Agilent

Software Status Bulletin


ChemStation     ChemStation (G2070BA)

Known Problem Report as of Mar 14 2017 10:27AM

Preface
-------------------------
This Software Status Bulletin (SSB) documents all known problems in the software
product designated above. The SSB is derived from Known Problem Reports (KPR)
which result from user problems that have been classified as documentation
problems or software defects. When a KPR is written, an identifying number is
assigned to it, and the KPR is added to the next edition of the SSB.

User inputs that have been classified as Enhancement Requests are not documented
in the SSB. User problems that have been submitted, but that have not been
classified by the time the SSB is generated are not included in the SSB.

How to use the SSB
-------------------------
When you experience a problem with a product, first check this SSB to see if the
problem has been reported already, and if there is a temporary workaround
available for the problem, or if the problem has already been fixed by a new
revision. If the problem is not listed in this SSB then you may wish to report
it to the Response Center or to your field support representative.


To determine if your problem is documented in this SSB, first look in the
Keyword Index section of the SSB. Under each keyword is a listing of one-line
descriptions of related KPRs. If any of these sound like yours, locate the KPR
# in the Known Problem Reports section of the SSB, and read the full KPR. The
KPRs in the Known Problem Reports section are sorted by KPR #.


There are two sections in the SSB:
Keyword Index: This index is categorized by keyword. For each KPR there is a
brief description and a KPR #. A KPR may be associated with more than one
keyword.


Known Problem Reports: This section contains KPRs, with all the available
information relevant to the problem. KPRs in this section are sorted by KPR #.


Keyword Glossary



1046 Control
1100 Control
1200 Control
1200 Control , Method Scouting Wizard
35900 Control
5890 Control
6850 Control
6850 Control , 6890 Control
6890 Control
68xx Driver
7890A Control
78xx Driver
ACQ
Audit Trail
Batch
Calibration
CE Calibration
CE Control
ColumnComp Configuration , ColumnComp Driver , ColumnComp Status
ColumnComp Configuration , Pump Configuration
Common Infrastructure
Companion
Config Editor
Custom Fields
Custom Reports
DAD/MWD Driver , DAD/MWD Status
DAD/MWD Driver , FLD Driver
Data Analysis
Data Analysis , Navigation Table
Data Analysis , Programming
Documentation
Easy Sequence
ECM Integration/OpenLab Option
GC ACQ
GC Analysis
GC Autosampler
Help
Help , Logbook
Initialization
Initialization , CE Control
Installation
Instrument Control
Integration
ISOPlot
Localization (Japanese)
Logbook
Method
Method , SEQ
Method Scouting Wizard
MSD Analysis
MSD Analyst
MSD Control
Navigation Pane
Navigation Table
Output
Peak Purity
Programming
Pump Configuration
Pump Driver
Reports
Reports , Custom Fields
Reports , ECM Integration/OpenLab Option
Reports , MSD Analysis
Reports , SEQ
RTL
Run Control
Run Control , ECM Integration/OpenLab Option
Sample Overlap
Sampler Configuration , Sampler Status
Sampler Driver
Sampler Driver , Sampler Method
Sampler Method
SEQ
SEQ , ACQ , Method Scouting Wizard
Spectra Analysis
User Interface
User Interface , Data Analysis
User Interface , SEQ , ACQ , ECM Integration/OpenLab Option
Utilities
Utilities , Spectra Analysis , MSD Analysis

Keyword Index


Keyword: 1046 Control

KPR#:1153 1046 FLD cannot be set up from the GUI

Keyword: 1100 Control

KPR#:38 Signal change in LC Detector during a run
KPR#:43 two MWD in one instrument: problem to control MWD 2 via GUI
KPR#:64 'Fast Reconditioning' enabled in Normal Mode of capillary pump
KPR#:89 with Column flow of 0.00 ul/min no data acquisition
KPR#:135 EMF indicators do not work
KPR#:179 Using flow gradients during Fraction Collection results in Wrong times in Report
KPR#:213 DAD Diagnosis: lamp indicate green, even when lamps off
KPR#:287 Pressure and Leak test in prep. Pump will not finish, unless Press. released
KPR#:290 Flow composition not updated in GUI when loading new method
KPR#:407 G1321A FLD: cuvette is not correctly recognized
KPR#:416 Starlight and FIA: double start pulsing of G1315C
KPR#:451 G1323B colum setting calles negative value in ChemStation column setting
KPR#:454 DAD Calibration history information logged in chemstation.ini
KPR#:457 Using same detector type twice no selection can be set diagnostic
KPR#:471 Sequence table not accepting options of Next Location/Plate
KPR#:472 G1314C VWD Detector diagnostic test results issue
KPR#:474 Fast Comp. Change / Reconditioning disabled after initialization
KPR#:475 "next plate" option using Vial Trays in Fraction Collection mode
KPR#:476 Customized methods for Noise, Wander, Drift issue
KPR#:487 Aborting System Test on LC systems might perform error state
KPR#:561 micro pressure test failed randomly
KPR#:715 injection number in column tag not update for non-switching valve TCC
KPR#:787 ECD detector scaling in online plot 10x larger than in the data file
KPR#:1189 G1362A RID Calibration Value cannot be entered

Keyword: 1200 Control

KPR#:513 Autobalance not accesible after stopping a sequence with AutoBalance off
KPR#:670 Method Overview: wrong value shown for TCC temperature for "as detector cell"
KPR#:775 impossible to enter a maximum pressure of 600 bar into column information
KPR#:781 Fraction Collector settings of the Peak Detectors table are not saved
KPR#:1064 Sample scan not acquired in case of spectral range mismatch for VWD G1314D/E
KPR#:1065 attenuation of 4000 cannot be specified for the Analog Output of VWD G1314D/E
KPR#:1103 G1365C/D Detectors not reporting wavelength scan summary
KPR#:1181 when solvent recycling for the RID is used the bottle filling still changes
KPR#:1213 Column Compartment G1316A/G1316B not available if it is part of cluster
KPR#:1227 CTC: Injection volume not displayed on sampler GUI
KPR#:1228 CTC: UI does not display correct plate
KPR#:1365 Solvents not available for selection in the pump timetable for a Preparative Pump
KPR#:1389 signal description for the RID differs between signals acquired with classic and RC.NET drivers
KPR#:1399 "Waste" or "Coulmn bypass" runs are causing injection counters to be incremented
KPR#:1411 the valve does not switch to by-pass position for the last injection when using overlapped injection
KPR#:1418 run cannot be stopped/aborted when the injector program is longer than the run time
KPR#:1421 LC RC.NET drivers: progress bar not updated during post-conditioning
KPR#:1429 No flow adjustment in case of temperature screening only campaigns
KPR#:1516 Modul of 1260 family in classic driver may be misinterpreted as member of 1100 family
KPR#:1521 incorrect firmware revision displayed for the ELSD
KPR#:1526 column switching valve: positions "Back-flushing on" and "Back-flushing off" are not kept
KPR#:1539 Wrong firmware revision displayed in Instrument Configuration dialog after firmware upgrade
KPR#:1560 TCC cluster cannot be unclustered in RC.Net configuration
KPR#:1596 Conditioning On cannot be switched off with FW B.06.30
KPR#:1719 Lamp required option in FLD method is not properly migrated
KPR#:1720 Cluster information can be overwritten from the offline session

Keyword: 1200 Control , Method Scouting Wizard

KPR#:1720 Cluster information can be overwritten from the offline session

Keyword: 35900 Control

KPR#:1570 All named Defined Events in Configuration Editor for ADC are rejected
KPR#:1623 35900E - Error The Current Event Name is the same for a special event
KPR#:1677 Cannot start two instruments using the two channels of one ADC at the same time
KPR#:1707 35900: Unable to set up Timed Events
KPR#:1721 View ACQ method from Data Analysis does not list acquisition parameters for the 35900 AD converter.
KPR#:1739 ADC can be non-responsive after long reprocessing
KPR#:1804 Queue method generated status “error” when using 35900E

Keyword: 5890 Control

KPR#:1047 5890: Data rate 10Hz used, Real Time Plot display is doubled

Keyword: 6850 Control

KPR#:707 6850 GC feature loss, such as Prep Run, when controlled by Chemstation
KPR#:1511 6850 instrument type does not show up on GLP report and ACAML file is incomplete
KPR#:1472 B.01.02 Lab Advisor is not compatible with ChemStation B.04.02 and higher

Keyword: 6850 Control , 6890 Control

KPR#:1472 B.01.02 Lab Advisor is not compatible with ChemStation B.04.02 and higher

Keyword: 6890 Control

KPR#:1472 B.01.02 Lab Advisor is not compatible with ChemStation B.04.02 and higher
KPR#:415 6890 Workstation Internal Software Error
KPR#:422 GC system crashes when view/instrument diagram active in Offline
KPR#:848 6890 FPD Classic driver is missing the "Air" and "Makeup" Actuals in Edit Parameters
KPR#:1167 IQ1.CFG file of 6890/6850 instrument corrupted after regional setting change
KPR#:1352 6890 RC.NET driver: instrument diagram not updated to show configuration
KPR#:1412 6890 RC.NET driver: instrument diagram not updated to show configuration
KPR#:1461 6890N with no tray or ALS is seeing GC Injector Front/Back/Dual as an injection source
KPR#:1464 Addition of new columns to the column catalog (database) is not possible after upgrade
KPR#:1684 6890 always opens to Configuration tab when opening Edit Agilent 6890 GC Parameters
KPR#:1784 6890 RC.NET: Edit a running Method by making the oven hold time longer and click ok, then the Sequence/Single run stops

Keyword: 68xx Driver

KPR#:16 Vial Missing Error Message
KPR#:18 If One Signal Needs to be Saved, it Must be Assigned as Signal 1
KPR#:20 Instrument Diagram Does Not Match Installed Column in Method
KPR#:22 In 68XX driver revision 5.02 - 6.11, it is not possible to Modify a Method During a Run to Extend the Run Time or Change the Oven Temp
KPR#:23 Info in Edit instrument parameters\Columns is incomplete
KPR#:31 Ignore missing vial option is not present in GC
KPR#:33 Instrument Setup graph has incorrect column legend
KPR#:38 ALS errors abort sequence; no Retry/Abort dialog box option offered with 68xx RC.NET Driver
KPR#:41 Method parameters from B.01.03 changed to defaults after upgrade to B0402 RC.net
KPR#:55 Opening 68xx RC.NET GC types result in 'FetchCPVar cannot talk to application'
KPR#:59 In 68XX Driver Version 5.02, it is not Possible to Start the 6890 RC.NET from the Front Panel Start Key
KPR#:66 CIS4 inlet is not recognized with 6890 GC RC.NET
KPR#:83 Only Fast Plunger Speed is available for PTV Solvent Vent multiple injections
KPR#:87 IQ1.CFG file of 6890/6850 instrument corrupted after regional setting change
KPR#:89 Incorrect injection volume shown in method report if PTV solvent vent is used
KPR#:91 B.04.03 DSP1 [52] Memory Leak may cause GC ChemStation with 68xx RC.Net driver to stop working
KPR#:92 Instrument Diagram Does Not Match Column 2's Inlet Configuration
KPR#:94 68xx Method Download Times Out
KPR#:95 "Method modified after loading" occurs when running a sequence (6890 RC.NET)
KPR#:98 GC Column Catalog File Cannot be Opened When Using the 68xx Driver Version 6.10
KPR#:99 Sequence continues while GC temperature is not ready
KPR#:100 A Method Can be Created Which Will Not Download to the Instrument and the Sequence Does Not Run
KPR#:101 GC Back Injector Not Present in Edit Method Screen
KPR#:102 Validation failed shown on left corner of Resolve Method screen
KPR#:103 Injector Panels in GC Edit Method Show as Uninstalled When an Injector is Installed.
KPR#:105 Plunger Speed is Not Saved with 6850 GC Methods
KPR#:106 Opening the GC method parameters to only look (no change made) and selecting OK to close stops the sequence or single run
KPR#:109 Method Contents Report Displays the Incorrect Configuration of Front and Back Detectors for 6890 GC
KPR#:113 Cannot Run a Packed GC Column with a 6890
KPR#:116 Pulsed Split/Splitless Injection Does Not Wait For Ready on a 6890.
KPR#:117 In 68XX GC Driver Version 6.01, 2 Injections are Made When Using the PTV
KPR#:119 Vial left in gripper; no message shown to alert user about the problem
KPR#:120 6850 vial numbers are not consistent with standard numbering system.
KPR#:122 Method Resolution Changes Oven Temperature Program
KPR#:123 Column info is missing with 6890
KPR#:124 Method Modification Flag is Set after Edit Method: Occurs on first connection or after upload from GC.
KPR#:128 Method Ddownload Failure Occurs Due to Column Calibration Limits
KPR#:129 Instrument actual value for COC inlet does not display correctly
KPR#:130 Instrument actual value for anode purge flow on an ECD does not display correctly
KPR#:134 Easy SamplePrep reset of one or any group of resources creates log entry: 'User reset all resources'
KPR#:136 Data Files do not Contain the Correct Signals if the Front Inlet is Connected to Back Detector

Keyword: 7890A Control

KPR#:911 Initial valve position set to ON before start when it should be OFF; Meth change
KPR#:914 7890A: ALS errors abort sequence; no Retry/Abort option offered
KPR#:985 7890A: Auto Start Macro fails -9999
KPR#:1049 7890A: Method modified after loading if run started from GC keyboard start butto
KPR#:1110 7890A Full report does not contain ALS module type information
KPR#:1117 Can Accept Missing Location Vial in Sequence Table
KPR#:1151 7890A: Signal data is not acquired if oven hold time is set to 0.0
KPR#:1169 7890A seq aborts part way through run: "AgGC7890Drv" already has ownership"
KPR#:1198 7890A does not show the vial number being injected on the sequence screen
KPR#:1239 7890A Runs cannot be started until the GC and injector are idle, view GC Status
KPR#:1255 7890A Parameter Mismatch error after upgrade
KPR#:1257 GCI1 did not start a run shown in instrument logbook
KPR#:1283 7890A Two inlets with one detector and one injector causes method resolution
KPR#:1329 ALS errors abort sequence; no Retry/Abort dialog box option offered
KPR#:1462 7890A with no tray will display the Sampling Diagram
KPR#:1463 7890A Extended Parameters enables Bar Code Reader section when BCR is not present
KPR#:1569 Sequence aborts if vial not found with Overlap injection enabled (7890A only)

Keyword: 78xx Driver

KPR#:5 Error Message when certain methodsare loaded onto an offline configured 7890.
KPR#:15 Shutdown events are not displayed on the software Status screen
KPR#:17 Barcode mismatch error
KPR#:26 Sample Prep Program with only a 7693A Back Injector does not save
KPR#:37 ALS Errors Abort Sequence; no Retry/Abort dialog box option offered
KPR#:54 Method edit: Must click twice with the mouse on each data field to change number
KPR#:67 ChemStation does not show "Hydrogen Safety Shutdown" in 7890A status
KPR#:75 7890A valve names added in the config screen but are not available in the method
KPR#:76 7820 method printout includes parameters and setpoints unavailable for PP inlet
KPR#:77 7820 method printout inserts negative number for Septum Purge Flow, invalid parameter
KPR#:80 If 7890 gets into an error status (flame off) this is NOT reported within the 7890 status window
KPR#:84 Read barcode just before injection and use barcode as sample name in report
KPR#:85 7890A: Column part number is not included in the method report
KPR#:90 Partial data collection is not available on the signals screen
KPR#:91 Need ability to do Digital switching for detector signals in Runtime events
KPR#:93 Read barcode just before injection and use barcode as the sample name in the report
KPR#:98 Park during sequence with Overlap causes unresponsiveness
KPR#:100 7890GC: inlet pressure unit inconsistency - method audit trail vs instrument setup
KPR#:103 RTL: New calculated pressure doesn’t update column 2 - 6 on current 7890A methods
KPR#:115 7890A: Shutdown events are not published to the software GC Status screen
KPR#:117 Upgrading B.01.03 to B.04.02 GC ChemStation: Addition of new columns to the column catalog (database) is not possible; error occurs
KPR#:118 Electrometer on/off not needed for FID or NPD
KPR#:124 Freeze Signal Value and Resume Signal Value missing from runtime events
KPR#:132 Error in flow setpoint upload when method has more than 100 events
KPR#:134 Running a sequence requires the arm from the tray, G4514A to be in the UN-PARKED position
KPR#:137 Extended Parameters does not disable the BCR when no hardware is present
KPR#:139 Switched and standard mode for septum purge not available on MMI
KPR#:140 Negative Split Splitless inlet Pressure values during Post Run Backflush
KPR#:154 Column compensation menu is available during the sequence
KPR#:166 Column: need to include Ramped and Constant pressure and flow modes as selection
KPR#:167 GC goes offline momentarily because of SOAP Exception
KPR#:170 No indication of inlet pressure shutdown in software status with 7890A
KPR#:174 GC Keypad Lock Behavior
KPR#:182 GC Instrument Status for Detectors
KPR#:190 FID Flame out; no feedback in software
KPR#:195 uECD and TCD report printout for makeup flows is wrong
KPR#:196 Control of PCM used as inlet EPC does not work in B.04.03 [52]
KPR#:200 GC continues when STOP on 7820 is pressed
KPR#:203 Open GC edit parameters and close it without changes; method modification flag is set
KPR#:204 7890A logbook shows 7693A turret, plunger errors as GCI1 did not start a run
KPR#:209 Unable to set equilibration time for LTM columns from Edit Parameters
KPR#:211 Changes made to 7890 method from GC instrument diagram result do not load.
KPR#:220 Modifying the Syringe Size When Using a 7890A GC
KPR#:223 Backflush icon is not displayed when the outlet of the column configuration is set as "Other".
KPR#:227 When running Backflush from ChemStation with segmented columns only the in-segment is taken into account in the calculation.
KPR#:228 The display of the parameter setting screen of the method setup is small.
KPR#:229 After editing the GC method and selecting ok - validation error message is displayed.
KPR#:230 7820 Back Injector Only Resets Injection Source to Manual
KPR#:238 PCM Inlet Heater Shutdown
KPR#:247 Multiposition valve resets to position # 1 between injections
KPR#:252 MMI in Split or Pulsed Split Mode does not allow editing of Total Flow value
KPR#:269 ALS Behavior When Aborting a Sequence
KPR#:277 Logging of Injection Tower Location
KPR#:282 Instrument Actuals Show "GC is Offline" Quickly at the Method or Sequence Start, then Returns Back Online and the Method Does Not Run
KPR#:348 B.01.01 driver on Multi-Technique ChemStation (MTCS) B.04.03.105 SP2 fails IQT: 76 Invalid files
KPR#:357 Inlet pressure setpoint displayed assuming all thermal zones are turned ON, results in confusing mismatch when heated zones are Off.
KPR#:358 IQT report failed on MT ChemStation B.04.03 SP2 with GC Driver B.01.01
KPR#:360 Method Editor for PP inlet & Packed Column incorrectly allows both flow and pressure setpoints to be entered
KPR#:1196 7890A: Shutdown events are not published to the software GC Status screen

Keyword: ACQ

KPR#:12 Time stamp errors in LC Chemstation & Chemstore, Win 2000 time format ignored.
KPR#:299 error "no string 8008" in sample information
KPR#:308 'Active signals" comment is chopped in MSD TIC title.
KPR#:375 Agilent Active Splitter will sometimes not start up correctly after installation
KPR#:448 Snapshot is not created if method or sequence name is of 40 characters long
KPR#:491 FIA runs with Dual Loop Autosampler are not reproducible
KPR#:492 Acquiring multi-signal data with low SIM Dwell times causes erroneous data.
KPR#:777 Main root path as data path in Preferences does not work for data acquisition
KPR#:1316 ChemStation switches to default path without warning if data path not available
KPR#:1321 sporadically the next run in a sequences will not start
KPR#:1564 Windows menu not responding after running ChemStation for long time
KPR#:1641 Not possible to modify data analysis parameters of method during a single run in the offline; cannot save
KPR#:1660 No special characters should be used for names of methods, data files or sequences
KPR#:1701 Terminal Service can close under high workload
KPR#:1732 Running dual tower sequence with all methods saved as Dual, get “Injector doesn’t match the sequence’s DUAL Injector”
KPR#:1742 Changes in sequence methods made during acquisition are not saved automatically
KPR#:1747 Queue Method vial range reflects range of loaded method, not the method browsed to
KPR#:1841 Run time for unsynchronized modules is shortened
KPR#:1842 Memory leak in RDPShell.exe may cause significant performance decrease on OpenLAB CDS ChemStation AICs
KPR#:1960 Chromatograms of manually stopped runs are shorter than expected
KPR#:1964 Data system uses the currently loaded method to validate the vial range in the Sequence table
KPR#:1528 blank runs are deleted from disk because of a setting in Method Scouting Wizard
KPR#:1763 In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.

Keyword: Audit Trail

KPR#:1122 report preview or printout from Report Layout does not trigger report history
KPR#:1133 deletion of compound group not tracked in method audit trail
KPR#:1749 Method Audit Trail does not include change of Injection source when HS/PAL is no longer configured
KPR#:1819 UIB II - set a timetable and control "Method Audit Trail"

Keyword: Batch

KPR#:1861 Cannot open batch - specified range parameter is bad

Keyword: CE Calibration

KPR#:263 CE system: Pressure are not applied correctly
KPR#:302 CE: Online plot does not adjust into negative not into y axis
KPR#:307 Not ready condition won't go away after applying external pressure
KPR#:536 CE: Partial Sequence initializes very slowly directly after loading a method
KPR#:682 CEMS configuration editor issue removing the MS part
KPR#:881 CE replenishment will stop sequence
KPR#:1015 Number format causes autostart macro error #16289 during CE ChemStation start

Keyword: CE Control

KPR#:1432 Method and Run Control not displayed correctly after reconfiguring the instrument from G1600 CE to G7100 CE
KPR#:1587 The VialTable check does not occur during the startup of the CE ChemSation
KPR#:1856 Device Driver Error with G7100 CE after update to B.04.03 SP2 [105]
KPR#:1664 CE G1600 instrument cannot be launched

Keyword: Calibration

KPR#:9 Calibrated compound found as uncalibrated one using library search report
KPR#:889 CE ChemStation: Update calibration level ignores area correction
KPR#:1089 Run a sequence with recalibration, only the first ISTD is updated
KPR#:1766 Calibration method printout does not print compound level's Heights when quantitation is based on height
KPR#:1793 Re-establish the old layout for calibration reports
KPR#:1799 Changing calibration table in Batch review generates an error
KPR#:1807 Calibration method printout does not print compound level heights when quantitation is based on height
KPR#:1809 C.01.04 HotFix03 : Amount units in calibration report printout is not updated

Keyword: ColumnComp Configuration

KPR#:22 TCC Cluster, TCC2 shows always valve position 1
KPR#:11 PumpValve-Cluster & TCC-Cluster halts sequence once methods change valve position

Keyword: ColumnComp Configuration , ColumnComp Driver , ColumnComp Status

KPR#:22 TCC Cluster, TCC2 shows always valve position 1

Keyword: ColumnComp Configuration , Pump Configuration

KPR#:11 PumpValve-Cluster & TCC-Cluster halts sequence once methods change valve position

Keyword: ColumnComp Driver

KPR#:22 TCC Cluster, TCC2 shows always valve position 1

Keyword: ColumnComp Status

KPR#:22 TCC Cluster, TCC2 shows always valve position 1

Keyword: Common Infrastructure

KPR#:18 High load on CAN may cause “Spectra runbuffer overflow” entries and a loss of spectra
KPR#:23 LC instrument remains offline
KPR#:25 Wrong Flow setting displayed when re-opening the Method User Interface

Keyword: Companion

KPR#:1200 Companion: Adding runs to run table is not possible while a sample is running
KPR#:1278 Companion: operator names cannot have spaces; invalid entry error
KPR#:1403 Companion: Adding runs to run table is not possible while a sample is running
KPR#:1474 Cannot see data generated by Companion in the ChemStation explorer view
KPR#:1609 Companion fails to run blank run for valved/ALS injections

Keyword: Config Editor

KPR#:1 Sequence and method path cannot be changed from the Configuration Editor
KPR#:275 LC Chemstation allows configuring a GC6890

Keyword: Custom Fields

KPR#:1344 XML import does not support Custom Fields
KPR#:1771 Custom Field content might get lost when a sequence is edited
KPR#:1789 Custom fields are not transferred to new result set
KPR#:1810 Missing Sample Custom fields in the report
KPR#:1837 Empty Custom Fields are not printed in Intelligent Report

Keyword: Custom Reports

KPR#:6 Add ability to insert JPG, GIF or BMP graphical file formats in Report Layout.
KPR#:208 Custom Reports General cal.Signals always reports uncalibrated peaks
KPR#:261 Excel area % report based on hparea00.xls not created when no peaks identified
KPR#:980 Relative Retention Time solution does not work for uncalibrated peaks
KPR#:1242 Response range from Signal Options not applied to custom reports

Keyword: DAD/MWD Driver

KPR#:21 Incomplete signal description in the online plot window
KPR#:10 G1315A/B do not show spectral data and G1321A/B show wrong time stamps for spectra

Keyword: DAD/MWD Driver , DAD/MWD Status

KPR#:21 Incomplete signal description in the online plot window

Keyword: DAD/MWD Driver , FLD Driver

KPR#:10 G1315A/B do not show spectral data and G1321A/B show wrong time stamps for spectra

Keyword: DAD/MWD Status

KPR#:21 Incomplete signal description in the online plot window

Keyword: Data Analysis

KPR#:15 Cannot resize chromatogram after you have resized it to a little box
KPR#:20 Chemstation Data Analysis Overlay function shows limited signals per window
KPR#:365 Multiplier difference if used in sequence table or calibration settings
KPR#:421 hpistd00.xls fails to produce ISTD report
KPR#:447 DA.M available for "Aquisition only methods" run in a sequence table
KPR#:493 Sequence Summary Reports do not work with Offline DA
KPR#:494 Running with Offline Data Analysis does not allow calibration .
KPR#:531 Signal Overlay toggle buttons work inverse
KPR#:555 error message "Not a range" when opening a data file with very long sample info
KPR#:721 sequence container is not loaded if Blank(s) in the sequence name
KPR#:745 Configured colors set in configuration editor not taken
KPR#:793 Calibration Table: From/To column not correctly updated
KPR#:846 Tic marks of fraction collector disappear after integration
KPR#:913 Autoscale doesn't work in GC Chemstation when loading next file
KPR#:972 data files cannot be loaded when the complete data path exceeds 140 characters
KPR#:1007 Certain chromatograms cannot be displayed in user scale
KPR#:1032 ChemStation explorer content may disappear
KPR#:1039 Additions to compound name do not show / print correctly
KPR#:1069 Data analysis autoscales incorrectly
KPR#:1113 error messages when switching tasks during reprocessing in Data Analysis view
KPR#:1191 Need to use different data analysis method for second signal, 3rd signal, etc
KPR#:1249 Cannot reprocess from Run Time Checklist
KPR#:1354 Chromatogram will rescale when switching between integration and calibration
KPR#:1361 Isoplot for FLD data shows wrong range for EM wavelength
KPR#:1375 impossible to load only one signal from all runs of current sequence and then overlay with runs from another sequence
KPR#:1467 Automatic GPC calculation might not be triggered on fast PCs
KPR#:1578 Reprocessing copy starting as a background window.
KPR#:1639 Data System will not "Autoscale" a TCD data file to the correct peak
KPR#:1651 Summed Peaks calculation error in the Summed Peaks Report
KPR#:1652 Y-axis labeled incorrectly when loading GC signals acquired with MSD ChemStation
KPR#:1654 ChemStation B.04.03 [54] will not "Autoscale" the TCD data file
KPR#:1706 Red X mark displays in data explorer of offline data analysis
KPR#:1724 Last report of a reprocessed sequence is not visible.
KPR#:1762 Error "An additional register must be added to the list of locked registers"
KPR#:1768 Baseline not displayed properly when using ChemStation AIC’s
KPR#:1796 Report template not found when reprocessing with a master method
KPR#:1813 Data analysis: Fractions are not displayed in UIB chromatogram
KPR#:1843 Snapshot runs cannot be integrated manually
KPR#:1857 Error message: "An additional register must be added to the list of locked registers".
KPR#:1727 User is not prompted to save manual integrations to file, when there are manual events in the method.
KPR#:1845 Garbage characters can appear in English LCMS reports
KPR#:1688 Preferences set in offline overwrite online Preferences and vice versa
KPR#:1717 Buttons for manual integration are not accessible, while they can be selected through the menu

Keyword: Data Analysis , Navigation Table

KPR#:1727 User is not prompted to save manual integrations to file, when there are manual events in the method.

Keyword: Data Analysis , Programming

KPR#:1845 Garbage characters can appear in English LCMS reports

Keyword: Documentation

KPR#:1575 HTML Help requires Internet Explorer revision 6.0.28.00.1106CO or higher
KPR#:1734 Understanding Your ChemStation manual needs more information about peak width calculations
KPR#:1783 G2613A 7683A injector firmware is incorrect in the Firmware guide

Keyword: ECM Integration/OpenLab Option

KPR#:585 wrong file extension for SSizip after navigating in the dialog for saving to ECM
KPR#:593 ChemStation reports with CyberPrinter limited to 30 characters as file name
KPR#:744 ECM->Save Data As changes default SSIZIP file name
KPR#:774 data review does not update the .acaml file
KPR#:966 CE .ACAML file: measured mobility and corrected area are missing
KPR#:1091 ChemStation Administration tool cannot be started right after installation
KPR#:1101 upload process to ECM checks for server name in case sensitive way
KPR#:1102 user without corresponding privileges can save report template
KPR#:1107 user without privileges can preview and print reports in Report Layout view
KPR#:1128 incomplete view initialization when unlocking ChemStation
KPR#:1136 "System out of memory" exception on uploading large sequence containers to ECM
KPR#:1156 ECM Operator in case of data modification during acquisition and forced upload
KPR#:1204 ECM path dropdown box does not support tokens
KPR#:1210 OpenLAB Intelligence report can't be created for single runs from ChemStation UI
KPR#:1212 OpenLAB Option: Difficulties to lock the ChemStation during reprocessing
KPR#:1215 Problems after moving items from ChemStation to a different LCDF path in ECM
KPR#:1235 upgrade installation resets the settings in the Administration Tool
KPR#:1301 ChemStation instrument not responding during upload of large data sets to ECM
KPR#:1452 report history may display more peaks than the actual report
KPR#:1453 chromatogram data not included in the ACAML file when a sequence is acquired in "Acquisition only" mode
KPR#:1458 incorrect injection volume in the .ACAML file when using injector program
KPR#:1479 there is no possibility to process items in the queue that are not related to the one that halts the queue
KPR#:1487 there is no indication in the ECM Logon dialog whether an account was disabled
KPR#:1496 ECM Audit Trail contain a large number of audit trail entries for each file e.g. inside a sequence container
KPR#:1523 user privileges not applied when logging on to ECM in ChemStation
KPR#:1525 message about failed upload to ECM sometimes not displayed
KPR#:1546 ECM related information missing on report for interactive reports created while sequence is still running
KPR#:1592 container change flag is cleared after rejecting automatic upload to ECM
KPR#:1599 Startup time of ChemStation is long when there are many items in the OpenLAB Option spooler queue
KPR#:1611 Path definition in OpenLab Control Panel is not used in ChemStation
KPR#:1659 The storage path for files in ECM is user-specific
KPR#:1672 it is not possible to remove AIC from networked workstation
KPR#:1714 ChemStore legacy data doesn't get reprocessed automatically
KPR#:1753 Wrong error message when downloading a method from ECM
KPR#:1806 Update methods from ECM dialog locks up
KPR#:1818 Submitting multiple sequences to queue, the User token in upload path is resolved to current user - not to submit user
KPR#:1833 In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:1545 OpenLAB Option: data cleanup during shut down ignores pending .PDF reports
KPR#:1635 Using the Abort action will leave the system in an undefined state
KPR#:1763 In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.

Keyword: Easy Sequence

KPR#:1568 HeadSpace does not get number of samples from modified .S sequence originally run from Queue

Keyword: FLD Driver

KPR#:10 G1315A/B do not show spectral data and G1321A/B show wrong time stamps for spectra

Keyword: GC ACQ

KPR#:667 Displaying method in control module prevents other meth download during seq
KPR#:1184 Offline crash requires reboot of PC
KPR#:1277 6850/6890 logbook does not advise of invalid inj volume overrides from seq table
KPR#:1423 Low Thermal Mass (LTM) logging should be in Seconds and not Minutes
KPR#:1459 ChemStation displays -1.0 during CTC cycle process pre and post run
KPR#:1460 When look-ahead vial is not found in a sequence using overlap, the currently running sample and sequence will abort
KPR#:1608 FetchCPVar Cannot Talk to Application Error
KPR#:1628 FetchCPVar Cannot Talk to Application Error
KPR#:1630 Column compensation cannot be started from the Data System with 68xx RC.NET
KPR#:1638 G1888A - Handshake Abort - Vial Status does not update
KPR#:1683 OpenLab CDS, C.01.02, 7820A run continues when STOP on 7820A is pressed
KPR#:1700 7890A stops run, ChemStation keeps going; GC Connection state is MONITOR in Status
KPR#:1759 Capacity factor (k') is not calculated; column information not printed in Performance report

Keyword: GC Analysis

KPR#:1741 Self-assembled result sets do not have full functionality for GC data
KPR#:1825 Customized GC ChemStation validation may fail

Keyword: GC Autosampler

KPR#:1738 Unclear error message: Undefined symbol: License$

Keyword: Help

KPR#:549 Selecting RTL Help Contents brings up ChemStation Help not RTL Help
KPR#:1105 Help for the Purification Software can not be displayed on Windows Vista
KPR#:1515 GPC help does not open on Win7 32bit
KPR#:1613 Help entry links from the logbook are sometimes unrelated

Keyword: Help , Logbook

KPR#:1613 Help entry links from the logbook are sometimes unrelated

Keyword: ISOPlot

KPR#:325 Isoabsorbance and 3D plots not available in data acquired with two DADs

Keyword: Initialization

KPR#:33 Opening Reprocessing copy opens also C:\WINNT\System32\CMD.exe
KPR#:59 unexpected information box when initializing Dual Loop Sampler
KPR#:289 ChemStation does not launch when two MWD detectors are configured
KPR#:294 Problems with initialising the 1100 instrument when a DAD G1315C is present
KPR#:372 ChemStation startup failure in case of printer name exceeding 92 characters
KPR#:1345 CE: macro error in instrument logbook after startup without modules
KPR#:1488 configuration of 1120 Compact LC lost after upgrade to B.04.02 DSP3
KPR#:1618 Sporadically launch of an online session will not proceed
KPR#:1664 CE G1600 instrument cannot be launched

Keyword: Initialization , CE Control

KPR#:1664 CE G1600 instrument cannot be launched

Keyword: Installation

KPR#:18 Installed two A/D licences, one instrument appears
KPR#:327 IQ fails after upgrade to B.01.03 if LCMS patches are previously installed.
KPR#:377 Multimode Ion Source Report Will Sometimes Not Print Out
KPR#:505 Retention Time Locking B.01.02 SW will not install on GC Chem B.02.01
KPR#:554 uninstallation of B.02.01 SR1 (upgrade installation from B.xx.xx)
KPR#:565 IQ fails after upgrade of CE/MS ChemStation B.01.03 to B.02.01 SR1
KPR#:578 Upgrade of G2730BA LC/MS Data Analysis only from B.01.01 fails
KPR#:582 duplicate Start Menu entries after upgrade of 2D/3D Data Analysis only
KPR#:591 problems with CE/MS upgrade installations
KPR#:702 GPC software does not work if installed on a path containing a blank
KPR#:806 CE IQ is failing after G2201BA CEMS addon uninstallation
KPR#:820 IQ fails after an upgrade to B.03.01 (LC 3D ChemStation)
KPR#:865 SR2 for B.02.01 does not update Help -> About for Config Editor and IQ tool
KPR#:899 With Chinese ChemStation B.03.01 SR1 IQT fails for G2090BA, G2190BA, and G2730BA
KPR#:1004 Chinese ChemStation: IQ fails after upgrade from B.03.01 to B.03.02
KPR#:1097 system files may appear as different in the IQT report
KPR#:1098 Cancelling the installation does not rename the ChemStation backup folder
KPR#:1124 ChemStation SR Entries in Add/Remove Programs after upgrade installation
KPR#:1219 B.04.01SP1 upgrade installation removes icons of add-on products from Start Menu
KPR#:1236 Uninstallation of ChemStation does not detect addons
KPR#:1269 Could not install ChemStation after removing previous ChemStation Installation
KPR#:1381 ChemStation B.04.02 on bundle computers cannot be upgraded to DSP1
KPR#:1404 Installing B.04.02 sometimes corrupts the Add/Remove programs list
KPR#:1431 after upgrade from ChemStation B.01.03, screen of UV/Vis ChemStation is displayed during startup
KPR#:1445 Upgrade Installation fails after changing the installation path
KPR#:1512 Configuration Assistant hangs when another controller is connected to the instrument
KPR#:1589 missing link and no automatic installation of the localized readme.txt of the CDS ed.
KPR#:1627 AIC ports not opened for 7890A acquisition
KPR#:1646 Ugrade to B.04.02 SP1 Hotfix 2, the preference setting "integrate after load" is lost in some instances

Keyword: Instrument Control

KPR#:90 Methods Must be Stored on the C Drive

Keyword: Integration

KPR#:298 manual integration not displayed when separate zooming is enabled
KPR#:413 Height-Reject is not taken in account during manual integration
KPR#:690 Reporting of negative peaks resulting from Area Summing
KPR#:694 Integration events values not saved by edit entire method
KPR#:981 Integration unable to handle complex chromatogram for area sum integration event
KPR#:1010 Reintegrate "all signals" with Int. Events open only applied for one signal
KPR#:1050 Batch.M does not display special area annotation for manually integrated peaks
KPR#:1121 integration results of manually split peaks may differ from results before split
KPR#:1123 riders found using new exponential baseline mode can not be deleted manually
KPR#:1252 timed area reject values ignored by manual integration
KPR#:1369 timed area reject values are ignored by manual integration
KPR#:1471 Integration unable to handle complex chromatogram for area summation
KPR#:1534 User is not asked to save manual integration events when manual events are stored in the method
KPR#:1607 Manual integration can cause unexpected results
KPR#:1642 Data lost when using area sum integration event, must have 0.06 minute space between groups
KPR#:1718 Some peaks in a chromatogram cannot be manually deleted
KPR#:1744 Misleading display of tangent peak width
KPR#:1760 Last Result Mode may not reflect interactive changes of integration events
KPR#:1777 Manual deletion of skimmed peak does not work
KPR#:1800 Timed integration event "Height Reject" cannot be configured.
KPR#:1838 GC compound peaks might be labelled as solvent peaks

Keyword: Localization (Japanese)

KPR#:962 impossible to input Japanese Kanji in Name Pattern field on Preferences dialog
KPR#:967 Japanese ChemStation: VWD Dark Current test fails
KPR#:1115 msupdate.exe cannot be executed on Japanese Windows Vista
KPR#:1668 Japanese characters from saved pdf report are garbled when copied and pasted to another application

Keyword: Logbook

KPR#:1613 Help entry links from the logbook are sometimes unrelated
KPR#:557 DDE Comm Failure when starting 6890 with G1888 HeadSpace
KPR#:803 Only 20000 lines are kept in the sequence log
KPR#:815 First run of 7890A always generates error in Logbook: undefined symbol, 41276
KPR#:1405 column temperature and pressure not in the logbook when using RC.NET drivers for LC
KPR#:1547 ChemLaunch: ChemStation sequence logbook and instrument logbook are not accessible
KPR#:1637 'Method not downloaded to device' message when PAL is present

Keyword: MSD Analysis

KPR#:4 Error "Identifier (MSO_PPwidth) is illegal or variable can't be manipulated"
KPR#:103 Cannot install G1043A NIST Library Search Program on B.01.01 LC/MSD ChemStation
KPR#:124 Data Browser rev. A.01.01 not supported with rev. B.01.01 LC/MSD ChemStation
KPR#:125 G2720BA Bioanalysis SW not available at release of rev B.01.01 LC/MSD ChemStati
KPR#:267 Batch review, no manual integration id smoothing is selected in signal details
KPR#:310 Sample Purity affects integration events settings, sets method change flag
KPR#:314 TIC report displays only one MS parameter when in FIA mode
KPR#:1665 Deconvolution tools are not available for LC/MS
KPR#:2125 Peak Performance calculations are off when data points are not equidistant
KPR#:1616 Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group

Keyword: MSD Analyst

KPR#:312 .wiff file cannot be opened in Analyst if the method has a Post time set .
KPR#:313 Analyst does not launch automatically if root directory is not default one.
KPR#:319 Analyst conversion log file shows warning that CRC doesn't match

Keyword: MSD Control

KPR#:14 Get application error in DLLHOST.EXE when setting up for running Multiple FIA
KPR#:123 Easy Access rev. A.03.00 not supported with rev. B.01.01 LC/MSD ChemStation
KPR#:309 Abundance Values missing from Lens Ramps
KPR#:318 Display of Time Delay is incorrect for MS acquisition while SIM on Target Mass.
KPR#:320 G2730BA Standalone DA displays Sample Purity Options dialog.
KPR#:1437 ChemStation with CTC sampler hangs up when stopping or aborting a sequence
KPR#:1675 Automated Delay Calibration of the MS signal for Fraction Collection is not supported in the RC.Net environment
KPR#:1686 MSD Tune View will awlays load ATUNES.TUN if tune file loading is cancelled
KPR#:1730 LC/MS may remain in a run state after Method/Sequence termination due to an HPLC hardware error

Keyword: Method

KPR#:10 AM/PM in time for entries in Method History, also with 24 hour time format
KPR#:23 Pressure unit is not correct on CE ChemStation method printout
KPR#:30 1050 VWD method printout shows too many digits for the Peak Width setting
KPR#:114 During Edit Entire Method, two dialog boxes come up simultaneously
KPR#:697 Major error at line 433 in file X:\Astxdrivers\iq\idd\src\ui\ssl1.c
KPR#:1202 6850 Workstation Internal SW error occurred when configuring a column in 6890
KPR#:1343 "Error occurred during loading method" after rejecting method resolution
KPR#:1353 GC Offline Data Analyisis checks for ALS location when loading method
KPR#:1466 different order of instrument parameters in method printout with classic and RC.NET LC drivers
KPR#:1469 Misleading parameter mismatch message in Data Analysis during data review
KPR#:1473 CTC portion of method printout does not include overlap settings
KPR#:1490 method for Compact LC is marked as changed after loading
KPR#:1517 "Method changed flag" set after silent LC method update
KPR#:1558 Changing trays in Well-Plate Samplers does not trigger LC method resolution
KPR#:1598 Update Master Method does not update the GPC settings
KPR#:1624 Undefined symbol: PrDARep error displayed when selecting Export e-Method
KPR#:1643 Open GC edit parameters and close it without changes; method modification flag is set
KPR#:1650 Headspace or PAL Sampler sequence with methods saved to a location other than C:\ does not run
KPR#:1663 No information around handling for special characters
KPR#:1694 Method marked Dirty on initial "Specify Report ..." with DEF_GC.M
KPR#:1710 Cannot transfer an English method to a Japanese or Chinese ChemStation and vice versa
KPR#:1748 Add/Removal of 7697 HS from 7890 does not show a change occurred; not displayed in Method Load options; method not shown as modified
KPR#:1750 Cannot save offline method if configuration is different: 'Method is not consistent with the configuration'
KPR#:1765 "Chemmain has encountered a problem and needs to close" when selecting back/dual inj position for GC manual inj
KPR#:1802 C.01.03 upgraded to C.01.04: 'Edit Entire Method' not working
KPR#:2021 ACQ.txt file contains Column(s) information that does not match actual column used and leads to confusion
KPR#:1621 sequence method of additional sequence lines may be overwritten by another method

Keyword: Method , SEQ

KPR#:1621 sequence method of additional sequence lines may be overwritten by another method

Keyword: Method Scouting Wizard

KPR#:1720 Cluster information can be overwritten from the offline session
KPR#:6 Memory Consumption
KPR#:7 Memory Problem when clicking online help
KPR#:8 Pressure spikes seen when switching columns
KPR#:9 Error message when starting Method Scouting wizard about not available paths
KPR#:1333 Method scouting sequence does not start due to missing methods
KPR#:1482 Misleading UI for equalibration and column storage solvent flows
KPR#:1483 Error message when generating a printout of the campaign
KPR#:1484 Unexpected program termination while reviewing an existing campaign
KPR#:1555 Option to Change Vial after n injections might not work in a sequence
KPR#:1556 Injection counters on tagged columns do not get updated
KPR#:1561 Sample info for blank runs might not be accurate
KPR#:1528 blank runs are deleted from disk because of a setting in Method Scouting Wizard

Keyword: Navigation Pane

KPR#:719 Performance problem if many data files to be shown in Navigation Table

Keyword: Navigation Table

KPR#:1727 User is not prompted to save manual integrations to file, when there are manual events in the method.
KPR#:836 Navigation table cannot display this date time format 20-03-2007 11:41:22
KPR#:1752 Changing the sample type in a sequence with cyclic calibration corrupts the sequence structure

Keyword: Output

KPR#:1612 Printing fails if printer name exceeds 100 characters

Keyword: Peak Purity

KPR#:7 A09.03 Peak Purity calculation
KPR#:453 not all decimals displayed for the Volume in printout of Fraction Details
KPR#:784 Purify: Only the study creator and administrators can access a study
KPR#:977 impure peak reported as pure under special conditions

Keyword: Programming

KPR#:1845 Garbage characters can appear in English LCMS reports
KPR#:171 Uni code problem with scheduler
KPR#:1617 AIC can experience communication loss if Application Experience service is running
KPR#:1755 The command "macro" does not list *.mac files
KPR#:1761 The command "macro " cannot be used to select *.mac files
KPR#:1831 Scheduler does not work on an AIC

Keyword: Pump Configuration

KPR#:11 PumpValve-Cluster & TCC-Cluster halts sequence once methods change valve position
KPR#:27 Misrepresentation of Bottlefillings in Dashboad of PumpValveCluster
KPR#:32 Cannot Check “Pump and solvent selection valve installed” Box for Flex Cube if Once Clear it

Keyword: Pump Driver

KPR#:24 Flow unit in Online Plot for Nano pump is incorrect.
KPR#:43 Memory leak in pump method dialog may cause performance drop

Keyword: RTL

KPR#:1392 Retention Time Locking: can only use kPa unit

Keyword: Reports

KPR#:21 spectra options are not printed correctly
KPR#:60 prefix and counter are not reported in hypersequence printout
KPR#:75 Study Run Summary Report: WPS: Tray description is missing
KPR#:81 Instrument curves printed twice, when set in "Sig. Details" and "inst.curves"
KPR#:82 Report: wrong behaviour in fract.table when injec.prog. longer than runtime
KPR#:116 time stamp issues for "inj. time" and "last changed time" for Rev.A Data
KPR#:164 Time Ranges not applied correctly in short report
KPR#:238 after manual integration and peak splitting RT and area counts double
KPR#:240 "Print all Windows" function is not printing correctly
KPR#:281 print to file: no csv file generated using performance and noise report style
KPR#:283 No report shown after method run
KPR#:285 XLS Performance and Noise Rep. : drift given in mAU/min
KPR#:315 Long FIA runs does not list all the injections in the report.
KPR#:316 No report due to "Quantification out of memory" command line error for FIA run.
KPR#:328 in the method printout two DADs have the same header
KPR#:349 "Use Multiplier and Dilution Factor with ISTDs" displayed in all reports
KPR#:373 Printing problems: black bars on the report
KPR#:420 Batch Report: "justify" for alignment is not taken in report/preview
KPR#:449 Group Summary report missing in CSV reporting
KPR#:450 Batch Report: duplicated results for multiple signals in Cal table
KPR#:458 Wrong unit printed on Delay Calibration Reprot
KPR#:459 The use of the manual injector not reported in method report
KPR#:464 page numbers on diagnostic reports are cut of on print out
KPR#:490 Chromatogram might not printed like defined in Graphics
KPR#:533 number of decimals for Capillary and Nano Pump flow in method printout
KPR#:590 Margins are used differently for the Print out of ChemStation Windows Printing
KPR#:722 Path naming on Standard Reports and Ex. Perf. Report not giving ECM info
KPR#:748 Report Layout: Footer print date earlier than last change date for calibrations
KPR#:755 7683 autosampler appears as 7673 on the ChemStation GLP report & meth print
KPR#:776 Report including Calibration Curves displays usage of height instead of area
KPR#:783 problems with chromatogram time and response axes in customized reports
KPR#:811 A report cannot be printed if you have overlaid signals
KPR#:812 7673 AutoSampler is printed instead of HeadSpace for Run Info in ChemStore
KPR#:864 Sequence Summary report does not include peaks that are grouped
KPR#:895 With HP LJ 4250 selection of output orientation does not work
KPR#:897 Not found peaks reported different in different report styles
KPR#:915 7890A Report heading does not include Injection Volume; always 'External'
KPR#:942 additional column "ISTD used" has no header in ISTD report of CE ChemStation
KPR#:949 wrong compound numbering in Extended Performance Report after removing compounds
KPR#:953 Japanese/Chinese CS: garbage characters in .DIF output of sequence summary
KPR#:996 inconsistent results in statistics report
KPR#:1011 ChemStation does not report resolution as defined in European Pharmacopoeia
KPR#:1108 6890N Full report displays 7683 ALS incorrectly as 7673
KPR#:1109 ChemStation unable to print to screen; Failure at end of run with Vista SP1
KPR#:1302 problem with page numbering in a long sequence summary report
KPR#:1305 Group sum results are not exported to the result.xml
KPR#:1348 "View Method" does not report sample related custom fields
KPR#:1355 Grouped calculations/sums will only work for reports using calibration curves
KPR#:1435 Sample Report is truncating long barcodes to 16 characters
KPR#:1440 acquisition parameters missing in the Method section of the sequence summary report when Unique Folder Creation is switched off
KPR#:1509 Paper Source selection is not kept
KPR#:1531 .pdf report misaligned when default font for the report is missing
KPR#:1563 problem when printing more than 8 overlaid chromatograms via Print Selected Window
KPR#:1572 Batch mode does not allow usage of intelligent reports
KPR#:1579 Report Templates newly added to a Result Set do not appear in the list of available templates
KPR#:1582 Mismatch in drawing of peak performance between Classic Report and Intelligent Report
KPR#:1619 Intelligent Reports do not contain ECM specific information in Review mode
KPR#:1634 B.04.03 DSP1 [54] Data Analysis, Graphics, response range does not work
KPR#:1661 USP tailing factor can be reported incorrectly with a value <0.
KPR#:1669 ACAML: GC injection volume not correctly given; it is showing more significant figures
KPR#:1673 CSV, DIF, and EMF files are not generated for some Classic report styles
KPR#:1679 sample name not reported with Intelligent Reporting in non-result set data that have been acquired prior to C.01.01
KPR#:1692 Date format is changed in sequence summary report from month/day/year to day/month/year
KPR#:1703 ChemMain.exe can crash when a long file path is used
KPR#:1705 Chinese/Japanese characters cannot be searched in the Report Viewer
KPR#:1711 The legend for calibration curves does not react on the calculation type
KPR#:1716 Report information such as chromatogram and results are truncated or cut off when saving to htm
KPR#:1722 Tangent peak width not correctly evaluated.
KPR#:1733 Mass- Based Purification (Fraction Collection) report missing information & shows inaccurate Fraction Tick marks (RC.NET only)
KPR#:1751 eSignature data file privilege is missing from the ChemStation lab manager, analyst, and operator roles
KPR#:1754 The structure of the reports to file has changed
KPR#:1756 CP3: Number format error in: %d when printing report to screen in Data Analysis
KPR#:1757 Error 41265 occurred at end of run; Number format error in: %d shown in logbook
KPR#:1775 Preview of classic reports fails with a macro error. 'Number Format error' in Sequence log file.
KPR#:1776 Chromatogram in On-Screen Classic report is not displayed properly
KPR#:1778 'Number Format error' in Sequence log file after printing report to screen
KPR#:1780 Sequence summary report sorts the data by injection time
KPR#:1786 A Sequence Summary Report after Partial Sequence reports all the runs of the full sequence
KPR#:1791 Macro error occurs (Parameter 2: Table not found) when printing a calibration table
KPR#:1792 LCMS specific report style options are missing from the Report dialog (Classic Reports)
KPR#:1801 C.01.03: XML results: The 'ESTD %' and 'ISTD %' result types in the Quant Calc node are incorrectly 'ESTD' and 'ISTD'
KPR#:1803 Acquisition method name is not displayed in an intelligent report
KPR#:1820 Missing Report Style Options for LCMS and LCMS Qualitative
KPR#:1835 Calibration data modification time stamp is set-off by 12 hours
KPR#:1839 File corruption may cause sequence runs appearing as single runs
KPR#:1974 Cannot change the Report Name for the Signal in the Classic Sequence Summary Report
KPR#:2119 Single Peak's Peak area is different from Total area for classical built in report
KPR#:2152 Classic Report area% calculation can provide incorrect results
KPR#:1837 Empty Custom Fields are not printed in Intelligent Report
KPR#:1545 OpenLAB Option: data cleanup during shut down ignores pending .PDF reports
KPR#:2125 Peak Performance calculations are off when data points are not equidistant
KPR#:1529 for very long sequence summary reports, error 41066 may occur

Keyword: Reports , Custom Fields

KPR#:1837 Empty Custom Fields are not printed in Intelligent Report

Keyword: Reports , ECM Integration/OpenLab Option

KPR#:1545 OpenLAB Option: data cleanup during shut down ignores pending .PDF reports

Keyword: Reports , MSD Analysis

KPR#:2125 Peak Performance calculations are off when data points are not equidistant

Keyword: Reports , SEQ

KPR#:1529 for very long sequence summary reports, error 41066 may occur

Keyword: Run Control

KPR#:293 ADC: "Location" field in "Sample Info" goes blank after first method run.
KPR#:841 Headspace instr will fail to start if HeadSpace is used as the injection source
KPR#:1851 Unexpected shifts in the Online Plot
KPR#:1635 Using the Abort action will leave the system in an undefined state

Keyword: Run Control , ECM Integration/OpenLab Option

KPR#:1635 Using the Abort action will leave the system in an undefined state

Keyword: SEQ

KPR#:1621 sequence method of additional sequence lines may be overwritten by another method
KPR#:1529 for very long sequence summary reports, error 41066 may occur
KPR#:54 Printout of Partial Sequence table misses essential information
KPR#:154 The data file name in a sequence summary compound report is limited to 8 charact
KPR#:241 No divisor available for samples in the sequence table
KPR#:259 sample name of a run changed when reprocessing parts of a sequence
KPR#:274 performance problem when many sequences and methods are present
KPR#:300 changing a plate in hypersequence table during leads to an error message
KPR#:356 Sample amount ignored for blank runs when entered from sequence table
KPR#:480 Method Name field blank in sequence table (in Data Analysis)
KPR#:778 Change of sample type during sequence acquisition is ignored
KPR#:780 Default hypersequence Def_LC.hyp can be overwritten
KPR#:813 Front injector is not installed but it is still available in the sequence table
KPR#:823 Sequence stops if vial number is 0 on an 1100/1200 system without error message
KPR#:927 Filldown wizard for CTC Increment Directions mixes up row/column directions
KPR#:938 Problems running more than 422 sequence lines with CTC sampler
KPR#:941 Online session closes when copy/paste is used to build seq table
KPR#:973 Sample location not displayed correctly in Data Analysis with high plate number
KPR#:1317 Mismatch for sequence completed time stamp in sequence and instrument logbook
KPR#:1322 sporadic ChemStation crashes after aborting a sequence
KPR#:1414 Purify: error message at the end of study
KPR#:1441 ChemStation crashes when trying to use a method containing an ' in the name in the sequence table
KPR#:1442 an error may occur when stopping a sequence immeadiately after start
KPR#:1470 ACQNOTREADYWAIT time cannot be changed in the user interface, but only via the command line
KPR#:1486 not possible to create an interactive report during post-time
KPR#:1501 wrong data file naming after missing vial with Prefix/Counter naming
KPR#:1514 The path to a data file has a limited length
KPR#:1541 sequence template may be marked as changed after modifying the sequence file in a sequence container
KPR#:1542 Sequence table ignores the 40 characters limit for file names
KPR#:1543 sequence data set shown as "in progress" although the sequence is longer acquiring
KPR#:1552 Sequence is marked as running although it is finished
KPR#:1554 Sequence reprocessing with methods from old revision fails
KPR#:1571 Missing Vial during sequences causes CTC ChemStation to hang
KPR#:1604 HeadSpace sequence method will not load when it follows a Valve sequence
KPR#:1636 Headspace or PAL Sampler sequence with methods saved to a location other than C:\ does not run
KPR#:1648 Sequence parameter subdirectory overwrites sample info parameters during run
KPR#:1687 Missing ‘Reprocessing Only’ in the Dropdown List for Unique Folder Creation OFF
KPR#:1708 B.04.03 DSP1 [54] Partial Sequence Acquisition Does Not Work with Headspace with the Back Sequence Table
KPR#:1712 Opening the front sequence table, click on a cell, make no changes, and select OK causes sequence to stop with 7697A Headspace Sampler
KPR#:1713 Opening the front sequence table, click on a cell, make no changes, and select OK causes sequence to stop with 7697A Headspace Sampler
KPR#:1725 Sample statistics are not done, when the first compond in the calibration table is not detected.
KPR#:1728 Pausing and resuming a sequence can change the data path.
KPR#:1729 Headspace or PAL Sampler sequence with methods saved to a location other than C does not run
KPR#:1735 Partial sequence run with methods saved to a location other than C:\ doesn’t run with headspace and PAL Samplers
KPR#:1736 Partial sequence with methods saved to a location other than C: does not run with headspace and PAL Samplers
KPR#:1737 Partial sequence with methods saved to a location other than on C does not run with headspace and PAL Samplers
KPR#:1767 GC: sequence does not continue if there is a missing vial, uploads data to ECM and goes to next sequence
KPR#:1787 78xx: Sequence stops when pushing the GC STOP key during sequence
KPR#:1788 Custom field window not prompted for Back injector on first run of the sequence
KPR#:1798 Sequence Reprocessing: Not possible to stop a sequence that is reprocessing
KPR#:1805 Sequence Import Mapping is missing LIMS ID columns
KPR#:1822 Sequence parameters: Changes to Fraction start location ignored
KPR#:1824 After upgrade, Sequence modified flag is set even if no changes are made before reprocessing
KPR#:1830 'Number Format error' in Sequence log file after reprocessing
KPR#:1870 Well Plate locations are not accepted in offline sequences
KPR#:2165 Partial sequence does not support run-to-run relations
KPR#:1528 blank runs are deleted from disk because of a setting in Method Scouting Wizard
KPR#:1763 In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.

