Specification Compliance Decisions with respect
to "ISO17025 Calibration"
An
uncertainty is a figure of merit associated with the actual measured
value; the boundary limits within which the ‘true’ value lies. Contributors
to this “potential for inaccuracy” include the performance of the
equipment used to make the measurement, the test process or technique
itself and environmental effects.
Additional imprecision may result from behaviour of the phenomenon
or item being measured. A skilled metrologist will assess and combine
these various components in an uncertainty budget. To prove that
a product complies with specification (or doesn’t), the uncertainty
must be less than the unknown’s specification.
In the following example, a measurement is performed on five different
items: A, B, C, D and E. The results of these measurements are shown
in the illustration below.

Without taking measurement uncertainty into account we could state
that:
- Devices A and B are within specification.
- Device C is borderline.
- Devices D and E are exceed specification.
However, when taking uncertainty into account:
- We can say with high-confidence that device A is within specification
(that is, even if the uncertainty is added to the measured value,
the tolerance is not exceeded).
- Devices B, C and D could be either in or out-of-specification
(that is, the status of the product's compliance is "indeterminate").
- Device E is out-of-specification.
Agilent’s Position on ‘Indeterminate’ Points
In the real world, the ‘indeterminate’ situation is largely impractical.
Agilent has considered this situation at great length, in conjunction
with input from its customers. The result is an approach that strikes
a good balance between what is metrologically precise and what customers
really want and so creating a realistic, commercially acceptable
service.
Consequently, for Agilent's "ISO17025 Calibration" service,
the assessment of A, B, C, D and E will be:
A = PASS
Performance is better than or equal to product specification. It
is labelled PASS and a statement of conformity is issued.
B, C = PASS #
Performance falls within the measurement uncertainty guard band
on the ‘in-specification’ side. The customer is notified on the
measurement report that:
”All measured values indicated by this data report lie within
specification. Due to measurement uncertainty, the true values
of one or more of these points may lie outside of the specification,
but there is a high probability that these values lie within specification.”
A certificate of calibration stating that the instrument meets
specifications is issued. B and C are not adjusted to be closer
to center.
D, E = FAIL
Performance falls within measurement uncertainty guard band on the
‘out-of-specification’ side and is considered a FAIL. A statement
of conformity is issued only if D and E are adjusted to be in specification.
|