Uncertainty Made Easy
About this Article
This paper by Ian Instone was first presented at the Institution
of Electrical Engineers, London in October 1996 at their Colloquium
entitled "Uncertainties made easy".
Note: Details of the current versions
of recommended uncertainty guidance publications can be found on
the Uncertainty Resources page in this section.
Simplified Method for Assessing Uncertainties
in a Commercial, Production Environment
Introduction
With the introduction of Edition 8 of NAMAS
document NIS3003 (1) and the
inclusion of the principles outlined in the ISO Guide to the Expression
of Uncertainty in Measurement (2)
the assessment of uncertainties of measurement has become a task
more suited to a mathematician rather than the average calibration
engineer. In some companies with small calibration departments it
might be possible for all of the engineers to be re-educated in
assessment of uncertainties, however, in larger laboratories it
is more usual for various engineers to become specialist in certain
aspects of the calibration process. This paper aims to demonstrate
a simplified approach to uncertainty assessment which falls broadly
within the guidelines set out in both NIS3003 and the ISO Guide.
One of the first stumbling blocks in NIS3003
is the necessity to derive a measurement equation. Whilst it is
agreed that this is a useful skill which might demonstrate a more
thorough understanding of the measurement principles, it seems only
to serve as an additional step in the uncertainty assessment process,
steps which were not thought necessary in previous 7 editions of
NIS3003. The next step, deriving sensitivity coefficients by the
use of partial differentiation will cause most calibration engineers
to reach for the mathematics text book. Fortunately, in many cases
these two steps can be replaced using a more practical approach.
A list of contributions to the uncertainty budget can be used in
place of the measurement equation and each term may be partially
differentiated by varying the quantity over its range and measuring
its influence on the measurand. For instance, it may have been determined
that temperature variations have an influence upon the quantity
being measured then, rather than produce a measurement equation
which includes temperature and partially differentiate it, one can
simply perform the measurement, change the temperature by the specified
amount and re-measure. The resultant change in the measurand becomes
a contribution to the uncertainty budget. There are also cases where
the same approach may be used but where there may be no necessity
to perform the measurements to obtain the data. For instance, many
resistors have a temperature coefficient specification, in the form
of ±N parts per million per degree Celsius. Assuming the temperature
is controlled to within ±2°C the change in the value of the resistor
due to temperature fluctuations will be given by:-
N parts per million
x 2°C
Most contributions to an uncertainty budget
can be assessed using either method. The practical method described
will often yield smaller values because they are based on measurements
performed on only a small quantity of items, where as the latter
method is based upon the equipment specification which should cover
the entire population of that instrument and so will normally produce
larger contributions to the uncertainty budget.
Type-A Uncertainties
In a commercial calibration laboratory often
it is not economical to perform several sets of measurements on
a particular instrument solely to produce a value for the random
(Type-A) uncertainty contribution. The alternative method shown
in NIS3003 is preferred and usually employed where possible. In
cases where multiple measurements are performed it is usual practice
to calculate the standard deviation of the population. The estimated
standard deviation of the uncorrected mean of the measurand is then
calculated using:
Esd
= Psd / sqrt(N)
Where:
Esd is the estimated
standard deviation of the uncorrected mean of the measurand
Psd is the standard deviation
of the population of values
N is the quantity of repeated measurements
When the quantity of measurements to be
performed on the equipment being calibrated is limited to one set
of measurements then N in the equation
above will be 1. The standard deviation of the population Psd
will previously have been determined from an earlier Type-A evaluation
based upon a large number of repeated measurements. In an ideal
world the measurements would be repeated on several instruments
within the family and the worst case standard deviation used in
the Type-A assessment. In practice however, providing the assessment
techniques outlined in this paper are employed, the Type-A contribution
to the uncertainty budget can often be shown to be negligible so
the need to make a very large number of repeated measurements is
reduced.
