updated value for Sampler is not displayed in reopened method.
KPR#:1092105 VICI 2 and VICI 3 related entries display on method audit trail even instrument has only VICI 1
KPR#:1173424 Thermo Fisher SII: Open file dialog moved to back when tooltip is shown
KPR#:1375236 PE Flexar: Incorrect Driver Version in Module List


KPR#:1  Product:OpenLAB Scheduler  OpenLAB Scheduler  A.02.01

Keyword: 

One-line Description:

Resultset locked after run started by Sample Scheduler

Problem:

Resultsets that were run from Scheduler were blocked from being reprocessed until another run was started on the same instrument

Temporary Solution:

Schedule a dummy run on the same instrument to release the resultset

Fix Information:

Fixed in Sample Scheduler for OpenLAB CDS EZChrom A.02.02


KPR#:1  Product:Prep LC  Varian Prep LC  A.01.01

Keyword: 

One-line Description:

Pump settings not set to initial value after run and/or after loading a Method

Problem:

During run a stop is requested to the Software.
Opening the User Interface where the system is shown the Initial pump parameters were not loaded.
It was still  A: 99.84%  B: 0.16% instead of A:100% and B:0%.
Loading the Default method still the gradient is wrong displayed in UI.

Temporary Solution:

n/a

Fix Information:

there's any fix for this issue.
The simptoms was very seldom and random.
Restarting the PC and the Modules will erase the wrong information displayed on the User Interface.


KPR#:2  Product:OpenLAB Scheduler  OpenLAB Scheduler  A.02.01

Keyword: 

One-line Description:

Scheduling Back Tower run on a 6890 or 7890 GC with only a Back Tower installed causes error and hanging EZChrom

Problem:

Scheduling a Back Tower run on a 6890 or 7890 GC with only a Back Tower installed causes error and hanging of EZChrom

Temporary Solution:

As a workaround schedule the Back Tower run as Front Tower run with the correct vial number (e.g. 201)

Fix Information:

A fix is planned for a later version of Sample Scheduler together with a later version of CDS


KPR#:3  Product:OpenLAB Scheduler  OpenLAB Scheduler  A.02.01

Keyword: 

One-line Description:

Scheduling a dual Gas Sampling Valve injection on a 7890B

Problem:

Scheduling a dual Gas Sampling Valve injection on a 7890B get stuck in "in progress" state with same vial number for Front and Back.
When scheduling with 2 different valid vial numbers the run gets scheduled a second time and the data from the first run is lost.

Temporary Solution:

Schedule a dual injection from Gas or Liquid Sampling Valves without vial numbers. 
To test the method in OpenLAB EZChrom create a sequence, and select Dual as Tower.

Fix Information:

Fixed with GC driver B.01.01
Scheduling one dual injection from Gas or Liquid Sampling Valves with 2 different vial numbers is injecting dual once.
Scheduling a dual injection from Gas or Liquid Sampling Valves with the same vial number for both the GC is running twice.


KPR#:4  Product:OpenLAB Scheduler  OpenLAB Scheduler  A.02.01

Keyword: 

One-line Description:

LIMS Agent can't write to trash if LimsCommandId has the wrong type

Problem:

If in the SQL query, the LimsCommandId gets mapped to a string, the LIMS Agent fails to write to the trash file because of the way it constructs the path.

Temporary Solution:

n/a

Fix Information:

This defect is planned to be fixed in a later revision of Sample Scheduler


KPR#:4  Product:MicroGC490  MicroGC490  N/A

Keyword: 

One-line Description:

Detector status not shown in OLCDS driver when detector is set to 'off' or "PeakSimulation"

Problem:

TCD status check in the OpenLab, when TCD is turned off, but it still has a number. 
When TCD is setup as Peak simulation, it shows the "Det off" in Openlab but not "peak sim".

Temporary Solution:

n/a

Fix Information:

It has fixed in 1.11.2.


KPR#:5  Product:MicroGC490  MicroGC490  1.11

Keyword: 

One-line Description:

OpenLabCS C.01.07 Print method does not include any 490 parameters

Problem:

OLCS C.01.07.107 on CBC-Win81 with 490 Driver 1.11.2.1048. Print method. The method parameters of the 490 are not printed in the file.

Temporary Solution:

n/a

Fix Information:

Fixed in 1.11.2.1109.


KPR#:5  Product:OpenLAB Scheduler  OpenLAB Scheduler  A.02.02

Keyword: 

One-line Description:

Uninstall of the Sample Scheduler with the installer msi is sometimes not successful

Problem:

The uninstallation of the Sample Scheduler with the installer msi is sometimes not successful.

Temporary Solution:

Do the uninstallation of the Sample Scheduler in Windows Programs and Features.

Fix Information:

A fix is planned for later releases.


KPR#:6  Product:MicroGC490  MicroGC490  1.11

Keyword: 

One-line Description:

490 on C.01.07 has invalid injection source: only Manual

Problem:

OLCS C.01.07.107 on CBC-Win81 with 490 Driver 1.11.2.1048. With Autosampler set to None in instrument config, the only option to start a run is Manual, which holds it limbo forever waiting. Can't abort.

Temporary Solution:

n/a

Fix Information:

Fixed in 1.11.2.1171


KPR#:7  Product:MicroGC490  MicroGC490  1.12

Keyword: 

One-line Description:

Invert signal function disable when upload method from Micro GC

Problem:

Edit a methods and enable invert signal, now download methods to micro gc 490, then using another PC with OpenLab to control micro gc 490 and upload method, the invert signal is disabled.

Temporary Solution:

n/a

Fix Information:

Fixed in 1.12.2


KPR#:9  Product:MicroGC490  MicroGC490  1.09

Keyword: 

One-line Description:

Sequence hang up after first run is finished

Problem:

The sequence process hang up after the first run is finished, GC is ready but the next run won't start at all.
The problem is found in both Openlab EZChrom and Chemstation software.

Temporary Solution:

n/a

Fix Information:

Fix in B 01.12.2


KPR#:10  Product:MicroGC490  MicroGC490  1.12

Keyword: 

One-line Description:

No Sampling time checkbox in OpenLab Chemstation/EZChrom when using continuous flow

Problem:

Configure sample mode to Continuous flow, the sample time is disabled, it's not comparable with PROStation.
Add the sampling time checkbox in OpenLab Chemstation / EZChrom software.

Temporary Solution:

n/a

Fix Information:

Add the sampling time checkbox in 1.12.2


KPR#:11  Product:MicroGC490  MicroGC490  N/A

Keyword: 

One-line Description:

Invert signal function is not working properly

Problem:

Check the invert signal check box, save method and then download to 490 Micro GC. Start a run, it is unable to invert the signal of the TCD.

Temporary Solution:

n/a

Fix Information:

Fix in 1.12.2


KPR#:11  Product:CTC PAL3  CTC PAL3  A.01.01

Keyword: 

One-line Description:

PAL3 method fails at runtime when an inject amount is out of range for the installed syringe

Problem:

If you have a 10 ul syringe installed and specify a 10 ul injection volume, the system will not indicate any problem until runtime. This is because the driver cannot look into the script to determine total volume (ie sample volume and air gap). This calculation is done at run and the error shows then.

Temporary Solution:

Use has to calculate total volume required when setting up the method.

Fix Information:

n/a


KPR#:12  Product:MicroGC490  MicroGC490  N/A

Keyword: 

One-line Description:

Check Remote access with UserDataStore in control panel, pops up "Unable to active UserDataStore, MFC71.dll is missing "

Problem:

Check Remote access with UserDataStore in control panel, pops up "Unable to active UserDataStore, MFC71.dll is missing "

Temporary Solution:

n/a

Fix Information:

Fix in 1.12.2


KPR#:13  Product:MicroGC490  MicroGC490  1.12

Keyword: 

One-line Description:

Online signal freeze after network reconnection at Openlab platform

Problem:

Cable disconnect during the running, online signal stopped.  After reconnecting the cable, communication came back normally, but the online signal still freezed.

Also checked the same issue on Openlab Chemstation.

Temporary Solution:

n/a

Fix Information:

Fix in 1.12.2


KPR#:15  Product:MicroGC490  MicroGC490  1.12

Keyword: 

One-line Description:

Legacy mainboard can't drive the new VICI valve (EUTA)

Problem:

Legacy mainboard with Firmware less than 2.44 can't support the new VICI valve (EUTA)

Temporary Solution:

Upgrade the driver

Fix Information:

n/a


KPR#:17  Product:MicroGC 990  MicroGC 990  2.02

Keyword: 

One-line Description:

Failed to configure 990 in Chemstation C.01.10

Problem:

OpenLab Chemstation C.01.10.239
Mirco GC Driver 2.2.2
Try to configure 990, in pops up window, enter correct IP, clcik "OK" button, back to instrument configuration, at the bottom, there is warning message "unconfigured moduel here" (actually configured already).

Temporary Solution:

NA

Fix Information:

Fixed in Driver 2.2.5


KPR#:18  Product:MicroGC 990  MicroGC 990  1.03

Keyword: 

One-line Description:

Driver 2.2.1 running at 990 FW 1.03.71: Failed to send the configuration to the GC

Problem:

Communication established successfully with 990, if any configuration (disable channel, flush cycles, pressure unit...) changed and applied, a warning message pops up: communication lost with XX: failed to send the configuration to GC.
Driver 2.2.1
FW 1.03.71
Openlab 2.5

Temporary Solution:

NA

Fix Information:

Fixed in 2.2.5


KPR#:19  Product:MicroGC490  MicroGC490  1.12

Keyword: 

One-line Description:

TCD stays OFF even method download

Problem:

TCD stays OFF even method download, but web page indicates the TCD has been calibrated well.

Temporary Solution:

n/a

Fix Information:

Fixed in 1.12.3


KPR#:26  Product:MicroGC 990  MicroGC 990  2.02

Keyword: 

One-line Description:

Gasifier can't be found in 990 configuration

Problem:

Configure Gasifier from 990 Micro GC, Openlab CDS is 2..6, driver is 2.2.7.0, but Gasifier can't be found in configuration.

Temporary Solution:

n/a

Fix Information:

Fixed in 2.2.11


KPR#:27  Product:MicroGC490  MicroGC490  2.02

Keyword: 

One-line Description:

990 Micro GC can not configure VICI valve

Problem:

Try to configure VICI valve, and click "Check VICI communication", warning message pops up "Gasifier configured, do not allow to configure VICI valve" see below screenshot. Actually, no gasifier connected.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:28  Product:MicroGC 990  MicroGC 990  2.01

Keyword: 

One-line Description:

GC goes offline when starting a run while the VICI valve physically disconnect

Problem:

Configure VICI valve but not physically connect to a valve, the Instrument Status goes offline when starting a single run or sequence.

Temporary Solution:

n/a

Fix Information:

Fixed in 2.3.6


KPR#:29  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Easy Sequence Setup not refreshing the Injection Location

Problem:

7890/7693: Create methods for front, back, and dual ALS injection sources.
Easy Seqeunce Setup does not refresh to show the correct injection location of the loaded method until user loads the corresponding method in ChemStation.

Even loading a different method, the previous injection source is kept.

Temporary Solution:

If the injection source has been changed, use the "Clear All" button and reload the method.

Fix Information:

n/a


KPR#:30  Product:MicroGC490  MicroGC490  1.12

Keyword: 

One-line Description:

VICI failed the first activation with uGC, need 2nd time to activate

Problem:

Driver 1.12.3.008
Using the VICI valve EUTA model, open configuration to activate the VICI, it failed at first time. need to re-activate the VICI.
It will result in the broken communication between VICI and uGC after system rebooting, SW will crash

Temporary Solution:

n/a

Fix Information:

Fixed in Driver 2.4.8


KPR#:31  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

No error checking for vial ranges in Easy Sequence Setup; fails at runtime

Problem:

There is no error checking within Easy Sequence Setup to determine if the vial range of the loaded method's injection source is accurate.

The method can be saved and submitted to the queue.  It fails at this time with:

'---------------------------
Error
---------------------------
Invalid location found in FRONT sequence, line 1
There may be others
---------------------------
OK   
---------------------------'

This defect occurs because there is a difference between the loaded method's injection source and the method that is selected in the Easy Sequence Template.

Temporary Solution:

Load the method in ChemStation from Method > Load Method menu that will be used to create the Easy Sequence Setup Template.

Fix Information:

n/a


KPR#:32  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Pause/Resume button is in the wrong state after a sequence is edited and then pauses

Problem:

Send a sequence to the Queue and while it is running, edit it. This modified/changed sequence pauses the Queue, but the first blue button is still in a 'Pause||' state. Click the button and change it to Resume, but it does not do anything as it is waiting for the current sequence to be saved before starting the next sequence.

In this case, because the sequence has been modified, the ChemStation is not accepting sequences. If you click pause || button, you pause the queue and it does not do any thing. The message on the active queue is red and the message says the sequence has been modified.

Temporary Solution:

Save the sequence that has been modified.  Once the sequence has been saved, the queue will resume running the next sequence.

Fix Information:

n/a


KPR#:33  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

7890A with no ALS shows Read Barcode button in Easy Sequence Sample List

Problem:

Load an Easy Sequence.es for a valve or HeadSpace Injection Source and the Sample Name cell in the Sample List shows a Read barcode button.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:34  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

HeadSpace Simple Calibration Blank type should only allow 'Blank with injection'

Problem:

HeadSpace software will not run a sequence without a vial location.  Therefore, the Easy Sequence Setup\Calibration\Simple Calibration options should only allow Blanks with Injection and not allow blanks without injection.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:35  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Calibration Blank with no injection loads with Inj Vol as 0

Problem:

Create a setup with a blank with injection and a blank with no injection.
Save and load to Easy Sequence. The Injection volume for the Blank with no injection is entered as 0. Native sequence table also shows this as zero.
This is not consistent with the volume for the blank with injection

Temporary Solution:

n/a

Fix Information:

Fixed in revision 2.2


KPR#:39  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Calibration table clears on method reload from Samples tab

Problem:

Create a Setup with Calibration. Return to Sample tab and load the same or a different method with the same injection location. Return to Calibration tab and all has been cleared. Reloading a method does not clear entries made on the Samples tab.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:40  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Queue displays running vials incorrectly when initial HeadSpace vial not found

Problem:

Start a 3 line headspace sequence from the queue, but remove the FIRST vial in the sequence. Queue does not update correctly, showing the first vial as running, though the HeadSpace has moved on to vial 2.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:41  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Aborted HeadSpace run displays Status as 'Complete' in Queue History

Problem:

HeadSpace aborts, but ChemStation does not. Select the Abort Run button from the Waiting for ready dialog and ChemStation aborts the sequence.
Queue History shows this run as 'Complete' instead of 'Aborted'.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:44  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Easy Sequence does not advise of invalid vial locations until sequence is fully loaded into Queue

Problem:

There is no advisory about the invalid vial locations until the sequence has been added to the queue.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 2.3


KPR#:45  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Sequence submitted to queue but will not run and can not be removed/deleted

Problem:

Sequence submitted to the queue but will not start and cannot be deleted.

Temporary Solution:

As a workaround, delete the following file:  SequenceQueue.que from the C:\Chem32\4\TEMP directory.

Fix Information:

Fixed in revision 2.4.


KPR#:47  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Injection Volume on ES Setup should show units (uL); valved inj are in mL and user must convert

Problem:

The Easy Sequence Setup dialog should show Injection volume units, which in ChemStation always seem to be uL. 

When the user uses EasySequence with a Valved injection, the volume, which for valves is mL, must be multiplied by 1000 to get a value that will be accepted.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:48  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Easy Sequence Setup should check for Sample Name length and not allow invalid entries

Problem:

Enter a long Sample Name in Easy Sequence Setup and add a counter.
Save and load to Easy Sequence. Fill Samples. Not until then is the user warned (with a red X) in the Sample list that there is a 40 character limit.
It is possible to save this but it will not be possible to add to the Queue.

Temporary Solution:

n/a

Fix Information:

Fixed in Easy Sequence 2.4[014].


KPR#:51  Product:Easy Sequence  Easy Sequence  2.2

Keyword: 

One-line Description:

Vial Status shows first run as completed for Easy Sequence following Run Method

Problem:

Run a single vial using Run Method.

Send sequence to queue vials 2-4. When the run method completes and the sequence loads and then the first vial in the sequence starts, Vial status shows the first line (vial 2) as immediately 'completed'.

The system stalls at Sequence running.  The GC is in prerun state and the system cannot start another run until it is shutdown and restarted.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:52  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Easy sequence actions are not included in the audit trail and no separate audit log exists

Problem:

Any changes done on Easy Sequence are not included in the audit trail; there is no separate audit log present for Easy Sequence.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:53  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Vial not found reported in software Status not consistent with instrument status

Problem:

Run an Easy Sequence using a HeadSpace injection source. When a Vial not Found event happens, the final Status for this run, which did not start a run, is reported as 'Completed'. The status being reported by the HS device is 'Run Aborted'.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:63  Product:Headspace Driver Software  RC.Net  B.01.01

Keyword: 

One-line Description:

The GC Enters Prerun Too Early for Splitless Modes and Pulsed Split

Problem:

Set up a splitless (or pulsed split) method. The 7890 goes into pre-run at the beginning of the sample movement, rather than just prior to injection. This means for long vial equilibration times that the inlet is not being purged.

Temporary Solution:

Do not use splitless mode with the 7694 and G1888A headspace samplers.

Fix Information:

This will work in the 7697A headspace sampler with a minimum firmware revision of A.01.02.  The user needs to deselect the "Manual Run on Prep Run Request" on the 7890 configuration screen to make it work.


KPR#:101  Product:Headspace Driver Software  RC.Net  B.01.01

Keyword: 

One-line Description:

In EZChrom, it is Possible in Some Cases that the 7890A GC Aborts a Sequence but tne Headspace Will Continue On

Problem:

This occurs in Open Lab EZChrom Version A.04.02 workstation and earlier when  Sequence Action Continue for the 7697A or Proceed with the G1888/7694B in EZChrom

Temporary Solution:

Do not use Sequence Action Continue for the 7697A or Proceed with the G1888/7694B in EZChrom

Fix Information:

n/a


KPR#:158  Product:68xx Driver Software  6890  B.01.01

Keyword: 

One-line Description:

User is not able to save 'Column Parameter' changes at the same time when change is made

Problem:

User is not able to save 'Column Parameter' changes at the same time when change is made unless they click on another field on the screen

Temporary Solution:

n/a

Fix Information:

Resolved in B.01.02


KPR#:160  Product:68xx Driver Software  ICF (68xx)  6.33

Keyword: 

One-line Description:

Invalid Resource Item implementation for oven program

Problem:

Invalid Resource Item implementation for oven program

Temporary Solution:

n/a

Fix Information:

Resolved in B.01.02


KPR#:617  Product:RTE  RTE  A.02.04

Keyword: 

One-line Description:

Exponential calibration curve plot isn't shown in Calibration Report when using Intelligent Reporting.

Problem:

Exponential calibration curve plots cannot be reported with Intelligent Reporting. An empty plot is shown instead.

Temporary Solution:

If exponential calibration curves plots need to be reported, the (old) classic reporting needs to be used in ChemStation. 
EZChrom users need to use the EZChrom standard reporting.

Fix Information:

A potentially fix is currently under investigation.


KPR#:138601  Product:    

Keyword: 

One-line Description:

LC detector lamps and column ovens do not turn off when using "Shutdown" sequence run types.

Problem:

LC detector lamps and column ovens do not turn off when using "Shutdown" sequence run types.

Temporary Solution:

Manually shutdown lamps and column ovens when required.

Fix Information:

Fixed in OpenLAB CDS A.04.05.


KPR#:138603  Product:    

Keyword: 

One-line Description:

"Additional Reports" cannot be viewed from EZChrom instrument interface

Problem:

When specifying "print additional reports" as a sequence runtype, aquiring a sequence and then reviewing it, there are no sequence reports listed when selecting Reports, View, Sequence from the client.

Temporary Solution:

Browse to the folder where the report was stored and open the report to view.

Fix Information:

Fix is currently under investigation.


KPR#:138604  Product:    

Keyword: 

One-line Description:

GC triggered by external device starts prior to platform readiness

Problem:

When using an Agilent GC that triggers an external device that is not part of the configuration, the trigger may happen prior to the system being ready to acquire data. This may result in the data system not correctly triggering for acquisition.

Temporary Solution:

Use samplers and devices supported within the configuration of the GC.

Fix Information:

Fix is currently under investigation.


KPR#:138605  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Values for the Instrument Modules snippet not shown in the report

Problem:

Added 'Instrument Modules' snippet to report template.  The part number, serial number, and firmware values are not shown in the report.

Temporary Solution:

n/a

Fix Information:

Resolved in OpenLAB CDS EZChrom Edition A.04.07.


KPR#:138607  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Intelligent Report InjOrderNumber issue

Problem:

In the Intelligent Report, the InjOrderNumber increses from 1..2..3. till the number of runs in the sequence. This is acting as an injection counter instead of acting as the number of injections/vial.

Temporary Solution:

n/a

Fix Information:

Resolved in OpenLAB CDS EZChrom Edition A.04.07.


KPR#:138608  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

OLCDS EZChrom IR should report both parameters: Peak Amount and Peak Concentration (corrected by multipliers)

Problem:

Right now only Peak Amount is recorded in IR.  Peak Concentration is not recorded.

Temporary Solution:

n/a

Fix Information:

This is planned to be fixed in OpenLAB CDS EZChrom A.04.05 Update 1.


KPR#:138609  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

6890 GC ALS cannot be deactivated by entering blank vial number

Problem:

When using GC 6890 with ALS and built-in valves, the ALS injection  works fine.

While using the valves, we give the input of a vial "0" or leave vail position blank so that the ALS wil not expect a vial/sample and the GC expects a manual start from the GC keyboard (Start run). However the ALS does not wait for trigger and immediately starts the run.

Temporary Solution:

Configure another instrument by disconnecting the ALS and then submit run while using the valves. Then the GC waits for trigger and the run starts only when the GC start button is pressed.

Fix Information:

This issue is resolved with an update to the GC driver.  The driver that is included with the A.04.06 release of EZChrom contains the update.  Using the EZChrom application, look for the drivers version currently in use.  It should say 68xx GC 6.23 (009).  The update is included in version 6.21 and higher.  This update is not currently available in any driver version that is validated with EZChrom A.04.05 (with any software patch) or earlier.  Although additional Software Updates may be available for A.04.05 in the future, there are none currently scheduled.


KPR#:138611  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Cannot do dual channel data collection on 5890

Problem:

When you want to collect both the front and back detector channels into a single data file, you set the signal source in Instrument set up for detector 1 and detector 2.  In both A.04.04 and A.04.05, both signals will change back to the channel that you last setup when you save the method.

Temporary Solution:

It may be necessary to use EZChrom Elite in a Workstation mode to control the 5890 in this manner.  The instrument is long out of support, and the intention for OpenLAB CDS support is only for use in a Read Only manner.

Fix Information:

There are currently no plans to enhance the OpenLAB CDS support for the 5890 to be able to fully control the instrument.


KPR#:138612  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

A.04.05 - Inst Client Crash on Windows 7 When Processing Result with Method Report window open

Problem:

When reprocessing with the Method Report window InstClient crashes with an error saying "Instrument client has stopped working".

Temporary Solution:

Make sure the Method Report Window is closed during reprocessing

Fix Information:

This will be fixed in OpenLAB CDS EZChrom A.04.05 Software Update 3.


KPR#:138614  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Intelligent Reporter Not Displaying per Peak Custom Parameter for Unnamed Peaks

Problem:

Intelligent reports in EZChrom (IR and OLIR) do not display values for per peak custom parameters for unnamed peaks (unknown peaks). The custom parameter value displays correctly in the EZChrom method report. This behavior is different from the behavior in OpenLAB ICM where values for per peak custom parameters displayed for all peaks in an injection, named and unnamed.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB CDS EZChrom Edition A.04.06.


KPR#:138615  Product:    A.04.04 (Lamorak)

Keyword: 

One-line Description:

The Intelligent Report cannot generate a report for a Single Data File

Problem:

When user opens the single Data file in EzChrom and try to preview the related data in an Intelligent Report(IR), the IR rendered the demo data instead of data in loaded data file. 

This is because the IR required the Result Set (ACAML file) to render the data in loaded data file.

This is the expected behavior in EZChrom system.

Temporary Solution:

Intelligent reporting should always be used in a sequence summary context. The *.RDL files can still be of "single run" type, but you would always reports across the entire sequence.

In order to get a single run Intelligent Reports for single, dedicated runs use the following workflow:
1. Load a result set
2. Change the run type of the samples runs you like to report to Begin summary (and select the template you like to use), Summary run and/or End Summary
3. Reprocess the entire sequence or the selected runs only (this will update the sequence ACAML file)
4. Go to the Reports tab> View > Sequence Report:  The dialog would display all the runs you have selected for the summary and the report template name
5. Press Preview report

This workflow allows you to use a specific summary template for e.g. System suitability runs and e.g. Single run templates for the unknowns.

Fix Information:

A solution is being evaluated for a future release of OpenLAB CDS EZChrom.


KPR#:138616  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

The Spectrum Similarity Table Does Not Show The Correct Colors

Problem:

The Spectrum Similarity Table does not show the correct colors when overlaying different spectra.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB CDS EZChrom Edition A.04.06.


KPR#:138617  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Saved PDF reports inside SSIZIP May not Reflect The Latest Revision Inside ECM

Problem:

When reprocessing saving PDF reports inside a result set is an optional field. If a user does not choose to save the report inside the result set during reprocessing but the option was selected during acquisition, you will only have reports from the previous revision in ECM.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB CDS EZChrom A.04.05 Software Update 5.


KPR#:138618  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Report prinout sizes were different based on different printing options in EZChrom

Problem:

The difference in printout sizes were observed in EZChrom A.04.05 when manually printing versus automatically printing.  The following setup was used to produce the printouts.

For Manual printout:
1. The "Result review mode" window is opened and displayed
2. Open the report of the method (method -custom report)
3. Print the report by doing Report>Print>Method report.

For automatic printout:
1. Reprocess the sequence and select "print hard copy" of the"method report"
2. The report is printed also but with a different format. The report is smaller, and the margins are bigger.

In conclusion there are two different report sizes.

Temporary Solution:

n/a

Fix Information:

A fix is planned for a future release of OpenLab CDS EZChrom.


KPR#:138619  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Automation Guide instructions for no-vial analysis does not work

Problem:

The Agilent OpenLAB CDS EZChrom Edition (version A.04.05 and A.04.06) Automation Guide includes a couple of ways to submit a no-vial analysis using the SubmitSingleRun method.  This is in page 58:

PropInitialVial Property - Sets or returns the vial that will be used for the first injection of a sample in a single run by the SubmitSingleRun() method. This is a member of the
SingleRun object. To specify no vial, the ezAutosamplerNoVial constant may be used.

PropInitialVialText Property - Sets or returns the vial that will be used for the first injection of a sample in a single run by the SubmitSingleRun() method. This is a member of the
SingleRun object. To specify no vial, an empty string may be used.

Both scenarios does not work.

Temporary Solution:

The ability to do no vial runs using automation was a feature in EZChrom Elite.  That went away with the new drivers so the feature is no longer available in the Open CDS EZChrom Editions.

A no-vial analysis can still be done, but in the OpenLAB CDS EZChrom Edition application by submitting a run leaving the vial field blank in the sequence.

Fix Information:

There will be a correction to remove the appropiate sections in the manuals in a future release of OpenLAB CDS EZChrom Edition.


KPR#:138620  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Doing two RefreshInstruments causes the instruments object to disappear

Problem:

We found that RefreshProjects and RefreshInstruments functions in the automation interface do not work as expected. Those function can only be called once.

We found that the Enterprise object must be cached after logging on. When getting the Enterprise object multiple times from the Application object, the automation interface does not close properly.

Since we know what not to do on the automation interface, we can work around this in our code. If we close this issue it is likely that the scenarios described here are not resolved or tested. I consider this up to you since we have workarounds in place but from my perspective the functions are broken.

Temporary Solution:

Do not do two RefreshInstruments.

Fix Information:

Resolved in OpenLAB CDS EZChrom A.04.07 SR2.


KPR#:138621  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

When performing "Add Runs to a Result" , the activity does not show in Result Set Sequence Audit Trail.

Problem:

The Result Set Sequence Audit trail does not show any entries when adding runs to the existing result sequence, even though the result sequence has changed.  Other changes in entries such as Sample ID, Run type, etc. do show up.   Some customers use this feature to keep using one Result Set to perform multiple runs without having to have separate files.

Temporary Solution:

For auditing and traceability purposes, the user will have to rely on a combination of the Sequence Last Modified Time and the Sequence Table on a Sequence Report that is generated at the end of each sequence run.  When creating a new standard report include both the Sequence Creation Time and the Sequence Last Modified Time.  Also include a sequence table by right clicking the mouse and choose Insert Report and then Sequence Table (customized) or (full). Either table shows the complete list of samples or runs in the sequence, and can be compared to the previous report for changes made.

Note that additional activities or changes made  when running the added Sequence will be recorded in the Sequence Audit Trail and not the Result Sequence Audit Trail because the changes occurred during acquisition.

Fix Information:

Fixed in OpenLAB CDS EZChrom Edition A.04.06.


KPR#:138623  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Result Set Shows Up as Not Being Saved Eventhough It has been Saved.

Problem:

When printing a report using Reports--> Print Method Report after reprocessing a result set you get a message indicating that the results have not been saved eventhough they have been saved.

Temporary Solution:

To print a report in this scenario load another result set after reprocessing and then load the result set reprocessed earlier.

Fix Information:

This is fixed in OpenLAB CDS A.04.05 Hotfix 05.


KPR#:138625  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

method changes not saving in results set

Problem:

What I am saying is that if you only change the traces that show up on a method custom report and save the method, the new method custom report is not saved into the result set. If we make any other changes to the method or to the method custom report, and save the method, those changes are saved into the method in the result set (or the master method depending on what we choose to save)

But if you only change what trace (or channel) is shown on the method custom report and save the method, the method in the result set is not saved.

Just open a result set
Change the trace shown on the method custom report
Save the method
Process the result set
Open the method custom report
You will see the trace was not changed

Temporary Solution:

There is no workaround at this time.

Fix Information:

n/a


KPR#:138626  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Number Of Reps in A Single Run Is Not Being Handled Correctly

Problem:

You get an error message when you try to start a single run when using reps.  This is due to the fact that reps are not being handled correctly when using a Single Run.

Temporary Solution:

Do not use reps when using a single run.

Fix Information:

Email thread:
-------------------------------------------
From: LINDAHL,JOHN (A-USA,ex1) 
Sent: Saturday, April 23, 2016 8:09 AM
To: MCCULLEY,JENNIFER (A-USA,ex1) <jennifer_mcculley@agilent.com>
Cc: JAIN,BABITA (A-Americas,ex1) <babita.jain@agilent.com> GOLDBERG,SETH (A-SantaClara,ex1) <seth_goldberg@agilent.com>
Subject: RE: Waiting for verification.xlsx

Correction. They are closing as verified from our review here.


From: LINDAHL,JOHN (A-USA,ex1) 
Sent: Saturday, April 23, 2016 8:05 AM
To: MCCULLEY,JENNIFER (A-USA,ex1) <jennifer_mcculley@agilent.com>
Cc: JAIN,BABITA (A-Americas,ex1) <babita.jain@agilent.com> GOLDBERG,SETH (A-SantaClara,ex1) <seth_goldberg@agilent.com>
Subject: RE: Waiting for verification.xlsx

Hi Jennifer,

I reviewed the list and found one already closed and one that I feel needs verification (222978). I assigned that over to Nitin for the HF6 project.

The rest are closing as DNF.

115120,116117,130473,142304,146057,147748,157630,161372,162010,174035,182152,184930,186547,186549,186550,186553,195173,197027,198564,202789,207114,209281,211805,213045,220260,220931

John


 
From: MCCULLEY,JENNIFER (A-USA,ex1) 
Sent: Wednesday, April 20, 2016 6:32 AM
To: LINDAHL,JOHN (A-USA,ex1) <john_lindahl@agilent.com>
Cc: JAIN,BABITA (A-Americas,ex1) <babita.jain@agilent.com> GOLDBERG,SETH (A-SantaClara,ex1) <seth_goldberg@agilent.com>
Subject: RE: Waiting for verification.xlsx

Most if not all are DNF that we missed to close them out entirely.   Therefore we should just close them if DNF works than I am fine with this.  

Regards,

Jennifer


 
From: MCCULLEY,JENNIFER (A-USA,ex1) 
Sent: Tuesday, April 19, 2016 2:11 PM
To: JAIN,BABITA (A-Americas,ex1) <babita.jain@agilent.com> LINDAHL,JOHN (A-USA,ex1) <john_lindahl@agilent.com> GOLDBERG,SETH (A-SantaClara,ex1) <seth_goldberg@agilent.com>
Subject: Waiting for verification.xlsx
 
There are defects waiting for verification due to the defect clean-up project we did last year.   Since we just had a full suite of testing done on a.04.07 SR2 and no response to the defects, please do a mass update with the x-functional team approval to close the defects.   
 
Thank you.
 
Jennifer
-------------------------------------------


KPR#:138627  Product:    A.04.06 (Tristan 3B)

Keyword: 

One-line Description:

Entry for error, when failing to upload results to Data Store,  is not logged within Instrument Activity Log while acquiring data.

Problem:

A log entry does not get inserted into the Instrument Activity log when data does not get uploaded due to the Data Store path not being found or the Alfresco Tomcat Service being stopped.

Temporary Solution:

None

Fix Information:

A fix is being evaluated for a future release of OpenLAB CDS EZChrom.


KPR#:138628  Product:    A.04.06 (Tristan 3B)

Keyword: 

One-line Description:

Only the latest revision of a Method Can Be Used For Acquisition

Problem:

When using OpenLAB CDS EZChrom with ECM or Datastore a user can only use the latest revision of a method for acquisition.

Temporary Solution:

OpenLAB CDS EZChrom allows running only the latest revision of a method for acquisition. If a user wishes to perform acquisition with an older revision of a method, the user must save the method before the run.

Fix Information:

n/a


KPR#:138629  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

When adding samples to a running sequence while the Instrument Monitor is OFF, the added new samples do not get processed in the actively running sequence.

Problem:

When editing a running sequence such as adding samples while the Instrument Monitor is set to OFF, the added samples do not get processed in the actively running sequence.  But in the EZChrom Elite version, changes made while the Instrument Monitor is OFF are taken into effect for the active sequence, and the added samples are processed. 

Here's an example of a workflow that describes the problem in OpenLAB CDS EZChrom Edition:
1. Run the sequence
2. Disable the Instrument Monitor (Done by going to Control and uncheck Monitor "Instrument Name"
3. Add new samples in a timely manner
4. Save the modified sequence
5. Enable the Instrument Monitor
6. Although the new samples are visible, they DO NOT get processed
7. But new samples added while Instrument Monitor is ON does get processed.

Temporary Solution:

The current workflow of not being able to edit a sequence with the instrument monitor off is by design in OpenLAB CDS EZChrom Edition.  It was made different from the old version, because in EZChrom Elite the function presented a risk and a compliance issue; a user from any other instrument was able to make accidental or unwanted changes to the running sequence.

In the OpenLAB CDS EZChrom Edition the sequence was separated so that one is running on the AIC, and the other one, when Instrument Monitor is OFF, is opened from storage.  In addition to this change, another new feature allowed different instruments to run the same sequence concurrently.

Suggested Workaround for Issue:  Make changes to the sequence while the Instrument Monitor is ON.  The user should be able to perform all other tasks while Instrument Monitor is OFF.

Fix Information:

There is no plan to re-design the sequence editing function for changes made while the Instrument Monitor is OFF to be implemented and processed in the running sequence.

But we do recognize the confusion of the user interface when the Instrument Monitor is back ON, which shows the changes made to the sequence while the Instrument Monitor is OFF.  We will evaluate a change to make the user interface less confusing in a future version of OpenLAB CDS EZChrom Edition.


KPR#:138630  Product:    A.04.06 (Tristan 3B)

Keyword: 

One-line Description:

Hitachi Primaide Detectors may not Get Configured

Problem:

A Hitachi Primaide detector is detected when you run the Hitachi Maintenance program but may not get configured when you run autoconfigure when configuring an instrument on OpenLAB CDS EZChrom A.04.06.

Temporary Solution:

n/a

Fix Information:

There is an updated Hiatchi Primaide driver coming out shortly for OpenLAB CDS EZChrom A.04.06 which will fix this.  Inquire about this update before installing a Hitachi Primaide with OpenLAB CDS EZChrom A.04.06.


KPR#:138631  Product:    A.04.06 (Tristan 3B)

Keyword: 

One-line Description:

Error trying to display the Instrument Status Screen For Hitachi LaChrom Elite

Problem:

The instrument status screen for Hitachi LaChrom Elite on OpenLAB CDS EZChrom A.04.06 opens but a blank message box displays on the screen causing the instrument status screen to become disfunctional.

Temporary Solution:

n/a

Fix Information:

There is an updated Hiatchi LaChrom Elite driver coming out shortly for OpenLAB CDS EZChrom A.04.06 which will fix this.  Inquire about this update before installing a Hitachi LaChrom Elite with OpenLAB CDS EZChrom A.04.06.


KPR#:138632  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Use external program before run doesn't work

Problem:

In the Method\Advaned\Files menu, there is a checkbox for Before Run. One can enter an executable here that will be called up before the run. Something like calculator.exe works fine for this. It is called up for EZChrom Elite, however it doesn't work on my OLCDS EZChrom work station.

Temporary Solution:

None as yet

Fix Information:

n/a


KPR#:138633  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Wrong signal runtime acquired if the last sample from a previous sequence is aborted or it is defined as a shutdown sample.

Problem:

Each LC module has the possibility to define its own runtime. That feature should not be used with OpenLAB CDS EZChrom A.04.05 . Set a runtime in the pump and set all other modules with the "as pump runtime"  option.

Temporary Solution:

Do not set method runtime at different LC modules.

Fix Information:

n/a


KPR#:138634  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Instclient.exe crashing during sequence reprocessing due to long Sample ID length while printing in landscape mode from the Method Report

Problem:

Reprocessing of sequence failed at line that has a long Sample ID name in the Method Report..  

The crash occurred with specific settings in the OpenLAB CDS EZChrom's Method Report. For example, choose the Landscape option in the View / Preferences / Page settings options, Load the Result Set and go to Result Set / Process, Set Processing to Use Last Results, and Report to Save as PDF / check the Method box.  Then start reprocessing and wait so see the failure at the line with the long Sample ID.

The following settings will not fail: page setting is not Landscape, user changes the "SAMPLE ID" to something equal or less than a certain number of characters, and if it is not printing or saving to PDF.

Temporary Solution:

Below are two workarounds:

1. Shorten the Sample ID, or other fields that may be causing the problem.

2. Reconfigure the Method Report to allow sufficient space for each field.  In the SSB Problem example, the method report contained a line with multiple fields: Sample ID, Computer Name and User who Acquired the Data.  If the Sample ID had its own line in the method report, there will be no crashing.

The length that is causing problem is not only the sample ID length but the length of the whole line which includes the sample ID, Computer name, User who acquired the data, etc.  It appears that in certain case the initial calculation of the length of the line with respect to available length for that orientation appears to be OK but while rendering the line it is causing the problem.  This rendering is also specific to the way the Method report is configured.

Fix Information:

A fix will be evaluated for a future release of OpenLAB CDS EZChrom.


KPR#:138635  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Strings in exported CDF file produced by EZChrom are truncated

Problem:

Users have reported that several strings in the CDF file that EZChrom generates get truncated so that the file name may not be totally visible. 

The following strings are truncated to certain character lengths when the file path names and or file names are too long:

detection_method_name : truncated to 63 characters
peak_processing_method_name: truncated to 63 characters
sample_name: truncated to 29 characters
sample_id : truncated to 29 characters

Temporary Solution:

Reduce the combined strings of the file path names and file names to be below the character limits for each of the following fields:

detection_method_name : below 63 characters
peak_processing_method_name: below 63 characters
sample_name: below 29 characters
sample_id : below 29 characters

Note that the strings for detection_method_name and peak_processing_method_name are the ones that include the file path.  The other ones do not.

Fix Information:

A solution is being evaluated for a future release of OpenLAB CDS EZChrom.


KPR#:138636  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Error in RCNet_RCASTrayUI when opening a tray view before instrument connection.

Problem:

This is regarding the error: "Error in RCNet_RCASTrayUI:: SetVialColor : Invalid Location Vial 81", that we are seeing in an Agilent LC.  

Software: OpenLAB CDS EZChrom A.04.05
Configuration:  The LC is a 1290 with a G4226A sampler that contains a 100 vial tray.

The error stems from  not being able to set the vial color on the  Control/Autosampler Trays/HiP Sampler view. 
If one opens the instrument and injects from vial 1-10, open the tray view, close and reopen the instrument, no error will occur. 

However, if one injects from a vial outside of the 96Agilent tray range like 81, close the instrument with the tray view still open then reopen the instrument, the error will occur.

Temporary Solution:

The simple workaround would be either to not open the tray view or wait until the instrument connects before opening it and being sure to close the view before closing the instrument. 

Upon opening the instrument, this default tray view (Agilent96) is seen until a connection with the ALS is made, then the correct tray is seen.