Keyword: SEQ , ACQ , Method Scouting Wizard

KPR#:1528 blank runs are deleted from disk because of a setting in Method Scouting Wizard

Keyword: Sample Overlap

KPR#:808 Overlapped injections omits run when using the same vials for front and back inj

Keyword: Sampler Configuration

KPR#:9 Control of Tray illumination missing for new G1329A

Keyword: Sampler Configuration , Sampler Status

KPR#:9 Control of Tray illumination missing for new G1329A

Keyword: Sampler Driver

KPR#:16 Control of tray illumination missing for G1329A/B & G4303A in Japanese and Chinese ChemStation
KPR#:19 Injection valve cleaning, entering new values does not work

Keyword: Sampler Driver , Sampler Method

KPR#:19 Injection valve cleaning, entering new values does not work

Keyword: Sampler Method

KPR#:19 Injection valve cleaning, entering new values does not work
KPR#:30 Injection Volume not shown when loading a method (Inactive)
KPR#:31 ICF A.01.04 - No method download, if method contains pretreatment/injector program steps.
KPR#:33 Dua Loop Sampler: ACQ method shows wrong loop with overlap injection

Keyword: Sampler Status

KPR#:9 Control of Tray illumination missing for new G1329A

Keyword: Spectra Analysis

KPR#:920 Apex spectra not correctly reported
KPR#:1036 unexpected question whether user wants to save spectra library with the method
KPR#:1323 Issues with scaling of online plot for DAD G4212A
KPR#:1790 Mass spectra drawn with strange shapes
KPR#:1616 Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group

Keyword: User Interface

KPR#:13 Fraction collection module error message given at different Fraction Collector
KPR#:44 Signal Task: Time reference lines only visible with 256 colors on Compaq bundle
KPR#:62 Operator field in the Change access level dialog box retains only 29 characters
KPR#:68 instrument remains in running status after a diagnose test
KPR#:69 Fraction time table entries corrected by the biggest delay time
KPR#:74 "EMPV Cleaning Procedure" displays misleading status
KPR#:76 Inconsistent update of peak detectors table in Set Up Fraction Collector
KPR#:95 Fraction Preview: Shifted marker in Preview
KPR#:121 Datafile name in File information not updated for overlayed datafiles
KPR#:126 Hardware related zooming issue
KPR#:232 Using "Print window" is not clear what is the active window to print
KPR#:234 Minimize Logbook option is not available
KPR#:239 "switch off pump and lamp" question is coming even if alreay switched off
KPR#:286 Switching between OQ/PV to Data Anaylsis view impact the fraction table GUI
KPR#:322 in online plot it is difficult to differentiate between two DAD detectors
KPR#:446 MultiMode Source is not listed as a possible detector for the OQ/PV
KPR#:455 Recovery Mode not updated in Instrument Actuals GUI
KPR#:462 Incorrect status message of Chip Cube in case of lost synchronization
KPR#:466 Manual injector use is not present in User Interface
KPR#:486 Sample info desciption for OQPV Fraction Collector tests contains symbols
KPR#:496 System.NullReferenceException error may occur in ChemStation
KPR#:501 toolbars can only be configured when displayed in separate window
KPR#:517 option to create folders from Preferences/Paths only available in Windows 2000
KPR#:693 wrong error message: pressure sensor not connected
KPR#:723 pressure signal plot for the PrepPump is missing in Online Plot
KPR#:731 Error Fetching CP Variable observed sometimes when starting the chemstation
KPR#:809 Splitless Mode - GC displays 'Not Ready" - Chemstation displays "Ready"
KPR#:1045 Function keys not displayed on screen
KPR#:1046 Initial Screen Window Size is not kept
KPR#:1125 network datapath not available in ChemStation Explorer after network outage
KPR#:1194 HeadSpace software error box is not shown until the HeadSpace SW has the focus
KPR#:1230 ChemStation displays -1.0 during CTC cycle process pre and post run
KPR#:1319 ChemStation uses mixture of color settings from Windows and Configuration Editor
KPR#:1336 signal description in the online plot does not provide wavelength information
KPR#:1342 Verification view does not initialize properly
KPR#:1376 not possible to move Edit Integration Events dialog
KPR#:1380 wording mismatch between menu and dialog for saving sequence templates
KPR#:1524 Dropdown box for signal selection does not have a scrollbar
KPR#:1574 OpenLAB Control Panel does not prevent connecting to an instrument already in use
KPR#:1577 Slow performances observed while loading datafiles when a networked printer is not accessible
KPR#:1583 Additional information about the [ChemStation:Sequence][Save Sequence] privilege
KPR#:1584 ALS Sampling diagram shows vial usage when using HeadSpace
KPR#:1605 The UI for the CTC sampler does not display the current location
KPR#:1631 Users have restiricted menus, although their privileges should allow all options
KPR#:1644 PAL Sampling diagram always shows 'Vial Status: Normal' for every vial
KPR#:1674 Graphic artifacts/software crashes caused by graphic settings
KPR#:1682 7890A, 7820A missing sampling diagram when only a rear injector is installed
KPR#:1695 Change Access Level... is not available from menu bar
KPR#:1740 The logged-on user is not displayed in lock dialog
KPR#:1745 Finish Queue Sequence Help is hidden behind dialog
KPR#:1746 Sampling diagram for 68xx Classic driver appears, but if selected and reselected, disappears; also when switching betw Meth&Run and DA
KPR#:1764 Run-time error ‘282’: No foreign application responded to a DDE initiate
KPR#:1808 Values under Overview Table are not aligned with the parameter heading
KPR#:1855 No sampling diagram displayed with a 7890B on B.04.03 SP1 [87] or B.04.03 SP2 [105]
KPR#:1688 Preferences set in offline overwrite online Preferences and vice versa
KPR#:1717 Buttons for manual integration are not accessible, while they can be selected through the menu
KPR#:1763 In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.

Keyword: User Interface , Data Analysis

KPR#:1688 Preferences set in offline overwrite online Preferences and vice versa
KPR#:1717 Buttons for manual integration are not accessible, while they can be selected through the menu

Keyword: User Interface , SEQ , ACQ , ECM Integration/OpenLab Option

KPR#:1763 In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.

Keyword: Utilities

KPR#:77 Hyperseq continues running between multiple sequences for one plate after stop
KPR#:199 B.01.01 and PC Anywhere is giving conflict
KPR#:209 Excel workbook issue with "special links" and chemstation.xls reports
KPR#:246 wrong pressure description using Japanese fonts
KPR#:247 Would like to be able to view LC/MS fraction and spectra in same View
KPR#:264 XML Import: Import of LimsId fails if sample amount has been set in import XML f
KPR#:389 no annotation or move annotation in 3D overlay window possible
KPR#:477 OQ/PV test Injector Precision test show pass instead of failed
KPR#:484 Holmium test passed although hardware fails
KPR#:488 Cap. Pump Flow Sensor Calibration required pump switched on
KPR#:489 Setup window for modules grey out after performing OQPV tests
KPR#:541 not possible to open a Purify study when the Purify Data Base Path is set to C:\
KPR#:588 Verification: unit for RID noise values
KPR#:589 Extracted Chromatograph changes when setting Bandwitdh and exporting
KPR#:594 1100 Diagnosis: Draw action performs red message
KPR#:602 K Prime may not be calculated using void volume
KPR#:605 zoomed peaks may be printed incompletely on reports
KPR#:648 Annotation Menu disables some of the Graphic, Signal options
KPR#:673 Wrong compressibility setting in ChemStation when using Handheld Controler
KPR#:675 Use of external contacts board on 1367C injector no binary output possible
KPR#:678 Invalid fraction start location
KPR#:679 Print report to XLS file strange characters are present
KPR#:680 Purify: sample description only once by wizard
KPR#:681 Started sequence with wrong vial position after changing trays
KPR#:720 zooming reset when copying to clipboard
KPR#:737 No > 10 is not accepted as vial number for Diagnosis single step tests for AFC
KPR#:791 Excel Seqence summary macro is failing
KPR#:854 Prep-Pumps online plot signals need to set again after changing views
KPR#:884 cannot load batch method, if method (on another drive) is specified
KPR#:958 ChemStation crash during MS data AIA export to network drive
KPR#:1012 English Acrobat Reader install package on ChemStation CD 2 not working
KPR#:1134 Modified or corrupt base reference file IQTRef.xml causes crash of IQT Report
KPR#:1135 with missing base reference file IQTXref.xml the IQT is still reported as passed
KPR#:1233 Chinese/Japanese ChemStation: cannot import .csv file to sequence
KPR#:1263 IQT Report may fail after switching to Diagnosis with invalid files
KPR#:1419 compound custom fields are only exported to result.xml when the compound was found
KPR#:1772 FPD Checkout Macros no longer work since there is no "autoint.mac" with OL CDS ChemStation C.01.04
KPR#:1616 Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group

Keyword: Utilities , Spectra Analysis , MSD Analysis

KPR#:1616 Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group


KPR#:2  Product:Easy Sequence  Easy Sequence  A.01.01

Keyword: 

One-line Description:

LC-3D, CE, and MSD technique-specific columns missing

Problem:

The columns that are missing include:

LC-3D and MSD:  Target Masses (number, range, list) and AutoBalance (blank or ‘OFF’)

CE: Voltage (number), StopTime (number), and User1-10 (1-48 or blank)

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:3  Product:Easy SamplePrep  Easy SamplePrep (G7300)  2.01

Keyword: 

One-line Description:

6890: Add\Advanced Tower Selection Syring Volume does not update

Problem:

6890: Add\Advanced Tower Selection Syring Volume does not update until session is closed/reopened.

For any Add step, select Advanced. The volume determines which tower will be used if automatically select by volume is selected (default). The Syringe Volume of that injector is displayed read-only below it.
If you go to the Configuration and change the syringe size for any tower, then add a new Add step, save and return, this syringe Volume change is not detected or displayed. It is only refreshed by closing and reopening the online session of ChemStation.

Temporary Solution:

Close and reopen the session to update the syringe volume.

Fix Information:

n/a


KPR#:4  Product:Easy SamplePrep  Easy SamplePrep (G7300)  2.01

Keyword: 

One-line Description:

Calibration Blanks from Easy Sequence using an Easy SamplePrep program with 'Use for unknowns only' checked reserves resources

Problem:

No Injection Calibration Blanks from Easy Sequence using an Easy SamplePrep program with 'Use for unknowns only' checked reserves resources, even though the Easy SamplePrep program does not run in this case.

Temporary Solution:

Supply an excess of resources to allow the sequence to be submitted to the Queue.

Fix Information:

n/a


KPR#:5  Product:Easy Sequence  Easy Sequence  A.01.01

Keyword: 

One-line Description:

Sequence location directory selection is not correct

Problem:

When creating an Easy Sequence template and select the sequence location funcion, the software displays the normal windows function for selecting a directory / sub-directlry. This is wrong. The software should show the selection of sequence locations as specified in the sequence preferences.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:6  Product:Easy Sequence  Easy Sequence  A.01.01

Keyword: 

One-line Description:

ChemStation: Vials are validated against the currently loaded method

Problem:

In the Easy Sequence Setup, a method is selected and then the vial numbers are entered. These vial numbers are passed to the data system to be validated.  

The validation occurs against the currently loaded method, not the method used in the template that will actually use the vials.

Temporary Solution:

n/a

Fix Information:

Fixed in Easy Sequence 2.4[002].


KPR#:6  Product:Easy SamplePrep  Easy SamplePrep (G7300)  2.01

Keyword: 

One-line Description:

Calibration Blanks with injection run Sample Prep when 'Use for unknowns only' is checked

Problem:

Calibration Blanks with injection run Sample Prep when 'Use for unknowns only' is checked

Temporary Solution:

Use a different methods for blanks that has 'Use Sample Prep program' unchecked

Fix Information:

n/a


KPR#:7  Product:Easy Sequence  Easy Sequence  A.01.01

Keyword: 

One-line Description:

Custom Fields are missing in Extended Parameters dialog

Problem:

Sample Custom Fields and Compound Custom Fields are missing for Extended Parameters.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:9  Product:Easy Sequence  Easy Sequence  A.01.01

Keyword: 

One-line Description:

Bracketing with unique folders off

Problem:

When bracketing is used with Easy Sequence with Unique Folder Creation OFF, data file names can not be generated for the calibration runs. The potential exists to overwrite files when using Bracketing with Unique Folder Creation OFF with Easy Sequence.

Temporary Solution:

When using Bracketing with Easy Sequence, use Unique Folder Creation ON from Preferences.

Fix Information:

n/a


KPR#:10  Product:Easy SamplePrep  Easy SamplePrep (G7300)  2.10

Keyword: 

One-line Description:

6890 Easy SamplePrep Add step fails for both by use and by volume resources

Problem:

If there is an ‘Add’ step in Easy SamplePrep program with just the ‘Solvent Wash’; even though the Back injector is to be used, it may not use the Back injector.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:11  Product:Easy SamplePrep  Easy SamplePrep (G7300)  A.02.02

Keyword: 

One-line Description:

Resource layout on 7890A GC instrument dialog is not translated

Problem:

Resource layout on 7890A GC instrument dialog is not translated into Chinese and Japanese.

Temporary Solution:

n/a

Fix Information:

Fixed in revision A.02.03.


KPR#:12  Product:Easy SamplePrep  Easy SamplePrep (G7300)  A.02.02

Keyword: 

One-line Description:

Easy SamplePrep Method printout is not translated

Problem:

Easy SamplePrep Method printout is not translated.

Temporary Solution:

n/a

Fix Information:

Fixed in Revision A.02.03.


KPR#:12  Product:Easy Sequence  Easy Sequence  A.01.01

Keyword: 

One-line Description:

Manual/Valved/HS templates cannot be loaded if inj source does not match loaded method

Problem:

6890 with no ALS or tray. Configured to a HeadSpace.
Create a front valve method, a back manual method, and a HeadSpace (front) method.

Depending on the method loaded, load an .est and this fails with:
'---------------------------
Inconsistent Easy Sequence Setup
---------------------------
Sequence requires a back injector not present on the instrument.

---------------------------
OK   
---------------------------'

The message really does not advise the user of what should be done to correct this situation.

Temporary Solution:

This is current ChemStation behavior.  The template created in Easy Sequence must use the currently loaded method.

Fix Information:

n/a


KPR#:12  Product:LC and CE Drivers  RC 1200.NET  N/A

Keyword: 

One-line Description:

LC driver starts run with ALS Chiller turned off

Problem:

Using the RC.NET Drivers to control the Autosampler Chiller the systems is injecting with the Autosampler Chiller turned off. This is a different behavior compared to the legacy Rapid Control drivers for EZChrom which forced an ON of the Autosampler Chiller.

Temporary Solution:

n/a

Fix Information:

This issue has been fixed with driver release A.02.01 SP1.


KPR#:13  Product:Easy SamplePrep  Easy SamplePrep (G7300)  A.02.02

Keyword: 

One-line Description:

Resources replenished on 7890A GC instrument dialog not translated

Problem:

Resources replenished on 7890A GC instrument dialog not translated

Temporary Solution:

n/a

Fix Information:

Fixed in revision A.02.03.


KPR#:14  Product:Easy SamplePrep  Easy SamplePrep (G7300)  A.02.03

Keyword: 

One-line Description:

Certain combination of setpoints cause setpoint error from tower

Problem:

The software allows the user to enter invalid parameters for the draw and dispense speeds for the syringe.

Using an "Add" command it generates a "Command to XXX Tower failed: XXX Tower Prep Failure Code: 142: Autosampler setpoint error..."

The error message is limited and it takes effort to find the "invalid" parameter.

Temporary Solution:

n/a

Fix Information:

Fixed in A.03.01.


KPR#:14  Product:Easy Sequence  Easy Sequence  A.01.01

Keyword: 

One-line Description:

Easy Sequence Setup does not display dual sources on method load; must Clear All to activate

Problem:

Open online session to a 6890 with dual ALS (Plasticman 6890A/7693).
DEF_GC.M is configured by default to front injector.
Set injection source to Dual and save as a new method, now loaded.

Select Easy Sequence Setup. No injection sources are displayed, (presumably because of the injection source selection of the previously loaded method had one injector.)

Load the Dual method. Still no injection sources.

Temporary Solution:

The user must select 'Clear All' to refresh the screen in order to display the Front/Back injection source checkboxes after loading the dual injector method.

Fix Information:

n/a


KPR#:15  Product:Easy SamplePrep  Easy SamplePrep (G7300)  2.01

Keyword: 

One-line Description:

Need ability to 'shake' with Mix command

Problem:

The Mix command needs some more detailed control. The mixer to is used to shake a vial (Mix 2500 rpm 1 sec, stop 1 sec , and change direction mix 2500 1 sec, repeat), this cannot be done with the current mix command.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 2.3.


KPR#:16  Product:Easy SamplePrep  Easy SamplePrep (G7300)  2.01

Keyword: 

One-line Description:

Wash/waste vials do not cycle; only A2 and B2 wash vials are used

Problem:

With a multi-line sample prep method, the wash and waste vials do not cycle through.  Only A2 and B2 wash vials are used.  Configuration was set to use A1-A6 and B1-B4.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 2.01.


KPR#:17  Product:Easy SamplePrep  Easy SamplePrep (G7300)  2.01

Keyword: 

One-line Description:

Cannot mix/heat resources in the resource layout

Problem:

It is not possible to heat or mix any resource in the resource layout.

Functionality is missing in the mix and heat command where only the sample (front or back) or newly made sample in an empty vial can be selected.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:17  Product:Easy Sequence  Easy Sequence  A.01.01

Keyword: 

One-line Description:

Easy Sequence Setup does not validate HeadSpace vial ranges

Problem:

HeadSpace Method loaded in ChemStation.
The Injection Source is set to HeadSpace Sampler

Easy Sequence Setup
Sample Information
Starting Vial Location

Any number entered into this location will error with the following error message:
The vial position could not be validated against the current configuration

Temporary Solution:

n/a

Fix Information:

Fixed in Easy Sequence 2.4[012].


KPR#:18  Product:Easy SamplePrep  Easy SamplePrep (G7300)  A.02.02

Keyword: 

One-line Description:

Add step fails to execute due to mismatch in # injections/volume/syringe size/solvent allowed by firmware

Problem:

Method with a single add step of 1000 uL. Syringe size is 10 uL. Vials are moved to turret and then error message displays: method not running.

This is a mismatch in # injections/volume/syringe size/solvent allowed by firmware.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:19  Product:Easy SamplePrep  Easy SamplePrep (G7300)  A.02.03

Keyword: 

One-line Description:

A method using Easy Sample Prep cannot be used to create a method not using Easy Sample Prep

Problem:

Create a method which has the "Use Sample Prep Program" checkbox enabled. Use this method to create a method which does not have the "Use Sample Prep Program" checkbox enabled. Load the method which does not have the checkbox enabled. Error message displays: Configuration mismatch: Default parameters loaded for 'Agilent 7890'.

Temporary Solution:

n/a

Fix Information:

Fixed in 3.1[21]. This revision will only fix with Openlab Chemstation C.01.03.


KPR#:24  Product:Easy Sequence  Easy Sequence  A.01.01

Keyword: 

One-line Description:

HeadSpace Injection Volume is accepting negative numbers; no checking; no meaning

Problem:

Create an Easy Sequence Template for a G1888/6890. There is no error checking for Injection volume for this injection source in either the Setup or Sample List. All values are accepted, including negative numbers.
There is no place to setup an 'injection volume' for a HeadSpace. There is only a time and pressure.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:25  Product:Easy Sequence  Easy Sequence  A.01.01

Keyword: 

One-line Description:

Easy Seq and Setup Starting Vial Location and Number of Samples do not check for limits

Problem:

Starting Vial Location and Number of Samples do not do any range checking against any method or configuration.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 2.3


KPR#:26  Product:Easy Sequence  Easy Sequence  A.01.01

Keyword: 

One-line Description:

Easy Seq Setup: Sample Name text entry should check for a Counter if used as Data File name

Problem:

Easy Sequence Setup: Entry of text in the Sample Name combo box does not check for and advise that a counter token is needed if the user also uses the token 'Sample Name' for the Data File Name.

The user will find out when sending to the queue 'Error: Sample Name xxxx occurs more than once.'

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:27  Product:Easy SamplePrep  Easy SamplePrep (G7300)  A.03.01

Keyword: 

One-line Description:

Sequence containing 4 or more samples aborts the run on the last sample

Problem:

Using ChemStation B.04.03 SP1, Easy Sample Prep A.03.01, and a 6890 GC instrument: 

Create a sequence that runs four or more samples. The first three samples complete successfully, but the last sample aborts as an undefined error.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:28  Product:LC and CE Drivers  RC 1200.NET  A.02.03

Keyword: 

One-line Description:

Lamps and cells of G1314F show incorrect dates in the RFID tags

Problem:

Lamps and cells of G1314F show incorrect dates in the RFID tags

Temporary Solution:

n/a

Fix Information:

Fixed with the driver version A.02.06


KPR#:29  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Easy Sequence Setup not refreshing the Injection Location

Problem:

7890/7693: Create methods for front, back, and dual ALS injection sources.
Easy Seqeunce Setup does not refresh to show the correct injection location of the loaded method until user loads the corresponding method in ChemStation.

Even loading a different method, the previous injection source is kept.

Temporary Solution:

If the injection source has been changed, use the "Clear All" button and reload the method.

Fix Information:

n/a


KPR#:31  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

No error checking for vial ranges in Easy Sequence Setup; fails at runtime

Problem:

There is no error checking within Easy Sequence Setup to determine if the vial range of the loaded method's injection source is accurate.

The method can be saved and submitted to the queue.  It fails at this time with:

'---------------------------
Error
---------------------------
Invalid location found in FRONT sequence, line 1
There may be others
---------------------------
OK   
---------------------------'

This defect occurs because there is a difference between the loaded method's injection source and the method that is selected in the Easy Sequence Template.

Temporary Solution:

Load the method in ChemStation from Method > Load Method menu that will be used to create the Easy Sequence Setup Template.

Fix Information:

n/a


KPR#:32  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Pause/Resume button is in the wrong state after a sequence is edited and then pauses

Problem:

Send a sequence to the Queue and while it is running, edit it. This modified/changed sequence pauses the Queue, but the first blue button is still in a 'Pause||' state. Click the button and change it to Resume, but it does not do anything as it is waiting for the current sequence to be saved before starting the next sequence.

In this case, because the sequence has been modified, the ChemStation is not accepting sequences. If you click pause || button, you pause the queue and it does not do any thing. The message on the active queue is red and the message says the sequence has been modified.

Temporary Solution:

Save the sequence that has been modified.  Once the sequence has been saved, the queue will resume running the next sequence.

Fix Information:

n/a


KPR#:33  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

7890A with no ALS shows Read Barcode button in Easy Sequence Sample List

Problem:

Load an Easy Sequence.es for a valve or HeadSpace Injection Source and the Sample Name cell in the Sample List shows a Read barcode button.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:34  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

HeadSpace Simple Calibration Blank type should only allow 'Blank with injection'

Problem:

HeadSpace software will not run a sequence without a vial location.  Therefore, the Easy Sequence Setup\Calibration\Simple Calibration options should only allow Blanks with Injection and not allow blanks without injection.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:35  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Calibration Blank with no injection loads with Inj Vol as 0

Problem:

Create a setup with a blank with injection and a blank with no injection.
Save and load to Easy Sequence. The Injection volume for the Blank with no injection is entered as 0. Native sequence table also shows this as zero.
This is not consistent with the volume for the blank with injection

Temporary Solution:

n/a

Fix Information:

Fixed in revision 2.2


KPR#:39  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Calibration table clears on method reload from Samples tab

Problem:

Create a Setup with Calibration. Return to Sample tab and load the same or a different method with the same injection location. Return to Calibration tab and all has been cleared. Reloading a method does not clear entries made on the Samples tab.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:40  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Queue displays running vials incorrectly when initial HeadSpace vial not found

Problem:

Start a 3 line headspace sequence from the queue, but remove the FIRST vial in the sequence. Queue does not update correctly, showing the first vial as running, though the HeadSpace has moved on to vial 2.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:41  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Aborted HeadSpace run displays Status as 'Complete' in Queue History

Problem:

HeadSpace aborts, but ChemStation does not. Select the Abort Run button from the Waiting for ready dialog and ChemStation aborts the sequence.
Queue History shows this run as 'Complete' instead of 'Aborted'.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:44  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Easy Sequence does not advise of invalid vial locations until sequence is fully loaded into Queue

Problem:

There is no advisory about the invalid vial locations until the sequence has been added to the queue.

Temporary Solution:

n/a

Fix Information:

Fixed in revision 2.3


KPR#:45  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Sequence submitted to queue but will not run and can not be removed/deleted

Problem:

Sequence submitted to the queue but will not start and cannot be deleted.

Temporary Solution:

As a workaround, delete the following file:  SequenceQueue.que from the C:\Chem32\4\TEMP directory.

Fix Information:

Fixed in revision 2.4.


KPR#:47  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Injection Volume on ES Setup should show units (uL); valved inj are in mL and user must convert

Problem:

The Easy Sequence Setup dialog should show Injection volume units, which in ChemStation always seem to be uL. 

When the user uses EasySequence with a Valved injection, the volume, which for valves is mL, must be multiplied by 1000 to get a value that will be accepted.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:48  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Easy Sequence Setup should check for Sample Name length and not allow invalid entries

Problem:

Enter a long Sample Name in Easy Sequence Setup and add a counter.
Save and load to Easy Sequence. Fill Samples. Not until then is the user warned (with a red X) in the Sample list that there is a 40 character limit.
It is possible to save this but it will not be possible to add to the Queue.

Temporary Solution:

n/a

Fix Information:

Fixed in Easy Sequence 2.4[014].


KPR#:52  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Easy sequence actions are not included in the audit trail and no separate audit log exists

Problem:

Any changes done on Easy Sequence are not included in the audit trail; there is no separate audit log present for Easy Sequence.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:53  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

Vial not found reported in software Status not consistent with instrument status

Problem:

Run an Easy Sequence using a HeadSpace injection source. When a Vial not Found event happens, the final Status for this run, which did not start a run, is reported as 'Completed'. The status being reported by the HS device is 'Run Aborted'.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:54  Product:Easy Sequence  Easy Sequence  2.01

Keyword: 

One-line Description:

PAL autosampler plate/vial locations not accepted in Setup location box

Problem:

Easy Sequence 2.01.41 on Chemstation B.04.03 DSP1 will not accept plate designations of PAL injection source in the Setup location combo box. (P1A1)
The user should be able to choose the location designation.

Temporary Solution:

n/a

Fix Information:

Resolved in revision 2.4 [12].


KPR#:60  Product:Easy Sequence  Easy Sequence  2.4

Keyword: 

One-line Description:

Print preview screen does not appear when blank is added to calibration program

Problem:

1. Method and Run control screen> easy sequence settings tab > calibration tab
2. Select "simple calibration" for calibration mode, add Blank to the sequence. click print button to print preview easy sequence settings. 

Print preview screen does not appear.

Temporary Solution:

n/a

Fix Information:

Fixed in Easy Sequence 2.4[30].


KPR#:352  Product:78xx Driver Software  7890B  B.01.01

Keyword: 

One-line Description:

Cannot Add User-Defined EMFs on Intuvo 9000 GC

Problem:

Unable to created a user defined EMF Intuvo 9000 GC. Enter a name and tracking type, but nothing appears on the User-defined panel or any other panel.  There are five user defined (seemingly blank) entries on that panel.

Temporary Solution:

none

Fix Information:

To Be fixed in B.01.02


KPR#:356  Product:78xx Driver Software  7890B  B.01.01

Keyword: 

One-line Description:

Method print is showing Solvent wash mode as A-A2, B-B2 when it should be A,B since it is a 16 vial turret

Problem:

Intuvo 9000 with a ALS and a 16 vial turret.  MassHunter software, the configuration of the ALS is set to Solvent Wash Mode A,B.  Saved and printed the method.  Solvent Was Mode on the method print shows A-A2, B-B2.

Temporary Solution:

none

Fix Information:

to be resolved in B.01.02


KPR#:1723  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: 

One-line Description:

Viewing the Signal Options modifies the method.

Problem:

Viewing the Signal Options is possible from either the Graphics menu or the specify report dialog. Changing any of the options will mark the method as modified, even if the dialog has been closed by pressing the 'Cancel' button.
Even though nothing has been changed the user sees this an an unexplained modification.

Temporary Solution:

Do not save the method after leaving the signal options, when no changes were intended.

Fix Information:

Fixed in revision B.04.03 SP1.


KPR#:1726  Product:ChemStation  ChemStation (G2070BA)  B.04.03 SP1

Keyword: 

One-line Description:

Wrong Flow setting displayed when re-opening the Method User Interface

Problem:

Setting the Flow on the G1376A using the Method User Interface.
After pressing "OK" the flow is applied but opening again the Method User Interface the Flow=0 is displayed.

Temporary Solution:

none

Fix Information:

fixed in Driver revision A.02.04 SP1


KPR#:1731  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: 

One-line Description:

LC/MS spray chamber temperatures are no longer be applied to the spray chamber heated zones following a spray chamber open/close action.

Problem:

After an LC/MS spray chamber has been opened and closed, temperature will no longer be applied to the Spray Chamber heated zones until the spray chamber parameters are  updated (by clicking "ok" in the Spray Chamber Parameters dialog box or loading a method) or the system is turned "off" or to "standby" and then back to on.  This occurs with the APCI, APPI and Agilent Jet Stream spray chambers.  The MultiMode source will set the system to "Standby" if opened & closed, however.

Temporary Solution:

Turn the instrument to "off" or to "standby" in ChemStation, then back to "on" if this condition is encountered.

Fix Information:

n/a


KPR#:1781  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: 

One-line Description:

Intelligent Report Templates generated with ChemStation C.01.03 may require update

Problem:

Intelligent Reporter templates designed for multi-signal methods may need to be adjusted after updating to ChemStation C.01.04. 
Non-calibrated peaks of signals calibrated in a main signal do no longer appear as calibrated compounds (as they are not quantified on the 2nd signal) but as un-identified peaks.

Note: Calibration tables setup to "calculate signals separatly" are not affected by this design change.

Temporary Solution:

In order to display the results of the mentioned peaks, the peak filter of the report template should be turned off completely (remove all check marks in the "show peak types" screen).

Fix Information:

In rev. C.01.05. the mentioned peaks are treated as identified, but not quantified peaks.


KPR#:1782  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: 

One-line Description:

UNC paths are not supported as report file storage loaction

Problem:

The "copy to" function of the intelligent report method setup does not support usage of UNC paths.

Temporary Solution:

Map the network path as a drive letter and select the new drive letter as a report destination.

Fix Information:

There is no fix planned.


KPR#:1943  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.01 SP1

Keyword: 

One-line Description:

78xx Stop button pressed no longer continues to the next run in the sequence

Problem:

Pushing the stop key at the GC keyboard will stop the current run and generate a report but does not continue to the next line in the sequence.  

This occurs beginning on ChemStation version B.04.01 SP1.

In revisions B.03.01 - B.04.01, pushing the stop key at the GC Keyboard would stop the current run and generate a report and continue to the next line in the sequence.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB CDS C.01.06.

Setup conditional System Behaviors for the 78xx in Manage Rules & Alerts found under the Instrument menu
-Continue
-Abort


KPR#:1971  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.05 HF01

Keyword: 

One-line Description:

The Active Splitter does not always stop at the end of a sequence when "Run Sync" is enabled in the Active Splitter Software

Problem:

The Active Splitter is not always stopping at the end of a sequence when Run Sync is enabled .

With "Run Sync" enabled, the operation of the Active Splitter should be controlled through ChemStation/Active Splitter software.  Intermittently, the software control of the Active Splitter does not stop its operation at the conclusion of a sequence/run, even with "Run Sync" enabled.

Temporary Solution:

An external Contact Closure card, G1351A, can be used to Start/Stop the Active Splitter independently of the Active Splitter software.

Fix Information:

n/a


KPR#:1153  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: 1046 Control

One-line Description:

1046 FLD cannot be set up from the GUI

Problem:

1046 FLD cannot be set up from the GUI. If you click the corresponding item in the System Diagram, the following red error message will be displayed:
view.floatingmenu "menu[1]."FLD1046" failed. Menu table contains incorrect menu rows.

Temporary Solution:

Use the corresponding entry in the Instrument menu to configure the 1046 FLD.

Fix Information:

No fix is currently intended.


KPR#:38  Product:ChemStation  ChemStation (G2070BA)  A.09.03

Keyword: 1100 Control

One-line Description:

Signal change in LC Detector during a run

Problem:

Running LC Sequences in operator mode it is possible to change signal parameter for various LC detectors. Multiple changes during acquisition are possible and the change will be documented on the Report. A change history list is missing.

Temporary Solution:

The change will be documented by the Chemstation using the method changed flag/ time stamp and by the sequence logbook (method changed).

Fix Information:

n/a


KPR#:43  Product:ChemStation  ChemStation (G2070BA)  A.10.02

Keyword: 1100 Control

One-line Description:

two MWD in one instrument: problem to control MWD 2 via GUI

Problem:

For ChemStaiton A.10.x revsion following issue has been reported for systems using a stack including 2 times a MWD:

The graphical user interface control for the secound MWD in the configuration is controlling the settings of the first MWD of the configuration. This issue only applies fo the graphical user interface. The Instrument menu allows to control the VWD in the correct way.

Temporary Solution:

Use the instrument menu to control two MWDs in one 1100 stack.

Fix Information:

 No fix Planned


KPR#:64  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: 1100 Control

One-line Description:

'Fast Reconditioning' enabled in Normal Mode of capillary pump

Problem:

For a Micro Wellplate Sampler, the "Fast Reconditioning" option should not be available in normal mode. Under the following conditions, "Fast Reconditioning" gets nevertheless enabled.

In the method "Injection Overlapping" is not enabled for the Micro Wellplate Sampler. If "Fast Reconditioning" and "Solvent Custom Calibration" get enabled in micro mode and afterwards the WPS is switched to normal mode, the "Fast Recondition" option is accessible in normal mode.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:89  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: 1100 Control

One-line Description:

with Column flow of 0.00 ul/min no data acquisition

Problem:

If a flow of 0.00 ul/min is applied to run a sequence on a system with capillary pump with stop time of e.g. 1 minute, the acquisition of detector signals stops correctly after 1 minute, but data acquisition will nevertheless only complete after 3 hours.

Temporary Solution:

Use flows larger than 0.00 ul/min.

Fix Information:

 No fix Planned


KPR#:135  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: 1100 Control

One-line Description:

EMF indicators do not work

Problem:

Using the EMF functionality of the Rev. B.01.01 ChemStation the EMF indicators for the needle movements on DLA is not functioning.
 
The Needle movements have exceeded limits and the limit fields are red in Diagnosis View. The EMF icon is green in Method and Run Control.
 

Temporary Solution:

n/a

Fix Information:

There is no fix planned for this problem.


KPR#:179  Product:ChemStation  ChemStation (G2070BA)  A.10.02

Keyword: 1100 Control

One-line Description:

Using flow gradients during Fraction Collection results in Wrong times in Report

Problem:

Following issue has been reported, using ChemStation A.10.0x and higher: 

Using flow gradients during Fraction Collection results in wrong times for the collected fractions in the Report (in the fraction table). 
However, tickmarks are displayed and generated correctly, so there is a mismatch betqween tickmarks and the reported fraction table times.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:213  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: 1100 Control

One-line Description:

DAD Diagnosis: lamp indicate green, even when lamps off

Problem:

In order to run DAD tests (e.g. DAD Wavelenght test) the DAD has to be switched on. However, the icon for the DAD always is green and ready for test, even when the DAD is switched off.

Temporary Solution:

Check the status of the DAD at the instrument itself.

Fix Information:

There is no fix planned for this problem.


KPR#:287  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: 1100 Control

One-line Description:

Pressure and Leak test in prep. Pump will not finish, unless Press. released

Problem:

The pressure and Leak test in G1361A prep. Pump will not finish, unless the Press. is released manually by the user, by opening a capillary.

Background: The EMPV valve is not allowed to switch, unless there is low pressure (safety feature). 

Temporary Solution:

Please manually release the pressure at the end of the test by opening a capillary.

Fix Information:

No fix is planned for this issue.


KPR#:290  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: 1100 Control

One-line Description:

Flow composition not updated in GUI when loading new method

Problem:

Following issue can be seen in ChemStation B.01.03 using methods with flow composition.

Modifying and saving flow composition values are not correctly updated in the "Instrument Actuals" of the corresponding pump, when a new method or default method with different settings is loaded.

Temporary Solution:

Either close and reopen the instrument actuals

or

use the "LC Parameter" view in the method and run control to monitor these values.

Fix Information:

n/a


KPR#:407  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: 1100 Control

One-line Description:

G1321A FLD: cuvette is not correctly recognized

Problem:

Using the G1321A FLD detector following problem has been reported using ChemStation B.01.01x:

Changing from a standard flow cell to a cuvette
(e.g. to perform OQ), the system will respond on the report with a "micro flow cell".

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:416  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: 1100 Control

One-line Description:

Starlight and FIA: double start pulsing of G1315C

Problem:

Using B.01.03 LC/MS system with the G13315C Detector, methods using FIA will see following behaviour:

Running x injections in one FIA run, the system will send y start pulses, so the the report might describe
additional runs.

For the each injection, two start pulses are sended.
Problem will be resolved with the next G1315C firmware revision.

Temporary Solution:

The issue will be fixed in the next firmware revision.

As a current workaround a macor solution can be used, please contact your Agilent Representative.

Fix Information:

 No fix Planned


KPR#:451  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: 1100 Control

One-line Description:

G1323B colum setting calles negative value in ChemStation column setting

Problem:

Using the control module G1323B to define the column compensition in the timetable and defined "off" for the temperature. the system will show -274 when opening  the colum comp. parameter using the ChemStation.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:454  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: 1100 Control

One-line Description:

DAD Calibration history information logged in chemstation.ini

Problem:

Using ChemStation please note that the  DAD Calibration history is read out of the 
ldadhistory information in the PCS section of the ChemStation. Attaching a different DAD detector to an existing ChemStation will result in a loss of the DAD calibration history.

Temporary Solution:

n/a

Fix Information:

No fix is planned for this issue.


KPR#:457  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: 1100 Control

One-line Description:

Using same detector type twice no selection can be set diagnostic

Problem:

Using ChemStation Rev. B follwing issue has been seen using the same detector type multiple times:
In diagostic, the system will list only one detectora for one type.

Temporary Solution:

In oder to perfom test, please run the test with only one detector of the same type configured at time.

Fix Information:

n/a


KPR#:471  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: 1100 Control

One-line Description:

Sequence table not accepting options of Next Location/Plate

Problem:

Following issue has been seen running ChemStation B.02.01 using Fraction Collection:
Defining a sequence and specify in the sequence table column Frac. start column the otion of next plate or next location for the first line, the system will give a warning when starting the sequence that that position is not available.

Temporary Solution:

The first sequence line need to have a specific Fraction Start location placed or it has to be left blank, so the system is taking the next free location
known by the Fraction collector isself.

Fix Information:

No fix is planned for this problem.


KPR#:472  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: 1100 Control

One-line Description:

G1314C VWD Detector diagnostic test results issue

Problem:

Using ChemStation B.02.01 performing verious diagnostis tests using the G1314C VWD detector, the test results are not present in the results file.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:474  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: 1100 Control

One-line Description:

Fast Comp. Change / Reconditioning disabled after initialization

Problem:

Following issue has been seen using ChemStation B.02.10 using fast compensition change/reconditiong option:
Rebooting the system and entering the part Fast Composition Change / Reconditioning the first time, the dialog is disabled though the pump is running in micro mode.

Temporary Solution:

Please enter and edit/cancel any other dialog (e.g. setup or auxiliary) first, then the options will be available.

Fix Information:

n/a


KPR#:475  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: 1100 Control

One-line Description:

"next plate" option using Vial Trays in Fraction Collection mode

Problem:

Following issue is seen on ChemStation B.02.01 running Fraction Collectors with Vial plates configured: 
Defining the position 1 and 101 for fraction collection on a Vial plate, and define the option "next plate" in the Frac. Start column of a sequence the field is ignored without error message.

Temporary Solution:

The Next Plate function is a valid function for Well Plates, not for vial trays. Please use the option "next location" in the Frac. Start column in the sequence table.

Fix Information:

No fix planned


KPR#:476  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: 1100 Control

One-line Description:

Customized methods for Noise, Wander, Drift issue

Problem:

Following issue has been seen running ChemStation B.02.01 and execute the OQ/PV Noise and Temperature Stability using a customized method:
Decreasing the stoptime to e.g. to 1 min. the ASTM Noise, Wander and Drift getting not calculated, but the system reports that the instrument verification test passes. 

Temporary Solution:

The test summary will show the test Noise, Wander  and Drift as not passed. 

Fix Information:

No fix is planned for this problem.


KPR#:487  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: 1100 Control

One-line Description:

Aborting System Test on LC systems might perform error state

Problem:

Following issue has been reported using ChemStaiton B.02.01 performing some HPLC System Test. If running system tests, e.g. the HPLC Chip System Test of a Chip Cube system is aborted while running, an error is displayed in the command line: "Fatal System Error: Check system resoureces..."

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:561  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: 1100 Control

One-line Description:

micro pressure test  failed randomly

Problem:

In some cases the micro pressure test failed whith a leak rate over the upper 
limit. This result was strongly affected by the volume in the flow path and 
the evaluation of the test, which was not done after the pressurization of 
the solvent.

Temporary Solution:

The micro pressure test was optimized with a longer pressurization time and 
the evaluation of the test is done at 8 minutes instead of 4.5 minutes. In 
addition the test runs with water on channel A1 instead of IPA on channel A2. 
These are the only changes done for the micro pressure test. The procedure to 
run the test, the evaluation and the limits are identical to the previous one.

Fix Information:

The issue is fixed with ChemStation B.02.01


KPR#:715  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: 1100 Control

One-line Description:

injection number in column tag not update for non-switching valve TCC

Problem:

Following behaviour has been reported for ChemStation using Thermostatted Column compartment (TCC) and the column ID Tag. The number of Injections, present in the column ID Tag is not updated in the following particular case:

TCC with column switching valve:

In the TCC set up, the user can determin which colum is going to be used, colum 1 - left hand,
colum 2 - right hand.
If you inject and you use column 1(left) the counter is updated for injection, if the Tag is present, because only the tag for the corresponding site is written to.
Placing the tag on the right site, but injecting on the left site, the counter does not increase,
because the injection update information is going to be sent only to the left site.


TCC without column switching valve:

The system is only increasing the number for the left site, not for the right site. For a TCC without a switching valve it should not matter where to place the column ID tag.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:787  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: 1100 Control

One-line Description:

ECD detector scaling in online plot 10x larger than in the data file

Problem:

Following problem has been reported: Running an ECD detector  on LC ChemStation, the online plot scaling is x10 larger than the scale in the data file.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1189  Product:ChemStation  ChemStation (G2070BA)  (None)

Keyword: 1100 Control

One-line Description:

G1362A RID Calibration Value cannot be entered

Problem:

During the calibration process for the RID, the flow cell is purged. Afterwards the calibration sample (Succrose) is injected into the flow cell. Then the calibration is started and the measured value is displayed. If not correct, the correct value of 521000 has to be entered.
In ChemStation B.04.01 this value will not accepted. Instead a message comes up specifying a wrong limit of values to be entered.

Temporary Solution:

Enter the calibration value via command line with command "NRIU 512000", or send the command via Instant Pilot.

Fix Information:

n/a


KPR#:513  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: 1200 Control

One-line Description:

Autobalance not accesible after stopping a sequence with AutoBalance off

Problem:

If you stop a sequence while a run is processed where "Off" is selected for the AutoBalance column, the Autobalance settings will remain inaccessible for the detector.

Temporary Solution:

Either restart the ChemStation or complete another run on the ChemStation.

