Software Status Bulletin
Headspace Driver Software ICF (Headspace)
Known Problem Report as of Dec 8 2025 4:00AM
Preface
-------------------------
This Software Status Bulletin (SSB) documents all known problems in the software
product designated above. The SSB is derived from Known Problem Reports (KPR)
which result from user problems that have been classified as documentation
problems or software defects. When a KPR is written, an identifying number is
assigned to it, and the KPR is added to the next edition of the SSB.
User inputs that have been classified as Enhancement Requests are not documented
in the SSB. User problems that have been submitted, but that have not been
classified by the time the SSB is generated are not included in the SSB.
How to use the SSB
-------------------------
When you experience a problem with a product, first check this SSB to see if the
problem has been reported already, and if there is a temporary workaround
available for the problem, or if the problem has already been fixed by a new
revision. If the problem is not listed in this SSB then you may wish to report
it to the Response Center or to your field support representative.
To determine if your problem is documented in this SSB, first look in the
Keyword Index section of the SSB. Under each keyword is a listing of one-line
descriptions of related KPRs. If any of these sound like yours, locate the KPR
# in the Known Problem Reports section of the SSB, and read the full KPR. The
KPRs in the Known Problem Reports section are sorted by KPR #.
There are two sections in the SSB:
Keyword Index: This index is categorized by keyword. For each KPR there is a
brief description and a KPR #. A KPR may be associated with more than one
keyword.
Known Problem Reports: This section contains KPRs, with all the available
information relevant to the problem. KPRs in this section are sorted by KPR #.
Keyword Glossary
Keyword Index
KPR#:176
ICF 2.03 and 2.04 with GC/HS driver B.01.04.044 throws Application Error on configuring G1888: FileNotFoundException
KPR#:177
Upgrade ICF GC/HS driver to B.01.01 from ICF 2.03 to ICF A.02.04 throws exception on selecting Configuration
KPR#:187
6890 with G1888 missing vial proceeds, but sequence gets out of sync
KPR#:193
Chromeleon: HSS Driver will never "release" monitor mode
KPR#:202
Indicated limits for Transfer Line Diameter in Chromeleon Instrument Configuration different to instrument
KPR#:210
ICF: Pressure Unit changes on Instrument Configuration is not saved for ICF/Chromeleon
KPR#:212
Chromeleon: HSS method migration: Changing Carrier parameter in HSS method UI does neither trigger dirty UI nor script refresh
KPR#:214
Empower 7697 Soft Configuration tab has no Follow GC option for communications capable GC's
KPR#:219
ICF Headspace GC Cycle Time advisory always advises incorrectly: Oven program reported by the GC is 0 minutes
KPR#:227
Headspace driver should remove context menu for upload/download method for Empower
KPR#:228
ICF Headspace Vial Status shows different/wrong Status from Activity Log
KPR#:231
ICF: Follow 7890 enable/disable not recorded to audit trail
KPR#:232
Configuration Pressure Unit change does not appear in Audit Trail
KPR#:235
Chromeleon: 8697 - Method Development Saving Issue
KPR#:237
ICF build: Cannot configure 7697, 'Unexpected headspace type; Exception type: NotImplementedException...'
