Software Status Bulletin
Hitachi LC M8227AA/BA Hitachi LC
Known Problem Report as of Dec 8 2025 4:00AM
Preface
-------------------------
This Software Status Bulletin (SSB) documents all known problems in the software
product designated above. The SSB is derived from Known Problem Reports (KPR)
which result from user problems that have been classified as documentation
problems or software defects. When a KPR is written, an identifying number is
assigned to it, and the KPR is added to the next edition of the SSB.
User inputs that have been classified as Enhancement Requests are not documented
in the SSB. User problems that have been submitted, but that have not been
classified by the time the SSB is generated are not included in the SSB.
How to use the SSB
-------------------------
When you experience a problem with a product, first check this SSB to see if the
problem has been reported already, and if there is a temporary workaround
available for the problem, or if the problem has already been fixed by a new
revision. If the problem is not listed in this SSB then you may wish to report
it to the Response Center or to your field support representative.
To determine if your problem is documented in this SSB, first look in the
Keyword Index section of the SSB. Under each keyword is a listing of one-line
descriptions of related KPRs. If any of these sound like yours, locate the KPR
# in the Known Problem Reports section of the SSB, and read the full KPR. The
KPRs in the Known Problem Reports section are sorted by KPR #.
There are two sections in the SSB:
Keyword Index: This index is categorized by keyword. For each KPR there is a
brief description and a KPR #. A KPR may be associated with more than one
keyword.
Known Problem Reports: This section contains KPRs, with all the available
information relevant to the problem. KPRs in this section are sorted by KPR #.
Keyword Glossary
Keyword Index
KPR#:138296
On-line signal for Hitachi Chromaster is only displayed during acquisition
KPR#:356008
Hitachi LCE and CM: Cannot Use Submit Button in Run Queue to Submit Runs
KPR#:401753
No error message for missing vial
KPR#:536644
Hitachi Chromaster LC: No Injection Happens if "No injection/Instrument blank" Used as the Injection Source
KPR#:695883
Hitachi ChromasterUltra RS: Displayed Runtime Doubles for Detectors Online Plot During a Run
KPR#:589099
Hitachi Chromaster Ultra: No System Activity Log Entries Upon Vial Settings Change
KPR#:548087
Hitachi Chromaster and LaChrom Elite LC: Differences in DAD channel naming in Data Analysis
KPR#:285964
Hitachi LaChrom Elite auto configuration fails on EZChrom with Win 10
KPR#:397013
Two different pressure units displayed in data analysis
KPR#:456679
Hitachi LCE and Chromaster LC: Comma as a Digital Separator Limitations
KPR#:391787
Hitachi's Sub CPU Firmware can Only be Visible from Hitachi Maintenance Tool
KPR#:601921
Hitachi Chromaster Ultra Rs: Document for the Hitachi Maintenance Software is not the latest
KPR#:701847
Hitachi Chromaster Ultra LC: Inconsistent driver version 3.20 reported in Software Verification Tool
KPR#:350455
Hitachi Chromaster can't Initialize after AIC reboot with "IFC cannot be found" error
KPR#:658322
Hitachi LC: Incorrect Display of Parameters in Maintenance Tool with Win 11
KPR#:708791
Hitachi LC: USB Drivers incompatible with OS Memory Integrity Option
KPR#:762177
Hitachi LC: USB Drivers incompatible with OS Memory Integrity Option
KPR#:1077556
Hitachi LC: Incorrect Display of Parameters in Maintenance Tool with Win 11
KPR#:592464
Hitachi Chromaster Ultra: Help Clarification for the Purge Options of the Pump
KPR#:592467
Hitachi Chromaster Ultra: Help Clarification to Rack Settings of Autosampler
KPR#:595483
Hitachi LaChrom Elite: Incorrect Error Message for Analog Input Device Issue
KPR#:147027
Uninstalling Hitachi Drivers may lead to failing SVT Report
KPR#:401780
"Switch Valve to Load" not available after "Switch Valve to Inject"; "Home arm" action needed first
KPR#:492287
Hitachi LC: Long Launching Time if Instrument Setup Window Left Open Prior To Exiting Instrument Window
KPR#:536651
Hitachi LaChrom Elite: Intermittent Waiting For Trigger Does Not Start the Run
KPR#:1129855
Hitachi Chromaster: Script Error Appears Under Certain Conditions
KPR#:738391
Hitachi Chromaster: Acquisition Window Crashes if Non-English OS Used for AIC/Client
KPR#:695899
Hitachi Chromaster Ultra LC: Direct Control "Turn on at" Functions Properly Only at UTC Time Zone Entry
KPR#:751372
Hitachi CMU: Method Resolution Error When Upgrading from CMU1.0 to CMU1.1
KPR#:781493
Hitachi Chromaster LC: Actual Run Time Can Be 10x Longer Than Set in the Method Due to Certain System Settings
KPR#:414461
Incorrect Absorption Value in the Online signal and Status of 6420 UV-VIS Detector When Non-Default Range Value Used.