In the ideal world where customers are willing
to pay unlimited amounts of money for their calibrations, or where
we have very large quantities of similar instruments to calibrate
it is a fairly simple matter to measure several instruments many
times and obtain a good reliable estimate for the standard deviation.
In reality, customers have limited budgets and calibration laboratories
rarely have even small quantities of particular instruments which
can be used for extensive testing to provide a good and reliable
estimate of the standard deviation. Another simpler, and more cost
effective method is required.
Before embarking upon the assessment of
uncertainties we need to understand exactly what our customer is
expecting of their calibration report and what use they will make
of it. For the majority of simple reference standards such as resistors,
standard cells, capacitors etc. it is likely that the measured values
will be used by the customer so an uncertainty assessment as defined
by NIS3003 will be required. For the great majority of instruments
it is often not possible to make any use of the values obtained
during its calibration so it is usually only necessary to provide
a calibration which demonstrates that the instrument is operating
within its specifications. In these cases it is usually not necessary
to provide measurements with the lowest measurement uncertainties,
which allows some compromises to be made.
ISO 10012-1 (3)
suggests that we should aim for an accuracy ratio between uncertainty
and the instrument being calibrated of greater than 3:1. The American
interpretation of ISO Guide 25 (4),
ANSI Z540-1 (5) suggests that
uncertainties become significant when the accuracy ratio is less
than 4:1. If we assume that the instrument specification has the
same coverage factor as the uncertainty the following expression
would describe the resultant combination of the uncertainty and
specification which should be used when the instrument is used to
make measurements:
§
= sqrt [ S 2 + U 2 ]
Where:
§ is the resultant expanded specification
resulting from the calibration
S is the specification of the parameter
being measured
U is the uncertainty of measurement
when performing the calibration
In the cases where S
>= 4U the effect of the uncertainty
upon the specification is shown to be negligible, for instance assume
that S = 8 and U
= 2 then:
§
= sqrt [ 8 2 + 2 2 ]
= sqrt [ 64 + 4 ]
= sqrt [ 68 ]
= 8.25
Therefore, with an accuracy ratio of 4:1
the effective specification expands by 3.1%. As most uncertainties
are only quoted using two figures it is unlikely that this small
increase would have any effect. Repeating the same with an accuracy
ratio of 3:1 produces an increase of only 5.2%.
The same analogy can be used when assessing
the significance of a particular uncertainty contribution. Type-A
uncertainties are those assessed using statistical methods usually
based on many sets of measurements, thereby making them the most
expensive to assess. Using the model above we can show that Type-A
uncertainties are insignificant when they are less than 30% of the
magnitude of the Type-B uncertainties:
Total Uncertainty = Type-B
uncertainties where Type-A < 0.3 Type-B
and:
Effective Specification =
Specification where Total Uncert < 0.3 Spec.
From above we can show that Type-A uncertainties
can be regarded as insignificant when they are less than 0.09 of
the specification being tested, or in approximate terms Type-A uncertainties
can be regarded as negligible when they are less than 10% of the
specification.
Verifying that an uncertainty contribution
is less than a given value is usually much easier than assessing
the precise magnitude of it. One method described in an earlier
paper (6) normally requires
only two complete sets of measurements to be made on the same instrument.
One set of measurements are then subtracted, one measurement at
a time from the other set. The largest difference is then assumed
to be a conservative estimate of the Type-A uncertainty contribution.
This technique has been verified many times against uncertainties
assessed in the traditional way and has always produced an acceptable
conservative estimate of the Type-A contribution, providing that
an adequate quantity of measurements are compared across the range.
Assuming that the comparison produces no values that are outside
the limits defined earlier (10% of the DUT specification or 30%
of the Type-B uncertainty estimate) it can be assumed that the Type-A
uncertainties are not significant. To provide good confidence and
consistency in the assessment process the value defined as insignificant
should always be included in the assessment.