Fix Information:

A fix is being evaluated for a future release of OpenLAB CDS EZChrom.


KPR#:138639  Product:    A.04.06 (Tristan 3B)

Keyword: 

One-line Description:

Sequence will NOT reprocess unless all data files reside in the same Result Path folder in a Project set as File mode

Problem:

In older OpenLAB CDS EZChrom versions, users were able to reprocess data files from many different subdirectories.

In OpenLAB CDS EZChrom A.04.06, a sequence will NOT reprocess data files unless they all happen to reside in the same Result Path.  This only occurs if the Project settings is in the "File Mode".


In the "Result Package Mode" for A.04.06, this is not an issue.

Temporary Solution:

When creating a project use the other Project setting: Result Package Mode.

If the project settings change is not acceptable and the user wants to reprocess an aggregate of data files from multiple subdirectories, the user has to copy the data files to a central location and then point the "Result Path" to this location.

Fix Information:

A fix is planned for a future release of OpenLAB CDS EZChrom.


KPR#:138641  Product:    A.04.06 (Tristan 3B)

Keyword: 

One-line Description:

New Sequence Has Read Only Fields When In Result Review Mode

Problem:

If you create a new sequence while you are in result review mode and a result sequence is open, the newly created sequence has read only fields that cannot be edited.

Temporary Solution:

The workaround for this is to close the open result sequence before creating a new sequence.  This will allow a new sequence to be created without read only fields.

Fix Information:

n/a


KPR#:138642  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Unable to report values of "SignalToNoise" and "SignalToNoise6Sigma" in Intelligent Report for Japanese version

Problem:

An Error message "An error occurred during report rendering. The report cannot be reprocessed�."  is displayed when a user previews an intelligent report template after adding the parameters "SignalToNoise_EP" and "SignalToNoise6Sigma" in the Japanese version.

Temporary Solution:

n/a

Fix Information:

A hotfix for A.04.05 will be made for this issue but we do not have any dates yet.


KPR#:138643  Product:    A.04.06 (Tristan 3B)

Keyword: 

One-line Description:

Unable to See Chromatogram when Processing a Sequence with Review-Result Review for Data Store only

Problem:

When the user performs Process Sequence with the Review-Result Review (Pause after each run) turned on, the chromatogram does not pause.  The user presses the down arrow to process the next sample, but the chromatogram pops up quickly for a second then dissappears.

This observation occurs only when the system using Data Store.  It does not occur for other OpenLAB CDS EZChrom systems.

Temporary Solution:

The reprocessed sample chromatograms can be seen at the end of reprocessing.

Fix Information:

Resolved in OpenLAB CDS EZChrom A.04.07 SR2.


KPR#:138645  Product:    A.04.06 (Tristan 3B)

Keyword: 

One-line Description:

Peak Names Truncated in CDF Files When Working With Double Byte Characters

Problem:

In OpenLAB CDS EZChrom A.04.06, peak names are getting truncated in CDF files when working with double byte characters.

Temporary Solution:

n/a

Fix Information:

This will be resolved in OpenLAB CDS EZChrom A.04.06 Software Update 3.


KPR#:138646  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

The Area Ratio Custom Parameter Does Not Work in OpenLAB CDS EZChrom A.04.05

Problem:

In OpenLAB CDS EZChrom A.04.05 the Area Ratio Custom Parameter does not work.  Results do not show up in the report.

The Area Ratio Custom Parameter does work in OpenLAB CDS EZChrom A.04.06.

Temporary Solution:

n/a

Fix Information:

The Area Ratio custom parameter will not be fixed in OpenLAB CDS EZChrom A.04.05.  If this fuctionality is needed, the recommendation is to upgrade to OpenLAB CDS EZChrom A.04.06.


KPR#:138649  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

When creating a sequence to reprocess data with the same names from different Results Set folders, data files are overwritten in the new Result Set folder..

Problem:

When a user creates a sequence that contains data files from different Result Set folders, and if the data files have the same name, those data files in the new Result Set folder are overwritten by the latter ones in the sequence.

For example, if a customer wants to evaluate the standard response of a data point from multiple folders to check the response factor variation, the data files need to have distinctive names.  If all the files names are exactly the same, there will only be one data point, and the new Result Set folder will contain only one data file.

OpenLAB CDS EZChrom does not prevent this and there is no warning to show the user.

A warning and or an Auto-renaming of those files when reprocessing will be a good idea.

Temporary Solution:

The workaround is to manually rename the files (or copy and rename the files to another folder) to be reprocessed.  

If the system is using an Advanced File System (AFS) mode, there is also a manual way to rename the files or create new folders.  The user will have to log into Windows with an Enterprise Service Account, or a member of the AFS group, that has access to the enterprise path under the restrictions of AFS.

Another workaround is to prevent the data files from having the same name by using the Date/Time naming convention when acquiring the data files.

Fix Information:

A solution will be evaluated for a future release of OpenLAB CDS EZChrom


KPR#:138651  Product:    A.04.06 (Tristan 3B)

Keyword: 

One-line Description:

The function Add Single Run to Result does not allow the use of replicates

Problem:

When you use the function "Add Single Run to Result" the replicates field is greyed out.  The use of replicates should be allowed with any type of run addition.

Temporary Solution:

The work-around for this workflow is to use "Add Runs to Result" instead of "Add Single Run to Result".  Create a new sequence with one run and enter in the number of replicates to use.  Then open the Result Set and use the function "Add Runs to Result" and use the new Sequence with the one run.  This will allow the one run to be added along with the replicates.

Fix Information:

A solution will be evaluated in a future release of OpenLAB CDS EZChrom.


KPR#:138652  Product:    A.04.06 (Tristan 3B)

Keyword: 

One-line Description:

Aborting the current run will abort all replicates in a run when using the Single Run option

Problem:

If the user is performing a Single Run with replicates and aborts only the current run, the current run will be aborted along with the rest of the replicate runs.  Only the current run should be aborted and the rest should continue to be processed.

Temporary Solution:

Instead of using the Single Run option, create a Sequence with one run, enter the desired number of replicates (Reps) and perform a Sequence Run.  Then aborting a run will stop only the current run and any replicate after will continue to be processed.

Fix Information:

A solution will be evaluated for a future release of OpenLAB CDS EZChrom.


KPR#:138654  Product:    A.04.05 (Tristan 3A)

Keyword: 

One-line Description:

Instrument window crashes when adding spaces in header of snandard report

Problem:

After adding spaces between a letter and adding spaces in the header of standard report instrument window crashes

Temporary Solution:

Instead of using spaces use TAB as space separator

Fix Information:

n/a


KPR#:138675  Product:    A.04.06 (Tristan 3B)

Keyword: 

One-line Description:

Single Sign on with Automation

Problem:

Single sign on with Automation does not work.

Temporary Solution:

n/a

Fix Information:

Resolved in OpenLAB CDS EZChrom A.04.07 SR2.


KPR#:138679  Product:    A.04.07 SR1

Keyword: 

One-line Description:

Edited sequence during "pause pending" not instantiated

Problem:

1. Create & launch any instrument online.

2. Create & submit two line sequence (with two replicates, Unknown as run type and with filename as 'Date & Time') for acquisition.

3. Pause sequence while first replicate acquisition is in progress.

4. In pending pause state, make some changes in 'Run type' column for pending pause line (Begin summary) and for next line (End summary) and save the sequence.

5. Resume run queue in processing (pending pause) state.

6. Result set completed successfully but result set is not created as per changes made in sequence.

7. From Report->View sequence report. Report is seen in it but it's not viewable to user.

8. Report is uploaded to respective backend.

Temporary Solution:

Edit only paused sequence

Fix Information:

Ready for development


KPR#:138682  Product:    A.04.07 (Tristan 3C)

Keyword: 

One-line Description:

Unable to set the calibration curve of Log amount & Log response scale

Problem:

Unable to set the calibration curve of Log amount & Log response scale in OpenLAB CDS EZChrom Compact A.04.07.  This only occurs in the Japanese version.

Temporary Solution:

n/a

Fix Information:

Resolved in OpenLAB CDS EZChrom A.04.07 SR2.


KPR#:138683  Product:    A.04.07 (Tristan 3C)

Keyword: 

One-line Description:

Double samples are submitted to headspace when using sequence bracketing when using a headspace with OpenLAB CDS EZChrom

Problem:

Double samples are submitted to headspace when using sequence bracketing when using a headspace with OpenLAB CDS EZChrom.

Temporary Solution:

n/a

Fix Information:

Resolved in OpenLAB CDS EZChrom A.04.07 SR2.


KPR#:138686  Product:    A.04.07 SR1

Keyword: 

One-line Description:

A User Can Edit A Running Sequence Without The Privilege

Problem:

In OpenLAB CDS EZChrom A.04.07 SR1 a user can edit a running sequence after pausing even if they do not have the privlege to do so.  The save sequence menu item is disabled; however, the user is prompted and able to save the sequence after restarting the run.

Temporary Solution:

n/a

Fix Information:

Resolved in OpenLAB CDS EZChrom A.04.07 SR2.


KPR#:138699  Product:    A.04.07 (Tristan 3C)

Keyword: 

One-line Description:

Method Resolution window not "finishing."

Problem:

When the Method Resolution window pops up, the oven tab is the default. A box is not being drawn completely, leaving the bottom portion missing. The window is also much larger than it needs to be.

Temporary Solution:

n/a

Fix Information:

Resolved in OpenLAB CDS EZChrom A.04.07 SR2.


KPR#:138705  Product:    A.04.07 (Tristan 3C)

Keyword: 

One-line Description:

OpenLAB Control Panel crashes after configuring an instrument for the 35900

Problem:

The OpenLAB Control Panel can crash when closing the configuration window after configuring a 35900 instrument with version B.01.01 of the 35900 driver.  This only occurs during the initial configuration.

Temporary Solution:

n/a

Fix Information:

To prevent the control panel from crashing install the Microsoft .NET fix using the link below.

https://download.microsoft.com/download/2/8/7/2870C339-3C77-49CF-8DDF-AD6189AB8597/NDP453-KB2969351-x86-x64-AllOS-ENU.exe


KPR#:138728  Product:    A.04.07 SR2 (Tristan 3D)

Keyword: 

One-line Description:

User not able to uninstall the "Agilent OpenLAB CDS EZChrom Edition-Agilent A/D" driver from Uninstall a Program on a Window 8.1 machine

Problem:

A user is not able to uninstall the "Agilent OpenLAB CDS EZChrom Edition-Agilent A/D" driver from Uninstall a Program on a OpenLAB CDS EZChrom A.04.07 Window 8.1 machine.

Temporary Solution:

n/a

Fix Information:

On the Windows 8 machine disable the following local security policy before uninstalling:  Security Settings->Local Policies->Security Options->User Account Control: Run all administrators in Admin Approval Mode


KPR#:138737  Product:  M8227AA/BA Hitachi LC  HitachiLC_EZC_February18Ed

Keyword: 

One-line Description:

Need to verify that Hitachi registry key is deleted after Hitachi Drivers Are Uninstalled

Problem:

If the Hitachi drivers installed with the OpenLab CDS EZChrom A.04.07 are uninstalled, please verify the registry key below is deleted from the registry after the uninstall.

HKEY_LOCAL_MACHINE\SOFTWARE\Wow6432Node\Hitachi\Analytical

Temporary Solution:

Manually delete the registry key if it persists after the driver uninstallation:
HKEY_LOCAL_MACHINE\SOFTWARE\Wow6432Node\Hitachi\Analytical

Fix Information:

There is no fix planned.


KPR#:138743  Product:    A.04.05 HF07

Keyword: 

One-line Description:

Software crash if Front Exponential Skim integration event is used and a peak is not defined.

Problem:

The Front Exponential Skim integration event can cause the instrument client to crash when you reprocess if the start and stop time range defined does not contain a valid peak.

Temporary Solution:

When using the Front Exponential Skim integration event make sure there is a valid peak in the defined start and stop time range.

Fix Information:

n/a


KPR#:138746  Product:    A.04.07 SR2 (Tristan 3D)

Keyword: 

One-line Description:

OpenLAB CDS EZChrom Antivirus Exclusions

Problem:

It is recommended that the following paths be excluded when using an antivirus program with OpenLAB CDS EZChrom:
The path that you use to store your data
C:\programdata\chromatography system\recovery data
C:\programdata\agilent
%programfiles%\agilent
%programfiles%\common files\agilent
%programfiles(x86)%\common files\agilent
%programfiles%\agilent technologies
%programfiles(x86)%\agilent technologies
%programfiles%\common files\agilent technologies
%programfiles(x86)%\common files\agilent technologies
%programfiles%\common files\agilent shared
%programfiles(x86)%\common files\agilent shared

Temporary Solution:

n/a

Fix Information:

Antivirus Exclusions are documented in the installation manuals of A.04.08


KPR#:138780  Product:    A.04.07 (Tristan 3C)

Keyword: 

One-line Description:

Manual integration fixes while data is being acquired get lost after data acquisition is complete.

Problem:

The manual integration fixes table is enabled during data acquisition which allows customers to make changes to this table while data is being acquired.  When changes are made to this table while data is being acquired the changes are lost after data acquisition is complete.

Temporary Solution:

If manual integration changes should be saved, they cannot be edited during running acquisition. Make the changes afterwards. 

Fix Information:

n/a


KPR#:138812  Product:    A.04.07 (Tristan 3C)

Keyword: 

One-line Description:

Missing Advanced Run Information in Method Report

Problem:

The LC Driver A.02.13 introduced the implementation of IRCControlAnalyticalResults to add advanced run info to the acquired data, such as temperature at start of run etc. In ChemStation these advanced run info are part of the ACAML and can therefore be reported via IR. Running a sample in EZChrom shows that no DiagnosticData are written to the ACAML result.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:138825  Product:    A.04.08

Keyword: 

One-line Description:

error on project creation in mixed mode when Project Group has a name with leading space

Problem:

When creating a project under a project group with leading space in mixed, an error message can show up: "An error prevented project changes from being saved". The project has not been created.

Temporary Solution:

There should not be leading spaces in front of Project Group Names

Fix Information:

A fix is planned for a later revision.


KPR#:138828  Product:    A.04.07 SR2 (Tristan 3D)

Keyword: 

One-line Description:

Unable to integrate old data after Upgrade from A.04.07 sr1 to A.04.07 sr2

Problem:

Unable to integrate old data after Upgrade from A.04.07 sr1 to A.04.07 sr2. LC drivers were also upgraded to latest version.

Temporary Solution:

Create a new instrument for reprocessing using the old data.

Fix Information:

n/a


KPR#:138830  Product:    A.04.07 SR2 (Tristan 3D)

Keyword: 

One-line Description:

After reprocessing, the result set name will be reported twice in IR instead of the result set and sequence template name

Problem:

Before reprocessing the sequence name field, will report the result set name as well as the sequence template name. 
After reprocessing the result set name will be reported twice and the sequence template name no more.

Temporary Solution:

n/a

Fix Information:

A fix is planned for a later revision.


KPR#:138860  Product:    A.04.07 SR2 HF03

Keyword: 

One-line Description:

Data lost by using Chinese special characters with ECM backend

Problem:

With A.04.07 SR2 HF05 there is a new separate location in the Enterprise Path with the name "RSTBackup" which is used with ECM backend when a file is unzippable. Files which can not be zipped and uploaded to ECM will be backed up to this separate location in the Enterprise Path: (\\EnterprisePath\RSTBackup\AICName\SSIZIP_Date-Time)
However since ECM does not support localization, when customers are using ECM, their OS needs to be an English OS with an English keyboard.

Temporary Solution:

n/a

Fix Information:

Fixed in A.04.07 SR2 HF05 and in A.04.08


KPR#:138870  Product:    A.04.07 SR2 (Tristan 3D)

Keyword: 

One-line Description:

Able to print sequence without saving

Problem:

Users are able to print a sequence without saving the sequence.

Temporary Solution:

n/a

Fix Information:

A fix is planned for a later revision.


KPR#:138874  Product:    A.04.07 SR2 HF01

Keyword: 

One-line Description:

Printouts for injection reports may not print in their injection order

Problem:

Reprocessing a sequence, each injection report creates its own pdf file and sometimes those pdf files are not sent to printer in the proper order (inj1, in2, inj3...). They get printed in the order they get processed by the printer server.

Analysis of the cause:
The difference in the order of sample in the sequence and the order of the physical printout is possibly caused by multiple factors, but mainly by various data complexity and network latency. These lead to a result that data reprocessing and report generation for later sample in the sequence may be completed faster than earlier samples. Or in a distributed setup, the report of later sample may reach the printer faster than earlier samples. 

 

Temporary Solution:

Recommendations to avoid printout out-of-order
Use the relevant best practice based on actual customer workflows:
1.	If the customer perform batch reprocess with automatic report generation and printing. The best practice is to define the printout in the method, so that the samples are reprocessed and report print out as soon as the data acquisition is completed. There is no racing of multiple reports reach to the printer, due to the gap introduced between reports (i.e. data acquisition time).
2.	If the customer reviews the result post acquisition, perform manual integration before print out. (It is very unlikely that the print out-of-order would happen in this scenario, but if it does), customer can use reprocess with review option (check the pause after each run). 

Fix Information:

DNF


KPR#:138876  Product:    A.04.08

Keyword: 

One-line Description:

IQT/SVT Fix for RC IF Deployment for OpenLAB EZChrom A.04.08

Problem:

Software verification report for Agilent Rapid Control .NET [3.0] [5] is FAIL. IQT/SVT Fix for RC IF Deployment for OpenLAB EZChrom A.04.08

Temporary Solution:

n/a

Fix Information:

Resolved in B.01.02.059


KPR#:138897  Product:    A.04.07 (Tristan 3C)

Keyword: 

One-line Description:

CDS EZChrom with ECM doesnot display all the subfolders when browsing for Project Path during project creation

Problem:

Problem - When the when creating projects and specifying project path, If you have more than 100 Sub folders, then you would not be able to set the project path correctly because ECM window displays only the first 100 folders.

Temporary Solution:

n/a

Fix Information:

Fixed in OLSS 2.1 - Hotfix 4 which does not support ECM 3.4.1


KPR#:139690  Product:    A.04.07 SR2 HF06

Keyword: 

One-line Description:

CDS is getting crashed as we trace large size data file (184 MB)

Problem:

The issue is to overlay traces (chromatograms) from very large data file (>184MB) in viewer. 
1.Create a sequence using large size data file.
2.Process the sequence twice to create two result set.
3.Open any one result set and perform Add Trace -> Other data file.
4.Choose data file from another result set created above.

Temporary Solution:

n/a

Fix Information:

Do not fix. Low risk and corner case.


KPR#:147026  Product:Shimadzu GC  M8232AA/BA Shimadzu GC  ShimadzuGC1.1

Keyword: 

One-line Description:

The online signal is missing after an acquisition run has started.

Problem:

The signal is present until the start signal of the injection. After the acquisition was started, the online signal is no longer displayed. This behavior was observed with a GC-2010 with FID-detector but can also happen with a GC-2014. The run will not be interrupted and the data is  completely collected and stored in a data file.

Temporary Solution:

The "Sampling rate" in the method must be in "Cycle" mode. The value of the "Sampling rate" must be the same as the "Base period" of the instrument configuration window of the detector.

Fix Information:

Fixed in Shimadzu GC v2.00


KPR#:149833  Product:LC and CE Drivers  RC 1200.NET  A.02.02 SP1

Keyword: 

One-line Description:

Flexible Cube G4227A does not work with HiP ALS G1367E

Problem:

On Chinese or Japanese CDS only the 1290 Infinity Flexible Cube G4227A does not work with High Performance Autosampler G1367E. 

On English CDS the operation of 1290 Infinity Flexible Cube G4227A with High Performance Autosampler G1367E is not affected.

Temporary Solution:

n/a

Fix Information:

Resolved in revision A.02.02


KPR#:149839  Product:LC and CE Drivers  RC 1200.NET  A.02.03

Keyword: 

One-line Description:

Lamps and cells of G1314F show incorrect dates in the RFID tags

Problem:

Lamps and cells of G1314F show incorrect dates in the RFID tags

Temporary Solution:

n/a

Fix Information:

Fixed with the driver version A.02.06


KPR#:149846  Product:LC and CE Drivers  RC 1200.NET  A.02.07

Keyword: 

One-line Description:

AFC & UIB II: AFC does not fractioning as expected

Problem:

System configuration:
1 DAD
1 UIB II
2 AFCs -> 1 AFC Cluster
Analog out of DAD connected to Analog In of UIB II
Simulated signal of DAD was used

Method settings for modules (expect threshold for UIB II signal): see attached method

1. Created a data file with the simulated signal recorded by the UIB II
2. Loaded data file into Fraction Preview
3. Used Fraction Preview to collect peakbased fractions with threshold > 0.12 V
4. The value 0.12 V is written into the table on the left side without conversion (the table shows mV as unit)

Using this method will cause that the AFC will fractioning all the time.

Cause:
The firmware interface expects the threshold in mV but the driver writes 0.12 (V) to the AFC
This causes that the AFC is fractioning all the time.

Temporary Solution:

n/a

Fix Information:

Fixed in UIB II signal is delivered as mV signal similar to  UIB I since LC Driver Release A.02.07.


KPR#:149848  Product:LC and CE Drivers  RC 1200.NET  A.02.07

Keyword: 

One-line Description:

Use Vial as Fraction collections

Problem:

Started a run with fraction collection.

The fraction collector config does not show 10 vials but only trays.

Temporary Solution:

If you have installed a 2 WP-Tray, vials are allways forbidden to use for fractionation, regardless if there are wellplates inserted or not.

Fix Information:

Issue fixed with LC and CE Driver revision A.02.07


KPR#:149849  Product:LC and CE Drivers  RC 1200.NET  A.02.07

Keyword: 

One-line Description:

Recovery volume and fraction volume: behaviour for reset volume

Problem:

I worked with one AFC.
Configured 2 wps +10 vials and 15 vial tray for recovery.

Started multiple runs. 
For each run a new Recovery location was used.
Then removed the recovery tray and entered it again.

Next run started with first recovery location --> this was already used.

Temporary Solution:

If you want to reset the Funnel Trays only, use "Reset Funnel Tray Volume"

If you want to reset Wellplates and vials (all except funnels), use "Reset Fraction Volume"

Fix Information:

Issue fixed with the LC and CE driver revision A.02.07


KPR#:149869  Product:LC and CE Drivers  RC 1200.NET  A.02.07

Keyword: 

One-line Description:

Fraction start location in a sequence parameter cannot accept the number "1" when using 2P+10Funnel tray at RC.net driver.

Problem:

The customers want to use 10 Funnels at the tray 2P+10Funnel without plates.

Temporary Solution:

Solved with LC & CE Drivers A.02.09

Fix Information:

N/A


KPR#:163394  Product:    A.04.07 SR2 HF05

Keyword: 

One-line Description:

After applying HF6, Customer workflow cannot be used any more due customer needs 2 user, samelevel, esign each datafile

Problem:

When a user with same level role tries to apply signature, user get following error:
“Failed to apply electronic signature. Data file is already signed by <user> (<User level>).”

Temporary Solution:

n/a

Fix Information:

OpenLAB CDS EZChrom Edition A.04.07 SR2 Update8 Point Patch 1
Fix is implemented around the e-signature functionality to resolve the issue wherein the multiple users having same signature levels are restricted from signing already signed data i.e. users are restricted from doing peer reviews. 


KPR#:177870  Product:    A.04.08

Keyword: 

One-line Description:

Windows 10 feature updates can corrupt existing EZChrom A.04.08 installation 

Problem:

When Windows 10 feature updates are performed on systems with existing OpenLab CDS EZChrom A.04.08 installations, the EZChrom installation can become corrupt. In some cases repair was successful, but with the later feature updates of Windows 10, like this to version 1709 or 1803, it happened frequently that EZChrom installations did break and could not be repaired. In such cases uninstall, reapply the necessary system settings and reinstall EZChrom is supposed to solve the problem. 

Temporary Solution:

Updates can be deferred for a different time depending on the Windows 10 servicing channel assigned. E.g.: Windows 10 Enterprise Long-Term Servicing Branch edition (LTSB)
For domain environments Windows 10 feature updates can be controlled with different Servicing Tools. E.g. the optional server role: Windows Server Update Services (WSUS)

Fix Information:

The identified defect will be fixed in OpenLab CDS EZChrom A.04.09. 
This should improve stability for future Windows feature updates 


KPR#:235060  Product:    A.04.07 SR2 HF01

Keyword: 

One-line Description:

Printouts for injection reports may not print in their injection order

Problem:

Reprocessing a sequence, each injection report creates its own pdf file and sometimes those pdf files are not sent to printer in the proper order (inj1, in2, inj3...). They get printed in the order they get processed by the printer server.

Analysis of the cause:
The difference in the order of sample in the sequence and the order of the physical printout is possibly caused by multiple factors, but mainly by various data complexity and network latency. These lead to a result that data reprocessing and report generation for later sample in the sequence may be completed faster than earlier samples. Or in a distributed setup, the report of later sample may reach the printer faster than earlier samples. 

 

Temporary Solution:

Recommendations to avoid printout out-of-order
Use the relevant best practice based on actual customer workflows:
1.	If the customer perform batch reprocess with automatic report generation and printing. The best practice is to define the printout in the method, so that the samples are reprocessed and report print out as soon as the data acquisition is completed. There is no racing of multiple reports reach to the printer, due to the gap introduced between reports (i.e. data acquisition time).
2.	If the customer reviews the result post acquisition, perform manual integration before print out. (It is very unlikely that the print out-of-order would happen in this scenario, but if it does), customer can use reprocess with review option (check the pause after each run). 

Fix Information:

DNF


KPR#:257114  Product:  M8244BA Thermo Fisher SII  1.0.0

Keyword: 

One-line Description:

Monitor Baseline disabled

Problem:

The Monitor Baseline feature accessible via the Instrument View is disabled.

Temporary Solution:

Create and execute a blank run to monitor the baseline

Fix Information:

n/a


KPR#:257121  Product:  M8244BA Thermo Fisher SII  1.0.0

Keyword: 

One-line Description:

No 3D data access

Problem:

3D field data, e.g., DAD or FLD spectral data is not accessible. Selecting 3D data acquisition in the instrument method has no effect on the recorded data.
Online spectra are not available and the result files do not contain 3D data.

Temporary Solution:

n/a

Fix Information:

This was implemented in SII 1.1


KPR#:278031  Product:    A.04.09 Update 03

Keyword: 

One-line Description:

Unable to retrieve method from ECM

Problem:

A.04.09 with ECM 3.6 and previous versions

Steps to reproduce:
 1. Launch and instrument online from a long project (Project name with 100 characters or more).
 2. Create a method and save it with long name. Method name with 75 characters or more).
 3. Open Blank New method from File > New > Method.
 4. Now try to open the method created at step 2.

Actual Result
• Error message “Unable to retrieve the file from content management” and method is not open.

ECM URL: If this URL encoded path length exceeds 245 characters, the file will not be uploaded. Read entire section, General Constraints on File Path Lengths, in ECM 3.5 ECM Administrator's Guide.pdf, for special character interpretation. 

UNC Path: Enterprise path length is limited to 245 characters by OpenLAB CDS EZChrom. If the enterprise path length exceeds 245 characters, the file will not download.
 

Temporary Solution:

Agilent recommends using as short as possible file names and/or comply with the Windows OS path limitations.

Fix Information:

n/a


KPR#:355640  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.0

Keyword: 

One-line Description:

UV/Vis Chromaster Ultra RS detector displays dual channel mode only during the run

Problem:

If a dual acquisition mode for the UV/Vis detector is selected in the method, the instrument status window for the detector shows only a single channel A until the run is started. On starting the run, both channel A and B are properly displayed in the Instrument Status window.

Temporary Solution:

n/a

Fix Information:

This will not be fixed due to technical difficulties of implementation. This is documented as a driver limitation in the User's Guide.


KPR#:368253  Product:    EZChrom Edition A.04.09

Keyword: 

One-line Description:

G3581A 490 Micro GC Windows 10 Compatibility

Problem:

As of A.04.08 and A.04.09, Windows 10 supported, the CDS_SupportedInstFirmware.pdf does not reflect compatibility for G3581A 490 Micro GC with Windows 10 64 bit. The Windows 10 reference is missing.

Temporary Solution:

n/a

Fix Information:

G3581A 490 Micro GC compatibility with Windows 10 64 bit tests passed for A.04.09. The 490 Micro GC PSP does also contain compatibility. Please verify documented supported firmware and drivers.


KPR#:368421  Product:  M8244BA Thermo Fisher SII  1.0.0

Keyword: 

One-line Description:

All Signals Are Named The Same in DA: By  The Instrument Name

Problem:

All detector signal traces are named SII1 in the Data Analysis. The signal names do not distinguish detectors.

Temporary Solution:

n/a

Fix Information:

This has been fixed in SII 1.1.
However, the fix breaks backwards compatibility in Data Analysis. See ReadMe document in the SubscribeNet package for SII 1.1 for details.


KPR#:371236  Product:  M8244BA Thermo Fisher SII  1.0.0

Keyword: 

One-line Description:

Take/Release Control Feature Does Not Work

Problem:

The Take/Release feature to gain exclusive access over an instrument does not work in this driver version. Clicking the Take/Release button does not have an effect. Also, direct control is always possible independently of the user permissions.

Temporary Solution:

There is no work around.

Fix Information:

Fixed in SII 1.1 with limitations: 
The direct control is now only possible for the instrument that is not taken via the "Instrument View" Instrument Status context menu. All other context menu options function as expected.


KPR#:375500  Product:  M8244BA Thermo Fisher SII  1.0.0

Keyword: 

One-line Description:

Entries in method audit trail are ambiguous

Problem:

In Method Audit Trail, the corresponding retention time is missing if changes are made in a table of a method. For example, in the gradient table, Method Audit Trail does not reflect to which row the changes refer.

Temporary Solution:

n/a

Fix Information:

This has been fixed in SII 1.1.


KPR#:462549  Product:  M8244BA Thermo Fisher SII  1.0.0

Keyword: 

One-line Description:

Missing scrollbar in Method Setup window

Problem:

Thermo Fisher SII v1.00 driver, Thermo Trace GC, OpenLab CDS 2.5
Detector tab in Method Setup window is missing a scrollbar and it prevents the user from accessing the bottom part of the screen. 

Temporary Solution:

n/a

Fix Information:

This has been fixed in SII 1.1.


KPR#:476099  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.1

Keyword: 

One-line Description:

SII: No 3D Signals Available for FLD

Problem:

No 3D signals are provided for FLD detector in Data Acquisition even though the method allows to select 3D data collection.

Temporary Solution:

None.

Fix Information:

To be addressed in the future revisions of SII driver.


KPR#:494402  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.0.0,1.1

Keyword: 

One-line Description:

SII: Inverse Gradient Feature Not Supported For Vanquish Instruments

Problem:

The Inverse Gradient feature is currently not available for Vanquish Instruments. However, it is supported by Chromeleon CDS. Online Help of OpenLab CDS mentions this feature as supported.

Temporary Solution:

None

Fix Information:

This is fixed in Thermo Fisher SII Control 1.4.


KPR#:497735  Product:  M8244BA Thermo Fisher SII  1.0.0.123396,1.1

Keyword: 

One-line Description:

SII: Instrument Activity Log Mixes Entries of Several Instruments

Problem:

Launching and operating two SII instruments at the same time can lead to an Instrument Activity Log which also contains some entries from the wrong instrument.

Temporary Solution:

None

Fix Information:

The activity log entries created by the utility drivers RS232 Input Port and Remote Inject Driver now appear in the appropriate Activity Log. However,
some Activity Log entries created during instrument configuration cannot be addressed and hence left unfixed.


KPR#:516800  Product:PerkinElmer LC  M8229BA PerkinElmer LC  PE_LC300_2.0

Keyword: 

One-line Description:

Perkin Elmer LC300: "No injection/instrument Blank" Injection Source Not Supported

Problem:

Injection source "No injection/instrument Blank" is not supported with Perkin Elmer LC 300 system.

Temporary Solution:

A possible workaround to perform an instrument blank run is to run the sample with the LC 300 Autosampler as the injection source but use 0 for the injection volume.

Fix Information:

To be considered in the future revisions of the driver.


KPR#:519615  Product:PerkinElmer LC  M8229BA PerkinElmer LC  PE_LC300_2.0

Keyword: 

One-line Description:

PE LC300: Uploading Method From Instrument Does Not Retrieve Pump Time Program

Problem:

When a method is uploaded from the instrument, only the initial conditions of the LC 300 pump are retrieved, not the entire pump time program (if any).

Temporary Solution:

Manually enter the desired pump program after the initial method creation from current instrument settings.

Fix Information:

To be considered for the future revisions of the driver.


KPR#:643040  Product:    uGC Driver - 2.2.7

Keyword: 

One-line Description:

Channel information is incorrect from Web Browser

Problem:

Check Channel information from web browser, the wording BF2D should be 'backflush to detector', not 'backflush to vent installed'.

Temporary Solution:

n/a

Fix Information:

Fixed in 2.3.6


KPR#:647335  Product:    uGC Driver - 2.3.3

Keyword: 

One-line Description:

Status always display "injecting" even instrument is idle

Problem:

Configure 990 Micro GC, create a method, then single run. But run can not stop even running is completed though Web Browser.

Temporary Solution:

n/a

Fix Information:

Fixed in 2.3.6


KPR#:725196  Product:    uGC Driver - 2.2.7

Keyword: 

One-line Description:

Software OpenLab CDS stays online even 990 Micro GC shut off already

Problem:

Software OpenLab CDS stays online even 990 Micro GC shut off already

Temporary Solution:

n/a

Fix Information:

2.6.8


KPR#:754535  Product:    uGC Driver - 2.4.2

Keyword: 

One-line Description:

 Help initial page showing folder content

Problem:

 Help initial page showing folder content

Temporary Solution:

n/a

Fix Information:

fixed and released in 2.5.5


KPR#:758156  Product:    uGC Driver - 2.4.3

Keyword: 

One-line Description:

Online Help initial page show folder content on Data System Configuration 

Problem:

Online Help initial page show folder content on Data System Configuration 

Temporary Solution:

n/a

Fix Information:

fixed in driver 2.5.5


KPR#:759268  Product:    uGC Driver - 2.4.3

Keyword: 

One-line Description:

The Help file of Driver in Configure Instrument need to be updated 

Problem:

The Help file of Driver in Configure Instrument need to be updated 

Temporary Solution:

n/a

Fix Information:

fixed in driver 2.5.5


KPR#:807723  Product:    uGC Driver - 2.4.8

Keyword: 

One-line Description:

The abnormal color for Gasifier status

Problem:

The abnormal color for Gasifier status

When switch off the transfer line heater and the Gasifier vaporizer heater, the color for Gasifier status 'OFF' is yellow. 

Temporary Solution:

n/a

Fix Information:

fixed in driver 2.5.5


KPR#:977013  Product:    uGC Driver - 2.4.8

Keyword: 

One-line Description:

The text of 'VICI stream position' need optimize.

Problem:

The text of 'VICI stream position' need optimize.

Temporary Solution:

n/a

Fix Information:

2.6.8


KPR#:1054977  Product:    uGC Driver - 2.4.8

Keyword: 

One-line Description:

No error happen if load method has invalid initial pressure

Problem:

Load a method include invalid initial pressure, try to download or run, no error message.

Temporary Solution:

n/a

Fix Information:

fixed in driver 2.5.5


KPR#:1074050  Product:OpenLAB - CDS2    CDS FR8

Keyword: 

One-line Description:

Hitachi ChromasterUltra RS: Selection Column by Column on Tray Display Not Supported

Problem:

For Hitachi CMU RS instrument, only row by row vial selection can be done in the Tray Display. Column by column vial selection pattern is not available.

Temporary Solution:

n/a

Fix Information:

To be addressed in the newer revision of the driver.


KPR#:1193684  Product:OpenLAB - CDS2    CDS 2.8

Keyword: 

One-line Description:

Hitachi ChromasterUltra RS: Selection in Tray A Display Also Reflects in Tray B of Microplates

Problem:

For Hitachi CMU RS instrument, when multiple racks are set, the vial selection made in the Tray display for the first rack is also reflected in the second rack. This does not affect data acquisition.

Temporary Solution:

N/A

Fix Information:

To be addressed in OpenLab CDS.


KPR#:1216156  Product:CTC PAL3 LC  M8223BA CTC PAL3 for LC  CTC_LC_PAL3_1.9.6_OpenLabCDS

Keyword: 

One-line Description:

CTC PAL 3: Acquisition Crashes Upon Vial Fill Down in Sequence Table

Problem:

In OpenLab Acquisition, in the Sequence editor, when selecting the option "Fill down - increment" for the column "Vial", an error message appears.

Once the error message is closed, the OpenLab Acquisition window closes. The instrument remains connected to the AIC in case of a Client/Server.

Temporary Solution:

Do not use Vila fill down function until the fix is issued.

Fix Information:

In investigation by the driver team.


KPR#:1387753  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.4

Keyword: 

One-line Description:

Thermo SII: Connection to the Instrument Failed in Method Editor

Problem:

SII 1.3: When creating a new method with a Thermo Ultimate instrument (possibly with other instruments, too), the system shows the error:
“Connection to the instrument "<Instrument_name>" on <PC_Name> failed”. 
Clicking "Retry" on the error message leads to the same result. Upon clicking "Cancel", the sampler method window is greyed out; starting a run typically fails or runs inconsistently.
The issue prevents normal method creation.

Temporary Solution:

None.

Fix Information:

The issue is being investigated by Thermo Fisher.


KPR#:213248  Product:    A.04.09

Keyword: 35900 Control

One-line Description:

Update to A.04.09 can break 35900E driver installation with CTC

Problem:

After Upgrade from Agilent OpenLAB CDS EZChrom Edition A.04.07 SR2 +CTC drivers  to A.04.09 , it happend that a 35900E can no more be configured as Generic instrument

Temporary Solution:

After Upgrade from Agilent OpenLAB CDS EZChrom Edition A.04.07 SR2 +CTC drivers  to A.04.09 , Repair  of Agilent A/D driver is required to configure a Generic instrument.

Fix Information:

n/a


KPR#:307590  Product:    EZChrom Edition A.04.09

Keyword: 35900 Control

One-line Description:

Preview run not working in A.04.09 & new RC.Net Driver functions differ

Problem:

The preview run feature is no longer available with the 35900E RC.net drivers.

There is a difference in 35900E behavior between OpenLab CDS EZChrom Edition A.04.08 and A.04.09, due to the native 35900E driver. The functional differences are between the native 35900E and new RC.Net 35900E drivers. 

A. OpenLAB CDS A.04.08 released build. (Agilent A/D :4.7.1.0 - Native) 

External: Run started immediately, but stops on “Pre-Run: Waiting for start” Manually hit start button on 35900 and data acquisition starts and the run completes. 

Manual: Run starts, “Start Data Acquisition” dialog displays. Use the PC mouse to press “Ok”. The acquisition starts and completes. 

None: Run starts immediately and completes without any error. 

B. OpenLAB CDS EZChrom Edition A.04.09 released build. (Agilent 35900 A/D Drivers: 2.3.53 - RC.Net)

External: Run starts immediately and complets without any error.

Manual: Run starts, “Start Data Acquisition” dialog displays. Use the PC mouse to press “Ok”. Acquisition stops on “Pre-Run: Waiting for start”.  Manually press the start button on 35900E. Data acquisition starts and the run completes.

None: Run starts immediately, but stops on “Pre-Run: Waiting for start”.
Manually press the start button on 35900. Data acquisition starts and completes.

Summary:
A.04.08: The run starts without pressing the start button. For 'External' trigger, the behavior is vice versa.

A.04.09: Using the RC.Net drivers requires pressing the start button for trigger 'None' and 'Manual'.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:381770  Product:    A.04.09 Update1

Keyword: 35900 Control

One-line Description:

Install manual(M8201-90038) references installing Classic Drivers for 35900 and is not correct

Problem:

35900 is supported with the supplied drivers and does not require installing classic drivers.  Installing classic can lead to unpredictable issues

Temporary Solution:

N/A

Fix Information:

Do not install classic drivers. Firmware and Instrument guide updated.


KPR#:138882  Product:    A.04.05 HF08

Keyword: 3rd Party Drivers

One-line Description:

The running sequence can not be updated

Problem:

With the EZChrom Edition A.04.05 Hitachi driver for AAA L-8800/L-8900 the sequence lines cannot be edited like with other instruments. 

Temporary Solution:

n/a

Fix Information:

With the EZChrom Edition A.04.05 Hitachi driver for AAA L-8800/L-8900 the sequence lines cannot be edited as with other instruments. The sequence can be paused and resumed but the driver does not fully support editing. 


KPR#:389878  Product:    A.04.08,A.04.09

Keyword: 3rd Party Drivers

One-line Description:

Hitachi LCE and CM: Cannot Use Submit Button in Run Queue to Submit Runs

Problem:

Hitachi (LaChromElite and Chromaster) LC does not start a single run from a Run Queue window via "Submit" button in OpenLab EZChrom edition.

Temporary Solution:

Submitting via "Advanced" option in the Run queue works. 