Fix Information:

n/a


KPR#:670  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: 1200 Control

One-line Description:

Method Overview: wrong value shown for TCC temperature for "as detector cell"

Problem:

The Method Overview box shows wrong values for the TCC temperature if the column thermostat is set to the value "as Detector cell". 
In this case, the Method Overview does not show the real temperature, but instead the inactive value from the most upper field of the Column Thermostat setup dialog.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:775  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: 1200 Control

One-line Description:

impossible to enter a maximum pressure of 600 bar into column information

Problem:

Select Instrument -> Columns.
In the Edit Column Identification Modules dialog you can enter the max. pressure in bar. Only values up to 400 bar are excepted, even if the 1200 LC hardware allows a maximum pressure of 600 mbar.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:781  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: 1200 Control

One-line Description:

Fraction Collector settings of the Peak Detectors table are not saved

Problem:

The following problem is observed with regard to Fraction Collector setup when the Working Mode is set to "Off":
All values for "Up Slope", "Down Slope", and "Threshold" are not saved upon closing the setup dialog.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1064  Product:ChemStation  ChemStation (G2070BA)  B.03.02 SR1

Keyword: 1200 Control

One-line Description:

Sample scan not acquired in case of spectral range mismatch for VWD G1314D/E

Problem:

If a blank scan and a sample scan are set up for the VWD G1314D/E are set up in the VWD's timetable, there will be no warning if there is a spectral range mismatch between the blank scan and the sample run.
When running the method, the blank scan will be recorded, but not the sample scan.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1065  Product:ChemStation  ChemStation (G2070BA)  B.03.02 SR1

Keyword: 1200 Control

One-line Description:

attenuation of 4000 cannot be specified for the Analog Output of VWD G1314D/E

Problem:

It is impossible to specify an attenuation of 4000 for the Analog Output of the VWD G1314D/E. In the corresponding dialog, the value 4000 is accepted, but it will not be kept.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1103  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: 1200 Control

One-line Description:

G1365C/D Detectors not reporting wavelength scan summary

Problem:

For the Multiple Wavelength Detectors G1365C and G1365D, the Wavelength Summary Report is not being shown in the in the Final ChemStation Compliance Report. No Spectrum or Results tables are shown.

Temporary Solution:

n/a

Fix Information:

It is currently not planned to fix this problem.


KPR#:1181  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR2

Keyword: 1200 Control

One-line Description:

when solvent recycling for the RID is used the bottle filling still changes

Problem:

With an RID in the 1100/1200 system, solvent recycling can be switched. In this case, nevertheless the bottle filling of the pump decreases. If the options to prevent the next analysis or to shut down the pump are enabled, the analysis will stop although the solvent is recycled.

Temporary Solution:

n/a

Fix Information:

There is currently no fix planned for this problem.


KPR#:1213  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: 1200 Control

One-line Description:

Column Compartment G1316A/G1316B not available if it is part of cluster

Problem:

If G1316A or G1316B column compartment is part of a cluster with G1316C column compartments, it is not available in Diagnosis view.

Temporary Solution:

Perform diagnosis on the column compartment while it is not part of a cluster.

Fix Information:

n/a


KPR#:1227  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: 1200 Control

One-line Description:

CTC: Injection volume not displayed on sampler GUI

Problem:

In the sampler GUI for the CTC the actual injection volume is not correctly displayed. Instead, always an injection volume of 10 ul will be indicated.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1228  Product:ChemStation  ChemStation (G2070BA)  B.04.01 SP1

Keyword: 1200 Control

One-line Description:

CTC: UI does not display correct plate

Problem:

For the CTC sampler, the GUI does not display the correct plate/tray after changing the plate.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1365  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: 1200 Control

One-line Description:

Solvents not available for selection in the pump timetable for a Preparative Pump

Problem:

For a preparative pump, the solvent cannot be selected in the Set Up Gradient dialog.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:1389  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: 1200 Control

One-line Description:

signal description for the RID differs between signals acquired with classic and RC.NET drivers

Problem:

With classic drivers, the signal description for the RID is "RID1 A, Refractive Index Signal".
With RC.NET drivers, the signal description for the RID is "RI1 A, Refractive Index Signal".
This difference in the signal description will prevent the peak identification for calibrated methods that have been migrated from classic drivers to RC.NET drivers.

Temporary Solution:

n/a

Fix Information:

This problem will be fixed in revision B.04.02 DSP2.


KPR#:1399  Product:ChemStation  ChemStation (G2070BA)  B.04.02 DSP 2 (FireAnt)

Keyword: 1200 Control

One-line Description:

"Waste" or "Coulmn bypass" runs are causing injection counters to be incremented

Problem:

The following defct affects all G1316C TCC clusters:
Injection counters for non tagged columns are incremented for  all waste-, bypass-, and storage methods. It affects all columns where the columns are marked as "installed" in the ChemStation column table.

As a result, the injection counter reports larger numbers compared to the injection performed on a certain column.

Temporary Solution:

n/a

Fix Information:

This defect planned to be fixed in ChemStation B.04.02 SP1.


KPR#:1411  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.01 SP1

Keyword: 1200 Control

One-line Description:

the valve does not switch to by-pass position for the last injection when using overlapped injection

Problem:

When using overlapped injection with an ALS, the valve does not switch to the by-pass position on the last injection of the sequence table.

Temporary Solution:

Add a blank run with the same method at the end of the sequence to ensure that the valve switches to the by-pass position for the last sample run.

Fix Information:

n/a


KPR#:1418  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: 1200 Control

One-line Description:

run cannot be stopped/aborted when the injector program is longer than the run time

Problem:

In case an injector program of a method is longer than the run time, a run with a method cannot be stopped or aborted during the time needed for the injector porgram.

Temporary Solution:

The run/sequence can be stopped or aborted after completion of the injector program.

Fix Information:

n/a


KPR#:1421  Product:ChemStation  ChemStation (G2070BA)  B.04.02 Aladdin

Keyword: 1200 Control

One-line Description:

LC RC.NET drivers: progress bar not updated during post-conditioning

Problem:

When using the RC.NEt drivers for LC ChemStation, the progress bar not updated during post-conditioning. The progress bar is only updated when post-Conditioning has completed.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1429  Product:Method Scouting Wizard  Method Scouting Wizard  A.02.01

Keyword: 1200 Control

One-line Description:

No flow adjustment in case of temperature screening only campaigns

Problem:

Method Scouting Wizard A.02.01 introduced the possibility to get the solvent flow automatically corrected for different column diameters.
Keeping the solvent velocity constant results in a comparable back pressure for a given solvent composition.
However, for isocratic base methods and screening campaigns with no gradient screening selected, the flow correction does not apply.
This may cause in a high back pressure on columns with a smaller diameter which may cause a damage of the column.

Temporary Solution:

In case of screening across different column diameters, it is recommend to have the gradient screening option always turned on and to define a single gradient only if no gradient screening is desired.
In case of iso-cratic base methods, the gradient screening should be done with a constant solvent composition over time.

Fix Information:

Not reproducible with version A.02.02 and higher.


KPR#:1516  Product:ChemStation  ChemStation (G2070BA)  B.04.02 SP1

Keyword: 1200 Control

One-line Description:

Modul of 1260 family in classic driver may be misinterpreted as member of 1100 family

Problem:

Some modules of 1260 family may be misinterpreted by classic drivers in ChemStation as members of 11oo family. The modules of this older family did lack a few recent features like e.g. illumination. This problem originates from a sideeffect caused by a new schema for serial numbers for the 1260 family.

Temporary Solution:

Where combination of LC modules permits, use RC.Net driver suite instead of classic drivers.

Fix Information:

Will be fixed in B.04.03


KPR#:1521  Product:ChemStation  ChemStation (G2070BA)  B.04.02 SP1

Keyword: 1200 Control

One-line Description:

incorrect firmware revision displayed for the ELSD

Problem:

In the Revisions and Serial Numbers dialog, the firmware revision for the ELSD is displayed incorrectly, e.g. "14" instead of "V1.4".

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1526  Product:ChemStation  ChemStation (G2070BA)  B.04.02 DSP 3

Keyword: 1200 Control

One-line Description:

column switching valve: positions "Back-flushing on" and "Back-flushing off" are not kept

Problem:

For a TCC column switching valve (classic drivers), the positions "Back-flushing on" and "Back-flushing off" are not kept when leaving the setup dialog of the TCC with OK. The positions switches back to "Column 1". A similar problem occurs when selecting "Back-flushing on" and "Back-flushing off" in the timetable of the setup dialog.

Temporary Solution:

n/a

Fix Information:

This problem will be fixed in revision B.04.03.


KPR#:1539  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02

Keyword: 1200 Control

One-line Description:

Wrong firmware revision displayed in Instrument Configuration dialog after firmware upgrade

Problem:

After a firmware upgrade for LC modules using RC.NET drivers, the old firmware revision (prior to upgrade) is still displayed in the Instrument Configuration dialog.

Temporary Solution:

Remove the affected modules from the configuration and re-add them.

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1560  Product:ChemStation  ChemStation (G2070BA)  B.04.02 DSP 2 (FireAnt)

Keyword: 1200 Control

One-line Description:

TCC cluster cannot be unclustered in RC.Net configuration

Problem:

Once a TCC cluster has been created from column compartments and valves there is no option inside the software to un-cluster it.

Temporary Solution:

If the cluster needs to be dissolved into individual modules, delete the configuration file RapidControl.InstrumentConfig.xml from the instrument folder.

Fix Information:

No fix planned.


KPR#:1596  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 DSP 3

Keyword: 1200 Control

One-line Description:

Conditioning On cannot be switched off with FW B.06.30

Problem:

The instrument control for the 1290 binary pumps offers the option to activate a Conditioning Mode. This mode cannot be switched off, when the pumps is on a firmware revision below B.06.32.

Temporary Solution:

The conditioning mode will switch off after a while.

Fix Information:

Upgrade the pump firmware to revision B.06.32 or higher.


KPR#:1719  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03 LC DSP

Keyword: 1200 Control

One-line Description:

Lamp required option in FLD method is not properly migrated

Problem:

When a method for the G1321A?B FLD is migrated from classic to RC.Net drivers, one setpoint is treated incorrectly.
The setting  "Enable analysis with lamp off" in classic is equivalent to the RC.Net setpoint  "Lamp on required for analysis".  This requires a translation from Yes to No and vv.
This translation is skipped during migration.

Temporary Solution:

Manually add or remove the tick mark when migrating a method.

Fix Information:

No fix available. See Previous Note.


KPR#:1720  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: 1200 Control , Method Scouting Wizard

One-line Description:

Cluster information can be overwritten from the offline session

Problem:

Changes in a TCC or pump cluster configuration (like e.g. changing a solvent name) usually cause method resolution. 
The changes will also be updated, even when the configuration dialog is closed by pressing Cancel. This is true for online and offline. This will consequently modify methods, although nothing has actually been changed.

Temporary Solution:

None

Fix Information:

Fixed in revision B.04.03 SP1.


KPR#:1570  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 SP1

Keyword: 35900 Control

One-line Description:

All named Defined Events in Configuration Editor for ADC are rejected

Problem:

Configure a 6890 classic mode and select advanced to configure a 35900E as well.  This is saves and opens fine.

However, in Configuration Editor, select Define Events\Change...
Enter any text or words in any blank High or Low cell, and this is rejected with pop-up:
'---------------------------
Application Startup
---------------------------
The current Event Name is the same as an event name for a special event.  Please change the current event name.
---------------------------
OK   
---------------------------'

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1623  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: 35900 Control

One-line Description:

35900E - Error The Current Event Name is the same for a special event

Problem:

B.04.03 GC ChemStation with a 35900E configured. Select the CHANGE, DEFINE EVENTS tab in device configuration and you get an error "The Current Event Name is the same for a special event. Please change the current event name." 
 
This error is reported on a clean install of B.04.03 [16].

ChemStation B.03.02 works correctly.

Temporary Solution:

n/a

Fix Information:

 Fixed in revision B.04.03 SP1.


KPR#:1677  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: 35900 Control

One-line Description:

Cannot start two instruments using the two channels of one ADC at the same time

Problem:

When two instruments are configured that use channel A and channel B of the same ADC converter, respectively, the two instruments cannot be running at the same time. When the second instrument is started, both ChemStation instruments will freeze and can no longer be operated.

Temporary Solution:

The two instruments will work when using only one of the instruments at a time.

Fix Information:

Resolved in ChemStation Edition C.01.07 SR1 Hotfix01 [113].  Use the 35900E RC.NET driver if you need to split the channels between 2 instruments on an AIC distributed system.


KPR#:1707  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: 35900 Control

One-line Description:

35900:  Unable to set up Timed Events

Problem:

In the 35900 configuration, the Timed Events table is populated with default values. 
In the method, with the injection source set to 35900 Pre-Run Events, the Setup ADC Timed Events window opened but would not permit setup of the events. All buttons and drop down menus were gray and inaccessible.

Temporary Solution:

n/a

Fix Information:

 Fixed in revision C.01.04


KPR#:1721  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: 35900 Control

One-line Description:

View ACQ method from Data Analysis does not list acquisition parameters for the 35900 AD converter.

Problem:

The acquisition parameters of an AD converter are not listed in the method printout from Data Analysis view.

Temporary Solution:

Use the Print Method... option from Method&Run Control view.

Fix Information:

Resolved in ChemStation Edition C.01.07 SR1 Hotfix01 [113] and with the Agilent 35900E RC.Net Add-On driver.


KPR#:1739  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: 35900 Control

One-line Description:

ADC can be non-responsive after long reprocessing

Problem:

The ADC might become NOT READY when the software is busy for a long time (e.g sequence reprocessing with IR to screen). When the system is no longer busy it cannot communicate with the ADC.

Temporary Solution:

Software and instrument need to be restarted.

Fix Information:

Fixed in revision C.01.04.


KPR#:1804  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: 35900 Control

One-line Description:

Queue method generated status “error” when using 35900E

Problem:

Queue method generated status “error” when using 35900E

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1047  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.02

Keyword: 5890 Control

One-line Description:

5890: Data rate 10Hz used, Real Time Plot display is doubled

Problem:

When the data rate of 10 Hz is used, the Real Time Plot display of the chromatogram elapsed time is doubled. 

When the data rate is changed to the default of 5 Hertz, the Real time Plot window was normal, displaying the same times.

Temporary Solution:

n/a

Fix Information:

No fix planned.


KPR#:707  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: 6850 Control

One-line Description:

6850 GC feature loss, such as Prep Run,  when controlled by Chemstation

Problem:

On the 6850 (with no hand held controller (HHC)) it is not possible (very difficult) for a user to activate the "Prep run" function.    It should be possible to turn the "Prep Run" function ON from the Chemstation user interface.  Customers without the HHC can not turn the "Prep Run" function ON.  The HHC must be used.

This functionality is important for customers using non-Agilent samplers (P&T, etc.)

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1511  Product:ChemStation  ChemStation (G2070BA)  B.04.02 SP1

Keyword: 6850 Control

One-line Description:

6850 instrument type does not show up on GLP report and ACAML file is incomplete

Problem:

The GC 6850 instrument type does not show up on ChemStation's GLP report and the reserved field in the ACAML is incomplete.

As a consequence the OpenLAB ECM Synchronization Service cannot process the 6850 data, as the database field does not allow a NULL entry.
The OpenLAB Intelligence Reporter cannot report any 6850 data.

Temporary Solution:

A manual modification of the config register is a temporary workaround.

Fix Information:

This problem will be fixed in revision B.04.03.


KPR#:1472  Product:ChemStation  ChemStation (G2070BA)  B.04.01 SP2

Keyword: 6850 Control , 6890 Control

One-line Description:

B.01.02 Lab Advisor is not compatible with ChemStation B.04.02 and higher

Problem:

B.01.02 Lab Advisor is not compatible with ChemStation B.04.02.  The GC will not start up.

Temporary Solution:

n/a

Fix Information:

B.01.03 Lab Advisor is compatible with B.04.02


KPR#:415  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: 6890 Control

One-line Description:

6890 Workstation Internal Software Error

Problem:

The ChemStation would open with an Error Dialog Box titled: 

  6890 Workstation Internal Software Error

Windows XP was set to auto update, when it did the Microsoft Hotfixes, one of them changed the decimal separator from a decimal point to a comma.

The Fix:  Change the comma back to a decimal point in Regional Settings, and it works as advertised.

Temporary Solution:

n/a

Fix Information:

The Fix:  Change the comma back to a decimal point in Regional Settings, and it works as advertised.


KPR#:422  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: 6890 Control

One-line Description:

GC system crashes when view/instrument diagram active in Offline

Problem:

When exiting the OFFLINE ChemStation with the view/instrument diagram actived, there's a graphical crash message that appears before closing.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:848  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: 6890 Control

One-line Description:

6890 FPD Classic driver is missing the "Air" and "Makeup" Actuals in Edit Parameters

Problem:

The "Air" and "Makeup" Actuals are missing from the 6890 FPD Detectors screen in Edit Parameters.

Temporary Solution:

The correct flows can be viewed from the GC front panel and also “Instrument Actuals”.  The FPD functions correctly.

Fix Information:

Fixed in 68xx RC.NET revision 6.10.


KPR#:1167  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: 6890 Control

One-line Description:

IQ1.CFG file of 6890/6850 instrument corrupted after regional setting change

Problem:

From Control panel\Regional settings: change the decimal separator from a period to a comma.

It is not possible to open a 6890/6850 ChemStation.

Temporary Solution:

1.  Backup data, methods, sequences, column databases (found in C:\Chem32\DRIVERS), etc 

Record GC IP addresses

2.  Delete IQ1.CFG located under the Chem32\1 directory for example

3.  Uninstall Chemstation

4.  Reinstall Chemstation

Fix Information:

n/a


KPR#:1352  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: 6890 Control

One-line Description:

6890 RC.NET driver:  instrument diagram not updated to show configuration

Problem:

In the 6890 RC.NET driver, changed the GC configuration to have the back inlet go to the back column to the back detector.  This change was not reflected in the instrument diagram as it is still showing the back inlet going to the front column to the back detector.

Temporary Solution:

The workaround is to turn off and on the instrument diagram to reflect the new configuration changes.

Fix Information:

n/a


KPR#:1412  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: 6890 Control

One-line Description:

6890 RC.NET driver:  instrument diagram not updated to show configuration

Problem:

In the 6890 RC.NET driver, changed the GC column configuration to have the back inlet go to the back column to the back detector.  This change was not reflected in the instrument diagram as it is still showing the back inlet going to the front column to the back detector.

Temporary Solution:

After the configuration change is made to the column connections, the workaround is to shut down the ChemStation software and restart to reflect the new configuration changes.

Beginning with  B.04.03 DSP1 [54], after updating the column configuration, save and reload the method will update the instrument diagram.

Classic 68xx driver works correctly.

Fix Information:

No fix Planned.


KPR#:1461  Product:ChemStation  ChemStation (G2070BA)  B.04.02 Aladdin

Keyword: 6890 Control

One-line Description:

6890N with no tray or ALS is seeing GC Injector Front/Back/Dual as an injection source

Problem:

6890N with no tray or autoinjectors installed.
Select Injection Source/Location.
GC Injector Front/Back/ and Dual is invalidly displayed as an injection source.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1464  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: 6890 Control

One-line Description:

Addition of new columns to the column catalog (database) is not possible after upgrade

Problem:

Upgrading from B.01.03 to B.04.02, it is no longer possible to add new columns to the GC column catalog (database) in Rev. B.04.02.

The following error is shown:

The Catalog File cannot be opened.  If needed, you may obtain an appropriate replacement file from Agilent.
C:\Chem32\CORE\...\DRIVERS\catalog.mdb

Temporary Solution:

The problem is that there are several columns that have the same manufacturer and model number. Our new database software doesn’t handle this.

The newer database software (in B.04.02 for example) no longer allows duplicates of manufacturer + model number. i.e., there can only be one HP HP-5 in the database.  Note: the new software will inform the user that the model number “xxx” is already in use and request that a different value be chosen. (The user is returned to the original entry screen with all the info filled in.) 

In theory, customers should not be creating more than one catalog entry with the same model number.  It’s like having two items in the LLBean catalog with the same order number!  But, of course, the older software allowed the duplicates.

Fix Information:

Before upgrading, fix the problem with the column database duplicates while running B.01.03 for example.  Review the column catalogs and make sure there are no duplicates and fix any that exist before the upgrade.


KPR#:1684  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: 6890 Control

One-line Description:

6890 always opens to Configuration tab when opening Edit Agilent 6890 GC Parameters

Problem:

Select Edit Agilent 6890 GC Parameters…
The tab will always open to the configuration tab.

Temporary Solution:

n/a

Fix Information:

 Fixed in revision B.04.03 SP1.


KPR#:1784  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02

Keyword: 6890 Control

One-line Description:

6890 RC.NET: Edit a running Method by making the oven hold time longer and click ok, then the Sequence/Single run stops

Problem:

During a run, the GC method parameters are opened.  Select OK after increasing the oven parameter hold time.  The sequence or single run will just stop and the method is dirty. 

Logbook shows:
Attempt to download method to GC failed.
Instrument Error - Method/Sequence stopped
Method Aborted
Instrument run aborted
Method Aborted
Sequence terminated due to an error

Temporary Solution:

n/a

Fix Information:

Issue fixed in OpenLAB C.01.04.


KPR#:16  Product:68xx Driver Software  6890  5.03

Keyword: 68xx Driver

One-line Description:

Vial Missing Error Message

Problem:

When the 6890 GC encounters a missing vial, the IALog displays Stop key pressed error #1 instead of "Missing Vial."

Temporary Solution:

None

Fix Information:

This is fixed in 68XX Driver version 6.21 which is included in C.01.05 OpenLAB ChemStation and A.04.05 OpenLAB EZChrom.


KPR#:18  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

If One Signal Needs to be Saved, it Must be Assigned as Signal 1

Problem:

If only one signal needs to be saved, you must assign it as signal 1.

Temporary Solution:

If only one signal needs to be saved, you must assign it as signal 1.

Fix Information:

Fixed in 68xx driver revision 6.11[008].  This driver is included in B.04.03 SP1 ChemStation and C.01.03 OpenLAB CDS.


KPR#:20  Product:68xx Driver Software  6890  5.10

Keyword: 68xx Driver

One-line Description:

Instrument Diagram Does Not Match Installed Column in Method

Problem:

The Instrument diagram of the GC instrumentation indicates a different column configuration than is set in the method.  The GC keyboard will reflect the new changes, only the software instrument diagram is affected.

Temporary Solution:

Beginning with B.04.03 DSP1 [54], after the configuration change is made to the columns, the workaround is to save the method and then load it.  This will update the displayed column configuration.  

Another workaround is to shut down the ChemStation software and restart to reflect the new configuration changes.  

These workarounds may not work for all configurations.

Classic 68xx driver works correctly.

Fix Information:

None


KPR#:22  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

In 68XX driver revision 5.02 - 6.11, it is not possible to Modify a Method During a Run to Extend the Run Time or Change the Oven Temp

Problem:

It is possible to the software edit parameters screen and change the oven hold time or oven temp and select OK.  The GC does not accept the updates.  The method in the software is changed to the new values; the GC ignores all.

Temporary Solution:

n/a

Fix Information:

68XX Driver 6.11 is included in B.04.03 SP1, OpenLAB ChemStation C.01.04 and OpenLAB EZChrom A.04.04.


KPR#:23  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

Info in Edit instrument parameters\Columns is incomplete

Problem:

Edit instrument parameters; columns

Applies to GC’s: 7890A, 6890N, 6850A

The window displayed is incomplete; it shows the column type and flow/pressure parameters, but does not give the length, ID or film thickness.  These are critical parameters, and should be displayed just as they were in the previous ChemStation revisions.  To find these parameters, the user now has to look on the configuration window.

Temporary Solution:

n/a

Fix Information:

Fixed in 68xx driver revision 6.11[008].  This driver is included with OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.


KPR#:31  Product:68xx Driver Software  6890  5.11

Keyword: 68xx Driver

One-line Description:

Ignore missing vial option is not present in GC

Problem:

Run a sequence with 11 runs and the 11th vial is missing.  The sequence aborts and SSIZip file was not uploaded to ECM as the 11th vial was missing.

The ignore missing vial option is not present in GC.

Temporary Solution:

n/a

Fix Information:

This is fixed in GC Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:33  Product:68xx Driver Software  6890  5.10

Keyword: 68xx Driver

One-line Description:

Instrument Setup graph has incorrect column legend

Problem:

In the 6890, the legend for the Instrument Setup graph says Column 2, irrespective of which column is highlighted.  The graph is correct, it is just the legend that is wrong.

Temporary Solution:

n/a

Fix Information:

This has been fixed in 68XX Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:38  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

ALS errors abort sequence; no Retry/Abort dialog box option offered with 68xx RC.NET Driver

Problem:

6890A RC.NET driver with ALS. 

If an ALS error occurs such as a Vial not found event, there is no dialog box option presented to Retry/Abort.  The sequence simply aborts.

Temporary Solution:

n/a

Fix Information:

This is fixed in GC Driver version 6.21, which is available with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:41  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

Method parameters from B.01.03 changed to defaults after upgrade to B0402 RC.net

Problem:

Methods from 6850 B.01.03 were loaded on the upgraded revision B.04.02 RC.NET driver. The 6850 GC methods were changed.

The method parameters that are changed are: 

· Oven temperature (switched to default 50 C)
· All gradient parameters are gone (I assume the mean temperature program)
· Gas type (switched from H2 to default He)

Temporary Solution:

n/a

Fix Information:

This problem is fixed in Software Driver Version: 6.01.  This driver is included with B.04.02 SP1, B.04.02 SP1 Hotfix 2, B.04.02 DSP3 and B.04.03 ChemStation.


KPR#:55  Product:68xx Driver Software  6890  6.01

Keyword: 68xx Driver

One-line Description:

Opening 68xx RC.NET GC types result in 'FetchCPVar cannot talk to application'

Problem:

Configure 6850 RC.NET or 6890N RC.NET GC. 

Open an online session, initial or any subsequent try. This results in a benign pop-up error:  'FetchCPVar cannot talk to application'.

Temporary Solution:

Go to view menu and turn off the sampling and instrument diagrams.  This will keep hpgcui00.exe from starting up.

Fix Information:

Fixed in driver revision 6.02.
Driver included in OpenLab C.01.01.


KPR#:59  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

In 68XX Driver Version 5.02, it is not Possible to Start the 6890 RC.NET from the Front Panel Start Key

Problem:

In 68XX Driver Version 5.02, it is not possible to start the 6890 RC.NET from the front panel Start key.  Software configuration is Injection source = manual.  6890 is READY, press start and get error message "ERROR Start Key disabled".

Temporary Solution:

The workaround is for the user to start the run from the "run control" then "run Method" in the Chemstation software and then press start at the GC.

Fix Information:

Fixed in driver revision 6.10.  This driver is included with B.04.03 DSP1 ChemStation.  It is also included with OpenLAB ChemStation C.01.02 and OpenLAB EZChrom A.04.02.


KPR#:66  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

CIS4 inlet is not recognized with 6890 GC RC.NET

Problem:

CIS4 inlet is not recognized with 6890 GC RC.NET

Temporary Solution:

n/a

Fix Information:

This is fixed in GC Driver version 6.21, which is available with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:83  Product:68xx Driver Software  6890  6.01

Keyword: 68xx Driver

One-line Description:

Only Fast Plunger Speed is available for PTV Solvent Vent multiple injections

Problem:

The fast plunger speed is only available when using multiple PTV Solvent Vent injections.  Most users of multiple PTV Solvent Vent injections need to use Slow or Variable plunger speed, not Fast.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 6.10.  This driver is included with B.04.03 DSP1 ChemStation and OpenLAB C.01.02 ChemStation and OpenLAB A.04.02 EZChrom.


KPR#:87  Product:68xx Driver Software  6890  6.02

Keyword: 68xx Driver

One-line Description:

IQ1.CFG file of 6890/6850 instrument corrupted after regional setting change

Problem:

From Control panel\Regional settings: change the decimal separator from a period to a comma.

It is not possible to open a 6890/6850 ChemStation.

Temporary Solution:

1.  Backup data, methods, sequences, column databases (found in C:\Chem32\DRIVERS), etc 

Record GC IP addresses

2.  Delete IQ1.CFG located under the Chem32\1 directory for example

3.  Uninstall Chemstation

4.  Reinstall Chemstation

Fix Information:

Period is required in Regional Settings.


KPR#:89  Product:68xx Driver Software  6850  5.10

Keyword: 68xx Driver

One-line Description:

Incorrect injection volume shown in method report if PTV solvent vent is used

Problem:

6850 with a PTV inlet

The method specifies PTV solvent vent mode with 3 injections, ten seconds apart.  The method volume is 5 µL.  Specify the method to use method volume for three injections.  Each injection is 5 µL.  However the report shows the injection volume as 1 µL.  This is incorrect.  The report should show the injection volume as either 5 µL or 5 µL x 3, or 15 µL.

Temporary Solution:

n/a

Fix Information:

Fixed in software driver 6.10 [023].  This fix is included in B.04.03 DSP1 and later versions of ChemStation.  It is included in OpenLAB ChemStation C.01.02 and OpenLAB EZChrom A.04.02 and later.


KPR#:91  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

B.04.03 DSP1 [52] Memory Leak may cause GC ChemStation with 68xx RC.Net driver to stop working

Problem:

A memory leak has been discovered in the 68xx RC.NET driver revision 6.10 [022]. The 6.10 [022] is included in B.04.03 DSP1.  As a consequence of the memory leak, the B.04.03 DSP1 [52] GC ChemStation software may stop working during a sequence run or when the system is left idle over long periods of time.  There is no error message.

Temporary Solution:

It is recommended to reboot the PC once per week until a fix is available.  This is needed only when using the 68xx RC.NET Driver.

Fix Information:

Fixed in 68xx driver revision 6.11[008].  This driver is included in B.04.03 SP1 ChemStation and C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom.


KPR#:92  Product:68xx Driver Software  6890  6.01

Keyword: 68xx Driver

One-line Description:

Instrument Diagram Does Not Match Column 2's Inlet Configuration

Problem:

Column 2 inlet configuration will never be correct if column 2 is connected to anything other than inlet 2.

Temporary Solution:

n/a

Fix Information:

Fixed in 68xx driver revision 6.10[044].  This fix is included in these minimum versions:B.04.03 SP1 ChemStation, OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.


KPR#:94  Product:68xx Driver Software  6890  6.01

Keyword: 68xx Driver

One-line Description:

68xx Method Download Times Out

Problem:

Connect to 6890N GC.

Within seconds of successfully opening an online session, logbook logs the following simultaneously with the loading of the method.

'Logbook      Welcome to ChemStation                         12:17:12 06/28/10
   0 1da6 4c28cb1f    0
6890       1 Timed out waiting for method download to fin>  12:17:35 06/28/10
   0 1da6 4c28cb1f    0
6890       1 ish.                                           12:17:35 06/28/10
7da 41e0 4c28cb1f    0
Method       Loading Method LCV_GCCONTROL.M                 12:17:35 06/28/10'

Temporary Solution:

n/a

Fix Information:

Fixed in GC Driver revision 6.10 [027].  This driver version is included in B.04.03 DSP1 ChemStation and C.01.02 OpenLAB ChemStation and A.04.02 OpenLAB EZChrom.


KPR#:95  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

"Method modified after loading" occurs when running a sequence (6890 RC.NET)

Problem:

The "Method modified after loading" flag occurs after lauching the sequence.

Started the sequence from the ChemStation software.  The method modification status indicates that the method has changed when it was not changed and the report will indicate 'modified after loading'.

The GC had a calibrated capillary column that was changed to a packed column.  When using this method that now has a packed column, the GC reports an error in the length parameter, and that is causing the method download not to reset the 'modified after loading' flag.

Temporary Solution:

There is a workaround: bring Method into MethodEdit UI, and "reconfigure" column (Packed to Packed would be OK) to get new structure.

Fix Information:

Fixed in GC Driver revision 6.10 [027].  This driver is included in B.04.03 DSP1 ChemStation and OpenLAB C.01.02 ChemStation and OpenLAB A.04.02 EZChrom.

Note, this fix will only apply to new methods, not to current methods.


KPR#:98  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

GC Column Catalog File Cannot be Opened When Using the 68xx Driver Version 6.10

Problem:

It is not possible to add new columns to the GC column catalog (database) using the RC.NET driver.  It is possible to add columns to the same column database using the Classic driver.  User switched back to classic on same system and could add columns without error.

The following error is shown:

The Catalog File cannot be opened.  If needed, you may obtain an appropriate replacement file from Agilent.
C:\Chem32\CORE\...\DRIVERS\catalog.mdb

The RC.NET driver is prevented from opening the db because you can’t have two column entries with the same model number!

Temporary Solution:

Start with a new, “stock” db column catalog file from another system.

Fix Information:

This has been fixed in 68XX Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:99  Product:68xx Driver Software  6890  1.01

Keyword: 68xx Driver

One-line Description:

Sequence continues while GC temperature is not ready

Problem:

When running a sequence where consecutive runs use the same method using the ALS on a data system containing 6890 driver version 6.10.027 or earlier, the data system may proceed to start a new run before the GC has equilibrated after the conclusion of the previous run.  This is more likely if the method contains a post run period of sufficient length that the GC equilibrates to the post-run conditions and the data system has completed data analysis well before the end of the post-run period.

This could result in a (retention time) RT shift.

Temporary Solution:

n/a

Fix Information:

Patch available for RC.NET GC driver revision 6.10 [027] for B.04.03 DSP1 [54].

Fixed beginning in: B.04.03 SP1 [87] and OpenLAB C.01.03 [37].


KPR#:100  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

A Method Can be Created Which Will Not Download to the Instrument and the Sequence Does Not Run

Problem:

Method download fails, sequence aborts after start; no injection made

Error reported in logbook:
Attempt to download method to GC failed.

The Data System will not allow the run to continue when the method download has failed.

Temporary Solution:

n/a

Fix Information:

This fix is in the 68xx driver version 6.10.032 and 6.11.001.  

This fix is included in B.04.03 SP1 and OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.


KPR#:101  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

GC Back Injector Not Present in Edit Method Screen

Problem:

After successfully running a sequence with a back method, another back method is loaded. The back injector is not present in the edit method screen.

Temporary Solution:

Change a parameter at the GC keyboard and click "upload from instrument" in the edit method screen. The back injector will show in the edit method screen.

Fix Information:

This has been fixed in 68XX Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:102  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Validation failed shown on left corner of Resolve Method screen

Problem:

Load a method which was developed for both detectors and has the front detector signal chosen under signals tab.
Load this method on a GC with no front detector - method resolution. 

Changes made to the detector signal tab are not detected, tab remains yellow (invalid).

Temporary Solution:

Click on a different tab and modify the value of any field to a valid value or click on the configuration tab and then click away. Change the signals tab to a valid value or turn it off to resolve the method.

Fix Information:

This has been fixed in 68XX Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:103  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Injector Panels in GC Edit Method Show as Uninstalled When an Injector is Installed.

Problem:

The front injector was unplugged. Close software, plug injector in, restart software.  Method Configuration/ALS panel and Injector panel both say: Front Injector Not Installed. The GC shows that it is installed.  Software inlet and injection parameters panel show that it is installed as well.

Temporary Solution:

n/a

Fix Information:

This has been fixed in 68XX Driver version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:105  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

Plunger Speed is Not Saved with 6850 GC Methods

Problem:

Plunger speed parameters are not saved in methods.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 6.11[002].  This fix is included in B.04.03 SP1 ChemStation, C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom.


KPR#:106  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Opening the GC method parameters to only look (no change made) and selecting OK to close stops the sequence or single run

Problem:

During a run, the GC method parameters are opened via the instrument diagram icon.  Select OK after looking at the oven parameter to close the screen.  The sequence or single run will just stop and the method is dirty.  See Log File for error message.

Log File:
  67 41e2 4e69f8b3    0
Method       Method started                                 07:29:55 09/09/11
  67 41e2 4e69f8b4    0
Method       Instrument running sample Vial 1 (front)       07:29:56 09/09/11
  67 41e2 4e69f8b4    0
Method       Instrument running sample Vial 2 (back)        07:29:56 09/09/11
   0 1da5 4e69f91a    0
6890       1 Attempt to download method to GC failed.       07:31:38 09/09/11
  67 41e2 4e69f91a    0
Method       Instrument Error - Method/Sequence stopped     07:31:38 09/09/11
  67 41e2 4e69f91a    0
Method       Method aborted                                 07:31:38 09/09/11

Temporary Solution:

Click the Cancel button rather than the OK button if the intent was to check the values of some of the instrument setpoints.  This would not interfere with any run which is in progress.

Fix Information:

Fixed in 68xx driver revision 6.11[008].  This is included in B.04.03 SP1 and OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.


KPR#:109  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Method Contents Report Displays the Incorrect Configuration of Front and Back Detectors for 6890 GC

Problem:

The report displays the incorrect configuration. With signal-1 selected as the back detector and integration events for back detector. The method contents report displays the integration events for the front detector and the remaining paramaters are for the back detector are configured to the front detector.

Temporary Solution:

n/a

Fix Information:

Fixed in 68xx driver revision 6.11[003].  This fix is included in B.04.03 SP1 ChemStation, C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom and later.


KPR#:113  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Cannot Run a Packed GC Column with a 6890

Problem:

Method download fails, sequence aborts after start when trying to run a 6890 with a packed column.  
  
Error reported in logbook:
Attempt to download method to GC failed.

The Data System will not allow the run to continue when the method download has failed.

Temporary Solution:

This can be run by setting the column as a capillary column but that changes the chromatography.

Fix Information:

Fixed in driver revision 6.10[032] and 6.11[001].

This fix is included with B.04.03 SP1 GC ChemStation, C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom


KPR#:116  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Pulsed Split/Splitless Injection Does Not Wait For Ready on a 6890.

Problem:

The GC does not wait for the inlet pressure to increase to and stabilize at the pulsed pressure before injection and data acquisition. The GC starts and injects sample at the same time, then starts data acquisition and increase of inlet pressure to pulsed pressure.

Temporary Solution:

If the 6890 is operated from the GC keyboard, the GC has the expected result.

Fix Information:

Fixed in 68xx driver revision 6.11[007].

This fix is included with B.04.03 SP1 ChemStation, OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03


KPR#:117  Product:68xx Driver Software  6890  6.01

Keyword: 68xx Driver

One-line Description:

In 68XX GC Driver Version 6.01, 2 Injections are Made When Using the PTV

Problem:

Set up the injector to make 1 ul injection x 1. The number of injections will revert to 2 even if the user enters 1.

Temporary Solution:

n/a

Fix Information:

This is fixed in driver revision 6.10 [27]  which is included in B.04.03 DSP1 [54] GC ChemStation, C.01.02 OpenLAB ChemStation and A.04.02 OpenLAB EZChrom.


KPR#:119  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Vial left in gripper; no message shown to alert user about the problem

Problem:

After a stop, if there is a vial in the gripper it is not returned to its original location, it is kept in the gripper until the start of a new run. In the interim, if a new vial is placed in the open location, during the next run the system runs into an error because the gripper wants to return the vial to its original location which is now occupied.

Temporary Solution:

n/a

Fix Information:

This is fixed in GC Driver version 6.21, which is available with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:120  Product:68xx Driver Software  6890  6.11

Keyword: 68xx Driver

One-line Description:

6850 vial numbers are not consistent with standard numbering system.

Problem:

A 6850 with a G2613A tower and a three vial turret will accept only single digit vial locations, ie., 1, 2, 3 instead of 101, 102, 103. Attempts to enter the correct vial number generates an error message "The vial 101 is not allowed with the current device configuration."

The also occurs when entering correct vial locations in a sequence table.

Temporary Solution:

n/a

Fix Information:

Fixedi n 68XX driver revision 6.11.  This driver is included with B.04.03 SP1 ChemStation, OpenLAB C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom.


KPR#:122  Product:68xx Driver Software  6890  6.10

Keyword: 68xx Driver

One-line Description:

Method Resolution Changes Oven Temperature Program

Problem:

Load a method from a 6890 GC with a fast oven on a 6890 GC with different hardware and a normal oven. Method resolution occurs. The oven temperature program from the fast oven GC has not been loaded on the normal oven GC.

Temporary Solution:

n/a

Fix Information:

Fixed in 68xx driver revision 6.11[002].  Fix is included in B.04.03 SP1 ChemStation, OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.


KPR#:123  Product:68xx Driver Software  6890  6.11

Keyword: 68xx Driver

One-line Description:

Column info is missing with 6890

Problem:

The column specified as being used is missing information: description, Manufacturer and Model Number.

Temporary Solution:

n/a

Fix Information:

Fixed in 68xx driver revision 6.11[005].

This fix is included in B.04.03 SP1 GC ChemStation, C.01.03 OpenLAB ChemStation and A.04.03 OpenLAB EZChrom.


KPR#:124  Product:68xx Driver Software  6890  6.11

Keyword: 68xx Driver

One-line Description:

Method Modification Flag is Set after Edit Method: Occurs on first connection or after upload from GC.

Problem:

1. First Connection to an online GC and save the method OR connect to an instrument and save the method, open method editor, click on ‘Upload from instrument’. Click ‘Ok’.
3. Click ‘Edit entire Method’, go on clicking ‘Ok’ without making any change.
4. Method is dirty at the end.

This only occurs on 6850 GCs with G4513A injectors.

Temporary Solution:

Save the method as a new name after initial connection or subsequent method upload.

Fix Information:

Fixed in 68xx driver revision 6.11[008].  This is included in B.04.03 SP1 and OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.


KPR#:128  Product:68xx Driver Software  6890  6.11

Keyword: 68xx Driver

One-line Description:

Method Ddownload Failure Occurs Due to Column Calibration Limits

Problem:

A valid column calibrated using calculate diameter function will result in a method download failure if the change from the original diameter is more than 20 microns. 

This occurs after OKing method changes that are downloaded, and will occur even if the column values are valid dimensions.

Temporary Solution:

n/a

Fix Information:

This is fixed in 68XX GC Driver Version 6.21, which is included with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:129  Product:68xx Driver Software  6890  6.11

Keyword: 68xx Driver

One-line Description:

Instrument actual value for COC inlet does not display correctly

Problem:

6890N GC with COC inlet connected to FID. The COC inlet pressure setpoint reflects the correct pressure set for the column, however the instrument actual value (blue text) in the setup method screen is displayed as 0.0.

Temporary Solution:

Enable setup actuals (inlet) in the setup method screen, the inlet pressure is properly displayed.

Fix Information:

Fixed in 68xx driver revision 6.11[018].  This driver is included in B.04.03 SP1 ChemStation and OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.


KPR#:130  Product:68xx Driver Software  6890  Monitoring

Keyword: 68xx Driver

One-line Description:

Instrument actual value for anode purge flow on an ECD does not display correctly

Problem:

The ECD anode purge flow setpoint reflects the correct value, however the instrument actual value (blue text) in the setup method screen is displayed as 0.0.

Temporary Solution:

Enable setup actuals in the setup method screen, the actual anode purge flow displayed properly.

Fix Information:

Fixed in 68xx driver revision 6.11[018].  This driver is included in B.04.03 SP1 ChemStation and OpenLAB ChemStation C.01.03 and OpenLAB EZChrom A.04.03.


KPR#:134  Product:68xx Driver Software  6890  6.11

Keyword: 68xx Driver

One-line Description:

Easy SamplePrep reset of one or any group of resources creates log entry:  'User reset all resources'

Problem:

Create a resource layout. Select menu item Reset Resources... From the list, select one or several resources and OK. The logbook always logs 'User reset all resources' no matter how many resources are actually reset.

Temporary Solution:

n/a

Fix Information:

There is no fix available.


KPR#:136  Product:68xx Driver Software  6890  5.02

Keyword: 68xx Driver

One-line Description:

Data Files do not Contain the Correct Signals if the Front Inlet is Connected to Back Detector

Problem:

Data files do not contain the correct signals if front inlet is connected to back detector

Temporary Solution:

n/a

Fix Information:

This is fixed in GC Driver version 6.21, which is available with OpenLAB ChemStation C.01.05 and OpenLAB EZChrom A.04.05.


KPR#:911  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: 7890A Control

One-line Description:

Initial valve position set to ON before start when it should be OFF; Meth change

Problem:

Setup method with 2 valves (1 and 2) with initial position to OFF and valve switching to on at 0.01 minutes. First time it runs the timing appears OK. The second time it runs the valves are initialized to ON (a method change ). The valve is checked or set to ‘on’ in the software (OFF on the 7890A). If you try to run the method, the software asks you to save the method, because it has changed in this manor.

This occurs with Manual injection only (Press start at 7890A).  This does not occur when using an ALS to start the run.

Temporary Solution:


The following workaround in now available beginning in B.03.02:  

Start the run from Chemstation by going to RunControl > Run Method, then press start on the 7890A.

Fix Information:

Fixed in 7890A Firmware A.01.10.


KPR#:914  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.01 SR1.1

Keyword: 7890A Control

One-line Description:

7890A: ALS errors abort sequence; no Retry/Abort option offered

Problem:

7890 with ALS. Create an ALS error, such as a Vial not found event.  Instead of offering a pop-up that offers a second chance to Retry/Abort with the named error, this simply aborts the sequence. The error is buried deep in the sequence log of that instrument:

'Method       Method started: (F) line# 2 at 2 inj# 1        09:11:51 09/20/07
  67 41e0 46f27198    0
Method       Instrument running sample Vial 2 (front)       09:11:52 09/20/07
   0 1da5 46f271b2  300
GC         1 There was no bottle in the gripper.            09:12:18 09/20/07
   0 1da6 46f271b2    0
Instrument   GCI1 did not start a run                       09:12:18 09/20/07
  67 41e0 46f271b2    0
Method       Instrument Error - Method/Sequence stopped     09:12:18 09/20/07
  67 41e0 46f271b2    0
Method       Method aborted                                 09:12:18 09/20/07
 43e 41df 46f271b2    0
Sequence     FRONTALS.S terminated due to an error          09:12:18 09/20/07'

Temporary Solution:

n/a

Fix Information:

Issue resolved in OpenLAB C.01.05.


KPR#:985  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: 7890A Control

One-line Description:

7890A: Auto Start Macro fails -9999

Problem:

Sometimes the software shuts down for no reason. A reboot then shows sometimes the “autostart macro failed, Error -999”. After some more tries the Chemstation starts without any problem.

Temporary Solution:

Reinstallation of the Chemstation has fixed this problem.

Fix Information:

n/a


KPR#:1049  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: 7890A Control

One-line Description:

7890A: Method modified after loading if run started from GC keyboard start butto

Problem:

Manual injection.

Started the run from the 7890A keyboard start button.  The method modification status indicates that the method has changed when it has not and the report will indicate 'modified after loading'.

Temporary Solution:

Start the run first by going to RunControl > Run Method and then press start at the 7890A.

Fix Information:

This problem no longer occurs beginning with B.04.02.


KPR#:1110  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: 7890A Control

One-line Description:

7890A Full report does not contain ALS module type information

Problem:

7890A with injectors

In Data Analysis, load a signal from this GC and Specify 'Full' Report.

Print Report.

The Module section does not recognize or print any ALS module types or Firmware for it.

Temporary Solution:

n/a

Fix Information:

No Fix Planned.


KPR#:1117  Product:ChemStation  ChemStation (G2070BA)  (None)

Keyword: 7890A Control

One-line Description:

Can Accept Missing Location Vial in Sequence Table

Problem:

On a 7890 GC without 100 vial tray it is nevertheless possible to input vial locations that aren't available when using the filldown wizard of the sequence table.

Temporary Solution:

n/a

Fix Information:

There is no plan to fix this issue.


KPR#:1151  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: 7890A Control

One-line Description:

7890A: Signal data is not acquired if oven hold time is set to 0.0

Problem:

When running an isothermal GC run, if the oven temp hold time is set to 0.0, signal data will not be acquired.

A data file is generated but does not contain the .ch file.  When this data file is opened in data analysis, the message no signal data is displayed.

Temporary Solution:

Enter a hold time greater than 0.0 in the hold time field will resolve this problem.

Fix Information:

n/a


KPR#:1169  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: 7890A Control

One-line Description:

7890A seq aborts part way through run: "AgGC7890Drv" already has ownership"

Problem:

Sequence aborts in the middle of the run, 7890A lost connection. 

In the instrument log, it states that the AgGC7890Drv already has ownership; request denied.

In the Chemstation software status view, in some cases the GC will come back in Monitor mode.

Temporary Solution:

n/a

Fix Information:

This defect is fixed beginning in 7890A Firmware revision A.01.09.  Lab Monitor & Diagnostic SW must be used to update the 7890A firmware.  If the old 'Ohana' Utility is used, this problem will not be solved even in firmware A.01.09.  This old utility does not update the 7890A firmware properly.


KPR#:1198  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: 7890A Control

One-line Description:

7890A does not show the vial number being injected on the sequence screen

Problem:

The vial number is not shown in the sequence screen.  Only a "#" sign is shown.  

Temporary Solution:

n/a

Fix Information:

Fixed in 7890A Firmware Rev A.01.10 with A.01.02 ALS.


KPR#:1239  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: 7890A Control

One-line Description:

7890A Runs cannot be started until the GC and injector are idle, view GC Status

Problem:

Occassionly, when trying to start a run or sequence on a GC whose front display and ALS are Ready, fails to start. Logbook shows:
'Runs cannot be started until the GC and injector are idle'.
This can occur after an abort or stop of sequence, or connecting to a GC that is controlled by another PC.

Temporary Solution:

Pressing "stop" on the GC front panel enabled the GC to makes runs again.  Power cycling the GC may be necessary.

Fix Information:

Fixed in revision B.04.03.


KPR#:1255  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: 7890A Control

One-line Description:

7890A Parameter Mismatch error after upgrade

Problem:

Parameter Mismatch in the Signal Descriptions between Detector Parameter and Calibration data after upgrading to B.04.01.  See attached screen shot.

Temporary Solution:

After the upgrade, set the signals to None, save the method and add the signals.

Another workaround could be to add the space manually to the calibration signal description.

Fix Information:

ChemStation GCI drivers which were delivered prior to B.04.01 composed an incorrect Signal Description (the space was missing).  Beginning with B.04.01, there is a space.


KPR#:1257  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: 7890A Control

One-line Description:

GCI1 did not start a run shown in instrument logbook

Problem:

Loaded method and method resolution screen is shown.  Selected 'Don't Resolve' button.

The following message is shown in the instrument log:
Method not downloaded to the device because the settings are not compatible with the device.

Save the method.

Start a run.

The following message is shown in the instrument log:
Method not downloaded to the device because the settings are not compatible with the device and GCI1 did not start a run.

Temporary Solution:

Print the method that will not run or view the 7890A method parameters from the software.  Go to Instrument > Agilent 7890A Configuration and correct the Configuration to match the method.  99% of the time it is a Column difference.

Fix Information:

n/a


KPR#:1283  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: 7890A Control

One-line Description:

7890A  Two inlets with one detector and one injector causes method resolution

Problem:

The system includes a 7890A with 2 inlets, single detector and a single tower.  When moving the tower from the front position to the back position and again to the front, with correct front method loaded results in method resolution.

Temporary Solution:

Workaround:  load the method and when meth resolution user interface is shown, select don’t resolve.  Then go to the GC Config and change the GC Config to match the just loaded method.

Fix Information:

n/a


KPR#:1329  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.01

Keyword: 7890A Control

One-line Description:

ALS errors abort sequence; no Retry/Abort dialog box option offered

Problem:

7890A with ALS. 

If an ALS error, occurs such as a Vial not found event, there is no dialog box option presented to Retry/Abort.  The sequence simply aborts. The error is shown in the sequence log of that instrument:

'Method       Method started: (F) line# 2 at 2 inj# 1        09:11:51 09/20/07
  67 41e0 46f27198    0
Method       Instrument running sample Vial 2 (front)       09:11:52 09/20/07
   0 1da5 46f271b2  300
GC         1 There was no bottle in the gripper.                09:12:18 09/20/07
   0 1da6 46f271b2    0
Instrument   GCI1 did not start a run                              09:12:18 09/20/07
  67 41e0 46f271b2    0
Method       Instrument Error - Method/Sequence stopped     09:12:18 09/20/07
  67 41e0 46f271b2    0
Method       Method aborted                                 09:12:18 09/20/07
43e 41df 46f271b2    0
Sequence     FRONTALS.S terminated due to an error          09:12:18 09/20/07'

GC goes back to "ready" state instead, and vials remain in the position they were when the injector failed (not returned to tray).