It is also possible to use values for the
Type-A assessment gained from other, related instruments providing
some knowledge of the construction of the instrument under test
is available. For instance, it may be that a laboratory has already
assessed a certain 50MHz to 18GHz signal generator and verified
that the Type-A uncertainty contribution meets the criteria outlined
above. A 12GHz signal generator from the same family is then submitted
for assessment. In this case, providing the two signal generators
share similar designs, and use similar hardware and layouts, and
the same test methods and equipment are used it would be reasonable
to employ the 18GHz Type-A assessment on both generators. In other
cases it might be possible to refer to published data for certain
Type-A contributions.
In cases where these techniques reveal that
the Type-A contributions are significant (as defined above) the
uncertainty assessment should be performed in the usual way using
many repeated measurements.
Sensitivity Coefficient
In most cases sensitivity coefficients can
be assumed to be 1. However there are some notable exceptions where
other values will be used. One of these relates to the measurement
of resolution bandwidth on a spectrum analyzer. In this case we
have measurement uncertainties expressed in two different units;
measurements of amplitude are expressed as an amplitude ratio (usually
in dB units) and measurements of frequency (in Hz.). The bandwidth
measurement is often performed by applying a "pure" signal
to the analyzer's input and setting the controls so that the signal
shown below is visible. The envelope describes the shape of the
filter and normally we would measure the 3dB or 30% (below the reference)
point of it (shown on the left of the figure below). To assess the
sensitivity coefficient we need to determine the gradient of the
graph at the measurement point. Spectrum analyzers often have an
amplitude specification of 0.1dB per 1dB, therefore the amplitude
uncertainty at 3dB will be ±0.3dB or ±7%. We then move ±7% from
the 70% point and read off the resultant change in frequency.
The resultant change in frequency
due to amplitude uncertainty is: ±3.8 frequency units. Since this
value has been found for an amplitude specification of ±0.3dB it
will have a sensitivity coefficient of 1.
On the right of the figure
is a similar construction for assessing the frequency uncertainty
due to the amplitude uncertainty when the 6dB (50%) point is measured.
In this case the amplitude uncertainty increases to ±0.6dB (0.1x6).
As a linear ratio this equates to ±13%.
Reading from the graph this represents a
frequency uncertainty of ±6 frequency units.
Assessing the uncertainty contributions
in this way greatly reduces the possibility of errors as might occur
if following the theory using partial differentiation. In addition
a practical technique such as this is preferred by most calibration
engineers.
Other empirical means of obtaining values
for the uncertainty budget may also be employed. For instance it
might be possible to establish a value for temperature coefficient
by changing the environmental temperature by a few degrees. In this
case we could derive a sensitivity coefficient for the output signal
in terms of temperature change.
Total Uncertainty Budget
One of the principle benefits of the latest
revision of NIS3003 is the strong suggestion that all of the uncertainty
contributions should be listed in a table along with their probability
distribution. Whilst at first sight this seems a tedious task, it
pays dividends in the future because it makes the contributors to
the budget absolutely clear. The table below shows a typical example
of an uncertainty assessment for a microwave power measurement at
18GHz using a thermocouple power sensor. These types of power sensor
measure power levels relative to a known power so a 1mW, 50MHz power
reference is included on the power meter for this purpose. In most
cases it is simpler and more correct to use a measuring instruments
specification rather than try to apply corrections and assess the
resultant uncertainty. For the majority of measurements it is not
possible to make corrections based upon a calibration report as
that report only indicates the instruments calibration status at
the time it was measured and only when operated in that particular
mode described on the certificate. It is not possible to predict
the errors at any other points.