Fix Information:

Under Investigation


KPR#:394168  Product:    A.04.08,A.04.09

Keyword: 3rd Party Drivers

One-line Description:

Hitachi LCE and CM: Cannot Use Submit Button in Run Queue to Submit Runs

Problem:

Hitachi (LaChromElite and Chromaster) LC does not start a single run from a Run Queue window via "Submit" button in OpenLab EZChrom edition.

Temporary Solution:

Submitting via "Advanced" option in the Run queue works. 

Fix Information:

Under Investigation


KPR#:16  Product:68xx Driver Software  6890  5.03

Keyword: 68xx Driver

One-line Description:

Vial Missing Error Message

Problem:

When the 6890 GC encounters a missing vial, the IALog displays Stop key pressed error #1 instead of "Missing Vial."

Temporary Solution:

None

Fix Information:

This is fixed in 68XX Driver version 6.21 which is included in C.01.05 OpenLAB ChemStation and A.04.05 OpenLAB EZChrom.


KPR#:17  Product:68xx Driver Software  6890  1.01

Keyword: 68xx Driver

One-line Description:

How to Change Column Dimensions

Problem:

Changing Column Dimensions requires being logged in as an administrator.

Temporary Solution:

Column dimensions are part of the GC Configuration and the end user must be logged on as an Administrator to change these parameters.

Fix Information:

This is fixed in GC Driver version 6.01/6.02 which is included in OpenLAB ChemStation C.01.01 and OpenLAB EZChrom A.04.01 and later.


KPR#:18  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

If One Signal Needs to be Saved, it Must be Assigned as Signal 1

Problem:

If only one signal needs to be saved, you must assign it as signal 1.

Temporary Solution:

If only one signal needs to be saved, you must assign it as signal 1.

Fix Information:

Fixed in 68xx driver revision 6.11[008].  This driver is included in B.04.03 SP1 ChemStation and C.01.03 OpenLAB CDS.


KPR#:22  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

In 68XX driver revision 5.02 - 6.11, it is not possible to Modify a Method During a Run to Extend the Run Time or Change the Oven Temp

Problem:

It is possible to the software edit parameters screen and change the oven hold time or oven temp and select OK.  The GC does not accept the updates.  The method in the software is changed to the new values; the GC ignores all.

Temporary Solution:

n/a

Fix Information:

68XX Driver 6.11 is included in B.04.03 SP1, OpenLAB ChemStation C.01.04 and OpenLAB EZChrom A.04.04.


KPR#:23  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

Info in Edit instrument parameters\Columns is incomplete

Problem:

Edit instrument parameters; columns

Applies to GC’s: 7890A, 6890N, 6850A

The window displayed is incomplete; it shows the column type and flow/pressure parameters, but does not give the length, ID or film thickness.  These are critical parameters, and should be displayed just as they were in the previous ChemStation revisions.  To find these parameters, the user now has to look on the configuration window.

Temporary Solution:

n/a

Fix Information:

Fixed in 68xx driver revision 6.11[008].  This driver is included with OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.


KPR#:31  Product:68xx Driver Software  6890  5.11

Keyword: 68xx Driver

One-line Description:

Ignore missing vial option is not present in GC

Problem:

Run a sequence with 11 runs and the 11th vial is missing.  The sequence aborts and SSIZip file was not uploaded to ECM as the 11th vial was missing.

The ignore missing vial option is not present in GC.

Temporary Solution:

n/a

Fix Information:

This is fixed in GC Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:33  Product:68xx Driver Software  6890  5.10

Keyword: 68xx Driver

One-line Description:

Instrument Setup graph has incorrect column legend

Problem:

In the 6890, the legend for the Instrument Setup graph says Column 2, irrespective of which column is highlighted.  The graph is correct, it is just the legend that is wrong.

Temporary Solution:

n/a

Fix Information:

This has been fixed in 68XX Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:38  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

ALS errors abort sequence; no Retry/Abort dialog box option offered with 68xx RC.NET Driver

Problem:

6890A RC.NET driver with ALS. 

If an ALS error occurs such as a Vial not found event, there is no dialog box option presented to Retry/Abort.  The sequence simply aborts.

Temporary Solution:

n/a

Fix Information:

This is fixed in GC Driver version 6.21, which is available with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:59  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

In 68XX Driver Version 5.02, it is not Possible to Start the 6890 RC.NET from the Front Panel Start Key

Problem:

In 68XX Driver Version 5.02, it is not possible to start the 6890 RC.NET from the front panel Start key.  Software configuration is Injection source = manual.  6890 is READY, press start and get error message "ERROR Start Key disabled".

Temporary Solution:

The workaround is for the user to start the run from the "run control" then "run Method" in the Chemstation software and then press start at the GC.

Fix Information:

Fixed in driver revision 6.10.  This driver is included with B.04.03 DSP1 ChemStation.  It is also included with OpenLAB ChemStation C.01.02 and OpenLAB EZChrom A.04.02.


KPR#:66  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

CIS4 inlet is not recognized with 6890 GC RC.NET

Problem:

CIS4 inlet is not recognized with 6890 GC RC.NET

Temporary Solution:

n/a

Fix Information:

This is fixed in GC Driver version 6.21, which is available with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:83  Product:68xx Driver Software  6890  6.01

Keyword: 68xx Driver

One-line Description:

Only Fast Plunger Speed is available for PTV Solvent Vent multiple injections

Problem:

The fast plunger speed is only available when using multiple PTV Solvent Vent injections.  Most users of multiple PTV Solvent Vent injections need to use Slow or Variable plunger speed, not Fast.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 6.10.  This driver is included with B.04.03 DSP1 ChemStation and OpenLAB C.01.02 ChemStation and OpenLAB A.04.02 EZChrom.


KPR#:89  Product:68xx Driver Software  6850  5.10

Keyword: 68xx Driver

One-line Description:

Incorrect injection volume shown in method report if PTV solvent vent is used

Problem:

6850 with a PTV inlet

The method specifies PTV solvent vent mode with 3 injections, ten seconds apart.  The method volume is 5 µL.  Specify the method to use method volume for three injections.  Each injection is 5 µL.  However the report shows the injection volume as 1 µL.  This is incorrect.  The report should show the injection volume as either 5 µL or 5 µL x 3, or 15 µL.

Temporary Solution:

n/a

Fix Information:

Fixed in software driver 6.10 [023].  This fix is included in B.04.03 DSP1 and later versions of ChemStation.  It is included in OpenLAB ChemStation C.01.02 and OpenLAB EZChrom A.04.02 and later.


KPR#:94  Product:68xx Driver Software  6890  6.01

Keyword: 68xx Driver

One-line Description:

68xx Method Download Times Out

Problem:

Connect to 6890N GC.

Within seconds of successfully opening an online session, logbook logs the following simultaneously with the loading of the method.

'Logbook      Welcome to ChemStation                         12:17:12 06/28/10
   0 1da6 4c28cb1f    0
6890       1 Timed out waiting for method download to fin>  12:17:35 06/28/10
   0 1da6 4c28cb1f    0
6890       1 ish.                                           12:17:35 06/28/10
7da 41e0 4c28cb1f    0
Method       Loading Method LCV_GCCONTROL.M                 12:17:35 06/28/10'

Temporary Solution:

n/a

Fix Information:

Fixed in GC Driver revision 6.10 [027].  This driver version is included in B.04.03 DSP1 ChemStation and C.01.02 OpenLAB ChemStation and A.04.02 OpenLAB EZChrom.


KPR#:95  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

"Method modified after loading" occurs when running a sequence (6890 RC.NET)

Problem:

The "Method modified after loading" flag occurs after lauching the sequence.

Started the sequence from the ChemStation software.  The method modification status indicates that the method has changed when it was not changed and the report will indicate 'modified after loading'.

The GC had a calibrated capillary column that was changed to a packed column.  When using this method that now has a packed column, the GC reports an error in the length parameter, and that is causing the method download not to reset the 'modified after loading' flag.

Temporary Solution:

There is a workaround: bring Method into MethodEdit UI, and "reconfigure" column (Packed to Packed would be OK) to get new structure.

Fix Information:

Fixed in GC Driver revision 6.10 [027].  This driver is included in B.04.03 DSP1 ChemStation and OpenLAB C.01.02 ChemStation and OpenLAB A.04.02 EZChrom.

Note, this fix will only apply to new methods, not to current methods.


KPR#:99  Product:68xx Driver Software  6890  1.01

Keyword: 68xx Driver

One-line Description:

Sequence continues while GC temperature is not ready

Problem:

When running a sequence where consecutive runs use the same method using the ALS on a data system containing 6890 driver version 6.10.027 or earlier, the data system may proceed to start a new run before the GC has equilibrated after the conclusion of the previous run.  This is more likely if the method contains a post run period of sufficient length that the GC equilibrates to the post-run conditions and the data system has completed data analysis well before the end of the post-run period.

This could result in a (retention time) RT shift.

Temporary Solution:

n/a

Fix Information:

Patch available for RC.NET GC driver revision 6.10 [027] for B.04.03 DSP1 [54].

Fixed beginning in: B.04.03 SP1 [87] and OpenLAB C.01.03 [37].


KPR#:100  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

A Method Can be Created Which Will Not Download to the Instrument and the Sequence Does Not Run

Problem:

Method download fails, sequence aborts after start; no injection made

Error reported in logbook:
Attempt to download method to GC failed.

The Data System will not allow the run to continue when the method download has failed.

Temporary Solution:

n/a

Fix Information:

This fix is in the 68xx driver version 6.10.032 and 6.11.001.  

This fix is included in B.04.03 SP1 and OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.


KPR#:101  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

GC Back Injector Not Present in Edit Method Screen

Problem:

After successfully running a sequence with a back method, another back method is loaded. The back injector is not present in the edit method screen.

Temporary Solution:

Change a parameter at the GC keyboard and click "upload from instrument" in the edit method screen. The back injector will show in the edit method screen.

Fix Information:

This has been fixed in 68XX Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:102  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Validation failed shown on left corner of Resolve Method screen

Problem:

Load a method which was developed for both detectors and has the front detector signal chosen under signals tab.
Load this method on a GC with no front detector - method resolution. 

Changes made to the detector signal tab are not detected, tab remains yellow (invalid).

Temporary Solution:

Click on a different tab and modify the value of any field to a valid value or click on the configuration tab and then click away. Change the signals tab to a valid value or turn it off to resolve the method.

Fix Information:

This has been fixed in 68XX Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:103  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Injector Panels in GC Edit Method Show as Uninstalled When an Injector is Installed.

Problem:

The front injector was unplugged. Close software, plug injector in, restart software.  Method Configuration/ALS panel and Injector panel both say: Front Injector Not Installed. The GC shows that it is installed.  Software inlet and injection parameters panel show that it is installed as well.

Temporary Solution:

n/a

Fix Information:

This has been fixed in 68XX Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:105  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

Plunger Speed is Not Saved with 6850 GC Methods

Problem:

Plunger speed parameters are not saved in methods.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 6.11[002].  This fix is included in B.04.03 SP1 ChemStation, C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom.


KPR#:106  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Opening the GC method parameters to only look (no change made) and selecting OK to close stops the sequence or single run

Problem:

During a run, the GC method parameters are opened via the instrument diagram icon.  Select OK after looking at the oven parameter to close the screen.  The sequence or single run will just stop and the method is dirty.  See Log File for error message.

Log File:
  67 41e2 4e69f8b3    0
Method       Method started                                 07:29:55 09/09/11
  67 41e2 4e69f8b4    0
Method       Instrument running sample Vial 1 (front)       07:29:56 09/09/11
  67 41e2 4e69f8b4    0
Method       Instrument running sample Vial 2 (back)        07:29:56 09/09/11
   0 1da5 4e69f91a    0
6890       1 Attempt to download method to GC failed.       07:31:38 09/09/11
  67 41e2 4e69f91a    0
Method       Instrument Error - Method/Sequence stopped     07:31:38 09/09/11
  67 41e2 4e69f91a    0
Method       Method aborted                                 07:31:38 09/09/11

Temporary Solution:

Click the Cancel button rather than the OK button if the intent was to check the values of some of the instrument setpoints.  This would not interfere with any run which is in progress.

Fix Information:

Fixed in 68xx driver revision 6.11[008].  This is included in B.04.03 SP1 and OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.


KPR#:109  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Method Contents Report Displays the Incorrect Configuration of Front and Back Detectors for 6890 GC

Problem:

The report displays the incorrect configuration. With signal-1 selected as the back detector and integration events for back detector. The method contents report displays the integration events for the front detector and the remaining paramaters are for the back detector are configured to the front detector.

Temporary Solution:

n/a

Fix Information:

Fixed in 68xx driver revision 6.11[003].  This fix is included in B.04.03 SP1 ChemStation, C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom and later.


KPR#:113  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Cannot Run a Packed GC Column with a 6890

Problem:

Method download fails, sequence aborts after start when trying to run a 6890 with a packed column.  
  
Error reported in logbook:
Attempt to download method to GC failed.

The Data System will not allow the run to continue when the method download has failed.

Temporary Solution:

This can be run by setting the column as a capillary column but that changes the chromatography.

Fix Information:

Fixed in driver revision 6.10[032] and 6.11[001].

This fix is included with B.04.03 SP1 GC ChemStation, C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom


KPR#:116  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Pulsed Split/Splitless Injection Does Not Wait For Ready on a 6890.

Problem:

The GC does not wait for the inlet pressure to increase to and stabilize at the pulsed pressure before injection and data acquisition. The GC starts and injects sample at the same time, then starts data acquisition and increase of inlet pressure to pulsed pressure.

Temporary Solution:

If the 6890 is operated from the GC keyboard, the GC has the expected result.

Fix Information:

Fixed in 68xx driver revision 6.11[007].

This fix is included with B.04.03 SP1 ChemStation, OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03


KPR#:117  Product:68xx Driver Software  6890  6.01

Keyword: 68xx Driver

One-line Description:

In 68XX GC Driver Version 6.01, 2 Injections are Made When Using the PTV

Problem:

Set up the injector to make 1 ul injection x 1. The number of injections will revert to 2 even if the user enters 1.

Temporary Solution:

n/a

Fix Information:

This is fixed in driver revision 6.10 [27]  which is included in B.04.03 DSP1 [54] GC ChemStation, C.01.02 OpenLAB ChemStation and A.04.02 OpenLAB EZChrom.


KPR#:119  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Vial left in gripper; no message shown to alert user about the problem

Problem:

After a stop, if there is a vial in the gripper it is not returned to its original location, it is kept in the gripper until the start of a new run. In the interim, if a new vial is placed in the open location, during the next run the system runs into an error because the gripper wants to return the vial to its original location which is now occupied.

Temporary Solution:

n/a

Fix Information:

This is fixed in GC Driver version 6.21, which is available with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:120  Product:68xx Driver Software  6890  6.11

Keyword: 68xx Driver

One-line Description:

6850 vial numbers are not consistent with standard numbering system.

Problem:

A 6850 with a G2613A tower and a three vial turret will accept only single digit vial locations, ie., 1, 2, 3 instead of 101, 102, 103. Attempts to enter the correct vial number generates an error message "The vial 101 is not allowed with the current device configuration."

The also occurs when entering correct vial locations in a sequence table.

Temporary Solution:

n/a

Fix Information:

Fixedi n 68XX driver revision 6.11.  This driver is included with B.04.03 SP1 ChemStation, OpenLAB C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom.


KPR#:122  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Method Resolution Changes Oven Temperature Program

Problem:

Load a method from a 6890 GC with a fast oven on a 6890 GC with different hardware and a normal oven. Method resolution occurs. The oven temperature program from the fast oven GC has not been loaded on the normal oven GC.

Temporary Solution:

n/a

Fix Information:

Fixed in 68xx driver revision 6.11[002].  Fix is included in B.04.03 SP1 ChemStation, OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.


KPR#:123  Product:68xx Driver Software  6890  6.11

Keyword: 68xx Driver

One-line Description:

Column info is missing with 6890

Problem:

The column specified as being used is missing information: description, Manufacturer and Model Number.

Temporary Solution:

n/a

Fix Information:

Fixed in 68xx driver revision 6.11[005].

This fix is included in B.04.03 SP1 GC ChemStation, C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom.


KPR#:124  Product:68xx Driver Software  6890  6.11

Keyword: 68xx Driver

One-line Description:

Method Modification Flag is Set after Edit Method: Occurs on first connection or after upload from GC.

Problem:

1. First Connection to an online GC and save the method OR connect to an instrument and save the method, open method editor, click on ‘Upload from instrument’. Click ‘Ok’.
3. Click ‘Edit entire Method’, go on clicking ‘Ok’ without making any change.
4. Method is dirty at the end.

This only occurs on 6850 GCs with G4513A injectors.

Temporary Solution:

Save the method as a new name after initial connection or subsequent method upload.

Fix Information:

Fixed in 68xx driver revision 6.11[008].  This is included in B.04.03 SP1 and OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.


KPR#:128  Product:68xx Driver Software  6890  6.11

Keyword: 68xx Driver

One-line Description:

Method Ddownload Failure Occurs Due to Column Calibration Limits

Problem:

A valid column calibrated using calculate diameter function will result in a method download failure if the change from the original diameter is more than 20 microns. 

This occurs after OKing method changes that are downloaded, and will occur even if the column values are valid dimensions.

Temporary Solution:

n/a

Fix Information:

This is fixed in 68XX GC Driver Version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:129  Product:68xx Driver Software  6890  6.11

Keyword: 68xx Driver

One-line Description:

Instrument actual value for COC inlet does not display correctly

Problem:

6890N GC with COC inlet connected to FID. The COC inlet pressure setpoint reflects the correct pressure set for the column, however the instrument actual value (blue text) in the setup method screen is displayed as 0.0.

Temporary Solution:

Enable setup actuals (inlet) in the setup method screen, the inlet pressure is properly displayed.

Fix Information:

Fixed in 68xx driver revision 6.11[018].  This driver is included in B.04.03 SP1 ChemStation and OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.


KPR#:130  Product:68xx Driver Software  6890  Monitoring

Keyword: 68xx Driver

One-line Description:

Instrument actual value for anode purge flow on an ECD does not display correctly

Problem:

The ECD anode purge flow setpoint reflects the correct value, however the instrument actual value (blue text) in the setup method screen is displayed as 0.0.

Temporary Solution:

Enable setup actuals in the setup method screen, the actual anode purge flow displayed properly.

Fix Information:

Fixed in 68xx driver revision 6.11[018].  This driver is included in B.04.03 SP1 ChemStation and OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.


KPR#:132  Product:68xx Driver Software  6890  6.21

Keyword: 68xx Driver

One-line Description:

6850 Support of Column Compensation

Problem:

The 6850 GC does not support column compensation from the software. This is only able to be done with the G2629A handheld controller which is now obsolete.

Temporary Solution:

n/a

Fix Information:

This is fixed in 68XX GC Driver Version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:136  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

Data Files do not Contain the Correct Signals if the Front Inlet is Connected to Back Detector

Problem:

Data files do not contain the correct signals if front inlet is connected to back detector

Temporary Solution:

n/a

Fix Information:

This is fixed in GC Driver version 6.21, which is available with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:140  Product:68xx Driver Software  6890  6.23

Keyword: 68xx Driver

One-line Description:

The Oven Ramp table does not display correctly in IR with Acquisition method selected.

Problem:

When you specify an Intellignet Report with Method Info report added to the  Sample Summary report to get the Acquisition method it is formatted incorrectly..

Temporary Solution:

n/a

Fix Information:

The rows in tables now have the same names and IDs so that the data system can organize them into a table. Fixed in revision 6.25 (GC Driver revision A.02.05).


KPR#:141  Product:68xx Driver Software  6890  Monitoring

Keyword: 68xx Driver

One-line Description:

6890 PTV Inlet not recognizable in OpenLAB EZChrom

Problem:

The Extracted configuration shows PTV back inlet and AIB Back Detector. Launching the instrument offline shows the column connected to the Back Inlet and the Back Detector.The Module Configuration Lists the PTV Back Inlet with Blank Back Detector, however the Back Inlet is seen as  Unrecognized inlet type.

Temporary Solution:

n/a

Fix Information:

The code wasn't checking for the right inlet type code for a PTV mounted in the back. Fixed in revision 6.25


KPR#:142  Product:68xx Driver Software  6890  6.23

Keyword: 68xx Driver

One-line Description:

Multiple errors when the 6890 doesn't have any inlets

Problem:

The OpenLAB CDS ChemStation gives errors when the 6890 doesn't have inlets installed and both inlets are configured as none on the 6890 GC.

Temporary Solution:

n/a

Fix Information:

The error was occurring in the code which set up the possible inlet connections for columns. Fixed in revision 6.25


KPR#:143  Product:68xx Driver Software  6890  6.23

Keyword: 68xx Driver

One-line Description:

Sequence table Inj Vol Override goes to wrong stops for 6890/7683: C.01.05, A.04.05, C.01.06, A.04.06

Problem:

If an Injection volume is entered into the Sequence which overrides the method injection volume setting the injector goes to the wrong stops for a 6890 with a 7683 ALS. The injection volume will be incorrect for versions C.01.05, A.04.05, C.01.06, A.04.06 of the software

Temporary Solution:

n/a

Fix Information:

Fixed in 68xx Driver Version 6.23 [011].


KPR#:145  Product:68xx Driver Software  6890  6.22

Keyword: 68xx Driver

One-line Description:

Often get two or more ChemMain.exe's running following shutdown of Control Panel or online session.

Problem:

Run Task manager always on top.  With no ChemMain.exe's running, open a 6890 online session. Close session and the ChemMain.exe does NOT go away.

Temporary Solution:

n/a

Fix Information:

Fixed in A.02.05


KPR#:147  Product:68xx Driver Software  6890  6.25

Keyword: 68xx Driver

One-line Description:

6890N with 7693 barcode scanning checksum mismatch does not return vial to tray and skips a vial.

Problem:

There was a timeout in the driver; if the reading of the bar code was not complete by the time that the timer timed out, the driver would declare that the reading had terminated.  This was misleading the data system into advancing to the next vial.  The time required to finish reading the bar code or declare that the attempt had finished is indeterminate.  If the tray is parked, the system will sit there waiting to move until the tray is unparked.  The timeout was removed from the code.

Temporary Solution:

n/a

Fix Information:

Fixed in A.03.01


KPR#:148  Product:68xx Driver Software  6890  5.03

Keyword: 68xx Driver

One-line Description:

Customer can't install colum information from catalog if column description is blank

Problem:

OpenLAB EZChrom A.04.06 and 6890N 
 
Step:
1, Customer selects 122-10BE column. this column doesn't have description.
 
2, Error message appears when you try to install this column.

Temporary Solution:

n/a

Fix Information:

Fixed in A.03.01


KPR#:149  Product:68xx Driver Software  6890  6.31

Keyword: 68xx Driver

One-line Description:

EZChrom: Installation of  GC Drivers A.03.02 breaks a 7697 configuration if installed

Problem:

Installation of  GC Drivers A.03.01 breaks a 7697 configuration if installed.

Temporary Solution:

Uninstall the 7697 HS driver and install the latest HS driver.

Fix Information:

None


KPR#:150  Product:68xx Driver Software  6890  A.03.01

Keyword: 68xx Driver

One-line Description:

Setting a Hold time of 0.0 does not work correctly with GC6890

Problem:

Setting up Hold Time of 0.0 does not work correctly with GC6890.

Temporary Solution:

No fix planned.

Fix Information:

None


KPR#:154  Product:68xx Driver Software  6890  6.32

Keyword: 68xx Driver

One-line Description:

Elite: G1888 status only shows connecting after GC driver installation

Problem:

1. Configured G1888 using original Elite driver after GC Driver installation
2. Selected Instrument Status under Control menu and chose HS tab.
3. Status window shows "Connecting" (see pic)
4. Other elements in window are inactive.
5. Attempts to create, save, and run a method were successful and data was returned.
6. Instrument Status of the same G1888 prior to GC driver installation was active and usable.

Temporary Solution:

Workaround 
  
1) Open up administrator command prompt
2) Navigate to EZChrom directory, e.g. cd "c:\EZChrom Elite"
3) run: regsvr32 HPCTUtil.dll

Fix Information:

n/a


KPR#:155  Product:68xx Driver Software  6890  6.23

Keyword: 68xx Driver

One-line Description:

Instrument session continues to run in background after session is closed.

Problem:

When an instrument session is closed the ChemMain.exe process continues to run in the background when an instrument connection isn't made or lost. This can lead to multiple ChemMain.exe processes running at the same time causing performance issues.

Temporary Solution:

Close the ChemMain.exe process in Task Manager.

Fix Information:

Fixed in GC Drivers version A.02.05.024


KPR#:159  Product:68xx Driver Software  6890  B.01.01

Keyword: 68xx Driver

One-line Description:

6890 online help has the bad strings in Injector page.

Problem:

6890 online help has the bad strings in Injector page.

Temporary Solution:

n/a

Fix Information:

Resolved in B.01.02


KPR#:272474  Product:    EZChrom Edition A.04.09

Keyword: 7890A Control

One-line Description:

Instrument Type does not include 8860 and 8890 GC

Problem:

When creating a new instrument for the Agilent 8860 GC System or Agilent 8890 GC System, execute the workflow below. 
In order to create an new Agilent 8890 GC System instrument, execute the steps below. 
1. Create a new instrument 
2. Type content for the "Name:" and/or "Description:"
3. Pull down the menu for the "Instrument type:"
4. Select the "Agilent 7890 & 9000 GC Systems" 
5. Select a project if required and Save
6. Select "Configure Instrument" 
7. In "Available Modules" window, move the "Agilent 8890 GC System" to the right and configure. 

For the Agilent 8860 GC System, steps 2 and 4 above, select the "Agilent 7820 GC System" 
The module names might vary in a minor way, depending on the OpenLab CDS EZChrom Edition version and GC Driver version.

Temporary Solution:

n/a

Fix Information:

This might be a coded fix in the future to change the Instrument Type menu.


KPR#:183609  Product:    A.04.07 SR1,A.04.08

Keyword: 7890A Control,Acquisition

One-line Description:

Dual channel issue online plot

Problem:

When the sequence is started the active windows show up as one window for each detector channel (FID Front and FID back). In the front detector signal window the front detector signal is displayed and in the back detector signal window the back detector signal is displayed. However, the sample ID in both windows is the same!

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:268  Product:DSA  7890B  5.01

Keyword: 78xx Driver

One-line Description:

Behavior of the Maintenance Screen When Adding Modules to the GC

Problem:

You will not see the module ID that is installed, but the module location will be identified 

If you add a back injector, the back injector tab will come up and all relevant counters will show.  The module ID will not be shown.  To make it show up, you have to close the data system and reopen the datasystem.

Temporary Solution:

Add hardware modules when the hardware is turned off and the data system is disconnected.

Fix Information:

n/a


KPR#:269  Product:DSA  7890A  5.01

Keyword: 78xx Driver

One-line Description:

ALS Behavior When Aborting a Sequence

Problem:

If a sequence is aborted in the middle of a sequence, the vial will be left in the turret until the next run is started.  At that time, it will be put back in the tray.

Temporary Solution:

This is expected behavior.

Fix Information:

n/a


KPR#:270  Product:DSA  7890B  5.01

Keyword: 78xx Driver

One-line Description:

7890B GC Column Maintenance Counter Behavior

Problem:

If you move a column from one GC to another that is using a different database, you will not see the current Maintenace Counters for the column.  

If you return the column to the original GC, it will have the old values.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:271  Product:DSA  7890B  5.01

Keyword: 78xx Driver

One-line Description:

Update GC Column Catalog dialog is limited to 10 items; no way to display the rest (...)

Problem:

Update Column Catalog dialog is limited to 10 items; no way to display the rest (...)

When you view the catalog, all items will appear.

Temporary Solution:

Description of behavior in driver version 5.01:  when you view the catalog, all items will appear.

Fix Information:

n/a


KPR#:272  Product:DSA  7890B  5.01

Keyword: 78xx Driver

One-line Description:

Method Differences Viewer Behavior with Scanned Columns

Problem:

When you scan a column in to the GC when you are not connected to the datasystem, and you connect with the data system, you will see differences in your method viewer upon opening the data system.  

The change will not be shown in highlight differences.

No matter what is selected, the column that was scanned into the GC will be the column that is configured.

Temporary Solution:

Scan the column in with the data system connected.

Fix Information:

n/a


KPR#:273  Product:DSA  7890A  4.02

Keyword: 78xx Driver

One-line Description:

7890B GC Method Compatibility

Problem:

The7890B GC driver included in the new data systems (A.01.05 OpenLAB CDS and B.07.00 MassHunter) supports up to 12 valves.  The GC driver in previous data systems only reported 8 valves.  Because of this difference, you cannot develop a method on a new data system and run it on an old data sytem.  You can take a method loaded on an old data system and run it on the new data system.

If you try to load a new data system method on an old data system, the behavior that you will see is that after starting a run, the software and the GC just sit there waiting for prerun to start, but nothing happens.

Temporary Solution:

n/a

Fix Information:

Documented Behavior for GC Methods Developed Using A.01.05 OpenLAB CDS and B.07.00 MassHunter.


KPR#:274  Product:DSA  7890B  5.01

Keyword: 78xx Driver

One-line Description:

Logbook Messages for Scanned Consumables Items

Problem:

There is no logbook message about scanned consumables or reset of Maintenance Counter values if the consumable is scanned when the Data System is Offline.

Temporary Solution:

In order for the logbook to capture messages about the scanned consumables or reset of Maintenance Counter values, make sure the Data System is connected when scanning the barcode of the consumable.

Fix Information:

n/a


KPR#:275  Product:DSA  7890B  5.01

Keyword: 78xx Driver

One-line Description:

The Syringe and Liner are Not Shown in the Method

Problem:

The syringe and liner not shown in method.  This is expected behavior for the 5.01 GC driver that is included in the C.01.05 OpenLAB CDS and B.07.00 MassHunter.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:276  Product:DSA  7890B  5.01

Keyword: 78xx Driver

One-line Description:

The PC and GC Clocks Need to be Synchronized to Account for Daylight Savings Time

Problem:

For the resource conservation feature, the GC and PC clocks need to be set to the same time(synchronized).  After daylight savings time, make sure the GC and PC clocks are synchronized by pressing the synchronize clocks button, which is available from the configuration screen and from the maintenance screen.  The times will not be synchronized automatically.

Temporary Solution:

Synchronize the clocks by opening the Maintenance Window, select the Maintenance Actions tab and synchronize the clocks.

Fix Information:

n/a


KPR#:277  Product:DSA  7890A  3.01

Keyword: 78xx Driver

One-line Description:

Logging of Injection Tower Location

Problem:

When the injector tower is moved from front to back position (or the opposite way) the method and sequence run correctly.  However, on the report there is still the original (back) position printed, the change of the position is not tracked.

Temporary Solution:

Recommended steps:  Move the injector to the new location, click "Don't resolve" (DO NOT RESOLVE THE METHOD) and then load the method for that injector position.

Fix Information:

n/a


KPR#:279  Product:DSA  7890B  5.01

Keyword: 78xx Driver

One-line Description:

Explanation for How Default Early Maintenance Feedback Counters Work

Problem:

When "set default recommended values for limits" is selected for the first time, and if those values are not changed, the Maintenance window needs to be closed and re-opened for the values to take effect.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:281  Product:DSA  7890B  5.01

Keyword: 78xx Driver

One-line Description:

The 7890 GC Goes Offline When Non-English Characters are Entered into the Column Information

Problem:

The 7890 GC goes offline when non-english characters are entered into the column information screen.  The issue has to do with the number of characters that the column description accepts.

Temporary Solution:

If non-english text is entered in the column information field and the GC can not connect to the data system, reboot the GC by holding the clear button down.  Do not enter the column information in non-english text.

Fix Information:

n/a


KPR#:282  Product:DSA  7890B  5.01

Keyword: 78xx Driver

One-line Description:

Instrument Actuals Show "GC is Offline" Quickly at the Method or Sequence Start, then Returns Back Online and the Method Does Not Run

Problem:

Instrument actuals show that the "GC is Offline" quickly at the start of the method and/or sequence and then returns back online.  The method and/or sequence does not run.  This is likely caused by a difference in the method parameters and what is loaded on the GC.

Temporary Solution:

n/a

Fix Information:

Open a brand new Def.M and create a new method and/or sequence.  Download the method to the GC from the software.


KPR#:283  Product:DSA  7890A  5.01

Keyword: 78xx Driver

One-line Description:

78xx GC configuration fails; Failed to Create Instance

Problem:

There is an issue with the GC Drivers Install Package that is included on the "July 2013" Disk 3 and Disk 5 of OpenLAB CDS.

When Configuring the 78xx GC, failed to create instance error message appears.

Temporary Solution:

n/a

Fix Information:

Use the August 2013 Disk 3 and Disk 5 when installing OpenLAB CDS with a GC.
 
Disk3_M8301-60035     OpenLAB ChemStation Drivers C.01.05 (Aug 2013)
Disk5_M8201-60029     OpenLAB CDS EZChrom Edition Drivers A.04.05 (Aug 2013)


KPR#:284  Product:DSA  7890B  5.01

Keyword: 78xx Driver

One-line Description:

Data not collected, sequence stopped when vial removed from turret after injection

Problem:

After the sample was injected, the vial was removed from the turrett.  The tray showed an error and the log file reported that there was no vial in gripper.  The Sequence was stopped when the vial was removed and the data was not collected.

Temporary Solution:

n/a

Fix Information:

Fixed in driver 5.03.


KPR#:285  Product:DSA  7890B  5.03

Keyword: 78xx Driver

One-line Description:

C.01.06: Import eMethod or Export eMethod freezes system

Problem:

In GC ChemStation C.01.06, the system will freeze when selecting either "Export eMethod" or "Import eMethod".  The status line in the ChemStation will show PrintMethod macro completed, but the system will be frozen and task manager must be used to close ChemStation.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 5.05


KPR#:287  Product:DSA  7890B  5.03

Keyword: 78xx Driver

One-line Description:

0-volume blank run in overlapped sequence causes injection

Problem:

In EZChrom, a blank run can be specified by specifying 0 as the injection volume.  If a blank run is so specified following an ALS non-blank run in a sequence and the sequence is run with overlap enabled, the injector makes a zero-volume injection for that run.  This means the sample will be moved from its location in the tray and subjected to any heating, mixing, bar code reading, sample washes, and sample pumps specified in the method, as well as having the vial's septum pierced and the syringe needle stuck into the sample.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 5.05 Fixed in revision 5.05


KPR#:294  Product:DSA  7890B  5.05

Keyword: 78xx Driver

One-line Description:

Standard and Easy Sample Prep programs run incorrectly in a calibration sequence (run for unknown samples only) when overlap is involved.

Problem:

Standard and Easy Sample Prep programs run incorrectly in a calibration sequence (run for unknown samples only) when overlap is involved. Sample Prep is correctly run for samples only when overlap is NOT involved. Also works fine with overlap inv lved and NO calibration. Also happens the same way in Easy Sample Prep. Calibrated sequences using Sample Prep/Run for Unknown Samples only/Overlap method do not execute correctly.

Temporary Solution:

None. Do not use 'don't prep calibration vials' and 'overlapped prep' together, as they don't work.

Fix Information:

n/a


KPR#:295  Product:DSA  7890B  5.03

Keyword: 78xx Driver

One-line Description:

7890B configuration, model number serial number and IP address change after restart

Problem:

During installation the 7890B is configured with an IP address, the instrument configured in ChemStation and the details are confirmed. The ChemStation is restarted and the GC rebooted. The 7890 IP address changes and the Model and serial numbers change in the configuration to 7890A and CNxxxxxxxx..

Temporary Solution:

n/a

Fix Information:

Fixed in revision 5.05 [019]


KPR#:296  Product:DSA  7890B  5.03

Keyword: 78xx Driver

One-line Description:

Makeup Gas mode terminology should be consistent with the GC front panel

Problem:

Makeup Gas mode terminology should be consistent with the GC front panel display. Causes confusion when setting up the system.

Temporary Solution:

n/a

Fix Information:

Also changed wording for ECD and TCD to be consistent with the GC display Fixed in revision 5.05


KPR#:297  Product:DSA  7890A  5.02

Keyword: 78xx Driver

One-line Description:

Agilent Solvent Elimination Wizard issue

Problem:

When select Solvent Vent mode in MMI, select 10uL syringe and Injection Volume 5uL in Solvent Elimination Calculator on first window. In second window, injected volume 5uL is out of the range. Because the range is 0.01-0.25uL. This is solvent elimination calculator issue.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 5.05


KPR#:298  Product:DSA  7890B  5.03

Keyword: 78xx Driver

One-line Description:

H2 Sensor button crashes EZChrom

Problem:

OL EZChrom A.04.05
GC driver update A.02.04.005 installed

1. Configured a 7890B with H2 sensor 
2. Opened instrument
3. Selected Agilent 7890B / Maintenance from Control Menu
4. Selected Maintenance Actions tab.
5. Clicked Set Sensor Information. (see pic)
6. No result evident. Could not close Maintenance window, nor Instrument window. EZChrom totally unresponsive.
7. Task Manager required to close instrument.
8. Also occurred when driver package was installed on EZChrom A.04.04

Temporary Solution:

n/a

Fix Information:

Fixed in revision 5.05


KPR#:299  Product:DSA  7890B  5.03

Keyword: 78xx Driver

One-line Description:

Clicking OK after making changes to Events does not save the changes made to the events.

Problem:

Clicking OK after making changes to Events does not save changes made to the events. If you click Apply first the events get saved.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 5.05


KPR#:301  Product:DSA  7890A  5.03

Keyword: 78xx Driver

One-line Description:

Can't select uECD makeup mode to "Does not affect Makeup Flow"

Problem:

Can't select uECD makeup mode to "Does not affect Makeup Flow"

If you select "Included in Makeup Flow" on uECD setting screen can't select "Does not affect Makeup Flow".    
  
If you remove column configuration from uECD both selection appear but, these are grayed out.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 5.05


KPR#:302  Product:DSA  7890A  5.03

Keyword: 78xx Driver

One-line Description:

Can't save "Included in Makeup Flow" on uECD setting screen.

Problem:

Can't save "Included in Makeup Flow" on uECD setting screen.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 5.05


KPR#:304  Product:DSA  7890B  5.03

Keyword: 78xx Driver

One-line Description:

The IQT is failing for GC Drivers version A.02.03.074(5.03) for Chinese and Japanese and passing for English in EZChrom A.04.05 with update 5 is installed.

Problem:

The IQT is failing for GC Drivers version A.02.03.074 for Chinese and Japanese and passing for English in EZChrom A.04.05 with update 5 is installed. Files generated for the IQT by the GC driver and EZChrom Update 5 conflict and cause the IQT failure.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 5.05.


KPR#:305  Product:DSA  7820A  5.03

Keyword: 78xx Driver

One-line Description:

7820 GC:  Method modification flag set after starting a sequence or method

Problem:

Method is saved.  Start the 7820 sequence.  The method modification flag is set indicating the method was changed, however, no changes were made.
The report will indicate the method was modified when it was not.

Temporary Solution:

n/a

Fix Information:

Fixed in 5.05.019.


KPR#:306  Product:DSA  7890B  5.05

Keyword: 78xx Driver

One-line Description:

EZChrom: Method returns to the "untitled.met" after running single or sequence.

Problem:

Method returns to the "untitled.met" after running single or sequence.

Temporary Solution:

n/a

Fix Information:

Fixed in A.02.05.024


KPR#:307  Product:DSA  7890B  5.05

Keyword: 78xx Driver

One-line Description:

EZChrom: Method Parameters incorrect after loading a new method. OpenLAB CDS EZChrom

Problem:

Method Parameters mismatch after loading new method. Load a method and then load a second method. Not all of the parameters are loaded.

Temporary Solution:

n/a

Fix Information:

Fixed in A.02.05.023


KPR#:308  Product:DSA  7890B  5.01

Keyword: 78xx Driver

One-line Description:

Method printout missing column parameters: flow or pressure, setpoint On, (Initial), and post run

Problem:

Using Print Method from the ChemStation Software C.01.05, the method printout is missing column parameters: flow or pressure, setpoint On, (Initial), and post run.

Column
Column #1
Flow or Pressure
Setpoint On
(Initial) 2.3922 mL/min
Post Run 2.1209 mL/min
Agilent 19091J-413: 496.46579
HP-5 5% Phenyl Methyl Siloxan

Temporary Solution:

n/a

Fix Information:

Resolved in A.02.05.024.


KPR#:309  Product:DSA  7890B  5.05

Keyword: 78xx Driver

One-line Description:

There is an incorrect character when displaying temperature information in the method report in Japanese

Problem:

There is an incorrect character when displaying temperature information in the method report in Japanese.  There is a question mark instead of a dash.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:310  Product:DSA  7890B  A.02.06

Keyword: 78xx Driver

One-line Description:

MMI Cryo Fast Cooldown not saved

Problem:

The MMI has a parameter called Cryo fast cool down that turns the coolant on as soon as the run is finished to cool the inlet down quicker.
When this parameter is checked in the Method set up screen and then applied and OK it is not downloaded to the GC and when Method edit is launched the check box is unchecked again.
There is a workaround which is to manually turn it on from the GC Keyboard, check the fast cool down box and Press Upload from Instrument. This is the only way that the parameter stays in both GC and ChemStation.

Temporary Solution:

n/a

Fix Information:

Fixed in a.03.01


KPR#:311  Product:DSA  7820A  5.03

Keyword: 78xx Driver

One-line Description:

Method is always seen as modified, nothing can be added to the queue.