KPR#:415474
Hitachi ChromasterUltra Rs: DAD Online Signal Wavelength Needs to be Selected in DAD Direct Control Option
KPR#:511004
Refresh period of the Online Plot Might Increase for Long Sequences
KPR#:658297
Hitachi Chromaster LC: Long Acquisition Launch Time
KPR#:309865
No Module Information is Reported in the Sequence Summary Report
KPR#:391501
Column Oven Pictogram Displays the Oven is ON in case of a door opened condition even if it is OFF
KPR#:763799
Hitachi ChromasterUltra RS: Selection in Tray A Display Also Reflects in Tray B of Microplates
KPR#:763901
Hitachi ChromasterUltra RS: Selection Column by Column on Tray Display Not Supported
KPR#:1043038
Hitachi Chromaster: Name of 2nd Channel Changes to None Outside Run in Online Signal View
Keyword:
One-line Description:
Need to verify that Hitachi registry key is deleted after Hitachi Drivers Are Uninstalled
Problem:
If the Hitachi drivers installed with the OpenLab CDS EZChrom A.04.07 are uninstalled, please verify the registry key below is deleted from the registry after the uninstall. HKEY_LOCAL_MACHINE\SOFTWARE\Wow6432Node\Hitachi\Analytical
Temporary Solution:
Manually delete the registry key if it persists after the driver uninstallation: HKEY_LOCAL_MACHINE\SOFTWARE\Wow6432Node\Hitachi\Analytical
Fix Information:
There is no fix planned.
Keyword:
One-line Description:
UV/Vis Chromaster Ultra RS detector displays dual channel mode only during the run
Problem:
If a dual acquisition mode for the UV/Vis detector is selected in the method, the instrument status window for the detector shows only a single channel A until the run is started. On starting the run, both channel A and B are properly displayed in the Instrument Status window.
Temporary Solution:
n/a
Fix Information:
This will not be fixed due to technical difficulties of implementation. This is documented as a driver limitation in the User's Guide.
Keyword: Acquisition
One-line Description:
On-line signal for Hitachi Chromaster is only displayed during acquisition
Problem:
The behavior of the Hitachi Chromaster differs from the other vendors' instruments: - The on-line signal for Hitachi Chromaster is only displayed during acquisition - The vertical red marker is not displayed at the start of acquisition in the on-line signal window. - The collected signal is only visible during the time of acquisition and disappears from the view upon completion of a run. However, it can be viewed from the Data Analysis after the acquisition.
Temporary Solution:
Use Data Analysis to view the chromatogram
Fix Information:
There is no fix
Keyword: Acquisition
One-line Description:
Hitachi LCE and CM: Cannot Use Submit Button in Run Queue to Submit Runs
Problem:
Hitachi (LaChromElite and Chromaster) LC does not start a single run from a Run Queue window via "Submit" button in OpenLab EZChrom edition.
Temporary Solution:
Submitting via "Advanced" option in the Run queue works.
Fix Information:
This was fixed with EZChrom A.04.09 Update 06.
Keyword: Acquisition
One-line Description:
No error message for missing vial
Problem:
As expected, when “Sense missing vial” option is not checked, the run/sequence will not abort in case of a missing vial, but no error message will inform the user about the issue. The instrument performs an acquisition with injecting air and a dummy acquisition data is generated.
Temporary Solution:
n/a
Fix Information:
Fixed in revision 1.1 of the driver.
Keyword: Acquisition
One-line Description:
Hitachi Chromaster LC: No Injection Happens if "No injection/Instrument blank" Used as the Injection Source
Problem:
If a Hitachi Chromaster LC Instrument is used without Autosampler, and "No injection/Instrument blank" used as the injection source, the instrument is waiting for trigger. No injection happens.