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1462  Product:ChemStation  ChemStation (G2070BA)  B.04.02 SP1

Keyword: 7890A Control

One-line Description:

7890A with no tray will display the Sampling Diagram

Problem:

The Sampling Diagram should not be displayed when there is not a tray installed.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1463  Product:ChemStation  ChemStation (G2070BA)  B.04.02 SP1

Keyword: 7890A Control

One-line Description:

7890A Extended Parameters enables Bar Code Reader section when BCR is not present

Problem:

The Extended Parameters for the native sequence table and Easy Sequence Setup (7890A) enable BCR section when that hardware does not exist. This behavior is benign.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1569  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 SP1

Keyword: 7890A Control

One-line Description:

Sequence aborts if vial not found with Overlap injection enabled (7890A only)

Problem:

7890A system using overlap injection feature

1. Define a sample overlap of 1 min.
2. Define a sequence of 5 runs
3. Insert 4 vials only

The 7890A GC starts the acquisition of Vial 4 correctly.  When it starts to prepare and pick up vial 5 which is missing, the acquisition of vial 4 is stopped and the sequence aborts.

When sample overlap is enabled and the vial to be prepared is not found, the currently running sample in the sequence will abort.  The sequence will stop and there will be errors displayed in the sequence logbook and GC status.

The error messages displayed are as follows:

- Command to Tray failed:  Tray Prep Failure Code: 250: Front injector has no 
  vail in turret...
- User stopped pretreatment program, finalizing.
- Instrument Error - Method/Sequence stopped
- Method aborted
- Sequence.S terminated due to an error
- Sequence.S completed

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:5  Product:78xx Driver Software  7890A  1.11

Keyword: 78xx Driver

One-line Description:

Error Message when certain methodsare loaded onto an offline configured 7890.

Problem:

If methods ISTD_EX.m or DEF_GC.M are loaded onto a configured 7890 offline session an error message (Exception in command "translateMethod$" occured).
These 2 methods are 6890 specific and should not be loaded onto a 7890.

Temporary Solution:

n/a

Fix Information:

Fixed in GC driver 4.01 [047].  

This driver was included in the following datasystem packages:
GC ChemStation B.04.03 SP1 [87] and OpenLab CDS C.01.02
GCMSD ChemStation E.02.02
OpenLab CDS EzChrom A.04.02


KPR#:15  Product:78xx Driver Software  7890A  1.10

Keyword: 78xx Driver

One-line Description:

Shutdown events are not displayed on the software Status screen

Problem:

Shutdown events, such as a thermal or pneumatic shutdown, displayed on the GC Keyboard are not reported to the software and therefore not displayed on the “Agilent GC Status” screen.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.01.  7890A requires a minimum Firmware revision of A.01.12.

This driver is included with the following datasystems:
GC ChemStationB.04.03 DSP1 [52] and OpenLAB C.01.02
GCMSD ChemStation E.02.02 
OpenLab CDS EzChrom A.04.02


KPR#:17  Product:78xx Driver Software  7890A  2.12

Keyword: 78xx Driver

One-line Description:

Barcode mismatch error

Problem:

When a barcode mismatch occurs with overlapped injections (option of do not inject on barcode mismatch), the mismatched vial is taken to the turret and chemstation stalls with the sequence running and waiting.  An abort is required to recover from this error.

Temporary Solution:

n/a

Fix Information:

Fixed in 78xx driver revision 4.02[023].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#:26  Product:78xx Driver Software  7890A  2.12

Keyword: 78xx Driver

One-line Description:

Sample Prep Program with only a 7693A Back Injector does not save

Problem:

The problem occurs when a 7693A Tray is installed and there is an injector in the Back and no injector in the Front. If a Sample Prep Program is created, the default values for a Move command are incorrect, and cause erratic behavior.

Temporary Solution:

See Service Notes 7693A-002 and 7693A-002A for available patch which is used with the 2.12.040 driver.

Fix Information:

n/a


KPR#:37  Product:78xx Driver Software  7890A  1.08

Keyword: 78xx Driver

One-line Description:

ALS Errors Abort Sequence; no Retry/Abort dialog box option offered

Problem:

The setting is selected by opening the Tray/Overlap tab in Method Editor.  

The Instrument Status must be opened to view the dialog box.

Pause is the default setting for ChemStation and MassHunter.
Skip is the default setting for EZChrom.

Temporary Solution:

n/a

Fix Information:

Fixed in the 5.01 Driver.
This driver is included in:
A.01.05 OpenLAB CDS
B.07.00 MassHunter


KPR#:54  Product:78xx Driver Software  7890A  2.13

Keyword: 78xx Driver

One-line Description:

Method edit: Must click twice with the mouse on each data field to change number

Problem:

In Method Edit of GC setpoints, the user has to click twice with the mouse on each data field to change a numerical entry.  The first click highlights the field, the second click highlights just the number, and then the number can be changed.

Temporary Solution:

n/a

Fix Information:

This problem is fixed in Software Driver Version: 3.01.

This driver is included in the following datasystem packages:
GC ChemStation B.04.02 SP1 [208]
Ezchrom Elite 332 SP2 
GCMSD ChemStation E.02.02


KPR#:67  Product:78xx Driver Software  7890A  2.12

Keyword: 78xx Driver

One-line Description:

ChemStation does not show "Hydrogen Safety Shutdown" in 7890A status

Problem:

When the 7890A goes into hydrogen safety shutdown the ChemStation shows "Front inlet flow shutdown" and other messages, but not  "Hydrogen Safety Shutdown".

The chemstation does not report the hydrogen shutdown but it does list the other causes. In addition, the heated zones can be turned on from the chemstation. Since a hydrogen shutdown should require operator intervention AT THE GC remote control from the (any) chemstation should be disabled.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.01.  7890A requires a minimum Firmware revision of A.01.12.

The 4.01 driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#:75  Product:78xx Driver Software  7890A  2.21

Keyword: 78xx Driver

One-line Description:

7890A valve names added in the config screen but are not available in the method

Problem:

The 7890A valve names can be added in the configure screen but the valve names are not available in the method.

Temporary Solution:

n/a

Fix Information:

This is fixed in GC driver version 3.01.

This driver is included in the following datasystem packages:
GC ChemStation B.04.02 SP1 [208]
Ezchrom Elite 332 SP2 
GCMSD ChemStation E.02.02


KPR#:76  Product:78xx Driver Software  7890A  2.13

Keyword: 78xx Driver

One-line Description:

7820 method printout includes parameters and setpoints unavailable for PP inlet

Problem:

Purged Packed Inlet includes parameters:
'Pressure  On  20.261 psi'
'Total Flow, On, 5.4077 mL/min'
'Septum Purge Flow On -0.0089673 mL/min'

Pressure, Total Flow, and Septum Purge Flow are parameters that are  not available for this GC.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.01.

This driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#:77  Product:78xx Driver Software  7890A  2.13

Keyword: 78xx Driver

One-line Description:

7820 method printout inserts negative number for Septum Purge Flow, invalid parameter

Problem:

7820 method printout for a Purged Packed (PP) inlet includes not only the invalid parameter 'Setptum Purge Flow', but inserts a negative number as its setpoint.

Temporary Solution:

n/a

Fix Information:

Fixed in Driver revision 3.02.

This driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.01
GCMSD ChemStation E.02.02


KPR#:80  Product:78xx Driver Software  7890A  2.12

Keyword: 78xx Driver

One-line Description:

If 7890 gets into an error status (flame off) this is NOT reported within the 7890 status window

Problem:

If the 7890 gets into an error status (such as flame off) this is NOT reported within the 7890 status window 

The flame failed to ignite at the FID but the status window ONLY 
displays that the GC is Not Ready.

Temporary Solution:

n/a

Fix Information:

Fixed in 78xx driver revision 4.02[009].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#:84  Product:78xx Driver Software  7890A  2.13

Keyword: 78xx Driver

One-line Description:

Read barcode just before injection and use barcode as sample name in report

Problem:

Users want to add vials to the tray as they go and have the vial bar code read just before injection and recorded as the name of the sample. 

Reading the vial ahead of time into the sequence table for each vial is time consuming.  Read the barcode just before injection since users are adding samples as they go.

This is working with the 6890 Classic driver.

Temporary Solution:

n/a

Fix Information:

This problem is fixed in Software Driver Version: 3.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.02 SP1 [208]
Ezchrom Elite 332 SP2 
GCMSD ChemStation E.02.02


KPR#:85  Product:78xx Driver Software  7890A  2.13

Keyword: 78xx Driver

One-line Description:

7890A: Column part number is not included in the method report

Problem:

The 7890A column part number is not printed in the method report (English, Japanese, Chinese).

Revision B.03.01 SR1.1 does include the column part number in the method report.  B.03.02, B.04.01, B.04.01 SP1, and B.04.02 do not include the part number of the column in the method report.

Temporary Solution:

n/a

Fix Information:

This problem is fixed in Software Driver Version: 3.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.02 SP1 [208]
Ezchrom Elite 332 SP2 
GCMSD ChemStation E.02.02


KPR#:90  Product:78xx Driver Software  7890A  1.05

Keyword: 78xx Driver

One-line Description:

Partial data collection is not available on the signals screen

Problem:

On the signals screen, need the ability to collect partial data for specified times.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.01 [31].
This driver version is included with the following data systems:
GC ChemStation B.04.03 DSP1 [54]
GCMSD ChemStation E.02.02
OpenLab CDS EzChrom A.04.02


KPR#:91  Product:78xx Driver Software  7890A  1.08

Keyword: 78xx Driver

One-line Description:

Need ability to do Digital switching for detector signals in Runtime events

Problem:

The ability to do digital switching of detector signals in the Runtime events table is needed.

It is only possible to do analog switching.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.01.  7890A requires a minimum Firmware revision of A.01.12.

Driver version 4.01 is included with the following data systems:
GC ChemStation B.04.03 DSP1 [52]
GCMSD ChemStation E.02.02
OpenLab CDS EzChrom A.04.02


KPR#:93  Product:78xx Driver Software  7890A  2.13

Keyword: 78xx Driver

One-line Description:

Read barcode just before injection and use barcode as the sample name in the report

Problem:

It is not possible to add vials to the tray and have the vial barcode read just before injection and recorded as the name of the sample. 

Reading the vial ahead of time into the sequence table for each vial is time consuming.  It is not possible to read the barcode just before injection and have the barcode used as the sample name in the report for 78xx.

Temporary Solution:

n/a

Fix Information:

Fixed in 78xx driver 3.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.02 SP1 [208]
Ezchrom Elite 332 SP2 
GCMSD ChemStation E.02.02


KPR#:98  Product:78xx Driver Software  7890A  2.13

Keyword: 78xx Driver

One-line Description:

Park during sequence with Overlap causes unresponsiveness

Problem:

Running ChemStation B.04.02.96. Start a sequence with overlapped injection set to immediately after injection. While the sample is being delivered to the turret, the Park button is pressed on the 7693A Tray. After delivering the vial, the tray is put in the Park position and the Injector completes the injection. At this point the ChemStation is unresponsive. (For example, when the ChemStation is in this state, any report printed to the screen can not be closed.)  

The ChemStation responds after the Park button is pushed and vial is returned.

Temporary Solution:

The ChemStation responds after the Park button is pushed and vial is returned.

Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#:100  Product:78xx Driver Software  7890A  3.01

Keyword: 78xx Driver

One-line Description:

7890GC: inlet pressure unit inconsistency - method audit trail vs instrument setup

Problem:

Problem with the Agilent 7890GC inlet pressure unit:

With method audit trail turned on, make changes to the inlet pressure (configured in psi in this case), save the method, check method audit trail log, it recorded the pressure in pascal.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 3.02.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
Ezchrom Elite 332 SP2 
GCMSD ChemStation E.02.02


KPR#:103  Product:78xx Driver Software  7890A  2.13

Keyword: 78xx Driver

One-line Description:

RTL: New calculated pressure doesn’t update column 2 - 6 on current 7890A methods

Problem:

RTL: New calculated pressure doesn’t update column 2 - 6 on current 7890A methods.

Only Column 1 is updated

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.02[034].

This driver is included with the following datasystems:
GC OpenLab CDS C.01.01
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.04


KPR#:115  Product:78xx Driver Software  7890A  3.01

Keyword: 78xx Driver

One-line Description:

7890A: Shutdown events are not published to the software GC Status screen

Problem:

The thermal inlet shutdown is not reported in the software “Agilent GC Status” screen.  

Only the first event, which is 'not ready', is reported to the Agilent GC Status screen.  The second event, which is the shutdown, is not displayed.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.01.  7890A requires a minimum Firmware revision of A.01.12.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#:117  Product:78xx Driver Software  7890A  2.10

Keyword: 78xx Driver

One-line Description:

Upgrading B.01.03 to B.04.02 GC ChemStation:  Addition of new columns to the column catalog (database) is not possible; error occurs

Problem:

Upon upgrading from B.01.03 to B.04.02 it is no longer possible to add new columns to the GC column catalog (database).

The following error is shown:

The Catalog File cannot be opened.  If needed, you may obtain an appropriate replacement file from Agilent.
C:\Chem32\CORE\...\DRIVERS\catalog.mdb

Temporary Solution:

The problem is that there are several columns that have the same manufacturer and model number. Our new database software doesn’t handle this.

The newer database software (in B.04.02 for example) no longer allows duplicates of manufacturer + model number. i.e., there can only be one HP, HP-5 in the database.  Note: the new software will inform the user that the model number “xxx” is already in use and request that a different value be chosen. (The user is returned to the original entry screen with all the info filled in.) 

Temporary Fix:

Before upgrading, fix the problem with the column database duplicates while running B.01.03 for example.  Review the column catalogs and make sure there are no duplicates and fix any that exist before the upgrade.

Fix Information:

This is fixed in GC Driver version 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#:118  Product:78xx Driver Software  7890A  0.0.1.2

Keyword: 78xx Driver

One-line Description:

Electrometer on/off not needed for FID or NPD

Problem:

The software shows an on/off checkbox for the electrometer on both the FID and NPD. This setting is not available in the normal local user screens in 7890A. The checkbox should be removed from the edit parameters screen.

Temporary Solution:

n/a

Fix Information:

This is fixed in driver version 3.01 which started to be included in the following data system packages:
GC ChemStation B.04.02 Driver Service Pack 3 (DSP3)  [256]
Ezchrom Elite 332 SP2 and OpenLab EZChrom
GCMSD ChemStation E.02.02


KPR#:124  Product:78xx Driver Software  7890A  0.0.1.5

Keyword: 78xx Driver

One-line Description:

Freeze Signal Value and Resume Signal Value missing from runtime events

Problem:

Freeze Signal Value and Resume Signal Value missing from runtime events

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.01.  7890A requires a minimum Firmware revision of A.01.12.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#:132  Product:78xx Driver Software  7890A  3.01

Keyword: 78xx Driver

One-line Description:

Error in flow setpoint upload when method has more than 100 events

Problem:

When  a method with more than 100 events is uploaded from the GC, all the steps above the 100th event that are related to flow (ml/min) are multiplied by a factor 1000.

Temporary Solution:

n/a

Fix Information:

Fixed in 7890A firmware A.01.11.  No software update is necessary.


KPR#:134  Product:78xx Driver Software  7890A  3.02

Keyword: 78xx Driver

One-line Description:

Running a sequence requires the arm from the tray, G4514A to be in the UN-PARKED position

Problem:

The tray "Park" status overrides the start of a sequence.

Temporary Solution:

n/a

Fix Information:

The G4514A arm from the tray must be UN-PARKED in order for a sequence to proceed. User must press Park button to UN-PARK arm.


KPR#:137  Product:78xx Driver Software  7890A  2.20

Keyword: 78xx Driver

One-line Description:

Extended Parameters does not disable the BCR when no hardware is present

Problem:

Extended Parameters for both the native sequence Parameters and for Easy Sequence Setup does not disable the Bar Code Reader section when no such hardware exists.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#:139  Product:78xx Driver Software  7890A  4.01

Keyword: 78xx Driver

One-line Description:

Switched and standard mode for septum purge not available on MMI

Problem:

For the split/splitless inlet, it is possible to run the septum purge in either standard mode or "switched" mode. In the switched mode, the septum purge is turned off during the injection cycle so that the only flow out of the inlet is through the column. The MMI would benefit from having this mode as well, particularly for splitless and solvent vent modes. In both modes on the MMI, the septum purge does not remain off until the inlet switches back to the purge state.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#:140  Product:78xx Driver Software  7890A  3.01

Keyword: 78xx Driver

One-line Description:

Negative Split Splitless inlet Pressure values during Post Run Backflush

Problem:

When using a 7890A GC with Split Splitless inlet to perform backflush the Inlet Setpoint and Actual Pressure assume negative values on the GC and Chemstation display during backflush. This occurs with a splitter (>2 columns) or with a purged union (Primary column plus restrictor). The columns must be running in "Flow" mode. 

A similar symptom is the inlet pressure gets set to "0.00" PSI during backflush rather than the wizard recommended value of "1.0" PSI.

Temporary Solution:

The workaround is to set the columns to "pressure" mode then back to flow and re-save the method.

Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#:154  Product:78xx Driver Software  7890A  1.02

Keyword: 78xx Driver

One-line Description:

Column compensation menu is available during the sequence

Problem:

The Column compensation menu is available during the sequence and if selected will stop the sequence.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#:166  Product:78xx Driver Software  7890A  1.08

Keyword: 78xx Driver

One-line Description:

Column:  need to include Ramped and Constant pressure and flow modes as selection

Problem:

User interface needs Constant Flow, Constant Pressure, Ramped Pressure, and Constant Flow for the Control Mode selections.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.01.

This driver was included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
GCMSD ChemStation E.02.02
OpenLab CDS EzChrom A.04.02


KPR#:167  Product:78xx Driver Software  7890A  1.10

Keyword: 78xx Driver

One-line Description:

GC goes offline momentarily because of SOAP Exception

Problem:

Setup 7693 injector and Tray. Put Tray in Park and then tried to start a run. The ChemStation showed that the GC went offline momentariily. The Event Log showed the following events :

Disconnecting because a System.Web.Services.Protocols.SoapException was caught when sending a command; 29: GC refused to do an injection
The connection state of the GC at 130.30.245.60 is now Offline
Attempting to connect to 130.30.245.60
In DataChannel[1].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000
In DataChannel[2].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000
In DataChannel[3].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000
In DataChannel[3].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000
The connection state of the GC at IP address is now Online
Connected to 7890 at 130.30.245.60, using license key 

Cannot start runs until GC and injector are idle.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#:170  Product:78xx Driver Software  7890A  1.02

Keyword: 78xx Driver

One-line Description:

No indication of inlet pressure shutdown in software status with 7890A

Problem:

Inlet pressure shutdown advisories indicating a problem with the instrument not shown in software status user interface.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.01.  7890A requires a minimum Firmware revision of A.01.12.

This driver is included with the following datasystems:
GC ChemStation B.04.03 DSP1 [52] and OpenLab CDS C.01.02
GCMSD ChemStation E.02.02
OpenLab CDS EzChrom A.04.02


KPR#:174  Product:78xx Driver Software  7890A  1.09

Keyword: 78xx Driver

One-line Description:

GC Keypad Lock Behavior

Problem:

In configuration we  now offer the option to configure how the keypad lock will behave.

Temporary Solution:

n/a

Fix Information:

The driver with this new behavior is included in:

A.01.05 OpenLAB CDS
B.07.00 MassHunter


KPR#:182  Product:78xx Driver Software  7890A  4.01

Keyword: 78xx Driver

One-line Description:

GC Instrument Status for Detectors

Problem:

In previous versions of software, only the temperature of the detectors was shown in instrument status.  In driver version 5.01, we show all GC detector status in instrument status.

Temporary Solution:

n/a

Fix Information:

Fixed in the 5.01 Driver.
This driver is included in:
A.01.05 OpenLAB CDS
B.07.00 MassHunter


KPR#:190  Product:78xx Driver Software  7890A  1.05

Keyword: 78xx Driver

One-line Description:

FID Flame out; no feedback in software

Problem:

The GC status and instrument actuals list the signal value as 0.0 but does not provide feedback if the flame is out.

Temporary Solution:

n/a

Fix Information:

Fixed in 78xx driver revision 4.02[001]

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#:195  Product:78xx Driver Software  7890A  4.01

Keyword: 78xx Driver

One-line Description:

uECD and TCD report printout for makeup flows is wrong

Problem:

7890A has a uECD detector with flows configured.
With the following method settings...
    Makeup Flow(N2)               :  30 mL/min
    Constant Column + Makeup : 31 mL/min

... the method printout reads:
    Makeup Flow(N2)               :  31 mL/min
    Constant Column + Makeup :    0 mL/min

Temporary Solution:

n/a

Fix Information:

Fixed in 7890A SW Driver 4.01 [048].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#:196  Product:78xx Driver Software  7890A  4.01

Keyword: 78xx Driver

One-line Description:

Control of PCM used as inlet EPC does not work in B.04.03 [52]

Problem:

If PCM is configured with the inlet heater as PCM A or PCM B controlling the column flows/pressures, the temperature and pressure/flow setpoints cannot be set. CS shows a septum purge value and the instrument goes into shutdown, due to "septum purge not ready".

Temporary Solution:

n/a

Fix Information:

Fixed in 78xx driver revision 4.02[007].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#:200  Product:78xx Driver Software  7820A  4.01

Keyword: 78xx Driver

One-line Description:

GC continues when STOP on 7820 is pressed

Problem:

User makes a manual injection and waits until the peaks of interest elute, then hits the STOP on the 7820 front panel the GC stops (and resets the initial GC settings) but the chemstation continues and reports the system is running data analysis but does not go into data analysis. System appears to hang up waiting on the expected run time and report is generated at that point rather than directly after STOP is pressed.

Temporary Solution:

Use STOP run from the software.

Fix Information:

Fixed in driver revision 4.02[11].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#:203  Product:78xx Driver Software  7890A  4.01

Keyword: 78xx Driver

One-line Description:

Open GC edit parameters and close it without changes; method modification flag is set

Problem:

Saved a method.  Open the GC edit parameters to review a setpoint and  made no changes or changed the view.  Selected OK and the method modification flag indicates a change was made.  No changes were made.

Temporary Solution:

n/a

Fix Information:

Fixed in 78xx driver revision 4.02[019].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#:204  Product:78xx Driver Software  7890A  3.02

Keyword: 78xx Driver

One-line Description:

7890A logbook shows 7693A turret, plunger errors as GCI1 did not start a run

Problem:

7693A ALS errors, e.g. turret, plunger, door open, etc. show a non-specific logbook entry: GCI1 did not start a run.

Temporary Solution:

n/a

Fix Information:

Fixed in 78xx driver revision 4.02[023].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#:209  Product:78xx Driver Software  7890A  4.01

Keyword: 78xx Driver

One-line Description:

Unable to set equilibration time for LTM columns from Edit Parameters

Problem:

The LTM column's equilibration time is shown as 0. Change the equilibration time by entering 0.5 and downloading the method. This does not change the value at the GC.

Temporary Solution:

n/a

Fix Information:

Fixed in 78xx driver revision 4.02[005] and GC Firmware revision A.01.13.

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#:211  Product:78xx Driver Software  7890A  4.02

Keyword: 78xx Driver

One-line Description:

Changes made to 7890 method from GC instrument diagram result do not load.

Problem:

Open the instrument parameter from GC Instrument Diagram and confirm sample prep is off. Set inlet heater - save changes. Open the instrument parameter from the GC Instrument Diagram and confirm sample prep is off. Change the inlet heater temp and save as a new method. Load the new method  - a configuration mismatch message is displayed.  

Requires Easy Sample Prep

Temporary Solution:

n/a

Fix Information:

Fixed in 78xx driver revision 4.02[004].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#:220  Product:78xx Driver Software  7890A  4.02

Keyword: 78xx Driver

One-line Description:

Modifying the Syringe Size When Using a 7890A GC

Problem:

After modifying and saving the method the injector screen will not accept the syringe size of 10 uL.  The menu will pull down and 10uL is accepted but when the screen is OK'ed the size changes to blank.

Temporary Solution:

Enter the syringe size from the GC front panel, upload the configuration and save the method.

Fix Information:

Fixed in 78xx driver revision 4.02[019].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#:223  Product:78xx Driver Software  7890A  0.0.1.2

Keyword: 78xx Driver

One-line Description:

Backflush icon is not displayed when the outlet of the column configuration is set as "Other".

Problem:

When the outlet of the column configuration is set as "Other", there is no backflush icon displayed in the 7890 Edit Parameter Screen.

Temporary Solution:

n/a

Fix Information:

Fixed in 78xx driver revision 4.02[009].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03

Please note, when the configuration is changed from backflush not available to backflush available, the user must exit the method editor and begin the edit a second time in order for the backflush icon to be shown.


KPR#:227  Product:78xx Driver Software  7890A  4.02

Keyword: 78xx Driver

One-line Description:

When running Backflush from ChemStation with segmented columns only the in-segment is taken into account in the calculation.

Problem:

If you run Backflush from Chemstation with segmented columns (usually associated with LTM), only the in-segment portion of the configuration is used in the backflush flow rate and void volume calculations. This leaves the user unable to calculate the backflush time.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.02[023].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#:228  Product:78xx Driver Software  7890A  4.01

Keyword: 78xx Driver

One-line Description:

The display of the parameter setting screen of the method setup is small.

Problem:

In the Edit Parameter screen for the oven and the inlets, the display of this screen is small. Adjusting the method setup screen will not resize this screen.

Temporary Solution:

Right click on  parameter setting screen, and go to "show graphical plots". The display will resize to a normal size.

Fix Information:

Fixed in 78xx driver revision 4.02[006].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#:229  Product:78xx Driver Software  7890A  4.01

Keyword: 78xx Driver

One-line Description:

After editing the GC method and selecting ok - validation error message is displayed.

Problem:

In driver revision 4.01, after editing the GC method and selecting ok, a "Validation Error" error message flashes in the left hand corner of the screen. User must select cancel to exit the GC parameter screen. 

No method resolution occurs when loading method
Software loads fine and can make runs. Methods cannot be edited in online or offline mode.  

The configuration includes 4 packed columns that flow to the Back TCD Detector. Column 1's inlet is PCM C-2 (@ const pressure); 4's inlet is Aux EPC 5 (@ const pressure); and both 2's and 3's inlet is PCM C-1. Column 2's mode is constant flow; and 3, since it is also connected to PCM C-1, is reported as being controlled by #2.

Temporary Solution:

Workaround for customers with driver revision 4.01:  replace columns #2 and #3 with a single undefined (e.g., packed) column, suitably described as being a parallel combination. The user can then set either a flow or a pressure setpoint to the new combined column #2.

Fix Information:

Fixed in driver revision 4.02.  

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03
OpenLab CDS ChemStation C.01.04


KPR#:230  Product:78xx Driver Software  7890A  4.02

Keyword: 78xx Driver

One-line Description:

7820 Back Injector Only Resets Injection Source to Manual

Problem:

On a 7820, with only a Back Injector Installed the following message is displayed when launching the 7820: 

Upon launching the 7820A, the following message appears:

DEF_GC.M: 7820
Resetting to Manual injection.
For method 'DEF_GC.M, injection source 'GC Injector' conflicts with injection location.

This does not happen when there is only a Front Injector installed.

Temporary Solution:

Open the injection source and change it to GC Injector.

Fix Information:

Fixed in the 5.01 Driver.
This driver is included in:
A.01.05 OpenLAB CDS
B.07.00 MassHunter


KPR#:238  Product:78xx Driver Software  7890A  4.02

Keyword: 78xx Driver

One-line Description:

PCM Inlet Heater Shutdown

Problem:

7890 GC with PCM controlled inlet heater. Open Edit Parameters, enable PCM inlet heater checkbox and make a change to the heater settings. Download the method to the GC or hit apply and the inlet heater is turned OFF at the GC. If the user closes edit parameters, and then re-opens edit parameters the PCM inlet heater checkbox is disabled.

Temporary Solution:

Create your method, save, and download to the GC (the PCM inlet heater will shutdown). At the GC, set the PCM inlet heater to the desired setting. Upload the method from the GC (Instrument, Upload Method), and save your method as a new method (Save method as …). This method can now be used and loaded successfully in the future.

The workaround will need to be performed any time changes are made to the method.

Fix Information:

This driver is included with the following datasystems:
OpenLab GC ChemStation C.01.04
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.04


KPR#:247  Product:78xx Driver Software  7890A  4.02

Keyword: 78xx Driver

One-line Description:

Multiposition valve resets to position # 1 between injections

Problem:

The set position of the gas streams used for a multiposition valve is controlled in the sequence. The user can run from different gas streams by entering the different numbers in the sequence table. The position is not stored as part of the method. The MPV set position is always reset to start from position #1 between runs.

Temporary Solution:

n/a

Fix Information:

This is fixed in 7890A firmware revision A.01.14.  No software update is required.


KPR#:252  Product:78xx Driver Software  7890A  4.02

Keyword: 78xx Driver

One-line Description:

MMI in Split or Pulsed Split Mode does not allow editing of Total Flow value

Problem:

MMI in Split or Pulsed Split Mode does not allow editing of Total Flow value. The Total Flow combo box is active, but any entry made there reverts to its original entry.

Temporary Solution:

Total Flow is editable in this mode from the GC keypad in driver versions prior to 5.01.

Fix Information:

Fixed in the 5.01 Driver.
This driver is included in:
A.01.05 OpenLAB CDS
B.07.00 MassHunter


KPR#:269  Product:78xx Driver Software  7890A  5.01

Keyword: 78xx Driver

One-line Description:

ALS Behavior When Aborting a Sequence

Problem:

If a sequence is aborted in the middle of a sequence, the vial will be left in the turret until the next run is started.  At that time, it will be put back in the tray.

Temporary Solution:

This is expected behavior.

Fix Information:

n/a


KPR#:277  Product:78xx Driver Software  7890A  3.01

Keyword: 78xx Driver

One-line Description:

Logging of Injection Tower Location

Problem:

When the injector tower is moved from front to back position (or the opposite way) the method and sequence run correctly.  However, on the report there is still the original (back) position printed, the change of the position is not tracked.

Temporary Solution:

Recommended steps:  Move the injector to the new location, click "Don't resolve" (DO NOT RESOLVE THE METHOD) and then load the method for that injector position.

Fix Information:

n/a


KPR#:282  Product:78xx Driver Software  7890B  5.01

Keyword: 78xx Driver

One-line Description:

Instrument Actuals Show "GC is Offline" Quickly at the Method or Sequence Start, then Returns Back Online and the Method Does Not Run

Problem:

Instrument actuals show that the "GC is Offline" quickly at the start of the method and/or sequence and then returns back online.  The method and/or sequence does not run.  This is likely caused by a difference in the method parameters and what is loaded on the GC.

Temporary Solution:

n/a

Fix Information:

Open a brand new Def.M and create a new method and/or sequence.  Download the method to the GC from the software.


KPR#:348  Product:78xx Driver Software  7890B  6.01

Keyword: 78xx Driver

One-line Description:

B.01.01 driver on Multi-Technique ChemStation (MTCS) B.04.03.105 SP2 fails IQT: 76 Invalid files

Problem:

MTCS B.04.03.105 uses an older version of the IQT/SVT Tool than our modern data systems & drivers.  As a result, it was agreed that the latest GC Drivers installer would not be IQT compatible with MTCS B.04.03.105.

Temporary Solution:

None.

Fix Information:

n/a


KPR#:357  Product:78xx Driver Software  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

Inlet pressure setpoint displayed assuming all thermal zones are turned ON, results in confusing mismatch when heated zones are Off.

Problem:

CDS displays inlet pressure setpoint assuming all thermal zones are turned ON, results in confusing mismatch when heated zones are Off.

Temporary Solution:

none

Fix Information:

To be resolved in B.01.02


KPR#:358  Product:78xx Driver Software  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

IQT report failed on MT ChemStation B.04.03 SP2 with GC Driver B.01.01

Problem:

IQT report failed on MT ChemStation B.04.03 SP2 with GC Driver B.01.01. There is a technology change between from IQT to SVT that causes this to happen..

Temporary Solution:

None

Fix Information:

n/a


KPR#:360  Product:78xx Driver Software  7890B  B.01.01

Keyword: 78xx Driver

One-line Description:

Method Editor for PP inlet & Packed Column incorrectly allows both flow and pressure setpoints to be entered

Problem:

When in flow mode, pressure should not be allowed. Entering a value messes up the flow setpoint.
When in pressure mode, flow should not be allowed. Entering a value messes up the pressure setpoint.

Temporary Solution:

Workaround is to only enter the appropriate setpoint.

Fix Information:

To be resolved in B.01.02


KPR#:1196  Product:78xx Driver Software  7890A  3.01

Keyword: 78xx Driver

One-line Description:

7890A: Shutdown events are not published to the software GC Status screen

Problem:

The thermal inlet shutdown is not reported in the software “Agilent GC Status” screen.  

Only the first event, which is 'not ready', is reported to the Agilent GC Status screen.  The second event, which is the shutdown, is not displayed.

Temporary Solution:

n/a

Fix Information:

Fixed in driver revision 4.01.  7890A requires a minimum Firmware revision of A.01.12.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#:12  Product:ChemStation  ChemStation (G2070BA)  A.09.03

Keyword: ACQ

One-line Description:

Time stamp errors in LC Chemstation & Chemstore, Win 2000 time format ignored.

Problem:

LC  ChemStation A0903  ignore the Win 2000 time format HH:mm:ss and will put always a PM stamp on the injection time for injections done in the morning onto the report.  The rawdata information will show the correct injection time.

LC Data spooled to Chemstore will use the right time stamp and will give the correct injection, processing & transfer times.

Temporary Solution:

Remove the PM symbol in Start>Settings>Control panel> Regional Options, Time Tab.
All injections times for  LC  runs are correctly shown  now without the PM stamp.

Fix Information:

 No fix Planned


KPR#:299  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: ACQ

One-line Description:

error "no string 8008" in sample information

Problem:

In the Sample Information dialog it is possible to specify the subdirectory of the DATA folder where the data file of a single run is to be saved in. In case the DATA directory contains a file without any extension (like e.g. .txt) the name of this file cannot be used for the subdirectory. An error message "no string 8008" is displayed.
This situation might occur e.g. when you ran sequence before with Sequence Summary Report and for the name of the report specified a name without file extension.

Temporary Solution:

n/a

Fix Information:

No fix is planned for this issue.


KPR#:308  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: ACQ

One-line Description:

'Active signals" comment is chopped in MSD TIC title.

Problem:

In "Setup MSD Signals" panel, set the comment for Active signals (e.g for signal 1 as "Scan #1a").
On running a method or sequence this label is seen incompletely as the MSD TIC title.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:375  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: ACQ

One-line Description:

Agilent Active Splitter will sometimes not start up correctly after installation

Problem:

The following problem was discovered in the B.01.03 Agilent Active Splitter software.

The characteristics of the problem are that when the user starts up the Agilent Active Splitter using the Active Splitter software, it will sometimes cause the Active Splitter to move about 20 times, then it will stop and not communicate with the program. Other times, the Active Splitter will not move at all when it is started with the software.

This problem usually occurs upon start up, although sometimes the Active Splitter will function normally for a while. The Agilent Active Splitter program will often put up error messages like "START Timeout after 15 retries."

Temporary Solution:

n/a

Fix Information:

An updated program is available which fixes this problem.  It is described in Service Note G2710-55A.


KPR#:448  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: ACQ

One-line Description:

Snapshot is not created if method or sequence name is of 40 characters long

Problem:

If you run a method with a name of 40 characters or a sequence with a name of 40 characters, it is not possible to create a snapshot during the run.

Temporary Solution:

Use method and sequence names with less than 40 characters if you want to create snapshots.

Fix Information:

n/a


KPR#:491  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: ACQ

One-line Description:

FIA runs with Dual Loop Autosampler are not reproducible

Problem:

When a customer does a six injection FIA series with a Dual Loop Autosampler (DLA) the first two injections are of much lower abundance than the other four.  If he does a sequence of six injections with the DLA and no FIA, the injections are all of the same height.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:492  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: ACQ

One-line Description:

Acquiring multi-signal data with low SIM Dwell times causes erroneous data.

Problem:

If you run a multisignal acquisition with one Scan and three SIM signals under conditions that create a very low (< 20 msec) Dwell time, then the highest m/z SIM ion will have an much greater response than it should.  This creates errroneous, artificial data which could mislead the customer.  On the customer methods that were provided to us, she had the MSD Acquisition Peakwidth set to 0.10 and the cycle times for the SIM ions set to 5% each.  This resulted in a Dwell time of 1 msec for signal MSD4.

Changing the parameters (Peakwidth 0.20, Cycle time 25%) makes this problem go away--the highest m/z ion does not have an artificially high response.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:777  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: ACQ

One-line Description:

Main root path as data path in Preferences does not work for data acquisition

Problem:

If a main root path (e.g. d:\) is added as a data path on the Paths tab of the Preferences dialog, any acquired data will not be saved to this path even when it is selected in the Sequence Parameters or Sample Information.
Any other paths work fine.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1316  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.01

Keyword: ACQ

One-line Description:

ChemStation switches to default path without warning if data path not available

Problem:

If in addition to the default data path, additional data paths have been specified in the Preferences one of the additional paths can be selected in the sequence template as data path for acquisition.
In case the selected data path is not available when the sequence is started (e.g. network problems in case of a network path), the data will be acquired to the default path without any warning to the user.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1321  Product:ChemStation  ChemStation (G2070BA)  B.04.02 Pre-Release

Keyword: ACQ

One-line Description:

sporadically the next run in a sequences will not start

Problem:

In ChemStation B.04.02 (pre-release version) it is sporadically observed that sequences do not continue with the next run.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1564  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: ACQ

One-line Description:

Windows menu not responding after running ChemStation for long time

Problem:

When running ChemStation for a long time (several days) while Windows Explorer is open and observing any of the instrument directories in the \chem32 directory, it may happen that the Windows menu is no longer responding.

Temporary Solution:

Close and restart Windows Explorer. Do not keep Windows Explorer open for longer time for observing the ChemStation instrument directories.

Fix Information:

n/a


KPR#:1641  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: ACQ

One-line Description:

Not possible to modify data analysis parameters of method during a single run in the offline; cannot save

Problem:

It is not possible to change integration parameters in the offline for a single run and save the new parameters in the method file.

Temporary Solution:

Data analysis parameters can be updated and saved if using a sequence.  It is recommended to prepare a single run sequence.

Fix Information:

 Fixed in revision C.01.04


KPR#:1660  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03 LC DSP

Keyword: ACQ

One-line Description:

No special characters should be used for names of methods, data files or sequences

Problem:

Names for data files, methods and sequences should only contain alphanumeric characters. Special characters like ",#,$, %, &, @, ? and others should not be used and can in general cause trouble. The same is true for trailing blanks.
While this is outlined in the written documentation several times, it is not enforced for all dialogs inside the software. There are still the occasional loop hole where such a character can be entered or imported.

Temporary Solution:

The possiblity of entering special characters does by no means imply that it is safe to do it.

Fix Information:

 See Previous Note.


KPR#:1701  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: ACQ

One-line Description:

Terminal  Service can close under high workload

Problem:

It has been observed that the terminal service on an AIC can terminate under high workload.
This is a limitation of Windows 2008 server. Since this only affects the remote session control to the ChemStation Client, no data will be lost.

Temporary Solution:

No fix available.

Fix Information:

Will not be fixed for Windows 2008.


KPR#:1732  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: ACQ

One-line Description:

Running dual tower sequence with all methods saved as Dual, get “Injector doesn’t match the sequence’s DUAL Injector”

Problem:

1. Create seq table, with one run (Method A) for the front table, one run (Method B) for the back table this must be done first
2. Before line one of the sequence completes, add a run2 to the back table only (BACK.M) by append
3. When the back run is in process, add a sequence line to the front table, use a different method (method A) 
4. Message stating that Injector doesn’t match the sequence’s  DUAL Injector occurs even though all methods are saved with Dual injection selected as the injection source

Temporary Solution:

n/a

Fix Information:

Fixed in Revision C.01.04.

Point patch available for B.04.03 SP1.


KPR#:1742  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: ACQ

One-line Description:

Changes in sequence methods made during acquisition are not saved automatically

Problem:

Run a sequence and change the method during the acquisition (e.g. change the runtime). There is no possibility to save the method. The method should get saved at the end of the run automatically, but it does not. This issue occurs in C.01.01 - C.01.03 versions.

Temporary Solution:

n/a

Fix Information:

Resolved in Revision C.01.04.


KPR#:1747  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: ACQ

One-line Description:

Queue Method vial range reflects range of loaded method, not the method browsed to

Problem:

Load a method in ChemStation. Select Queue Method. Within that dialog, Browse... to load another method with a different vial range. The vial range (Sample Parameters: Vial/Location) is not updated to reflect the range of the method loaded in Queue Method, though the Signal (1 and/or 2) Data File info is correctly loaded from the browsed to method.

Temporary Solution:

n/a

Fix Information:

Resolved in C.01.07 SR2


KPR#:1841  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: ACQ

One-line Description:

Run time for unsynchronized modules is shortened

Problem:

The runtimes of modules on a CAN bus (1260, 7100, integrated LCs) are handled internally to synchronize start and stop. The same applies to devices attached to the remote bus (e.g. 35900 or MSD). 
Devices that are not synchronized might be either stopped too early which leads to data loss, or too late which causes time-out errors. This affects ELSD detectors, SFC modules, modules with GCI drivers or 3rd party modules with RC.Net drivers, e.g Waters Alliance modules.

Temporary Solution:

Make sure the runtimes of all modules producing data are identical and no other module has a longer runtime.

Fix Information:

Fixed in C.01.05 and C.01.04 HF04 by introducing a master runtime for all modules.


KPR#:1842  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: ACQ

One-line Description:

Memory leak in RDPShell.exe may cause significant performance decrease on OpenLAB CDS ChemStation AICs

Problem:

The following problem has been reported for OpenLAB CDS ChemStation C.01.0x Acquisition Instrument Controllers (AIC) operated on Windows 2008 R2 SP1 operating systems. Windows 7 or Windows XP operating systems are not affected.

A memory leak in RDPShell.exe may cause significant performance decrease as the overall memory usage may exceed 80% after several weeks of operation. This unintended and permanent memory allocation is triggered by moving the ChemStation application window over the client desktop. 

This problem is acknowledged by Microsoft (http://support.microsoft.com/kb/2636613/en-us) and hot fix is available on request. 
NOTE:  The hot fix is not deployed as part of the automated update procedure.

Temporary Solution:

Unpatched AIC machines should be rebooted every 2nd week to release the allocated memory. Note: ChemStation instances need to be shutdown (rather than disconnected) before rebooting the AIC.
Use the remote application window in full screen mode only and avoid moving the application window on the client screen.

Shutting down the ChemStation instances does not release the memory on the Windows 2008 server as the RDP sessions remain in an idle state untill they are re-used.

Fix Information:

Microsoft offers a hot fix that is not autmatically distributed by the update process.  The patch needs to be explicetly requested from http://support.microsoft.com/kb/2636613/en-us.
Please follow the back-up and installation recommendations from Microsoft.

Agilent performed limited tests with the mentioned hot fix and no negative impact was found.


KPR#:1960  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.06

Keyword: ACQ

One-line Description:

Chromatograms of manually stopped runs are shorter than expected

Problem:

User sometimes stop runs as soon as the expected peaks have eluted as seen from the online plot.
The run on disk might not contain those last peaks.
The online plot uses a different channel which is unfiltered. Stopping a run will finalize the chromatogram immediately. When the detector has a high filter width (response time), the last data points are not yet stored to the chromatogram.

Temporary Solution:

With low data rates (and long filter widths) the online plot might be up to 30 seconds ahead of the data collection in the data file. Consider adding this time span before stopping a run.

Fix Information:

N/A


KPR#:1964  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 DSP 2 (FireAnt)

Keyword: ACQ

One-line Description:

Data system uses the currently loaded method to validate the vial range in the Sequence table

Problem:

Sequence does not run or will abort if the currently loaded method vial range is different than the vail range of the next method in the sequence table.

Set up two methods:
Method E with injection source set to headspace (valid vials 1-111)
Method A with injection source set to ALS sampler with tray. (valid vials 1-150)

Construct a sequence
Vial      Method
1            A
1            E
130        A
2            E

If method E is the currently loaded active method then this sequence will not validate because the first method in the sequence (method A) has a vial range that is valid for an ALS.  The vial numbers are checked against method E which has a vial range for the headspace. The sequence cannot be run.

If method A is the currently loaded active method, then this sequence will validate and start running. When the datasystem attempts to move to the 3rd line it validates the vial against E with a headspace injectorID and fails, aborting the sequence.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB CDS C.01.06.


KPR#:1122  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Audit Trail

One-line Description:

report preview or printout from Report Layout does not trigger report history

Problem:

Creating a report preview or printout from Report Layout view does not trigger an entry in the Report History.

Temporary Solution:

n/a

Fix Information:

It is currently not intended to change this behavior.


KPR#:1133  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Audit Trail

One-line Description:

deletion of compound group not tracked in method audit trail

Problem:

When a compound group is deleted by deleting all the compounds in the group, the deletion of the compounds is tracked in the method audit trail, but the removal of the compound group is not tracked.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:1749  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Audit Trail

One-line Description:

Method Audit Trail does not include change of Injection source when HS/PAL is no longer configured

Problem:

GC with HS or PAL as the injection source. Close the session and remove the external device. Open the session and select Compare. No differences are seen in the method.

Download the original method. The  injection source has been changed to Manual. No change is logged to the Method Audit Trail.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1819  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Audit Trail

One-line Description:

UIB II - set a timetable and control "Method Audit Trail"

Problem:

If a timetable of UIB II  is set and stored via "Save Method as..", the time of the change should be stored in the "Audit Trail".
But the time of the method change is not correct.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1861  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 SP1

Keyword: Batch

One-line Description:

Cannot open batch - specified range parameter is bad

Problem:

An error occurs when loading newly modified and saved batch files.

During a sequence a batch file is created. 

To reproduce the error:

Once the sequence has completed, go the Data Analysis view and, select Batch > Load Batch and select the batch file .B (select all files).  Make a change to one of the parameters (modify integration, multiplier, or dilution, etc.) and print a report.  Save this modified batch as a new .B file in the result set folder.  The original batch files loads without error.  But the modified .B file will not load and gives an "Invalid Parameter Range" error message.

The original batch file will load without error, only the newly modified and saved batch file will not load.

Temporary Solution:

Resolved in C.01.04 Hotfix06 and C.01.05 Hotfix04.

The Fix will prevent the problem from appearing after installation. Existing batches will still be incomplete and cannot be repaired.

Fix Information:

Resolved in C.01.06.

The resolution will prevent the problem from appearing for new batches. Existing batches will still be incomplete and cannot be repaired.


KPR#:9  Product:ChemStation  ChemStation (G2070BA)  A.09.03

Keyword: Calibration

One-line Description:

Calibrated compound found as uncalibrated one using library search report

Problem:

In case a multi-wavelength calibration table is setup to calculate signals separately, it can happen that a calibrated compound appears in the listing of uncalibrated compounds of a library search report, while it is not listed in the calibrated compounds part.

Temporary Solution:

The problem can be avoided either by
- not calculating signals separately when using a library search report
or
- calibrating all signals where a compound actually shows a response

Fix Information:

 No fix Planned


KPR#:889  Product:ChemStation  ChemStation (G2070BA)  B.03.01 SR1

Keyword: Calibration

One-line Description:

CE ChemStation:  Update calibration level ignores area correction

Problem:

The interactive calibration level update function of ChemStation does not work for CE calibration tables using corrected areas.
The real area is added to the calibration table, instead.

Temporary Solution:

Delete the level and use the interactive 'Add level' function. This includes the area correction.

Fix Information:

n/a


KPR#:1089  Product:OpenLAB CDS ChemStation Edition  M83xx  B.01.01

Keyword: Calibration

One-line Description:

Run a sequence with recalibration, only the first ISTD is updated

Problem:

The ASTM method for biodiesel uses two internal standards in the calibration table, but when you run a sequence with recalibration only the first ISTD gets updated.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB CDS C.01.06. Up to 8 ISTD compounds can be updated during one run.


KPR#:1766  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Calibration

One-line Description:

Calibration method printout does not print compound level's Heights when quantitation is based on height

Problem:

Calibration method printout does not print compound level's Heights when quantitation is based on height

Temporary Solution:

Available fix in C.01.04 Hotfix03

Fix Information:

Fixed in C.01.05.


KPR#:1793  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Calibration

One-line Description:

Re-establish the old layout for calibration reports

Problem:

Re-establish the old layout for calibration reports

Temporary Solution:

n/a

Fix Information:

Fixed in C.01.04 Hotfix03.


KPR#:1799  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.01

Keyword: Calibration

One-line Description:

Changing calibration table in Batch review generates an error

Problem:

When run is spooled in Batch Review, click on menu Calibration –> New Caibration Table, select OK and overwrite existing table
Click OK on calibration table when it is displayed
For a brief moment an error is displayed

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1807  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Calibration

One-line Description:

Calibration method printout does not print compound level heights when quantitation is based on height

Problem:

Can't show calibration table based on Height when printing calibration table curves

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1809  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04 HF03

Keyword: Calibration

One-line Description:

C.01.04 HotFix03 : Amount units in calibration report printout is not updated

Problem:

Enter “ug/ml” for “Amount Units” in the “Calibration settings”. The calibration table was updated with “Amt (ug/ml)” but the calibration table printout for the Overview table indicated “Amt (wt%)”. This also happens with “Amt (kg)” and the calibration report printout still reflects as “Amt (wt%)”.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:263  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: CE Calibration

One-line Description:

CE system: Pressure are not applied correctly

Problem:

Running A.10.02 CE ChemStation in serial mode,
using external pressure following behavior is seen:

1. without pre or post conditioning
   the correct high pressure from the main method
   setting is used

2. As soon as pre or post conditioning is used and a 
   pressure setting is defined in the timetable, that pressure remains during the whole run and the system is not going back to the system pressure.
--This problem is intermittend 

3. as soon as pre and post conditing are used, the problem is always using the pre condition set pressure and changes only the pressure for the post conditing.

Temporary Solution:

Following workaround should be used to avoid this issue:
1. using pre and post condition
Please add an additional line in the pre conditioning using exactly the same pressure like it should be applied during the run,
even a short time will be enough. By doing so the correct value will be used to run the method and it works as well in the sequence.

2. only post condition
Please add an additional line in the post conditioning using exactly the same pressure like it should be applied during the run, 
even a short time will be enough. By doing so the correct value will be used to run the method and it works as well in the sequence.

3. only pre condition 
 this failure is intermittent,
To avoid that intermittent failure, please add an additional line in the p reconditioning using exactly the same pressure like it should be applied during the run, 
even a short time will be enough. 



Fix Information:

 No fix Planned


KPR#:302  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: CE Calibration

One-line Description:

CE: Online plot does not adjust into negative not into y axis

Problem:

Using a CE online chemstation B.01.03.
following issue has been reported using the online plot: If a negative current and voltage are set and the "adjust" button in the online plot might be used to adjust the signal, the adjust button has no impact,  the y-axis goes not into "negative".

Temporary Solution:

Use the  "Change" button to  select both signals and change manually  into the necessary negative value.

Fix Information:

n/a


KPR#:307  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: CE Calibration

One-line Description:

Not ready condition won't go away after applying external pressure

Problem:

If in the CE online ChemStationan external pressure is applied without connecting an gas supply, a not-ready condition "High pressure" is displayed and is not cleared after the end of the specified time for application of the external pressure. Only loading another method will clear the not-ready state of ChemStation.

Temporary Solution:

Load another method to clear the not-ready state of ChemStation.

Fix Information:

n/a


KPR#:536  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: CE Calibration

One-line Description:

CE: Partial Sequence initializes very slowly directly after loading a method

Problem:

When you open the Partial Sequence dialog immediately after you loaded a method, the Partial Sequence dialog takes very long to open (up to 2 minutes).

Temporary Solution:

n/a

Fix Information:

No fix is planned for this problem.