Problem:

The septum purge on/off state and flow value were being considered as part of the method, although the 7820 does not allow these items to be changed by software. The flow value for the septum purge (and total flow, which has the purge as a component), obtained by uploading the method is thus an actual and not a setpoint.

Temporary Solution:

n/a

Fix Information:

Fixed in A.02.05


KPR#:312  Product:DSA  7890B  5.01

Keyword: 78xx Driver

One-line Description:

Oven Slow fan will be off after method is downloaded the slow oven mode on to 7890

Problem:

OL CDS EZChrom A,04.05 and 7890B 02.03.1
step1. turn on Oven Slow fan, 
step2. click "Download Method to GC".
step3. 7890B turned Oven Slow fan on correctly. then "Method Download Complete" message appeared on OL EZChrom.
step4. instrument set up screen refreshed.
step5. Oven Slow fan check box turned off.

Temporary Solution:

n/a

Fix Information:

Fixed in A.03.01


KPR#:313  Product:DSA  7890B  5.04

Keyword: 78xx Driver

One-line Description:

During a pulsed splitless run, the purge time shorter than the pulse time cannot be entered.

Problem:

On a 7890B one of the modes of operation is pulsed splitless. In some cases the purge time needs to be shorter than the pulse time.
Recent changes in GC driver has invalidated this parameter change. Many customers need to turn the purge time shorter than the pulse time.

Temporary Solution:

n/a

Fix Information:

Fixed in A.03.01


KPR#:314  Product:DSA  7890B  5.03

Keyword: 78xx Driver

One-line Description:

Method print out with 7890 configured does not show total GCc run time.

Problem:

With 7890 configured; printing the method does not show total GC run time, however if 6890 configured it does show total run time.

Temporary Solution:

n/a

Fix Information:

Fixed in A.03.01


KPR#:315  Product:DSA  7820A  5.05

Keyword: 78xx Driver

One-line Description:

There is no  "Inlet Septum Purge" event in Runtime Events table - EZChrom

Problem:

Method setting;
On OpenLAB EZChrom Online session, 
Click dropdown list on Runtime events table, there is no "Inlet Septum Purge" event in this list.

Temporary Solution:

n/a

Fix Information:

Fixed in A.03.01


KPR#:316  Product:DSA  7890B  A.02.05

Keyword: 78xx Driver

One-line Description:

External Device "Not Ready" ignored by  OpenLAB CDS ChemSataion and EZChrom

Problem:

When doing a valve injection with a Multi-Position valve and an external device the system ignores the external device readiness. When the external device is "Not Ready" the sequence will continue to make the injection and ignore the APG readiness. If a sequence is initiated locally at the front panel of the 7890 with the external device being "Not Ready" the sequence will wait for the external device to become ready prior to making the injection and starting the run. (This is the correct behavior.)

Temporary Solution:

n/a

Fix Information:

Fixed in A.03.01


KPR#:318  Product:DSA  ICF (DSA)  5.02

Keyword: 78xx Driver

One-line Description:

Aux Flow Limited to 150 mL/min even when pressure is less than Maximum

Problem:

Aux Flow Limited to 150 mL/min even when pressure is less than Maximum

Temporary Solution:

n/a

Fix Information:

Fixed in A.02.05


KPR#:319  Product:DSA  ICF (DSA)  5.02

Keyword: 78xx Driver

One-line Description:

Post Run Flow Rate has default that out of range and value seems not to be stored

Problem:

Post Run gas flow defaults to 250 mL min even though the highest allowed value is 150 mL min. 
Even if you change it to save the analysis method as soon as you re-open the method it defaults back to 250 mL min.

Temporary Solution:

n/a

Fix Information:

Fixed in A.03.01


KPR#:320  Product:DSA  ICF (DSA)  5.02

Keyword: 78xx Driver

One-line Description:

Method User Interface does not correctly load the content of a method.

Problem:

The content of a loaded method is only partly reflected in the method user interface. I seems that only the visible tab is updates.   
Steps to reproduce:
1.) Create a new method 
2.) Change Oven Ramp, Events table or Signals 
3.) Go back to the ALS Tab and save the method 
4.) Create new method or load a different one
5.) Load the method saved in step 3


Verify the content of Oven Ramp, Events table or Signals. It does not display the correct content. 
See also attached document. 


This issue was found during the integration in Waters UNIFI and could also be reproduced in ICF TestApp and Shimadzu LabSolutions.

Temporary Solution:

n/a

Fix Information:

Fixed in A.03.01


KPR#:324  Product:DSA  7890A  A.02.05

Keyword: 78xx Driver

One-line Description:

Does not delete previous open GC method event table when load method. Wrong injections can be created.

Problem:

Steps:
Open instrument setup panel
Create method with some lines into the event table. 
Save method
Create a second method with empty event table.
Click detectors tab to see detectors info.
Save this method.
Open the first method, detectors tab is shown.
Go to events table and it is blank.

Temporary Solution:

n/a

Fix Information:

Fixed in A.03.02


KPR#:325  Product:DSA  7890B  5.05

Keyword: 78xx Driver

One-line Description:

Method download to GC while in sequence causes it to disconnect and sequence stops.

Problem:

When a method is downloaded to the GC while running a sequence the GC disconnects and the sequence stops.

Temporary Solution:

n/a

Fix Information:

Fixed in A.03.02


KPR#:326  Product:DSA  7890B  6.03

Keyword: 78xx Driver

One-line Description:

7890 method print doesn't include post run time

Problem:

When printing a method the GC setpoints section does not include the post run time as expected.

Temporary Solution:

n/a

Fix Information:

Fixed in A.03.02


KPR#:335  Product:DSA  7890B  5.03

Keyword: 78xx Driver

One-line Description:

Crash of Software after check "Show Threshold Value as remaining time" in EMF Options

Problem:

Once you check the "show threshold value as remaining time" in EMF options, and close the EMF dialog. Once you re-enter the dialog, software freezes and has to be closed by task manager.

Temporary Solution:

n/a

Fix Information:

Fixed in A.03.02


KPR#:339  Product:DSA  7890A  A.03.02

Keyword: 78xx Driver

One-line Description:

Not allowed  to set proper GC column max temp because it is less than 150degrees.

Problem:

Problem when a column has the following restrictions:
The maximum allowable operating temperature Isothermal: 130 °C
The maximum allowable operating temperature Programmed: 150 °C
When they want to enter this number in this configuration the field becomes yellow.

This value can can't be entered in the field due to value limitations of related field. Maximum value should be in the range of 150 ≤ °C ≤ 450.

Temporary Solution:

n/a

Fix Information:

Fixed in B.01.01


KPR#:340  Product:DSA  7890B  A.03.02

Keyword: 78xx Driver

One-line Description:

FID signal subtraction setting changes spontaneously from "Nothing" to "Column Compensation" during run.

Problem:

When customer set FID - > Subtraction from signal value to "Nothing". Save method and upload method to instrument. Function is automatically changed to option "Column compensation run. During measure, there is activity log settings that the settings for method was changed in Subtract from signal function.

Temporary Solution:

n/a

Fix Information:

Fixed in B.01.01


KPR#:341  Product:DSA  7890B  A.03.02

Keyword: 78xx Driver

One-line Description:

"Extend Run" in GC settings does not work correctly

Problem:

Setting the run time to extend the run for a length of time is not handled correctly. The length of time is longer than the setpoint.

Temporary Solution:

n/a

Fix Information:

Fixed in B.01.01


KPR#:342  Product:DSA  7890B  A.03.02

Keyword: 78xx Driver

One-line Description:

Cannot configure a 7890B on an existing client server system

Problem:

Error in configuration panel  "There are unconfigured modules"

Temporary Solution:

n/a

Fix Information:

Fixed in B.01.01


KPR#:343  Product:DSA  7890B  5.32

Keyword: 78xx Driver

One-line Description:

Copy to Clipboard Function in Instrument Actuals throws exception error

Problem:

Customers have reported encountering an exception when using the "Copy to Clipboard" function.
Intermittent errors when clicking the ‘copy to clipboard’ button in the Status UI on the 7890 – which gives an exception error.

Temporary Solution:

n/a

Fix Information:

Fixed in B.01.01


KPR#:344  Product:DSA  7890B  A.03.02

Keyword: 78xx Driver

One-line Description:

The column flow calculator calculation is off by almost 3x the flow rate for the system

Problem:

The values which are setup for the column and the flow rates which were entered. The problem is when you look at the GC for the values they don't match the column flow calculator.

Temporary Solution:

n/a

Fix Information:

Fixed in B.01.01


KPR#:345  Product:DSA  7890A  5.32

Keyword: 78xx Driver

One-line Description:

Sample Overlap settings can not be edited after driver update, but are still part of the method.

Problem:

With OpenLAB EZChrom A.04.07 SR1 Update2 and GC driver version A.02.05.021 (from OpenLAB installation media) on a 7890A+ with 7693 sampler under ALS/ [Tray/Other] Sample Overlap parameters can be configured and saved to the method. Those settings are still active after upgrading to GC Drivers A.03.02.019 in the method and can cause that 2 methods behave in terms of sample overlap totally different. But they can no more been edited with the new driver.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:346  Product:DSA  7890B  5.32

Keyword: 78xx Driver

One-line Description:

Error Unhandled Exception when changing columns and a Gerstel CIS4 in installed in the 7890.

Problem:

If e a Gerstel CIS4 installed and the columns is reconfigured, a msinctl error occurs and the MassHunter software shuts down.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:347  Product:DSA  7890B  5.04

Keyword: 78xx Driver

One-line Description:

Incomplete Method Parameters for xCD in Acquisition Method Report

Problem:

When we do an Acquisition report, not all the xCD parameters are stored. The Reaction Cell Pressure and the Flow through the Ozone generator should also be reported. Can be confusing to customer.

Temporary Solution:

None

Fix Information:

n/a


KPR#:349  Product:DSA  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

Actual flows displayed by CDS aren't the same as the setpoint flows

Problem:

Examples:

Setpoint     Actual

3.000          3
2.998          3
2.95            3
2.92            2.9

The display for the 7890 shows three digits to the right of the decimal for both the setpoint and the actual. 
As shown above, the data system sometimes shows one digit to the right of the decimal and sometimes nothing to the right of the decimal (e.g. just "3")

Temporary Solution:

none

Fix Information:

Resolved in revision B.01.02


KPR#:350  Product:DSA  All GCs  B.01.01

Keyword: 78xx Driver

One-line Description:

Print method does not include all MMI setpoints: Purge time, Post Run Temp, Time, Flow

Problem:

Print method for the Intuvo 9000 GC doesn't have any MMI Purge time or Post Run temp or Post Run total Flow or time.

Temporary Solution:

n/a

Fix Information:

Resolved in revision B.01.02


KPR#:351  Product:DSA  All GCs  B.01.01

Keyword: 78xx Driver

One-line Description:

Extend run does not extend the acquisition of the Intuvo 9000 GC

Problem:

Intuvo 9000 GC on C.01.07.255 SR2. Extend run does not extend the run. LUI shows Idle during the extended part of the run.

Temporary Solution:

none

Fix Information:

Resolved in revision B.01.02


KPR#:352  Product:DSA  All GCs  B.01.01

Keyword: 78xx Driver

One-line Description:

Cannot Add User-Defined EMFs on Intuvo 9000 GC

Problem:

Unable to created a user defined EMF Intuvo 9000 GC. Enter a name and tracking type, but nothing appears on the User-defined panel or any other panel.  There are five user defined (seemingly blank) entries on that panel.

Temporary Solution:

none

Fix Information:

To Be fixed in B.01.02


KPR#:353  Product:DSA  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

EZChrom: Some settings not accepted after method migration

Problem:

OL EZChrom A.04.07 SR2
GC Drivers--> B.01.01

Connected to Intuvo 130.30.246.219

1. Loaded a 7890B method with inlet mode set to Split and ALS Error set to Skip on a 9000GC

2. After resolving the method, found most settings had migrated over (configurations were similar) 

3. The inlet mode retained its original Splitless setting.

4. The ALS Errors setting of PAUSE was also retained.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:355  Product:DSA  All GCs  B.01.01

Keyword: 78xx Driver

One-line Description:

ChemStation Instrument actuals showing D2 Bus Temperature with no D2 bus installed

Problem:

ChemStation Instrument actuals showing Second Detector Bus Temperature with no second Detector.

Temporary Solution:

n/a

Fix Information:

Resolved in revision B.01.02


KPR#:356  Product:DSA  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

Method print is showing Solvent wash mode as A-A2, B-B2 when it should be A,B since it is a 16 vial turret

Problem:

Intuvo 9000 with a ALS and a 16 vial turret.  MassHunter software, the configuration of the ALS is set to Solvent Wash Mode A,B.  Saved and printed the method.  Solvent Was Mode on the method print shows A-A2, B-B2.

Temporary Solution:

n/a

Fix Information:

Resolved in revision B.01.02


KPR#:357  Product:DSA  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

Inlet pressure setpoint displayed assuming all thermal zones are turned ON, results in confusing mismatch when heated zones are Off.

Problem:

CDS displays inlet pressure setpoint assuming all thermal zones are turned ON, results in confusing mismatch when heated zones are Off.

Temporary Solution:

n/a

Fix Information:

Resolved in revision B.01.02


KPR#:359  Product:DSA  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

IQT report failed on EZChrom Elite 3.3.2 SP2 with GC Driver B.01.01.

Problem:

IQT report failed on EZChrom Elite 3.3.2 SP2 with GC Driver B.01.01. There is a technology change between from IQT to SVT that causes this to happen.

Temporary Solution:

none

Fix Information:

n/a


KPR#:360  Product:DSA  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

Method Editor for PP inlet & Packed Column incorrectly allows both flow and pressure setpoints to be entered

Problem:

When in flow mode, pressure should not be allowed. Entering a value messes up the flow setpoint.
When in pressure mode, flow should not be allowed. Entering a value messes up the pressure setpoint.

Temporary Solution:

n/a

Fix Information:

Resolved in revision B.01.02


KPR#:361  Product:DSA  7890B  B.01.02

Keyword: 78xx Driver

One-line Description:

ChemStation can lock up when changing Intuvo Bus Flow Chip types with ChemStation session running.

Problem:

ChemStation can lock up when changing Intuvo Bus Flow Chip types with ChemStation session running. If replacing the flow chip with the same type then the problem will not occur.

Temporary Solution:

Workaround for now is to disconnect ChemStation, then turn off GC, do work, turn on GC, then connect ChemStation.

Fix Information:

n/a


KPR#:364  Product:DSA  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

Selection box remains after selection made on Events page

Problem:

Selection box remains after selection made on Events page. Drop down box remains open.

Temporary Solution:

n/a

Fix Information:

Resolved in revision B.01.02


KPR#:365  Product:DSA  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

Gray bars hide setpoints after context menu download of method

Problem:

Gray bars hide setpoints after context menu download of method after editing

Temporary Solution:

n/a

Fix Information:

Resolved in revision B.01.02


KPR#:366  Product:DSA  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

Multiple clicks necessary for combo box to list choices.

Problem:

Multiple clicks necessary for combo box to list choices.

Temporary Solution:

n/a

Fix Information:

Resolved in revision B.01.02


KPR#:367  Product:DSA  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

Right Clicking on Method Editor makes UI hide behind Instrument Screen

Problem:

Right Clicking on Method Editor makes UI hide behind Instrument Screen

Temporary Solution:

n/a

Fix Information:

Resolved in revision B.01.02


KPR#:368  Product:DSA  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

Incorrect Values in 9000 Pressure/Flow Calculator

Problem:

The values within the pressure/flow calculator when initializing from "Inlet 1" still do not match the instrument parameters.

Temporary Solution:

n/a

Fix Information:

Resolved in revision B.01.02


KPR#:369  Product:DSA  All GCs  B.01.01

Keyword: 78xx Driver

One-line Description:

Make changes to guard chip on Intuvo 9000 GC, audit trail just says temperature changed

Problem:

With Audit trail enabled, I make changes to the GC parameters panel of the Intuvo 9000 GC.  Changed the guard chip temperature and hold time and saved the method.  Looked at the audit trail and it showed events but it just said Temperature setpoint changed and Temperature hold time changed.  There was no reference to the guard chip.

Temporary Solution:

n/a

Fix Information:

Resolved in revision B.01.02


KPR#:370  Product:DSA  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

Backflush Wizard illogically sets trial backflush times after the elution time of last peak of interest

Problem:

Backflush Wizard illogically sets trial backflush times after the elution time of last peak of interest

Temporary Solution:

n/a

Fix Information:

Resolved in revision B.01.02


KPR#:371  Product:DSA  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

Should not be able to download a method while a in a GC run or sequence.

Problem:

Downloading a method while a in a GC run or sequence can cause issues.

Temporary Solution:

n/a

Fix Information:

Resolved in revision B.01.02


KPR#:375  Product:DSA  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

No indication that the AutoSampler tray is in park mode.

Problem:

From the data system there isn't a way to determine that the AutoSampler tray is in park mode.

Temporary Solution:

n/a

Fix Information:

Resolved in B.01.03


KPR#:376  Product:DSA  All GCs  B.01.01

Keyword: 78xx Driver

One-line Description:

The Local User Interface(LUI) on the Intuvo 9000 GC shows long name for connected data system.

Problem:

The front panel LUI shows the PC name with other information appended to it. Customers would like to see just the PC name.

Temporary Solution:

n/a

Fix Information:

Fixed in B.01.03


KPR#:377  Product:DSA  7890B  Future

Keyword: 78xx Driver

One-line Description:

For both the 7890 and Intuvo 9000 there is no change in the UI to show the user that a rack is not present.

Problem:

For both the 7890 and Intuvo 9000 there is no change in the UI to show the user that a rack is not present.

Temporary Solution:

n/a

Fix Information:

Resolved in B.01.03


KPR#:378  Product:DSA  All GCs  B.01.02

Keyword: 78xx Driver

One-line Description:

Post Run Backflush. The Guard Chip temperature was not set to the Post Run Oven Temperature.

Problem:

Using an Intuvo 9000 with Post Column Restrictor. The Backflush Wizard to set the Post Run Backflush. The Guard Chip temperature was not set to the Post Run Oven Temperature.

Temporary Solution:

n/a

Fix Information:

Resolved in B.01.03


KPR#:379  Product:DSA  All GCs  B.01.02

Keyword: 78xx Driver

One-line Description:

Location of Intuvo 9000 GC System Health Report should be moved

Problem:

Location of Intuvo 9000 GC System Health Report is not consistent with the location on the LUI of the GC.

Temporary Solution:

The System Health Report location has been moved to the  Maintenance/Diagnostics screen.

Fix Information:

Resolved in B.01.03


KPR#:380  Product:DSA  7890B  B.01.02

Keyword: 78xx Driver

One-line Description:

Link to Intuvo 9000 GC help does not work.

Problem:

Click on options > Show 9000 Help and Information. This gives a message that the 9000 cannot be found.

Temporary Solution:

n/a

Fix Information:

Resolved in B.01.03


KPR#:381  Product:DSA  All GCs  B.01.02

Keyword: 78xx Driver

One-line Description:

Event drop downs are black when creating a new method

Problem:

When creating a new GCMS method and adding events, drop downs are black. If the screen is switched and you go back to events, it works.

Temporary Solution:

n/a

Fix Information:

Resolved in B.01.03


KPR#:383  Product:DSA  7890B  B.01.03

Keyword: 78xx Driver

One-line Description:

PCM will not control in constant flow.

Problem:

PCM will not control in constant flow. Configured two packed columns using PCM Module to a single detector. The PCM will control in pressure mode but not flow mode. This was reported by a customer using OpenLAB 2.2 and tested using OpenLAB CDS ChemStation. .

Temporary Solution:

Use pressure control mode and ramp the pressure.

Fix Information:

Resolved in B.01.04


KPR#:388  Product:DSA  7890B  B.01.03

Keyword: 78xx Driver

One-line Description:

RTL for OpenLAB CDS 2.X does not calculate the pressure correctly when locking or re-locking the method when setting a Target retention time.

Problem:

RTL for OpenLAB CDS 2.X does not calculate the pressure correctly when locking or re-locking the method. This only occurs when the Target Retention field is set to a different retention time than the peak in the chromatogram.

Temporary Solution:

The work-around is to not update target retention time when selecting the peak for RTL locking. Use the retention time of the peak selection and the calculations are correct.

Fix Information:

Resolved in B.01.05


KPR#:422  Product:DSA  7890A  5.01

Keyword: 78xx Driver

One-line Description:

HS Sleep/Wake mode should sync with GC Sleep/Wake mode

Problem:

When a 7697 is configured to a 7890B, it should communicate such that the Sleep or Wake state of either one should be mimicked on the other.

Temporary Solution:

n/a

Fix Information:

Resolved in B.01.05


KPR#:431  Product:DSA  All GCs  B.01.04

Keyword: 78xx Driver

One-line Description:

Overlap injections on the 9000 uses only the injection volume from the first injection for subsequent injection.

Problem:

Overlap injections on the 9000 uses only the injection volume from the first injection for subsequent injection.

Temporary Solution:

n/a

Fix Information:

Resolved in B.01.05


KPR#:439  Product:DSA  All GCs  3.0

Keyword: 78xx Driver

One-line Description:

Downloading method parameters (like oven temperature) on a 6890A does not work with OpenLab CDS 2.4

Problem:

Downloading method parameters (like oven temperature) on a 6890 does not work with OpenLab CDS 2.4. The GC is connected (current values are correctly displayed) and setting these parameters from the keyboard does work. Settings are updated in the method when selecting "Upload method".

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:472  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: Intuvo trigger two instrument notification for configuration changed

Problem:

Chromeleon: Intuvo trigger two instrument notification for configuration changed. Configure an Intuvo with guard chip feature and open the online method in context menu of status panel and get two config changed events with a timely distance of 30 seconds.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:473  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: GC: Change of column length (and related parameters like flow etc) via calibration in online method sometimes reported in audit trail sometimes not

Problem:

Chromeleon: GC: Change of column length (and related parameters like flow etc) via calibration in online method sometimes reported in audit trail sometimes not

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:474  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: GC: Column length can only be calibrated once in Online Method Editor

Problem:

Chromeleon: GC: Column length can only be calibrated once in Online Method Editor

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:475  Product:DSA  ICF (DSA)  B.01.05

Keyword: 78xx Driver

One-line Description:

Chromeleon: Setting Liner on Intuvo triggers two config change messages in audit trail

Problem:

Chromeleon: Setting Liner on Intuvo triggers two config change messages in audit trail

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:476  Product:DSA  All GCs  3.0

Keyword: 78xx Driver

One-line Description:

8860 Open Lab CDS EZChrom compact license is invalid.

Problem:

8860 Open Lab CDS EZChrom compact license is invalid.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:477  Product:DSA  All GCs  3.1

Keyword: 78xx Driver

One-line Description:

Selecting Partial Data for the 6890 causes the retention times and areas to be wrong.

Problem:

Selecting Partial Data for the 6890 causes the retention times and areas to be wrong.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:478  Product:DSA  All GCs  3.1

Keyword: 78xx Driver

One-line Description:

6890 - cannot set post run flow for packed column

Problem:

6890 - cannot set post run flow for packed column

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:479  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: GC 6850A and 6890A: GC online method is not editable

Problem:

Chromeleon: GC 6850A and 6890A: GC online method is not editable

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:480  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: GC 6890A: Wrong temperature shown in chromatogram for signal GC_OvTempe_Act

Problem:

Chromeleon: GC 6890A: Wrong temperature shown in chromatogram for signal GC_OvTempe_Act

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:481  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: GC 6850A: Cannot create instrument method

Problem:

Chromeleon: GC 6850A: Cannot create instrument method

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:482  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: GC8890: Current Signal Value in 'Blank evaluation' shows incorrect value in Method editor

Problem:

Chromeleon: GC8890: Current Signal Value in 'Blank evaluation' shows incorrect value in Method editor

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:483  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: GC8890: Blank evaluation setpoints missing in method report

Problem:

Chromeleon: GC8890: Blank evaluation setpoints missing in method report

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:484  Product:DSA  All GCs  3.1

Keyword: 78xx Driver

One-line Description:

Signal Scaling is incorrect for the 6890

Problem:

The real time plot scaling is incorrect. The offset shown in the GC Status is 10.2. The Real time plot shows the signal level to be about 7.8 x10^4. It appears that the signal is not being scaled correctly, but is using the signal directly.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:485  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: Intuvo: Help button in window 'Configure Agilent Intuvo 9000 GC' opens 78xx Series GC Help

Problem:

Chromeleon: Intuvo: Help button in window 'Configure Agilent Intuvo 9000 GC' opens 78xx Series GC Help

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:486  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: GC8890: Issues during sequence execution

Problem:

Chromeleon: GC8890: Issues during sequence execution. During the execution of a sequence (using the blank evaluation) I observed some issues:
1.no audit trail entry if blank evaluation was used or not ( was performed as blank run, blank evaluation option enabled)
2.during execution of sequence the GC8890 reported high CPU usage
3.The blank evaluation failure was reported during : 16.04.2019 17:20:29 +02:00 2,059 Agilent Blank Evaluation Failed, Failure Action is Continue

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:487  Product:DSA  ICF (DSA)  3.0.2

Keyword: 78xx Driver

One-line Description:

Chromeleon: Changes made in 68xx Edit Online Method not applied or reported

Problem:

Chromeleon: Changes made in 68xx Edit Online Method not applied or reported.
1. With Chromeleon open, selected to open Instrument Actuals
2. Right click in window and select Edit Online Method
3. Attempted several changes to online method.
4. Clicked Set Method Button.
5. No change in instrument actuals reflecting changes to method
6. No audit trail entries for changes made.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:488  Product:DSA  ICF (DSA)  3.1

Keyword: 78xx Driver

One-line Description:

Error when loading GC method without Blank Run Analysis setpoint

Problem:

Error when loading GC method without Blank Run Analysis setpoint

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:489  Product:DSA  All GCs  3.00

Keyword: 78xx Driver

One-line Description:

Chromeleon: Corrected column length will not save after manual method resolution

Problem:

Chromeleon: Corrected column length will not save after manual method resolution

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:490  Product:DSA  ICF (DSA)  3.00

Keyword: 78xx Driver

One-line Description:

Chromeleon: Signals from previous method persist in new method

Problem:

Chromeleon: Signals from previous method persist in new method

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:491  Product:DSA  ICF (DSA)  3.00

Keyword: 78xx Driver

One-line Description:

Chromeleon: ECD Help page cannot be accessed

Problem:

Chromeleon: ECD Help page cannot be accessed

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:492  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: All 8860 signals not displayed or available in report

Problem:

Chromeleon: All 8860 signals not displayed or available in report.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:493  Product:DSA  ICF (DSA)  3.00

Keyword: 78xx Driver

One-line Description:

Chromeleon: Error occurred at several points of operation

Problem:

Chromeleon: Error occurred at several points of operation.

1. Open instrument, select instrument (8890) and choose Create Method

2. Error pops up (see pic)

3. Click OK and proceed setting method parameters

4. Finish method; new instrument method opens. 

5. Error appears again

6. Click OK, check method and save. Method saves

7. Right click on Status window and select Edit Online Method

8. Error appears again

9. Click OK and proceed with method edit

10. No time in this instance did the error impede actual saving method or running a sequence

Expected behavior: Method creation and editing proceed without error

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:494  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: GC6890: Front Injector parameter change in Online method not applied

Problem:

Chromeleon: GC6890: Front Injector parameter change in Online method not applied.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:495  Product:DSA  ICF (DSA)  3.00

Keyword: 78xx Driver

One-line Description:

Chromeleon: GC6890A: Method download fails

Problem:

Chromeleon: GC6890A: Method download fails

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:496  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

8860 on Chromeleon -Unhandled exception in Chromeleon Driver Host (CmDDKHost): AgilentTechnologies.DdkIcfAdapter|Agilent

Problem:

8860 on Chromeleon -Unhandled exception in Chromeleon Driver Host (CmDDKHost): AgilentTechnologies.DdkIcfAdapter|Agilent

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:497  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon does not recognize the tray with Back Injector installed.

Problem:

Connected to a Back inlet only 8890 with a Back Injector and Tray. Was not able to use a vial position on the tray as the vial number in the sequence for Chromeleon.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:498  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: Automatic de-activation of not supported feature blank evaluation cannot be saved

Problem:

Chromeleon: Automatic de-activation of not supported feature blank evaluation cannot be saved

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:499  Product:DSA  ICF (DSA)  3.00

Keyword: 78xx Driver

One-line Description:

Getting multiple Errors in the PC Event Log for AgGC7890 after disconnecting and Reconnecting LAN cable.

Problem:

Getting multiple Errors in the PC Event Log for AgGC7890 after disconnecting and Reconnecting LAN cable.

Temporary Solution:

n/a

Fix Information:

Unable to duplicate the issue.


KPR#:500  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

GC6890: Valid vial ranges are not correct for 6890 in some cases

Problem:

GC6890: Valid vial ranges are not correct for 6890 in some cases

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:501  Product:DSA  ICF (DSA)  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: GC6890A: Error 'The communication FIFO of the DDK driver is corrupted.' after connecting and disconnecting several times.

Problem:

Chromeleon: GC6890A: Error 'The communication FIFO of the DDK driver is corrupted.' after connecting and disconnecting several times.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:502  Product:DSA  All GCs  3.1

Keyword: 78xx Driver

One-line Description:

When click the "Apply the Option" in GC Chromatographic Column Directory under Column Configuration, an exception occur on Openlab 2.4

Problem:

When click the "Apply the Option" in GC Chromatographic Column Directory under Column Configuration, an exception occur on Openlab 2.4.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:503  Product:DSA  ICF (DSA)  3.0.2

Keyword: 78xx Driver

One-line Description:

Upgrade installation from GC B.01.03 to 3.0 SR2 fails to install Ag7697WS.dll (and .pdb)

Problem:

Upgrade installation from GC B.01.03 to 3.0 SR2 fails to install Ag7697WS.dll (and .pdb)

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:504  Product:DSA  All GCs  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: GC6850A: Changing method parameter creates method inconsistent to configuration

Problem:

Chromeleon: GC6850A: Changing method parameter creates method inconsistent to configuration

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:505  Product:DSA  All GCs  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon:  GC6850: Partial Data option causes issues during sequence execution

Problem:

Chromeleon:  GC6850: Partial Data option causes issues during sequence execution

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:506  Product:DSA  All GCs  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: GC6890: Unexpected audit trail entries during endurance run

Problem:

Chromeleon: GC6890: Unexpected audit trail entries during endurance run

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:507  Product:DSA  All GCs  3.0

Keyword: 78xx Driver

One-line Description:

SR1 for the GC3.0 drivers breaks the connection between an 8890 and 7000D transferline temperature reading in MH ACQ 10.

Problem:

SR1 for the GC3.0 drivers breaks the connection between an 8890 and 7000D transferline temperature reading in MH ACQ 10.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:508  Product:DSA  All GCs  3.1

Keyword: 78xx Driver

One-line Description:

Not able to download flow setpoints to 6890 VI inlet.

Problem:

Not able to download flow setpoints to 6890 VI inlet. Set up a method at the 6890. The column flow as set to 3 ml/min and the split ratio to 20 giving a total flow of 66.26 ml/min. Changed the split ratio to 10 and downloaded the method. The total flow did not change.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:509  Product:DSA  7890B  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: Incorrect conversion and missing unit when using pressure setting bar

Problem:

Chromeleon: Incorrect conversion and missing unit when using pressure setting bar

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:510  Product:DSA  All GCs  3.0.2

Keyword: 78xx Driver

One-line Description:

68xx: Chromeleon: System with G4513 won't make 1.1 uL injection

Problem:

68xx: Chromeleon: System with G4513 won't make 1.1 uL injection

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:511  Product:DSA  7890B  3.0

Keyword: 78xx Driver

One-line Description:

Chromeleon: System crash: system slow and then crashing

Problem:

Chromeleon: System crash: system slow and then crashing. Chromeleon was unreactive and then suddenly the instrument connection was lost. After restarting the instrument controller the instrument was online again.

Audit trail showns an OutOfMemory exception.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:512  Product:DSA  7890B  3.0

Keyword: 78xx Driver

One-line Description:

8860 with Headspace/PAL does not consume correct VL license for OpenLAB 2.4

Problem:

8860 with Headspace/PAL does not consume correct VL license for OpenLAB 2.4

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:513  Product:DSA  7890B  3.0

Keyword: 78xx Driver

One-line Description:

B,04.03 SP2 will not connect to an 8890 GC.

Problem:

B,04.03 SP2 will not connect to an 8890 GC.There is no way to configure an 8890 and if you try to connect as a 7890 it will not start the instrument session.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:514  Product:DSA  7890B  3.0.2

Keyword: 78xx Driver

One-line Description:

Chromeleon: 6890GC Oven temperature disabled in method - Run does not start.

Problem:

Chromeleon: 6890GC Oven temperature disabled in method - Run does not start.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:515  Product:DSA  All GCs  3.00

Keyword: 78xx Driver

One-line Description:

3.0 SR2 driver fails SVT on OL CS C.01.09.144

Problem:

3.0 SR2 driver fails SVT on OL CS C.01.09.144

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:516  Product:DSA  All GCs  3.00

Keyword: 78xx Driver

One-line Description:

OLCS C.01.09 with Driver 3.0.650 will not launch any instrument.

Problem:

OLCS C.01.09 with Driver 3.0.650 will not launch any instrument.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:517  Product:DSA  7890B  3.0.2

Keyword: 78xx Driver

One-line Description:

Chromeleon: SVT Chameleon 2.1.30 Fails

Problem:

Chromeleon: SVT Chameleon 2.1.30 Fails

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:518  Product:DSA  All GCs  3.0

Keyword: 78xx Driver

One-line Description:

Function_build number not display in SVT report in EzChrom and OpenLab CDS 2.4

Problem:

Function_build number not display in SVT report in EzChrom and OpenLab CDS 2.4

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:519  Product:DSA  All GCs  3.00

Keyword: 78xx Driver

One-line Description:

8890: the MS transfer line temp displayed is 0 C but its actual temp is 250 C.

Problem:

8890: the MS transfer line temp displayed is 0 C but its actual temp is 250 C.

Temporary Solution:

n/a

Fix Information:

Resolved in version 3.0.652


KPR#:599  Product:DSA  7890B  B.01.04

Keyword: 78xx Driver

One-line Description:

Instrument Status shows HS Comm as 'OFF' when it is not.

Problem:

Instrument Status shows HS Comm as 'OFF' when it is not.

Temporary Solution:

n/a

Fix Information:

Resolved in GC Drivers Version 3.2


KPR#:1196  Product:DSA  7890A  3.01

Keyword: 78xx Driver

One-line Description:

7890A: Shutdown events are not published to the software GC Status screen

Problem:

The thermal inlet shutdown is not reported in the software “Agilent GC Status” screen.  

Only the first event, which is 'not ready', is reported to the Agilent GC Status screen.  The second event, which is the shutdown, is not displayed.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.01.  7890A requires a minimum Firmware revision of A.01.12.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#:345674  Product:    A.04.09

Keyword: 88xx GC

One-line Description:

8860 GC System: there is no core license available

Problem:

Two scenarios:
Scenario #1:
1. In Open LAB EZChrom A.04.09 with original driver 3.0.SR1, add a valid license.
2. Update GC driver to 3.0.SR2 (3.0.652).
3. Remove and add the same license, and the status of license is valid.
4. Configure and launch 8860, an error message pops up 'there is no core license available.

If add the license and then update GC driver to 3.0.SR2, reinstall of the OpenLab CDS EZChrom Edition or reimage the PC is required to solve the problem.

Scenario #2:
Another issue reported can be reproduced by steps below.
1. Install GC 3.0.620 SR1 with corresponding licenses. Configure instrument and launch.  
2. Now upgrade to GC 3.0.652 SR2 and install corresponding licenses.
3. Launch instrument created in step #1. Instrument cannot be  launched due to error  ‘no core licenses’.
[Solution/Workaround]
Reconfigure the same instrument and then it can be successfully launched.

Temporary Solution:

If you update GC driver from 3.0.SR1 to 3.0.SR2 before adding the license, 8860 GC can be launched successfully, without any error message.

Fix Information:

n/a


KPR#:138296  Product:Hitachi LC  M8227AA/BA Hitachi LC  HitachiChromaster_2.1.10

Keyword: Acquisition

One-line Description:

 On-line signal for Hitachi Chromaster is only displayed during acquisition 

Problem:

The behavior of the Hitachi Chromaster differs from the other vendors' instruments: 
- The on-line signal for Hitachi Chromaster is only displayed during acquisition 
- The vertical red marker is not displayed at the start of acquisition in the on-line signal window.
- The collected signal is only visible during the time of acquisition and disappears from the view upon completion of a run. However, it can be viewed from the Data Analysis after the acquisition.

Temporary Solution:

Use Data Analysis to view the chromatogram

Fix Information:

There is no fix


KPR#:146968  Product:Thermo Fisher LC  M8236AA/BA Thermo Fisher Surveyor/Accela  ThermoLC 4.1.2.1

Keyword: Acquisition

One-line Description:

Certain auxiliary traces of Accela pump are not supported

Problem:

The auxiliary traces for flow and solvent proportions of the Accela pump are not supported in OpenLAB CDS Thermo Fisher Accela Surveyor Driver Rev. 4.1.2.0

Temporary Solution:

N/A

Fix Information:

No fix


KPR#:147007  Product:Shimadzu LC  M8240AA/BA Shimadzu LC  ShimadzuLC 1.41

Keyword: Acquisition

One-line Description:

The default "Use Method" setting is not working as expected.

Problem:

If using the default "Use Method" setting, the sampler will perform a blank run, because the method contains no defined injection volume.

Temporary Solution:

You must define an injection volume the have an integer value. This limitation is written in the Shimadzu Release Note. 

Fix Information:

Fixed with Shimadzu LC 1.5


KPR#:147010  Product:Thermo Fisher GC  M8241AA Thermo Fisher GC  2015R2

Keyword: Acquisition

One-line Description:

Mismatched channel/peak data for Thermo Trace 1310GC

Problem:

When peaks are defined for a specific signal, they appear in the peak table but not in the annotation or the report.

Temporary Solution:

n/a

Fix Information:

This problem as fixed by Thermo via a release of an updated Thermo GC driver (2016 R1).


KPR#:147797  Product:    A.04.07 SR2 HF02

Keyword: Acquisition

One-line Description:

LC does not abort sequence after pressure exceeds upper limit

Problem:

Some newer LC modules do not abort the sequence after the pressure exceeds the upper limit

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB CDS EZChrom Edition A.04.07 SR2 Update 7


KPR#:200819  Product:    A.04.08 Update2

Keyword: Acquisition

One-line Description:

Delay observed when comparing the Run Time and Analysis Time of the data.

Problem:

Comparing the Run Time and Analysis Time from the method report and from the data -> properties -> general tab, the run time does not match the actual runtime of the data. There is always a delay of 10 seconds or more.

Temporary Solution:

N/A

Fix Information:

N/A


KPR#:220292  Product:Shimadzu GC  M8232AA/BA Shimadzu GC  ShimadzuLCGCv2.02

Keyword: Acquisition

One-line Description:

Shimadzy 2010 GC and HS-20 synchronization issue

Problem:

When Headspace HS-20 is in pre-run, Shimadzu 2010 GC module has to be ready but GC also goes to a running state.

Temporary Solution:

None

Fix Information:

This is addressed by the Shimadzu GC v2.03 driver.


KPR#:316749  Product:    A.04.08 Update2

Keyword: Acquisition

One-line Description:

Delay observed when comparing the Run Time and Analysis Time of the data.

Problem:

Comparing the Run Time and Analysis Time from the method report and from the data -> properties -> general tab, the run time does not match the actual runtime of the data. There is always a delay of 10 seconds or more.

Temporary Solution:

N/A

Fix Information:

N/A


KPR#:356008  Product:Hitachi LC  M8227AA/BA Hitachi LC  HitachiLC_EZC_February18Ed

Keyword: Acquisition

One-line Description:

Hitachi LCE and CM: Cannot Use Submit Button in Run Queue to Submit Runs

Problem:

Hitachi (LaChromElite and Chromaster) LC does not start a single run from a Run Queue window via "Submit" button in OpenLab EZChrom edition.

Temporary Solution:

Submitting via "Advanced" option in the Run queue works. 

Fix Information:

This was fixed with EZChrom A.04.09 Update 06.


KPR#:380636  Product:Bruker GC  M8237AA/BA Bruker GC  BrukerGC_A.02.03,BrukerGC_A.02.04,BrukerGC_A.02.04_Upd1

Keyword: Acquisition

One-line Description:

Bruker GC: Changing the current time of the computer makes the driver go offline sporadically

Problem:

Observed in OpenLab CDS 2.4, Bruker GC A.02.03

Changing the current time of the computer makes the driver going offline sporadically. That results in connection loss and abortion of the sequence.

Temporary Solution:

N/A

Fix Information:

Fixed in 2.5 version of the driver: the information on how to avoid the issue was added in the "Known issues" chapter of the User's Guides.


KPR#:401753  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.0

Keyword: Acquisition

One-line Description:

No error message for missing vial

Problem:

As expected, when “Sense missing vial” option is  not checked,  the run/sequence will not abort in case of a missing vial, but no error message will inform the user about the issue.
The instrument performs an acquisition with injecting air and a dummy acquisition data is generated.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 1.1 of the driver.