|
Symbol
|
Source of Uncertainty
|
Value
± %
|
Probability Distribution
|
Divisor
|
Ci
|
Ui
± %
|
|
K
|
Calibration factor at 18 GHz
|
2.5
|
normal
|
2
|
1
|
1.25
|
|
D
|
Drift since last calibration
|
0.5
|
rectangular
|
sqrt(3)
|
1
|
0.29
|
|
I
|
Instrumentation Uncertainty
|
0.5
|
normal
|
2
|
1
|
0.25
|
|
R
|
50 MHz Reference spec.
|
1.2
|
rectangular
|
sqrt(3)
|
1
|
0.69
|
|
Mismatch loss uncertainties:
|
|
M1
|
Sensor to 50 MHz Reference
|
0.2
|
U-shaped
|
sqrt(2)
|
1
|
0.14
|
|
M2
|
Sensor to 18 GHz Generator
|
5.9
|
U-shaped
|
sqrt(2)
|
1
|
4.17
|
| |
|
A
|
Type-A Uncertainties
|
2.1
|
normal
|
2
|
1
|
1.05
|
| |
|
|
|
|
|
|
|
UC
|
Combined Standard Uncertainty
|
|
normal
|
|
|
4.55
|
|
U
|
Expanded Uncertainty
|
|
normal (k=2)
|
|
|
9.10
|
Where:
Ci
is the sensitivity coefficient used to multiply the input quantities
to express them in terms of the output quantity.
Ui
is the standard uncertainty resulting from the input quantity.
The standard uncertainties are combined
using the usual root-sum-squares method and then multiplied be the
appropriate coverage factor (in this case k=2). In some cases it
will be appropriate to use a different coverage factor, perhaps
when a 95% confidence level is not adequate, or sometimes when the
input quantities are shown to be "unreliable". The Vi
(degrees of freedom of the standard uncertainty) or Veff
(effective degrees of freedom) column has not been included in the
table above in order to simplify the assessment process.
Degrees of Freedom
Degrees of freedom is a term used to indicate
confidence in the quality of the estimate of a particular input
quantity to the uncertainty budget. For the majority of calibrations
performed under controlled conditions there will be no need to consider
degrees of freedom and a coverage factor of k=2 will be used.
In cases where the Type-A uncertainty has been assessed using very
few measurements a different coverage factor, using the degrees
of freedom, would normally be calculated. However, whilst the assessment
method proposed in this paper is based on only two sets of measurements
being performed experimental data confirms that this treatment (taking
the worst case difference) produces a reliable, conservative estimate
of the Type-A uncertainties. In most cases the degrees of freedom
can be assumed to be infinite and the evaluation of the t
factor using the Welch-Satterwaite equation would not be necessary.
NIS3003 provides some guidance on using these methods but does stress
that normally it is not necessary to employ them.
Conclusion
The uncertainty assessment method described
in this paper have been employed at Hewlett-Packard's UK Service
Center for several years. External, internal and informal measurement
audits have in every case provided confirmation that the uncertainties
are being estimated with the expected level of confidence. This
simplified approach is easier to understand and use which enables
more calibration engineers to contribute fully to the uncertainty
assessments.
- The expression of Uncertainty and
Confidence in Measurement for Calibrations, NIS3003 Edition
8 May 1995.
- Guide to the Expression of Uncertainty
in Measurement. BIPM, IEC, IFCC, ISO, IUPAC, OIML. International
Organization for Standardization. ISBN 92-67-10188-9. BSI Equivalent:
"Vocabulary for Metrology, Part 3. Guide to the Expression
of Uncertainty in Measurement", BSI PD 6461: 1995.
- Quality assurance requirements for
measuring equipment, Part 1. Metrological confirmation system
for measuring equipment, ISO 10012-1:1992.
- Calibration Laboratories and Measuring
and Test Equipment - General Requirements ANSI/NCSL Z540-1-1994.
- General requirements for the competence
of calibration and testing laboratories, International Organization
for Standardization, ISO Guide 25:1990.
- Calculating
the Uncertainty of a Single Measurement from IEE
Colloquium on "Uncertainties in Electrical Measurements",
11 May 1993, Author Ian Instone, Hewlett-Packard.
|