KPR#:682  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: CE Calibration

One-line Description:

CEMS configuration editor issue removing the MS part

Problem:

Following problem has been reported using CE/MS ChemStation B.02.01 SR1:

In the Device Configuration of an CE/MS delete the installed MSD, the 3D DAD will be removed as well.

Temporary Solution:

To configure the instrument as a CE only you have to delete the MSD a 2nd time and then to re-add the DAD.

Fix Information:

n/a


KPR#:881  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR2

Keyword: CE Calibration

One-line Description:

CE replenishment will stop sequence

Problem:

Following problem has been reported using Capillary Electrophorese ChemStation and using the replenishment functionality:
Replenish two vials after the other, the replenishment is executed, but the second vial is not recognized to be placed back after replenishment, so the sequence is stopping, with an message "specify outlet vial". After pressing that message, the system is able to continue.

Temporary Solution:

Please inject using the “injection table” and specify an outlet vial in the set up.

Fix Information:

n/a


KPR#:1015  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: CE Calibration

One-line Description:

Number format causes autostart macro error #16289 during CE ChemStation start

Problem:

During the start-up of a CE ChemStation a user may face the following error messages:

"Method not compatible with instrument! Current method is now not valid! Please load valid method!"

followed by an

"Autostart Macro failed, error #16289 occured".

These error messages are caused by number formates using a comma as a decimal delimiter.

Temporary Solution:

Shut down the ChemStation and change the regional settings of the operating system to English (US) and adjust time,date and currency  formats to your needs, but keep the number format.

Fix Information:

ChemStation requires a number format using a dot as decimal delimiter.
There is no design change planned.


KPR#:1432  Product:ChemStation  ChemStation (G2070BA)  B.04.02 Aladdin

Keyword: CE Control

One-line Description:

Method and Run Control not displayed correctly after reconfiguring the instrument from G1600 CE to G7100 CE

Problem:

When a CE ChemStation instrument is reconfigured from G1600 CE to G7100 CE, the Method and Run Control view is not correctly displayed, e.g. the tray is not displayed, incorrent summary status window is displayed, etc.

Temporary Solution:

1. Shut down online and offlien session of the CE instrument.
2. Using Windows Explorer, delete the files
C:\Chem32\x\config.reg
and
C:\Chem32\x\conf_off.reg,

where x is the number of the ChemStationn instrument.

Fix Information:

n/a


KPR#:1587  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: CE Control

One-line Description:

The VialTable check does not occur during the startup of the CE ChemSation

Problem:

During the startup of the CE ChemStation, the System VialTable is not loaded therefore when a method is loaded conflict checks with the method vialtable is not performed.

Temporary Solution:

If you are using an Agilent 1600 Capillary Electrophoresis instrument, the following post installations steps are required.

1. Insert the OpenLAB CDS Disk 6 
2. Open the folder called \Release Updates\OpenLAB CDS ChemStation Edition\CE\
3. Copy the file HPCE.MAC to the folder \CE\
   Note> The default chemStation path is c:\Chem32

It's recommended to create a backup copy of the original HPCE.MAC before replacing it.

Fix Information:

This issue is resolved in the software revision C.01.02


KPR#:1856  Product:ChemStation  ChemStation (G2070BA)  B.04.02 SP2

Keyword: CE Control

One-line Description:

Device Driver Error with G7100 CE after update to B.04.03 SP2 [105]

Problem:

After an upgrade from ChemStation B.04.03 SP1 to SP2 the CE 7100 instrument session would no longer work.
The software reports a driver error and certain elements are missing from the display.
This is caused by a missing registration of the instrument with the new driver interface.

Temporary Solution:

Continue working with revision B.04.02 SP1.

Fix Information:

Fixed in B.04.03 SP2 [106], July 2013.


KPR#:22  Product:LC and CE Drivers  RC 1200.NET  A.02.01

Keyword: ColumnComp Configuration , ColumnComp Driver , ColumnComp Status

One-line Description:

TCC Cluster, TCC2 shows always valve position 1

Problem:

In a TCC cluster of 2 G1316C TCC, the TCC2 shows always valve position 1. 
The TCC1 shows the correct position.

Temporary Solution:

n/a

Fix Information:

Resolved in Rc.net Driver Revision: A.02.03


KPR#:11  Product:LC and CE Drivers  RC 1200.NET  A.02.03

Keyword: ColumnComp Configuration , Pump Configuration

One-line Description:

PumpValve-Cluster & TCC-Cluster halts sequence once methods change valve position

Problem:

During sequence, if method change causes valve to be switched, the ChemStation hangs.

Temporary Solution:

n/a

Fix Information:

Fixed in A.02.04


KPR#:18  Product:LC and CE Drivers  RC 1200.NET  N/A

Keyword: Common Infrastructure

One-line Description:

High load on CAN may cause “Spectra runbuffer overflow” entries and a loss of spectra

Problem:

If LAN is connected not to LC module generating most data (i.e. to Pump in spite of DAD's presence) then interconnecting CAN connection between modules might get overloaded. Potential consecutive loss of spectra may be observed and is flagged as “Spectra runbuffer overflow” in logbook and on report.

Temporary Solution:

Installation documentation for LC HW states that the LAN access should be in the module with the highest workload, i.e. by DAD/MWD – FLD – VWD – RID – Pump – ALS (in this order) and independent whether it is a 1100 – 1200 – 1290 – 1260 system.

Fix Information:

n/a


KPR#:23  Product:LC and CE Drivers  RC 1200.NET  A.02.04

Keyword: Common Infrastructure

One-line Description:

LC instrument remains offline

Problem:

If during initialization of communication between RC.Net LC Drivers in ChemStation and the connected LC module a communication problem appeared, the LC driver might not be able to recover from any further communciation problems during initialization. Instead the instrument in ChemStation would (falsely) remain offline.

Temporary Solution:

If the network connction between ChemStation and LC is stable and operational, either a reboot of the ChemStation or of the connected LC module is able to resolve the problem arising from consecutive communication errors locking out the instrument into offline.

Fix Information:

n/a


KPR#:25  Product:LC and CE Drivers  RC 1200.NET  A.02.04

Keyword: Common Infrastructure

One-line Description:

Wrong Flow setting displayed when re-opening the Method User Interface

Problem:

Setting the Flow on the G1376A using the Method User Interface.
After pressing "OK" the flow is applied but opening again the Method User Interface the Flow=0 is displayed.

Temporary Solution:

None

Fix Information:

Fixed in Driver revision A.02.04 SP1


KPR#:1200  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Companion

One-line Description:

Companion:  Adding runs to run table is not possible while a sample is running

Problem:

It is not possible to add a sample to the run table while there is a sample being run.  New samples can only be added after pausing the run table and waiting for the run to complete or after all the runs have completed.

The icon to ‘add run’ is grayed out during a sample run.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1278  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Companion

One-line Description:

Companion: operator names cannot have spaces; invalid entry error

Problem:

At the Run table, select an operator whose name has a space. Get the following error:

'Invalid Entry
Invalid Subdirectory name:
Limit Length to 40 characters and avoid /\:.*<>[]|" '

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:1403  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Companion

One-line Description:

Companion:  Adding runs to run table is not possible while a sample is running

Problem:

It is not possible to add a sample to the run table while there is a sample being run.  New samples can only be added after pausing the run table and waiting for the run to complete or after all the runs have completed.

The icon to ‘add run’ is grayed out during a sample run.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:1474  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02

Keyword: Companion

One-line Description:

Cannot see data generated by Companion in the ChemStation explorer view

Problem:

Any data that is generated by Companion does NOT appear in the Data Analysis Explorer listing data files.

Temporary Solution:

WORKAROUND-1 
One way to get that subdirectory to show in the Data Analysis Explorer list was to do a simple run from the "normal" ChemStation which put its data file in the SAME subdirectory.  You can then see the data file just generated along with the Companion generated data files. 

WORKAROUND-2 
In the Companion data subdirectory there is a file “acquiring.txt”.  Delete this file manually and the data subdirectory appears in the Data Analysis Navigation table.

Fix Information:

n/a


KPR#:1609  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: Companion

One-line Description:

Companion fails to run blank run for valved/ALS injections

Problem:

Create a run table that uses valid locations and some that use 'blank' as the injection location (a valid choice in Companion).
The runs associated with this location fail to run with a pop-up:
'Vial range must be in range 1-32'.
This also happened for an ALS method.

Temporary Solution:

n/a

Fix Information:

Fixed in Revision B.04.01.


KPR#:1  Product:ChemStation  ChemStation (G2070BA)  A.09.01

Keyword: Config Editor

One-line Description:

Sequence and method path cannot be changed from the Configuration Editor

Problem:

If you change the path for the sequence and the method in Configuration Editor (Configure - Paths) it is not changed in the software. Going to Load Method or Method Save As the path has not changed.
Same with sequence path.

Temporary Solution:

Load a method or sequence from the location you want to use.  The ChemStation shall use this last location.

Fix Information:

No fix is planned for this problem.


KPR#:275  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Config Editor

One-line Description:

LC Chemstation allows configuring a GC6890

Problem:

In cases where a LC and GC ChemStation are installed on the same workstation, the configuration editor offers all LC and GC instruments to be configured as instrument 1 or 2.

If a GC6890 gets configured in an LC ChemStation, the startup of the program fails when trying to locate the DEF_GC method.

Temporary Solution:

In the Configuration Editor, remove the GC6890 from the LC ChemStation and add it to the GC instrument instead.

Fix Information:

There is no fix planned for this issue.


KPR#:1344  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.01

Keyword: Custom Fields

One-line Description:

XML import does not support Custom Fields

Problem:

Custom fields are used for ECM, LIMs, Easy Sequence.  XML import does not support Custom Fields.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1771  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: Custom Fields

One-line Description:

Custom Field content might get lost when a sequence is edited

Problem:

An issie with a sequence has been observed, when it uses methods with custom fields.
All fields have been populated before the sequence started. When the sequence is edited by adding runs or when the custom fields are checked during a run, the content of the fields is wiped.
The behaviour was partially fixed in revision C.01.03. When the master methods are stored in a subfolder of the methods directory, the problem still occurred.

Temporary Solution:

For OpenLAB CDS ChemStation C.01.02 a hotfix04 is available.

Fix Information:

Fixed in revision C.01.04


KPR#:1789  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Custom Fields

One-line Description:

Custom fields are not transferred to new result set

Problem:

When a new result set is created from a sequence with compound custom fields, the new sequence is missing the compound custom fields.

Temporary Solution:

n/a

Fix Information:

Issues fixed in OpenLAB C.01.05.


KPR#:1810  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Custom Fields

One-line Description:

Missing Sample Custom fields in the report

Problem:

When printing the final Intelligent Report, not all the fields are reported (for example the field for injection line 13 is present in the Custom Fields but not in the report).

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:6  Product:ChemStation  ChemStation (G2070BA)  A.09.01

Keyword: Custom Reports

One-line Description:

Add ability to insert JPG, GIF or BMP graphical file formats in Report Layout.

Problem:

Enhancement request to add support for GIF and JPG import into custom reports.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:208  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Custom Reports

One-line Description:

Custom Reports General cal.Signals always reports uncalibrated peaks

Problem:

Following issue has been reported using the Report Layout in ChemStation B.01.01 to create own report styles:

Using the "General" or "General cal.signals" sections these options do not appear to work for uncalibrated peak settings in Method in Report -Specify report options. When set to "Do Not Report" the uncalibrated peaks are always reported. 

Temporary Solution:

It is possible to create a new section for "Cal.signals" not "General cal.signals" and this will only report the calibrated peaks.

Fix Information:

No fix is planned for this problem.


KPR#:261  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Custom Reports

One-line Description:

Excel area % report based on hparea00.xls not created when no peaks identified

Problem:

When Custom Reports with Excel are used to report the ChemStation results, the following problem is observed when no peaks are found in the chromatogram:
With the Excel template HPAREA00.XLS the report is not printed (containing a message "No peaks found"), but the Excel template gives run-time error 1004 "Unable to set the Weight property of the Border class", instead.

Temporary Solution:

n/a

Fix Information:

No fix is planned for this problem.


KPR#:980  Product:OpenLAB CDS ChemStation Edition  M83xx  B.02.01 SR2

Keyword: Custom Reports

One-line Description:

Relative Retention Time solution does not work for uncalibrated peaks

Problem:

The Relative Retention Time solution available on the Agilent web site in the FAQ section only works for calibrated peaks.
So does the solution that has been implemented in C.01.04 and Intelligent Reporting.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1242  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: Custom Reports

One-line Description:

Response range from Signal Options not applied to custom reports

Problem:

For report styles based on custom report templats (.frp) the settings for the response ranges in the Signal Options are not applied.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:21  Product:LC and CE Drivers  RC 1200.NET  4.02.90

Keyword: DAD/MWD Driver , DAD/MWD Status

One-line Description:

Incomplete signal description in the online plot window

Problem:

DAD 1A (no additional information) is shown

DAD 1A, wavelenght,bandwitdh,reference WL, ref. BW is requested

Temporary Solution:

n/a

Fix Information:

Resolved in Revision: A.02.03


KPR#:10  Product:LC and CE Drivers  RC 1200.NET  A.02.03

Keyword: DAD/MWD Driver , FLD Driver

One-line Description:

G1315A/B do not show spectral data and G1321A/B show wrong time stamps for spectra

Problem:

RC.Net Driver's Interpretation of timestamp associated with spectra sent from device is wrong. So G1315A/B DADs appear to show no spectal data, while all spectral data of G1321A/B FLD appears to have a timestamp of "0".

Temporary Solution:

n/a

Fix Information:

ChemStation B.04.03 DSP2 was issued


KPR#:15  Product:ChemStation  ChemStation (G2070BA)  A.09.03

Keyword: Data Analysis

One-line Description:

Cannot resize chromatogram after you have resized it to a little box

Problem:

In LC/MSD Data Analysis, in spectral task mode, if you load a MSD signal, you will have a chromatogram displayed in the upper have of the data analysis.  If you then put your cursor on the lower right corner of the chromatogram window, and click and drag the window to the upper left, you will have a small box in the upper left corner.  At that point you cannot resize the window back to its original diminesion.  You put the cursor on the little box and try to click and drag it out, it does not expand.

The only solution is to shut down chemstation and delete config.reg and restart chemstation.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:20  Product:ChemStation  ChemStation (G2070BA)  A.10.01

Keyword: Data Analysis

One-line Description:

Chemstation Data Analysis Overlay function shows limited signals per window

Problem:

With ChemStation system A.09.01 or higher follwing issue has been reported during printing  multiple overlayed signals:
 
E.g. six signal overlay will be printed out in 2 different chromatograms when you request a Print Report. The first 4 signals print on the first page, and the second 2 print on a separate chromatogram on the second page.
 

Temporary Solution:

This issue has been seen sporadicly and depends on the used printer driver.

Fix Information:

 No fix Planned


KPR#:365  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Data Analysis

One-line Description:

Multiplier difference if used in sequence table or calibration settings

Problem:

The following problem has been reported using  ChemStation B.01.03 using the multiplier/dilution factor:

Entering a multiplier in sequence with 4 (or more) 
decimals the results are  calculated depending 
on the set up used in the Calibration Settings:

Use Sample data: From Samle Defaults Below

1. Enter multiplier in calibration setting 
   with a leading number except Zero and 4 digits
   - the system is cutting the multiplier to 
     3 digits e.g. 1.234 value

2. Enter multiplier in calibration setting 
   with a leading Zero and 4 digits
   - the system is using and the  and reporting 
     the X. xxxx value, using all entered digits
  
The used multiplier/dilution factor is printed on the report.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:421  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Data Analysis

One-line Description:

hpistd00.xls fails to produce ISTD report

Problem:

The following issue has been reported using ChemStation B.01.0x revision and the implemented custom reports to excel for Internal Standarads: HPISTD00.xls

The excel file hpistd00.xls fails to produce ISTD report, because the VBA macro stops fails during execution at:
Selection.Borders(xlInsideHorizontal).Weight = xlThin


The error message is:

Run-Time Error '1004':
Unable to set weigth property of the Border class

Temporary Solution:

The following line exists twice in the hpistd00.xls
Visiul Basic macro:
Selection.Borders(xlInsideHorizontal).Weight = xlThin.

Workaround is to comment these lines out usingwith a single quote.

Fix Information:

 No fix Planned


KPR#:447  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Data Analysis

One-line Description:

DA.M available for "Aquisition only methods" run in a sequence table

Problem:

If you run a sequence using methods with "Acquisition only" specified in the Run Time Checklist, a DA.M will nevertheless be saved in each data file.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:493  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Data Analysis

One-line Description:

Sequence Summary Reports do not work with Offline DA

Problem:

When switching from doing data processing online to offline, no reports are printed. The user will see a message indicating Summary Report requested but no Data Analysis was scheduled.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:494  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Data Analysis

One-line Description:

Running with Offline Data Analysis does not allow calibration .

Problem:

If the user is running sequences with the first sample being a calibration sample and they select Offline Data Analysis, the sample type is ignored and all samples are treated as Sample and no recalibration is performed.  With standard Data Analysis, the calibration sample gets recalibrated then the samples are run  in the normal fashion.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:531  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Data Analysis

One-line Description:

Signal Overlay toggle buttons work inverse

Problem:

When using the Overlay function of the batch review the toggle buttons for overlaid and separated display work inverse. When pressing the overlay button the signals are displayed separated and vice versa.
These buttons are only available in the signal tasks.

Temporary Solution:

Click the opposite button to achieve the desired display.

Fix Information:

There is currently no fix planned.


KPR#:555  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Data Analysis

One-line Description:

error message "Not a range" when opening a data file with very long sample info

Problem:

The error message "Not a range" is observed in the message line when opening a data file in Data Analysis that contains a very long Sample Info. This happens if in the method currently loaded the option "Sample info on each page" is selected in the report settings.
If the Sample Info is very long, the option would not allow to display any chromatogram on a single report page. This causes the error described above.

Temporary Solution:

Either use shorter Sample Info or deselect to print the Sample Info on each page.

Fix Information:

No fix is planned for this problem.


KPR#:721  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR2

Keyword: Data Analysis

One-line Description:

sequence container is not loaded if Blank(s) in the sequence name

Problem:

Following issue have been reported using ChemStation B.02.01 and higher using one or more BLANK characters in a sequence name: If the sequence name containing blanks have been used to create the sequence data container,  then the sequence will not load in DataAnalysis. No error message is given.

Temporary Solution:

Spaces/Blanks are not allowed for the sequence name if this name is used to build the sequence container.

To load the sequence data container, close the Chemstation. Than use the windows explorer to delete the blanks in the sequence and batch file name. Restart the session.
The sequence and batch file can be found in the unique data file container.

Fix Information:

The ChemStation Manuals starting with B.03.01 will contain the information about allowed used characters for datafiles, subdirectories, sequence data containers.


KPR#:745  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR2

Keyword: Data Analysis

One-line Description:

Configured colors set in configuration editor not taken

Problem:

For ChemStation Rev. B.02.01 SR1 following problem has been reported: In the configuration editor is the option present to configure colors (like backbroung, Ret Time, etc).
Specifying a color and saving the cahnges doesn't effect the ChemStation for the following screen elements :
    - Window Title
    - Retention Times
    - Compound Names
    - Annotation
    - Frame

    - Background , this feature is not used anymore, Background color setting is done using the Windows Operating system features.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:793  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: Data Analysis

One-line Description:

Calibration Table: From/To column not correctly updated

Problem:

The following problem occurs when different retention time windows have been specified for reference and other peaks:
When the calibration table is open in the "Identification Details" mode, you can review the actual retention time windows of each compound. When the retention time of a reference compound its retention time window is not updated correctly in the display. However, correct windows are applied during all the calculations.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:846  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR2

Keyword: Data Analysis

One-line Description:

Tic marks of fraction collector disappear after integration

Problem:

Following problem has been seen using ChemStation Rev. B and Fraction collection: Loading a datafile including two signals, eg. 35900 ADC and MWD, the Tick marks of the fraction collector appear in the MWD window. After using the menu function  Integration ==> Integrate, the fraction ticks disappear, but will be present in the print out.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:913  Product:ChemStation  ChemStation (G2070BA)  B.03.01 SR1

Keyword: Data Analysis

One-line Description:

Autoscale doesn't work in GC Chemstation when loading next file

Problem:

Load a sequence or single data file.  The scaling is set to full ranges.  Select Graphics > signal options and change it to autoscale. This chromatogram is correct, but now click on any other data file in the navigation table with a different method and you see that it still keeps the scaling of the other method after switching to autoscale.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:972  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: Data Analysis

One-line Description:

data files cannot be loaded when the complete data path exceeds 140 characters

Problem:

Data files cannot be loaded from the Naviagtion Table. when the complete data path exceeds 140 characters. A red macro error occurs "Could not open file: \".

Temporary Solution:

The data files can still be loaded via Load Signal from the File menu.

Fix Information:

n/a


KPR#:1007  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: Data Analysis

One-line Description:

Certain chromatograms cannot be displayed in user scale

Problem:

Certain GC chromatograms can only be displayed in full scale, autoscale or a zoomed scale.
This is very inconvenient in case of blank runs where the fixed user scale is not applied and the scale is zoomed to the highest peak. It makes carry over very visible.

After loading the datafile, it is not possible to zoom out or apply any larger user scale to suppress the trace peaks to a flat line.

LC data doesn't seem to be affected.

Temporary Solution:

Set the multi-signal settings to “overlaid” and the ChemStation will make use of the user settings.

Fix Information:

n/a


KPR#:1032  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR2

Keyword: Data Analysis

One-line Description:

ChemStation explorer content may disappear

Problem:

After sequence reprocessing of legacy 16-bit ChemStation data the content of the ChemStation B.02/B.03 explorer may vanish.
This happens if the raw data file is stored under \\chem32\x\data with no further subdirectory. Sequence reprocessing generates a new batch (.B) file which is used as an identifier for a sequence container.
As the ChemStation explorer is supposed to list only sequence container names, but not the content, all other folders besides the .B file are hidden.
If the legacy data is stored in a sequence specific subdirectory , the .B file gets generated there and the other containers do not get hidden.

Other possible problem causes are under investigation.

Temporary Solution:

Use the Windows Explorer to inspect the \\chem32\x\data folder for .B files. Create a new subdirectory and move the .B files, data files (.D) and methods (.M) to this new subdirectory, but keep the sequence containers generated by ChemStation B.02.01 or higher.

Fix Information:

There is currently no fix planned


KPR#:1039  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: Data Analysis

One-line Description:

Additions to compound name do not show / print correctly

Problem:

In Japanese ChemStation part of compound names like e.g. Roman numbers entered in Calibration Table do not show correctly in graphic's annotation. Instead the same indistinguishable glyph is shown for these various special parts of the compound name. Similarly the printout of the chromatogram as report on paper shows the same defect.

This is caused by utilizing special characters of the Japanese font like e.g. encircled numbers or Roman numbers, which are not properly drawn or printed.

Temporary Solution:

Roman numbers can be entered and formed by individual ASCII characters 'i', 'V', 'X', 'L', 'C', 'D', 'M' like e.g. 'IV'  instead of distinct two-byte UNICODE character combining it into one character.

Fix Information:

There is currently no plan to fix this problem.


KPR#:1069  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: Data Analysis

One-line Description:

Data analysis autoscales incorrectly

Problem:

Data analysis will either open in full scale or the correct scale depending on the setting of the MULTI CHROMATOGRAM settings, even when there is only one signal. 

GRAPHICS then SIGNAL OPTIONS, select Multi Chromatogram “Separated” and “Each in Full Scale” overrides the time range selection even with only one chromatogram.

Temporary Solution:

n/a

Fix Information:

There is no plan to adress this issue in future.


KPR#:1113  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.01

Keyword: Data Analysis

One-line Description:

error messages when switching tasks during reprocessing in Data Analysis view

Problem:

If you switch between tasks in Data Analysis view during sequence reprocessing, the following error messages are observed in the message line: "Job file not found" and "Couldn't end job".

Temporary Solution:

n/a

Fix Information:

 See Previous Note.


KPR#:1191  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Data Analysis

One-line Description:

Need to use different data analysis method for second signal, 3rd signal, etc

Problem:

GCs have more than one signal and customers often split their analysis to more than one detector, typically the second signal.  Customers need to be able to specify a different data analysis method for the second, third, or forth signal.  GCs do support up to 4 signals. For Dual injection and also for single sampler injections.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1249  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Data Analysis

One-line Description:

Cannot reprocess from Run Time Checklist

Problem:

When reprocessing from the Data Analysis view in the offline chemstation, ALL items in the runtime checklist are ignored, except the Data Analysis and corresponding Analysis Method for Second Signal.

This causes a problem if using macros to send data to LIMS, etc.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1354  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Data Analysis

One-line Description:

Chromatogram will rescale when switching between integration and calibration

Problem:

Chromatographic Display

In data analyis, when switching between integration and calibration the chromatogram will rescale.  This can be a real headache.  If you have the scaling set up to work with the calibration table (tabbing through target compounds), it would be very convenient to enter into the integration mode and fix and integration without “finding” your peaks again.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1361  Product:ChemStation  ChemStation (G2070BA)  (None)

Keyword: Data Analysis

One-line Description:

Isoplot for FLD data shows wrong range for EM wavelength

Problem:

IsoPlot of FLD data shows wrong range for EM-wavelength. While it is about the expected sprad of wavelength, a wrong descrition for EM-axis (x-axis) by some offset is given (e.g. 12-44nm instead of 668-7000nm).

Temporary Solution:

n/a

Fix Information:

Firmware A.06.11 [005] for the FLD solves this problem.


KPR#:1375  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02

Keyword: Data Analysis

One-line Description:

impossible to load only one signal from all runs of current sequence and then overlay with runs from another sequence

Problem:

It is impossible to load only one signal from all runs of current sequence and then overlay with runs from another sequence.
It is possible to load e.g. only signal A from each run in the Navigation and then to overlay a signal from a run of a different sequence, however when starting to review the runs via auto-stepping in the Navigation table ChemStation will again load all signals of the runs that originally were in the Navigation Table.
A specific signal has to be loaded for each file separately in this workflow scenario.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1467  Product:ChemStation  ChemStation (G2070BA)  B.04.01 SP1

Keyword: Data Analysis

One-line Description:

Automatic GPC calculation might not be triggered on fast PCs

Problem:

On faster computers (3 GHz) it has been observed that the automatic creation of GPC reports has not been triggered. Manual operation worked fine.

Temporary Solution:

To get ariound the timing issue, activate the GPC software screen first.

Fix Information:

Currently no fix available.


KPR#:1578  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: Data Analysis

One-line Description:

Reprocessing copy starting as a background window.

Problem:

When using "Reprocessing Copy" from the view menu, it may happen that the reprocessing session is hidden behind the Online session.

Temporary Solution:

You can minimize the online session to access the Reprocessing Session

Fix Information:

 See Previous Note.


KPR#:1639  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: Data Analysis

One-line Description:

Data System will not "Autoscale" a TCD data file to the correct peak

Problem:

Data analysis will not "Autoscale" a TCD data file.
AUTOSCALE autoscales to the first (smallest) peak. 

NOTE, there is no solvent, first peak at 0.798 minutes is very small, software identifies peak 4 at 2.158 minutes as the solvent.

Temporary Solution:

Full Scale works correctly.

Use Ranges can be made to work correctly.

Fix Information:

n/a


KPR#:1651  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: Data Analysis

One-line Description:

Summed Peaks calculation error in the Summed Peaks Report

Problem:

A Peak Summing Table containing two segments and each segment has a Use Reference for the response factor and an ISTD Reference.  The Chemstation does not report the amounts for each segment correctly.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1652  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: Data Analysis

One-line Description:

Y-axis labeled incorrectly when loading GC signals acquired with MSD ChemStation

Problem:

When loading GC-signals acquired with the MSD ChemStation, the Y-axis is labeled: mAu. This should be pA.

The signal units are read directly from the rawdata file "FID1A.ch"
which are written by the acquiring software.

The acquiring software shows that mAU was written.
In order to be correct this, it must be fixed by the acquiring SW.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1654  Product:ChemStation  ChemStation (G2070BA)  B.04.03 SP1

Keyword: Data Analysis

One-line Description:

ChemStation B.04.03 [54] will not "Autoscale" the TCD data file

Problem:

There is no solvent peak and the first peak is very small so the software identifies the wrong peak as the solvent.  The Chromatogram is not autoscaled correctly.

Temporary Solution:

Full Scale works correctly.

Use Ranges can be made to work correctly.

Fix Information:

n/a


KPR#:1706  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: Data Analysis

One-line Description:

Red X mark displays in data explorer of offline data analysis

Problem:

A red X mark displays in data explorer of offline data analysis during sequence.

Temporary Solution:

During data acquisition, don't display the subdirectory in the offline data explorer window.

Fix Information:

 Fixed in revision C.01.04


KPR#:1724  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: Data Analysis

One-line Description:

Last report of a reprocessed sequence is not visible.

Problem:

The report to screen can be hidden behind other screens. This has been observed when the review option is set to "use method from data file DA.M". When a sequence is reprocessed with this setting, the windows in Data Analysis get reset, when reprocessing is finished. The last on-screen report is hidden by the chromatogram window.

Temporary Solution:

Install macro tools.mac from the LC_UCL #5 , which is located on the ChemStation DVD. This offers menu entries that allow cycling through the available windows.

Fix Information:

Fixed in revision B.04.03 SP1.


KPR#:1762  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: Data Analysis

One-line Description:

Error "An additional register must be added to the list of locked registers"

Problem:

During various operations in the ChemStation (e.g. repeated manual integration of data files) a lock-up situation could occur. Internal registers were no longer accessible and the software might either no longer respond or produce error messages like "An additional register must be added to the list of locked registers"

Temporary Solution:

A fix for this issue is available for C.01.03 with Hotfix03 and any other Hotfixes released after  Hotfix03

Fix Information:

Fixed in C.01.04


KPR#:1768  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: Data Analysis

One-line Description:

Baseline not displayed properly when using ChemStation AIC’s

Problem:

A nearly horizontal baseline is not displayed or printed completely when doing a Remote Desktop connection to a ChemStation WorkStation or to a ChemStation AIC.

Temporary Solution:

None

Fix Information:

The graphics library from the ChemStation revision C.01.04, which uses the latest GDI+ API and a new GDI smoothing mode, has been implemented. A variable in the ChemStation.ini file is used to toggle between the GDI smoothing mode only or the full GDI+ features. The fix affects the appearance of chromatograms. The changes do not affect results, only the display and printing of chromatographic curves and baselines.

This defect is fixed in revision C.01.05 or higher,  C.01.04 HF03 or higher and C.01.03 Hotfix 04 and higher.


KPR#:1796  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.05

Keyword: Data Analysis

One-line Description:

Report template not found when reprocessing with a master method

Problem:

When a result set is analyzed with a master method, i.e a method from the method path, it will use a report template from the result set.
The template name is taken from the method. Since the template is not copied to the result set automatically, no report will be created.

Temporary Solution:

N/A

Fix Information:

That is the intended workflow. Users who intend to use master methods might want to use report templates that are not part of the result set.
The results of the analysis with different methods can be compared in a common template in the Review mode.


KPR#:1813  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Data Analysis

One-line Description:

Data analysis: Fractions are not displayed in UIB chromatogram

Problem:

Completed a single run with manual fraction collection.
In Chemstation Data Analysis, loaded the datafile and switched to "Purify".
The UIB1 chromatogram does not show the fractions.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1843  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03 LC DSP

Keyword: Data Analysis

One-line Description:

Snapshot runs cannot be integrated manually

Problem:

Under specific circumstances it is not possible to integrate a snapshot run manually. The manual integration buttons are greyed out and the menu is also greyed out.
This issue occurs if a sequence container is loaded while the snapshot is taken in off-line ChemStation.

Temporary Solution:

The recommended workflow is:
- Double-click the single run node
- Take the snapshot

If you are in a situation where you encounter this issue:

- Double-click the single run node
- Double-click the run with file name SNAPSHOT.D

The manual integration events are available, then.

Fix Information:

Fixed in OpenLAB CDS ChemStation C.01.04.


KPR#:1857  Product:ChemStation  ChemStation (G2070BA)  B.04.03SP2

Keyword: Data Analysis

One-line Description:

Error message: "An additional register must be added to the list of locked registers".

Problem:

When a Sequence is running in Online and another sequence is reintegrated or processed in the Offline session at the same time the system can hang up and the error message appears: "An additional register must be added to the list of locked registers". After that the session has to be restarted.

Temporary Solution:

None available.

Fix Information:

N/A


KPR#:1727  Product:ChemStation  ChemStation (G2070BA)  B.04.03 SP1

Keyword: Data Analysis , Navigation Table

One-line Description:

User is not prompted to save manual integrations to file, when there are manual events in the method.

Problem:

When users work with manual integration events the software behaviour depends on the method settings.

When the user performs manual integrations, he has two options to save his work: Save the events from one file to the method (A) or save them to the individual data file (B).
If he goes for option (A) the manual activities become a method setting. They can be automatically applied by ticking the checkbox 'Manual Events' in the integration Events dialog.
If the checkbox is not ticked, the user will see a dialog, when he loads a new file. He will be asked whether he wants to save the event to the data file or discard the changes.
If the checkbox is ticked, the dialog will not come up and additional manual events are lost.

Temporary Solution:

This is the intended behaviour. The user should save the method instead. Typically these changes go into the DA.M method inside the data file.

Fix Information:

No fix planned.


KPR#:1845  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: Data Analysis , Programming

One-line Description:

Garbage characters can appear in English LCMS reports

Problem:

Some sample parameters from LCMS data files (like e.g. Operator Name) are not printed properly. Instead unreadable text appears, like a mix of Chinese and Japanese characters. Those parameters are incorrectly extracted from the data.ms files. 
If the data file also contains DAD traces, the sample parameters are taken from there. 
There is still the risk that other applications access the incorrect values from the LCMS objects.

Temporary Solution:

Fix planned for C.01.03.

Fix Information:

Fixed in C.01.04.


KPR#:1575  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: Documentation

One-line Description:

HTML Help requires Internet Explorer revision 6.0.28.00.1106CO or higher

Problem:

The HTML Help requires Internet Explorer Rev. 6.0.2.800.1106CO or higher.
If a lower IE revision is used, it is possible that executing the 
'Reference' button on the top of the Help screen gives an 
error message: '...a Script Error ...'

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1734  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: Documentation

One-line Description:

Understanding Your ChemStation manual needs more information about peak width calculations

Problem:

The manual needs updated to include more information on the peak width calculation. Is the peak width calculation meant to approach, approximate, or enhance the standard half-height method of calculation.

Temporary Solution:

n/a

Fix Information:

Resolved in OpenLAB C.01.04.


KPR#:1783  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: Documentation

One-line Description:

G2613A 7683A injector firmware is incorrect in the Firmware guide

Problem:

The CDS Supported Instruments and Firmware Guide does not list the correct firmware for the 7683A injector.  The correct firmware is A.10.07.

Temporary Solution:

n/a

Fix Information:

CDS Supported Instruments and Firmware Guide has been updated with the correct firmware revision in OpenLAB C.01.05.


KPR#:1568  Product:ChemStation  ChemStation (G2070BA)  B.04.02 SP1

Keyword: Easy Sequence

One-line Description:

HeadSpace does not get number of samples from modified .S sequence originally run from Queue

Problem:

Run a 3 line sequence from the Queue. This puts the most recent sequence in C:\Chem32\1\TEMP\AESEQ\

Select native sequence table and modify this loaded sequence to have differently named methods (that have only GC parameter differences): M1On.M, M1onA.M, etc. for FOUR lines.
Save and run sequence. The HS gets the correct methods downloaded, but only three samples, not four. Vial Status shows the three lines with correct methods, but no fourth sample.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:585  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: ECM Integration/OpenLab Option

One-line Description:

wrong file extension for SSizip after navigating in the dialog for saving to ECM

Problem:

Data, methods, and sequences that are saved in ECM via the ChemStation ECM Integration, are stored as .SSIzip archives. Depending on the type of file, the extension is enhanced, e.g. .m.SSIzip for methods.
When the dialog for saving ChemStation files to ECM is opened, a folder is already preselected. If you navigate via the ECM root node to another folder, the extension of the file will be changed to read .m.m.SSizip.

Temporary Solution:

Check the file extension in the Save dialog before pressing the Save button. If necessary, delete any wrong additions manually.

Fix Information:

n/a


KPR#:593  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: ECM Integration/OpenLab Option

One-line Description:

ChemStation reports with CyberPrinter limited to 30 characters as file name

Problem:

When the CyberPrinter XP is selected as printer in the ChemStation, any report printed to the CyberPrinter is converted to a *.pdf file and placed in the monitored folder and transferred by the print service to the specified location in ECM. During conversion to .pdf, the name supplied by the system is fixed, and defined as
• single runs: .pdf
• sequence runs: _ .pdf
The maximum file length for this automatically generated report naming is 30 characters; longer files names are truncated. If, due to the truncation, file names are the same, the files will be saved in ECM as different versions of the same .pdf file.

Temporary Solution:

Make sure the naming scheme for files, sequences, and sequence containers is set up in a way that the limit is not exceeded.

Fix Information:

 No fix Planned


KPR#:744  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR2

Keyword: ECM Integration/OpenLab Option

One-line Description:

ECM->Save Data As changes default SSIZIP file name

Problem:

When using the ECM Save Data As function in the ChemStation the default file name should be the sequence container name, like :
Save as Opens in a folder:
- Default Name = ChemStation Container name

Under following circumstances, this is  not the case:

Navigating to a folder that contains already SSIZIP files:
 - Default Name = First SSIZIP file name in Folder

Creating a new folder:
 - Default Name = Becomes empty



Temporary Solution:

Please enter the file name manually.

Fix Information:

n/a


KPR#:774  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: ECM Integration/OpenLab Option

One-line Description:

data review does not update the .acaml file

Problem:

The .acaml file of an already acquired sequence is only updated when the sequence is reprocessed. Reviewing the data (e.g. load a data file, change the method, and print a report) does not update the .acaml file.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:966  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.02

Keyword: ECM Integration/OpenLab Option

One-line Description:

CE .ACAML file: measured mobility and corrected area are missing

Problem:

Although HPCE Mobility report style is selected, the .ACAML file adoes not contain values for measured mobility and corrected area.

Temporary Solution:

The values are exported to .ACAML if report style Short is selected.

Fix Information:

n/a


KPR#:1091  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: ECM Integration/OpenLab Option

One-line Description:

ChemStation Administration tool cannot be started right after installation

Problem:

Starting the ChemStation Admin Tool as part of the ECM integration will be launched once.
If the installation user tries to start it up again right
after the installation, it fails with an error message saying:
"Windows cannot access the specified device, path , or file. You may not have the appropriate permission to access the item."

Temporary Solution:

In order to access the Admin Tool it is necessary to log off the operating system and log on again.

Fix Information:

This issues will be described in the next edition of the installation manual.


KPR#:1101  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: ECM Integration/OpenLab Option

One-line Description:

upload process to ECM checks for server name in case sensitive way

Problem:

Check for server name for upload process to ECM is case sensitive.
Therefore, if during logon to ECM at startup of the ChemStation the server name is specified in wrong case, the upload will fail.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1102  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: ECM Integration/OpenLab Option

One-line Description:

user without corresponding privileges can save report template

Problem:

ChemStation with OpenLAB Option:
A user without corresponding privileges modifies an existing report template. When the user switches to another view, he gets asked whether he wants to save the changes. If he confirms with 'yes', the template gets saved.

Temporary Solution:

If a user should not have permission to modify a report template, do not grant him access to the Report Layout View, either.

Fix Information:

n/a


KPR#:1107  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: ECM Integration/OpenLab Option

One-line Description:

user without privileges can preview and print reports in Report Layout view

Problem:

In Report Layout View, a user without privileges to print or preview a report, can nevertheless print a report when using the corresponding entry in the File menu.

Temporary Solution:

Only grant users access to Report Layout View, who are allowed to print reports.

Fix Information:

n/a


KPR#:1128  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.01

Keyword: ECM Integration/OpenLab Option

One-line Description:

incomplete view initialization when unlocking ChemStation

Problem:

When a user unlocks ChemStation, but has insufficient prvileges to access the current ChemStation view, the ChemStation will switch to a view that he has appropriate privileges to access. However, this view is not correctly initialized.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1136  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: ECM Integration/OpenLab Option

One-line Description:

"System out of memory" exception on uploading large sequence containers to ECM

Problem:

Upload of large sequence containers to ECM fails. The limit is 100 MB of .acaml files being present in the sequence container. The size of these files directly depend on number of injections in the sequence and the number of peaks found in the chromatogram.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:1156  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: ECM Integration/OpenLab Option

One-line Description:

ECM Operator in case of data modification during acquisition and forced upload

Problem:

As with any currently acquired sequence it is also possible to review the data set that is currently acquired in the Offline ChemStation, when automatic upload to ECM is specified. In this case, when the sequence is completed the user will be informed that the second version of the data set will be uploaded when releasing the sequence container in the Offline  session ("forced upload").
For both versions (the one after acquisition and the one from the parallel offline review) the ECM operator is the user who has run the acquisition, even if another user has performed the review in the Offline ChemStation.

Temporary Solution:

n/a

Fix Information:

There currently no plans to change this behavior.


KPR#:1204  Product:ChemStation  ChemStation (G2070BA)  B.04.01 SP1

Keyword: ECM Integration/OpenLab Option

One-line Description:

ECM path dropdown box does not support tokens

Problem:

In Method and Run Control view, a dropdown list is available listing the ten last transfer paths for automatic upload to ECM. When tokens were used to specify the path in the Preferences dialog, the dropdown list will always display the Paths without the tokens, but as the paths were finally used for upload with the final values for the token fields.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1210  Product:ChemStation  ChemStation (G2070BA)  B.04.01 SP1

Keyword: ECM Integration/OpenLab Option

One-line Description:

OpenLAB Intelligence report can't be created for single runs from ChemStation UI

Problem:

An OpenLAB Intelligence report cannot be created for single runs directly from the ChemStation UI via the corresponding icon.

Temporary Solution:

Create the report in OpenLAB Intelligence Reporter client.

Fix Information:

n/a


KPR#:1212  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.01

Keyword: ECM Integration/OpenLab Option

One-line Description:

OpenLAB Option: Difficulties to lock the ChemStation during reprocessing

Problem:

During reprocessing (especially with complex reports as e.g. the Performance reports) the menu items and icon for locking the ChemStation are frequently not accessible because ChemStation is busy. The user has to wait for a point of time where ChemStation is not busy.

Temporary Solution:

n/a

Fix Information:

Fixed in revision C.01.05.


KPR#:1215  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.01

Keyword: ECM Integration/OpenLab Option

One-line Description:

Problems after moving items from ChemStation to a different LCDF path in ECM

Problem:

If a ChemStation master method, sequence template, or data set is downloaded from ECM and afterwards the copy in ECM is moved to a different LCDF path, the local copy in the ChemStation will not be linked to the item in ECM anymore. This will lead to errors during automatic data upload, local method/sequence update from ECM, viewing the ECM Properties, etc.

Temporary Solution:

n/a

Fix Information:

Fixed in revision C.01.04


KPR#:1235  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.01 SP1

Keyword: ECM Integration/OpenLab Option

One-line Description:

upgrade installation resets the settings in the Administration Tool

Problem:

If a system with B.04.01 ChemStation including G2189BA OpenLAB Option is upgraded to revision B.04.01 SP1, all settings in the Administration Tool are reset to the default values.
All configurations have to be repeated afterwards. That may include the transfer settings of each ChemStation instrument in case the Administration Tool allowed configuration separately for each instrument.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1301  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.01 SP1

Keyword: ECM Integration/OpenLab Option

One-line Description:

ChemStation instrument not responding during upload of large data sets to ECM

Problem:

When large amounts of data are uploaded to ECM via the OpenLAB Option (e.g. the sequence container of a long sequence), the ChemStation will not responding until upload is completed.

Temporary Solution:

Do not terminate ChemStation via Task Manager during the upload, but wait until upload has completed.

Fix Information:

n/a


KPR#:1452  Product:ChemStation  ChemStation (G2070BA)  B.04.01 SP1

Keyword: ECM Integration/OpenLab Option

One-line Description:

report history may display more peaks than the actual report

Problem:

In case a report style is selected that only reports calibrated peaks (e.g. Performance Report), the Report History will also display the uncalibrated peaks. This means that more peaks are listed in the Report History than in the actual report.

Temporary Solution:

n/a

Fix Information:

It is not planned to fix this problem.


KPR#:1453  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02

Keyword: ECM Integration/OpenLab Option

One-line Description:

chromatogram data not included in the ACAML file when a sequence is acquired in "Acquisition only" mode

Problem:

The chromatogram data not included in the ACAML file when a sequence is acquired in "Acquisition only" mode.

Temporary Solution:

n/a

Fix Information:

Fixed in revision C.01.04


KPR#:1458  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 SP1

Keyword: ECM Integration/OpenLab Option

One-line Description:

incorrect injection volume in the .ACAML file when using injector program

Problem:

When an injector program is used in the method, the injection volume will not be written correctly into the .ACAML file. A negative value will be reported instead of the correct volume.

Temporary Solution:

The numbers correspond as follows:

-1: no injection
-2: manual injection
-3: injector program
-4: external injector

Fix Information:

n/a


KPR#:1479  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 DSP 2 (FireAnt)

Keyword: ECM Integration/OpenLab Option

One-line Description:

there is no possibility to process items in the queue that are not related to the one that halts the queue

Problem:

In case an item in the ECM queue cannot be uploaded to ECM the complete queue cannot be processed.
If the problem of the first item is not general in nature (e.g. ECM server is down), but only the first item has a problem, it would be useful to process items in the queue that are not related to the same data set that is causing the ECM queue to halt.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1487  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 DSP 2 (FireAnt)

Keyword: ECM Integration/OpenLab Option

One-line Description:

there is no indication in the ECM Logon dialog whether an account was disabled

Problem:

If the user locks out his ECM account by entering a wrong password several times in the ECM Logon dialog for the ChemStation, there is no indication about this in the Logon dialog informing the user to contact an administrator.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1496  Product:ChemStation  ChemStation (G2070BA)  B.04.02 SP1

Keyword: ECM Integration/OpenLab Option

One-line Description:

ECM Audit Trail contain a large number of audit trail entries for each file e.g. inside a sequence container

Problem:

ECM (all versions until Version 3.3.2 Sp1) generate a very detailed audit trail on every file operation within the system. This level of auditing is also performed on SSIZIP Archives. These SSIZIP files are WINZIP like containers for a set of files. They are often used to keep a set of files together in a consistent and compressed manner.  For example, the collection of physical files that contain data for a single ChemStation Analysis, or even a sequence of Analyses, are usually stored together in an SSIZIP in the ECM, rather than as individual files.  When SSIZIP files are uploaded into the ECM, an inventory of the SSIZIP content is performed and System Audit Trail data is generated for every file it contains.  The audit trail generated can be enormous and, in some cases, unnecessary.
One typical example for this is a ChemStation Sequence Container file (SC.SSIZIP) that contains multiple runs and associated Methods, each consisting of many files. A typical ChemStation Sequence can consist of more than 1000 files. This will cause 1000 records to be added to the audit trail tables. This will lead to a enormous Audit trail tables and slow response time when querying the audit trail records of a particular file. System owners may wish to turn this feature off.

Temporary Solution:

In order to turn off the audit trail record generation for SSIZIP contents there is a configuration file that can be edited on each ECM Application Server.
This file is named AppServiceOptions.xml and can be found under:
     \Agilent Technologies\ECM Application Service\

Procedure:
1) Edit the file with a Text Editor (e.g Notepad) and find the SpecialFileHandling section.


  
    
      
        True
        True
      
      
        False
        False
      
    
  

2) Change the AuditTrailContents line to False.
True
Changes To
False

3) Restart the Application Server Service.
4) Repeat these steps for all Application Servers.

Fix Information:

n/a


KPR#:1523  Product:ChemStation  ChemStation (G2070BA)  B.04.02 SP1

Keyword: ECM Integration/OpenLab Option

One-line Description:

user privileges not applied when logging on to ECM in ChemStation

Problem:

Under the following conditions the user privileges are not applied correctly in the ChemStation when logging on to ECM:
IF the ChemStation session is started for the first time and the user who logs on on this occasion does not have the privilege to access the Command Line, the ChemStation privileges will not applied correctly to any user logging on in this ChemStation session afterwards.

The same problem can occur when logging without the Command Line privilege after deleting the files chem32/x/config.reg or chem32/x/conf_off.reg for the online or offline ChemStation respectively (x corresponds to the instrument number).

Temporary Solution:

To prevent this problem, make sure you log on as a user with the Command Line privilege when you start the ChemStation for the first time (or after deleting config.reg or conf_off.reg).
If the problem already has occured the following steps can resolve the situation:
1. Shut down the ChemStation session.
2. Delete config.reg for an online session or conf_off.reg in case of a offline session.
3. Log on to the ChemStation session as a user with the Command Line privilege.

Fix Information:

n/a


KPR#:1525  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: ECM Integration/OpenLab Option

One-line Description:

message about failed upload to ECM sometimes not displayed

Problem:

In case of unavailability of the ECM server, the message about failed automatic data upload to ECM is sometimes not displayed.

Temporary Solution:

The icon of the ECM queue displays the information about the items in the queue.

Fix Information:

n/a


KPR#:1546  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 DSP 3

Keyword: ECM Integration/OpenLab Option

One-line Description:

ECM related information missing on report for interactive reports created while sequence is still running

Problem:

While a sequence is still running, it is possible to interactively create reports for the runs that are already completed. On these reports, ECM related information (e.g. ECM upload path) are missing.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1592  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 Hotfix-2

Keyword: ECM Integration/OpenLab Option

One-line Description:

container change flag is cleared after rejecting automatic upload to ECM

Problem:

When the local copy of a data set that is related to ECM is modified, it is flag as modified (yellow color in the ChemStation Explorer). When the user selects a different data set, he is asked whether he would like to save the changed data set to ECM. If he selects "No", the change flag is nevertheless cleared (green color in ChemStation Explorer).

Temporary Solution:

Always upload changes before loading a new data set.

Fix Information:

n/a


KPR#:1599  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 DSP 2 (FireAnt)

Keyword: ECM Integration/OpenLab Option

One-line Description:

Startup time of ChemStation is long when there are many items in the OpenLAB Option spooler queue

Problem:

The startup of a ChemStation instrument may be long when there are many items in the OpenLAB Option queue, e.g. after an ECM server failure.

Temporary Solution:

This behavior is caused by the option "Manage queue on connect". As a temporary solution, diable this option and manage the queue manually until the runs have been transfered.

Fix Information:

n/a


KPR#:1611  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: ECM Integration/OpenLab Option

One-line Description:

Path definition in OpenLab Control Panel is not used in ChemStation

Problem:

The storage path that is entered in the control panel is not supposed to be the default storage location in ChemStation.
The storage path is related to EZChrom CDS projects.

Temporary Solution:

None

Fix Information:

No fix planned. See Previous Note.


KPR#:1659  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02

Keyword: ECM Integration/OpenLab Option

One-line Description:

The storage path for files in ECM is user-specific

Problem:

The LCDF storage path for saving methods to ECM is user specific.A user that logs on to ECM from different instruments will always save to "his" path by default.
While this is the intended behaviour, it can be confusing. Method developers would prefer a workflow that makes the path instrument specific.

Temporary Solution:

Check the folder path before uploading methods.
Consider including the instrument name in the folder name.

Fix Information:

Fix considered for a future revision.


KPR#:1672  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: ECM Integration/OpenLab Option

One-line Description:

it is not possible to remove AIC from networked workstation

Problem:

If a workstation is removed from a networked system, the AIC remains in OpenLAB Shared Services and cannot be removed.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1714  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: ECM Integration/OpenLab Option

One-line Description:

ChemStore legacy data doesn't get reprocessed automatically

Problem:

Migrated ChemStore data acquired with ChemStation rev. A..01.03 and earlier does not get automatically reprocessed when downloading the files from ECM.
The automatic reprocessing is requried to populate the ChemStation navigation table. The reprocessing stops with error message: 
Sequence was aborted due to a file access error. Please restart your sequence.