KPR#:536644  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CM_2.3,Hitachi_LCE_2.5

Keyword: Acquisition

One-line Description:

Hitachi Chromaster LC: No Injection Happens if "No injection/Instrument blank" Used as the Injection Source

Problem:

If a Hitachi Chromaster LC Instrument is used without Autosampler, and "No injection/Instrument blank" used as the injection source,
the instrument is waiting for trigger. No injection happens.

Temporary Solution:

N/A

Fix Information:

Fixed in driver v2.6


KPR#:695883  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.1

Keyword: Acquisition

One-line Description:

Hitachi ChromasterUltra RS: Displayed Runtime Doubles for Detectors Online Plot During a Run

Problem:

In case the Hitachi ChromasterUltra RS is configured with 2 detectors of the same type, the Online plots of the detectors are displayed incorrectly by doubling the runtime. The data results look correct in the Data Analysis.

Temporary Solution:

Ask your Hitachi support to get the IFC Firmware 8898214-04 installed.
This is documented with the driver version 1.2

Fix Information:

Hitachi fixed this defect with a new version of the IFC Firmware. Ask your Hitachi support to get the IFC Firmware 8898214-04 installed.


KPR#:949857  Product:    A.04.10 Update 03

Keyword: Acquisition

One-line Description:

Data is not being saved when executing a sequence using the PAL 3 RTC with Tool changes

Problem:

The problem is when a tool change is being made in a sequence, and numeric values are used for injection volume for the different tools, the sequence will not produce the data. This also will cause the instrument to hang, because the injection volume ranges do not match.  

Temporary Solution:

n/a

Fix Information:

Use method  must be used for the Injection Volume for each of the methods based on a different tool.


KPR#:1192808  Product:Shimadzu LC    ShimadzuLCGC v3.50

Keyword: Acquisition

One-line Description:

Shimadzu LC: Manual Injection Starts Without Waiting for Trigger

Problem:

With Open Lab CDS Ver. 2.8 and Shimadzu LC Driver Ver. 2.20, the run starts without waiting for an external signal. The same issue does not happen with previous versions of OpenLab CDS.

This happens with Shimadzu LC20 and LC40 series.

Temporary Solution:

Avoid using manual injection in OpenLab CDS 2.8 until a fix is issued.

Fix Information:

A point patch 2 has been issued by Shimadzu for this issue in OpenLab CDS 2.8. It is distributed upon request to 3PIC (Third Party Instrument Control Team) and is planned to be included in the next Shimadzu LC driver (after version 2.20). 


KPR#:1382228  Product:CTC PAL3 LC  M8223BA CTC PAL3 for LC  CTC_LC_PAL3_1.10.4

Keyword: Acquisition

One-line Description:

CTC PAL: Activity Log: “Could not read barcode” message does not identify the vial/tray position

Problem:

When a barcode read fails, the Activity Log shows a generic message: “Barcode could not be read".
The entry does not include the vial context (sampler, tray, position, sequence line / sample name etc.). 

Temporary Solution:

None.

Fix Information:

To be considered in the future revision of the driver.


KPR#:36  Product:OpenLAB Data Store  OpenLAB Data Store  A.02.01

Keyword: Administration

One-line Description:

Windows Domain tab is grayed out in OpenLAB Server Utility

Problem:

When the authentication provider is set to domain authentication, the user account used to browse the active directory is stored in the OpenLAB Shared Services database and can be modified or updated using the OpenLAB Server Utility.
The OpenLAB Server Utility will then update the databse with the new information. 
When starting the OpenLAB Server Utility as a domain user which is not a database administrator, the information in the windows domain tab will be greyed out because the tool can't connect to the database.

Temporary Solution:

Login to windows as an user account which is granted database administration privileges

Fix Information:

n/a


KPR#:39  Product:OpenLAB Data Store  OpenLAB Data Store  A.02.01

Keyword: Administration

One-line Description:

Cannot stop Shared Serviceshost service on computer after changing the Domain Account password

Problem:

After changing the Domain Account password on the OpenLAB Server Utility, you may occasionally get an error message, “Cannot stop Shared Serviceshost service on computer”.

Temporary Solution:

As a temporary solution, restart the Agilent OpenLAB Shared Services service or reboot the shared services server.

Fix Information:

n/a


KPR#:179143  Product:    A.04.08 Update2

Keyword: Agilent LC Driver

One-line Description:

Error in RCNet_RCPluginProvider::SetConfigSettings...-Exception

Problem:

After reconfiguration of an LC module, in some cases an error message can occur once in the online session when opening the instrument status window “Error in RCNet_RCPluginProvider::SetConfigSettings : Failed to instantiate the plug in for module <module type> Exception has been thrown by the target of an invocation”. This error can appear once per saved instrument configuration and will not reappear with this instrument until reconfiguration. 

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:400080  Product:    A.04.08 Update2

Keyword: Agilent LC Driver

One-line Description:

Warning message “The installed OpenLAB software is incomplete” or “Repair is not possible as no complete

Problem:

If OpenLAB CDS EZChrom Edition was installed by User A (Windows Administrator) it is not possible to run “OpenLAB
Additional Software and Drivers" or "Repair OpenLAB CDS” with User B (Windows Administrator). An error “The installed
OpenLAB software is incomplete” appears.

Temporary Solution:

n/a

Fix Information:

Install the “Agilent OpenLAB CDS Control Panel Help” manually with User B. The installable .msi is located on installation
media “Disk1\OLSS\<SupportedOLSSFolder>\Agilent OpenLAB CDS Control Panel Help.msi”. Double-click on this file to
install “Agilent OpenLAB CDS Control Panel Help”. Afterwards “OpenLAB Additional Software and Drivers" or "Repair
OpenLAB CDS” will work as expected.
Note this will also add a new Program into Windows Programs and features. It is ok to leave this in place and not remove it.


KPR#:146969  Product:Thermo Fisher LC  M8236AA/BA Thermo Fisher Surveyor/Accela  ThermoLC 4.1.2.1

Keyword: Audit Trail

One-line Description:

Timetable changes audit traces on general level only

Problem:

Performing time table changes in a method, the audit trail reflects a general statement only that the timetable has changed.
Currently there is no granularity listing each changed value of the timetable in the audit trail, as the system does with other method changes.

Temporary Solution:

Using OpenLAB CDS EZChrom Edition, the user can set the Audit Trail to be prompt for a reason at every change and prompt for a reason saving the method.  This will allow the user to type in the reason, in this case the timetable modification.

Fix Information:

N/A


KPR#:146973  Product:Waters LC  M8505AA/BA Waters ACQUITY/Alliance  Alliance 6.1.2.1

Keyword: Audit Trail

One-line Description:

PDA 2998/ UV 2489 : Audit trail messages do not contain unit information when values are reported

Problem:

The audit trail messages for the PDA 2998/ UV 2489 do not contain any unit information. For example, missing units are observed for run time such as "mins" , or data frequency such as "Hz". Other units in the audit trail relating to the detectors are similarly affected. 
This applies to the driver version for OpenLAB CDS EZChrom Edition and ChemStation Edition.

Temporary Solution:

n/a

Fix Information:

Resolved in driver version 6.1.3.


KPR#:191802  Product:    A.04.07 SR2 Update 8 Point Patch 1

Keyword: Audit Trail

One-line Description:

Error message for e-signing twice can be unclear on Citrix Clients

Problem:

In a Client Server system with Citrix Clients, if a user tries to sign a file the second time, it may show the message: "Apply/Revoke Signature failed for some of the data files. Please look at the audit trail log for details." instead of the expected message: "Failed to apply electronic signature. Datafile is already signed by <user><level>"

Temporary Solution:

Interpret the error message "Apply/Revoke Signature failed for some of the data files. Please look at the audit trail log for details." as it could mean in some cases: "Failed to apply electronic signature. Datafile is already signed by <user><level>"

Fix Information:

A fix is considered for a later revision.


KPR#:210882  Product:    A.04.07 SR2 Update 08

Keyword: Audit Trail

One-line Description:

Level changes in Peak Table are not captured in the Method Audit Trail

Problem:

Level changes in the Peak Table are not captured in the Audit Trail log.Make sure all the audit trail were enabled 
Load any method then Add one peak to the "peak/group" table,  
Input name and level 1, level 2 level 3 , level 4 and level 5
Save method

Inspect the Method Audit Trail log. Level changes are not present.

Temporary Solution:

Work around: 
Clear the first row in the peak table. 
After the row is cleared, method should be saved. 
Method should be saved after clearing the peak and then further changes should be done. A change to the name of the peak, changes to additional values will be written to the Method Audit Trail.

If the values for levels are same and not changed, they are not reported in the Method Audit Trail. 

Clear the first row and add new values. A change to existing saved values is detected as a difference. Subsequent changes are captured to the Method Audit Trail. Then, changes to existing values are detected and identified as a difference to be written to the Method Audit Trail.  

Fix Information:

Will not fix. Workaround is available.


KPR#:251903  Product:    A.04.07 SR2 HF05

Keyword: Audit Trail

One-line Description:

Method audit trail will be lost once loading method in different instruments

Problem:

An instrument method created on one instrument is opened in another dissimilar instrument and loses the audit trail history, when saved. 

Temporary Solution:

n/a

Fix Information:

Every module/device determines the unique instrument composition. Driver version and firmware may impact the method compatibility among instruments, even on instruments with the same apparent composition. Any detected instrument difference could trigger a instrument method resolve. A resolve does not guarantee the same instrument control parameters and acquisition parameters. Methods are instrument specific and a resolved method should have the parameters manually verified.

Automatic method resolution happens when the driver can automatically resolve. User interactive manual resolution is an option, but the driver will automatically attempt to resolve parameters. A pop-up message box appears indicating resolved settings required. Always review and verify the instrument parameters and save the method. This designed workflow results in a new instrument method audit trail and new method name.

The industry standard design of instrument specific method will not be changed.


KPR#:292456  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_OLCDS_V1.0.20.1

Keyword: Audit Trail

One-line Description:

Units for carrier A timetable

Problem:

The UI of the carrier A timetable is always displaying the unit "psi".
Changes in the Rate and Setpoint columns are creating Audit trail entries which always use the unit "psi".
This is only correct if the Control is set to "Pressure" and the system is currently configured to use "psi" as pressure unit.

Temporary Solution:

n/a

Fix Information:

Fixed in PE Clarus GC Control 2.0


KPR#:297766  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_OLCDS_V1.0.21.1

Keyword: Audit Trail

One-line Description:

Method Audit Trail Incorrectly References Detector A Gas 2 as Gas 3  

Problem:

Method report only contains a reference to Detector A Gas 1 and 2. However, method audit trail incorrectly references Detector A Gas 2 as Gas 3.

Temporary Solution:

N/A

Fix Information:

Fixed in PE Clarus GC Control 2.0


KPR#:589099  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.1

Keyword: Audit Trail

One-line Description:

Hitachi Chromaster Ultra: No System Activity Log Entries Upon Vial Settings Change

Problem:

A change of the vial settings ("Sense missing vial" and "Abort sequence by missing vial") under the direct control options for the autosampler is not recorded in the System Activity Log.

Temporary Solution:

n/a

Fix Information:

This is fixed in revision 1.2 of the driver.


KPR#:256374  Product:    A.04.09 Update1

Keyword: Automation

One-line Description:

“Unable to Create UILess IR control” error message in Instrument activity after launching instrument from Automation tool.

Problem:

Error message: “Unable to Create UILess IR control” in Instrument activity after launching instrument from EZChrom Automation Tool Kit.

Temporary Solution:

n/a

Fix Information:

A fix is provided in OpenLAB CDS EZChrom Edition A.04.09 Update 02


KPR#:338  Product:DSA  7890B  A.03.02

Keyword: Autosampler

One-line Description:

EZChrom: PAL XT module not available after installation

Problem:

1. Uninstalled existing PALxt driver. Driver was removed from Programs and Features

2. Installed PALxt B.01.07 using Gold Master disk. (dated 10/1/15)

3.  Driver appeared in Programs and Features.

4. Restarted because it is EZChrom.

5. Opened EZChrom Control Panel and selected to configure.

6. PAL XT (CTC Sampler) was not present in modules window. (see pic)

7. PAL3 and HS were already present.

Temporary Solution:

n/a

Fix Information:

Fixed in B.01.01


KPR#:146999  Product:Thermo Fisher LC  M8236AA/BA Thermo Fisher Surveyor/Accela  ThermoLC 4.1.4

Keyword: CDS

One-line Description:

EZChrom Edition only: Increased memory consumption after a sequence.

Problem:

OpenLAB CDS EZChrom Edition only!
After a sequence is finished the memory consumption of a specific process increases.
This is observed for Accela and Surveyor and was already present in previous versions of the driver.

Temporary Solution:

N/A

Fix Information:

Will not fix


KPR#:516793  Product:PerkinElmer LC  M8229BA PerkinElmer LC  PE_LC300_2.0

Keyword: CDS

One-line Description:

PE LC300: Incorrect Status of Non-Detector Modules While Running Shutdown Method

Problem:

While running a Shutdown method, the non-detector modules (pump, column and oven) incorrectly display ‘Idle’ as the status in the Instrument Status dashboard  while detectors show "Running".

Temporary Solution:

None.

Fix Information:

Fixed in driver version 3.1.


KPR#:517638  Product:Bruker GC  M8237AA/BA Bruker GC  BrukerGC_A.02.04_Upd1

Keyword: CDS

One-line Description:

Bruker GC456: Stepwise Online Signal

Problem:

The online plot does not look smooth and has steps. However, the signal is not affected when looked at from the Data Analysis.

Temporary Solution:

n/a

Fix Information:

Fixed in 2.5 version of the driver.


KPR#:548087  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CM_2.5,Hitachi_LCE_2.4,Hitachi_CM_2.6,Hitachi_LCE_2.5,Hitachi_LCE_2.6

Keyword: CDS

One-line Description:

Hitachi Chromaster and LaChrom Elite LC: Differences in DAD channel naming in Data Analysis

Problem:

Signal name in Acquisition and Data Analysis is different.

Temporary Solution:

N/A

Fix Information:

Hitachi improved the signal names to match the naming in OpenLab Data Analysis in driver revision 2.7.


KPR#:554454  Product:  M8223BA CTC PAL3 for LC  CTC_LC_PAL3_1.3.0_OpenLabCDS

Keyword: CDS

One-line Description:

CTC PAL3 for LC Cannot be Launched with OpenLab CDS 2.6

Problem:

Driver version 1.3.0 for CTC PAL3 for LC cannot be launched with OpenLab CDS 2.6. "Failure while initializing modules" exception is thrown upon launching the instrument.

Temporary Solution:

None.

Fix Information:

This is addressed by the CTC PAL3 driver 1.7.4.0.


KPR#:1378523  Product:Shimadzu LC  M8240AA/BA Shimadzu LC  ShimadzuLC_2.20_PP1

Keyword: CDS

One-line Description:

Shimadzu LC: A Task with Shutdown Functions Not Executed

Problem:

With CDS 3.0, a Task with shutdown functions submitted at the end of a sequence is not executed. After the sequence is completed, the pump, DAD and UV-lamps still stay turned on, so the actions defined in the task are not applied.

Temporary Solution:

Do not use a Task feature to perform shutdown of the system at the end of a sequence. Instead, append a separate line to the end of the sequence with a method that turns the modules off.

Fix Information:

This is to be addressed in the newer revision of Shimadzu LC driver.


KPR#:46  Product:OpenLAB Intelligent Reporting  OpenLAB Intelligent Reporting  A.01.04

Keyword: Client Report Rendering

One-line Description:

Intelligent report of English openlab chemstation can not show chinese name in PDF report

Problem:

Intelligent report of English openlab chemstation can not show chinese name in PDF report,but priview and export report as word report is ok ,only PDF report use "?" instead of chinese characters.

Temporary Solution:

n/a

Fix Information:

Customer needs to install the SimSun font on the English windows system


KPR#:628  Product:RTE  RTE  A.02.04

Keyword: Client Report Rendering

One-line Description:

Large matrix reports may produce additional empty pages

Problem:

Large matrix reports may produce additional empty pages.
The issue is in the Microsoft Report Viewer control and cannot be fixed in the current product.

Temporary Solution:

Matrix reports easily exceed A4 or Letter sized paper formats and printing to a larger PDF paper format could be used as alternative.

Fix Information:

There is currently no fix planned.


KPR#:653  Product:OpenLAB ECM Intelligent Reporter  G4635AA  A.02.01 SP1 HF01

Keyword: Client Report Rendering

One-line Description:

Calculating the standard deviation of identical numbers may return NaN

Problem:

When calculating the standard deviation or relative standard deviation of multiple identical numbers, the calculation may return "NaN" rather than 0.

The problem is caused by the small delta derived from the rounded result of the average calculation and the individual values. Summing up these small numbers may result in a floating point exception.

Temporary Solution:

Do not apply standard deviation calculations to series of identical  values.

Fix Information:

There is currently no fix planned.


KPR#:657  Product:OpenLAB ECM Intelligent Reporter  G4635AA  A.02.04

Keyword: Client Report Rendering

One-line Description:

Changes in report templates are not immediately shown on report printouts

Problem:

Changes in report templates are not immediately  shown on the printout if the user does not preview the report and save the template first.

Temporary Solution:

Always preview the report printout after applying changes to the template.
Or save and reload the template from disk, if a preview should be avoided.

Fix Information:

There is no fix planned.


KPR#:659  Product:OpenLAB ECM Intelligent Reporter  G4635AA  A.02.04

Keyword: Client Report Rendering

One-line Description:

Rendering error when using a template containing a large result matrix

Problem:

A template having a matrix inside a list may show a rendering error - The value 'x' is invalid. Valid values are between '0' and 'x'
This error occurs when there is large data used for rendering and due to this the matrix data spans across multiple pages.

Temporary Solution:

For large result matrices it is recommended to print the results to PDF files, rather than paper.
The paper format should adjusted to display large matrices on a single page.

Fix Information:

There is currently no fix planned.


KPR#:22100  Product:Waters LC  M8239AA Waters Alliance HPLC Driver  6.1.2.1

Keyword: Configuration

One-line Description:

Auto-configuration for instruments is not supported

Problem:

The Auto-Configuration function for configuring instruments for both OpenLAB CDS Chemstation and EZChrom Editions is not supported and therefore manual configuration is required. For more details , consult the Waters Alliance HPLC AddOn Installation Upgrade Guide.

Temporary Solution:

User needs to manually configure the instrument.

Fix Information:

n/a


KPR#:146924  Product:Waters LC  M8505AA/BA Waters ACQUITY/Alliance  ACQUITY_A.01.01

Keyword: Configuration

One-line Description:

Starting Acquity instrument for first time after an AIC reboot has errors

Problem:

After reboot of the dedicated Acquity "AIC," when the Acquity instrument is started, there are numerous errors generated in pop up panels. Click okay through the errors, and the instrument user interface closes. Starting the instrument the second time and it works without errors. If the AIC is rebooted, the first startup of the instrument generates the errors again.

The errors also occur if the Waters Acquity driver software is not stopped after closing a previous instrument session and the instrument in OpenLAB CDS Ezchrom is subsequently launched.

Temporary Solution:

n/a

Fix Information:

Fixed with Agilent penLAB CDS Waters Acquity Driver A.01.02


KPR#:146964  Product:Thermo Fisher LC  M8236AA/BA Thermo Fisher Surveyor/Accela  ThermoLC 4.1.2.1

Keyword: Configuration

One-line Description:

Some Autosampler parameters are not uploaded e.g. loop size.

Problem:

Some Autosampler configuration parameters are not uploaded manually, e.g. loop size.

Temporary Solution:

The user needs to specify missing configuration parameters, e.g. the loop size, manually.

Fix Information:

n/a


KPR#:146966  Product:Thermo Fisher LC  M8236AA/BA Thermo Fisher Surveyor/Accela  ThermoLC 4.1.2.1

Keyword: Configuration

One-line Description:

Manual tray type configuration required

Problem:

The tray type display is incorrect under the following conditions.
If user selects '1 or 2 ml deep wellplates + Short microwell carrier' or '96 well microplate + Short microwell carrier' or '96 well PCR plate + Cooling adapter + Short microwell carrier' and downloads method, selected tray type gets downloaded. However, when user uploads configuration, tray type entry is displayed as '1 or 2 ml deep wellplates + Short microwell carrier'

Temporary Solution:

The user needs to specify the correct tray type manually.

Fix Information:

n/a


KPR#:146970  Product:Waters LC  M8505AA/BA Waters ACQUITY/Alliance  Alliance 6.1.2.1

Keyword: Configuration

One-line Description:

Auto-configuration for instruments is not supported

Problem:

The Auto-Configuration function for configuring instruments for both OpenLAB CDS Chemstation and EZChrom Editions is not supported and therefore manual configuration is required. For more details , consult the Waters Alliance HPLC AddOn Installation Upgrade Guide.

Temporary Solution:

User needs to manually configure the instrument.

Fix Information:

n/a


KPR#:146988  Product:Shimadzu LC  M8240AA/BA Shimadzu LC  Shimadzu 1.30 SP1

Keyword: Configuration

One-line Description:

It is not possible to configure a Shimadzu SPD-M20A detector in the same system configuration as a Shimadzu Nexera integrated system.

Problem:

When the instrument configuration is accessed and you select the Nexera instrument modules, there is no selection/module available for the SPD-M20A in the list of "Available Modules" field.

Temporary Solution:

Windows Registry needs to be modified as a workaround. For assistance please contact the 3rd party driver support team of Agilent Technologies under the Emailaddress: techsupport.3rd-party-ic@agilent.com

Fix Information:

There is no fix.


KPR#:147000  Product:Waters LC  M8505AA/BA Waters ACQUITY/Alliance  Alliance 6.1.2.1

Keyword: Configuration

One-line Description:

200uL Sample loop not configurable in Alliance instrument configuration

Problem:

When configuring the 200UL sample loop size in the Alliance instrument configuration within OpenLAB CDS C.01.07 SR1, no drop-down selection for this loop size is available.
Other CDS versions and Editions are impacted similarly.

Temporary Solution:

n/a

Fix Information:

Fixed in Alliance driver 6.2.1.


KPR#:147014  Product:Bruker GC  M8237AA/BA Bruker GC  BrukerGC_A.02.01,BrukerGC_A.02.02

Keyword: Configuration

One-line Description:

For the GC column properties, there are missing units for column length and internal diameter.

Problem:

The column properties tab containing the entries for column length and internal column diameter are displayed without any indication of what units are being used.

Temporary Solution:

This is currently documented in the Bruker GC Install Guide as a known limitation.
The current expectation is column id value should be entered in microns and column length in meters.

Fix Information:

This is fixed by the latest rev. of the driver.


KPR#:285964  Product:Hitachi LC  M8227AA/BA Hitachi LC  HitachiLaChromEliteUltra_4.84

Keyword: Configuration

One-line Description:

Hitachi LaChrom Elite auto configuration fails on EZChrom with Win 10

Problem:

When configuring a Hitachi LaChrom Elite instrument on OpenLab EZChrom Edition 04.09 system with Windows 10, the Auto Configuration functionality may lead to an error.


Temporary Solution:

n/a

Fix Information:

This is fixed in the Jun-2019 version of the drivers.


KPR#:288817  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_V1.3.0.16

Keyword: Configuration

One-line Description:

Instrument status shows temperature value for non-configured injector

Problem:

In the Oven/Inlets tab of the Instrument Status window, injector B shows up the temperature even though it is not configured.

Temporary Solution:

n/a

Fix Information:

Fixed in PE Clarus GC Control 2.0


KPR#:397013  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.0

Keyword: Configuration

One-line Description:

Two different pressure units displayed in data analysis

Problem:

Sometimes the pressure units displayed in Data Analysis are different than in Acquisition (bar vs MPa). 

Temporary Solution:

n/a

Fix Information:

Fixed in revision 1.1 of the driver.


KPR#:456679  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CM_2.4

Keyword: Configuration

One-line Description:

Hitachi LCE and Chromaster LC: Comma as a Digital Separator Limitations

Problem:

In OpenLab CDS, when comma is used as a decimal separator (instead of the period), the numbers come out incorrectly when setting up the parameters in the Instrument Setup window. E.g. 0.2 ml/min might show up as 2 ml/min. Follow the section below that specifies the conditions under which comma is supported as a decimal separator.

Temporary Solution:

N/A

Fix Information:

The comma decimal separator is supported under specific conditions only:
•	Regional Setting other than English or Japanese
•	Decimal symbol: Comma (“,”)
•	Digit Grouping Symbol: Period (“.”)
•	List Separator: Semicolon (“;”)


KPR#:486957  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PELegacy_EZC_RQ

Keyword: Configuration

One-line Description:

PE AutoSystem XL GC: cannot add VICI valves

Problem:

When configuring an instrument, one cannot add VICI sampling valves.

Temporary Solution:

Use a workaround to add VICI valves:
An AutoSystem XL GC Registry Key has to be manually edited:
In Regedit find the registers corresponding to the AutoSystem XL GC and 
under the CategoryType change the original value 1 to 11.
After that one should be able to add VICI valves to the configuration.

Fix Information:

There is no fix for this.


KPR#:551474  Product:PerkinElmer GC  M8242BA PerkinElmer TurboMatrix HS  PETurboMatrixHS_V1.1

Keyword: Configuration

One-line Description:

PE TurboMatrix HS with Agilent GC Cannot be Configured More Than Once

Problem:

The first time configuration of Agilent GC with PE TurboMatrix always succeeds but the subsequent reconfiguration of the instrument sometimes fails to open the configuration dialog -- a busy cursor is shown while doing it. The OpenLab Control Panel does not react any more and has to be closed with Windows Task Manager.

Temporary Solution:

The issue doesn't happen if the TurboMatrix HS module is configured before the Agilent GC module in the instrument configuration window in OpenLab Control Panel. Configure the PE TurboMatrix HS module first and Agilent GC last.

Fix Information:

This issue was addressed with OpenLab CDS 2.6 Update 11 and in OpenLab CDS 2.7.


KPR#:715153  Product:PerkinElmer LC  M8229BA PerkinElmer LC  PEFlexarLC_1.0-RQ1

Keyword: Configuration

One-line Description:

PE Flexar LC: Configuration of Flexar Column Oven with Peltier Fails

Problem:

Peltier Oven cannot be configured and controlled by OpenLab CDS. The User's guide, the Release Notes from PE state it is supported.

Temporary Solution:

Avoid using Peltier oven. Use Heat only oven.

Fix Information:

The Peltier oven is not supported any longer.


KPR#:1216158  Product:CTC PAL3 LC  M8223BA CTC PAL3 for LC  CTC_LC_PAL3_1.7.4_OpenLabCDS,CTC_LC_PAL3_1.9.6_OpenLabCDS

Keyword: Configuration

One-line Description:

CTC PAL: Imported/removed scripts are not reflected in Method editor unless acquisition is relaunched

Problem:

In OpenLab Acquisition with a PAL3 sampler, when importing or removing scripts in the CTC Script editor dialogue box, these changes are not visible in the Method editor unless Acquisition window is relaunched after that.
This is working as designed.

Temporary Solution:

N/A

Fix Information:

The workaround is to be documented in the User's guide.


KPR#:732721  Product:PerkinElmer GC  M8242BA PerkinElmer TurboMatrix HS  PETurboMatrixHS_V1.1

Keyword: Configuration,Instrument Control

One-line Description:

PE TurboMatrix HS: High Pressure Injection Not Usable Without Vial Pressure Sensor

Problem:

The high pressure injection method setting is not usable when the vial leak sensor is not installed.

Temporary Solution:

None

Fix Information:

Fixed in M8242BA V1.1 point patch 1.


KPR#:800415  Product:PerkinElmer GC  M8242BA PerkinElmer TurboMatrix HS  PETurboMatrixHS_V1.1

Keyword: Configuration,Instrument Control

One-line Description:

PE TurboMatrix HS: High Pressure Injection Not Usable Without Vial Pressure Sensor

Problem:

The high pressure injection method setting is not usable when the vial leak sensor is not installed.

Temporary Solution:

None

Fix Information:

Fixed in M8242BA V1.1 point patch 1.


KPR#:117721  Product:OpenLAB  Shared Services  A.02.02

Keyword: Control Panel

One-line Description:

Exception when login in OpenLab Control Panel. ECM does not display all the subfolders when browsing for Project Path during project creation; exception displayed

Problem:

When the user creates projects, the standard procedure is to browse to the folder Location when selecting the Project Path. However, there are more than 100 sub-folders, it is not possible to set the project path correctly, because when browsing, the ECM window displays only the first 100 folders.
The following exception is displayed: "Agilent.OpenLab.SharedServices.SharedServicesInternalException: The incoming request has too many parameters. The server supports a maximum of 2100 parameters..."
An exception is also displayed when logging in OpenLab Control Panel: "Agilent.OpenLab.SharedServices.SharedServicesInternalException: The incoming request has too many parameters..."
 

Temporary Solution:

n/a

Fix Information:

The fix is OpenLAB Shared Services (OLSS) 2.1 Hotfix 4 and is distributed with the CDS.

OpenLAB Shared Services 2.1 Hotfix4 is included with C.01.07 SR4 (Disk1) and is installed. Resolved in C.01.08.


 


KPR#:1995  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.06

Keyword: Custom Reports

One-line Description:

FIltering syntax has changed

Problem:

In OpenLAB CDS A.02.01 expressions can be used for filtering. As a result of this new feature filtering for values has to use expressions as well. If the values are entered as plain numbers, the report will print an error:

An error occured during local report processing.
An error has occured during report processing.
The processing of FilterExpression for the table 'OLIR_Table_XX' cannot be performed. Cannot compare data of types System.Int32 and System.String. Please check the data type returned by the FilterExpression.

Temporary Solution:

The correct syntax for A.01.05 and older for filtering peaks was:
=Peak_RetentionTime     >    1
when peaks with a retention time above one minute should be displayed.

In A.02.01and higher the syntax needs to be:
=Peak_RetentionTime     >    =1

Fix Information:

Existing report templates written with the old syntax will show up correctly in new software. 
No fix required.


KPR#:149830  Product:LC and CE Drivers  RC 1200.NET  

Keyword: DAD/MWD Driver

One-line Description:

High load on CAN may cause "Spectra runbuffer overflow" entries and a loss of spectra

Problem:

If LAN is connected not to LC module generating most data (i.e. to Pump in spite of DAD's presence) then interconnecting CAN connection between modules might get overloaded. Potential consecutive loss of spectra may be observed and is flagged as "Spectra runbuffer overflow" in logbook and on report.

Temporary Solution:

Installation documentation for LC HW states that the LAN access should be in the module with the highest workload, i.e. by DAD/MWD - FLD - VWD - RID - Pump - ALS (in this order) and independent whether it is a 1100 - 1200 - 1290 - 1260 system.

Fix Information:

n/a


KPR#:149837  Product:LC and CE Drivers  RC 1200.NET  A.02.05

Keyword: DAD/MWD Driver

One-line Description:

Autobalace executed although in Sequence table set to "OFF"

Problem:

1) A method XYZ.M is saved with following DAD settings: Autobalance -> Prerun
2) A sequence table is saved using the method XYZ.M and AutoBalace set to "OFF"
3) The sequence gets started and although AutoBalace set to "OFF" it is executed.

Temporary Solution:

still under investigation

Fix Information:

none


KPR#:149898  Product:LC and CE Drivers  RC 1200.NET  A.02.09 SP1

Keyword: DAD/MWD Driver

One-line Description:

Lowest possible DAD wavelength is 190.5 nm

Problem:

Lowest possible DAD wavelength is 190.5 nm. In earlier EZChrom and ChemStation revisions it was 190 nm. The midpoint wavelength or bandwidth need to be changed such that the lowest wavelength is 190.5 nm.

Temporary Solution:

Modify methods and set the wavelength and bandwidth such that their lowest wavelength is 190.5 nm.

Fix Information:

Has been fixed with driver A.02.11


KPR#:139618  Product:    A.04.07 SR2 HF05

Keyword: Data Analysis

One-line Description:

Incorrect value entered at peak group table if GC method parameter window is open using Agilent GC driver B.01.02

Problem:

If GC instrument method parameter window is open, an incorrect value may appear in the peak group table.
For example, click a cell and press 1.2 using the keyboard, then "1.002" is entered. 

Temporary Solution:

You need to use a mouse to click on the decimal position you want and then enter the desired value.   Alt. to close the method parameter window. 

Fix Information:

Fixed in Agilent GC Driver B.01.03.092 


KPR#:153192  Product:    A.04.07 SR2 (Tristan 3D)

Keyword: Data Analysis

One-line Description:

Chromatogram Annotation for Purity does not reflect actual value less than 1

Problem:

The decimals field for Purity annotation have no effect on the presented value in the chromatograms window and does not match the value reported in Intelligence Report (IR).

Temporary Solution:

Reprocessing all data of a result set is accurately representing the value of purity.

Fix Information:

The ACAML file containing metadata is not updated if only one data file of a result set is reprocessed. Consequently, the IR reporting does not match the Method Report. All data files of a result set must be reprocessed in order to accurately update the metadata in the ACAML file used by Intelligence Reporting.


KPR#:379394  Product:    A.04.07 SR2 HF01

Keyword: Data Analysis

One-line Description:

Sequence audit trail create time differs from report

Problem:

Create time in sequence audit trail differs from the create time in the sequence report by 8 minutes. There are four example scenarios below.

Scenario -1
Title: New sequence save as
Behavior: Mismatch in sequence creation time and Audit Trail sequence creation entry.
Steps: 	
-Create a new sequence and make changes to update the Audit Trail.
-Save as a new sequence with the same sequence name.

Scenario -2
Title: Existing sequence updated and save as the same name
Behavior: No mismatch in sequence creation time and Audit Trail sequence creation entry

Scenario -3
Title: Existing sequence updated and saved.
Behavior: No mismatch in sequence creation time and Audit Trail sequence creation entry.

Scenario -4
Title: Existing sequence save as with different existing sequence name.
Behavior: No sequence creation entry in Audit Trail of updated file. All previous Audit Trail entries are erased with new Audit Trail entry ‘Saved from {sequence path}’. Found mismatch in sequence creation time and Audit Trail entry.

Temporary Solution:

n/a

Fix Information:

Will not fix


KPR#:408720  Product:    A.04.09 Update 05

Keyword: Data Analysis,Installation

One-line Description:

Analyses turns into error state  "HRESULT: 0x80010105 (RPC_E_SERVERFAULT)"

Problem:

During analyses, the error information is that it could not be connected to the EZChrom client with an exception "HRESULT: 0x80010105 (RPC_E_SERVERFAULT)". In the log file for the instrument client it is stated that the URI is empty as there is an UriFormatException thrown.

Temporary Solution:

Do not exceed a thirty one (31) character username.

Fix Information:

The error will occur in all Agilent OpenLab CDS EZChrom Edition versions, with or without Agilent OpenLab Sample Scheduler where the username exceeds 31 characters. The 31 character limitation will not be changed.


KPR#:191223  Product:    A.04.07 SR2 HF03

Keyword: Data Store Integration

One-line Description:

The Users and Groups report generated on a A.02.02 client from a 2.0 Datastore Server is incorrect

Problem:

The Users and Groups report generated on a A.02.02 client from a 2.0 Datastore Server is incorrect. If privileges are assigned only using groups, the users don't show up on the report. 

Temporary Solution:

A report generated on the server works correctly.

Fix Information:

The problem will be resolved with later Control Panel and later OpenLab Server revisions


KPR#:339790  Product:    A.04.08 Update 05

Keyword: Data Store Integration,ECM Integration

One-line Description:

Mismatch between result set modification time and result set creation time on CM and ECM3.5 backend 

Problem:

Content Management 2.1 or ECM 3.5 update06 backend
Steps to reproduce:
1. Launch any instrument online and create a method which includes method report containing Result set creation time, Result set last Modification time.   
2. Submit a single run /sequence run with above created method including hard copy and PDF printing.
3. Verify Result set creation time and result set last Modification time in method report available in result set, on Hard copy print, in pdf print and on CDS.

Observation
Discrepancy observed between method report exported and method report on CDS, for result set modification time.
-Whenever result set is opened on CDS, the result set modification time updates even though no change is made in the result set.
-Discrepancy observed between method report exported and method report on CDS, for result set creation time and result set modification time. 

Temporary Solution:

n/a

Fix Information:

Do not fix


KPR#:31  Product:Waters LC  M8239AA Waters Alliance HPLC Driver  6.1.2.1

Keyword: Documentation

One-line Description:

The module type of the Alliance instrument is reported as a 'Detector' in OpenLAB CDS ChemStation Edition reports

Problem:

Using the OpenLAB CDS ChemStation Edition reporting functions, the module type of the Alliance instruments is incorrectly reported as a 'Detector'.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:146997  Product:Waters LC  M8505AA/BA Waters ACQUITY/Alliance  Alliance 6.1.3

Keyword: Documentation

One-line Description:

Instrument Configuration report can not be printed when offline.

Problem:

Instrument Configuration report can not be printed when offline.

Temporary Solution:

n/a

Fix Information:

Fixed in Driver version 6.1.3


KPR#:147024  Product:Waters LC  M8505AA/BA Waters ACQUITY/Alliance  Alliance 6.1.2.1

Keyword: Documentation

One-line Description:

The firmware version required by the driver is incorrectly stated as v1.41.6164 for the Waters Alliance 2489 detector in the User Guide

Problem:

The firmware version required by the driver is incorrectly stated as v1.41.6164 for the Waters Alliance 2489 detector in previous versions of the driver User guide.
 it should read v 1.41.2272

Temporary Solution:

n/a

Fix Information:

This has now been corrected in the guide for Alliance driver v6.1.4


KPR#:177259  Product:    A.04.08

Keyword: Documentation

One-line Description:

No help for reason type in Method Audit Trail

Problem:

The issue may be observied when one enables audit trail in the projectand sets prompt for a reason when saving a method. Then launch an instrument session. Create a method and save as or modify an existing method. After a change to the method, File > Save > Method. A dialog panel displays, heading is "Method Audit Trail" pops "Please select a reason or type in your own:" Pressing the "Help" button renders a popup message "VB (Offline)" and includes the message "Help not found."

Temporary Solution:

n/a

Fix Information:

If information is required, launch an instrument session and press the blue colored question mark for EZChrom Help.

In the Contents tab go to About Methods > Method Properties > Method Properties Audit Trail.


KPR#:260862  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_OLCDS_V1.0

Keyword: Documentation

One-line Description:

PE Clarus GC: Some Method parameters are missing in the method report

Problem:

Some Method parameters are missing in the method report that appear in Agilent GC method report.

Temporary Solution:

NA

Fix Information:

Method parameters are now present in PE Clarus GC reports with driver version 1.3.0.21 (OpenLab EZChrom system) and 1.0.21.1 (OpenLab CDS 2.x system)


KPR#:391787  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.0

Keyword: Documentation

One-line Description:

Hitachi's Sub CPU Firmware can Only be Visible from Hitachi Maintenance Tool

Problem:

Hitachi LC's Main Board firmware can be visible in the Instrument Configuration report and in the Module List that is accessed from the Instrument Status window. However, Hitachi LC's Sub CPU firmware can only be accessed from the Hitachi Maintenance Tool that is used for configuring the instrument.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 1.1 of the driver.


KPR#:601921  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.1

Keyword: Documentation

One-line Description:

Hitachi Chromaster Ultra Rs: Document for the Hitachi Maintenance Software is not the latest

Problem:

The "CM6000Maintenance" document in the driver package is from EZChrom platform, not OpenLab CDS. The Hitachi driver information is still valid except a few mentions of EZChrom.

Temporary Solution:

n/a

Fix Information:

The document is replaced by the proper one in the driver revision 1.2.


KPR#:634039  Product:Shimadzu LC  M8240AA/BA Shimadzu LC  ShimadzuLCGC v3.20

Keyword: Documentation

One-line Description:

Shimadzu LC:  No Information about Wavelength for UV Detector in Chromatogram in DA

Problem:

Shimadzu UV detectors (except for LC-40) do not display the wavelength used in the data acquisition in the chromatogram in the Data Analysis.

Temporary Solution:

The wavelength for the detector can be found in Acquisition Setpoints.

Fix Information:

LC2050/2060 is fixed in revision 3.40 of Shimadzu LCGC driver. The same will not be addressed for LC20/LC30/LC2030/LC2040 and a workaround described below to be used instead.


KPR#:651995  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_V1.3.0.21,PEClarusGC_OLCDS_V1.0.22.1

Keyword: Documentation

One-line Description:

PE Clarus GC 500: Min Firmware for Mainboard Missing in Training

Problem:

The minimal main control board firmware version for PE Clarus GC 500 is not specified in the documentation.

Temporary Solution:

The main control board firmware version 4.3.026 (min.) is required to use PE Clarus GC 500 Control (v1.3.0.22 and older) for OpenLab CDS.

Fix Information:

PE Clarus GC 500 is now obsoleted (starting with Clarus GC Control 2.0).


KPR#:701847  Product:Hitachi LC  M8227AA/BA Hitachi LC  HitachiLC_EZC_June19Ed,HitachiLC_EZC_March22Ed

Keyword: Documentation

One-line Description:

Hitachi Chromaster Ultra LC: Inconsistent driver version 3.20 reported in Software Verification Tool

Problem:

When the Hitachi CMU driver version 3.19 is installed, it is shown 
* as version 3.19 in Windows Program and Features (correct)
* as version 3.20 in Software Verification Report (false)
This issue has no impact on the driver usage.