KPR#:243
Chromeleon: 8697 - The default of the Bryce Method Development should be "off"
KPR#:244
Chromeleon: 8697 - Vial status, lock appears too late
KPR#:245
Chromeleon: 8697 - Audit trail shows incorrect incrementing entries
KPR#:246
Chromeleon: 8697: "Sequence ends in" Clock runs down to 0 before 1st injection while cooling oven
KPR#:248
ICF: Configuration\7697 Audit Trail sees no changes made to Preferences or Schedule; Config Report also shows no notation of these items
KPR#:251
Chromemeon: 8697 - Typo on the HSS 8697 Tile "Off Line"
KPR#:265
Chromeleon: HSS Status Dashboard, Statement "Online" is missing
KPR#:268
Chromeleon: HSS Status shows online after Terminate button is clicked
KPR#:276
Chromeleon: Validation of a full sequence for a GC/HSS System does not complain about the fact that HSS does not support blank runs
KPR#:277
Chromeleon: GC/HSS: Error when starting a sequence without using HSS
KPR#:278
7697A HCV: ICF logs additional 'Error: leak test failure -Sample Skipped' after 'Warning: Leak test failed - Sample Continuing'; MHE sample to same vial abort incorrectly
KPR#:291
Chromeleon: Headspace status dashboard display scrolled automatically down
KPR#:295
Chromeleon: Sequence Validation error message for blank run is not correct
KPR#:296
Barcode Mismatch Actions Are Sequence Scoped, not sample Scoped
KPR#:297
ICF: Wrong IP address for HS instrument results in Application Error
KPR#:299
ICF: trying to configure a 8697 HS when the instrument is in use causes exception
KPR#:301
Chromeleon: OutOfmemory Chromeleon.exe while processing highest data rate on GC
Keyword:
One-line Description:
ICF: Configuration Preferences do not take in a session
Problem:
Open a HS instrument. Go to Configuration->Preferences and disable the timeline in the method editor. The timeline is still shown in the method editor.
Temporary Solution:
n/a
Fix Information:
The problem is fixed in B.01.06.
Keyword:
One-line Description:
Chromeleon: In Status Interface, Vial bar in maximized "HSS 8697" Tile expands over scroll and progress bar.
Problem:
This issue is found in Chromeleon: When the "HSS 8697 " tile of the status UI is maximized, while the Chromeleon application is not in Full screen, the Vial display expands over scroll and progress bar\.
Temporary Solution:
N/A
Fix Information:
Fixed in HS driver 3.2.
Keyword:
One-line Description:
Chromeleon: 8697 - Remove "Keyboard Lock" from the configuration report for 8697
Problem:
The HSS 8697 does not have a Keypad/Keyboard. It follows GC. Please remove the line "Keyboard Lock: OFF" from Configuration report.
Temporary Solution:
N/A
Fix Information:
Fixed in HS driver 3.2.
Keyword:
One-line Description:
Chromeleon: 8697: Remove Method Editor Timeline section from online help
Problem:
The Graphical Method Timeline is not available for Headspace in HSS 3.1. But the online help still states that Timeline is available.
Temporary Solution:
N/A
Fix Information:
Fixed in HS driver 3.2.
Keyword: Headspace
One-line Description:
ICF 2.03 and 2.04 with GC/HS driver B.01.04.044 throws Application Error on configuring G1888: FileNotFoundException
Problem:
With ICF 2.03 or 2.04 with GC/HS driver B.01.04.044 installed, ICF 2.03 and 2.04 with GC/HS driver B.01.04.044
Temporary Solution:
N/A
Fix Information:
Fixed in Headspace driver B.01.07.03.
Keyword: Headspace
One-line Description:
Upgrade ICF GC/HS driver to B.01.01 from ICF 2.03 to ICF A.02.04 throws exception on selecting Configuration
Problem:
Upgrade ICF GC/HS driver to B.01.01 from ICF 2.03 to ICF A.02.04 throws exception on selecting ConfigurationReconfigure G1888A, Select Configuration and get a large Application Error, ThreadException.
Temporary Solution:
N/A
Fix Information:
Fixed in Headspace driver B.01.07.03.
Keyword: Headspace
One-line Description:
6890 with G1888 missing vial proceeds, but sequence gets out of sync
Problem:
Start a 4 line sequence to G1888 with vial 2 missing. The hardware skips to vial 3, etc, correctly. The Instrument Log correctly shows that Vial 2 was missing, skipping to vial 3. However, the sequence does not abort vial 2, instead putting vial 3's data into vial 2's line in Run Automation. At completion of the sequence which seemingly runs for the time of the four runs, vial 2 shows Error: unexpected exception, Vial 3 is noted as 'State: Executing' and Vial 4 is 'Scheduled.'