Restarting the sequence reprocessing does not resolve the error.

Temporary Solution:

Affected sequences cannot be reporcessed automatically, but the runs are available as single runs.
Re-calibration using the calibration samples is possible interactively. Result review can be automated by using the auto-stepping feature of recalculation mode.

Fix Information:

This defect will be fixed in ChemStation rev. B.04.03 Service Pack 1 as well as rev. C.01.04.


KPR#:1753  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: ECM Integration/OpenLab Option

One-line Description:

Wrong error message when downloading a method from ECM

Problem:

When a method is downloaded from ECM and this download is repeated in the same session an error message appears:

The Server Path is not identical.

This message is incorrect. One of the two path strings has a leading backslash, whole the other does not. 
There is also a warning that the method already exists. This message is correct.

Temporary Solution:

Repeat the download attempt. It works the second time.

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1806  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: ECM Integration/OpenLab Option

One-line Description:

Update methods from ECM dialog locks up

Problem:

When selecting ECM -> Update Methods.. in the ChemStation Method and Run Control View, the dialog box locks up if more than a certain number of methods are linked to ECM which can be as low as 20.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1818  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.05

Keyword: ECM Integration/OpenLab Option

One-line Description:

Submitting multiple sequences to queue, the User token in upload path is resolved to current user - not to submit user

Problem:

When submitting multiple sequences to the sequence queue, each sequence data path containing the  token will be resolved to the currently logged in user.

Temporary Solution:

n/a

Fix Information:

Resolved in OpenLAB CDS C.01.07 SR2.
When submitting multiple sequences to the sequence queue, each sequence data path containing the  token will be resolved to the user that submitted the sequences.


KPR#:1833  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: ECM Integration/OpenLab Option

One-line Description:

In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.

Problem:

By OpenLAB ChemStation Edition C.01.04 the requirement for ECM's remote data path (LCDF) became more strict than before. If for editing the Sequence Parameters dialog is opened, and a backslash at the LCDF's begin is missing, the builtin check of this dialog-box will demand a manual correction by the user. This can happen especially once loading a sequence template file from an earlier version of OpenLAB ChemStation Edition.

The later release of OpenLAB ChemStation Edition (like e.g. C.01.05) will tolerate and correct the missing leading backslash at ECM's remote datat path (LCDF) in Sequence Parameters dialog automatically.

Execution of sequences by OpenLAB ChemStaation Edition is not affected. I.e. sequence templates do NOT have to be corrected, neither for sequence execution nor for sequence reprocessing.

Temporary Solution:

Leaving the opened Sequence Parameters Dialog by OK is interdicted until all parameters are in valid ranges. This now for OpenLAB ChemStation C.01.04 requires the remote data path (LCDF) for ECM to start with a leading backslash. If at some sequence template (e.g. from an earlier version of OpenLab ChemStation Edition) this character was missing, it now has to be inserted. Only then the Sequnece Parameters dialog can be successfully closed and left with OK-button.

Fix Information:

n/a


KPR#:667  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: GC ACQ

One-line Description:

Displaying method in control module prevents other meth download during seq

Problem:

The handheld control module was used to view and display Method B parameters while running a sequence.  The handheld control module was left connected to the 6850 and was displaying the Method B parameters.  Leaving the control module displaying the parameters of Method B will cause the sequence to use Method B parameters such as runtime for each line in the sequence.  Other methods were not downloaded.

The reports however, will have the correct method that was supposed to be used printed on them but the ChemStation used Method B parameters.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:1184  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: GC ACQ

One-line Description:

Offline crash requires reboot of PC

Problem:

After a crash in the Offline part of the chemStation, a reboot of the PC is required to recover.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1277  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.01

Keyword: GC ACQ

One-line Description:

6850/6890 logbook does not advise of invalid inj volume overrides from seq table

Problem:

In the sequence table\Injection volume cell of a 6850 or 6890, enter an injection volume that exceeds the valid range for the installed syringe. This value is ignored and uses the method's injection volume.

Temporary Solution:

n/a

Fix Information:

This behavior is true only for the classic driver.  The RC .Net driver does check volumes entered in the sequence table and will not allow an invalid volume.


KPR#:1423  Product:ChemStation  ChemStation (G2070BA)  B.04.02 SP1

Keyword: GC ACQ

One-line Description:

Low Thermal Mass (LTM) logging should be in Seconds and not Minutes

Problem:

LTM reporting should be in seconds and not in minutes as it currently implemented.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1459  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: GC ACQ

One-line Description:

ChemStation displays -1.0 during CTC cycle process pre and post run

Problem:

During the CTC process cycles of pre and post run, ChemStation Elapsed time is showing -1.0

This is not while the GC is in a run. 

Set up a CTC run with pre and post run cycle activities.  (Prewashes and Postwashes)

Run Method.

During the pre run before the GC starts, the elapsed time will show -1.0
After the GC run is completed, and the Post run CTC events begin (post washes), the GC elapsed time will show -1.0.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1460  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.02

Keyword: GC ACQ

One-line Description:

When look-ahead vial is not found in a sequence using overlap, the currently running sample and sequence will abort

Problem:

Create a sequence using a method specifying sample overlap. If a non-initial vial is not found, the currently running sample will abort, the sequence will abort, and there will be only an error displayed in the logbook and GC Status.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1608  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: GC ACQ

One-line Description:

FetchCPVar Cannot Talk to Application Error

Problem:

Start up Chemstation, but no instrument or sampling diagrams are shown.  
FetchCPVar Cannot Talk to Application Error occurs.

Systems use 7683 ALS.

Temporary Solution:

n/a

Fix Information:

Fixed in revision B.04.03 DSP1.


KPR#:1628  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: GC ACQ

One-line Description:

FetchCPVar Cannot Talk to Application Error

Problem:

Open an online session, initial or any subsequent try. This results in a benign pop-up error:  'FetchCPVar cannot talk to application'.

Temporary Solution:

n/a

Fix Information:

Fixed in B.04.03 DSP1 [54].


KPR#:1630  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: GC ACQ

One-line Description:

Column compensation cannot be started from the Data System with 68xx RC.NET

Problem:

Column compensation (comp) cannot be started from the Data System with 68xx RC.NET

Temporary Solution:

Column compenstation must be started via the GC keyboard.

Fix Information:

No Fix Planned.


KPR#:1638  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03 GC DSP

Keyword: GC ACQ

One-line Description:

G1888A - Handshake Abort - Vial Status does not update

Problem:

G1888A configuration 
Set Handshake to Abort

Start Sequence
Open GC oven door to prior to end of vial equilibration time.

GC Logbook records:
G1888 - Job Stopped prematurely.

However,
Detailed Vial status never updates for vial one.
It will say Preparing for injection is 'In Process' and is stuck
at 40%.

Status will go to Ready.

ChemStation software has to be manually aborted.

An error in headspace (HS) that should abort the datasystem, no longer aborts.  

Abort does work using a 7697A.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1683  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: GC ACQ

One-line Description:

OpenLab CDS, C.01.02, 7820A run continues when STOP on 7820A is pressed

Problem:

A manual injection is done and if the STOP on the 7820A front panel is selected, the GC stops (and resets the initial GC settings) but the chemstation continues.

When the STOP button is pushed on the 7820A, the "current logbook" reports the end of run correctly, and reports the system is running data analysis but does not go into Data Analysis right away.

If this is a 30 minute run, and the STOP button is pressed on the 7820A after 10 minutes, AND YOU WAIT 20 more minutes until the Chemstation gets to the end of the run, a report is generated.  

The system appears to hang up waiting on the expected run time.

Temporary Solution:

Workaround is to use STOP run from the software.

Fix Information:

n/a


KPR#:1700  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: GC ACQ

One-line Description:

7890A stops run, ChemStation keeps going; GC Connection state is MONITOR in Status

Problem:

7890A with A.01.12.1 firmware connected to Chemstation B.04.03 and C.01.02 and when the 7890A reached the run time (10 minutes) the GC stopped and reset the setpoints, but the chemstation continued and did not stop. It is possible to abort the run.  Created a new method, same problem. Noticed that the dirty method flag was set every time even when there were no changes.

The 7890A is finished the run and GC ready state is Ready but ChemStation Status is BLUE, Run in progress, DATA ACQUISITION, but nRdy Wait.

GC Connection state shown in Status is Online MONITOR which is WRONG if the ChemStation is collecting data.

Temporary Solution:

Workaround:  Power cycle the 7890A, shut down chemstation, power on GC, restart chemstation and everything worked.

Fix Information:

This issue could not be reproduced in C.01.03.


KPR#:1759  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: GC ACQ

One-line Description:

Capacity factor (k') is not calculated; column information not printed in Performance report

Problem:

From the Configure Instrument user interface, there are configurable modules such as 7890A, 7697A, 7694B.  If the GC’s location in the “Selected Modules” table in not listed first or as module 1, k’ is not calculated.  Column info is not printed in the performance report.  The file acqres.reg on the .D directory contains no column information
 
For example, if the 7697 is module 1 and the 7890A is module 2 then the requests for column info would fail and k' (Capacity factor) is not calculated.  Column information will not be included in reports.

Temporary Solution:

The order in the module table is set by the order in configuration. The workaround  is to re-order configuration so that 78xx is first.
After re-order of the configuration, delete config.reg and conf_off.reg and restart the system.

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1741  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: GC Analysis

One-line Description:

Self-assembled result sets do not have full functionality for GC data

Problem:

A result set that is built from individual runs does not provide the full functionality for GC runs. The resulting sequence does not support front or back injections. Reprocessing is not possible, because there is a mismatch with the method.

Temporary Solution:

N/A

Fix Information:

Fixed in revision C.01.04.


KPR#:1825  Product:ChemStation  ChemStation (G2070BA)  B.04.03 SP1

Keyword: GC Analysis

One-line Description:

Customized GC ChemStation validation may fail

Problem:

The following defect has been observed with rev. B.04.03 SP1 of G2070BA GC ChemStation.

The validation test "default.val" fails due to a difference in the sample name. The original name contains a space after the name, but the space gets truncated during the validation. The difference is recorded as a verification failure.
Running the same test from the data archive directory still causes the same verification failure.

Note: The default.val is NOT used by the standard validation procedure.

Temporary Solution:

Follow the standard procedure for ChemStation validation or select any other validation test, like GCarea.val.

Note: In case of an update installation, the latest reference files used for verification are stored in the data archive directory.

Fix Information:

There is no fix planned.


KPR#:1738  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: GC Autosampler

One-line Description:

Unclear error message: Undefined symbol:  License$

Problem:

When a GC method is run that uses a vial number not valid for the configuration, an error message "Undefined symbol:  License$" appears.

Temporary Solution:

The error message is misleading. Check the vial range whether it makes sense for the selected injection source.

Fix Information:

Fixed in revision C.01.04.


KPR#:549  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Help

One-line Description:

Selecting RTL Help Contents brings up ChemStation Help not RTL Help

Problem:

From the RTLock or RTSearch menus, select RTL Help Contents.  ChemStation Help is shown.  RTL Help should be shown instead of the ChemStation Help.

Temporary Solution:

n/a

Fix Information:

No fix is planned for this issue.


KPR#:1105  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Help

One-line Description:

Help for the Purification Software can not be displayed on Windows Vista

Problem:

Help for the Purification Software can not be displayed on Windows Vista.

Temporary Solution:

Install WinHlp32.exe for Windows Vista.

Fix Information:

 No fix Planned


KPR#:1515  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: Help

One-line Description:

GPC help does not open on Win7 32bit

Problem:

GPC help is not accessible on a Win7 32-bit OS, because of the help file format. A message pops up, leading the user to the Microsoft update page.

Temporary Solution:

n/a

Fix Information:

After installing the Microsoft update, help is accessible.


KPR#:1613  Product:ChemStation  ChemStation (G2070BA)  B.04.03 GC DSP

Keyword: Help , Logbook

One-line Description:

Help entry links from the logbook are sometimes unrelated

Problem:

A double click on an entry in the instrument logbook calls an entry from the online help. In some cases the help pages are unrelated to the log entry (e.g. a headspace entry calls the help for UV-spectra).

Temporary Solution:

No fix available.

Fix Information:

The functionality is not implemented for all instruments and situations and will be switched off in future revisions.
No fix planned.


KPR#:33  Product:ChemStation  ChemStation (G2070BA)  A.09.03

Keyword: Initialization

One-line Description:

Opening Reprocessing copy opens also C:\WINNT\System32\CMD.exe

Problem:

When opening a reprocessing copy of Chemstation A.09.03 (View - Reprocessing copy) it opens also C:\WINNT\System32\CMD.exe. 
It appears in the toolbar and when opening it, an empty black window appears and cannot be edited.

Chemstation sessions are not effected, the cmd.exe window can be closed.
When closing the reprosessing session, also the cmd.exe window dissappears.

Temporary Solution:

Either open the Offline Version of the ChemStation 
via  Start>Programs>ChemStation>offline version
or close the cmd.exe  window.
For revision B.01.01 the cmd window needs to be closed manually.

Regarding CFR 21 part 11security:
The CMD window will be only displayed.
The user (even adminsitrator)  cannot  type or execute commands in the window.

So there is no violation of CFR 21 part 11 securtiy.

Fix Information:

n/a


KPR#:59  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Initialization

One-line Description:

unexpected information box when initializing Dual Loop Sampler

Problem:

When a Dual Loop Sampler was powercycled, after initialization an info box is displayed "The tray configuration has been changed! Please check locations used in method and injector program", although the tray configuration in fact was not changed.

Temporary Solution:

Click OK in the information box and ignore the warning.

Fix Information:

 No fix Planned


KPR#:289  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Initialization

One-line Description:

ChemStation does not launch when two MWD detectors are configured

Problem:

When two MWD detectors are configured in an 1100 instrument, the ChemStation session will not start. The error message "Autostart Macro failed, error #16289 occured" is displayed.

Temporary Solution:

Remove one of the MWD detectors from your instrument and restart the ChemStation session.

Fix Information:

No fix is planned for this issue.


KPR#:294  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Initialization

One-line Description:

Problems with initialising the 1100 instrument when a DAD G1315C is present

Problem:

In 1100 systems containing the Diode Array Detector G1315C the following problems are observed if the DAD G1315C is not the LAN connection module:
At initialisation of the instrument the error message "1100 DAD detector 1 not found. Check that 1100 DAD 1 is correctly configured" is obtained. A power failure for the DAD is observed in the ChemStation GUI afterwards.

Temporary Solution:

Connect the 1100 system to the LAN with the LAN card installed in the DAD G1315C.

Fix Information:

n/a


KPR#:372  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Initialization

One-line Description:

ChemStation startup failure in case of printer name exceeding 92 characters

Problem:

Following issue has been seen using ChemStation B.01.03 : The ChemStation has a printer name limitation of 92 characters. 
If the name is longer ChemStation might freezes during program start and turn into a non responding status in task manager.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:1345  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: Initialization

One-line Description:

CE: macro error in instrument logbook after startup without modules

Problem:

When starting a G7100A CE ChemStation instrument without any modules configured, the instrument logbook will contain a macro error: "Undefined scalar (or scalar function): vtCheckVialTable".

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1488  Product:ChemStation  ChemStation (G2070BA)  B.04.02 DSP 3

Keyword: Initialization

One-line Description:

configuration of 1120 Compact LC lost after upgrade to B.04.02 DSP3

Problem:

When upgrading a previously configured ChemStation B.04.02 SP1 with a 1120 Compact LC, the configuration gets lost. At startup of the ChemStation the user will receive an "Error while loading configuration".

Temporary Solution:

Reconfigure the instrument by executing autoconfiguration again.

Fix Information:

n/a


KPR#:1618  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: Initialization

One-line Description:

Sporadically launch of an online session will not proceed

Problem:

Sporadically the launch of an online session will not proceed in a networked setup.

Temporary Solution:

Launch the online session a second time. The second attempt usually works.

Fix Information:

No fix planned


KPR#:1664  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: Initialization , CE Control

One-line Description:

CE G1600 instrument cannot be launched

Problem:

In ChemStation revision C.01.02, an instrument configured as G1600 CE cannot be launched. The startup process hangs during "Initialize Instrument".

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:18  Product:ChemStation  ChemStation (G2070BA)  A.09.03

Keyword: Installation

One-line Description:

Installed two A/D licences, one instrument appears

Problem:

During the installation of ChemStation Rev. A.09.03, A.10.01 and A.10.02 following issue has been reported: Installing Product G2077aa (A/D licence) as instrument 1 and  Product G2073aa (A/D control module) as instrument 2, the configuration editor will install and set up instrument one only.

Temporary Solution:

The installation needs to be done in two steps:
First install G2077a A/D system as instrument one and finish the installation. Then add the second instrument G2073a and finish the installation.

Both instruments are available in the configuration editor.

Fix Information:

n/a


KPR#:327  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Installation

One-line Description:

IQ fails after upgrade to B.01.03 if LCMS patches are previously installed.

Problem:

If the B.01.01 ChemStation for LC/MSD with patches (e.g. Multimode source) is upgraded to B.01.03 the Installation Qualification will fail.

Temporary Solution:

Uninstall any LC/MS patches prior to upgrading to B.01.03 ChemStation.

Fix Information:

No fix is planned for this problem.


KPR#:377  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Installation

One-line Description:

Multimode Ion Source Report Will Sometimes Not Print Out

Problem:

The following problem was reported in the rev. B.01.03 LC/MSD ChemStation with a G1978A Multimode Source attached to the LC/MSD.

On pages 45-46 of the Agilent G1978A Multimode Source User's Guide (P/N G1978-90006) the checkout for the "Multiple FIA method for G1978A Interface in Mixed
Mode operation" is described. At the end of the procedure, the ChemStation should print out the Multimode Ion Source Report. Due to an error in the macro which generates the Multimode Ion Source Report, the report will sometimes not print out.

Temporary Solution:

A patch for rev. B.01.03 has been released that fixes this problem.  It is described in Service Note G2710-57.

Fix Information:

 No fix Planned


KPR#:505  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Installation

One-line Description:

Retention Time Locking B.01.02 SW will not install on GC Chem B.02.01

Problem:

The following error is presented when installing Retention Time Locking add-on software B.01.02 on the GC Chemstation:

“The ChemStation version that exists on this computer is not compatible with the product being installed.  Contact Agilent Sales to obtain the correct ChemStation version.”

The Retention Time Locking add-on software B.01.02 will not install on GC ChemStation B.02.01.

Temporary Solution:

n/a

Fix Information:

Install RTL revision B.01.03.


KPR#:554  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Installation

One-line Description:

uninstallation of B.02.01 SR1 (upgrade installation from B.xx.xx)

Problem:

This problem is only related to uninstallation of a B.02.01 SR1 ChemStation that originates from an upgrade installation from any other B.xx.xx revision.
For uninstallation via Add/Remove Programs, the ChemStation B.02.01 SR1 CD has to be inserted into the CD-ROM drive.

If your ChemStation B.02.01 SR1 is not an upgrade uninstallation, it can be uninstalled without the CD.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:565  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Installation

One-line Description:

IQ fails after upgrade of CE/MS ChemStation B.01.03 to B.02.01 SR1

Problem:

When a CE/MS ChemStation B.01.03 is upgraded to B.02.01 SR1, the IQ will fail with the following files marked as changed:

sys\agilentlcmsdata.dll
sys\chemrtd.exe

Temporary Solution:

Uninstall CE/MS ChemStation B.01.03 and install B.02.01 SR1 afterwards.

Fix Information:

 No fix Planned


KPR#:578  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Installation

One-line Description:

Upgrade of G2730BA LC/MS Data Analysis only from B.01.01 fails

Problem:

During an upgrade installation of G2730BA LC/MS Data Analysis only ChemStation, an error message "An on-line core must be purchased in order to install additional instrument modules." will be observed. The upgrade process will not result in a working ChemStation installation.

Temporary Solution:

Back up your ChemStation data, then uninstall ChemStation and perform a clean install of your new ChemStation revision.

Fix Information:

 No fix Planned


KPR#:582  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Installation

One-line Description:

duplicate Start Menu entries after upgrade of 2D/3D Data Analysis only

Problem:

After an upgrade installation of 2D/3D Data Analysis only ChemStation (G2090BA/G2190BA) to B.02.01 SR1 or B.03.01, the start menu entries for the instrument instances, the license editor and the scheduler are available twice - both identical and working.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:591  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Installation

One-line Description:

problems with CE/MS upgrade installations

Problem:

The following types of problems have been observed when upgrading CE/MS systems to B.02.01 SR1:
1. Installation Verification does not pass.
2. Error messages during the upgrade installation process.

Temporary Solution:

Back up your ChemStation data, then uninstall ChemStation and perform a clean install of ChemStation B.02.01 SR1.

Fix Information:

 No fix Planned


KPR#:702  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Installation

One-line Description:

GPC software does not work if installed on a path containing a blank

Problem:

If ChemStation is installed on a path containing blanks, the GPC addon software does not work.

Temporary Solution:

The ChemStation (and so the GPC addon software) must be installed on a path that dos not contain blanks.

Fix Information:

 No fix Planned


KPR#:806  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: Installation

One-line Description:

CE IQ is failing after G2201BA CEMS addon uninstallation

Problem:

The Installation Verification will fail under the following condition:
After deinstalling the G2201BA MS ChemStation addon part of a CE/MS ChemStation, the Installation Verification will report missing files.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:820  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: Installation

One-line Description:

IQ fails after an upgrade to B.03.01 (LC 3D ChemStation)

Problem:

The IQ fails after an upgrade of an LC 3D ChemStation to revision B.03.01 with several missing files.

Temporary Solution:

Backup all user created files (e.g. methods, sequences, result data), uninstall the previous ChemStation revision and then install revision B.03.01.

Fix Information:

No fix is planned for this problem.


KPR#:865  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR2

Keyword: Installation

One-line Description:

SR2 for B.02.01 does not update Help -> About for Config Editor and IQ tool

Problem:

When Service Release 2 (SR2) is installed on top of ChemStation B.02.01 SR1, the version string in Help/About is updated for the ChemStation instruments.
However, for the Configuration Editor and Installation Verification tool the version string is not updated.

Temporary Solution:

The improper revision information in the Configuration Editor and the Installation Verification tool does not influence the functionality.
The successful completion of the IQT wizard and the end of the SR2 installation proves the correct installation of all components of the SR2.

Fix Information:

 No fix Planned


KPR#:899  Product:ChemStation  ChemStation (G2070BA)  B.03.01 SR1

Keyword: Installation

One-line Description:

With Chinese ChemStation B.03.01 SR1 IQT fails for G2090BA, G2190BA, and G2730BA

Problem:

ChemStation revision B.03.01 SR1 August Edition was released August 16, 2007.

The SR1 is coming with a built-in IQT capacity for its qualification. However, with Chinese ChemStation IQT fails for G2090BA, G2190BA, and G2730BA with missing files. Please note that this problem only occurs with Chinese ChemStation.

Temporary Solution:

When qualifying Chinese ChemStation B.03.01 SR1 G2090BA, G2190BA, or G2730BA with IQT, the IQT wizard will report missing files.

A service note CHEMSTATION-40 was released documenting the affected files. Please aks your Agilent contact to obtain the service note. To document successful validation of the affected systems, attach the service note and the appropriate attachment .pdf file with the IQT qualification result.

Fix Information:

There is no plan to fix this problem for B.03.01 SR1.


KPR#:1004  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: Installation

One-line Description:

Chinese ChemStation: IQ fails after upgrade from B.03.01 to B.03.02

Problem:

For Chinese ChemStation only, the IQ fails after upgrade from revision B.03.01 to B.03.02.
One changed file is reported:
drivers\rapid\language\altcsrcamoduledrv.chs (Version 32.3.9.0)

Temporary Solution:

Delete the file drivers\rapid\language\altcsrcamoduledrv.chs before upgrade.

Fix Information:

 No fix Planned


KPR#:1097  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Installation

One-line Description:

system files may appear as different in the IQT report

Problem:

If system files are present that differ from the versions listed in the IQT reference files, this difference will be noted in the IQT report. However, the report will denote a "Pass" state.
The difference in the system file may most probably be caused by the fact that another application on the ChemStation computer installed a different version of that system file.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1098  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Installation

One-line Description:

Cancelling the installation does not rename the ChemStation backup folder

Problem:

The following problem occurs when cancelling upgrade installation to revision B.04.01:
When the upgrade installation is cancelled and the installation folder has already been moved to the backup location, this backup folder is renamed to the original ChemStation installation directory on cancellation.

Temporary Solution:

Manually rename the backup folder to the original ChemStation installation folder name.

Fix Information:

n/a


KPR#:1124  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Installation

One-line Description:

ChemStation SR Entries in Add/Remove Programs after upgrade installation

Problem:

When ChemStation service releases had been installed on top of a full version B.0x.0x, a separate entry on this service release can be found in Add/Remove Programs of Windows.
An upgrade installation to ChemStation B.04.01 or higher can be performed without uninstalling the ChemStation service releases, but it may happen that the entries for them are not removed from the list of Add/Remove Programs during the upgrade installation.
If this is the case, do not try to remove them afterwards!

Temporary Solution:

In order to avoid remaining entries of outdated service releases, uninstall them before performing the upgrade installation.

Fix Information:

No fix for this issue is planned.


KPR#:1219  Product:ChemStation  ChemStation (G2070BA)  B.04.01 SP1

Keyword: Installation

One-line Description:

B.04.01SP1 upgrade installation removes icons of add-on products from Start Menu

Problem:

An upgrade installation from B.04.01 ChemStation to B.04.01 SP1 can be performed without uninstalling add-on programs. However, during this upgrade installation the complete Agilent ChemStation folder is deleted from the Start Menu including the icons of add-on programs like ChemStore.

Temporary Solution:

n/a

Fix Information:

It is not planned to fix this issue.


KPR#:1236  Product:ChemStation  ChemStation (G2070BA)  B.04.01 SP1

Keyword: Installation

One-line Description:

Uninstallation of ChemStation does not detect addons

Problem:

When ChemStation is uninstalled, the user will not be informed in the case that add-on programs may be installed, e.g. ChemStore. Uninstallation of ChemStation will not be prevented.

Temporary Solution:

Verify the presence of add-on programs before uninstallation of ChemStation. Uninstall add-on programs first (follow the uninstallation instructions of each add-on program).

Fix Information:

n/a


KPR#:1269  Product:ChemStation  ChemStation (G2070BA)  B.04.01 SP1

Keyword: Installation

One-line Description:

Could not install ChemStation after removing previous ChemStation Installation

Problem:

ChemStation B.04.01 installed.

From the Add/Remove Programs
Remove ChemStation

Tried to install B.04.01 SP1, however it failed with the following error.

---------------------------------------------
Agilent ChemStation Setup
---------------------------------------------
Failed to rename installation directory C:\Chem32\ to C:Chem32_001. Unknown error (-1)

Installation fails and will not continue.

Temporary Solution:

Restart the PC.

Fix Information:

n/a


KPR#:1381  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: Installation

One-line Description:

ChemStation B.04.02 on bundle computers cannot be upgraded to DSP1

Problem:

Driver Service Pack 1 (DSP1) for ChemStation B.04.02 introduces support the G7100A CE instrument and the 1120 Compact LC.

The DSP1 can be installed on top of an existing ChemStation B.04.02. However,
on Agilent bundle PCs this upgrade on top of a pre-installed B.04.02 ChemStation will not work. Although at the end of the attempted upgrade a message "Setup completed the upgrade installation" is displayed, the DSP1 files have not been installed. The revision code will still be "Agilent ChemStation B.04.02 [96]".

The defect is caused by an entry in the following registry key:
HKEY_LOCAL_MACHINE\SOFTWARE\Classes\Installer\Products\1F180CCCAEA42FC4B93468945CDBB67C\SourceList\Media\MediaPackage

For a successful installation from the DSP1 DVD, the value of this key has to be "\Install\" which is not the case on the bundle PCs with B.04.02 pre-installed.

The defect NOT affect e.g. the following situations:
- upgrading a bundle system from any version prior to B.04.02
- upgrading from B.04.02 in case ChemStation B.04.02 was installed from the
  product DVD
- installation of B.04.02 DSP1 on a clean system without prior version of
  ChemStation

On the other hand, the defect will also occur, when the installation was performed from a different location as the DVD.

Temporary Solution:

Change the registry key to the correct value before upgrading to DSP1.

Fix Information:

The process for preparation of the bundle computers will be changed. Bundle computers shipping after this change will not exhibit the defect anymore.


KPR#:1404  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: Installation

One-line Description:

Installing B.04.02 sometimes corrupts the Add/Remove programs list

Problem:

Installing B.04.02 will sometimes corrupt the Add/Remove programs list.  See attached. 

Once corrupted, the user must use the CD to remove the chemstation.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:1431  Product:ChemStation  ChemStation (G2070BA)  B.04.02 SP1

Keyword: Installation

One-line Description:

after upgrade from ChemStation B.01.03, screen of UV/Vis ChemStation is displayed during startup

Problem:

After upgrading from ChemStation for LC, GC, CE, and LC/MS, revision B.01.03 to ChemStation B.04.02, during startup of the ChemStation the screen for the UV/Vis ChemStation will be displayed.

Temporary Solution:

This affaects only the screen displayed during startup, functionality is not affected.

Fix Information:

Currently no fix is planned.


KPR#:1445  Product:ChemStation  ChemStation (G2070BA)  B.04.02 SP1

Keyword: Installation

One-line Description:

Upgrade Installation fails after changing the installation path

Problem:

When using the Back button on the "Ready to install" dialog, the user can afterwards change the installation path even for an upgrade installation. However, changing the path for an upgrade installation will cause the upgrade to fail.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1512  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: Installation

One-line Description:

Configuration Assistant hangs when another controller is connected to the instrument

Problem:

The Configurationt Assistant provides the possibility to check which driver set can be used for LC instruments. After specifying the IP address of the instrument, the application will hang up without any error message if another controller is already connected to the instrument.

Temporary Solution:

n/a

Fix Information:

It is not planned to fix this problem.


KPR#:1589  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: Installation

One-line Description:

missing link and no automatic installation of the localized readme.txt of the CDS ed.

Problem:

The localized readme.txt of the CDS ed. are present on the root of the CD/DVD holding the software (OpenLAB CDS ChemStation ) .
Nevertheless for the localized readme.txt  no automatic installation of the file to the harddisk will take place, the link in the programs folder links to the english readme.txt.

Temporary Solution:

n/a

Fix Information:

 Fixed in revision C.01.04


KPR#:1627  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: Installation

One-line Description:

AIC ports not opened for 7890A acquisition

Problem:

It was not possible to run any 7890A from either the client or AIC. It was discovered that communication ports were not opened upon installation of the AIC version of C.01.02. 

file "Exceptions.AgtChemRCNetAdapter.Online"

Temporary Solution:

Manually create and open a port.

Fix Information:

n/a


KPR#:1646  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 Hotfix-2

Keyword: Installation

One-line Description:

Ugrade to B.04.02 SP1 Hotfix 2, the preference setting "integrate after load" is lost in some instances

Problem:

During upgrade to B.04.02 SP1 Hotfix 2, some instances have lost the preference setting "integrate after load".

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:90  Product:Headspace Driver Software  G292xAA  B.01.01

Keyword: Instrument Control

One-line Description:

Methods Must be Stored on the C Drive

Problem:

The headspace sampler will run a method, but it will not run a sequence if the methods are saved to a location other than C:\ 
This defect affects all headspace hardware models running with B.0x.0x series headspace Sampler Software.

Temporary Solution:

The C Drive must be selected.

Fix Information:

Point patch available for B.04.03 SP1 [87].

This is fixed in C.01.04 OpenLab CDS.


KPR#:298  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Integration

One-line Description:

manual integration not displayed when separate zooming is enabled

Problem:

The following problem is observed when "Zoom Separately" is selected in the "Signal Options" dialog when you integrate a data file with several signals:
Perform a manual integration not in Signal A, but in one of the following signals. The result is not displayed before you switch the view, e.g. to from Integration Task to Signal Task.

Temporary Solution:

Perform manual integration only without separate zoom.

Fix Information:

n/a


KPR#:413  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Integration

One-line Description:

Height-Reject is not taken in account during manual integration

Problem:

Using ChemStation B.01.0x following issue has been reported:

For manually integrated peaks the area reject is taken into account, but not the height reject.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:690  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Integration

One-line Description:

Reporting of negative peaks resulting from Area Summing

Problem:

Negative peaks should integrated using the "Negative Peak Integration" Event. The peaks will then be reported as positive values in the UI and report in order to allow for amount calculation.
However, using the "Area Sum" event, there are cases where the baseline is above the chromatogram line. The area is reported to be negative in the UI, but positive in the report.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:694  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Integration

One-line Description:

Integration events values not saved by edit entire method

Problem:

With the def_gc.m loaded, begin to Edit Entire Method...

At Data Analysis, the default events table always seems to be uECD and is grayed out (unchangeable). The integration events for that table are:
500 Slope Sens, .08 Peak width, 1 Area and Height Reject.
Save the method to a new name.

Select Data Analysis. Load the method and select Integration\Integration Events. Now the Events table drop-down is active and TST Default is loaded for test plot. Int events are 1 Slope Sens, .04 Peak width, 1 Area reject, 1.7 Height Reject. uECD is not even available from the drop-down.

If a method is edited and saved from Edit entire method, it should stay that way and not be overwritten.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:981  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.01

Keyword: Integration

One-line Description:

Integration unable to handle complex chromatogram for area sum integration event

Problem:

When creating area sum integration event, the 'off' time for one group and the 'on' time for the next group must be separated by a minimum of 0.06 minutes. In a complex chromatogram this is a significant loss of data.

Temporary Solution:

n/a

Fix Information:

Fixed in revision C.01.04


KPR#:1010  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: Integration

One-line Description:

Reintegrate "all signals" with Int. Events open only applied for one signal

Problem:

When several signals are loaded, displayed separately, and all selected via "all loaded signals" option, the following problem is observed when editing the default integration events of a detector type (e.g. "DAD Default"):

Changes in the default integration events are only applied to the signal that was the current signal, when the Integration Events dialog was opened. Only after closing the dialog, the changed events are also applied to the other signals.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1050  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: Integration

One-line Description:

Batch.M does not display special area annotation for manually integrated peaks

Problem:

The method BATCH.M does not display the special area annotation for manually integrated peaks.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1121  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Integration

One-line Description:

integration results of manually split peaks may differ from results before split

Problem:

If an integrated peak is split by using the corresponding manual integration event, the results for area and height may differ from the results of the original unsplit peak.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:1123  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Integration

One-line Description:

riders found using new exponential baseline mode can not be deleted manually

Problem:

riders found during integration using the new exponential baseline mode can not be deleted manually with the remove peak tool.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1252  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Integration

One-line Description:

timed area reject values ignored by manual integration

Problem:

A timed integration event for "Area Reject" will be ignored by manual integration. Manual Integration always refers to the initial area reject value. The initial area reject is valid for manually integrated peaks for the complete chromatogram.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1369  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Integration

One-line Description:

timed area reject values are ignored by manual integration

Problem:

When an area reject is specified as a time integration event, this area reject will be ignored by any manual integration event, i.e. a manually integrated peak will still be found if its area is below the area reject value.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1471  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.01

Keyword: Integration

One-line Description:

Integration unable to handle complex chromatogram for area summation

Problem:

When creating area summation groups, the 'off' time for one group and the 'on' time for the next group must be separated by a minimum of 0.06 minutes. In a complex chromatogram this is a significant loss of data.

Temporary Solution:

n/a

Fix Information:

Resolved in Revision C.01.04.


KPR#:1534  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 Hotfix-2

Keyword: Integration

One-line Description:

User is not asked to save manual integration events when manual events are stored in the method

Problem:

When in the Integration Events Table the option "Method Manual Events" is enabled, the ChemStation assumes the default for storing manual integration events is the method.
When the user manually integrates the runs with this setting, he will not get asked whether he wants to save them when he changes the method. In order to trigger this he would have to first select "update manual events of method" from the Integration menu.
This behavior differs from the situation when the events are stored in the data file: In that case, the user always gets asked when switching to the next data file without any additional activity from the user.
This defect implies that it could easily happen that the events are not stored. When working with OpenLAB Option, the manual events do not get uploaded to ECM.

Temporary Solution:

Traceability is nevertheless given, since the events are described in the data file log, and the report is in the report history.

Fix Information:

Fixed in OpenLAB CDS C.01.06.


KPR#:1607  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: Integration

One-line Description:

Manual integration can cause unexpected results

Problem:

Drawing a manual baseline across an existing peak typically replaces the existing integration with a manually drawn baseline.
When Tangent Skim is switched on in the integration event, a manual baseline across an existing peak can be interpreted as drawing a tangent baseline. This leads to two peaks with the same retention time. This might escape the operator's attention, if he did not zoom into the baseline.

Temporary Solution:

To be on the safe side, delete existing peaks, before you integrate again.

Fix Information:

It is planned to change this behaviour.


KPR#:1642  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: Integration

One-line Description:

Data lost when using area sum integration event, must have 0.06 minute space between groups

Problem:

When creating area sum integration event, the 'off' time for one group and the 'on' time for the next group must be separated by a minimum of 0.06 minutes. In a complex chromatogram this is a significant loss of data.

Temporary Solution:

n/a

Fix Information:

 Fixed in revision C.01.04


KPR#:1718  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: Integration

One-line Description:

Some peaks in a chromatogram cannot be manually deleted

Problem:

For some small peaks that are close to a more intense peak it might not be possible to delete their baselines manually.
In all reported the peaks were not baseline separated from bigger peaks, so that the small peaks were interpreted as 'riders' on the big peak. In all observed cases the tangent skim mode was 'New Exponential"

Temporary Solution:

Use a different Tangent Skim Mode (Straight, Exponential or Standard). If that is not possible, use a timed area reject event targeting the peak in question.

Fix Information:

No fix planned.


KPR#:1744  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Integration

One-line Description:

Misleading display of tangent peak width

Problem:

The annotation of the tangent peak width in the performance tool is done by highlighting the part the baseline defined by the intersection with the tangents. It has been considered as misleading, because the drawing suggests that the tangent width is the length of the baseline segment. The mathematical truth is the projection of the segment on the time axis. This is what the tool calculates.

Temporary Solution:

The calculation is done correctly. 
All other peaks widths are drawn as horizontal lines. Only the tangent width is drawn following the baseline  which better describes how the tangent width is constructed.

Fix Information:

N/A See Previous Note.


KPR#:1760  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: Integration

One-line Description:

Last Result Mode may not reflect interactive changes of integration events

Problem:

The following defects affects specific workflows and configurations of ChemStation version B.02.01 and higher and C.01.01 and higher:

When changing integration events while navigating through a list of samples, the changes of the integration events are not preserved in the DA.M file inside the run specific data folder (.D). This data analysis method is used when using the ChemStation "last result" review feature. The interactively changed integration events are not restored and the latest result is not shown.

This defect does not affect ChemStation systems connected to OpenLAB ECM with result audit trail turned on. Stepping through the runs without closing the integration events table is not possible in this case.
This defect does also not affect ChemStation systems connected to ChemStore, as the ChemStore spooler stores an additional copy of the method right from memory, including all interactive changes.

However, this defect affects sequence containers reloaded from ChemStore in the case where a specific DA.M method should be used for reprocessing.
It also affects sequence containers extracted from ChemStore in the context of a ChemStore to OpenLAB ECM data migration.

Temporary Solution:

For ChemStation systems not connected to ECM or ChemStore, the integration events table should always be closed before stepping to another run and if the "last result" mode is important.

ChemStore users would find the updated method in the Run information of the ChemStore Review Client, by explicitly downloading the method associated with a specific run.

For ChemStore legacy data extracted to a flat file system, the method version originally stored by the ChemStore spooler mechanism is stored as CS.M inside each data folder .D. For all data that is not affected by the described workflow defect, the DA.M and CS.M method associated with each analysis are identical and redundant.
In cases where original results cannot be restored by applying integration parameters of DA.M, the method CS.M should be used as an alternative.

Fix Information:

n/a


KPR#:1777  Product:ChemStation  ChemStation (G2070BA)  B.04.03 SP1

Keyword: Integration

One-line Description:

Manual deletion of skimmed peak does not work

Problem:

Sometimes a skimmed peak riding on top of a bigger peak cannot be manually deleted. 
This behavior seems to be limited to peaks where the skim mode is exponential. Applying the manual event Delete Peak has no effect.

Temporary Solution:

Use the manual event Tangent Skim Peak to replace the existing skimmed peak. This replacement rider can be deleted manually.

Fix Information:

Fixed in revision C.01.05.
No fix planned for the B ChemStation.


KPR#:1800  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Integration

One-line Description:

Timed integration event "Height Reject" cannot be configured.

Problem:

In C.01.04 one new timed event appeared: Height Reject.
The expected behaviour for an event parameter would be a numeric value for the minimum peak height. Instead a selection box allows values TAN and DROP. 
The timed event Height Reject should not have been in the selection list. Do not use Timed height reject in C.01.04.

Temporary Solution:

Don't use this event. Timed Area Reject works as expected.

Fix Information:

Fixed in C..01.05 by adding the event with the expected functionality.


KPR#:1838  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.05

Keyword: Integration

One-line Description:

GC compound peaks might be labelled as solvent peaks

Problem:

The integrator algorithm annotates many compound peaks with Peak Code S, which should only be given to solvent peaks.

Temporary Solution:

n/a

Fix Information:

No fix planned


KPR#:325  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: ISOPlot

One-line Description:

Isoabsorbance and 3D plots not available in data acquired with two DADs

Problem:

It was observed that with a system containing a DAD G1315C and another DAD G1315A or G1315B, the following problem occurs when spectra were acquired with the DAD G1315C:
The Isoabsorbance and 3D plots can't be viewed.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:962  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: Localization (Japanese)

One-line Description:

impossible to input Japanese Kanji in Name Pattern field on Preferences dialog

Problem:

Japanese Kanji characters cannot be entered into the Name Pattern field of the Preferences dialog.
Hiragana (fullwidth) and Katakana (both full and half width) characters can be entered.

Temporary Solution:

Use a text editor, e.g. Notepad to create the desired text with Kanji characters and paste it into the Name Pattern field.

Fix Information:

n/a


KPR#:967  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR2

Keyword: Localization (Japanese)

One-line Description:

Japanese ChemStation: VWD Dark Current test fails

Problem:

Japanese ChemStation: The VWD Dark Current test fails with error message "No sample data" in step "Performing dark current noise measurement".

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1115  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Localization (Japanese)

One-line Description:

msupdate.exe cannot be executed on Japanese Windows Vista

Problem:

Msupdate.exe (version 3.9) cannot be executed on Japanese Windows Vista.

Temporary Solution:

Just find a Japanese VISTA and found the return if the PING utility is localized and the key characters identifying the response are not there so that msupdate canüft connect to the client.
Put un-localized ping.exe file in \Chem32\MS\Firmware folder. After doing that, the msupdate should work fine.

Fix Information:

n/a


KPR#:1668  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: Localization (Japanese)

One-line Description:

Japanese characters from saved pdf report are garbled when copied and pasted to another application

Problem:

Japanese characters from a saved ChemStation .pdf report are garbled when copied and pasted to another application (e.g. Notepad, Word, Paint, etc.).

Temporary Solution:

n/a

Fix Information:

No fix is planned for this issue.


KPR#:557  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Logbook

One-line Description:

DDE Comm Failure when starting 6890 with G1888 HeadSpace

Problem:

GC ChemStation B 01.03, 6890N
HSS control soft A 01.04 controlling G1888A

Occasionally, when the 6890 GC with an G1888 HS is started, the following lines appears in the instrument logbook:

"Error in Hook LoadMethod : HS_HookLoadMethod

DDE Comm Failure: Remote execution of (method|load|C:\CHEM32\4\METHODS\LOADTEST.M\) failed. [Channel 459260]
Error 41339 occurred:
at line 486 in file Headspace
In Command method|load|C:\CHEM32\4\METHODS\LOADTEST.M\ in macro HS_Chat"

Temporary Solution:

This problem is due to timing issues between the Chemstation/HS software and the Operating system. Minor changes in the Chemstation and Windows operating system are affecting the (DDE) technology.  These types of DDE errors appear to be events associated with starting and shutting down a software application. This error is showing a timing problem and should not be captured in the Instrument logbook but it is.

The Agilent Chemstation/HeadSpace SW and Headspace Instrument will operate correctly.

Fix Information:

 No fix Planned


KPR#:803  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.01

Keyword: Logbook

One-line Description:

Only 20000 lines are kept in the sequence log

Problem:

Only 20000 lines are kept in the sequence log. When more lines have to be added, the oldest lines are overwritten.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB CDS C.01.06.


KPR#:815  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: Logbook

One-line Description:

First run of 7890A always generates error in Logbook: undefined symbol, 41276

Problem:

First run of a 7890A (only), no matter what the injection source, always generates an error in the logbook. This does not happen for subsequenct runs. Printer was configured.

'CP Macro     ========================================       10:17:48 02/21/07
  70 41de 45dc629c    0
CP Macro     In Command IF in macro HS_HookPrintInstMethod  10:17:48 02/21/07
  70 41de 45dc629c    0
CP Macro     at line 65 in file C:\CHEM32\HS\G292XAA.MCX    10:17:48 02/21/07
  70 41de 45dc629c    0
CP Macro     Error 41276 occurred:                          10:17:48 02/21/07
  70 41de 45dc629c    0
CP Macro     Undefined symbol: prInjSrcAndLoc               10:17:48 02/21/07
  70 41de 45dc629c    0'

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1405  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02

Keyword: Logbook

One-line Description:

column temperature and pressure not in the logbook when using RC.NET drivers for LC

Problem:

When using LC RC.NET drivers, the column temperature and pressure at start and end of the run are not included in the run.log and sequence.log files.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1547  Product:ChemStation  ChemStation (G2070BA)  B.04.02 SP1

Keyword: Logbook

One-line Description:

ChemLaunch: ChemStation sequence logbook and instrument logbook are not accessible

Problem:

In a ChemStation installation with ChemLaunch, the instrument logbook and the sequence logbook are not accessible.

Temporary Solution:

n/a

Fix Information:

This problem will be fixed in revision B.04.03.


KPR#:1637  Product:OpenLAB CDS ChemStation Edition  M83xx  B.01.01

Keyword: Logbook

One-line Description:

'Method not downloaded to device' message when PAL is present

Problem:

Configure 7697A to 7890A. Close and reopen session and configure a PAL, as well.

With the headspace (HS) method still loaded, Run Method to the still cabled device. This runs, but logs message to the logbook:
'G3382AA    1 Method not downloaded to the device, because>  12:12:53 03/18/11
   0 1da9 4d838485    0
G3382AA    1  the settings are not compatible with the de>  12:12:53 03/18/11
   0 1da9 4d838485    0
G3382AA    1 vice configuration.                            12:12:53 03/18/11
  67 41e2 4d838485    0
Method       Instrument running sample Vial 1 (front)       12:12:53 03/18/11
   0 1da7 4d83848f    0
7697A      1 Sequence Started                               12:13:03 03/18/11'

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:10  Product:ChemStation  ChemStation (G2070BA)  A.09.03

Keyword: Method

One-line Description:

AM/PM in time for entries in Method History, also with 24 hour time format

Problem:

The time format for entries in Method History is taken from Windows, according to the settings in control Panel, Regional Options, Time. 

Regarding this settings "AM l" and "PM l"  symbol will appear in Method History.
The behavoir remains when the  24 hour time format is configured.

Temporary Solution:

To get the proper 24 time stamp  just delete "PM" "PM symbol" (and AM from "AM symbol") in Regional Options. 
Then the above mentioned time will be displayed in 24h time format.

This  applies for Windows 2000 and Windows XP Professional.

Fix Information:

No fix planned


KPR#:23  Product:ChemStation  ChemStation (G2070BA)  A.09.01

Keyword: Method

One-line Description:

Pressure unit is not correct on CE ChemStation method printout

Problem:

The printout of a CE method shows mbar instead of bar for the High Pressure Value.

Temporary Solution:

Printout this Known Problem Report and keep it with your method printouts.

Fix Information:

This defect is planned to be fixed with ChemStation A.10.01


KPR#:30  Product:ChemStation  ChemStation (G2070BA)  A.09.03

Keyword: Method

One-line Description:

1050 VWD method printout shows too many digits for the Peak Width setting

Problem:

The method printout for the 1050 VWD shows more significant figures compared to the selection list of the VWD user interface.
This does not affect the ability to restore the method using the printed information, since the peakwidth can only be choosen from a selection list.

Temporary Solution:

A printout of this Known Problem Report should be stored along with the method printouts.

The following table shows the corresponding printed values to the setting in the user interface. It can be used to recreate the method starting from the printout.

UI setting                      method printout
-------------------------------------------------------------
0.013 min                      0.013 min
0.026 min                      0.027 min
0.053 min                      0.053 min
0.13  min                       0.133 min
0.27  min                       0.266 min
0.53  min                       0.532 min
1.06  min                       1.064 min
1.60  min                       1.596 min

Fix Information:

No fix planned.


KPR#:114  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Method

One-line Description:

During Edit Entire Method, two dialog boxes come up simultaneously

Problem:

In Method and Run Control, select the Method > Edit Entire Method... menu item.  Select all four sections to edit, and go through the dialog boxes clicking OK until you get to the Sample Purity Options dialog box.  Move it over and notice that underneath it there is the Signal Details dialog box.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:697  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Method

One-line Description:

Major error at line 433 in file X:\Astxdrivers\iq\idd\src\ui\ssl1.c

Problem:

Error with chemstation B.01.03 SR1 controlling a GC + HeadSpace. Error occurs when clicking on the inlet button: There is a major error at line 433 in file X:\Astxdrivers\iq\idd\src\ui\ssl1.c.

Temporary Solution:

It was discovered that method resolution was disabled by adding the disable6890methres=1 in chemstation.ini.

There is an error on the back configuration and using disable6890methres=1 does not allow the flows to be corrected. The error disappeared after deleting  disable6890methres=1 from the chemstation.ini. The Inlet flow could now be read properly.

This error looks bad but the system works correctly.

Fix Information:

 No fix Planned


KPR#:1202  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Method

One-line Description:

6850 Workstation Internal SW error occurred when configuring a column in 6890

Problem:

Installed a A-012 column that was already in the column catalog and is also installed in another GC

Inventory#  Model Number
652.63131   Agilent 19091A-012

Calibration Information
Inventory# 652.63131
0.00 m x 0 um x 0.17 um calibrated
Last Calibrated: 1/1/1980

On the columns screen, select Change and Install and select the A-012 column and the following error is shown:
6850 Workstation Internal Software Error

A software error has been detected within the 6850 workstation software ....  This box is shown again a couple more times after selecting OK.  You must select cancel.  

The GC is a 6890.

Temporary Solution:

Workaround: add the A-012 column with a new inventory number and then install it.

Fix Information:

n/a


KPR#:1343  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02

Keyword: Method

One-line Description:

"Error occurred during loading method" after rejecting method resolution

Problem:

When a method is loaded that requires resolution, the user has the possibility to cancel loading of this method and return to the previous method.
In this case, a macro error is observed "error occurred during loading method", and the previous method is loaded, but is marked as changed.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1353  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.01 SP1

Keyword: Method

One-line Description:

GC Offline Data Analyisis checks for ALS location when loading method

Problem:

7890 or 6890 Offline session checks for the ALS location when loading a method or processing a sequence.  The dialog box, 'resetting to Manual injection' is diplayed after loading a method with dual injectors to a GC chemstation that only has a back injector or on the same GC but you just removed one of the autosamplers.  This should not occur in offline data analysis.

Temporary Solution:

n/a

Fix Information:

Resolved in OpenLAB CDS ChemStation Edition C.01.06.