Temporary Solution:

N/A

Fix Information:

This was fixed with the newer driver version 3.22-- now the version is reported consistently.


KPR#:251997  Product:    A.04.08

Keyword: ECM Integration

One-line Description:

ECM LCDF name ending with a period "." prevents resultset download

Problem:

Topology: OLCDS EZChrom A.04.xx and ECM 3.5

When the last character of the folder name in ECM is a "." dot the associated result set can't be loaded in EZChrom. This was observed with ECM 3.5 update 5 and OpenLab CDS EZChrom A.04.08

Temporary Solution:

n/a

Fix Information:

Will not fix


KPR#:277494  Product:    A.04.09 Update 03

Keyword: ECM Integration

One-line Description:

Report Template Long File Name Not Uploaded and Causes InstClient crash

Problem:

A.04.09 with ECM 3.6 and other versions. 
Example steps to reproduce:
1.Launch an instrument online.
2.Open Blank .rdl template from File > New > Report.
3.Add some snippet and/or make changes and save the report template with a name having more than 200 characters.

Result: InstClient crash and the report template is not saved to the Template folder of the Project in ECM. 

Note: Above limits are based on the path length of the test system. The path length will vary from system to system.

Regarding the number of characters, the number will vary, depending on the URL file path constraints defined in ECM Administrator's Guide.pdf, section "General Constraints on File Path Lengths". Read the entire section to fully understand how the ECM URL path is formed. Additionally, read how special characters are interpreted.

If the Windows OS limitation of URL encoded path length exceeds 245 characters, the file will not be uploaded. When this path length is exceeded, the OpenLAB CDS EZChrom can crash or result in unexpected behavior.

Temporary Solution:

Agilent recommends using as short as possible file names and/or comply with the Windows OS path limitations.

Fix Information:

n/a


KPR#:278036  Product:    A.04.09 Update 03

Keyword: ECM Integration

One-line Description:

Long File Name Prevents Single Level Calibration

Problem:

A.04.09 with ECM 3.6 and other versions.

Steps to reproduce:
1.Launch an instrument online from a long project (Enter Project name with 100 characters or longer).
2.Open a data file.
3.Perform single level calibration using long result name (Enter result name with 50 characters or longer).
4.Observe the CDS behavior.

Result: Single level calibration does not work. Nothing happens on CDS UI.

Expected result: Single level calibration should work without any issue. Result set should uploaded to backend without any issue. If Path name is too long it should give proper error message for long path, but does not.

Note: Above values are based on the path length limit of 245 characters for each unique system. The path length will vary from system to system.

Long file names can impact either an upload to ECM, download or copy to the enterprise path.

ECM URL: If this URL encoded path length exceeds 245 characters, the file will not be uploaded. Read entire section, General Constraints on File Path Lengths, in ECM 3.5 ECM Administrator's Guide.pdf, for interpretation of special characters. The ECM URL path is constrained by the rules of the Windows OS. 

UNC Path: Enterprise path limit is limited to 245 characters by OpenLAB CDS EZChrom. 

The OLCDS EZChrom should catch long file names that could exceed UNC and ECM URL path limitation, but one should not rely only on the detection. 

Temporary Solution:

Agilent recommends using as short as possible file names and/or comply with the Windows OS path limitations.

Fix Information:

n/a


KPR#:350455  Product:Hitachi LC  M8227AA/BA Hitachi LC  HitachiLC_EZC_June19Ed

Keyword: External Component

One-line Description:

Hitachi Chromaster can't Initialize after AIC reboot with "IFC cannot be found" error

Problem:

With Hitachi Chromaster v7.21 with OpenLab CDS EZChrom A.04.08/09
rebooting the AIC and launching the instrument on-line cannot
initialize the instrument:
All modules come up in red color and the error message "IFC cannot be found" is displayed.


Temporary Solution:

n/a

Fix Information:

The issue is fixed by upgrading the firmware for the IFC board to 8928190-10; See Chromaster Release Note document delivered with Hitachi LC October 2024 Edition for OpenLab EZChrom


KPR#:658322  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.1,Hitachi_CM_2.6,Hitachi_LCE_2.6

Keyword: External Component

One-line Description:

Hitachi LC: Incorrect Display of Parameters in Maintenance Tool with Win 11

Problem:

This only happens with Windows 11 OS: The border of editable fields in the Hitachi Maintenance Tool is not displayed correctly after the screensaver was activated once. However, it is still possible to enter values and save them.

Temporary Solution:

N/A

Fix Information:

The workaround is documented in the Release Notes in revision 2.8 of the driver.


KPR#:708791  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.1,Hitachi_CM_2.6,Hitachi_LCE_2.6

Keyword: External Component

One-line Description:

Hitachi LC: USB Drivers incompatible with OS Memory Integrity Option

Problem:

When a Hitachi LC driver for OpenLab is installed on an Windows OS with activated "Memory integrity" option, the installed Hitachi USB drivers do not become  valid.
Therefore, it will not be possible to connect with the Hitachi LC instrument. The following error is displayed in the configuration process: Unable to connect to IFC.
Alternatively, the Memory Integrity option cannot be enabled with Hitachi USB drivers already installed. "Incompatible Drivers" message is shown in such case.

Temporary Solution:

n/a

Fix Information:

This is fixed in driver version 2.7.


KPR#:762177  Product:Hitachi LC  M8227AA/BA Hitachi LC  HitachiLC_EZC_March22Ed

Keyword: External Component

One-line Description:

Hitachi LC: USB Drivers incompatible with OS Memory Integrity Option

Problem:

When a Hitachi LC driver for OpenLab is installed on an Windows OS with activated "Memory integrity" option, the installed Hitachi USB drivers do not become  valid.
Therefore, it will not be possible to connect with the Hitachi LC instrument. The following error is displayed in the configuration process: Unable to connect to IFC.
Alternatively, the Memory Integrity option cannot be enabled with Hitachi USB drivers already installed. "Incompatible Drivers" message is shown in such case.

Temporary Solution:

Uninstall the USB driver for Hitachi from Device Manager/Driver menu. Uninstall the Hitachi driver. Disable the Memory Integrity option. Reinstall the Hitachi driver afresh. The USB drivers should work properly this time.

Fix Information:

Fixed in Hitachi LC October 2024 Edition for OpenLab EZChrom; New compatible USB driver version will be installed with the LC driver.


KPR#:788048  Product:Shimadzu LC  M8240AA/BA Shimadzu LC  ShimadzuLCGC v3.20

Keyword: External Component

One-line Description:

Shimadzu LC: Current Value of Wavelength Accuracy 656nm (NAR) Not Displayed in Shimadzu LC System Check Report

Problem:

When running a system check, in the W lamp wavelength accuracy check, the 656nm (NAR) item is "Passed", but the current value (= actual measured value) is blank. 

Temporary Solution:

N/A

Fix Information:

This is planned to be fixed in the next revision of Shimadzu driver


KPR#:1036196  Product:Shimadzu LC  M8240AA/BA Shimadzu LC  ShimadzuLCGC v2.10

Keyword: External Component

One-line Description:

Shimadzu LC: SPD-M20A Excluded from the System Check Report if Upgraded to SPD-M40

Problem:

If SPD-M20A upgraded to SPD-M40, it does not show up in the System Check report anymore.
During the system check, the detector goes offline and is excluded from the System Check report.

Temporary Solution:

Run System check without the wavelength checked for the detector to be visible in the report, as a workaround.

Fix Information:

This is caused by the SPD-M40's "M20A compatibility mode” feature. To be addressed in the instrument firmware by Shimadzu.


KPR#:1077556  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.1

Keyword: External Component

One-line Description:

Hitachi LC: Incorrect Display of Parameters in Maintenance Tool with Win 11

Problem:

This only happens with Windows 11 OS: The border of editable fields in the Hitachi Maintenance Tool is not displayed correctly after the screensaver was activated once. However, it is still possible to enter values and save them.

Temporary Solution:

N/A

Fix Information:

N/A


KPR#:1092976  Product:CTC PAL3 LC  M8223BA CTC PAL3 for LC  CTC_LC_PAL3_1.7.4_OpenLabCDS

Keyword: External Component

One-line Description:

CTC PAL 3 for LC: Importing Pre-installed LC Injection Script Leads to Exception

Problem:

Re-importing the pre-installed LC injection script from C:\Program Files (x86)\Agilent Technologies\OpenLab Acquisition\CTCAnalyticsPal3Driver, then closing the connection and re-opening the Acquisition again leads to an exception.

Temporary Solution:

The default scripts are already imported and should not be re-imported to avoid the issue.

Fix Information:

There is no fix.


KPR#:180  Product:Headspace Driver Software  G2924AA  B.01.07.3

Keyword: Headspace

One-line Description:

EZChrom: Original method not restored when sequence run from EZChrom

Problem:

At 7697 keypad, set Restore Method after sequence to YES.
But after sequence completion, temps remained with last line method of sequence.
Method Restore worked when sequence runs from 7697 keypad.

Temporary Solution:

N/A

Fix Information:

Fixed in Headspace driver B.01.07.03.


KPR#:22140  Product:Waters LC  M8239AA Waters Alliance HPLC Driver  6.2.1

Keyword: Help

One-line Description:

Error with detectors help

Problem:

For the 2489/2998 detectors in the method parameters UI pressing F1 for raising help an error message  appears. Nevertheless the correct help for these detectors is shown.

Temporary Solution:

The correct help is shown but also an error message appears. 
This error message and the related "Windows Help and Support" windows can be closed without consequence for the workflow.

Fix Information:

n/a


KPR#:90310  Product:  M8236AA/BA Thermo Fisher Surveyor/Accela  ThermoLC 4.1.4

Keyword: Help

One-line Description:

Incorrect units for autosampler method setup shown in Help

Problem:

For previous Thermo driver versions prior to v4.1.4, the Instrument setup section for Surveyor/Accela Autosampler methods within the Online Help displayed the wrong units (mL instead of µL). This does not impact functionality within the driver.

Temporary Solution:

n/a

Fix Information:

Fixed in the latest revision of the driver.


KPR#:592464  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.1

Keyword: Help

One-line Description:

Hitachi Chromaster Ultra: Help Clarification for the Purge Options of the Pump

Problem:

The Help incorrectly says that after the purge starts the flow will decrease; it should say the flow will increase.
The Help also says the valve of the Auto purge will stay open at the end of the purge. This is not correct: the valve is automatically closed as expected.

Temporary Solution:

n/a

Fix Information:

This is fixed in revision 1.2 of the driver.


KPR#:592467  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.1

Keyword: Help

One-line Description:

Hitachi Chromaster Ultra: Help Clarification to Rack Settings of Autosampler

Problem:

The Help says " If the chosen rack parameters are different than those stored in the Autosampler configuration, an error will occur during the run."
This happens only in case the Rack configuration of the OpenLab instrument is very different as the real used Rack.

Temporary Solution:

n/a

Fix Information:

Fixed in the driver version 1.2


KPR#:595483  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_LCE_2.5,Hitachi_LCE_2.6

Keyword: Help

One-line Description:

Hitachi LaChrom Elite: Incorrect Error Message for Analog Input Device Issue

Problem:

An incorrect error message for Analog Input Device can be seen upon encountering an issue: _A was received from the is in ADC calibrating_

Temporary Solution:

n/a

Fix Information:

Fixed in driver version 2.7.


KPR#:22  Product:OpenLAB Data Store  OpenLAB Data Store  A.01.01

Keyword: Installation

One-line Description:

FTP protocol is enabled by default on OpenLAB Data Store A.01.02 systems

Problem:

When OpenLAB Data Store A.01.02 is installed, any user configured in OpenLAB Shared Services for Instrument operation has the possibility to authenticate to the Data Store File system with their credentials. OpenLAB CDS makes sure that file uploads are maintained by the audit trail of Data Store. Users can also upload files via FTP connection directly to Data Store (without OpenLAB CDS). With the FTP connection enabled, it is possible to bypass the Data Store versioning features and overwrite existing files in Data Store without any audit trail notification.

Temporary Solution:

To disable Data Store FTP server "ftp.enabled" parameter in C:\Program Files (x86)\Agilent Technologies\OpenLAB Data Store\tomcat\shared\classes\alfresco-global.properties should be set to "false".

Fix Information:

FTP and FTPS  Upload are supported with OpenLAB Data Store A.02.02 or later


KPR#:38  Product:OpenLAB Data Store  OpenLAB Data Store  A.02.01

Keyword: Installation

One-line Description:

Project activity logs not displayed after upgrading to A.02.01

Problem:

The project activity log entries displayed in the project pane are nor displayed after upgrading to OpenLAB Data Store A.02.01 or OpenLAB Shared Services A.02.01.
The global activity log in the administration pane is not affected and will contain all entries after the upgrade.
The Project Activity log only shows the entries of the last 7 days.

Temporary Solution:

N/A

Fix Information:

n/a


KPR#:42  Product:OpenLAB Data Store  OpenLAB Data Store  A.02.02

Keyword: Installation

One-line Description:

OpenLAB Control Panel does not open after upgrading Windows Server 2008 R2 to Windows Server 2012 R2

Problem:

OpenLAB Control Panel does not open after upgrading Windows Server 2008 R2 to Windows Server 2012 R2.
The following error message is shown "Could not load type 'System.Runtime.Diagnostics.ITraceSourceStringProvider' from assembly 'System.ServiceModel.Internals, Version=4.0.0.0, Culture=neutral, PublicKeyToken=31bf385ad364e35"

Temporary Solution:

n/a

Fix Information:

Affected revision(s): OpenLAB CDS A.02.02

Microsoft .Net Framework 4.5.2 needs to be applied to fix the issue


KPR#:94355  Product:  M8236AA/BA Thermo Fisher Surveyor/Accela  ThermoLC 4.1.4

Keyword: Installation

One-line Description:

Installer & Uninstaller wizards still show incorrect references to the driver name as an "Addon" in the installer screens

Problem:

When running the install or un-install wizard for the Thermo RC.NET driver v4.1.4 or lower, references to the old driver product name are still seen on some screens e.g. "Addon". No impact on driver functionality is observed.

Temporary Solution:

n/a

Fix Information:

There is no fix.


KPR#:130809  Product:    A.04.07 SR1

Keyword: Installation

One-line Description:

If a privilege is removed from the default roles in OLSS the application registration fails at installation

Problem:

If a privilege is removed from the default roles in OLSS the application registration fails during installation. 

Temporary Solution:

Don't change the default roles in OLSS, when needed copy them to a new name and edit this new role.

Fix Information:

n/a


KPR#:175952  Product:Thermo Fisher GC  M8241AA Thermo Fisher GC  ThermoGC_2.3

Keyword: Installation

One-line Description:

SVT report does not work when installing the driver in non-default location

Problem:

If the driver is installed to a location different from the CDS installation directory, the driver is not listed in the list of modules identified at the bottom of the SVT-report.

Temporary Solution:

n/a

Fix Information:

Use only the default location when installing the driver, same as CDS.


KPR#:175999  Product:Thermo Fisher GC  M8241AA Thermo Fisher GC  ThermoGC_2.3

Keyword: Installation

One-line Description:

SVT Report Does Not Check 34 Files for Thermo Fisher Trace GC Driver

Problem:

SVT report does not check 34 files for Thermo Fisher Trace GC driver. For details, see the Readme that accompanies the installation zip in SubscribeNet.

Temporary Solution:

A manual verification of the 34 files is required for compliancy.

Fix Information:

There is no permanent fix.


KPR#:203907  Product:    A.04.09

Keyword: Installation

One-line Description:

Not all SVT reports get displayed on the screen when manually run SVT on Server

Problem:

Some of the Software Verification Reports are intermittently not displayed or popped up on the screen when manually run SVT tool on Server.

Temporary Solution:

In this case, please look in 'SVReports' folder on the installation drive.

Fix Information:

n/a


KPR#:222652  Product:Shimadzu LC  M8240AA/BA Shimadzu LC  ShimadzuLCGC v2.04

Keyword: Installation

One-line Description:

Missing file in the reference file in the MSI Package for Shimadzu LC in EZChrom Edition

Problem:

One file is missing in the reference file: shimadzu.lcdriver.optionalsettings.xml in the MSI Package for Shimadzu LC in OpenLab EZChrom Edition. 

Temporary Solution:

n/a

Fix Information:

This is addressed by the Shimadzu LC v1.63 driver.


KPR#:281960  Product:Bruker GC  M8237AA/BA Bruker GC  BrukerGC_A.02.03

Keyword: Installation

One-line Description:

SVT report has 41 files not found in reference

Problem:

When running MSI_SVT_Checker_1.0.2 to validate the Bruker Drivers A.02.03, it mentions 41 Files not found in Reference. This does not affect the driver functionality and performance.

Temporary Solution:

There is no work around.

Fix Information:

There is no fix.


KPR#:290491  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_OLCDS_V1.0.19.1

Keyword: Installation

One-line Description:

No warning when installing later PE GC driver version with the preinstalled older version might lead to malfunction

Problem:

Upon installation of the most recent version of the driver, there is no warning if uninstallation of the older version of the driver is not performed manually. This leads to coexistence of the two versions of the driver that might cause malfunctioning in the instrument control later. This issue happens with OpenLab CDS 2.x.

Temporary Solution:

n/a

Fix Information:

Fixed in PE Clarus GC Control 2.0


KPR#:297111  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_OLCDS_V1.0.21.1

Keyword: Installation

One-line Description:

SVT for OpenLab CDS disappears upon uninstalation of PE Clarus GC Drivers

Problem:

When Clarus GC drivers are uninstalled, the OpenLab CDS node in the Software verification Tool is not available any more and one cannot run the software verification report.

Temporary Solution:

n/a

Fix Information:

Fixed in PE Clarus GC Control 2.0


KPR#:298681  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_OLCDS_V1.0.21.1

Keyword: Installation

One-line Description:

There is no Warning When Another Version of PE Driver is Installed Over The Previous One

Problem:

There is no warning that another version of the driver already exists if another version is being installed. This leads to the co-existence of both driver versions that is also displayed in "Programs and features" . This might cause issues in instrument control. 

Temporary Solution:

n/a

Fix Information:

Fixed in PE Clarus GC Control 2.0


KPR#:303159  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_V1.3.0.22

Keyword: Installation

One-line Description:

SVT report does not show the installed driver for non-default path CDS installation

Problem:

The SVT report does not show the installed driver for non-default path CDS installation.

Temporary Solution:

The Windows Control panel does show the driver.

Fix Information:

There is no fix.


KPR#:605033  Product:    A.04.10

Keyword: Installation

One-line Description:

Mixed Mode OpenLab EZChrom A.04.10 and CDS 2.5 Warning 

Problem:

While installing OpenLab EZChrom A.04.10 into a OpenLab CDS 2.5 Mixed mode environment. A warning is displayed "Potentially incompatible Shared Services version detected. Shared Services Server version is higher than current version of Shared Services being installed on this system. Please refer to the installation documentation. Do you want to proceed?"

Temporary Solution:

n/a

Fix Information:

Despite the warning, OpenLab EZChrom A.04.10 is compatible and supported in a mixed environment with OpenLab CDS 2.5. Accept the warning by selecting Yes and continue with the installation. 


KPR#:606448  Product:Thermo Fisher LC  M8236AA/BA Thermo Fisher Surveyor/Accela  ThermoLC 4.1.4-UPD1

Keyword: Installation

One-line Description:

Error During Driver Installation

Problem:

The driver cannot be installed as described in the User's Guide by clicking on it. An error is thrown.

Temporary Solution:

Open a Command Prompt as Administrator.
Use the following command to start the installation:
msiexec /i ThermoLCAddOn.Drivers.OLCDS2.Setup-4.1.4.28.msi
Follow the Wizard instructions.

Fix Information:

No plan to fix it currently.


KPR#:2121  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.07 SR1 HF01

Keyword: Installation , 35900 Control

One-line Description:

35900E ADC-instrument cannot be configured with RC.Net driver after reinstalling the OpenLAB CDS application

Problem:

The 35900E cannot be configured with the RC.Net driver if the OpenLAB CDS application is uninstalled and reinstalled.

Temporary Solution:

If the OpenLAB CDS application needs to be reinstalled, additional drivers such as the Agilent 35000E RC.Net driver must be uninstalled first using the add/remove programs from Windows Control Panel.  
The Agilent 35900E RC.Net driver will need to be reinstalled after completing the OpenLAB CDS installation.

Fix Information:

n/a


KPR#:32  Product:Waters LC  M8239AA Waters Alliance HPLC Driver  6.1.2.1

Keyword: Instrument Control

One-line Description:

OpenLAB CDS ChemStation Edition sequence shutdown macro PUMP ALL OFF is not executed.

Problem:

Using the OpenLAB CDS ChemStation sequence shutdown macro in the sequence parameter screen and executing the command "PUMP ALL OFF" - the macro is not executed for the Waters Alliance instrument.

Temporary Solution:

The pump needs to be switched off manually.

Fix Information:

n/a


KPR#:146967  Product:Thermo Fisher LC  M8236AA/BA Thermo Fisher Surveyor/Accela  ThermoLC 4.1.2.1

Keyword: Instrument Control

One-line Description:

Thermo Accela pump communication LED remains active

Problem:

The Thermo Accela pump communication LED remains active for 5-6 seconds after disconnection. This a graphical impact only, the update of the interface is slightly delayed.

Temporary Solution:

N/A

Fix Information:

There is no fix.


KPR#:146992  Product:Waters LC  M8505AA/BA Waters ACQUITY/Alliance  Alliance 6.1.2.1

Keyword: Instrument Control

One-line Description:

During acquistion of aWaters Alliance HPLC connected to OpenLab CDS EZChrom Edition, the software allows access/change instrument parameters even if roles/privileges for manual control are disabled.

Problem:

Manual control privileges defined in OL Control Panel /Administration are not honored thus allowing users to change instrument parameters during acquisition.

Temporary Solution:

n/a

Fix Information:

Fixed in driver version 6.1.3


KPR#:146993  Product:Waters LC  M8505AA/BA Waters ACQUITY/Alliance  Alliance 6.1.2.1

Keyword: Instrument Control

One-line Description:

Unexpected change of Alliance pump flow rate and solvent composition after a run observed on a French operating system

Problem:

This problem was observed with a French Windows operating system with French locale settings set.
When a series of runs was queued, it was noticed that once the analytical run had completed, the Alliance pump would be set to a different flow rate and/or different solvent composition other than the initial strating conditions specified in the method while preparing for the next run. Only the equilibration was impacted. Methods were not impacted.

Temporary Solution:

n/a

Fix Information:

Fixed in driver version 6.1.3: Updated .DLL in the driver now allows French O/S and French locale settings to be used with this version (6.1.3) of the driver.


KPR#:147009  Product:Shimadzu LC  M8240AA/BA Shimadzu LC  ShimadzuLC 1.41

Keyword: Instrument Control

One-line Description:

Shimadzu HPLC shows "Ready" even if Pump or Lamp is set to Off

Problem:

At the first start of the instrument, it is observed that the OpenLab CDS instrument status shows a different status when compared to the actual LC status. For example: OpenLab shows a "Not Ready" status while the instrument shows "Ready". This can happen when Shimadzu LC pump or lamp are set to off. With some instrument configurations (such as with SPD-M20A), the OpenLab CDS Status always shows "not ready" outside of a run.

Temporary Solution:

Downloading a valid method to the instrument that has both the lamp and the pump turned on, adjusts the status bar in OpenLab CDS to "Ready".

Fix Information:

Shimadzu informed Agilent this will not be fixed.  


KPR#:211507  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEAutosystem4.2.0.1

Keyword: Instrument Control

One-line Description:

Inlet time program does not work properly

Problem:

If one enters the flow of for example 1.5ml/min and goes to the next column, it is automatically rounded to the next number 2, without digits.

Temporary Solution:

N/A

Fix Information:

This has been fixed in driver version 4.2.1.0


KPR#:212506  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEAutosystem4.2.0.1

Keyword: Instrument Control

One-line Description:

Only one valve functions

Problem:

In the Method Editor, the valves can be time programmed. But only Valve 1 has an effect.

Temporary Solution:

n/a

Fix Information:

This has been fixed in v. 4.2.1.0 of the driver


KPR#:212515  Product:PerkinElmer GC  M8230AA PerkinElmer AutoSystem GC  PEAutosystem4.2.0.1

Keyword: Instrument Control

One-line Description:

GSV and Split valves cannot be controlled from the events table

Problem:

GSV and Split valves cannot be controlled from the events table.
Caution: split valves connected to split/splitless inlets are not concerned and should not be controlled using the events table.

Temporary Solution:

n/a

Fix Information:

This has been fixed in v. 4.2.1.0 of the driver


KPR#:286720  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_V1.3.0.16

Keyword: Instrument Control

One-line Description:

Status of Oven and Autosampler is "Ready" when the GC is switched off 

Problem:

When the instrument is off-line, the status of the oven and the autosampler might incorrectly display them to be in a "Ready" state.

Temporary Solution:

n/a

Fix Information:

Fixed in PE Clarus GC Control 2.0


KPR#:288479  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_V1.3.0.22

Keyword: Instrument Control

One-line Description:

PE Clarus GC: at the end of a run ""The current method has changed" message

Problem:

When acquisition of a single run is finished, a dialog appears "The current method has changed - would you like to save it?" 
At this point, the method which was loaded before is not displayed in the header of the EZChrom UI any more, but "untitled.met" is displayed.
 When "No" is clicked in the dialog, the window disappears and the previously loaded method is shown again.

Temporary Solution:

There is no work around available.

Fix Information:

Will not be fixed.


KPR#:394374  Product:Shimadzu GC  M8232AA/BA Shimadzu GC  ShimadzuLCGC v2.10,ShimadzuLCGC v2.11

Keyword: Instrument Control

One-line Description:

Shimadzu 2010 Plus GC: instrument connection not closing gracefully

Problem:

After pressing the “Close Connection” button in OpenLab CDS, instrument status changes to “not connected” (which is fine), but shortly after that, the status turns into “not ready” or “standby”, which is not correct – the instrument should stay offline.

After that, it is not possible to launch the instrument again or try to close the connection again. A restart of the AIC is required to be able to launch the instrument again.

Temporary Solution:

NA

Fix Information:

Fixed with OpenLab CDS 2.4 Update 7 and in OpenLab CDS 2.5.


KPR#:401780  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.0

Keyword: Instrument Control

One-line Description:

"Switch Valve to Load" not available after "Switch Valve to Inject"; "Home arm" action needed first

Problem:

After a "Switch Valve to Inject" action, the "Switch Valve to Load" function is displayed in Direct control menu but is not active. The "Home arm" function has to be executed first, then "Switch Valve to Load" is activated and can be called.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 1.1 of the driver.


KPR#:492287  Product:Hitachi LC  M8227AA/BA Hitachi LC  HitachiLC_EZC_June19Ed

Keyword: Instrument Control

One-line Description:

Hitachi LC: Long Launching Time if Instrument Setup Window Left Open Prior To Exiting Instrument Window

Problem:

If the Instrument Setup window left open upon exiting the Instrument window, launching the same instrument might take up to 3 min next time.

Temporary Solution:

As a workaround, close the Instrument Setup window prior to exiting the Instrument window.

Fix Information:

N/A


KPR#:516796  Product:PerkinElmer LC  M8229BA PerkinElmer LC  PE_LC300_2.0

Keyword: Instrument Control

One-line Description:

PE LC300: Uploading Method to the Instrument with Turned off Pump Switches Pump On

Problem:

The LC 300 pump starts the flow when an acquisition method is sent to the instrument with a pump flow > 0 mL/min even if the pump was turned off on the instrument.

Temporary Solution:

There is no fix for this

Fix Information:

N/A


KPR#:536651  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_LCE_2.2,Hitachi_LCE_2.4,Hitachi_LCE_2.5,Hitachi_LCE_2.6

Keyword: Instrument Control

One-line Description:

Hitachi LaChrom Elite: Intermittent Waiting For Trigger Does Not Start the Run

Problem:

Waiting for Trigger happens repeatedly upon performing an injection on LCE.

Temporary Solution:

N/A

Fix Information:

This is fixed in driver revision 2.7.


KPR#:543340  Product:VICI Valves  M8224BA VICI Valves  VICIValve_A.01.02

Keyword: Instrument Control

One-line Description:

VICI Valves: Acquisition Crash on Pressing Change Button in VICI Dashboard

Problem:

The Acquisition process gets an exception when trying to use the Direct Control functionality of a VICI Valve.
In the Dashboard of the VICI Valve the Change button allows to open a popup window but the system crashes when clicking on it. 
This only happens with OpenLab CDS 2.6.

Temporary Solution:

Avoid using the Change button in direct control window for VICI valves.

Fix Information:

Fixed by the Update 01 in OpenLab CDS 2.6.


KPR#:645291  Product:Bruker GC  M8237AA/BA Bruker GC  BrukerGC_A.02.03

Keyword: Instrument Control

One-line Description:

Bruker 456 GC: FID Flame status inconsistent

Problem:

Observed with Bruker driver A.02.03:
When the flame is turned off on the GC because gases are cut down, the driver still shows "Flame ON" instead of "Flame OFF".

Temporary Solution:

Use visual confirmation for the flame ON/OFF

Fix Information:

Cannot be addressed due to the old age of the driver (it comes from Varian) and complexity of the change (it is a high risk change)


KPR#:723437  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_OLCDS_V1.0.22.1,PEClarusGC_V1.3.0.22

Keyword: Instrument Control

One-line Description:

PE Clarus 480 GC Cannot Be Connected to the AIC Due to Communication Error

Problem:

PE Clarus 480 GC cannot be connected to the AIC due to communication error.

Temporary Solution:

There is no workaround.

Fix Information:

Perkin Elmer stopped the support of Clarus 480 GC.


KPR#:939675  Product:PerkinElmer LC  M8229BA PerkinElmer LC  PEFlexarLC_1.0

Keyword: Instrument Control

One-line Description:

PerkinElmer Flexar LC: Communication to Instrument Lost During Long Sequence

Problem:

The execution of a long sequence might be sporadically aborted due to communication loss with one of the modules of the Perkin Flexar LC stack (usually pump or autosampler).

Temporary Solution:

None

Fix Information:

Currently, there is no plan from PerkinElmer to release a new version of the Flexar driver to fix this defect.


KPR#:947158  Product:VICI Valves  M8224BA VICI Valves  VICIValve_A.01.02

Keyword: Instrument Control

One-line Description:

VICI +Agilent 7890: Changing Position of VICI Valve during a Sequence Can Cause Its Abortion

Problem:

Changing positions of VICI valve during a sequence might lead to an abortion in case more than one Valve is used.
VICI Valve module goes to Offline state as position change takes too much time and acquiring run/sequence is aborted.

Temporary Solution:

A point patch for this issue is released in August 2023.

Fix Information:

To be addressed in the future revisions of the VICI driver.


KPR#:962546  Product:Shimadzu LC  M8240AA/BA Shimadzu LC  ShimadzuLCGC v3.20,ShimadzuLCGC v3.40

Keyword: Instrument Control

One-line Description:

Shimadzu PDA Cannot Be Configured with Agilent LC

Problem:

Agilent LC cannot be run with Shimadzu PDA as the run aborts immediately after the run is submitted. This issue was introduced with Shimadzu LC driver 2.10 and also happens in 2.20 (this does not happen with 2.00).

Temporary Solution:

Utilize version 2.00 as the countermeasure until this is fixed in the newer revision of the driver. Or apply Shimadzu LC 2.20 Point Patch 1 until the newer driver release.

Fix Information:

This is addressed by Shimadzu LC 2.20 Point Patch 1 (distributed by request only). This will also be included in the future Shimadzu LC driver (around Q1/Q2 2025).


KPR#:1129855  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CM_2.6,Hitachi_CM_2.7

Keyword: Instrument Control

One-line Description:

Hitachi Chromaster: Script Error Appears Under Certain Conditions

Problem:

When working with a Chromaster instrument, a script error can show up in the Status UI under certain conditions. 
This happens in a client/server environment if comma (",") is used as decimal separator. After the script error, a user without privilege can access the "Auto Purge Wash" & "Plunger Wash" functions in the Pump Status UI.

Temporary Solution:

There's no impact on data acquisition.
Confirm the script error by clicking the Ok button.
Use the "Auto Purge Wash" & "Plunger Wash" only if your role allows it.
Do NOT use the "Auto Purge Wash" & "Plunger Wash" when the state of the instrument is set to “Release” for “Take/Release” function.

Fix Information:

This is fixed with revision 2.8 of Chromaster driver.


KPR#:1185772  Product:Shimadzu LC  M8240AA/BA Shimadzu LC  ShimadzuLCGC v3.50

Keyword: Instrument Control

One-line Description:

Shimadzu LC: Manual Injection Starts Without Waiting for Trigger

Problem:

With Open Lab CDS Ver. 2.8 and Shimadzu LC Driver Ver. 2.20, the run starts without waiting for an external signal. The same issue does not happen with previous versions of OpenLab CDS.

This happens with Shimadzu LC20 and LC40 series.

Temporary Solution:

Avoid using manual injection in OpenLab CDS 2.8 until a fix is issued.

Fix Information:

This is fixed with a Shimadzu LC Control 2.20 Point Patch 2. Available upon request to the third party instrument control team.


KPR#:138745  Product:    A.04.07 SR1 Update 2

Keyword: Integration

One-line Description:

Application crash if back data files resultset is reprocessed with the clear calibration flag

Problem:

If using a GC with a back tower only and you reprocess with the Clear Calibration flag, the application can crash.

Temporary Solution:

Do not use the Clear Calibration flag when reprocessing using only the back tower. Use 100% calibWeight instead.

Fix Information:

n/a


KPR#:263508  Product:    A.04.08,A.04.08 Update2,A.04.09

Keyword: Integration

One-line Description:

Reprocess or analyze of data file causes instclient.exe crash

Problem:

In this case, one peak at noise level is integrated and reported twice. Noise level peaks are integrated and this conflict can cause the program to crash. 

Temporary Solution:

Zooming in to the data at noise level, there is no relevant peak seen at the detection point. One can use integration events to avoid unnecessary peaks used for calculations. The "Trace Annotation Properties" allows entry of an area limit for peak detection. Peaks areas that fall below this minimum area will not be integrated and reported as peaks. This event is useful for eliminating noise or contaminant peaks from your report and eliminate the InstClient crash.

Fix Information:

n/a


KPR#:287503  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_V1.3.0.16

Keyword: Localization

One-line Description:

PE Clarus GC: Decimal symbol for syringe volume

Problem:

The Method parameter "Syringe capacity" is not following the Regional Settings for the Decimal Symbol and will be always displayed with a dot (Options are 0.5 mL, 5.0 mL and 50 ml). As consequence, the "Syringe capacity" is also always printed with a dot in the Method Report. This doesn't have any impact on the instrument functionality.

Temporary Solution:

n/a

Fix Information:

Fixed in PE Clarus GC Control 2.0


KPR#:738391  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CM_2.3

Keyword: Localization

One-line Description:

Hitachi CM and LCE: Acquisition Window Crashes if Non-English OS Used for AIC/Client

Problem:

Once a Chromaster or LaChrom Elite instrument is launched, after 5 ~ 10 minutes OpenLab CDS is generating error messages in Activity log:
"Le format de la chaîne d'entrée est incorrect"
(Input String is not in correct format)
until the system crashes.

Temporary Solution:

Change the format of system account to English (United States). In Region settings select [Administrative][Copy current settings] to [Welcome screen and system accounts], and re-boot the AIC/Client. 

Fix Information:

Fixed in revisions 2.8 of both drivers.


KPR#:146963  Product:Thermo Fisher LC  M8236AA/BA Thermo Fisher Surveyor/Accela  ThermoLC 4.1.2.1

Keyword: Method

One-line Description:

Pretreatment methods are saving invalid parameter

Problem:

It is possible to save a pretreatment method with invalid parameters.

Temporary Solution:

As soon as the system provides an error message during the sample prep method creation, the user should abort and not save the method.

If user still saves the invalid sample prep and starts the run, run will be aborted.

Fix Information:

There is no fix.


KPR#:146971  Product:Waters LC  M8505AA/BA Waters ACQUITY/Alliance  Alliance 6.1.2.1

Keyword: Method

One-line Description:

Method upload functionality is not supported for both OpenLAB CDS Chemstation and EZChrom Editions

Problem:

The selectable method options "Upload from Instrument" and "New Method from Instrument" are still shown as being available without any errors or information being displayed but this functionality does not work. An entry in the instrument activity log states that this upload could not be performed.

Temporary Solution:

n/a

Fix Information:

This issue has been documented in the Installation and User's Guide for this driver.


KPR#:146977  Product:Waters LC  M8505AA/BA Waters ACQUITY/Alliance  Alliance 6.1.2.1

Keyword: Method

One-line Description:

Run Time Mismatch after running a Gradient Method

Problem:

The run time define for a isocratic method executes with the run time of a previous used gradient method, only when the following conditions are met:
-  isocratic method is using the pump auxiliary channels (switched on) 
- a gradient method was executed just before executing the isocratic method
- OpenLAB CDS was not rebooted in between run

This defect has been reported on OpenLAB CDS EZChrom only.

Temporary Solution:

To temporarily resolve the issue 
-  reboot the OpenLAB CDS system
- add a timetable entry for the isocratic pump and enter the run time in there

Fix Information:

The issue is fixed in OpenLAB CDS Waters Alliance 6.1.2.1 HF1 and available from SubscribeNet.


KPR#:146995  Product:Bruker GC  M8237AA/BA Bruker GC  BrukerGC_A.01.01

Keyword: Method

One-line Description:

The 450GC oven does not cool down as expected when the run finishes.

Problem:

When running the 450GC with either the Bruker GC A.01.01 or A.01.02 instrument drivers, the ramp down of the oven tempertaure once the run finishes takes a very long time to achieve the initial oven temp setpoint again. This can take upto several hours to achieve the initial oven temperature setpoint again for the subsequent injection.

Temporary Solution:

n/a

Fix Information:

Fixed in A.01.03.


KPR#:146996  Product:Bruker GC  M8237AA/BA Bruker GC  BrukerGC_A.01.02

Keyword: Method

One-line Description:

Not Ready in Run status/warning appears during run of 450GC and stops run

Problem:

If a very steep, unrealistic oven ramp is programmed (e.g. in excess of 50C/min where the GC is not physically able to achieve), the GC while able to start the run is then not able to manage the steep oven ramp  profile thus this warning is displayed during the run. Subsequently the run is stopped on the GC and on Openlab EZChom Edition.

Temporary Solution:

n/a

Fix Information:

Fixed in A.01.03 driver revision.
This warning is now ignored by the GC and the run now continues. Methods should be checked prior to starting the run to ensure realistic oven ramp profiles are entered.


KPR#:147008  Product:Shimadzu LC  M8240AA/BA Shimadzu LC  ShimadzuLC 1.41

Keyword: Method

One-line Description:

Missing instrument configuration information in the method reports.

Problem:

It has been observed that when printing an OpenLab standard or custom report, the instrument configuration information is missing from the method. 

Temporary Solution:

It is possible to obtain a list of the instrument configuration including modules and serial numbers by selecting the Information ("i") button at the bottom of the "Instrument Status" window.

Fix Information:

Fixed in Shimadzu LC 1.5


KPR#:147013  Product:Bruker GC  M8237AA/BA Bruker GC  BrukerGC_A.02.02

Keyword: Method

One-line Description:

Instrument Configuration information is not contained within method report.

Problem:

It has been observed that when printing an OpenLAB standard or custom report, the instrument configuration information is missing from the method. This is a known issue and expect this to be fixed in a future OpenLAB CDS version.

Temporary Solution:

Workaround: It is possible to obtain a list of the instrument configuration including modules and serial numbers by selecting the Information ("i" button at the bottom of the Instrument Status window.

Fix Information:

n/a


KPR#:147017  Product:VICI Valves  M8224BA VICI Valves  VICIValve_A.01.01

Keyword: Method

One-line Description:

Instrument Configuration information is not contained within method report.

Problem:

When printing a Method report with the Instrument details listed, no actual information for the instrument configuration is available or printed.

Temporary Solution:

Workaround: It is possible to obtain a list of the instrument configuration including modules and serial numbers by selecting the Information ("i")button at the bottom of the Instrument Status window.

Fix Information:

There is no fix for this


KPR#:279725  Product:Shimadzu LC  M8240AA/BA Shimadzu LC  ShimadzuLCGC v2.05,ShimadzuLCGC v2.10,ShimadzuLCGC v2.11,ShimadzuLCGC v3.00,ShimadzuLCGC v3.10,ShimadzuLCGC v3.20

Keyword: Method

One-line Description:

Shimadzu LC Vial message: "Object reference not set to an instance of an object"

Problem:

With Shimadzu LC v1.63 driver (or later) in OpenLab CDS, after launching the instrument the message "Object reference not set to an instance of an object" appears in field validation for "Vial" in "Single Sample", although the default value is correct.

Temporary Solution:

Vial number has to be re-entered to be valid.

Fix Information:

Will not fix.


KPR#:288808  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_V1.3.0.16

Keyword: Method

One-line Description:

Split flow in instrument status is always X mL/min higher than in instrument setup

Problem:

When one enters a value for split flow in Instrument Setup and downloads the method to the instrument, the split flow displayed in the instrument status is always X ml/min higher than the value entered, where X is the septum purge flow plus the column flow.
 