Temporary Solution:
n/a
Fix Information:
fixed in B.01.08
Keyword: Headspace
One-line Description:
Chromeleon: HSS Driver will never "release" monitor mode
Problem:
We have had various defects describing a problem with HSS Driver being in monitor mode and never come out from this
Temporary Solution:
n/a
Fix Information:
Fixed in revision B.01.10.
Keyword: Headspace
One-line Description:
Indicated limits for Transfer Line Diameter in Chromeleon Instrument Configuration different to instrument
Problem:
The indicated allowed range for Transfer Line Diameter in ICF’s Configuration is 0.25 to 0.53 mm (250 to 530 microns). But on the instrument front panel the range can be set from 200 to 600 microns. When a value outside the 0.25 to 0.53 range is set in the instrument configuration, the enter box background becomes yellow.
Temporary Solution:
n/a
Fix Information:
Fixed with HS driver 3.1
Keyword: Headspace
One-line Description:
ICF: Pressure Unit changes on Instrument Configuration is not saved for ICF/Chromeleon
Problem:
Not possible to change HSS7697A pressure unit in Instrument Configuration Manager even if a drop down menu is provided.
Temporary Solution:
n/a
Fix Information:
Fixed with HS driver 3.1.
Keyword: Headspace
One-line Description:
Chromeleon: HSS method migration: Changing Carrier parameter in HSS method UI does neither trigger dirty UI nor script refresh
Problem:
After migrating a method (GC7890 controlled by native Thermo driver and HSS7697A controlled by Thermo ICF) to Agilent drivers for Thermo Chromeleon a change in the HSS7697A method UI carrier settings does not trigger a dirty UI. Possibly therefore, no HSS inject command is added to the script.
Temporary Solution:
n/a
Fix Information:
Fixed with HS driver 3.1.
Keyword: Headspace
One-line Description:
Empower 7697 Soft Configuration tab has no Follow GC option for communications capable GC's
Problem:
In Empower, select Module Options\Module Options\Agilent 7697A\Configuration tab. There is no 'Follow GC' text or button, though there should be for a 7890B.
Temporary Solution:
n/a
Fix Information:
Fixed with HS driver 3.1.
Keyword: Headspace
One-line Description:
ICF Headspace GC Cycle Time advisory always advises incorrectly: Oven program reported by the GC is 0 minutes
Problem:
ICF 2.6.30 Update 3 with GC/HS driver 3.1.205 repackaged connected to 8860 with 7697 12 vial sampler. Select HS method Times view. With a GC oven time set, the text next to the GC Cycle time always reads incorrectly: 'Oven program reported by the GC is 0 minutes long.'
Temporary Solution:
n/a
Fix Information:
Fixed with HS driver 3.1.
Keyword: Headspace
One-line Description:
Headspace driver should remove context menu for upload/download method for Empower
Problem:
The headspace driver should mirror the GC driver in that it should remove the Upload\download method context menu from the method edit for headspace in Empower. B.01.10.007
Temporary Solution:
n/a
Fix Information:
Fixed with HS driver 3.1.
Keyword: Headspace
One-line Description:
ICF Headspace Vial Status shows different/wrong Status from Activity Log
Problem:
Set all Sequence actions to Skip. Vial 1 is wrong size, vial 2 is missing, during vial 3's equilibration, open the GC oven door, vial 4 has no septum. The activity log shows correctly 'Incorrect vial size found on vial 1', but HS Vial Status shows 'Timed out waiting for datasystem ready on vial 1'. Activity Log shows correctly ;Timed out waiting for datasystem ready on vial 3', but the Vial Status shows 'Run Aborted'.
Temporary Solution:
n/a
Fix Information:
Fixed with HS driver 3.1.