KPR#:1466  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: Method

One-line Description:

different order of instrument parameters in method printout with classic and RC.NET LC drivers

Problem:

The order of the LC instrument parameters in the method printout differs between methods based on classic drivers and methods based on RC.NET drivers.

Temporary Solution:

At the least the order of the LC instrument modules can be changed. The order in the method printout is determined by the order of the modules in the Instrument Configuration dialog.

Fix Information:

A fix for this problem is currently not planned.


KPR#:1469  Product:ChemStation  ChemStation (G2070BA)  B.04.02 DSP 2 (FireAnt)

Keyword: Method

One-line Description:

Misleading parameter mismatch message in Data Analysis during data review

Problem:

Under the following conditions a parameter mismatch message is displayed when saving a calibrated method in Data Analysis view, although the parameters for the calibration table and the detector do match:
1. In Method and Run Control a method is loaded that specifies certain detector signals for acquisition. 
2. The user afterwards switches to Data Analysis view to review some data. He loads a different method, that contains with different signals for the calibration table. 
3. When saving DA parameter changes to this second method, a parameter mismatch message will be displayed, even if there is
no mismatch in the parameters of the second method.

Temporary Solution:

As there is no real parameter mismatch in the method, the parameter mismatch message can be ignored.

Fix Information:

This problem will be fixed in revision B.04.03.


KPR#:1473  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 SP1

Keyword: Method

One-line Description:

CTC portion of method printout does not include overlap settings

Problem:

CTC portion of the method printout does not include the Setup Injector Overlap settings.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1490  Product:ChemStation  ChemStation (G2070BA)  B.04.02 DSP 3

Keyword: Method

One-line Description:

method for Compact LC is marked as changed after loading

Problem:

When a ChemStation instrument controlling a Compact LC is upgraded from a previous version to version B.04.02 DSP3, any method for the Compact LC will be flagged as modified directly after loading.

Temporary Solution:

Save the method when it was flagged as modified directly after loading.

Fix Information:

It is not planned to change this behavior.


KPR#:1517  Product:ChemStation  ChemStation (G2070BA)  B.04.02 Hotfix-2

Keyword: Method

One-line Description:

"Method changed flag" set after silent LC method update

Problem:

After upgrading a LC ChemStation to rev. B.04.02 SP1 or DSP3, the user might be asked to save method changes when loading a different method or when closing the application.
After saving the changes, the method audit trail contains a new comment like:
"G1367D mismatch between configuration and method settings automatically resolved".

This behavior can be observed with all LC methods based on RC.Net drivers that got set up with an earlier revision of ChemStation software.

In case of rev. B.04.02 SP1 it affects all types of liquid samplers.
In case of rev. B.04.02 DSP 3 it affects the generic valve of the Column Compartments.

The configuration mismatch is caused by enhancements to the method schema adding new functionality. Previously saved parameters do not get changed.

Temporary Solution:

Print out a copy of this KPR and attach it electronically (PDF) or physically to the printout of your validated method.

Fix Information:

A design improvement is considered for a future release. 08-Feb-2012 Susanne Kramer, Gerd Pfaffmann, Uwe Nassal, Katja Johann


KPR#:1558  Product:ChemStation  ChemStation (G2070BA)  B.04.02 DSP 2 (FireAnt)

Keyword: Method

One-line Description:

Changing trays in Well-Plate Samplers does not trigger LC method resolution

Problem:

Changing the tray types in a well-plate sampler can cause incompatibilities with the method. The injector program, specified wash vials or other settings will refer to positions that are no longer valid with a different tray type.
Method resolution will not kick in. It will cause an exception error, when the method is run.

Temporary Solution:

N/A

Fix Information:

N/A


KPR#:1598  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: Method

One-line Description:

Update Master Method does not update the GPC settings

Problem:

The option to update a master method from data analysis view does not update the method with any modified GPC settings.

Temporary Solution:

Update master method only updates data analysis settings from the ChemStation Core. For update of GPC settings, save the method manually.

Fix Information:

No fix planned


KPR#:1624  Product:ChemStation  ChemStation (G2070BA)  B.04.03 GC DSP

Keyword: Method

One-line Description:

Undefined symbol: PrDARep error displayed when selecting Export e-Method

Problem:

An error occurs when selecting “Export e-method”.  The error that is displayed on the bottom left side of software is “Undefined symbol: PrDARep”.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1643  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: Method

One-line Description:

Open GC edit parameters and close it without changes; method modification flag is set

Problem:

Saved a method.  Open the GC edit parameters to review a setpoint and made no changes or changed the view.  Selected OK and the method modification flag indicates a change was made.  No changes were made.

This does not occur with the Classic Driver.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1650  Product:ChemStation  ChemStation (G2070BA)  B.04.03 GC DSP

Keyword: Method

One-line Description:

Headspace or PAL Sampler sequence with methods saved to a location other than C:\ does not run

Problem:

Save method on D drive, and start sequence with headspace or PAL Sampler.
    GC stays in "Prep Run".
    headspace and PAL Samplers do not work.

Headspace or PAL Sampler sequence does not start when method is saved to the D drive

This defect affects all of the B.0x.0x series headspace and PAL Sampler software.

It is possible to run a method.

Temporary Solution:

Workaround is save and use methods from the C:\ drive.

Fix Information:

Point patch available for B.04.03 SP1 [87]

Patch has been tested with:

G2181BA ChemStore rev. B.04.02 SP2
G2183BA Security Pack   rev. B.04.02 SP2
Headspace Control Software B.01.02.1


KPR#:1663  Product:ChemStation  ChemStation (G2070BA)  A.09.01

Keyword: Method

One-line Description:

No information around handling for special characters

Problem:

ChemStation online/offline handles special characters in the following way:

(1) The (dot) '.' is accepted when entered once at the end of the name. However, more than one ‘.’ at the end of the file name is not accepted.

(2) The following characters are not accepted but there is no error message: '*' and '?'. This could be because they are wildcard characters, but an error message is encountered when trying to accept '%', '<', '>', '+', '=' and ',' as special characters.

Details will be outlined in the ChemStation Manual.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1694  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: Method

One-line Description:

Method marked Dirty on initial "Specify Report ..." with DEF_GC.M

Problem:

This problem is reproducible if you load DEF_GC.M. Once loaded, execute 'Edit Entire Method' and check 'Data Anylsis'. Close all dialogs with OK without doing changes. Then method becomes dirty [Error!].

Once the modifed DEF_GC.M is saved, the problem does not occur.

Also happens in DA View. If DEF-GC.M is loaded and "Specify report" dialog is entered and closed without a change, then the Method dirty flag is set.

Temporary Solution:

n/a

Fix Information:

Resolved in Revision C.01.04.


KPR#:1710  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: Method

One-line Description:

Cannot transfer an English method to a Japanese or Chinese ChemStation and vice versa

Problem:

It is not possible to take a method from a ChemStation running one language and move it to a ChemStation running a different language (English -> Japanese, for instance) without visiting this user interface and checking / fixing any values.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1748  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Method

One-line Description:

Add/Removal of 7697 HS from 7890 does not show a change occurred; not displayed in Method Load options; method not shown as modified

Problem:

Removed the HS in Control Panel. Open online session and select Compare at method download. The HS is not seen as a module. Method Load Options Compare does not include the HS as a module or detect any difference in the methods.  Download to instrument and no method resolution occurs. The method is not marked as modified, though the Instr.log shows the configuration changes.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1750  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Method

One-line Description:

Cannot save offline method if configuration is different: 'Method is not consistent with the configuration'

Problem:

GC with 7693 tray and dual injectors: create and save a method for that configuration. Close session, unplug the tray, select to Configure... in Control Panel and Get GC Configuration.

Open an offline session of the instrument. User cannot save any method now. OK to the Method Editor and get:
'---------------------------

---------------------------
Method is not consistent with the configuration. Reason: Configuration mismatch:  
Tray

---------------------------
OK   
---------------------------'

Temporary Solution:

n/a

Fix Information:

Fixed in revision C.01.05.


KPR#:1765  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Method

One-line Description:

"Chemmain has encountered a problem and needs to close" when selecting back/dual inj position for GC manual inj

Problem:

On a GC system with no injectors (manual injection), select injection source from menu and select Back or Dual,  ChemMain.exe error occurs.  7820A , 7890A, 6890 have no injectors installed.

Temporary Solution:

Fix included in C.01.04 Hotfix03

Fix Information:

Fixed in OpenLAB Revision C.01.05.


KPR#:1802  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Method

One-line Description:

C.01.03 upgraded to C.01.04: 'Edit Entire Method' not working

Problem:

When the user select any items from “Check method sections to edit” and click on “OK”, there is no action.  The dialog box remains at its position.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:2021  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03SP2

Keyword: Method

One-line Description:

ACQ.txt file contains Column(s) information that does not match actual column used and leads to confusion

Problem:

The ACQ.txt file found in the data directory includes a Column(s) section at the bottom of the file and this does not match the column information in the GC section.

Temporary Solution:

Print the method from the Method menu > print method and the column info at the bottom is no longer included.

Fix Information:

No fix planned.


KPR#:1621  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: Method , SEQ

One-line Description:

sequence method of additional sequence lines may be overwritten by another method

Problem:

The defect occurs when new lines are added to a currently running acquisition sequence. If lines are added directly after the currently running line and a different method is used in the new lines, the new method is overwritten by the previous method, if editing of the sequence table is completed after the current run has completed.

Temporary Solution:

Always pause the sequence when editing the sequence table.

Fix Information:

Point patches are available for ChemStation revision B.04.03 [16] and [54] and for revision B.04.02 SP1 [212] and [256]. They are available for download on the Agilent website:
http://www.chem.agilent.com/en-US/Support/Downloads/Patches/MultiInstrumentChemStation/Pages/default.aspx Fixed in revision B.04.03 SP1.


KPR#:6  Product:Method Scouting Wizard  Method Scouting Wizard  A.02.06

Keyword: Method Scouting Wizard

One-line Description:

Memory Consumption

Problem:

The Method Scouting Wizard Add-On have been tested successfully in our software QA labs. 
ChemStation requires memory for its own functionality. In addition, memory is consumed by drivers per module, by add-ons and tools. Depending on system complexity and parallel activities, the memory limit of 2 GB for Windows processes may be reached resulting in an out-of-memory error or a crash. 

Might be  affected at Method Scouting Wizard:  graphical gradient editing UI

Temporary Solution:

In order to avoid such issues we recommend for MSW usage
Do not excessively zoom or re-size the window of the graphical gradient editing UI.

In general we recommend:

•	Only use required configuration ChemStation C.01.07 SR2 and LC driver A.02.14, which have been optimized for memory consumption
•	Use recommended stack configurations. Avoid very large systems by using multiple detectors, multiple column thermostats, pump-valve clusters, ISET and fraction collectors at the same time
•	Restart ChemStation from time to time, e.g. once per week or more often for complex systems
•	Save data before starting new tasks
•	Avoid high levels of interactivity during runs by editing methods, using signal plots, using online help, starting reports etc.
•	Increasing the PC memory does not help as this does not increase allowed memory consumption per Windows process (2 GB)

Fix Information:

Planned to be fixed in a next revision of ChemStation and  LC&CE Driver version.


KPR#:7  Product:Method Scouting Wizard  Method Scouting Wizard  A.02.06

Keyword: Method Scouting Wizard

One-line Description:

Memory Problem when clicking online help

Problem:

The Method Scouting Wizard Add-On have been tested successfully in our software QA labs. 
ChemStation requires memory for its own functionality. In addition, memory is consumed by drivers per module, by add-ons and tools. Depending on system complexity and parallel activities, the memory limit of 2 GB for Windows processes may be reached resulting in an out-of-memory error or a crash. 

What might be  affected at Method Scouting Wizard:   clicking for online help

Temporary Solution:

In order to avoid issues we recommend for MSW usage to save your work before clicking the online help button 

In general we recommend:

•	Only use required configuration ChemStation C.01.07 SR2 and LC driver A.02.14, which have been optimized for memory consumption
•	Use recommended stack configurations. Avoid very large systems by using multiple detectors, multiple column thermostats, pump-valve clusters, ISET and fraction collectors at the same time
•	Restart ChemStation from time to time, e.g. once per week or more often for complex systems
•	Save data before starting new tasks
•	Avoid high levels of interactivity during runs by editing methods, using signal plots, using online help, starting reports etc.
•	Increasing the PC memory does not help as this does not increase allowed memory consumption per Windows process (2 GB)

Fix Information:

Fixed with Method Scouting Wizard A.02.06 when using ChemStation C.01.07 SR2 Hotfix 1 and LC&CE Drivers A.02.15.
Alternatively:
Fixed with Method Scouting Wizard A.02.07 based on ChemStation C.01.07 SR3 and LC&CE Drivers A.02.16.


KPR#:8  Product:Method Scouting Wizard  Method Scouting Wizard  A.02.06

Keyword: Method Scouting Wizard

One-line Description:

Pressure spikes seen when switching columns

Problem:

Switching columns at full flow can cause pressure spikes, which can damage the columns and will over time reduce the life time of the instrument.
Method Scouting Wizard writes column screening sequences, which means the column selection valve is switching between different levels of pressure, e.g. from a flush runs with high flow rate against waste to an equilibration run for the next column typically with much lower flow rate. 
The valve might still switch at the high flow rate of the flush or of a previous column, which can push high pressure spikes against the next column. 

A protection is required to lower the flow and pressure before switching the to a different column.

Temporary Solution:

n/a

Fix Information:

This problem is in particular obvious when you use the MSW, but it is a general problem.
Therefore, there will be no fix in MSW.

Instead, the problem will be generally addressed by the instrument drivers.


KPR#:9  Product:Method Scouting Wizard  Method Scouting Wizard  A.02.07

Keyword: Method Scouting Wizard

One-line Description:

Error message when starting Method Scouting wizard about not available paths

Problem:

If you delete a configured ChemStation instrument and then create a new one again with the same instrument number, you might get warnings when starting MSW in this new ChemStation instrument about missing campaign directory (path). 
Reason is that the MSW is still following paths settings made for the previous ChemStation instrument configuration in order to find the campaign root directory.

Temporary Solution:

Workaround: 
BEFORE (!) deleting a configured ChemStation Instrument remove all path settings that point to the MSW campaign root directory. 
You can do this in the respective ChemStation Instrument under View\Preferences using the "Paths" tab.

Fix Information:

Fix planned in a future revision of the ChemStation


KPR#:1333  Product:ChemStation  ChemStation (G2070BA)  N/A

Keyword: Method Scouting Wizard

One-line Description:

Method scouting sequence does not start due to missing methods

Problem:

In cases where the ChemStation preferences "unique folder creation" are turned OFF, no method development sequence created with the Method Scouting Wizard add-on will run successfully. The sequences will stop due to missing methods that cannot be located at the specified path.

The "unique folder" data storage is mandatory for method scouting sequences generated by the Agilent Method Scouting Wizard add-on.

Temporary Solution:

Turn the "unique folder creation" option ON.
This setting is available in the ChemStation preferences menu.

Fix Information:

A fix is considered for the next version of Method Scouting Software.


KPR#:1482  Product:Method Scouting Wizard  Method Scouting Wizard  A.02.01

Keyword: Method Scouting Wizard

One-line Description:

Misleading UI for equalibration and column storage solvent flows

Problem:

The column storage and column equalibration tab of Method Scouting Wizard A.02.01 says that the base methods flow is used to prepare for the next run.
In case of different column diameters the flow gets effectively adjusted to keep the solvent velocity constant, in case where the flow adjusment option has been selected.

Temporary Solution:

N/A

Fix Information:

The user interace and on-line help of rev. A.02.02 got changed to avoid this missunderstanding.


KPR#:1483  Product:Method Scouting Wizard  Method Scouting Wizard  A.02.01

Keyword: Method Scouting Wizard

One-line Description:

Error message when generating a printout of the campaign

Problem:

Method Scouting Wizard A.02.01:
When printing the column table and the width of the table exceeds the page size, an error message is shown.

Temporary Solution:

N/A

Fix Information:

This defect is fixed in rev. A.02.02 of Method Scouting Wizard software.


KPR#:1484  Product:Method Scouting Wizard  Method Scouting Wizard  A.02.01

Keyword: Method Scouting Wizard

One-line Description:

Unexpected program termination while reviewing an existing campaign

Problem:

Reviewing an existing Method Scouting campaing with rev. A.02.01 may lead to an unexpected program termination. The shutdown is caused by a memory leak.

Temporary Solution:

N/A

Fix Information:

This defect is fixed in rev. A.02.02 of Method Scouting Wizard software.


KPR#:1555  Product:Method Scouting Wizard  Method Scouting Wizard  A.02.01

Keyword: Method Scouting Wizard

One-line Description:

Option to Change Vial after n injections might not work in a sequence

Problem:

When the combined volumes drawn from one vial exceed the vial volume, the software displays a warning. There is also the option to change the vial after n injections. 
There is one situation, where the option to change the vial cannot be applied:
When a sequence line has multiple injections (number injections>n) no warning message will be displayed.

Temporary Solution:

N/A

Fix Information:

Fixed with MSW A.02.05


KPR#:1556  Product:ChemStation  ChemStation (G2070BA)  B.04.02 DSP 2 (FireAnt)

Keyword: Method Scouting Wizard

One-line Description:

Injection counters on tagged columns do not get updated

Problem:

In a TCC cluster with several tagged columns the injection counters on the tags do not get updated.

Temporary Solution:

N/A

Fix Information:

N/A


KPR#:1561  Product:Method Scouting Wizard  Method Scouting Wizard  A.02.01

Keyword: Method Scouting Wizard

One-line Description:

Sample info for blank runs might not be accurate

Problem:

The method scouting wizard writes screening parameters like solvent composition into the sample info field.
For blank runs this approach can lead to incorrect results.

Temporary Solution:

Ignore the info. Blank runs are not used for screening.

Fix Information:

No fix planned.


KPR#:4  Product:ChemStation  ChemStation (G2070BA)  A.09.01

Keyword: MSD Analysis

One-line Description:

Error "Identifier (MSO_PPwidth) is illegal or variable can't be manipulated"

Problem:

After upgrading from rev. A.09.01 to A.09.03 LC/MSD ChemStation, then go into Data Analysis and select the Peak Purity tool and click on the first peak, the error  "Identifier (MSO_PPwidth) is illegal or variable can't be manipulated" can occur.

This does not occur with a clean load of rev. A.09.03.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:103  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: MSD Analysis

One-line Description:

Cannot install G1043A NIST Library Search Program on B.01.01 LC/MSD ChemStation

Problem:

When trying to install G1043A NIST Library Search Program rev. D.01.00 by running setup.exe from the CD root directory, the error "You must install the HP MSD ChemStation Software before installing this Database sofware" is generated.  An OK button is displayed, and when it is clicked the installation program stops. 

Temporary Solution:

n/a

Fix Information:

A fix is currently being investigated.


KPR#:124  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: MSD Analysis

One-line Description:

Data Browser rev. A.01.01 not supported with rev. B.01.01 LC/MSD ChemStation

Problem:

The currently shipping G2727AA Data Browser software rev. A.01.01 (Sept 2004) is not supported and does not work with the LC/MSD ChemStation revision B.01.01.

A new version of Data Browser software compatible with rev. B.01.01 LC/MSD ChemStation will be available in the future.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:125  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: MSD Analysis

One-line Description:

G2720BA Bioanalysis SW not available at release of rev B.01.01 LC/MSD ChemStati

Problem:

At the release of rev. B.01.01 LC/MSD ChemStation software (Sept 2004), the G2720BA Bioanalysis software (Peptide Tools + Deconvolution) is not available.  The Peptide Tools functionality is currently not supported with rev. B.01.01 LC/MSD ChemStation, and the Peptide Tools menu item in the Spectra menu in Data Analysis is greyed out.  

However, the Deconvolution program part of the Bioanalysis software does work and has been tested.  For customers upgrading from rev. A.xx.xx LC/MSD ChemStation + G2720AA Bioanalysis software to rev. B.01.01 LC/MSD ChemStation and who wish to continue to use Deconvolution, the G2720AA license number can be entered for the G2720BA product during the ChemStation software installation.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:267  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: MSD Analysis

One-line Description:

Batch review, no manual integration id smoothing is selected in signal details

Problem:

Following issue has been reported using B.01.01 LC/MS ChemStion only in conjuction with smoothing: 
The Signal Detail screen for LC-MS allows to select smoothing. If it is selected, then manual integration are not marked  on batch review results. The user can do a manual integration and can save it to batch. The first column will get the "s" indicator. If the data file is re-selected, then manual integration is not visible. (the "s" indicator is still in place). 

Temporary Solution:

The issue only occurs if the user selects 'gausian' smoothing and leaves the default value of '0' in the 'smooth parm' column of signal details. Please enter a non-zero value as a work-around.

Fix Information:

 No fix Planned


KPR#:310  Product:ChemStation  ChemStation (G2070BA)  (None)

Keyword: MSD Analysis

One-line Description:

Sample Purity affects integration events settings,  sets method change flag

Problem:

While developing a method containing "sample purity options" the system will change integration events during integration. In addition the method change flag will be set during running a sequence, and that method change flag will be present starting from sample 2 till the end of the sequence. The flag is triggered due to the changes of the integration events.

Temporary Solution:

n/a

Fix Information:

No fix is planned.


KPR#:314  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: MSD Analysis

One-line Description:

TIC report displays only one MS parameter when in FIA mode

Problem:

When you set a FIA method with two MSD parameters as being variables, in the final report only the first one in the FIA table is displayed.

Temporary Solution:

The user can set to have those 2 parameters visible in the report by going to Data Analysis>Graphics>FIA Display Options...

Fix Information:

 No fix Planned


KPR#:1665  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: MSD Analysis

One-line Description:

Deconvolution tools are not available for LC/MS

Problem:

The Deconvolution tools are not available in LC/MS ChemStation even if BioAnalysis Tools are configured for the instrument.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:312  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: MSD Analyst

One-line Description:

.wiff file cannot be opened in Analyst if the method has a Post time set .

Problem:

This happens with G2710 LCMS + Analyst 1.4 MR + the Analyst patch.
-Set up a method that has a post time set up in the pump.
-Set up the Analyst to do automatic wiff file translation, launch Analyst at the end of the run. Use the naming convention for wiff file same as that of data file.(the default option).
-Run this method.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:313  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: MSD Analyst

One-line Description:

Analyst does not launch automatically if  root directory is not default one.

Problem:

-Install G2710 LCMS and Analyst 1.4MR with Analyst patch .
-Launch Analyst  and create a new root directory e.g Build Test.
-In Analyst/Set Up Current Method menu select options  ‘Automatic wiff file translation’ and ‘Automatically Launch Analyst’ and  using the   select the above created root directory. Keep all other options as default.
-Run a Method/Sequence

At the end of the run, Analyst does not launch automatically and pops up 'translat' dialog box.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:319  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: MSD Analyst

One-line Description:

Analyst conversion log file shows warning that CRC doesn't match

Problem:

When running a conversion using the ChemStation->WIFF with a .csv file option, the log file shows a warning that the CRC doesn't match.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:14  Product:ChemStation  ChemStation (G2070BA)  A.09.01

Keyword: MSD Control

One-line Description:

Get application error in DLLHOST.EXE when setting up for running Multiple FIA

Problem:

If you into Run Control/Run Multiple FIA Methods, then select Edit/Group/Add and give it a group name (eg testgroup), then select Edit/Method/Append, intermittent you will get an error:

"An application error has occurred...

DLLHOST.EXE

Exception: access violation:
(0xc0000005, Address: 0x10011c5b

If you then close ChemStation, you will get an HPCORE error (no other details, no GPF message), then you click okay to that and ChemStaiton window will close.

It occurs with WIndows 2000 or WInNT

Temporary Solution:

Running Multiple FIA Methods is not supported with rev. B.01.01 LC/MSD ChemStation.  It will be supported and the defect fixed in a future revision.

Fix Information:

n/a


KPR#:123  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: MSD Control

One-line Description:

Easy Access rev. A.03.00 not supported with rev. B.01.01 LC/MSD ChemStation

Problem:

The currently shipping G2725AA Easy Access software rev. A.03.00 (Sept 2004) is not supported and does not work with the LC/MSD ChemStation revision B.01.01.

A new version of Easy Access software compatible with rev. B.01.01 LC/MSD ChemStation will be available in the future.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:309  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: MSD Control

One-line Description:

Abundance Values missing from Lens Ramps

Problem:

The abundance values are cut off in the Lens Ramp window from Manual Tune.  This makes it impossible to determine the maximum abundance range when ramping a lens element.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:318  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: MSD Control

One-line Description:

Display of Time Delay is incorrect for MS acquisition while SIM on Target Mass.

Problem:

If you set a time delay for start of the MS run, then select SIM on Target Masses, then enter a Target Mass, the time delay resets to 0.  This is for display purposes only--it actually does start after the delay time but the display says 0.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:320  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: MSD Control

One-line Description:

G2730BA Standalone DA displays Sample Purity Options dialog.

Problem:

If you install G2730BA LC/MSD Standalone Data Analysis you will be able to access the Sample Purity Options dialog under the Method menu.  However, the MSD options will be grayed out.

Since this is a Standalone Data Analysis and the Sample Purity Options are NOT post-acquisition parameters, that dialog should not be available in G2730BA.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:1437  Product:ChemStation  ChemStation (G2070BA)  B.04.02 DSP 2 (FireAnt)

Keyword: MSD Control

One-line Description:

ChemStation with CTC sampler hangs up when stopping or aborting a sequence

Problem:

When using the "DLW_Priming_Rev01" injector program in a configuration including a CTC sampler, ChemStation will hang up when stopping or aborting a sequence.

Temporary Solution:

The "DLW_Priming_Rev01" injector program originates from an earlier version of the CTC control software and has been removed from revision A.01.06 which is required for ChemStation B.04.02.

The following workaround is avaialble:

- Use the standard DLW injector program (LC-Inj DLW Standard_Rev03.cyx) and perform a blank injection. This injector program is included with the Agilent A.01.06 CTC Control CD.
- Since the DLW flushing times are method-dependent, they can use the standard DLW injector program to create a “Priming.M” method that will perform the DLW flushing steps and stator wash.  
- Because the standard DLW injector program does perform an injection, it can be used with ChemStation sequences and MassHunter worklists.

Fix Information:

n/a


KPR#:1675  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: MSD Control

One-line Description:

Automated Delay Calibration of the MS signal for Fraction Collection is not supported in the RC.Net environment

Problem:

Automated Delay Calibration of the MS signal for Fraction Collection is not supported in the  RC.Net environment.  In the Classic Environment, Automated Delay Calibration of the MS signal for Fraction Collection is supported and accessible through the Diagnosis View from the Maintenance | Fraction Collector ... | Delay Volume Calibration ... menu item.

Temporary Solution:

To perform a Delay Calibration of the MS signal for Fraction Collection, configure the system to operate in the Classic Environment.  This may require setting some LC Modules into "emulation mode" to be recognized in the Classic Environment.  Once the Delay Calibration(s) is/are completed, make note of the Delay Calibration value(s) and revert the system to RC.Net environment if so desired.

For systems that cannot be reverted to operate in the Classic Environment to perform Delay Calibration, please contact Agilent support for assistance.

Fix Information:

Delay Volume Calibration Wizard is introduced in OpenLAB ChemStation C.01.07 SR3


KPR#:1686  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: MSD Control

One-line Description:

MSD Tune View will awlays load ATUNES.TUN if tune file loading is cancelled

Problem:

In MSD Tune View, it is possible to manually load a tune file from the File | Load MSD Tune File... menu item.  If the "Cancel" button is clicked in the subsequent Dialog Box, the tune file ATUNES.TUN will be loaded, regardless of what tune file was previously loaded.  The loaded tune file is displayed in a drop-down box below the menu bar.

Temporary Solution:

In the event that the tune file loading process is cancelled as in the Problem Description, please ensure that the desired tune file is subsequently loaded in MSD Tune View.  Note that this problem does not impact data acquistion, as each unique data acquistion method will load the tune file that is specified as part of the method.

Fix Information:

n/a


KPR#:1730  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: MSD Control

One-line Description:

LC/MS may remain in a run state after Method/Sequence termination due to an HPLC hardware error

Problem:

After the HPLC experiences a hardware error (such as a liquid leak), the acquistion method/sequence will terminate.  However, the LC/MS  may remain in a run state.

Temporary Solution:

If the LC/MS remains in a run state after the HPLC experiences a hardware error, the ChemStation software must be closed and re-openned.  This will reset the LC/MS status to ready and bring it out of the run state.

Fix Information:

26-Apr-2012 Susanne Kramer, Heiko Fessenmayr, Achim Teubner, Uwe Nassal, Katja Johann


KPR#:719  Product:OpenLAB CDS ChemStation Edition  M83xx  B.02.01 SR1

Keyword: Navigation Pane

One-line Description:

Performance problem if many data files to be shown in Navigation Table

Problem:

A perfomance issue using ChemStation B.02.01 and higher has been reported under the following condition: 
If many single runs are in one folder, then the ChemStation needs quite some time to fill the Navgation Table with the necessary data.

Temporary Solution:

Split the content of one huge folder containing single runs into smaller directories.

Fix Information:

n/a


KPR#:836  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR2

Keyword: Navigation Table

One-line Description:

Navigation table cannot display this date time format 20-03-2007 11:41:22

Problem:

The Chemstation cannot display the following date time format 20-03-2007 11:41:22 in the Navigation table in Data analysis.  The following date is shown when this date time format 20-03-2007 11:41:22 is used in the regional settings:  01-01-0001 00:00:...

The date time format shown in the File Information is: 20-03-2007 11:41:22.   The report shows the date correct and not this strange date: 01-01-0001 00:00:....

The navigation table can display correctly the following date/time format:  25-Apr-07, 08:30:04

Temporary Solution:

Work around:  Use the following date time format in the regional settings:
25-Apr-07, 08:30:04

Fix Information:

 No fix Planned


KPR#:1752  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Navigation Table

One-line Description:

Changing the sample type in a sequence with cyclic calibration corrupts the sequence structure

Problem:

When a sequence has been run with cyclic calibration, it is not possible to change the sample type for the calibration runs to Sample.
When this is done and the sequence is reprocessed, the sequence structure is corrupted. The sequence cannot be loaded in the Navigatio Table. The Navigation Table will not be usable and the ChemStation has to be closed and reopened.
In Cyclic calibrations the structure of the sequence, the order of runs and the file naming depend very much on the sequence table setup. Later changes will cause conflicts.

Temporary Solution:

To reconstruct the original sequence the batch file needs to be removed from the sequence folder. Clear the Navigation Table and load the individual runs from the file menu. Then create a new result set from the runs using the functionality from the Sequence menu.

Fix Information:

N/A


KPR#:1612  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: Output

One-line Description:

Printing fails if printer name exceeds 100 characters

Problem:

Printing fails without error message, when the length of printer name plus port name exceeds 100 characters. For a TCP/IP port the printer name should not exceed 87 characters.

Temporary Solution:

Use a shorter printer name.

Fix Information:

 See Previous Note.


KPR#:7  Product:ChemStation  ChemStation (G2070BA)  A.09.03

Keyword: Peak Purity

One-line Description:

A09.03 Peak Purity calculation

Problem:

In Peak Purity Task the Purity Calculation of the peak will change after hitting the option button and leaving the window without changes. The Purity Factor of the Peak will be recalculated and the result can differ.

Temporary Solution:

When working with peak purity, Agilent strongly recomments to update to the new spectra/purify options.
After a new method is created based on def_lc.m and you are going to use the purify option
For the first time, the system will ask you to chose the newer version - select YES.
Methods already specified with the older spectra/purity options can be updated by
Typing following command into the command line:
	
	SPC_NewVersion=1

The system will update to the newer spectra/purify options.

Fix Information:

 No fix Planned


KPR#:453  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Peak Purity

One-line Description:

not all decimals displayed for the Volume in printout of Fraction Details

Problem:

When the Fraction Details are displayed in Data Analysis, a Print button is available to print these details. On the printout, the decimal places for the Volume are missing, although they displayed in ChemStation.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:784  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Peak Purity

One-line Description:

Purify: Only the study creator and administrators can access a study

Problem:

Only the study creator and members of the Administrators group can access a Purify study.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:977  Product:ChemStation  ChemStation (G2070BA)  (None)

Keyword: Peak Purity

One-line Description:

impure peak reported as pure under special conditions

Problem:

The results of the peak purity calculations are presented in the report in the following way:
1. The calculated values for the spectra a re presented graphically together with the threshold value. Three values above the threshold indicate an impure peak.
2. Average purity factor and threshold value based on the values that lie above the threshold. A text states whether the purity value is within the threshold.

The graphical presentation (1) is based on purity values that are not rounded, while the values described in (2) are rounded to 3 fractional digits.
In rare cases it may happen that the average of the purity values that lie above the shreshold is rounded to value equal to the threshold. In this case the graphical representation represents data points above the threshold while the text states that the values are within the limit.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:171  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Programming

One-line Description:

Uni code problem with scheduler

Problem:

Following issue has been reported using  macros to  set up  scheduler files. 

The macro is written in Unicode, but if the scheduler shows the macro, the first line is grayed out. Two extra characters  are add ( like Yp).

Temporary Solution:

There is an issue regarding the conversion between Unicode and ANSI files. In order to run the macro corret to create the scheduler files, the macro needs
to be converted to ANSI.

A tool is available on ChemStation "UnidcodeToAnsi.exe" in the Chem32\sys directory.
For more information refer to the "Upgrade preperation guide B.01.01" G2170-90222.

Fix Information:

No Fix planned


KPR#:1617  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: Programming

One-line Description:

AIC can experience communication loss if Application Experience service is running

Problem:

The likelihood of an irrecoverable communication loss between instrument, AIC and OLSS will increase, when the Application Experience service is running on the AIC.

Temporary Solution:

To increase stability of communication it is recommended to disable the Aero look on the AIC, install the SP1 for Windows Server 2008 R2 and disable the Application Experience Service 
See also:  http://support.microsoft.com/kb/902196

Fix Information:

N/A


KPR#:1755  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Programming

One-line Description:

The command "macro" does not list *.mac files

Problem:

The internal command "Macro" can be used to load a macro file from disk. Typing macro from the command line opens a dialog that points to the Chem32\Core\ folder and allows selecting *.mac and *.mcx files.
In C.01.04 *.mac files are no longer selectable.

Temporary Solution:

Use the macro name and its full location as parameter.
Example: 
Macro shutdown.mac   (equivalent to Macro C:\chem32\core\shutdown.mac)
Macro C:\chem32\core\addon\install.mac

Fix Information:

Fixed in revision C.01.05.


KPR#:1761  Product:ChemStation  ChemStation (G2070BA)  B.04.03 SP1

Keyword: Programming

One-line Description:

The command "macro " cannot be used to select *.mac files

Problem:

Occasionally the command MACRO that is used to select and load macro files from disk is not working properly.
In some casesonly *.mcx files could be loaded, but no *.mac files. There have been other cases where typing MACRO from the command line caused the ChemStation program to terminate.

Temporary Solution:

Use the command with the extended syntax. Rather than selecting the file from a dialog box, use the full path of the macro in the syntax.
Examples:
macro shutdown.mac
or
macro U:\Addons\mymacro.mac

Fix Information:

No fix planned on the B revisions.


KPR#:1831  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: Programming

One-line Description:

Scheduler does not work on an AIC

Problem:

The ChemStation scheduler cannot establish a connection with instruments on an AIC. 
When a command is processed, the message "Instrument not found" will be listed in the result column.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB revision C.01.05


KPR#:27  Product:LC and CE Drivers  RC 1200.NET  A.02.05

Keyword: Pump Configuration

One-line Description:

Misrepresentation of Bottlefillings in Dashboad of PumpValveCluster

Problem:

When a Valve is configured for Channel A in PVC the visual presentations of the BottleFillings of ChannelB, ChannelC and ChannelD are wrong.
It will not show the BottleFillings of the corresponding Channel but the BottleFilling of Channel A Valve Pos 2 etc.

Temporary Solution:

none

Fix Information:

Will be fixed with the Rc.NET Drivers Rev . A.02.06


KPR#:32  Product:LC and CE Drivers  RC 1200.NET  A.02.06

Keyword: Pump Configuration

One-line Description:

Cannot Check “Pump and solvent selection valve installed” Box for Flex Cube if Once Clear it

Problem:

LC Driver version: A.02.06 [363] (Test application/Install on Japanese ChemStation)
1. Open Instrument configuration dialog
2. Select Flexible Cube
3. Open Flexible Cube Configuration dialog
4. Clear the “Pump and Solvent selection valve installed” check box, then press OK
5. Re-open Flexible Cube Configuration dialog
6. Thick that check box and press OK
7. Open Flexible Cube Configuration dialog again -> check mark is gone

Temporary Solution:

Workaround: use autoconfig to set up the FlexCube

Fix Information:

Fix in the next rev RC.Net A.02.07.


KPR#:24  Product:LC and CE Drivers  RC 1200.NET  A.02.04 SP1

Keyword: Pump Driver

One-line Description:

Flow unit in Online Plot for Nano pump is incorrect.

Problem:

The flow unit in the Online Pot for the Nano Pump is displayed as ml/min.
The correct unit shoulf be µl/min.

Temporary Solution:

None.

Fix Information:

Fixed in driver revision A.02.04 SP1.


KPR#:43  Product:LC and CE Drivers  RC 1200.NET  A.02.04

Keyword: Pump Driver

One-line Description:

Memory leak in pump method dialog may cause performance drop

Problem:

A memory leak in pump method dialog may cause performance drop of the LC ChemStation (C.01.03 or earlier) application if methods get changed very often and/or the ChemStation instrument is operated several weeks without restarting the instrument.
The memory leak is about 1-2 MB large and this portion of memory does not get released after opening the instrument method dialog - which does always show the pump parameters, first.

Temporary Solution:

In order to avoid a performance decrease due to permanent memory allocation it is recommended to restart each ChemStation instrument after 2-3 weeks of operation.
A user macro is under development and testing to remind users of restarting a specific instrument after 2 weeks of operation.

Fix Information:

This defect is fixed in LC/CE driver version A.02.05 or higher included in OpenLAB CDS ChemStation C.01.04.
OpenLAB CDS ChemStation C.01.03 installations can be updated with update HF 08 (or higher) and LC/CE instrument drivers version A.02.06.  The Update HF 08 and  instrument drivers are available for download from https://agilent.subscribenet.com/control/agil/product?child_plneID=589223&ver=ARC.


KPR#:21  Product:ChemStation  ChemStation (G2070BA)  A.10.01

Keyword: Reports

One-line Description:

spectra options are not printed correctly

Problem:

n/a

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:60  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Reports

One-line Description:

prefix and counter are not reported in hypersequence printout

Problem:

The following defect is observed in a system with wellplate handler when running a hypersequence using prefix/counter option for the datafile subdirectories: In the hypersequence report the specified values for prefix and counter are not reported.

Temporary Solution:

n/a

Fix Information:

No fix is planned for this problem.


KPR#:75  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Reports

One-line Description:

Study Run Summary Report: WPS: Tray description is missing

Problem:

For a system with Dual Loop sampler configured to use wellplate the following problem is observed:

When a study is run using the Purify software, the Study Run Summary report does not show the tray description for the Dual Loop sampler. 

Temporary Solution:

n/a

Fix Information:

No fix is planned for this problem.


KPR#:81  Product:ChemStation  ChemStation (G2070BA)  A.10.02

Keyword: Reports

One-line Description:

Instrument curves printed twice, when set in "Sig. Details" and "inst.curves"

Problem:

Following issue has been reported on B.01.01  ChemStation printing a report including the instrument curves:
If the Instruments curves has been enabled twice, the Instrument curves will be printed twice on the reprot.

It is possible to define the "Instrument Curves" within the "Edit entire methods - Edit instrument curves" , or/and in the "Signal Details" at the 'calibration symbol' in the GUI of the method and Run Control.

Temporary Solution:

to print the instrument curves on the report, please select the option only once , 
either in "edit entire method - instrument curves"
or in " signal details" of the 'calibration symbol' in the GUI of the method and Run Control.

Once the instrument curves printing option have been set twice, the method will remain these parameters. In order to overcome this issue, the method needs to be rewritten using a method where these option not have been set twice.

Fix Information:

n/a


KPR#:82  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Reports

One-line Description:

Report: wrong behaviour in fract.table when injec.prog. longer than runtime

Problem:

In case a injector program is used it is important to have a run time specified for the method that is long enough to cover the time needed for all steps of the injector program. Otherwise problems like wrong information in the reports and runs not stopping are observed.

Temporary Solution:

Be sure that the runtime of the instrument modules is long enough to cover the whole needed to complete the injector program.

Fix Information:

There is no fix planned for this problem.


KPR#:116  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Reports

One-line Description:

time stamp issues for "inj. time" and "last changed time" for Rev.A  Data

Problem:

Data, Methods and Sequences taken from a Rev. A.xx.xx ChemStation and reprocessing them on Rev. B.01.01 ChemStation, the ChemStation reports will report different timestamps for "Injection Time" and "last changed" Time for Data Acquisition
and Data Analysis method compared to ChemStation Rev. A.xx.xx reports.

The Rev. A.xx.xx ChemStation stores times using standard C functionality that stores times as seconds since the 01.01.1970 00:00:00 GMT.
When a time (e.g. an injection time) is stored in a data file, GMT is calculated in repect to the standard time zone of the ChemStation: Pacific Standard Time.
The Rev. B.01.01 ChemStation is using the regional setting of the Windows system, there for the time stamps are updated to the local time after reprocessing.

ChemStation Plus system using the TZ variable will NOT see this issue.

Temporary Solution:

Detailed information about the time stamp issues are documented in the Readme.txt file. The Readme.txt file is present in your Chem32\Core directory.

Note: ChemStation Plus system using the TZ variable will NOT see this issue.

Fix Information:

It is not planned to fix this problem.


KPR#:164  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Reports

One-line Description:

Time Ranges not applied correctly in short report

Problem:

Within the Data Analysis of B.01.01 Chemstation following issue has been reported:
Defining User Ranges in the "Signal options" 
the Time Ranges are applied to the chromatogram
and displayed/printed accordingly, for Response ranges are not taken into account.

The issue can only be seen when 
in addition the Multi Chromatogramm Scaling is set to "each in full scale".

Temporary Solution:

Applying "all the same scale" for the Multi Chromatogramm  will fix the issue and the scaling is working correctly.

Fix Information:

n/a


KPR#:238  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Reports

One-line Description:

after manual integration and peak splitting RT and area counts double

Problem:

Using  ChemStation Data Analysis Software following issue has been reported using manual integration.

Load a datafile and integrate a peak manually. The RT and the Area (and the name) are written to the peak. Now use the manual "split peak" funktion and the RT and the Area are printed double on the screen as well on the print out.

This defect is most pronounced when you split the peak directly at the apex. The impact of this issue decrease with the distance to the apex.

Temporary Solution:

n/a

Fix Information:

No fix is planned for this issue.


KPR#:240  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Reports

One-line Description:

"Print all Windows" function is not printing correctly

Problem:

Following issue has been reported for the Spectra Part of the Data Analysis  in ChemStation by using the "print all window" funktion:

If a spectra and referecnce is called in on screen
the system will either not print all windows or is overlaying them. The issue depends on the print driver in use.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:281  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Reports

One-line Description:

print to file: no csv file generated using performance and noise report style

Problem:

Using the "Performance+Noise" report style it is observed that .csv reports are not created when they are selected in the "Specify Report" dialog.

Temporary Solution:

n/a

Fix Information:

No fix is planned for this problem.


KPR#:283  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Reports

One-line Description:

No report shown after method run

Problem:

If in the "Signal Details" dialog of the Data Analysis part of method no detector signals are selected, but only non-detecor signals like pump flow, no report will be created when running this method.

Temporary Solution:

There is a macro available in the LC User Contributed Libraries that allows to only save the instrument curves. It is macro #14 on the ChemStation B.02.01 SR1 CD.

Fix Information:

No fix planned.


KPR#:285  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Reports

One-line Description:

XLS Performance and Noise Rep. : drift given in  mAU/min

Problem:

Following issue has been reported printing the "Performance + Noise"  report to XLS files:

In Excel (sheet "noise") the drift is shown in mAU/min instead of mAU/h. 


Temporary Solution:

n/a

Fix Information:

No fix is planned for this issue.


KPR#:315  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Reports

One-line Description:

Long FIA runs does not list all the injections in the report.

Problem:

For example, in a run of 750 injections, the report lists only 400 flow injections with Inj , InjTime, Location and FIA Sample Name.
After the 400th line it just displays a line in bracket as    "…(total number of flow injections is 750)" .
i.e it does not list the details of remaining 350 flow injections.

Note that it does complete the 750 actual flow injections, but in the report it LISTS only 400 flow injections.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:316  Product:ChemStation  ChemStation (G2070BA)  (None)

Keyword: Reports

One-line Description:

No report due to "Quantification out of memory" command line error for FIA run.

Problem:

If you run a very long FIA run and then request a report or preview, the report is neither generated for preview nor gets printed. Instead it gives a command line error in red "Quantification out of memory".

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:328  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Reports

One-line Description:

in the method printout two DADs have the same header

Problem:

In a system with a DAD G1315C and an additional DAD G1315A or G1315B the following issue was observed in the method printout:
The sections about the detectors both have the header "Agilent 1100 Diode Array Detector 1".

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:349  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Reports

One-line Description:

"Use Multiplier and Dilution Factor with ISTDs" displayed in all reports

Problem:

The message about the usage of multiplier and dilution with ISTDs as specified in the Calibration Settings dialog is displayed in all kind of reports, but it should only be displayed in the relevant reports (i.e. ISTD and ISTD%).

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:373  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Reports

One-line Description:

Printing problems: black bars on the report

Problem:

Using B.01.03 or higher customers reporting black bars
on their print reports, the report on the screen is correct.

The problem only happens using
1. single signal
2. zoomed in 
3. zoomed in that one peak in the window is cut off,  and the peak is huge compared to the others

Depening how deep the zoom is, the problem might appear on the report.

Temporary Solution:

The problem is printer depended, PCL6 is giving problems. Please use either PLC 5e or the Postscript driver.

Fix Information:

n/a


KPR#:420  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Reports

One-line Description:

Batch Report: "justify" for alignment is not taken in report/preview

Problem:

Using the ChemStation batch function, the system
gives you the possibility to adjust the batch report by:
open th "options for Batch Review", "Report Table" tab,
and define the "justify" for each parameter.

This alignement is not applied to the preview/report.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:449  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Reports

One-line Description:

Group Summary report missing in CSV reporting

Problem:

If you perform peak grouping, the Group Summary report is missing when the report is printed to .csv file.

Temporary Solution:

n/a

Fix Information:

No fix is planned for this problem.


KPR#:450  Product:ChemStation  ChemStation (G2070BA)  A.10.01

Keyword: Reports

One-line Description:

Batch Report: duplicated results for multiple signals in Cal table

Problem:

If you setup an calibration table for multiple signals and treat the signals separately,  the batch output reports the multiple peaks of the same name,  but the reported data (RT and Amount) for each peak are duplicates of the first occurance of that peak.

Example Calibration table:
#    RT      Signals     Compound    Amount    Area     Respf 
1    1.001   DAD1 A   Peak1           2.000       200      0.0100
2    1.001   DAD1 B   Peak1           2.000       400      0.0050
3    2.001   DAD1 A   Peak2           2.000       300      0.0066
4    2.001   DAD1 B   Peak2           2.000       350      0.0057

Batch report:
the reported RT and Amount for Compound nr 2 and 4 are respectively a duplicate of nr 1 an

Temporary Solution:

Temporary Solution Notes:
This option is only used in case the user preferres are confirmation result rather than a qualifier ratio.
To get the confirmation result calculated in the batch table it's recommended to name the compounds different for each wavelength. (e.g. by adding the wavelength information in brackets).
This helps as well to differenciate the results on the batch report.

Fix Information:

 No fix Planned


KPR#:458  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Reports

One-line Description:

Wrong unit printed on Delay Calibration Reprot

Problem:

Using ChemStation REv. B.02.10 following issue has been
seen printing a Delay Calibration Report.

The printed report will state the "current delay" in time(sec): xx  instead of volume ul

Temporary Solution:

n/a

Fix Information:

No fix is planned for this problem.


KPR#:459  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Reports

One-line Description:

The use of the manual injector not reported in method report

Problem:

Following issue has been seen printing a method using a manual injetor. The method print out does not mention the use of the manual injector.

Temporary Solution:

n/a

Fix Information:

No fix is planned for this problem.


KPR#:464  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Reports

One-line Description:

page numbers on diagnostic reports are cut of on print out

Problem:

Following issue has been reported printing diagnostic reports: The page numbers (page 1 of ***) are cut off on the right side in the printouts. The "total numbers of pages" are not completetly printed.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:490  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Reports

One-line Description:

Chromatogram might not printed like defined in Graphics

Problem:

Following issue has been seen on ChemStation B.02.01 using the option "User Range". Only if the multi-chromatogram layout is select to be "separately", the Time axis is used on the printed report, but not the ranges.

Temporary Solution:

‘Each in same scale’ allows comparison of identical graphs and uses the Response Range. 

If using “Each in full scale’, it ignores the Response range setting and only uses the Time.

Fix Information:

n/a


KPR#:533  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Reports

One-line Description:

number of decimals for Capillary and Nano Pump flow in method printout

Problem:

With a Nano Pump (G2226A) or a Capillary Pump (G1376A), only two decimals can be configured for the pump flow. However, the method printout will show a third decimal that is 0.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:590  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Reports

One-line Description:

Margins are used differently for the Print out of ChemStation Windows Printing

Problem:

Following printing problem has been reported using ChemStation Rev. B using the "page set margins" for the manual printout in File:

If in data analysis a data file is selected and a "page set margin" is set, the margin is applied printing "all windows", but the bottom page margin is ignored for "print selected window".


 

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:722  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Reports

One-line Description:

Path naming on Standard Reports and Ex. Perf. Report not giving ECM info

Problem:

Following issue using ChemStation B.02.01 SR 1/SR2 with ECM integration has been reported: Printing a report using the extended Performance Report, as well on Customized Reports, the path information should state ECM as path information.

Temporary Solution:

A point patch will be available, please contact your Agilent Representative.

Fix Information:

 No fix Planned


KPR#:748  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Reports

One-line Description:

Report Layout: Footer print date earlier than last change date for calibrations

Problem:

Following problem has been reported for reports, genenrated with the Report Layout using ChemStation:
Performing a sequence, including calibration samples, the time stamp "last changed" time for the Data Analysis method will be 2-3 seconds earlier thant the footer print time stamp.

Temporary Solution:

n/a

Fix Information:

The print handling of the report is a taske, using multi-processes which are getting different time stamps. Therefore when the report is set together, it is possible that the header/footer information get a different timestamp that than the chromatogram/compound section.


KPR#:755  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Reports

One-line Description:

7683 autosampler appears as 7673 on the ChemStation GLP report & meth print

Problem:

The 7683 autosampler is installed on the GC but it shows up as a 7673 on the Chemstation GLP report and when the method is printed.

Temporary Solution:

The 7673 is the old family name and covers all of the Autosampler components including the various controllers, multiple different injector modules, and the tray(s).

Fix Information:

n/a


KPR#:776  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.01

Keyword: Reports

One-line Description:

Report including Calibration Curves displays usage of height instead of area

Problem:

If a report style is used that includes calibration curves, in the formula the report indicates the the Y axis represents Height although calculation based on Area has been selected.
The values of the parameters itself and all the calculations are correctly based on Area.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:783  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Reports

One-line Description:

problems with chromatogram time and response axes in customized reports

Problem:

The following problem is observed with customized report templates that contain chromatogram in the general section:
When defining to have the time axis at the top or the response axis at the right side of the chromatograms the numbers are displayed inside the chromatogram area.
Also when having the time axis at the top of the chromatogram, the signal description is displayed at the bottom and is upside down. The retention times are also turned 180 degrees in the latter case.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:811  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Reports

One-line Description:

A report cannot be printed if you have overlaid signals

Problem:

The dual injector GC gives two data files, one with ADC-1 and AIB-1, and one with ADC-2 and AIB-2.  It is not possible to overlay the signals and print them.  The following message is given: “A report cannot be printed if you have overlaid signals.  Use Load Signals and try again.”  