The value displayed as "Split flow" in the Status UI is the TOTAL flow value, i.e.septum purge flow plus the column flow plus the split flow, as it is displayed on the GC device UI.

Temporary Solution:

n/a

Fix Information:

Fixed in PE Clarus GC Control 2.0


KPR#:290303  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_OLCDS_V1.0.19.1

Keyword: Method

One-line Description:

Instrument Setup Window: Time constant for FIDW is displayed without a unit

Problem:

The field Time constant for FIDW is displayed without a unit in the Instrument setup window.

Temporary Solution:

n/a

Fix Information:

Fixed in PE Clarus GC Control 2.0


KPR#:297200  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_OLCDS_V1.0.21.1

Keyword: Method

One-line Description:

Incorrect Units (°C) for carrier gas settings in Method report

Problem:

In the method UI carrier settings are shown with unit psi in the time table (which is wrong when the flow control mode is selected).  When the method is printed in CDS 2.x, the carrier settings in the table are shown with the unit "°C", which is always incorrect independent of the mode.

Temporary Solution:

n/a

Fix Information:

Fixed in PE Clarus GC Control 2.0


KPR#:354452  Product:Shimadzu LC  M8240AA/BA Shimadzu LC  ShimadzuLCGC v2.10

Keyword: Method

One-line Description:

Shimadzu LC 2030C+: invalid range(0.1-50.0) error for syringe >50 uL

Problem:

The instrument is shown as 100 uL loop size(maximum injection volume) at configuration window, but it allows to enter up to 50 uL. If one enters more than 50 uL injection volume in the acquisition method, it is given an warning message "invalid range(0.1-50.0)step 0.1".
It is not possible to create any acquisition method with 100 uL injection volume either.


Temporary Solution:

NA

Fix Information:

This is fixed by Shimadzu in v 2.00 of LC driver (v3.00 of LCGC driver package)


KPR#:370796  Product:    EZChrom Edition A.04.09

Keyword: Method

One-line Description:

Method resolution is possible without privilege "Instrument Setup"

Problem:

When a method from a different instrument is loaded method resolution is activated and the method resolution panel is presented. Resolution and save method are possible.
- Modifies instrument parameters
- Allows the logged on user to modify instrument parameters
The expectation is the normal user without instrument setup should not have method save privileges.

Temporary Solution:

n/a

Fix Information:

Will not be fixed.


KPR#:530381  Product:  M8242BA PerkinElmer TurboMatrix HS  PETurboMatrixHS_V1.0

Keyword: Method

One-line Description:

PE TurboMatrix HS: New Method Throws an Error Message Upon Saving

Problem:

When saving a new acquisition method, an error message appears: " 'TurboMatrix HS' has one or more invalid settings",
even when there are no invalid settings.
The issue is only seen once after launching a newly configured instrument  for the first time and saving a method.

Temporary Solution:

N/A

Fix Information:

This is fixed in revision 1.1 of the driver.


KPR#:542480  Product:PerkinElmer GC  M8230AA PerkinElmer AutoSystem GC  PEAutosystem4.2.3Patch01

Keyword: Method

One-line Description:

PE AutoSystem XL: Method Downloading to the Instrument Not Logged in the Activity Log

Problem:

PE Autosystem XL: Downloading the method to the instrument is not logged in the activity log,.

Temporary Solution:

N/A

Fix Information:

Not to be addressed in the future.


KPR#:695899  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.1

Keyword: Method

One-line Description:

Hitachi Chromaster Ultra LC: Direct Control "Turn on at" Functions Properly Only at UTC Time Zone Entry

Problem:

The issue happens with Hitachi CMU driver v1.1.
The "Turn on at" function does not work with the current time zone the AIC is set up to but instead uses the UTC time.
This affects the "Turn on at" function of the instrument's pump, oven and detectors.

Temporary Solution:

n/a

Fix Information:

Fixed in version 1.2 of the driver.


KPR#:751372  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.1

Keyword: Method

One-line Description:

Hitachi CMU: Method Resolution Error When Upgrading from CMU1.0 to CMU1.1

Problem:

When a method that is saved using the CMU 1.0 driver version is opened with a system with the CMU 1.1 driver version, some parameters are not resolvable and set to default. A message is given by the system to notify the users about it.

Temporary Solution:

n/a

Fix Information:

This is addressed in the Hitachi Chromaster Ultra Rs driver 1.2.


KPR#:761728  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PETurboMatrixHS_V1.1

Keyword: Method

One-line Description:

PE TurboMatrix and Clarus GC: System Sometimes Prompts to Save Unmodified Method 

Problem:

When trying to load an existing method or creating a new one, the system sometimes asks the user to save the current method even it has not been modified. This only happens if TurboMatrix Headspace is part of the system.

Temporary Solution:

N/A

Fix Information:

Addressed in the Point Patch for TurboMatrix driver. Contact 3PIC for the patch.


KPR#:781493  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CM_2.6

Keyword: Method

One-line Description:

Hitachi Chromaster LC: Actual Run Time Can Be 10x Longer Than Set in the Method Due to Certain System Settings

Problem:

The actual method run time may occasionally (depending on the Region system settings) be ten times longer than specified in the instrument method. For example, specifying a time of 0.5 minutes may result in an actual run time of 5 minutes.

Instrument / Configuration / Setup

OpenLab CDS 2.7
Hitachi Chromaster driver 2.6.02

Temporary Solution:

 If the issue occurs, please reach out to the Third Party Driver support for a workaround that involves changing certain system settings.

Fix Information:

Fixed in revision 2.8 of the driver.


KPR#:800420  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PETurboMatrixHS_V1.1

Keyword: Method

One-line Description:

PE TurboMatrix and Clarus GC: System Sometimes Prompts to Save Unmodified Method 

Problem:

When trying to load an existing method or creating a new one, the system sometimes asks the user to save the current method even it has not been modified. This only happens if TurboMatrix Headspace is part of the system.

Temporary Solution:

N/A

Fix Information:

Fixed by the Point Patch for TurboMatrix driver. Contact 3PIC for the patch.


KPR#:1063041  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_2.0

Keyword: Method

One-line Description:

PE Clarus GC Default Method Setting Combination for Carrier Not Allowed with EZChrom

Problem:

Clarus GC, EZChrom:
Upon creating a brand-new method, one can encounter an issue where they cannot switch out of the Carrier tab without any error being displayed. If this happens, it means that the following two default settings are not simultaneously allowed: 
Mode / Carrier A Control = Programmed Flow
AND
Split Control / Mode = None

Temporary Solution:

Change the Split mode to a "Flow" from "None". Alternatively, change the Carrier Mode to "None" and keep Split Mode at "None".
This might only happen when the method is created. After that, the tooltip is properly displayed informing the user about the combinations that are not allowed.

Fix Information:

To be addressed in the future revision of the driver.


KPR#:1155105  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_2.0

Keyword: Method

One-line Description:

Clarus GC: Method UI Might Appear Hanging For Carrier Tab Under Certain Condition

Problem:

Under certain conditions, it can happen that the Clarus GC Method Editor appears to be hanging in the Carrier Tab.

Temporary Solution:

One should use a temporary workaround: switch to another tab in the method editor (e.g. Oven or Detector) and continue working. The current method is not affected by that and can be saved for further use.

Fix Information:

To be fixed with next Clarus GC Control driver version


KPR#:147025  Product:Shimadzu LC  M8240AA/BA Shimadzu LC  ShimadzuLC 1.41

Keyword: Method Usage

One-line Description:

After data has been acquired on Shimadzu LC-20 systems, a warning  "The current method has changed would you like to save it?" is displayed.

Problem:

After an a data acquisition has been run, the following warning will be displayed even the method has been saved before.
"The current method has changed. Would you like to save it?" You can quit this box with "Yes", "No" or "Cancel" buttons".

Temporary Solution:

Please do not open the Instrument setup UI during acquisition.

Fix Information:

Fixed in Shimadzu LC 1.51


KPR#:297768  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_OLCDS_V1.0.21.1

Keyword: Method Usage

One-line Description:

Incorrect Carrier Info in Method Report

Problem:

In the method report, under "Carrier B setting", there is "Carrier A Control: NONE"
 It should be "Carrier B Control: NONE"

Temporary Solution:

n/a

Fix Information:

Fixed in PE Clarus GC Control 2.0


KPR#:414461  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.0

Keyword: Method Usage

One-line Description:

Incorrect Absorption Value in the Online signal and Status of 6420 UV-VIS Detector When Non-Default Range Value Used.

Problem:

When the Absorbance range of the 6420 UV-VIS detector is not 1.0 (default value), the displayed value of the Online signal and the Status is not correct and have to be ignored. However, the acquired and stored data is not affected. The data corresponds to the selected Range value and will be displayed correctly in Data Analysis.

Temporary Solution:

n/a

Fix Information:

The failure is fixed in  revision 1.1 of the driver (the fix is in the firmware).


KPR#:415474  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.0

Keyword: Method Usage

One-line Description:

Hitachi ChromasterUltra Rs: DAD Online Signal  Wavelength Needs to be Selected in DAD Direct Control Option

Problem:

The DAD signal in the online plot does not reflect the signals selected in the method currently downloaded to the instrument. Instead, the option "Change Monitoring Wavelength" in DAD direct control context menu (in the Instrument Status window) controls what wavelength to display for the online plot.
This behavior applies to both  OpenLab CDS and OpenLab CDS EZChrom Edition for Hitachi ChromasterUltra Rs driver.

Temporary Solution:

None.

Fix Information:

Fixed in revision 1.1 of the driver.


KPR#:691353  Product:Shimadzu GC  M8232AA/BA Shimadzu GC  ShimadzuLCGC v3.10

Keyword: Method Usage

One-line Description:

Shimadzu GC: Method Resolution Error Does Not Provide Details

Problem:

Upon loading a Method, a method resolution error shows up but does not provide info on what needs to be adjusted based on changed instrument configuration. In this particular case a syringe volume was changed. There are possibly more cases when such incomplete error message on method resolution shows up.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 3.40 of Shimadzu LCGC driver. 


KPR#:246004  Product:    A.04.08

Keyword: Output

One-line Description:

EZChrom GPC/SEC uses a time slice algorithm that gives different results from other GPC software

Problem:

EZChrom GPC calculations use a time slice method of performing GPC calculations. This method was used by earlier generation GPC software packages because it is computationally efficient (faster with less memory required). Today, data storage and processing is no longer a concern and all modern GPC software packages simply use the available data points.
	
EZChrom GPC also performs calculations with integer values by rounding time values to the nearest second. Again, this is done for computational efficiency. For narrow slice widths, this has the effect of slightly increasing the time values which can result is significantly lower molecular weights. The magnitude of the difference depends on the selected slice width and peak shape. 

Temporary Solution:

EZChrom GPC/SEC software will continue to be supported by Agilent until the EZChrom version goes out of support. 
	
Note Agilent discontinued sales of new EZChrom GPC/SEC packages in November 2017.

Fix Information:

Agilent recommends users of EZChrom GPC/SEC plan to upgrade to OpenLab CDS with the GPC Add On. This package uses an updated algorithm that provides more accurate results.
	
For applications such as batch to batch verification, the change from EZChrom to OpenLab CDS may require a new specification be established. Agilent recommends users perform a side-by-side study with the analytes of interest between EZChrom GPC/SEC and OpenLab CDS with GPC. With the study results, users can establish a new specification using OpenLab CDS with the GPC Add On software.


KPR#:261090  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.0.0

Keyword: Performance

One-line Description:

Abort Current Run Sets Instrument in Error State

Problem:

Using the "Abort current run" option during sequence execution sets the instrument in the error state. Returning to idle state requires manual intervention.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 1.2 of Thermo Fisher SII.


KPR#:281316  Product:    A.04.08,A.04.09

Keyword: Performance

One-line Description:

Sporadic internal error "timeout expired" for activity log queries

Problem:

An "Internal server error - timeout expired" error message can occur when searching the System Activity Log with a user and/or description filter. This occurs more frequently with large Activity Logs and when the system is busy with a high load on the database. 
 
This occurs because user and description filters are full text queries of the database which are less efficient than other filters. When the query takes too long, the software aborts the operation and displays the error. The error does not indicate any data integrity issue.

A complete fix for this issue is being worked on. By then, the likelihood of the issue in some cases can be reduced by these means:
- Database index rebuilt and statistics update
- Further reduce the date/time range

Temporary Solution:

N/A

Fix Information:

Resolved in A.04.10


KPR#:377731  Product:  M8244BA Thermo Fisher SII  1.0.0

Keyword: Performance

One-line Description:

Thermo SII Driver: Corona Veo Not Supported in Agilent LC stack

Problem:

The Corona Veo CAD detector is currently not supported within an Agilent LC stack due to unreliable triggering into Run state of the detector.
Support for this configuration will be provided as soon as possible.

Temporary Solution:

n/a

Fix Information:

This issue is resolved with a ThermoFisher_SII_Control Hotfix-123396.


KPR#:511004  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_LCE_2.4

Keyword: Performance

One-line Description:

Refresh period of the Online Plot Might Increase for Long Sequences

Problem:

The period to refresh the Online Plot might expand by a fraction of a minute after running a few hundreds of injections in a single sequence.
The quality of the recorded data is not impacted by the issue.

Temporary Solution:

n/a

Fix Information:

This is fixed in Hitachi LaChrom Driver v2.6


KPR#:511440  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.1

Keyword: Performance

One-line Description:

Thermo Fisher SII: Submitting Shutdown Method in Acquisition Leads to Error for Corona Veo Detector

Problem:

If in the acquisition Run Queue, a Shutdown Method is defined and is submitted via the button in the run queue tab, the Corona Veo Detector goes into the error state with the error message:
"SII: This module expects to receive a hardware inject synchronization signal and is therefore not able to work in the NoInjection mode".

Temporary Solution:

Use a workaround: 
Add a shutdown method after a single run or as the last line of a sequence.

Fix Information:

This is fixed in Thermo Fisher SII Control 1.4.


KPR#:614017  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.1

Keyword: Performance

One-line Description:

Thermo Fisher Ultimate 3000: The Injection Collection Could Not Be Initialized

Problem:

SII 1.1: If the operator manually performs “Wash Needle Externally” before running a sequence, and the instrument directly runs the sequence after, the sequence operation will be interrupted with the following error:
SII: the injection collection could not be initialized (Not all injections have been run , the injections collection must not be finalized in that state)

Reason:
After executing the ”Wash needle externally ” program, the injection valve position does not return to the “injection position”.

Temporary Solution:

After completing the ”Wash needle externally ” program procedure, it is necessary to manually switch the injection valve to the injection position prior to starting a sequence.

Fix Information:

This issue is not reproducible in SII 1.3 revision of the driver.


KPR#:614020  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.1

Keyword: Performance

One-line Description:

Thermo Fisher SII: "Blank" Type Injections Cause Sequence to Abort

Problem:

Sequence goes into Error state and aborts if one adds/inserts blank type samples into a sequence during a run.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 1.2 of Thermo Fisher SII.


KPR#:658297  Product:Hitachi LC  M8227AA/BA Hitachi LC  OLCDS_2.7

Keyword: Performance

One-line Description:

Hitachi Chromaster LC: Long Acquisition Launch Time

Problem:

The Launch time of the Acquisition software can take more than 30 seconds when working with a Chromaster instrument.
The issue appears under the following conditions:
_ Previous Acquisition session has been closed while the Method Setup window was displayed
_ The instrument connection has been closed after the previous Acquisition session.

Temporary Solution:

Close the Method setup window prior to closing the instrument window. The delayed launch will be avoided this way upon the subsequent launch of the Hitachi instrument.

Fix Information:

This is fixed in OpenLab CDS 2.8.


KPR#:664341  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2

Keyword: Performance

One-line Description:

Thermo Fisher Trace GC, Integrion, Aquion: Method Part of the UI Not Accessible Due to Missing Scrollbar

Problem:

In Trace 1600/1610 method > Detector Time Program UI: When the Window is too narrow (e.g. when a second Window is open at its side) the Add/Delete/Sort buttons are not visible. There is no horizontal scrollbar to access this part of the UI. 
Similar issues happen with Aquion and Integrion.

Temporary Solution:

Workaround: close other window next to the method UI window. 

Fix Information:

To be addressed in the future revisions of the driver.


KPR#:667230  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2

Keyword: Performance

One-line Description:

Thermo Fisher Aquion Remains Offline After Reset

Problem:

Aquion remains offline after resetting it from the instrument dashboard context menu of the Pump ECD; the module goes offline and does not start again. 

Temporary Solution:

Use "Connect" option in the context menu of the module to bring the instrument back on-line.

Fix Information:

N/A


KPR#:667234  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2

Keyword: Performance

One-line Description:

Thermo Fisher Aquion: Error Due to "Non-Specified Flow Rate"

Problem:

For certain suppressor types, a warning about non-specified flow rate at the initial step is displayed when doing a method check, although the initial flow rate is specified. It is not possible to start a run with this method, the run is aborted due to the hardware error.

Temporary Solution:

The issue does not happen is no suppressor is used.

Fix Information:

N/A


KPR#:667333  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2

Keyword: Performance

One-line Description:

Thermo SII Integrion: Setpoint Column Temp Shown Instead of Actual Temp in OpenLab Instrument Status

Problem:

The column temperature shown in the OpenLab Instrument Status UI dashboard for the Pump_ECD is the setpoint temperature, not the actual temperature.

Temporary Solution:

As a workaround, use the Thermo Fisher Instrument Status window that can be opened from the context menu of the corresponding module of the OpenLab Instrument Status window (right click on module-> Open Instrument View)

Fix Information:

To be addressed in the future revisions of the driver.


KPR#:667424  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2,1.3

Keyword: Performance

One-line Description:

Termo SII: Integrion Method Editing Issues

Problem:

Method usability issues can arise when changing method settings in Integrion. E.g. when changing the type of an existing method in the Pump section from isocratic to multi-step gradient and entering a new row in the gradient table in the Gradient tab causes an error message (without clarification). Also, selecting a Flow tab or another method section causes an error message as well.
Some other issues may arise in other tabs and fields when editing a method.

Temporary Solution:

The only way to avoid the error message is to delete the new row in the gradient table, go back to the Flow tab, change the method type to isocratic and then to multi-step gradient and go back to the gradient tab. This way the table can be edited without error.
Some other issues may arise in other tabs and fields when editing a method and sometimes for them there is no graceful workaround and the only way to proceed is to create/load another method and discard all changes to the present method. 

Fix Information:

To be addressed in the future revisions of the driver.


KPR#:683446  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.1

Keyword: Performance

One-line Description:

Thermo CAD with Agilent LC: The Sequence Aborts if it Encounters No Injection/Blank Run

Problem:

The "no injection/blank run" type cannot be used while the CAD is configured with Agilent LC in CDS system. The sequence gets aborted if it encounters blank run.

Temporary Solution:

If the workflow allows, use 0 injection volume in a regular run type instead of a No injection/Blank run type.

Fix Information:

This is fixed in Thermo Fisher SII Control 1.4.


KPR#:695395  Product:Shimadzu GC  M8232AA/BA Shimadzu GC  ShimadzuLCGC v3.20

Keyword: Performance

One-line Description:

Shimadzu GC: Online Signal Does Not Work

Problem:

After installing OpenLab CDS 2.7 Update 01, there is still a bug in the online plot for Shimadzu GC. The Online Plot is only shown correctly during a run. Outside a run, no online plot signal is shown. 

Temporary Solution:

n/a

Fix Information:

Fixed in revision 3.40 of Shimadzu LCGC driver.


KPR#:801770  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2

Keyword: Performance

One-line Description:

Sequence aborts after updating running sequence

Problem:

Editing running sequences can lead to sequence abortion. The preconditions are:
- The first pending line is edited, that includes inserting new lines.
- The sequence update is not processed when the current run ends.

Temporary Solution:

N/A

Fix Information:

This is fixed in Thermo Fisher SII 1.3


KPR#:804696  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_OLCDS_V1.0.22.1

Keyword: Performance

One-line Description:

PE Clarus 680: Attenuation Time Change Events Do Not Work Properly

Problem:

Timed attenuation changes in the run cause variable peak areas and heights throughout the run, yet the ratio of the peaks remains consistent. 

Temporary Solution:

None

Fix Information:

Fixed in PE Clarus GC Control 2.0


KPR#:1029458  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2,1.3

Keyword: Performance

One-line Description:

Thermo Fisher SII: Take/Release Functionality Leads to an Unhandled Exception

Problem:

This happened sporadically with Ultimate HPLC but possibly happens with other instruments in a client/server environment.
Repetitive use of Take/Release functionality for the same instrument on different machines leads to an unhandled exception:
An unexpected error occurred and we cannot continue...

Temporary Solution:

Avoid using the Take/Release functionality if this happens on your system.

Fix Information:

The issue is fixed in OpenLab CDS 3.0


KPR#:1058303  Product:    uGC Driver - 2.4.8

Keyword: Performance

One-line Description:

Injector temp Error happened when method switch from high temp. to low

Problem:

Injector temp Error happened when method switch from high temp. to low

Temporary Solution:

n/a

Fix Information:

Fixed in driver 2.5.5


KPR#:1064776  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.3

Keyword: Performance

One-line Description:

Thermo SII 1.3 with CDS 2.8: Issue about log4net in Station IQ and some Thermo instruments cannot be configured

Problem:

Running a Station IQ on a system with SII 1.3 and OpenLab CDS 2.8 shows an error about a wrong version of log4net being installed.

This results in the hanging of the application upon configuration of some Thermo  instrument.

Temporary Solution:

N/A

Fix Information:

This has been addressed in OpenLab CDS 2.8 Update 01. This is also resolved in Thermo SII control 1.4.


KPR#:1073349  Product:    uGC Driver - 2.4.8

Keyword: Performance

One-line Description:

Sequence stalling and the time in log back to 1601 year

Problem:

Sequence stalled, check the log file, the injection date is back to year 1601 

Temporary Solution:

n/a

Fix Information:

Fixed in driver 2.5.5


KPR#:1076815  Product:    uGC Driver - 2.4.8

Keyword: Performance

One-line Description:

Run can't be aborted due to Auxiliary trace enabled

Problem:

abort a single run in the middle of time, state of Run Quene turns to Aborting untill the rest time of run consumed, the state will turn to Aborted completely, however, the uGC has been stopped and into idle state already.

For example,  run time of method is 600 sec, abort the run when running at 300 sec, uGC has been stopped at 300 sec and turn to idle, but CDS is kept Aborting state for another 300sec, it will wait for 600 sec completed then the run could be aborted.

The cause we found, if uncheck the Auxiliary Trace in Configuration, abort a run could get aborted in the quene in time, no need to wait for the rest of run time completed.

Temporary Solution:

n/a

Fix Information:

fixed in driver 2.5.5


KPR#:1087709  Product:    uGC Driver - 2.5.2

Keyword: Performance

One-line Description:

Unable to configure instrument in ChemStation

Problem:

Unable to configure instrument in ChemStation

Temporary Solution:

n/a

Fix Information:

fixed in driver 2.5.5


KPR#:1378700  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.4

Keyword: Performance

One-line Description:

Thermo Fisher SII: SQL engine not starting after installation on Win 11 and Win Server 2022 OS

Problem:

During SII for OpenLab installation on Windows 11 or Windows Server 2022, the SQL Server Database Engine fails to start. The installer proceeds, several package install then fail, and the final configuration step fails and errors messages will be found in the SQL Server error log or in the Windows Application Event Log. ?
SQL error symptoms may include:
* 0x851A001A with “Wait on the Database Engine recovery handle failed.”
* Error log entries about misaligned log I/O falling back to synchronous I/O.
* Application faults (e.g., sqlservr.exe faulting in ntdll.dll, exception 0xc0000005).
* LocalDB message about master files formatted for 4 KB now on a 16 KB sector-size volume.

This happens for any version of SQL Server including LocalDB. ?

Root cause: some Windows 11/Server 2022 storage stacks (notably NVMe) expose a physical sector size > 4 KB, which SQL Server does not support (supports 512 B & 4 KB only), preventing startup/recovery.

Temporary Solution:

There are several workarounds:
* Use a drive with supported sector size (512 B or 4 KB) for SQL database files; select that location after SQL installation. 
* Force 4 KB emulation via registry key (Windows 11/Server 2022 NVMe driver):
* Start SQL Server with trace flag 1800 to force 4 KB sector behavior for I/O.

Fix Information:

This is a Microsoft issue. There is no fix for it yet. Use a workaround described below.


KPR#:227940  Product:    A.04.08 HF01 PP1

Keyword: Project Configuration

One-line Description:

 "Edit running method" privilege has no effect on user privileges

Problem:

The privilege “Edit running method” does not have the expected impact on users in OpenLab CDS EZChrom. It should allow editing of the running method and without this privilege editing should be prevented. Editing of the running method is always possible.

Temporary Solution:

Use the "Manual Control" privileges to establish the necessary administration

Fix Information:

A fix is under investigation for a later release


KPR#:149838  Product:LC and CE Drivers  RC 1200.NET  A.02.05

Keyword: Pump Driver

One-line Description:

Misrepresentation of Bottlefillings in Dashboad of PumpValveCluster

Problem:

When a Valve is configured for Channel A in PVC the visual presentations of the BottleFillings of ChannelB, ChannelC and ChannelD are wrong.
It will not show the BottleFillings of the corresponding Channel but the BottleFilling of Channel A Valve Pos 2 etc.

Temporary Solution:

none

Fix Information:

Will be fixed with the Rc.NET Drivers Rev . A.02.06


KPR#:149845  Product:LC and CE Drivers  RC 1200.NET  A.02.07

Keyword: Pump Driver

One-line Description:

PumpValveCluster with ISET - Solvent composition: resulting %A is wrong (at 1 line in timetable)

Problem:

1. Working with a G4220A pump with ISET enabled.
2. The emulated pump was: G1312B no mixer.
3. Set up a timetable.
4. Added a G1160A valve into the system.
5. Reconfigured the instrument and setup a pump-valve-cluster.
6. During startup selected "Save instrument settings in new method".
7. Enabled ISET and selected G1311A pump this time.

-> the timetable of previous configuration still exists, but on one line "95.1%A and 5%B" is displayed instead of "95%A, 5%B

Temporary Solution:

Select Method -> New Method

Fix Information:

Solved with RC.Net A.02.07


KPR#:25  Product:OpenLAB Intelligent Reporting  OpenLAB Intelligent Reporting  A.01.04

Keyword: Report Template Design

One-line Description:

Intelligent Reporting : Last displayed digits might be cut off, in cases where the number is zero

Problem:

If the number formatting scheme "1234.000" is used, the selected number of digits is not shown in case the last digit is zero. 
A number such as 1.340 will be displayed as 1.34, which does not contain the same information as initially intended.

This issue affects also the rounding mechanism used for peak labels.

Temporary Solution:

The formatting scheme "1,234.000" can be used as this scheme is not affected by the defect.

For chromatogram labeling (e.g. retention time), the expression 
FORMATNUMBER(Fields!PEAK_RETENTIONTIME.Value,3,True,False,True)
can be copied to the expression editor of the peak label.

The FORMATNUMBER command is not available in the expression editor, but still a valid and supported command. 
It can also be applied to other data fields containing numbers.

Fix Information:

Issue fixed in OpenLAB CDS 2.1


KPR#:40  Product:OpenLAB Intelligent Reporting  OpenLAB Intelligent Reporting  A.01.03

Keyword: Report Template Design

One-line Description:

Missing named peaks are reported in Intelligent Reporting for all the channels or signals

Problem:

Missing named peaks are reported for all channels, even if the peak identification table lists only a single channel for this named peak (compound).
By design named peaks (compounds) are linked to a peak result in the XML file. If no peak is found the named peak has no link to a signal and is reported as 'not found' in any of the signals.

Temporary Solution:

N/A

Fix Information:

This issue is fixed in OpenLAB CDS 2.1 see comment:


KPR#:42  Product:OpenLAB Intelligent Reporting  OpenLAB Intelligent Reporting  A.01.07

Keyword: Report Template Design

One-line Description:

The Fixed Text field in IR Templates does not accept Chinese characters

Problem:

Input any Chinese character to the item "text" of Special objects,  will disappear after entering.

Temporary Solution:

The direct input does not work. As a workaround, access the value of the field via Properties. Then open the fx dialog.
Use the syntax
="Whatever text you like @#$%"

Local language characters can be entered in place of the text inside the quotes.

Fix Information:

Fixed in OpenLAB revision A.02.01.


KPR#:50  Product:OpenLAB Intelligent Reporting  OpenLAB Intelligent Reporting  A.01.07 HF01

Keyword: Report Template Design

One-line Description:

Intelligent Reporting - Spaces are stripped from filters when they are saved

Problem:

This information is relevant only for Intelligent Reporter template edition on OpenLAB CDS A.02.01 and A.02.02, ChemStation or EZChrom Edition.

The filters using a space in a string expression will appear stripped when the report template is closed and reopened. The filter itself will be still functional, but will not appear correct any longer. This is just a a wrong display of the expression in the report editor.

When editing a report template, and, using a space in a text filter expression such as : Compound_Name = "Compound 1" 
When the report template is saved, closed and then reopened, the expression will appear as following: Compound_Name = "Compound1" 

This expression will not be displayed correctly in the editor, but will be still functional with the correct "Compound 1 " expression filter when generating a report.

Temporary Solution:

Do not copy and paste the related filter without a space into another filter expression

Fix Information:

The corresponding filter expression fix is part of the OpenLAB CDS A.02.02 Service Release 2

Wait for this Service Release: 
OpenLAB CDS ChemStation Edition A.02.02 SR2. Once available, install the software service release.
OpenLAB CDS EZChrom Edition A.02.02 SR2. Once available, install the software service release.


KPR#:56  Product:OpenLAB Intelligent Reporting  OpenLAB Intelligent Reporting  A.01.05

Keyword: Report Template Design

One-line Description:

Error seen in Intelligent Report header (#error) if any related data field is used in an expression

Problem:

This note is about Intelligent Reporting template design

In the reporting editor, the report header could not used related injection data information such as "Sequence Name", "Sample name", "Vial location", or any information or results that are contained in the data file.
If the drag and drop functionality is used to placed one of this information inside the report template header, the editor will show that this is not allowed. But if a text field is added and edited with an expression like "=Sequence_Name". The report generated from this template will lead to an error message during the report generation.

Temporary Solution:

Such information cannot be displayed in any report header

Fix Information:

Will be announced in a separate note when this functionality will be available


KPR#:58  Product:OpenLAB Intelligent Reporting  OpenLAB Intelligent Reporting  A.01.06

Keyword: Report Template Design

One-line Description:

Intelligent Reporting - Stacked Chromatograms do not display the time axis in reports

Problem:

If a chromatogram display is set to show multiple chromatograms in a stacked display, the time axis is no longer displayed.  

Due to the 3D perspective effect that the sacked mode induces, the time axis is no longer displayed.

Temporary Solution:

The time can be displayed on the peak level by using the peak label functionality.

Fix Information:

n/a


KPR#:59  Product:OpenLAB Intelligent Reporting  OpenLAB Intelligent Reporting  A.01.06

Keyword: Report Template Design

One-line Description:

Intelligent Reporting - Square brackets cannot be used in filter expressions

Problem:

Intelligent reporting use Visual Basic expression in the report template edition. This note is related to the limitations of this language expressions. 
When a reporting item is filtered on a specific expression such as  Sample_Name like "Standard [1]", the filter will provide the desired effect to hide or show this specific sample name.
All expression using text condition with "[1]" will not be functional.

Temporary Solution:

Replace the all square brackets "[]" by round brackets "()".

Fix Information:

n/a


KPR#:60  Product:OpenLAB Intelligent Reporting  OpenLAB Intelligent Reporting  N/A

Keyword: Report Template Design

One-line Description:

Cannot calculate summary statistics on columns containing variable or aggregator

Problem:

When using variable or aggregator in a table column of a report template. It would not be possible to compute a summary calculation of the list of value in this column. 
The summary calculation results, such Standard deviation, or sum for example, will appear like "NaN".
This is a limitation of the Microsoft report renderer working underneath Intelligent Reporting to generate a report in OpenLAB Rev. A , ChemStation edition or EZChrom edition.

Temporary Solution:

To compute a summary calculation of a column using variable or aggregator, the table summary calculation functionality cannot be used. To be able to compute these summary calculations, a temporary solution is as following:
 
1. Save the value entered in the column value as an aggregator (Go to the "column properties", click on the expression editor "fx" in the "value" tab, and then on "Save Expression Result As")
2. Select "Aggregator", and type the name of the desired "Calculation Aggregator Name"
3. Click "ok" in all properties dialogs
4. Insert a single text field below the table from the report items tree
5. In the text field value expression, in the field tree, select the "variable" node, and click on "aggregator"
6. In this "aggregator", select the aggregator previously created, and select the summary calculation that would need to be computed (sum, avg, stdev, ...)
7. Select both report items, the single text box, and the table used, and click-right on the two selected items to select the "group" option
8. Once the group created, click-right on the frame surrounding the two report items grouped, and select properties
9. In the group properties, make sure to use the same "group repeat" than the table group repeat value (for example : Compound_Name), and validate all properties dialog
10. In the table "Peaks and Repeating" tab, for the table repeat option, select "None"

Fix Information:

This limitation is due to the Microsoft renderer. 
A new renderer is used in OpenLAB CDS 2.1, and this limitation is not observed.


KPR#:620  Product:RTE  RTE  A.02.07

Keyword: Report Template Design

One-line Description:

Error message when adding a chromatogram to a report template

Problem:

The following defect may affect the report template editor of OpenLAB CDS installations on Windows XP operating systems:

Whenever adding a chromatogram to a template or loading or applying a template with chromatograms, the following error message is shown:

Could not load file or assembly 'Agilent.ChromPlot.Infrastructure, Version=1.0.0.0, Culture=neutral, PublicKeyToken=26d0ba2f646bead1' or one of its dependencies. The located assembly's manifest definition does not match the assembly reference.

Temporary Solution:

n/a

Fix Information:

Install all updates from Windows Update for the .NET Framework 3.5 SP1 and .NET Framework 2.0 Service Pack 2 on Windows XP Service Pack 3 and Windows Server 2003 Service Pack 2 , in particular the security update 2518864 (MS11-044). 
Visit  http://update.microsoft.com to download and install the mentioned updates.


KPR#:621  Product:RTE  RTE  A.02.07

Keyword: Report Template Design

One-line Description:

Internal grouping of tables and usage of calculation variables cannot be applied at the same time

Problem:

Internal grouping of tables and usage of calculation variables cannot be applied at the same time.

In cases where the table contains an expression that is saved as a calculation variable it is not possible to define a table internal group. Whenever you setup a group it is not saved and not applied.

In cases where a grouped table gets extended by column that defines a new expression saved as a calculation variable, the expression is not calculated and the variable is not saved when previewing the report.

Temporary Solution:

In order to apply statistical calculations (average, standard deviation, RSD, etc.) on calculated results, the results need to be stored in an array rather than a single variable.
Consult the report template editor on-line on how to use category aggregators to define a result array.

Fix Information:

There is currently no fix planned.


KPR#:622  Product:RTE  RTE  A.02.01 M3

Keyword: Report Template Design

One-line Description:

Summary Calculations cannot be applied to table columns containing calculation variables

Problem:

Enhanced calculations often require storing results in variables to be used in further calculations.
If a table contains a calculation expression with previously defined variables, the statistical summary calculations on the column return "NaN".

Temporary Solution:

In order to calculate statistics on table columns including calculation expressions with variables, the results of this line-by-line calculation need to be stored in an array for further processing.
Consult the on-line help for the usage of "category aggregator" to set-up such an array.

Fix Information:

There is currently no fix planned.


KPR#:623  Product:RTE  RTE  A.02.07

Keyword: Report Template Design

One-line Description:

Adding a template description to a previously erased and saved description field causes a KeyNotFoundException

Problem:

Whenever a template description gets deleted from the template properties dialog a new description can't be entered if the template has been saved with the empty description field.

The user will see an error message:

--- System.Collections.Generic.KeyNotFoundException ---
   at System.ThrowHelper.ThrowKeyNotFoundException()
   at System.Collections.Generic.Dictionary`2.get_Item(TKey key)

Temporary Solution:

It is recommended to have the template description field always being populated with text.

Fix Information:

There is currently no fix planned.


KPR#:627  Product:RTE  RTE  A.02.04

Keyword: Report Template Design

One-line Description:

Customized composite groups should have a unique name

Problem:

Whenever naming a customized composite groups with the name of an existing, pre-defined snipped, the composite group name cannot not be deleted from the tree view.
The composite group itself is deleted, but name remains under the composite group fields.

Temporary Solution:

Use an individual and unique name for a customized compound group.

Fix Information:

There is currently no fix planned.


KPR#:662  Product:OpenLAB ECM Intelligent Reporter  G4635AA  A.02.03

Keyword: Report Template Design

One-line Description:

Filter definition does not always allow usage of OR operators

Problem:

The usage of OR operators in report filter conditions is only allowed when combining several different report parameters to one filter expression.
In cases where a report parameter may fulfill multiple filter conditions (e.g. Instrument Name = Instrument A or Instrument B), the report template editor does not allow using an OR operator.

This is a limitation due to the definition of the operators available for filtering in report rendering (Microsoft Reporting Services). http://msdn.microsoft.com/en-us/library/ms154029(v=sql.90).aspx.  In particular, there is no "OR" at all.

Temporary Solution:

The exception is the "IN" operator, which allows a FilterExpression to be compared against a list of FilterValues. 
The desired filtering may also be achieved by using wildcards (*) in combination with a CONTAINS condition.
For number, the usage of > and < conditions could be used to achieve the desired filtering.

For external report parameters (e.g. filtering for multiple sample types) the following JOIN statement could be used in the expression editor of the filter:
[=Join(Report_Parameter_0, " ").Contains(Sample_Type   1)    = True]   =  "True"

Fix Information:

There is currently no fix possible.


KPR#:668  Product:OpenLAB ECM Intelligent Reporter  G4635AA  A.02.03

Keyword: Report Template Design

One-line Description:

Reporting time format does not offer an option of (dd/mm/yyyy and time)

Problem:

The various time stamp formats offered by the Report Template Editor does not offer the combination of dd/mm/yyyy plus time stamp.

Temporary Solution:

The format of the time stamps can be modified by using Microsoft Business Intelligence Studio or simple a text editor, like Microsoft Notepad.

- open the report temaplte file (*.RDL)
- search the string "toString"
- Modify the following section like
      ToString("MM/dd/yyyy HH:mm:ss")</Value>
 to e.g.
     ToString("dd/mm/yyyy HH:mm:ss")</Value>

- save the RDL file and validate the changes.

Note:  The Report Template Editor cannot interpret the time stamp format correctly and will show an error filed in the editor window.  However, the time stamp format is correctly applied on the reports.

Fix Information:

RTE Support of the (dd/mm/yyyy and time) time stamp format is considered for a future relaese.


KPR#:36  Product:OpenLAB Intelligent Reporting  OpenLAB Intelligent Reporting  A.01.07

Keyword: Report Templates

One-line Description:

Min and Max column summary calculations report GMT and not local time for date fields

Problem:

Times stamps reported as part of a summary calculation, e.g. minimum or maximum of all time stamps in a table column, are reported in GMT time rather than your local time. The column values are reported in local time.

This defect is caused by a missing SYSTEM.TIMEZONE.CURRENTTIMEZONE.TOLOCALTIME(...) conversion in the .rdl template file.

Temporary Solution:

The missing function can be inserted into the template file using a text editor. This is only a temporary solution as the manual edits are overwritten by further table edits.

Fix Information:

A fix is planned for the next release of the Report Template Editor component of OpenLAB CDS.


KPR#:39  Product:OpenLAB Intelligent Reporting  OpenLAB Intelligent Reporting  N/A

Keyword: Report Templates

One-line Description:

The report date is printed incorrectly in the Agilent supplied intelligent reports

Problem:

The report date in the footer of the Agilent built-in intelligent reports incorrectly applies a conversion of the current time when a report is printed. The report time in the footer is entered as = String. Format("Printed: {0}", SYSTEM.TIIMEZONE.CURRENTTIMEZONE.TOLOCALTIME(Execution Time)). 
This is not a problem when displaying/previewing the report but only when printing the report.

Temporary Solution:

Modify the report templates to have the footer time to be =String.Format("Printed: {0}",(ExecutionTime))

Fix Information:

The built-in templates are corrected in OpenLAB CDS rev. A.02.01.


KPR#:43  Product:OpenLAB Intelligent Reporting  OpenLAB Intelligent Reporting  A.01.06

Keyword: Report Templates

One-line Description:

The template fields "SignalToNoise_EP" and "SignalToNoise6Sigma" do not work in Japanese version

Problem:

Error message “An error occurred during report rendering. The report cannot be reprocessed….”  is displayed when a user previews an IR template after adding parameters "SignalToNoise_EP" and "SignalToNoise6Sigma" in Japanese version.

Temporary Solution:

This is addressed in ChemStation C.01.05 SP1

Fix Information:

Fixed in ChemStation C.01.06.