Keyword: Headspace
One-line Description:
ICF: Follow 7890 enable/disable not recorded to audit trail
Problem:
ICF: Follow 7890 enable/disable not recorded to audit trail
Temporary Solution:
n/a
Fix Information:
Fixed in HS driver 3.1 SR1
Keyword: Headspace
One-line Description:
Configuration Pressure Unit change does not appear in Audit Trail
Problem:
When you change the pressure unit in the configuration dialog of 7697 and then call the audit trail, no changes are reported.
Temporary Solution:
n/a
Fix Information:
Fixed with HS driver 3.1 SR1.
Keyword: Headspace
One-line Description:
Chromeleon: 8697 - Method Development Saving Issue
Problem:
Open a method that has method development enabled, change the parameters of the method development like increment and max temperature, save the method, close it and open it again, the old, unchanged development parameters are in the method.
Temporary Solution:
n/a
Fix Information:
Fixed with HS driver 3.1 SR1.
Keyword: Headspace
One-line Description:
ICF build: Cannot configure 7697, 'Unexpected headspace type; Exception type: NotImplementedException...'
Problem:
ICF 3.0.37 with GC/HS driver 3.6.73. Configure a 7697 and get: Unexpected headspace type; Exception type: NotImplementedException... Cannot proceed.
Temporary Solution:
n/a
Fix Information:
Fixed with HS driver 3.1 SR1.
Keyword: Headspace
One-line Description:
Chromeleon: 8697 - The default of the Bryce Method Development should be "off"
Problem:
It was reported that after one user created a method using temperature increment in Method development UI without downloading to the instrument and disconnected from the instrument, the other user then connected to the same instrument and saw that method (with temperature increment) as default method.
Temporary Solution:
n/a
Fix Information:
Fixed with HS driver 3.1 SR1.
Keyword: Headspace
One-line Description:
Chromeleon: 8697 - Vial status, lock appears too late
Problem:
After having started the sequence, the "Vial Status" having a "Lock"-Symbol is not in sync with the messages we get from ICF regarding locked lines. So the Chromeleon sequence shows 3 lines locked (marked yellow), the Vial Status Window show only two locks for vial 1 and 2.
Temporary Solution:
n/a
Fix Information:
Fixed with HS driver 3.1 SR1.
Keyword: Headspace
One-line Description:
Chromeleon: 8697 - Audit trail shows incorrect incrementing entries
Problem:
Incorrect Incrementing info in audit TrailSummary: While running a full sequence with method development on HSS 8697 "Bryce", entries for incrementing the loop and transfer line temperatures are found, while both are switched off in the method.
Temporary Solution:
n/a
Fix Information:
Fixed with HS driver 3.1 SR1.
Keyword: Headspace
One-line Description:
Chromeleon: 8697: "Sequence ends in" Clock runs down to 0 before 1st injection while cooling oven
Problem:
The "Sequence ends in" Clock on the Status UI tile of HSS 8697 runs, while HSS is not ready (cooling oven)
Temporary Solution:
n/a
Fix Information:
Fixed with HS driver 3.1 SR1.
Keyword: Headspace
One-line Description:
ICF: Configuration\7697 Audit Trail sees no changes made to Preferences or Schedule; Config Report also shows no notation of these items
Problem:
ICF A.02.04.115 with GC/HS drivers A.03.02.098 on Win10 connected to 7890A.01.16 and 7697 12 vial or 7890B/7697 111 vial system. Select Configuration\7697. Change anything on the Preferences tab or the Schedule. Select Audit Trail and none of this is noted as changed. Select Report and none of the Schedule and Preferences are listed, either.
Temporary Solution:
N/A
Fix Information:
Fixed with HS driver 3.1 SR1.
Keyword: Headspace
One-line Description:
Chromemeon: 8697 - Typo on the HSS 8697 Tile "Off Line"
Problem:
there is a typo on the HSS 8697 tile in the status UI. "Off Line" should be "offline"
Temporary Solution:
N/A
Fix Information:
Fixed with HS driver 3.1 SR1.