Temporary Solution:
N/A
Fix Information:
Fixed in driver v2.6
Keyword: Acquisition
One-line Description:
Hitachi ChromasterUltra RS: Displayed Runtime Doubles for Detectors Online Plot During a Run
Problem:
In case the Hitachi ChromasterUltra RS is configured with 2 detectors of the same type, the Online plots of the detectors are displayed incorrectly by doubling the runtime. The data results look correct in the Data Analysis.
Temporary Solution:
Ask your Hitachi support to get the IFC Firmware 8898214-04 installed. This is documented with the driver version 1.2
Fix Information:
Hitachi fixed this defect with a new version of the IFC Firmware. Ask your Hitachi support to get the IFC Firmware 8898214-04 installed.
Keyword: Audit Trail
One-line Description:
Hitachi Chromaster Ultra: No System Activity Log Entries Upon Vial Settings Change
Problem:
A change of the vial settings ("Sense missing vial" and "Abort sequence by missing vial") under the direct control options for the autosampler is not recorded in the System Activity Log.
Temporary Solution:
n/a
Fix Information:
This is fixed in revision 1.2 of the driver.
Keyword: CDS
One-line Description:
Hitachi Chromaster and LaChrom Elite LC: Differences in DAD channel naming in Data Analysis
Problem:
Signal name in Acquisition and Data Analysis is different.
Temporary Solution:
N/A
Fix Information:
Hitachi improved the signal names to match the naming in OpenLab Data Analysis in driver revision 2.7.
Keyword: Configuration
One-line Description:
Hitachi LaChrom Elite auto configuration fails on EZChrom with Win 10
Problem:
When configuring a Hitachi LaChrom Elite instrument on OpenLab EZChrom Edition 04.09 system with Windows 10, the Auto Configuration functionality may lead to an error.
Temporary Solution:
n/a
Fix Information:
This is fixed in the Jun-2019 version of the drivers.
Keyword: Configuration
One-line Description:
Two different pressure units displayed in data analysis
Problem:
Sometimes the pressure units displayed in Data Analysis are different than in Acquisition (bar vs MPa).
Temporary Solution:
n/a
Fix Information:
Fixed in revision 1.1 of the driver.
Keyword: Configuration
One-line Description:
Hitachi LCE and Chromaster LC: Comma as a Digital Separator Limitations
Problem:
In OpenLab CDS, when comma is used as a decimal separator (instead of the period), the numbers come out incorrectly when setting up the parameters in the Instrument Setup window. E.g. 0.2 ml/min might show up as 2 ml/min. Follow the section below that specifies the conditions under which comma is supported as a decimal separator.
Temporary Solution:
N/A
Fix Information:
The comma decimal separator is supported under specific conditions only: • Regional Setting other than English or Japanese • Decimal symbol: Comma (“,”) • Digit Grouping Symbol: Period (“.”) • List Separator: Semicolon (“;”)
Keyword: Documentation
One-line Description:
Hitachi's Sub CPU Firmware can Only be Visible from Hitachi Maintenance Tool
Problem:
Hitachi LC's Main Board firmware can be visible in the Instrument Configuration report and in the Module List that is accessed from the Instrument Status window. However, Hitachi LC's Sub CPU firmware can only be accessed from the Hitachi Maintenance Tool that is used for configuring the instrument.
Temporary Solution:
n/a
Fix Information:
Fixed in revision 1.1 of the driver.
Keyword: Documentation
One-line Description:
Hitachi Chromaster Ultra Rs: Document for the Hitachi Maintenance Software is not the latest
Problem:
The "CM6000Maintenance" document in the driver package is from EZChrom platform, not OpenLab CDS. The Hitachi driver information is still valid except a few mentions of EZChrom.
Temporary Solution:
n/a
Fix Information:
The document is replaced by the proper one in the driver revision 1.2.
Keyword: Documentation
One-line Description:
Hitachi Chromaster Ultra LC: Inconsistent driver version 3.20 reported in Software Verification Tool
Problem:
When the Hitachi CMU driver version 3.19 is installed, it is shown * as version 3.19 in Windows Program and Features (correct) * as version 3.20 in Software Verification Report (false) This issue has no impact on the driver usage.
Temporary Solution:
N/A
Fix Information:
This was fixed with the newer driver version 3.22-- now the version is reported consistently.