KPR#:47  Product:OpenLAB Intelligent Reporting  OpenLAB Intelligent Reporting  2.0

Keyword: Report Templates

One-line Description:

Center of triangular calibration point labels do not represent the measured value

Problem:

The calibration curve plot offers 2 kind of triangular point labels. Unlike other applications the measured value is not represented by the center of the triangle but the the top corner or the bottom corner of the triangle, respectively.

Temporary Solution:

Consider using other point labels, like dot, rectangle or cross.

Fix Information:

For backward compatibility reasons the design of these markers will not be changed.


KPR#:48  Product:OpenLAB Intelligent Reporting  OpenLAB Intelligent Reporting  2.0

Keyword: Report Templates

One-line Description:

XLS export of intelligent reporter templates generate hidden rows

Problem:

When exporting an intelligent report to a XLS file format, the empty space on the report template gets transformed to line of zero height. 
This has no impact on XLS files used for data transfer to e.g. a LIMS system

Temporary Solution:

When opening the XLS file with an application like Microsoft Excel, the row height can be auto-adjusted by selecting the entire spreadsheet and double-clicking the split line between row indicator 1 and 2.

Fix Information:

There is currently no fix planned.


KPR#:658  Product:OpenLAB ECM Intelligent Reporter  G4635AA  A.02.04

Keyword: Report Templates

One-line Description:

Customized chromatogram or axis lables may not display correclty

Problem:

The correct syntax for using the and "&" function requires a blank space in front and behind the ampersand sign.
Example: An expression like 
=FieldName&”Response” needs to be written like 
=FieldName & ”Response”
if it is used as a label in a chromatogram plot.

Expressions like  =Str(FieldName)&”Response” are not supported in reporting service and it results in an  “#Error” in preview.

Temporary Solution:

Use a blank space in front and behind the ampersand sign.

Fix Information:

There is currently no fix planned.


KPR#:138370  Product:    A.04.07 SR2 (Tristan 3D)

Keyword: Report Templates

One-line Description:

Esignature table does not remember the formatting in Standard report

Problem:

1. Open a result set.
2. New Template - Import report - Area% report
3. Add - E-sign table - align margins as per requirement
4. Save template. Print - Found ok
5. Open next data - all margins set for e-sign table resets to default.

Temporary Solution:

Manually reset margins between result set.

Fix Information:

Currently being considered.


KPR#:138858  Product:    A.04.07 SR1

Keyword: Report Templates

One-line Description:

Only 50 .srp active template are reflecting under Report -> View option

Problem:

Cannot access (or it becomes grayed out) report template from "Report -> View" option if count reach beyond 50. System can shows only 50 as active template.

Temporary Solution:

All report template can be accessed from "File -> open -> Standard Report" dialog

Fix Information:

n/a


KPR#:241653  Product:VICI Valves  M8224BA VICI Valves  VICIValve_A.01.01,VICIValve_A.01.02

Keyword: Report Templates

One-line Description:

VICI Valves: No Serial Number and no Firmware in IR report

Problem:

There are no serial number and firmware in IR.

Temporary Solution:

The Firmware Revision is displayed on the "Information" Window in the Instrument Status Window.

Fix Information:

 IR report will be fixed with next revision.


KPR#:288580  Product:PerkinElmer GC  M8242BA PerkinElmer GC  PEClarusGC_V1.3.0.16

Keyword: Report Templates

One-line Description:

Carrier table in method report is incorrect

Problem:

Carrier table in method report displays the information for Carrier Ramp incorrectly. 

Temporary Solution:

n/a

Fix Information:

Fixed in PE Clarus GC Control 2.0


KPR#:301630  Product:    A.04.08 Update 2 Point Patch 1

Keyword: Report Templates

One-line Description:

Peaks and Repeating property in IR template does not present Missing Peaks

Problem:

Load a result set with named – peaks – some of those peak must be missing
Select any Peaks and Compounds Table snippet (Peak Results)
Select the Peaks and Repeating property of the peak result table.
Check the Missing Peaks box.
Preview the report and you will not see Missing Peaks.


Temporary Solution:

Return to IR template to make changes.
Group the "Peak Results" table
Select Group properties.
Set Group repeat value to -1.
Set Filtering properties and create the filter list.
Set the Filtered List to be Compound_Type = 1 and Compound_Type = 5  (Found and Missing.) 
Click in the Group on the table and select properties. In Peaks and Repeating uncheck all filter options > Ok
Preview results.

Fix Information:

n/a


KPR#:309865  Product:Hitachi LC  M8227AA/BA Hitachi LC  HitachiLC_EZC_June19Ed

Keyword: Report Templates

One-line Description:

No Module Information is Reported in the Sequence Summary Report

Problem:

Sequence Summary Report does not contain the Instrument Module Information.

Temporary Solution:

The information about the Instrument modules can be seen in the Standard Method report and can be printed from: File -> Print -> Configuration.

Fix Information:

This will not be fixed as a work around is available.


KPR#:138810  Product:    A.04.07 SR2 (Tristan 3D)

Keyword: Reports

One-line Description:

Versioning in the Method Report (Classic Reports)

Problem:

The following issues were found with versioning in the Method Report (Classic Reports) for OpenLAB CDS EZChrom A.04.07 SR2:

1. The path is incorrect for the report template name.

2. For a result set version the sequence version is reported instead of the result set version. There is no field that reports on result set version.

3. There is no field to report the version of the method inside the ssizip when the project is setup to add the method to the result set. Individual files inside the result set do not have a version. Only the master method is versioned. Any Method version information is only for the Master Method.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:138849  Product:    A.04.09

Keyword: Reports

One-line Description:

System hangs when trying to generate sequence summary reports with Primaide Modules

Problem:

With Primaide Modules it can happen that the system hangs when trying to generate sequence summary reports when the storage path is pointed to a non-standard drive (i.e. D drive)

Temporary Solution:

Install to the default path

Fix Information:

n/a


KPR#:224121  Product:    A.04.07 SR2 Update 08

Keyword: Reports

One-line Description:

QC Check Rounding limitation

Problem:

The rounding described below is observed for QC Check.
While generating report, CString.Format() function is being used to insert the numbers (float, double) as strings. In Format() function, "%.3f" is used to round off the double numbers. This converts 0.1 - 0.5 to 0(i.e.round down) and 0.6 - 0.9 to 1(i.e. round up).

%.3f 
   f - float
   all widths
   3 - precision

 In EZChrom most of the values are inserted in the above mentioned way. 
 This may effect most of the values in EZChrom reports.

Temporary Solution:

n/a

Fix Information:

DNF


KPR#:372183  Product:  M8244BA Thermo Fisher SII  1.0.0

Keyword: Reports

One-line Description:

SII: Wrong Entry in Instrument Module Report

Problem:

In the OpenLab CDS report, for some instrument modules there is an entry that incorrectly refers to the entire instrument with the driver name as a part number.

Temporary Solution:

N/A

Fix Information:

This is fixed in SII 1.1.
However, it only applies to newly configured instruments -> Old instruments (from SII 1.0) need to be deleted and reconfigured anew to get rid of this issue.


KPR#:414840  Product:  M8244BA Thermo Fisher SII  1.0.0.123396

Keyword: Reports

One-line Description:

An Additional Warning on Station IQ report for OpenLab CDS 2.5

Problem:

Thermo Station IQ expects an older version of Agilent SVT than the one actually installed with the CDS 2.5 release build (2.5.0.927). As a consequence, an additional warning to already existing four (described in the driver Release Notes)  is shown on the IQ report (the 5th one is for OpenLab CDS 2.5 only):
Agilent Software Verification Tool B.01.01: A newer version of the package is installed. Expected version 5.1.10.0

Temporary Solution:

n/a

Fix Information:

This has been fixed in SII 1.1.


KPR#:1370829  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.4

Keyword: Reports

One-line Description:

Thermo SII Instruments: Missing Information in Module List and Reports for Some Modules

Problem:

Some instrument modules do not display Part Number and Serial Number in the module list and reports.
This has been observed for AI/AS 1310 injectors configured with Trace 1310 GC and Ultimate DAD 3000RS detector. 

Temporary Solution:

None

Fix Information:

To be addressed in the future revision of Thermo Fisher SII driver.


KPR#:213496  Product:    A.04.07 SR2 HF01

Keyword: Reports - IR

One-line Description:

"Not-found compound" table in IR report not working

Problem:

"Not-found compound" table in IR report is not working as it is not reporting any missing or undetected peaks. 

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:242758  Product:    A.04.08,A.04.08 Update2

Keyword: Reports - IR

One-line Description:

Difference in compound group area reported on standard and intelligent reports

Problem:

There is a difference in the compound group area reported on a standard method report in OpenLab CDS EZChrom and the compound group area reported in an intelligent report. 

It appears that the issue only occurs when a high threshold value is used in the integration events.

Timed group or named group will impact on Group Area in Intelligent Report.

Temporary Solution:

Workaround is to use the current report type (either Classic or IR), and not to use the new report type, until the fix is released. Thus, the result is consistently reported from historical perspective. 

Fix Information:

n/a


KPR#:290069  Product:    A.04.08,A.04.08 Update2

Keyword: Reports - IR

One-line Description:

Difference in compound group area reported on standard and intelligent reports

Problem:

There is a difference in the compound group area reported on a standard method report in OpenLab CDS EZChrom and the compound group area reported in an intelligent report. 

It appears that the issue only occurs when a high threshold value is used in the integration events.

Timed group or named group will impact on Group Area in Intelligent Report.

Temporary Solution:

Workaround is to use the current report type (either Classic or IR), and not to use the new report type, until the fix is released. Thus, the result is consistently reported from historical perspective. 

Fix Information:

n/a


KPR#:291693  Product:    A.04.08

Keyword: Reports - IR

One-line Description:

Wrong value of Sample Type in IR - A.04.08

Problem:

1. Open result set
Initially the Run type is set as unknown. The IR report shows Sample type as “Sample” and the Standard report shows Run type as “Unknown”

2. After changing the run types to system suitability and reprocessing the result set, the standard report shows the Run type as “System suitability” however IR still reports sample type as “sample” even after refreshing the preview.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:316761  Product:    A.04.07 SR2 HF01

Keyword: Reports - IR

One-line Description:

"Not-found compound" table in IR report not working

Problem:

"Not-found compound" table in IR report is not working as it is not reporting any missing or undetected peaks. 

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:547223  Product:    A.01.12

Keyword: Reports - IR

One-line Description:

Zeroes-trimming for injection volumes lower than 1

Problem:

In report printout the injection volumes will be reported with a trimmed zero for injection volume lower than 1.

This formatting can be changed in the injection volume report item expression.

Temporary Solution:

The injection volume expression can be changed to the following to show the 0:

=IIf(First(Injection_Volume)>=0, FormatNumber(First(Injection_Volume),3,True,False,False), Iif(IsNothing(First(Injection_Volume)), "N/A", choose(First(Injection_Volume)+5,"external","Inj. program","manually", "No injector"))) + " " + Fields!Injection_VolumeUNIT.Value

Fix Information:

n/a


KPR#:709423  Product:    A.04.08

Keyword: Reports - IR

One-line Description:

Integration droplines may not be displayed on a report chromatogram when using Intelligent Report templates

Problem:

It has been observed that Integration droplines are not displayed on report chromatograms when using Intelligent Report templates if the dropline is on top of an exponential skim baseline. All other integration droplines should be displayed. 

Temporary Solution:

Use a Standard Method Report if integration droplines over an exponential skim baseline need to be displayed on a report chromatogram. 

Fix Information:

n/a


KPR#:177092  Product:    A.04.05 HF07

Keyword: Roles and Privileges

One-line Description:

Analyze tab not greyed out for those without the privilege

Problem:

User without the privilege of analyze data is able to do so as the analyze tab is not greyed out.

This is reproducible in A.04.05 HF7 and there is no way to disable 'analyze' button.

Temporary Solution:

N/A

Fix Information:

N/A


KPR#:508781  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.1

Keyword: Roles and Privileges

One-line Description:

SII: Direct Control Context Menu Disabled During a Run for User with All Permissions

Problem:

The Direct Control context menu is disabled for the user with all permissions during the run. However, those direct control options are enabled when the instrument is in Idle state.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 1.2 of Thermo Fisher SII.


KPR#:508896  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.1

Keyword: Roles and Privileges

One-line Description:

SII: Direct Control Context Menu Disabled for User with Manual Control (in Run) Permissions

Problem:

Although the user has the privilege “Manual control (in run)”, direct control is not possible neither in a run nor in idle state.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 1.2 of Thermo Fisher SII driver.


KPR#:509382  Product:  M8244BA Thermo Fisher SII  1.1

Keyword: Roles and Privileges

One-line Description:

SII: Instrument Control is Still Possible for User Without Manual Control Privileges

Problem:

A user without the privileges for manual control has all direct control options disabled in the OpenLab InstrumentStatus UI *except* the Open Instrument View option that opens the Thermo Fisher Instrument View. There, the user can change any settings of the instrument. The activity log tracks the changes, however, in some cases the computer name is logged instead of the user.

Temporary Solution:

Restrictions should be implemented via SOP to restrict users without permissions for direct control not to utilize SII Instrument View.

Fix Information:

To be addressed in the future revisions of the driver.


KPR#:933834  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2

Keyword: Roles and Privileges

One-line Description:

Thermo SII: Method "Approve" Function Does Not Allow Method Changes Anymore

Problem:

SII 1.2, Thermo Fisher Ultimate 3000 and possibly other instruments:
Using the Approve function for the Method status leads to a situation that does not allow to change the Method status again (independent of the user's privileges) because of pending changes. This happens even if the saved Approved method is just loaded without any other interaction. 

Temporary Solution:

N/A

Fix Information:

This is fixed in revision 1.3 of Thermo Fisher SII driver.


KPR#:149831  Product:LC and CE Drivers  RC 1200.NET  

Keyword: Sampler Driver

One-line Description:

LC driver starts run with ALS Chiller turned off

Problem:

Using the RC.NET Drivers to control the Autosampler Chiller the systems is injecting with the Autosampler Chiller turned off. This is a different behavior compared to the legacy Rapid Control drivers for EZChrom which forced an ON of the Autosampler Chiller.

Temporary Solution:

n/a

Fix Information:

This issue has been fixed with driver release A.02.01 SP1.


KPR#:149841  Product:LC and CE Drivers  RC 1200.NET  A.02.04

Keyword: Sampler Driver

One-line Description:

ICF A.01.04 - No method download, if method contains pretreatment/injector program steps.

Problem:

Methods including pretreatment/injector steps are not downloaded to the instrument, prior to a run.

Temporary Solution:

n/a

Fix Information:

The defect is fixed with the Driver Release A.02.06

In relation to the Instrument Control Framework, a PointPatch is available.
Please contact your Agilent Support Team.


KPR#:149880  Product:LC and CE Drivers  RC 1200.NET  A.02.06 SP1

Keyword: Sampler Driver

One-line Description:

For overlapped injections in vial samplers G1313A and G1329A/B, injection volumes and sample locations are not updated synchronously in the status UI

Problem:

For overlapped injections in vial samplers G1313A and G1329A/B, injection volumes and sample locations are not updated synchronously in the status UI. This means that the second sample of an overlapped injection is drawn up already and the UI displays the injection volume of that sample but still the location of the sample which is currently being analyzed is shown

Temporary Solution:

n/a

Fix Information:

With LC driver A.02.10 the UI displays injection volume and sample location of the sample currrently being analyzed.


KPR#:226629  Product:    A.04.07 SR1 Update 1

Keyword: Sequencing

One-line Description:

Error message: Unable to open method (2) when the sequence is started

Problem:

1. Submit sequence1, the sequence acquisition gets completed successfully and the resultset is saved to ECM. T
2. Reprocesses the resultset, successfully.
3. Launch the instrument online again. The run queue shows previous sequences as completed successfully. Nothing is pending in the run queue.
4. Submit sequence2. 

An error message of the below type is displayed on the screen and after clicking on OK the current sequence acquisition gets started:

(BUILT-IN\<User_Name>);Error;Instrument;Unable to open method (2):

This error message does not affect the current running sequence and all runs get completed for sequence2.seq. This error does not have any impact on data acquisition. However, the error is displayed in system activity and instrument activity logs.

The method mentioned in the error message is not used in the current running sequence2. This was the method used in the previous sequence which was acquired succesfully on this instrument. This resultset is succesfully uploaded to ECM as well.

This is not a workflow, instrument, client, sequence or method specific. It has been observed with multiple instruments, sequences and on different AIC's.

Temporary Solution:

This error message does not affect the current running sequence .

Fix Information:

Under investigation and fix will be considered.


KPR#:236585  Product:    A.04.05 HF08,A.04.08

Keyword: Sequencing

One-line Description:

Sequence Table Column width, reduced making column invisible cannot be restored

Problem:

If a column width in the sequence table is reduced to the point that the column disappears, it cannot be recovered using properties, creating a new sequence or loading a sequence. The properties are persistent.

Affects A.04.05 to A.04.08 and possibly newer.

Temporary Solution:

Workaround: Delete the instrument and create a new one. This will break the persistent column width changes.

Fix Information:

n/a


KPR#:266379  Product:    A.04.09 Update 02,A.04.07 SR2 Update 09

Keyword: Sequencing

One-line Description:

Saving Easy Sequence without change results in "Filed to upload..."

Problem:

Topology: Client Server setup with terminal server with ECM 3.5 Update 2 backend (With AFS_With GATE)

For Easy sequence setup:
Scenario: Without making changes, save Easy sequence setup:
1.Create Easy sequence setup & save it
2.Close Easy sequence setup window.
3.Open Easy sequence setup window again, open created easy sequence setup
4.Save it without making any changes by clicking on save easy sequence setup button.
5.Again save it without making any changes this time displayed is a popup message; “Failed to upload “filename” file to ECM. ECM exception “Unable to save the file to content management.”

Observation:
If no change to Easy Sequence file, subsequent saves without making any changes, even close & open again fresh window,  displays popup message; “Failed to upload “filename” file to ECM. ECM exception “Unable to save the file to content management.”

The Easy Sequence file will increment only if the file is updated with changes.


Temporary Solution:

n/a

Fix Information:

DNF


KPR#:22101  Product:Waters LC  M8239AA Waters Alliance HPLC Driver  6.1.2.1

Keyword: User Interface

One-line Description:

The alignment of the instrument module icons on the Instrument status display within OpenLAB CDS Chemstation Edition are not correctly aligned and in some cases, may be partially obscured.

Problem:

When viewing the instrument status, the various icons representing the Alliance system and connected modules may not be completely aligned within the status window. The Detector icon may also appear partially obscured or cut off .

Temporary Solution:

n/a

Fix Information:

This is a cosmetic issue. Functionality is still available. Will not be fixed.


KPR#:22136  Product:Waters LC  M8239AA Waters Alliance HPLC Driver  6.1.3

Keyword: User Interface

One-line Description:

Comment field is not available in the Instrument setup dialog for the 2998 PDA

Problem:

The Comments field is not fully visible and text can not be entered as a comment for the PDA 2998.

Temporary Solution:

Use another Comments field from another Alliance module

Fix Information:

n/a


KPR#:138857  Product:    A.04.08

Keyword: User Interface

One-line Description:

In Windows 10 Data Appearance Properties Schemas cannot be saved.

Problem:

In Windows 10 the Chromatogram Appearance Properties can not be saved as a new or existing Schema.

Temporary Solution:

On systems where it is feasible in terms of security risk, the User Account Control (UAC) policies for the entire system could be modified temporary for saving a new schema: 
Control Panel->Administrative Tools->Local Security Policy->Security Settings->Local Policies->Security Options->User Account Control: Run all administrators in Admin Approval Mode (Disable this option)
After saving the schema the User Account Control (UAC) can be set back to enabled as this is more safe.

Fix Information:

A fix is planned for a later revision


KPR#:146965  Product:Thermo Fisher LC  M8236AA/BA Thermo Fisher Surveyor/Accela  ThermoLC 4.1.2.1

Keyword: User Interface

One-line Description:

Tray image for 'Custom 384 well settings' is not displayed.

Problem:

The tray image for the 'Custom 384 well settings' is not displayed.

Temporary Solution:

n/a

Fix Information:

There is no fix.


KPR#:146976  Product:Waters LC  M8505AA/BA Waters ACQUITY/Alliance  Alliance 6.1.2.1

Keyword: User Interface

One-line Description:

The alignment of the instrument module icons on the Instrument status display within OpenLAB CDS Chemstation Edition are not correctly aligned and in some cases, may be partially obscured.

Problem:

When viewing the instrument status, the various icons representing the Alliance system and connected modules may not be completely aligned within the status window. The Detector icon may also appear partially obscured or cut off .

Temporary Solution:

n/a

Fix Information:

This is a cosmetic issue. Functionality is still available. Will not be fixed.


KPR#:146978  Product:Waters LC  M8505AA/BA Waters ACQUITY/Alliance  Alliance 6.1.2.1

Keyword: User Interface

One-line Description:

Waters Alliance: The elapsed run time dislayed on the instrument dashboard is not updated when a UV2489/PDA2998 detector is connected (ChemStation Edition only)

Problem:

In some cases, the elapsed run time value in the instrument status UI is not updated when running a Waters Alliance system with 2489 or 2998 PDA detector and OpenLAB CDS ChemStation Edition.

Temporary Solution:

n/a

Fix Information:

Fixed in driver 6.1.3


KPR#:146990  Product:Waters LC  M8505AA/BA Waters ACQUITY/Alliance  Alliance 6.1.2.1

Keyword: User Interface

One-line Description:

Improved handling of spectrum acquistion wavelength paremeters requried in user interface.

Problem:

Waters Alliance detector (2998, 996) configured with OpenLab EZChrom/OpenLab CDS.
The customer receives a displayed warning indicating "Out of Range- please enter a value divisible by 1.2" warning when entering certain spectrum acquisition wavelength parameters. It is not obvious what the corrective action should be from the customer viewpoint.
To resolve this warning, the customer has to, by trial and error, determine and enter a valid wavelength divisible by 1.2nm which is both confusing and tedious..

Background/Improvement
It has been determined the optical wavelength bandwidth for these detectors is 1.2nm, so the logic in use is correct. The user interface for entering both the start wavelength and end wavelength for defining the spectrum acquisition is now improved to pre-determine what usable wavelength values can be selected and provide feedback to the user via the same user interface.

Temporary Solution:

n/a

Fix Information:

Improved in driver version 6.1.3.


KPR#:146998  Product:Waters LC  M8505AA/BA Waters ACQUITY/Alliance  Alliance 6.1.3

Keyword: User Interface

One-line Description:

Comment field is not available in the Instrument setup dialog for the 2998 PDA

Problem:

The Comments field is not fully visible and text can not be entered as a comment for the PDA 2998.

Temporary Solution:

Use another Comments field from another Alliance module

Fix Information:

n/a


KPR#:257130  Product:  M8244BA Thermo Fisher SII  1.0.0

Keyword: User Interface

One-line Description:

Context Menu Entries in Instrument Status Appear Active During Run

Problem:

Some context menu entries may appear as active, i.e. not grayed out during a run. Selecting such a menu entry does not trigger the selected action. Outside of the run the menu entries work as expected.

Temporary Solution:

n/a

Fix Information:

This has been fixed in SII 1.1.


KPR#:257664  Product:  M8244BA Thermo Fisher SII  1.0.0

Keyword: User Interface

One-line Description:

No Online Help for Thermo Fisher Instruments Available in Status Dashboard

Problem:

The online help for Thermo Fisher instruments is not accessible via the status dashboard.

Temporary Solution:

Open Instrument View and access online help from there.

Fix Information:

This issue does not happen in SII 1.3


KPR#:391501  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.0

Keyword: User Interface

One-line Description:

Column Oven Pictogram Displays the Oven is ON in case of a door opened condition even if it is OFF

Problem:

The actual condition of column oven is OFF when the oven compartment door is opened, but, the pictogram of the oven shows it as ON.

Temporary Solution:

n/a

Fix Information:

This is fixed in revision 1.1 of the driver.


KPR#:415443  Product:  M8244BA Thermo Fisher SII  1.0.0.123396

Keyword: User Interface

One-line Description:

Entries in Method Audit Trail Are Unclear

Problem:

Entries in the method audit trail are sometimes unclear making the older method reconstruction impossible.

Temporary Solution:

n/a

Fix Information:

This has been fixed in SII 1.1.


KPR#:489974  Product:  M8244BA Thermo Fisher SII  1.1

Keyword: User Interface

One-line Description:

SII: Configuring The First Instrument for SII Might Display Two Modules, One being Empty

Problem:

After configuration of SII instrument, two Thermo Fisher instrument modules exist, one of them being empty. Launching such an instrument displays an Initialization Error. This only happens on the freshly set up system and when configuring the instrument for the first time. Reconfiguring the same instrument does not result in such error.

Temporary Solution:

After removing the empty item, the instrument can be launched without issues.

Fix Information:

To be addressed in the future revisions of the driver.


KPR#:502441  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.1

Keyword: User Interface

One-line Description:

Using Non Unique Method for Identical Instruments Prompts to Save Changes Even When No Changes Exist

Problem:

If the same method is opened on another instrument of identical (same) configuration, the system might prompt to save the changes in the method even if there are no changes in the method.
To avoid the “unsaved changes” dialog upon the run submission, use a unique method developed for each instrument.

Temporary Solution:

n/a

Fix Information:

This limitation of the SII driver is now documented in the driver documentation that comes with SII 1.4. Create methods unique to every instrument and do not use the identical methods interchangeably, even if the instrument configuration is identical.


KPR#:508118  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.1

Keyword: User Interface

One-line Description:

SII: Smart Startup, Smart Shutdown, Smart Standby Not Supported

Problem:

Smart Startup/Shutdown/Standby parameters are available in the method UI. However, these features are currently not supported in SII for OpenLab CDS. Method Audit trail does not rack the changes of these parameters either. But if changes are made in this section, Audit Trail generates a new file version without changes. 

Temporary Solution:

Instead of Smart shutdown, a dedicated shutdown method must be used at the end of the sequence or after a single run.
For the shutdown method that turns off the detector at the end of the run, in the Script Editor section of the method, perform the following changes:
1.	Delete all acquisition command entries under the Start Run and Stop Run sections. E.g., UV.UV_VIS_1.AcqOn and UV.UV_VIS_1.AcqOff and for all other data acquisition channels.
2.	Add the following commands to turn off the detector in the Stop Run section. For Example, UV.Visible_Lamp with a value "Off" and UV.UV_Lamp with a value "Off"

This method can be used as the last line of a sequence to shut down the detector lamp, or as a standalone Shutdown Method run from the Run Queue.

Fix Information:

Will not be fixed in SII as the Smart Startup/Shutdown feature is not supported in general. This is noted in the driver documentation.


KPR#:508900  Product:  M8244BA Thermo Fisher SII  1.1

Keyword: User Interface

One-line Description:

SII: Method Can Still Be Edited But Not Saved Without Create and Edit Acquisition Method Permission

Problem:

A method can still be edited by a user without the privilege to create and edit acquisition methods. It correctly cannot be saved after that.

Temporary Solution:

This does not affect the method since the Save function is not available in such case.

Fix Information:

Working as designed.


KPR#:664364  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2

Keyword: User Interface

One-line Description:

Thermo Fisher Trace GC: Method Parameters Out of Range Warnings Not Properly Displayed

Problem:

In Trace 1600/1610  GC, some out of range method parameters warnings are not properly displayed.
Sometimes the field is not colored red, UI is not reactive anymore, but without warning it is not obvious why. There is no warning dialog when clicking to another field.

Temporary Solution:

No workaround except thoroughly checking all fields for proper value entries. Use Help to see the ranges.

Fix Information:

To be addressed in the future revisions of the driver.


KPR#:667210  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2

Keyword: User Interface

One-line Description:

Thermo Fisher Trace GC: AI/AS 1600 Autosampler Missing Info in Module List

Problem:

For Trace 1600/1610 GC with an AI/AS 1600 autosampler, the Module List misses the details on the autosampler. Also, the sampler misses the tooltip showing the type of the module in the Status UI dashboard.

Temporary Solution:

N/A

Fix Information:

To be addressed in the future revisions of the driver.


KPR#:667219  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2

Keyword: User Interface

One-line Description:

Thermo Fisher Aquion: Imperfect Method Parameters Input Validation

Problem:

The data input validation is not consistent throughout the entire method: sometimes the data is not checked for boundary conditions and error messages are missing, sometimes the number of decimal places is not verified and the method does not catch these upon saving the method; sometimes the tooltips are missing. 

Temporary Solution:

Use Help feature for clarification of the input values and generally use common sense.

Fix Information:

To be considered for the future revisions of the driver.


KPR#:667312  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2

Keyword: User Interface

One-line Description:

Thermo Fisher Integrion: Waveforms Cannot Be Edited at First Trial

Problem:

An Integrion with AS-AP sampler a method with integrated amperometry does not allow to edit the waveform under EDet Option when a waveform selector is being edited for the first time: the waveform editor opens but the new name field is read-only.

Temporary Solution:

Workaround: Select “New waveform” in the waveform selector and click “Edit”. The waveform editor opens. Close it, select the waveform to be edited again and click “Edit”. The waveform editor opens and the name field becomes editable.

Fix Information:

To be considered for the future revisions of the driver.


KPR#:675175  Product:    uGC Driver - 2.3.6

Keyword: User Interface

One-line Description:

Acquisition delay is reset to 0 after reconnecting

Problem:

Set a value of Acquisition delay in method, save method and download to GC. No problem if proceed the run.

But if close the Openlab and disconnect the GC, reconnect the Openlab the method will be reset to zero.

Temporary Solution:

n/a

Fix Information:

Fixed in Driver 2.4.8


KPR#:682958  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.1

Keyword: User Interface

One-line Description:

 Thermo SII: ICS-6000 Acquisition Method Parameter Not Correctly Shown

Problem:

Upon loading, a saved acquisition method incorrectly shows a parameter for the Suppressor. However, it shows the correct value upon starting a run. Also, a message asking to save changes to the acquisition method appears even if no changes were performed.
 

Temporary Solution:

Navigating to the script, set the current value as desired  and verify that the value is correctly set.

Fix Information:

N/A


KPR#:684251  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.1

Keyword: User Interface

One-line Description:

Thermo SII:  Not Showing Online Signals Outside a Run

Problem:

Thermo SII instruments do not display the online signal outside of the run (there is no preview of a signal). This is an instrument limitation that cannot be addressed via the driver. See a workaround below.

Temporary Solution:

Signal display can be manually started and stopped via the Instrument View. 
For that:
Open Instrument View (context menu in Instrument Status)->Select Take Control-> Settings (gear icon on top, or F8)->Select the channel to be displayed (left window)->Commands Tab (right window)->AcqOn (check)

The signal display must be turned off in order to start a run, or else an error message will appear.
To turn off the signal before starting the acquisition, select AcqOff (check).

Fix Information:

This is a limitation of the Thermo Fisher hardware and cannot be fixed in the software.


KPR#:703570  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2,1.3

Keyword: User Interface

One-line Description:

Thermo Fisher Trace GC, Integrion, Aquion: Method Part of the UI Not Accessible Due to Missing Scrollbar

Problem:

The lower part of the method setup is truncated/not accessible (even if there is no other window below) for some of the modules/tabs. There is no vertical scrollbar to access the missing part of the UI.

Temporary Solution:

Use "script editor" to adjust the method parameters not accessible due to the missing scrollbars.

Fix Information:

To be addressed in the future revisions of the driver.


KPR#:708978  Product:    uGC Driver - 2.3.6

Keyword: User Interface

One-line Description:

uGC Channel status disappear from OpenLab CDS Aquisition 

Problem:

Micro GC via Openlab CDS, the channel status doesn't appears

Temporary Solution:

n/a

Fix Information:

Fixed in driver 2.4.8


KPR#:710021  Product:    uGC Driver - 2.3.6

Keyword: User Interface

One-line Description:

Don't support Aux trace feature on Openlab CDS

Problem:

Enable the Aux trace in configuration, but the feature can't be supported in Openlab CDS.

Openlab CDS 2.7

Temporary Solution:

n/a

Fix Information:

Fixed in Driver 2.4.8


KPR#:735528  Product:    uGC Driver - 2.4.1

Keyword: User Interface

One-line Description:

Method_Acqusition delay makes all peaks move backward

Problem:

Acquisition delay in the GC method is not working correctly. 
All peaks RT time will be delayed, but actually these peaks are eluting the same time.

Temporary Solution:

n/a

Fix Information:

Fixed in Driver 2.4.8


KPR#:758653  Product:    uGC Driver - 2.3.6

Keyword: User Interface

One-line Description:

VICI valve can't configure successfully

Problem:

I use the same instrument to configure VICI valve from FW 2.1.164 to 2.02.191, and found VICI can't be configured successfully with FW 2.02.191.

Downgrade to 2.1.164, all works well.

Upgrade to 2.2.191 again, same problem.

Temporary Solution:

n/a

Fix Information:

Fixed in Driver 2.4.8


KPR#:760039  Product:    uGC Driver - 2.4.3

Keyword: User Interface

One-line Description:

No injection delay in EZ chrom 04.10

Problem:

 No 'Injection delay'  function in EZ Chrom 04.10, but there is the description of 'Injection delay' in the help file .

Temporary Solution:

n/a

Fix Information:

fixed in driver 2.5.5


KPR#:763799  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.2

Keyword: User Interface

One-line Description:

Hitachi ChromasterUltra RS: Selection in Tray A Display Also Reflects in Tray B of Microplates

Problem:

For Hitachi CMU RS instrument, when multiple racks are set, the vial selection made in the Tray display for the first rack is also reflected in the second rack. This does not affect data acquisition.

Temporary Solution:

N/A

Fix Information:

To be addressed in OpenLab CDS.


KPR#:763901  Product:Hitachi LC  M8227AA/BA Hitachi LC  Hitachi_CMU_1.2

Keyword: User Interface

One-line Description:

Hitachi ChromasterUltra RS: Selection Column by Column on Tray Display Not Supported

Problem:

For Hitachi CMU RS instrument, only row by row vial selection can be done in the Tray Display. Column by column vial selection pattern is not available.

Temporary Solution:

n/a

Fix Information:

To be addressed in the newer revision of the driver.


KPR#:801393  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2

Keyword: User Interface

One-line Description:

SII Module Status May Show As Not Ready in Idle State

Problem:

SII controlled modules may show a Not Ready status when the instrument is idle. The overall status shown in OpenLab CDS is Idle. This defect does not affect the functionality of the instrument or the data system.

Temporary Solution:

N/A

Fix Information:

This is fixed in Thermo Fisher SII 1.3.


KPR#:948202  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2

Keyword: User Interface

One-line Description:

SII module cannot be selected during configuration

Problem:

In rare cases it is not possible to select the Thermo Fisher chromatography modules (Vanquish and Ultimate instruments, possibly other) via the USB serial number during configuration.
In these cases, the driver reports a code 39 error in the Windows device manager.
The cause for this is the Windows core isolation feature.

Temporary Solution:

Deactivate core isolation on all AICs which connect to Thermo Fisher instruments. Core isolation may be managed by customer IT via group policies. 

Fix Information:

To be addressed by Thermo Fisher in the future revision of the driver.


KPR#:978882  Product:    uGC Driver - 2.4.8

Keyword: User Interface

One-line Description:

The color of Actuals for Gasifier should be updated after plug the USB cable.

Problem:

The color of Actuals for Gasifier should be updated after plug the USB cable.

Temporary Solution:

n/a

Fix Information:

fixed in driver 2.5.5


KPR#:1038023  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.2

Keyword: User Interface

One-line Description:

Truncation in Pump Limit Field Makes Input Semi-Invisible

Problem:

The input numbers are not quite visible in the Pressure Limit fields in the Instrument Setup window of the Pump (upon entering 50--only 0 is seen).  This was observed with different LC instruments and only happens in Chinese OS, ENG OS does not have this issue. The Script editor shows all numbers correctly.

Temporary Solution:

Use Script editor to view the fields instead of Method Setup window.

Fix Information:

Reported to Thermo Fisher.


KPR#:1043038  Product:Hitachi LC  M8227AA/BA Hitachi LC  HitachiChromaster_2.1.14,Hitachi_CM_2.7

Keyword: User Interface

One-line Description:

Hitachi Chromaster: Name of 2nd Channel Changes to None Outside Run in Online Signal View

Problem:

For UV 5410 in Dual Wavelength mode, the online signal window of OpenLab CDS displays the 2nd wavelength of Hitachi Chromaster 5410 UVetector name as "None" outside the run. The view of the signals is correct in the DA window.
(The online signals do not reflect the true signal when not in a run--this is a known limitation)

Temporary Solution:

None.

Fix Information:

This limitation is documented in the Operation Manual in revision 2.8 of the driver.


KPR#:1052659  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.0.0,1.1,1.2,1.3

Keyword: User Interface

One-line Description:

Dashboard shows pump as idle when running out of rear seal wash solvent 

Problem:

A Thermo Fisher pump that has run out of rear seal wash solvent will show as idle in the OpenLab dashboard while showing not ready in the instrument view.
A sequence started in this situation will abort immediately.

Temporary Solution:

Refill the rear seal wash solvent before submitting a new sequence when it is low.

Fix Information:

To be addressed in the future revision of the driver.


KPR#:1074052  Product:    PL 1.8

Keyword: User Interface

One-line Description:

Hitachi ChromasterUltra RS: Selection in Tray A Display Also Reflects in Tray B of Microplates

Problem:

For Hitachi CMU RS instrument, when multiple racks are set, the vial selection made in the Tray display for the first rack is also reflected in the second rack. This does not affect data acquisition.

Temporary Solution:

N/A

Fix Information:

To be addressed in OpenLab CDS.


KPR#:1074053  Product:Hitachi LC    Hitachi_CMU_1.2

Keyword: User Interface

One-line Description:

Hitachi ChromasterUltra RS: Tray Display Vial Format Requirements

Problem:

For multiple racks, tray display only supports numbering with dash (for example A-1, B-2). Vial numbering for previous CMU driver version uses dot (for example A.1, B.2). Selection for sequence created with previous driver version is not visible in Tray display, however, the sequence can still be performed if dot is used.

Temporary Solution:

Use dash instead of dot format in vial designation for the proper Tray display.

Fix Information:

This is a limitation in the driver, RC.NET and OpenLab CDS. 
One could either use the existing sequences without tray visualization or recreate/updated the sequences if tray visualization is to be used.


KPR#:1074165  Product:    uGC Driver - 2.4.8

Keyword: User Interface

One-line Description:

OpenLab CDS--Method save must be before download.

Problem:

OpenLab CDS--Method save must be before download.

Temporary Solution:

n/a

Fix Information:

fixed in driver 2.5.5


KPR#:1080182  Product:    uGC Driver - 2.5.x

Keyword: User Interface

One-line Description:

Temperature of Gasifier is blank when Status is not Ready

Problem:

Temperature of Gasifier is blank when Status is not Ready

Temporary Solution:

n/a

Fix Information:

fixed in driver 2.5.5


KPR#:1082987  Product:    uGC Driver - 2.5.x

Keyword: User Interface

One-line Description:

Version on "About" page not consistent with actual installed version

Problem:

Version on "About" page not consistent with actual installed version

Temporary Solution:

n/a

Fix Information:

fixed in driver 2.5.5


KPR#:1084188  Product:    uGC Driver - 2.5.x

Keyword: User Interface

One-line Description:

Difference driver version between Configuration and Windows Control Panel (On ChemStation C.01.10 Update07)

Problem:

Difference driver version between Configuration and Windows Control Panel (On ChemStation C.01.10 Update07)

Temporary Solution:

n/a

Fix Information:

fixed in driver 2.5.5


KPR#:1091615  Product:    uGC Driver - 2.5.4

Keyword: User Interface

One-line Description:

The updated value for Sampler is not displayed in reopened method.

Problem:

The updated value for Sampler is not displayed in reopened method.

Temporary Solution:

n/a

Fix Information:

fixed in driver 2.5.5


KPR#:1092105  Product:    uGC Driver - 2.5.x

Keyword: User Interface

One-line Description:

VICI 2 and VICI 3 related entries display on method audit trail even instrument has only VICI 1

Problem:

VICI 2 and VICI 3 related entries display on method audit trail even instrument has only VICI 1

Temporary Solution:

n/a

Fix Information:

fixed in driver 2.5.5


KPR#:1173424  Product:Thermo Fisher SII  M8244BA Thermo Fisher SII  1.3

Keyword: User Interface

One-line Description:

Thermo Fisher SII: Open file dialog moved to back when tooltip is shown

Problem:

The dialog to open an acquisition method "vanishes" once the tooltip opens.
This was observed with ICS-6000 but could possibly happen with other instruments.

Temporary Solution:

Click on the grey area below the method editor to bring the file selection dialog back into focus.
Do not hover the mouse above an acquisition method file name to avoid the tooltip opening.

Fix Information:

This was reported to Thermo Fisher for investigation.


KPR#:1375236  Product:PerkinElmer LC  M8229BA PerkinElmer LC  PEFlexarLC_1.0

Keyword: User Interface

One-line Description:

PE Flexar: Incorrect Driver Version in Module List

Problem:

All modules of Flexar LC display an incorrect driver version in the Module List (launched from the Instrument Status window in OpenLab Acquisition). It should be 1.0 instead of 2.0.1904.21044.

Temporary Solution:

Use SVT report for OpenLab CDS to verify Flexar driver version instead.

Fix Information:

To be fixed in the future revision of the driver.