Keyword: Headspace
One-line Description:
Chromeleon: HSS Status Dashboard, Statement "Online" is missing
Problem:
The word "Online" is missing below the status bar when the status change from Offline to Online.
Temporary Solution:
N/A
Fix Information:
Fixed in HS driver 3.2.
Keyword: Headspace
One-line Description:
Chromeleon: HSS Status shows online after Terminate button is clicked
Problem:
After Terminate button is clicked, the HSS still shows as being online (in the dashboard) while the GC does go offline.
Temporary Solution:
N/A
Fix Information:
Fixed in HS driver 3.2.
Keyword: Headspace
One-line Description:
Chromeleon: Validation of a full sequence for a GC/HSS System does not complain about the fact that HSS does not support blank runs
Problem:
When Validating Sequence, the driver needs to return "Error" with a reasonable error description ("Blank run is not supported ...").
Temporary Solution:
N/A
Fix Information:
Fixed in HS driver version 3.3.
Keyword: Headspace
One-line Description:
Chromeleon: GC/HSS: Error when starting a sequence without using HSS
Problem:
An NullReference exception is thrown on call to StartFullSequenceMode when the HS7697 is offline.
Temporary Solution:
N/A
Fix Information:
Fixed in HS driver version 3.3.
Keyword: Headspace
One-line Description:
7697A HCV: ICF logs additional 'Error: leak test failure -Sample Skipped' after 'Warning: Leak test failed - Sample Continuing'; MHE sample to same vial abort incorrectly
Problem:
Start sequence using MHE method to the same vial. Sequence action is to continue on leak. Detailed Status shows both injections fail for leak test failure, Sample continuing. However, there is an additional 'Error: leak test failed - Sample Skipped' for both. Then the second injection aborts both the run (both injections) and the sequence. Run Automation and Sample ID shows 'unexpected exception'.
Temporary Solution:
N/A
Fix Information:
Fixed in HS driver version 3.3.
Keyword: Headspace
One-line Description:
Chromeleon: Headspace status dashboard display scrolled automatically down
Problem:
Status UI of Headspace Tab scrolled every time automatically down.
Temporary Solution:
N/A
Fix Information:
Fixed in Version 3.4.
Keyword: Headspace
One-line Description:
Chromeleon: Sequence Validation error message for blank run is not correct
Problem:
The Sequence Validation for Blank runs has been implemented on 3.3 (TT 284720). The Chromeleon team verified that the Blank Run issue is detected through sequence Validation. But they found the error message is presented as a formatting "template": "Blank Run are not supported on {0} Headspace Sampler".
Temporary Solution:
N/A
Fix Information:
Fixed in HS driver version 3.4.
Keyword: Headspace
One-line Description:
Barcode Mismatch Actions Are Sequence Scoped, not sample Scoped
Problem:
ICF supports changing the barcode mismatch option (inject, don't inject, etc) on a sample basis. The HS driver only supports this on a sequence basis as it constructs methods to send to the firmware and doesn't check to see if the selected action has changed between samples (which would require a new firmware method).
Temporary Solution:
N/A
Fix Information:
Fixed in HS driver version 3.4.
Keyword: Headspace
One-line Description:
ICF: Wrong IP address for HS instrument results in Application Error
Problem:
If wrong IP address is entered, an Application Error, ThreadExeption occured during uploading process. ICF has to be restarted.
Temporary Solution:
N/A
Fix Information:
Fixed in HS driver 3.4.
Keyword: Headspace
One-line Description:
ICF: trying to configure a 8697 HS when the instrument is in use causes exception
Problem:
when a 8697 HS is configured while the instrument is in use by empower an exception appears.
Temporary Solution:
N/A
Fix Information:
Fixed in HS driver version 3.4.
Keyword: Headspace
One-line Description:
Chromeleon: OutOfmemory Chromeleon.exe while processing highest data rate on GC
Problem:
The status XML is growing over time and break the longer sequences for Chromeleon
Temporary Solution:
N/A
Fix Information:
Fixed in version 3.5