Keyword: External Component
One-line Description:
Hitachi Chromaster can't Initialize after AIC reboot with "IFC cannot be found" error
Problem:
With Hitachi Chromaster v7.21 with OpenLab CDS EZChrom A.04.08/09 rebooting the AIC and launching the instrument on-line cannot initialize the instrument: All modules come up in red color and the error message "IFC cannot be found" is displayed.
Temporary Solution:
n/a
Fix Information:
The issue is fixed by upgrading the firmware for the IFC board to 8928190-10; See Chromaster Release Note document delivered with Hitachi LC October 2024 Edition for OpenLab EZChrom
Keyword: External Component
One-line Description:
Hitachi LC: Incorrect Display of Parameters in Maintenance Tool with Win 11
Problem:
This only happens with Windows 11 OS: The border of editable fields in the Hitachi Maintenance Tool is not displayed correctly after the screensaver was activated once. However, it is still possible to enter values and save them.
Temporary Solution:
N/A
Fix Information:
The workaround is documented in the Release Notes in revision 2.8 of the driver.
Keyword: External Component
One-line Description:
Hitachi LC: USB Drivers incompatible with OS Memory Integrity Option
Problem:
When a Hitachi LC driver for OpenLab is installed on an Windows OS with activated "Memory integrity" option, the installed Hitachi USB drivers do not become valid. Therefore, it will not be possible to connect with the Hitachi LC instrument. The following error is displayed in the configuration process: Unable to connect to IFC. Alternatively, the Memory Integrity option cannot be enabled with Hitachi USB drivers already installed. "Incompatible Drivers" message is shown in such case.
Temporary Solution:
n/a
Fix Information:
This is fixed in driver version 2.7.
Keyword: External Component
One-line Description:
Hitachi LC: USB Drivers incompatible with OS Memory Integrity Option
Problem:
When a Hitachi LC driver for OpenLab is installed on an Windows OS with activated "Memory integrity" option, the installed Hitachi USB drivers do not become valid. Therefore, it will not be possible to connect with the Hitachi LC instrument. The following error is displayed in the configuration process: Unable to connect to IFC. Alternatively, the Memory Integrity option cannot be enabled with Hitachi USB drivers already installed. "Incompatible Drivers" message is shown in such case.
Temporary Solution:
Uninstall the USB driver for Hitachi from Device Manager/Driver menu. Uninstall the Hitachi driver. Disable the Memory Integrity option. Reinstall the Hitachi driver afresh. The USB drivers should work properly this time.
Fix Information:
Fixed in Hitachi LC October 2024 Edition for OpenLab EZChrom; New compatible USB driver version will be installed with the LC driver.
Keyword: External Component
One-line Description:
Hitachi LC: Incorrect Display of Parameters in Maintenance Tool with Win 11
Problem:
This only happens with Windows 11 OS: The border of editable fields in the Hitachi Maintenance Tool is not displayed correctly after the screensaver was activated once. However, it is still possible to enter values and save them.
Temporary Solution:
N/A
Fix Information:
N/A
Keyword: Help
One-line Description:
Hitachi Chromaster Ultra: Help Clarification for the Purge Options of the Pump
Problem:
The Help incorrectly says that after the purge starts the flow will decrease; it should say the flow will increase. The Help also says the valve of the Auto purge will stay open at the end of the purge. This is not correct: the valve is automatically closed as expected.
Temporary Solution:
n/a
Fix Information:
This is fixed in revision 1.2 of the driver.
Keyword: Help
One-line Description:
Hitachi Chromaster Ultra: Help Clarification to Rack Settings of Autosampler
Problem:
The Help says " If the chosen rack parameters are different than those stored in the Autosampler configuration, an error will occur during the run." This happens only in case the Rack configuration of the OpenLab instrument is very different as the real used Rack.
Temporary Solution:
n/a
Fix Information:
Fixed in the driver version 1.2
Keyword: Help
One-line Description:
Hitachi LaChrom Elite: Incorrect Error Message for Analog Input Device Issue
Problem:
An incorrect error message for Analog Input Device can be seen upon encountering an issue: _A was received from the is in ADC calibrating_
Temporary Solution:
n/a
Fix Information:
Fixed in driver version 2.7.
Keyword: Installation
One-line Description:
Uninstalling Hitachi Drivers may lead to failing SVT Report
Problem:
Uninstalling Hitachi drivers may lead to failing SVT report due to hwinhelp.dll file not being properly handled. The uninstallation may cause this file not to be registered properly.
Temporary Solution:
n/a
Fix Information:
Fixed in Hitachi Chromaster 7.25 driver delivered with Hitachi LC October 2024 Edition for OpenLab EZChrom
Keyword: Instrument Control
One-line Description:
"Switch Valve to Load" not available after "Switch Valve to Inject"; "Home arm" action needed first
Problem:
After a "Switch Valve to Inject" action, the "Switch Valve to Load" function is displayed in Direct control menu but is not active. The "Home arm" function has to be executed first, then "Switch Valve to Load" is activated and can be called.
Temporary Solution:
n/a
Fix Information:
Fixed in revision 1.1 of the driver.
Keyword: Instrument Control
One-line Description:
Hitachi LC: Long Launching Time if Instrument Setup Window Left Open Prior To Exiting Instrument Window
Problem:
If the Instrument Setup window left open upon exiting the Instrument window, launching the same instrument might take up to 3 min next time.
Temporary Solution:
As a workaround, close the Instrument Setup window prior to exiting the Instrument window.
Fix Information:
N/A
Keyword: Instrument Control
One-line Description:
Hitachi LaChrom Elite: Intermittent Waiting For Trigger Does Not Start the Run
Problem:
Waiting for Trigger happens repeatedly upon performing an injection on LCE.
Temporary Solution:
N/A
Fix Information:
This is fixed in driver revision 2.7.
Keyword: Instrument Control
One-line Description:
Hitachi Chromaster: Script Error Appears Under Certain Conditions
Problem:
When working with a Chromaster instrument, a script error can show up in the Status UI under certain conditions.
This happens in a client/server environment if comma (",") is used as decimal separator. After the script error, a user without privilege can access the "Auto Purge Wash" & "Plunger Wash" functions in the Pump Status UI.
Temporary Solution:
There's no impact on data acquisition. Confirm the script error by clicking the Ok button. Use the "Auto Purge Wash" & "Plunger Wash" only if your role allows it. Do NOT use the "Auto Purge Wash" & "Plunger Wash" when the state of the instrument is set to “Release” for “Take/Release” function.
Fix Information:
This is fixed with revision 2.8 of Chromaster driver.
Keyword: Localization
One-line Description:
Hitachi Chromaster: Acquisition Window Crashes if Non-English OS Used for AIC/Client
Problem:
Once a Chromaster instrument is launched, after 5 ~ 10 minutes OpenLab CDS is generating error messages in Activity log: "Le format de la chaîne d'entrée est incorrect" (Input String is not in correct format) until the system crashes.
Temporary Solution:
Change the format of system account to English (United States). In Region settings select [Administrative][Copy current settings] to [Welcome screen and system accounts], and re-boot the AIC/Client.
Fix Information:
Fixed in revision 2.8 of the driver.
Keyword: Method
One-line Description:
Hitachi Chromaster Ultra LC: Direct Control "Turn on at" Functions Properly Only at UTC Time Zone Entry
Problem:
The issue happens with Hitachi CMU driver v1.1. The "Turn on at" function does not work with the current time zone the AIC is set up to but instead uses the UTC time. This affects the "Turn on at" function of the instrument's pump, oven and detectors.
Temporary Solution:
n/a
Fix Information:
Fixed in version 1.2 of the driver.
Keyword: Method
One-line Description:
Hitachi CMU: Method Resolution Error When Upgrading from CMU1.0 to CMU1.1
Problem:
When a method that is saved using the CMU 1.0 driver version is opened with a system with the CMU 1.1 driver version, some parameters are not resolvable and set to default. A message is given by the system to notify the users about it.
Temporary Solution:
n/a
Fix Information:
This is addressed in the Hitachi Chromaster Ultra Rs driver 1.2.
Keyword: Method
One-line Description:
Hitachi Chromaster LC: Actual Run Time Can Be 10x Longer Than Set in the Method Due to Certain System Settings
Problem:
The actual method run time may occasionally (depending on the Region system settings) be ten times longer than specified in the instrument method. For example, specifying a time of 0.5 minutes may result in an actual run time of 5 minutes. Instrument / Configuration / Setup OpenLab CDS 2.7 Hitachi Chromaster driver 2.6.02
Temporary Solution:
If the issue occurs, please reach out to the Third Party Driver support for a workaround that involves changing certain system settings.
Fix Information:
Fixed in revision 2.8 of the driver.
Keyword: Method Usage
One-line Description:
Incorrect Absorption Value in the Online signal and Status of 6420 UV-VIS Detector When Non-Default Range Value Used.
Problem:
When the Absorbance range of the 6420 UV-VIS detector is not 1.0 (default value), the displayed value of the Online signal and the Status is not correct and have to be ignored. However, the acquired and stored data is not affected. The data corresponds to the selected Range value and will be displayed correctly in Data Analysis.
Temporary Solution:
n/a
Fix Information:
The failure is fixed in revision 1.1 of the driver (the fix is in the firmware).
Keyword: Method Usage
One-line Description:
Hitachi ChromasterUltra Rs: DAD Online Signal Wavelength Needs to be Selected in DAD Direct Control Option
Problem:
The DAD signal in the online plot does not reflect the signals selected in the method currently downloaded to the instrument. Instead, the option "Change Monitoring Wavelength" in DAD direct control context menu (in the Instrument Status window) controls what wavelength to display for the online plot. This behavior applies to both OpenLab CDS and OpenLab CDS EZChrom Edition for Hitachi ChromasterUltra Rs driver.
Temporary Solution:
None.
Fix Information:
Fixed in revision 1.1 of the driver.
Keyword: Performance
One-line Description:
Refresh period of the Online Plot Might Increase for Long Sequences
Problem:
The period to refresh the Online Plot might expand by a fraction of a minute after running a few hundreds of injections in a single sequence. The quality of the recorded data is not impacted by the issue.
Temporary Solution:
n/a
Fix Information:
This is fixed in Hitachi LaChrom Driver v2.6
Keyword: Performance
One-line Description:
Hitachi Chromaster LC: Long Acquisition Launch Time
Problem:
The Launch time of the Acquisition software can take more than 30 seconds when working with a Chromaster instrument. The issue appears under the following conditions: _ Previous Acquisition session has been closed while the Method Setup window was displayed _ The instrument connection has been closed after the previous Acquisition session.
Temporary Solution:
Close the Method setup window prior to closing the instrument window. The delayed launch will be avoided this way upon the subsequent launch of the Hitachi instrument.
Fix Information:
This is fixed in OpenLab CDS 2.8.
Keyword: Report Templates
One-line Description:
No Module Information is Reported in the Sequence Summary Report
Problem:
Sequence Summary Report does not contain the Instrument Module Information.
Temporary Solution:
The information about the Instrument modules can be seen in the Standard Method report and can be printed from: File -> Print -> Configuration.
Fix Information:
This will not be fixed as a work around is available.
Keyword: User Interface
One-line Description:
Column Oven Pictogram Displays the Oven is ON in case of a door opened condition even if it is OFF
Problem:
The actual condition of column oven is OFF when the oven compartment door is opened, but, the pictogram of the oven shows it as ON.
Temporary Solution:
n/a
Fix Information:
This is fixed in revision 1.1 of the driver.
Keyword: User Interface
One-line Description:
Hitachi ChromasterUltra RS: Selection in Tray A Display Also Reflects in Tray B of Microplates
Problem:
For Hitachi CMU RS instrument, when multiple racks are set, the vial selection made in the Tray display for the first rack is also reflected in the second rack. This does not affect data acquisition.
Temporary Solution:
N/A
Fix Information:
To be addressed in OpenLab CDS.
Keyword: User Interface
One-line Description:
Hitachi ChromasterUltra RS: Selection Column by Column on Tray Display Not Supported
Problem:
For Hitachi CMU RS instrument, only row by row vial selection can be done in the Tray Display. Column by column vial selection pattern is not available.
Temporary Solution:
n/a
Fix Information:
To be addressed in the newer revision of the driver.
Keyword: User Interface
One-line Description:
Hitachi Chromaster: Name of 2nd Channel Changes to None Outside Run in Online Signal View
Problem:
For UV 5410 in Dual Wavelength mode, the online signal window of OpenLab CDS displays the 2nd wavelength of Hitachi Chromaster 5410 UVetector name as "None" outside the run. The view of the signals is correct in the DA window. (The online signals do not reflect the true signal when not in a run--this is a known limitation)
Temporary Solution:
None.
Fix Information:
This limitation is documented in the Operation Manual in revision 2.8 of the driver.