Temporary Solution:

After selecting the signals you want to overlay, go to FILE, select PRINT, then select “Selected Window”.

Fix Information:

 No fix Planned


KPR#:812  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR2

Keyword: Reports

One-line Description:

7673 AutoSampler is printed instead of HeadSpace for Run Info in ChemStore

Problem:

Information in ChemStore.  The HS Sampler and its serial number information are not being displayed.
Headspace is shown as a 7673 AutoSampler.  

Headspace runs correctly and puts the data in the study, but it does not include the G1888.

This was found using G2070BA B.02.01 SR2 with ChemStore/Security Pack B.03.03 SR1.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:864  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: Reports

One-line Description:

Sequence Summary report does not include peaks that are grouped

Problem:

Two peaks are grouped together.  The individual reports have the peaks grouped, but the peaks are not grouped in the Sequence Summary Report.

Temporary Solution:

n/a

Fix Information:

No fix Planned


KPR#:895  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: Reports

One-line Description:

With HP LJ 4250 selection of output orientation does not work

Problem:

The selection of printout's orientation from ChemStation by File -> Printer Setup does not have the desired effect. 

Instead the setting determined at the respective printer's perferences (e.g. by Window’s Start -> Settings -> Printers and Faxes) does rule the printers behaviour.

If e.g. portrait is selected there than every printout is portrait. This is independent of a landscape orientaion selected in ChemStation by File -> Printer Setup! The opposite (landscape stays landscape in spite of portrait selected in ChemStation) is true, too.

This behavior was observed with recent PLC5 or PCL6 printer drivers for HP LJ4250 obtained from www.hp.com.

Temporary Solution:

Utilizing the PCL5 printer driver deliverd with ChemStation on CD 2 (User Documentation and Accessories) under \Support\Printers\English\lj4050_pcl5e as lj1130en.exe solves the problem.

Installation of that printer driver can be checked from Control Panel -> System -> Device Manager by navigating to the respective communication port used for the LaserJet 4250. The Driver Details show Mirocosoft as Driver Provider, a Driver Date: 7/1/2001 and a Driver Version: 5.1.2600.0

Fix Information:

n/a


KPR#:897  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Reports

One-line Description:

Not found peaks reported different in different report styles

Problem:

The way a peak that is not found is reported is different in different report styles.

Built in Styles:
short: "-"
Full: "-"
Detail: "-"
Performance: "Not found"

Report Layout Templates:
EXTSTD.FRP: "0"
LIBRARY.FRP: "-"
AreaPCT.FRP: "0"
SIMPLE.FRP: "0"

Temporary Solution:

For the Report Layout templates this can easily be changed by editing the templates.

Fix Information:

It is not planned to fix this issue.


KPR#:915  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: Reports

One-line Description:

7890A Report heading does not include Injection Volume; always 'External'

Problem:

The 7890A Report heading always list the Injection Volume as 'External'.  The amount injected is correct.

Temporary Solution:

n/a

Fix Information:

Fixed in Revision B.04.01.


KPR#:942  Product:ChemStation  ChemStation (G2070BA)  B.03.01 SR1

Keyword: Reports

One-line Description:

additional column "ISTD used" has no header in ISTD report of CE ChemStation

Problem:

When ISTD report is created, the report will provide the information which ISTD is used for each reported compound.
In CE ChemStation, this value is included in the result table, but it is displayed without the corresponding header.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:949  Product:ChemStation  ChemStation (G2070BA)  (None)

Keyword: Reports

One-line Description:

wrong compound numbering in Extended Performance Report after removing compounds

Problem:

Compouinds that were once in the calibration table of a method, but have been deleted are still taken into account for numbering in the Extended Performance Report. This is only true for the section of the report that reports the performance parameters for the individual peaks.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:953  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: Reports

One-line Description:

Japanese/Chinese CS: garbage characters in .DIF output of sequence summary

Problem:

When printing the sequence summary report, it is possible to select file output as .dif file. When opening this file with e.g. Microsoft Word, Japanese and Chinese charcters will not be displayed characters.

Temporary Solution:

The defect is caused by the fact that the .dif file is stored in ANSI format and thus cannot handle Japanese and Chinese characters.
In order to view the content of the .dif file, open it with an application that is able to read ANSI files, e.g. Notepad.

Fix Information:

No fix is planned.


KPR#:996  Product:ChemStation  ChemStation (G2070BA)  A.10.02

Keyword: Reports

One-line Description:

inconsistent results in statistics report

Problem:

The following problem is observed for calibration tables with separate calculation for each signal. If several peaks have the same name, the sequence summary report may calculate the statistics from peaks in different signals and summarize them under the header of the first signal.

Temporary Solution:

If peaks with the same retention time in different signal have the same compound name, separate calculation for the signals has to be disabled.

Fix Information:

n/a


KPR#:1011  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.02

Keyword: Reports

One-line Description:

ChemStation does not report resolution as defined in European Pharmacopoeia

Problem:

ChemStation does not report resolution as defined in European Pharmacopoeia. EP requires the following formula to be used:
R = 1.18 * (Tr(b)-Tr(a)) / W50(b)+W50(a)


Only the following definitions are available in ChemStation for resolution:

USP, ASTM (in Extended Performance Report):
R = 2 * (Tr(b)-Tr(a)) / WB(b)+WB(a)

Half-width method as used in Performance Report:
R = (2.35/2) * (Tr(b)-Tr(a)) / W50(b)+W50(a)

JP (in Report Layout view only):
R = 1.18 * (Tr(b)-Tr(a)) / W50(b)+W50(a)

Temporary Solution:

In Report Layout view, use Resolution (JP) which uses the same formula.

Fix Information:

Fixed in revision C.01.04


KPR#:1108  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: Reports

One-line Description:

6890N Full report displays 7683 ALS incorrectly as 7673

Problem:

6890N.05.05 with 7683B ALS installed in back position.

Load a signal produced on this GC and specified 'Full' report.

Print Report.

The 7683B ALS is incorrectly interpreted as a 7673 with a Firmware of G1530N.05.05 as follows:

=======================================================
Module                                 Type   Firmware rev.   Serial number
--------------------------------------|------|---------------|---------------
68GC                                   68GC   N.05.05         CN10429002     
7673                                   7673   G1530N.05.05                   

A 6890A with two 7683A injectors installed prints this as '7683/G2912A.02.01'

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1109  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Reports

One-line Description:

ChemStation unable to print to screen; Failure at end of run with Vista SP1

Problem:

Running with no printer configured except Nova pdf and printing to screen.  The OS is Vista SP1.

At end of run, software hangs with pop-up:
'Due to a printer configuration failure, the ChemStation is unable to print to screen!'

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1302  Product:ChemStation  ChemStation (G2070BA)  B.04.01 SP1

Keyword: Reports

One-line Description:

problem with page numbering in a long sequence summary report

Problem:

In case of a very long sequence summary report (more than 1000 pages), the page numbers for the included Analysis reports in the Summary section are displayed for any page number higher than 999.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1305  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.01

Keyword: Reports

One-line Description:

Group sum results are not exported to the result.xml

Problem:

The group results in the normalized percent report are not exported to the result.xml, see attached JPG file. 

This problem occurs in 7890 or 68xx.

Temporary Solution:

Individual amounts and group membership are part of the acaml file. Including the sum as a separate item would be redundant information.

Fix Information:

No change planned.


KPR#:1348  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: Reports

One-line Description:

"View Method" does not report sample related custom fields

Problem:

When reviewing the method parameters for run via "View Method" in Data Analysis view, the sample-related custom fields are not included in the report.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:1355  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Reports

One-line Description:

Grouped calculations/sums will only work for reports using calibration curves

Problem:

The grouped calculations/sums will only work for reports using calibration curves.   They will not work in area percent mode.  Even when working with Chem2XL and try to bring over area percent and compound group information, the group information is not available.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1435  Product:ChemStation  ChemStation (G2070BA)  B.04.02 Aladdin

Keyword: Reports

One-line Description:

Sample Report is truncating long barcodes to 16 characters

Problem:

7890A GC with 7693A ALS

Sequence table displays and reads valid and lengthy barcodes correctly.  However, the report heading is truncating these to 16 characters.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1440  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.01 SP1

Keyword: Reports

One-line Description:

acquisition parameters missing in the Method section of the sequence summary report when Unique Folder Creation is switched off

Problem:

The acquisition parameters missing in the Method section of the sequence summary report when Unique Folder Creation is switched off. Only the run-time checklist, method information and data analysis parameters are reported.

Temporary Solution:

n/a

Fix Information:

Fixed in revision C.01.04


KPR#:1509  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 SP1

Keyword: Reports

One-line Description:

Paper Source selection is not kept

Problem:

In the Printer Setup in ChemStation, it is possible to select the Paper Source. However, this selection is not kept when pressing OK, it always switches back to the default setting.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1531  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: Reports

One-line Description:

.pdf report misaligned when default font for the report is missing

Problem:

ChemStation uses the font Consolas for the .pdf reports (Specify Report / Destination: File / .PDF). If this font is not present on the computer, the report will use another font and the report will be misaligned.

Temporary Solution:

Change the font for the .pdf reports to use a font that is installed, e.g. Lucida Console:
1. Close all ChemStation session.
2. Open the file C:\WINDOWS\CHEMSTATION.ini with Notepad.
3. In the [PCS] section, identify the line
pdf_ReportFontName=Consolas
4. Change the entry to read
pdf_ReportFontName=Lucida Console
5. Save the changed CHEMSTATION.ini file.

Fix Information:

It is not planned to change the current default font for the .pdf reports.


KPR#:1563  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 SP1

Keyword: Reports

One-line Description:

problem when printing more than 8 overlaid chromatograms via Print Selected Window

Problem:

When more than 8 signals are overlaid in one chromatogram window, the user zooms into the window and then the window is printed by the option print Selected Window, the chromatograms will not be printed on one single page, but only seven of them will be on each page of the printout.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1572  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: Reports

One-line Description:

Batch mode does not allow usage of intelligent reports

Problem:

If user works in batch mode and the batch method is setup to use intelligent reports, this selection is ignored, even if the batch options are set to use the batch method s report settings.

Temporary Solution:

N/A

Fix Information:

There is no plan to fix this issue in future.


KPR#:1579  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: Reports

One-line Description:

Report Templates newly added to a Result Set do not appear in the list of available templates

Problem:

When you add more than one report template to a result set, only the seletcted template in the browse dialog will be visible in the Specify Report list box under Report Templates or Report Templates for calibration samples

Temporary Solution:

After closing and reopening the Specify Report menu, all Report Templates added to the result set are visible.

Fix Information:

n/a


KPR#:1582  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: Reports

One-line Description:

Mismatch in drawing of peak performance between Classic Report and Intelligent Report

Problem:

The peak performance drawing in the Classic Reporting and in the Intelligent Reporting are slightly different. Thisi is only a graphical representation and this is not affecting the performance

Temporary Solution:

n/a

Fix Information:

 See Previous Note.


KPR#:1619  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: Reports

One-line Description:

Intelligent Reports do not contain ECM specific information in Review mode

Problem:

ECM specific fields like path, server, operator or version number are not printed in the Intelligent Report templates in the Review mode.

Temporary Solution:

Print the report from the Data Analysis view.

Fix Information:

No fix available.


KPR#:1634  Product:ChemStation  ChemStation (G2070BA)  B.04.03 GC DSP

Keyword: Reports

One-line Description:

B.04.03 DSP1 [54] Data Analysis, Graphics, response range does not work

Problem:

B.04.03, data analysis, load a data signal  (demo\100FID20 for example) and select Graphics. Select Use Ranges and set the time range to 0 to 6 minutes and the response range to 0 to 100 pA for example, then OK.  Time scale works (6 minutes) but the response range does not work.

Temporary Solution:

WORKAROUND 
Even though this is a single chromatogram, in Multi Chromatogram, LAYOUT select separated, and SCALE, select All the same scale and it works.

Fix Information:

 Fixed in revision B.04.03 SP1.


KPR#:1661  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02

Keyword: Reports

One-line Description:

USP tailing factor can be reported incorrectly with a value <0.

Problem:

The USP tailing factor must by definition result in a positive value. There are cases where the ChemStation report results in a negative value. These cases are linked to peaks where the integration baseline is drawn in a way where a peak width at 5% height is not well defined. This width is needed for the calculation.

Temporary Solution:

A tailing factor with a reported negative value should be interpreted as: "USP tailing not calculated"

In case of a customized report this can be automated by modifying the macro in the shortutl.mac.

Replace the call for the macro USR_CalcPerfShort$
with the following lines:

!*************************
name USR_CalcPerfShort$
!*************************
local counter
  if (Check(MACRO,evaluatePerformance) = 0)
	macro "report.mcx"
  endif
  evalperf  ,1
for counter =1 to  tabhdrval(chromres,peak,numberofrows)
 if tabval(chromres,peak,counter,tailfact)<=0
  settabval chromres,peak,counter,tailfact,-1
 endif
next counter
  return ""	
endmacro

Fix Information:

Resolved in B.04.03 DSP1.
Resolved in C.01.04


KPR#:1669  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: Reports

One-line Description:

ACAML: GC injection volume not correctly given; it is showing more significant figures

Problem:

Data created using dual tower 7890A GC, the injection volume defined in the method is not correct in the acaml file.

In the ChemStation report, the injection volume is 1ul but in the ACAML file, it is 0.99999998 ul.  The same is true for 3,4,5 - the ACAML file is always
showing more significant figures. (Injection volume 2 is fine)

Temporary Solution:

n/a

Fix Information:

Fixed in Revision B.04.01.


KPR#:1673  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: Reports

One-line Description:

CSV, DIF, and EMF files are not generated for some Classic report styles

Problem:

The CSV, DIF, and EMF files are not generated for some Classic report styles:
CSV and DIF files are not created for Library Search and Performance related report styles.
For the Extended Performance report style, also the EMF report is not created.

Temporary Solution:

n/a

Fix Information:

It is not planned to fix this issue.


KPR#:1679  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: Reports

One-line Description:

sample name not reported with Intelligent Reporting in non-result set data that have been acquired prior to C.01.01

Problem:

The sample name is not reported when using Intelligent Reporting when working on data that have been required in revisions prior to OpenLAB CDS ChemStation Edition C.01.01.

Temporary Solution:

Use the "result set migration" tool of the Chemstation, to migrate the sequence to container format.

Fix Information:

 Fixed in revision C.01.04


KPR#:1692  Product:ChemStation  ChemStation (G2070BA)  A.08.03

Keyword: Reports

One-line Description:

Date format is changed in sequence summary report from month/day/year to day/month/year

Problem:

The date format is changed from month/day/year to day/month/year intermittently during long sequences.

140 Vial 140 1 04.12.2011 22:29:00 apr110000140.D Rigi

141 Vial 141 1 13.04.2011 00:29:00 apr110000141.D Rigi
142 Vial 142 1 13.04.2011 02:28:59 apr110000142.D Rigi

The data files have dates of the form DD.MM.YYYY, chemstation tries to parse this as MM.DD.YYYY, the chemstation then prints the date as part of the report. If the parsing fails then the string stored in the data file is used

12.04.2011 -> Parsed as December 4, 2011 -> Printed in local time format as 04.12.11

13.04.2011 -> Fails the parse, string is just printed.

Temporary Solution:

n/a

Fix Information:

Fixed in B.01.01.

Data from the revision A software will still show the problem in the B.0x.0x installation but all new data created will not have this problem.  The month is parsed properly beginning in B.0x.0x.


KPR#:1703  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: Reports

One-line Description:

ChemMain.exe can crash when a long file path is used

Problem:

The file path in ChemStation has a hard coded limit of 256 characters.
In an ECM environment there are also hidden characters added to ensure unique file naming. As a consequence the complete lenth for  +  +  +  must not exceed a length of 175 characters.
Since revisions C.01.06 HF04 and C.01.07 SR1 also the data file names within a sequence can be composed from tokens.

Temporary Solution:

Make sure that the path length of a report from  -->  C:\Chem32\... to  ...pdf  <-- does not exceed 175 characters.

Fix Information:

No fix available.


KPR#:1705  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: Reports

One-line Description:

Chinese/Japanese characters cannot be searched in the Report Viewer

Problem:

The report viewer that was introduced with C.01.03 allows searching for text within a report.
Chinese and Japanese characters, however, cannot be searched.

Temporary Solution:

No fix available.

Fix Information:

n/a


KPR#:1711  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 SP1

Keyword: Reports

One-line Description:

The legend for calibration curves does not react on the calculation type

Problem:

The units of the axes of a calibration curve plot depend on the calculation type. For ISTD and ISTD% calculations the units are Area Ratio and Amount Ratio respectively. In all other cases it is Area and Amount, respectively. The labelling of the curve plot reacts on a change of the calculation type.
In some reports the calibration curve is printed together with the curve's equation and a legend. The legend description is always Area and Amount, independent of the calculation type. The affected report styles are Detail, GLP+Detail, Detail+Spectrum, Full and the Calibration Reports.

Temporary Solution:

Ignore the Legend for X: and Y:
They are redundant. The description in the curve plot provides the correct information.

Fix Information:

No fix planned.


KPR#:1716  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR2

Keyword: Reports

One-line Description:

Report information such as chromatogram and results are truncated or cut off when saving to htm

Problem:

Report graphics file gets truncated or cut off.  The report is missing part of the chromatogram and results.
May have something to do with the printer driver.

Temporary Solution:

n/a

Fix Information:

Fixed in Revision B.03.01.


KPR#:1722  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: Reports

One-line Description:

Tangent peak width not correctly evaluated.

Problem:

The peak performance parameters have been modified in ChemStation revision B.04.03 to provide better peak width evaluation for chromatograms with high data rates.
This new algorithm sometimes fails to find the correct inflection points. As a result peak width at baseline (tangent peak widths) can be reported incorrectly. Consequently parameters calculated based on this peak width (tangent resolution and tangent plates) are incorrect as well.

Temporary Solution:

A patch for B.04.03 is available on request.

Fix Information:

Fixed in revision B.04.03 SP1.


KPR#:1733  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Reports

One-line Description:

Mass- Based Purification (Fraction Collection) report missing information & shows inaccurate Fraction Tick marks (RC.NET only)

Problem:

The Mass-Based Purification (Fraction Collection) report is missing information in RC.NET mode; this same information is available in Classic mode.

The "Mass" column in the report should specify which detected mass(es) triggered the fraction collection; this feature is currently available in Classic mode only.

Additionally, the Fraction Tick Marks displayed are incorrect as is the BeginTime & EndTime information in the report Fraction Information section. The Fraction Tick Marks do not line up accurately with the collected fractions triggered by the LC/MS; these values have the MS Delay Time added to them, which is incorrect.

Temporary Solution:

To determine which masses triggered Fraction Collection or what the actual Fraction Collection BeginTime and EndTime values are, open the affected  ChemStation data file (.D folder) and open the file MBFC.ini.  This MBFC.INI file contains the Mass-Based Fraction Collection triggering events, including Mass, Collection ON, and Collection OFF information.  Do not modify this file.

The format of MBFC.INI must be interpreted linearly.

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1751  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Reports

One-line Description:

eSignature data file privilege is missing from the ChemStation lab manager, analyst, and operator roles

Problem:

ChemStation / Data Store integration:  the "eSignature data file" privilege is missing from the ChemStation lab manager, ChemStation analyst, and ChemStation operator roles.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1754  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Reports

One-line Description:

The structure of the reports to file has changed

Problem:

The structure of the reports in xls, csv and dif format has changed.
New columns have been introduced to hold additional information provided since C.01.04.

Temporary Solution:

Users who perform their own spreadsheet calculations based on those reports need to be aware that the order and content of columns is different.

Fix Information:

N/A


KPR#:1756  Product:ChemStation  ChemStation (G2070BA)  B.04.03 SP1

Keyword: Reports

One-line Description:

CP3: Number format error in: %d when printing report to screen in Data Analysis

Problem:

Completed a single run and a report was printed to screen.

Loaded a different method in DA
Select Report > Print Report
No report is printed to screen
Error shown in left hand corner: CP3: Number format error in: %d

Temporary Solution:

The problem was resolved by doing a clean uninstall and new installation.

Fix Information:

Not reproducible.


KPR#:1757  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Reports

One-line Description:

Error 41265 occurred at end of run; Number format error in: %d shown in logbook

Problem:

Method run completed. The following error message appeared at the end of the run in the Logbook.

CP Macro     ========================================       14:46:33 02/13/12
  70 41e0 4f396899    0
CP Macro     In Command sprintf$ in macro GetRunPdfRptFil>  14:46:33 02/13/12
  70 41e0 4f396899    0
CP Macro     eNameRoot$                                     14:46:33 02/13/12
  70 41e0 4f396899    0
CP Macro     at line 196 in file C:\CHEM32\CORE\REPORT.MC>  14:46:33 02/13/12
  70 41e0 4f396899    0
CP Macro     X                                              14:46:33 02/13/12
  70 41e0 4f396899    0
CP Macro     Error 41265 occurred:                          14:46:33 02/13/12
  70 41e0 4f396899    0
CP Macro     Number format error in: %d                     14:46:33 02/13/12
  70 41e0 4f396899    0
CP Macro     ========================================       14:46:33 02/13/12

Default report setting was used.

Temporary Solution:

The problem was resolved by doing a clean uninstall and new installation.

Fix Information:

Not reproducible.


KPR#:1775  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: Reports

One-line Description:

Preview of classic reports fails with a macro error. 'Number Format error' in Sequence log file.

Problem:

Doing preview of a report with the method settings to screen only works fine for 25 days. After that, the number used in the filename of  the temporary.pdf report file shown in the report preview is getting larger than MAX int (2.147.483.647) and causes a formatting error during the conversion to a string value. After again about 25 days it starts over with 0 and works again for 25 days.

Temporary Solution:

The number used for the temp file name is the windows OS tick count, and it gets larger than MAX int after 24 days. It is reset after every reboot.

Fix Information:

This issue is fixed with C.01.03 [45] HF06. Fixed with C.01.05. Fixed with B.04.03 SP2.


KPR#:1776  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: Reports

One-line Description:

Chromatogram in On-Screen Classic report is not displayed properly

Problem:

On systems accessed via  Remote Desktop, the on screen classic report may not be displayed properly. Part of the baseline, x-axis or y-axis labels may be missing. This is only affecting the chromatogram in the on-screen classic report. The printed report or display in the data analysis view will correctly show the baseline, axis markers and labels.

Affected revisions: C.01.03 and C.01.04.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB revision C.01.05.


KPR#:1778  Product:ChemStation  ChemStation (G2070BA)  B.04.03 SP1

Keyword: Reports

One-line Description:

'Number Format error' in Sequence log file after printing report to screen

Problem:

An error message appears when a report  is printed to screen on computers that have not been rebooted for more than 24 days.

Temporary Solution:

For the OpenLAB CDS ChemStation C.01.03 a hotfix 06 is available.
For the OpenLAB CDS ChemStation C.01.04 a hotfix 04 is available.

Fix Information:

Fixed in B.04.03 SP2. Fixed in C.01.05.


KPR#:1780  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Reports

One-line Description:

Sequence summary report sorts the data by injection time

Problem:

In case of self-assembled result sets, the order of acquisition sequence lines numbers might appear in a random order. This is caused by the default sorting criteria, which is the injection time.

Temporary Solution:

Customize the report template and adjust the sorting criteria to your needs and preferences.
In case of self-assembled result sets, the e.g. processing time stamp is a suitable sorting criteria.

Fix Information:

Revision C.01.04 introduced advanced sorting functions.


KPR#:1786  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Reports

One-line Description:

A Sequence Summary Report after Partial Sequence reports all the runs of the full sequence

Problem:

In IR Sequence Summary reports always the full sequence is reported, even in partial sequence.
Classic SSR reacts on the selection of partial sequence.

Temporary Solution:

To have only a subset of the sequence in the IR use the Review mode or create a self-assembled result set.

Fix Information:

n/a


KPR#:1791  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Reports

One-line Description:

Macro error occurs (Parameter 2: Table not found) when printing a calibration table

Problem:

Macro error occurs (Parameter 2: Table not found) when printing a calibration table

Temporary Solution:

n/a

Fix Information:

Fixed in Hotfix03 for C.01.04.


KPR#:1792  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Reports

One-line Description:

LCMS specific report style options are missing from the Report dialog (Classic Reports)

Problem:

C.01.04, Report Style LCMS and LCMS Qualitative:  Missing Report Style Options...:

-Background Subtract
-Spectrum Average Peak
-EIC

Temporary Solution:

n/a

Fix Information:

Fixed in C.01.04 Hotfix03.


KPR#:1801  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: Reports

One-line Description:

C.01.03: XML results: The 'ESTD %' and  'ISTD %' result types in the Quant Calc node are incorrectly 'ESTD' and 'ISTD'

Problem:

ChemStation C.01.03: XML results:  The 'ESTD %' and  'ISTD %' result types in the Quant Calc node are incorrectly reported as 'ESTD' and 'ISTD'. This is confusing the receiving LIMS system. The printed reports have correct titles.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1803  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Reports

One-line Description:

Acquisition method name is not displayed in an intelligent report

Problem:

Acquisition method name is not displayed in an intelligent report

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1820  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Reports

One-line Description:

Missing Report Style Options for LCMS and LCMS Qualitative

Problem:

Report Style LCMS and LCMS Qualitative:  Missing Report Style Options:

-Background Subtract
-Spectrum Average Peak
-EIC

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1835  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Reports

One-line Description:

Calibration data modification time stamp is set-off by 12 hours

Problem:

The Short_ESTD.RDL report template (and all related rdl templates) contains a data field Calibration data modification time.
The time stamp always shows the appendix AM, even in case recalibrations processed in the afternoon.
Switching the time format to 24:00 does not correct the problem.

Temporary Solution:

N/A

Fix Information:

This defect if fixed in HF 7 for  OpenLAB CDS ChemStation C.01.03 and OpenLAB CDS ChemStation C.01.04 and higher.

OpenLAB CDS ChemStation C.01.04 also introduced a new compound specific and more granular time stamp  "CalibrationCurveLastModifiedDate". This time stamp gets stored in the result file and remains with the result. Report templates need to be adjusted in order to print this new time stamp.


KPR#:1839  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Reports

One-line Description:

File corruption may cause sequence runs appearing as single runs

Problem:

Under specific circumstance sequence runs may appear as single runs on OpenLAB CDS Intelligent Reports. Sequence related information does not appear on the report and multi-injections appear as single injections.
The sequence context of an analysis run is stored in .b file inside the result set. In case of a file corruption the .b file is incomplete and the affected run gets a sample.acaml result file attached that indicates the run as a single run.

Temporary Solution:

To prevent file corruption:
- Exclude the \chem32 from active virus protection
- Check the computer harddisk for errors and repair/replace the disk in case of errors.

To correct the mentioned error:
- delete the sample.acaml file inside the .D folder of the affected run
- reprocess the sequence to re-generate a consistent .B file.

Fix Information:

With introduction of rev. C.01.05 of OpenLAB CDS ChemStation it is sufficient to reprocess the sequence in order to re-generate a consistent .B file and to replace the sample.acaml result file by a sequence.acaml result file.
With C.01.03 HF07 installed, it is sufficient to reprocess the sequence in order to re-generate a consistent .B file and to replace the sample.acaml result file by a sequence.acaml result file. HF 07 does not actively prevent file corruption.


KPR#:1974  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.06

Keyword: Reports

One-line Description:

Cannot change the Report Name for the Signal in the Classic Sequence Summary Report

Problem:

Select Sequence Parameters\Sequence Output and change the report names for both Signal 1 and Signal 2. Select OK.  Save and reopen. Signal 1 was not updated to the new report name and was changed back to the default GLPrptF.txt.

Temporary Solution:

n/a

Fix Information:

No fix planned.


KPR#:2119  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.06

Keyword: Reports

One-line Description:

Single Peak's Peak area is different from Total area for classical built in report

Problem:

In some situations OpenLAB CDS "internally" may calculate with more digits than visible on the GUI or listed on the final report. In some cases the number of digit for rounding a final total may even defer from the digit of those during previous operations. As a result the real sum of the visible single numbers on a report may differ from the result shown at the end. This could even result in a difference between e.g. the area “sum” of a single value (one peak only) and it´s “total”. This is not a OpenCDS ChemStation malfunction but a mathematical effect that may occur in every software that internally calculates more exactly than visible on a GUI or report for more convenient readability. This effect is typically limited on the very last digit. 

Example:
10.123507893450 + 
10.000569993551 = 
20.124077887001

If only 5 digits are displayed (cut off after the 5th digit) this would result in:
10.12350+
10.00056=
20.12407

If the result would be rounded a the 6th digit but the summands only at the 12th but all values are shown with 5 digits it would even be:
10.12350+
10.00056=
20.12408

A further issue is the fact that computers internally use binary formats for numbers which have to be transferred to decimal values. This brings some additionals issues (google for: Floating Point Numbers, Mantissa, Exponents, IEEE 754) which cannot be fixed in the ChemStation.

Temporary Solution:

n/a

Fix Information:

No Fix planned


KPR#:2152  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.05

Keyword: Reports

One-line Description:

Classic Report area% calculation can provide incorrect results

Problem:

The calculation in Area percent and height percent reports might be incorrect. The total sum of the response percent values do not add up to 100%. 

This occurs under the following conditions:
 - There is more than one signal.
 - In at least one of the signals no peaks are found.
 - The report setting specifies that peak results are sorted by retention time.

If the above conditions are met a value of 2 is subtracted from the total sum of areas and heights for every signal without peaks. This reduced sum is used in the calculation of the percent values. In the case of one single peak, or of few, small peaks, even the area( height) percent value of a single peak can be higher than 100%.

Intelligent Reports calculate the values correctly.

Temporary Solution:

The correct calculation can be viewed in Review mode of the OpenLAB ChemStation.
By loading only the signal with peaks a correct classic report can be generated.

Fix Information:

Fix planned.


KPR#:1837  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.05

Keyword: Reports , Custom Fields

One-line Description:

Empty Custom Fields are not printed in Intelligent Report

Problem:

If the content of a Custom Fields is empty this Custom Field will not appear in an IR at all.

Temporary Solution:

Use default values during CF setup.

Fix Information:

No fix Planned


KPR#:1545  Product:ChemStation  ChemStation (G2070BA)  B.04.02 DSP 3

Keyword: Reports , ECM Integration/OpenLab Option

One-line Description:

OpenLAB Option: data cleanup during shut down ignores pending .PDF reports

Problem:

When data cleanup at shutdown of the ChemStation is selected, the following problem may occur after performing data review with creation of .PDF reports:
If not all of the .PDF reports have been created when the ChemStation is shut down, the local data files will be deleted. The pending .PDF reports will be written to a sequence container and file structure that has been deleted during the shutdown. This results in a sequence container with .D files containg just those .PDF reports that had been generated after the shutdown.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:2125  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.07 SR1 HF01

Keyword: Reports , MSD Analysis

One-line Description:

Peak Performance calculations are off when data points are not equidistant

Problem:

ChemStation B.04.03 (and OpenLAB CDS ChemStation C.01.03) introduced a new algorithm for the calculation of peak performance parameters. The algorithm assumes that the data rate is constant across the signal. 
LCMS data with SIM groups that are not overlapping will cause a problem, becasue the data points are not at regular intervals across the whole chromatograms. There will be gaps without data points.
In these cases the peak tangents are calculated incorrectly. 

The peaks drawn in the classic performance report and the interactive peak performance tool show obvious issues:
The peaks are not centered at the reported retention time. 
The horizontal lines for peaks width and the vertical line for height do not match the peak outline.
The tangents are obviously wrong. Depending on the peak integration the tangents and peak height might not be on the report at all.

The algorithm used in ChemStation B.04.02 and earlier did also have a problem with this kind of data, but it has not been that obvious. The tangent calculations were still incorrect

Temporary Solution:

Define SIM groups across the whole chromatogram in the acquisition method. The time segments for the SIM groups should overlap.
For existing old data use the signal details to load only the time segment described by one SIM group.

Fix Information:

Resolved in revision OpenLAB CDS ChemStation C.01.07 SR2.


KPR#:1529  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: Reports , SEQ

One-line Description:

for very long sequence summary reports, error 41066 may occur

Problem:

For a very long sequence summary report (several thousand pages), error 41066 may occur:
"Output on external device Printer failed".

Temporary Solution:

Such long sequence summary reports are created when including the single run reports for a sequence with many runs. As a workaround, remove the single run reports from the sequence summary report.

Fix Information:

It is currently not planned to fix this problem.


KPR#:1392  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: RTL

One-line Description:

Retention Time Locking:  can only use kPa unit

Problem:

Retention Time Locking (RTL) only allows the unit of kPa. It is not possible to use psi or bar.

Temporary Solution:

n/a

Fix Information:

This problem is fixed in revision B.04.02 SP1.


KPR#:293  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Run Control

One-line Description:

ADC: "Location" field in "Sample Info" goes blank after first method run.

Problem:

In "Method and Run Control" view of A/D online session, "Location" field in "Sample Info" dialog box goes blank after first method run.

Temporary Solution:

n/a

Fix Information:

No fix is planned for this issue.


KPR#:841  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Run Control

One-line Description:

Headspace instr will fail to start if HeadSpace is used as the injection source

Problem:

The Headspace instrument will fail to start if HeadSpace is used as the injection source in GC Chemstation B.02.0x. 

HeadSpace as an "Injection Source" is now available beginning in ChemStation Revision B.02.01. However, the HeadSpace software, G2924AA Rev A.01.04, does not support this new injection source. If HeadSpace is used as the injection source, there is a failure to start the HeadSpace instrument.

Temporary Solution:

The workaround is to use the "Manual" choice as the Injection Source (as previously). The G2924AA HeadSpace Revision A.01.05 corrects 
this issue.

Fix Information:

This defect is fixed beginning in G2924AA HeadSpace Software revision A.01.05.


KPR#:1851  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03 SP1

Keyword: Run Control

One-line Description:

Unexpected shifts in the Online Plot

Problem:

Artifacts have been observed in the Online Plot for instruments with more than one module. 
These artifacts appear as inserted straight segments inside one of the observed traces. As a results traces may appear shifted from one injection the the next. Aside from the shift these insertions are not obvious.
The situation can occur when at least one module in the plot region has an RC.Net driver and the other has not. Examples are LC stacks or the 7100 CE in combination with an MSD, a 35900 or and ELSD.
The raw data are correct.

Temporary Solution:

Fixed in C.01.05 Hotfix 03.

Fix Information:

Fixed in OpenLAB CDS C.01.06.


KPR#:1635  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: Run Control , ECM Integration/OpenLab Option

One-line Description:

Using the Abort action will leave the system in an undefined state

Problem:

An Abort menu is available in several views of the ChemStation. It can be used to terminate the currently running task. 
This function is designed as an emergency stop. The current task is literally dropped and as a consequence other processes can end up unfinished or in an undefined state. This may result in unreliable instrument communication, lost data, unfinished ECM uploads or an inability to print reports.

Temporary Solution:

It is strongly recommended to exit the ChemStation software after an abort was executed. It is often necessary to also power cycle the online instruments.
A method or sequence should be saved under a new name.

Fix Information:

No fix planned.


KPR#:808  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: Sample Overlap

One-line Description:

Overlapped injections omits run when using the same vials for front and back inj

Problem:

Made up a 3-vial sequence using both ALS injectors and the same vial sequence (1, 2, and 5) for front and back.  In this case, the action should be:

Run 1:  Front: vial 1   Back: no injection
Run 2:  Front: vial 2   Back: vial 1
Run 3:  Front: vial 5   Back: vial 2
Run 4:  Front: no injection  Back: vial 5

Enabled overlapped injections in the sequence parameters.

The result was that the back injection from vial 1 was never made.  The following result is shown:

Run 1:  Front: vial 1        Back: no injection
Run 2:  Front: vial 2        Back: no injection
Run 3:  Front: vial 5        Back: vial 2
Run 4:  Front: no injection  Back: vial 5

Even though vial 1 has not been injected in back a datafile has been created: "001B0101.D".

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:9  Product:LC and CE Drivers  RC 1200.NET  A.02.02 SP1

Keyword: Sampler Configuration , Sampler Status

One-line Description:

Control of Tray illumination missing for new G1329A

Problem:

New G1329A ALS contain a tray illumination now. RC.Net driver for standard autosampler does not present possibility to switch on or off tray illumination of these new G1329A ALS.

Temporary Solution:

Check if G1329A is a new one and already has tray illumination.

On ChemStation's command line manually enter:
print SendModule$(LALS,"LGHT 1")
to switch on sampler's illumination and
print SendModule$(LALS,"LGHT 0")
to switch it off.

In ChemStation the use of a configuration with 'Classic' drivers might also be an option.

Fix Information:

fixed with LC Drivers A.02.03


KPR#:16  Product:LC and CE Drivers  RC 1200.NET  A.02.02 SP1

Keyword: Sampler Driver

One-line Description:

Control of tray illumination missing for G1329A/B & G4303A in Japanese and Chinese ChemStation

Problem:

The following problem occurs in Japanese and Chinese ChemStation only:

RC.Net drivers for Agilent 1260 Infinity Standard Autosampler G1329A/B or Agilent 1260 Infinity SFC Standard Autosampler G4303A do not provide any possibility - neither on ChemStation's menu nor on sampler's context menu - to switch on or off tray illumination inside G1329A/B or G4303A.

Control of Agilent 1260 Infinity Standard Autosampler G1329B by 'classic' drivers of ChemStation is not affected.

Agilent 1260 Infinity SFC Standard Autosampler G4303A requires the RC.Net Driver. No 'classic' driver for it exists.

Temporary Solution:

On ChemStation's command line enter:
print SendModule$(LALS,"LGHT 1")
to switch on sampler's illumination and:
print SendModule$(LALS,"LGHT 0")
to switch it off.

Fix Information:

The issue will be solved in driver version A.02.03.


KPR#:19  Product:LC and CE Drivers  RC 1200.NET  4.02.89

Keyword: Sampler Driver , Sampler Method

One-line Description:

Injection valve cleaning, entering new values does not work

Problem:

Go to setup -> injector cleaning -> check time1 to 4 and enter a time.
Close dialog and open again.
Entry fields for time is disabled.

Temporary Solution:

You have to uncheck and check the time 1 to 4 again to enter new values

Fix Information:

To be Resolved in Revision: future


KPR#:30  Product:LC and CE Drivers  RC 1200.NET  A.02.01

Keyword: Sampler Method

One-line Description:

Injection Volume not shown when loading a method (Inactive)

Problem:

The injection volume on a method is only shown correctly after the start of the sequence.
After just loading the method, the previous used inj vol is shown.
ChemStation B.04.02

Temporary Solution:

n/a

Fix Information:

fixed with release A.02.04 of the LC Drivers.


KPR#:31  Product:LC and CE Drivers  RC 1200.NET  A.02.04

Keyword: Sampler Method

One-line Description:

ICF A.01.04 - No method download, if method contains pretreatment/injector program steps.

Problem:

Methods including pretreatment/injector steps are not downloaded to the instrument, prior to a run.

Temporary Solution:

n/a

Fix Information:

The defect is fixed with the Driver Release A.02.06

In relation to the Instrument Control Framework, a PointPatch is available.
Please contact your Agilent Support Team.


KPR#:33  Product:LC and CE Drivers  RC 1200.NET  A.02.07

Keyword: Sampler Method

One-line Description:

Dua Loop Sampler: ACQ method shows wrong loop with overlap injection

Problem:

Pre Cond.: Overlapped injection wit overfill by factor with a DL with different loop sizes for upper and lower loop
Consequence: due FW automatism it happens that we will report the wrong loop and the wrong injection volume

Wrong injection volume reported in the case of 2 different loops
(does not make sense in the application).
A logbook entry: 
"Warning: Used injection volume can be different then specified in the method due to different loop size."

Temporary Solution:

no fix

Fix Information:

n/a


KPR#:54  Product:ChemStation  ChemStation (G2070BA)  A.09.03

Keyword: SEQ

One-line Description:

Printout of Partial Sequence table misses essential information

Problem:

The partial sequence dialog offers the option to print the run list.

This printout does not contain:
operator 
date/time information
page number
total page number
instrument name

Temporary Solution:

There is currently no adequate workaround available.

Fix Information:

 No fix Planned


KPR#:154  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: SEQ

One-line Description:

The data file name in a sequence summary compound report is limited to 8 charact

Problem:

With ChemStation Rev. B.01.01 following issue has been reported using Sequence Summary:

Using the Sequence Summary, Compound Summary option, the data file name within the resulted report is limited to 8 character.

Temporary Solution:

A macro is availabe to fix this issue. Please contact your Agilent support.

Fix Information:

n/a


KPR#:241  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: SEQ

One-line Description:

No divisor available for samples in the sequence table

Problem:

There are only multipliers available in the sequence table (Multiplier and Dilution), but no divisors.

Temporary Solution:

If the user wants to divide the results with a constant factor she/he needs to calculate reverse value first and enter that in one of the multipliers.

Fix Information:

For this issue currently no fix is planned.


KPR#:259  Product:ChemStation  ChemStation (G2070BA)  A.10.01

Keyword: SEQ

One-line Description:

sample name of a run changed when reprocessing parts of a sequence

Problem:

A sequence of several runs has been used to acquire data, and the sample names have been assigned to the runs using the corresponding column in the sequence table. When now the sequence is to be reprocessed, but some lines are deleted from the table, the sample names are applied to different runs than during the acquisition if the data file naming is not mastered by the sequence table.

Temporary Solution:

ChemStation provides the function "Partial Sequence" that allows the user to select only parts of a sequence with automatically assigning the original data file names to the runs even when in the original sequence the names are handled in the Sequence Parameter dialog.

Fix Information:

It is currently not planned to change this behavior.


KPR#:274  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: SEQ

One-line Description:

performance problem when many sequences and methods are present

Problem:

If many methods and sequences are present in the corresponding folders of an instrument, the following performance problem is observed:
The first time after starting your computer when you load a method or the sequence table of a sequence, it takes a longer time until the corresponding dialog finally will appear.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:300  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: SEQ

One-line Description:

changing a plate in hypersequence table during leads to an error message

Problem:

When a hypersequence is running it is nevertheless possible to change the plate of the currently running seuqence. However, this leads to the following error message at the end of this sequence: "CP_hyper2:Parameter3:A value is outside the legal range".

Temporary Solution:

n/a

Fix Information:

No fix is planned for this issue.


KPR#:356  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: SEQ

One-line Description:

Sample amount ignored for blank runs when entered from sequence table

Problem:

A sequence is reprocessed with the option "Use Sequence Table" checked in the Sequence Parameters dialog. When in the sequence table no vial number is specified (i.e. when the run is actually a blank run), a sample amount specified in the sequence table will not be applied to the run.

Temporary Solution:

If you want the sample amount to be applied with this setup, specify any vial number for the run in the sequence table.

Fix Information:

 No fix Planned


KPR#:480  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: SEQ

One-line Description:

Method Name field blank in sequence table (in Data Analysis)

Problem:

In Data Analysis, load a sequence. From the reprocessing tools, load the sequence table. In any line, select the Method field, select "Browse...", but than press Cancel.
Instead of falling back to the original method which was present before touching that field, the method field remaims empty. A warning is only displayed when you leave the table.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:778  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.01

Keyword: SEQ

One-line Description:

Change of sample type during sequence acquisition is ignored

Problem:

If you change the sample type of a run (that has not yet been executed) of a running sequence, the change in sample type will be ignored.

Temporary Solution:

Change the sample type after the sequence has been completed and reprocess the sequence data.

Fix Information:

Fixed in OpenLAB CDS C.01.06.


KPR#:780  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: SEQ

One-line Description:

Default hypersequence Def_LC.hyp can be overwritten

Problem:

The default hypersequence Def_LC.hyp can be overwritten after editing. The user is not forced to save the hypersequence with a different name.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:813  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.01

Keyword: SEQ

One-line Description:

Front injector is not installed but it is still available in the sequence table

Problem:

Running on a 7890A Back Injector only. Only the Back injector is available as the Injector selection, however, in the sequence table, both front and back are available for sequence input. The front will take back injector vial numbers and will not reject them. 

The front injector position should not be shown or accept vial numbers since it is not configured.

When only the front injector is configured, the back injector position will also be shown in the sequence table.  If the back injector is not configured, do not use the back injector in the sequence table.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:823  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR2

Keyword: SEQ

One-line Description:

Sequence stops if vial number is 0 on an 1100/1200 system without error message

Problem:

If on an 1100/1200 LC system a sequence is run that contains a line with vial number 0 (e.g. a sequence that has been created for a 1090 system), the sequence stops at this line without an error message that indicates the root cause.
The cause is only documented in the sequence logbook.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:927  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: SEQ

One-line Description:

Filldown wizard for CTC Increment Directions mixes up row/column directions

Problem:

Create a CTC method and open the Sequence table Filldown Wizard.

The three options for Increment direction seem to be mixing up the row and column for each selection such that they are creating a sequence that is the selection, but at a 90 degree angle from the selection.

It sees 'rows' as A-G (short side of the rectangle); 'columns' as 1-14 (long side of the rectangle).

For a VT-98 tray:
Option 1 claims it will access vials from left to right, which should be A1-G1, according to their definition. However, it creates a sequence going from A1-A14, B1-B14, etc., front to back, front to back.

Option 2 claims it will access vials from left to right in the first 'column', then right to left in the second, then left to right, winding back and forth.
Instead, this winds back and forth, but front to back and back to front (A1-14, B14-B1, C1-C14...)

Option 3 claims it will access vials from front to back, front to back across the 'rows'. Instead, it creates a sequence th at goes across each 'column' left to right (A1-G1, B2-G2). 

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:938  Product:ChemStation  ChemStation (G2070BA)  B.03.01 SR1

Keyword: SEQ

One-line Description:

Problems running more than 422 sequence lines with CTC sampler

Problem:

In a 1200 system with CTC sampler, sequences will stop after 422 sequence lines.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:941  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: SEQ

One-line Description:

Online session closes when copy/paste is used to build seq table

Problem:

Create a new seq from (DEF_GC.S) and add the first line to the sequence table. Copy and paste the next line causes the online session to shut down with no errors.

Temporary Solution:

There are 2 workarounds:

1. Clear the clipboard in Excel. Go to Edit > Office clipboard. You should see the button to clear it. Then create the sequence table using copy and paste.

2. Close all the Office Applications before creating a sequence table. Clear the clipboard before closing the application. 

Fix Information:

n/a


KPR#:973  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: SEQ

One-line Description:

Sample location not displayed correctly in Data Analysis with high plate number

Problem:

If you inject samples using a sampler that can handle more plates than the 1100/1200 well plate samplers (e.g. CTC sampler), the follwoing problem can be observed:
When injection from plates with plate numbers larger than 7, Data Analysis displays a strange sample location, like e.g. "- 20479".

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1317  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02

Keyword: SEQ

One-line Description:

Mismatch for sequence completed time stamp in sequence and instrument logbook

Problem:

The time stamps for sequence completion may differ between the sequence logbook and the instrument logbook:
The sequence logbook lists the time after completion of the acquisition, while the instrument logbook displays the time after completion of the sequence summary report.

Temporary Solution:

n/a

Fix Information:

Fixed in revision C.01.04


KPR#:1322  Product:ChemStation  ChemStation (G2070BA)  B.04.02 Pre-Release

Keyword: SEQ

One-line Description:

sporadic ChemStation crashes after aborting a sequence

Problem:

In the ChemStation B.04.02 Pre-Release it is sporadically observed that ChemStation crashes after a sequence got aborted.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1414  Product:ChemStation  ChemStation (G2070BA)  B.04.02 Aladdin

Keyword: SEQ

One-line Description:

Purify: error message at the end of study

Problem:

When running a study with Purify add-on software, the following error message will be displayed at the end of running a study:
"The system is unable to load a sequence while another sequence is running."

Temporary Solution:

Close the error message by pressing OK before continuing with the next study.

Fix Information:

n/a


KPR#:1441  Product:ChemStation  ChemStation (G2070BA)  B.04.02 SP1

Keyword: SEQ

One-line Description:

ChemStation crashes when trying to use a method containing an ' in the name in the sequence table

Problem:

When you try to use a method in the sequence table that contains an ' in its name, the ChemStation will crash when trying to leave the sequence table with the OK button.

Temporary Solution:

n/a

Fix Information:

This problem will be fixed in revision B.04.03.


KPR#:1442  Product:ChemStation  ChemStation (G2070BA)  B.04.02 SP1

Keyword: SEQ

One-line Description:

an error may occur when stopping a sequence immeadiately after start

Problem:

A Microsoft .NET Framework exception may occur when stopping a sequence immeadiately after start of the sequence.
Afterwards, the sequence will not restart.

Temporary Solution:

Shut donw and restart ChemStation.

Fix Information:

n/a


KPR#:1470  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 DSP 2 (FireAnt)

Keyword: SEQ

One-line Description:

ACQNOTREADYWAIT time cannot be changed in the user interface, but only via the command line

Problem:

ACQNOTREADYWAIT is a system variable that determines the time the sequence engine waits for the instrument modules to get ready. The variable can only be changed via a command in the ChemStation command line, and is not available in any of the ChemStation dialogs or the user interface.

ACQNOTREADYWAIT is not to be confused with the "Not ready Timeout" in the Sequence Parameters dialog which specifies the time the instrument waits for the ChemStation to complete data analysis tasks.

Temporary Solution:

Use the command
ACQNOTREADYWAIT=
to change ACQNOTREADYWAIT.

Fix Information:

n/a


KPR#:1486  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 DSP 2 (FireAnt)

Keyword: SEQ

One-line Description:

not possible to create an interactive report during post-time

Problem:

During the post-time for an acquisition run, it is not possible to load the corresponding data file in the offline ChemStation and to generate an interactive report. Only after the post-time, the run can be reviewed in the offline ChemStation.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1501  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: SEQ

One-line Description:

wrong data file naming after missing vial with Prefix/Counter naming

Problem:

When Prefix/Counter naming is used and several injections are done per sequence line, the names of the subsequent runs will be wrong after a line where the vial is missing: For the line with missing vial, the expected data file folders are created with the expected name. Afterwards, the first injection of the next line will be written in the last empty folder for the line with missing vial. All subsequent file names are wrong.

Temporary Solution:

n/a

Fix Information:

No fix is planned for this problem.


KPR#:1514  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: SEQ

One-line Description:

The path to a data file has a limited length

Problem:

The current path that points to files inside a data folder has a limit of 260 characters.
With a data file name of 18 characters (16 + .d) and the known files inside a data folder the effective length of a data path, including the container name cannot exceed 170 characters.

Temporary Solution:

None

Fix Information:

No fix planned


KPR#:1541  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: SEQ

One-line Description:

sequence template may be marked as changed after modifying the sequence file in a sequence container

Problem:

With Unique Folder creation switched, a copy of the sequence file *.S is copied into the sequence data container at the start of sequence acquisition. When the user modifies the sequence file during acquisition but does not save the change, the modified sequence is saved in the data container, but after acquisition the sequence template is loaded again with a change flag.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:1542  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: SEQ

One-line Description:

Sequence table ignores the 40 characters limit for file names

Problem:

Data file names in the chemstation are limited to 40 character (+2 for .d)
In the sequence table longer names can be entered by copy/paste or xml import.
The file names will be truncated to 40+2 character length. This can lead to two or more file names within a sequence folder that are identical, causing the later runs to overwrite the older runs.

Temporary Solution:

Make sure your filenames are unique within the first forty characters.

Fix Information:

This problem will be fixed in revision B.04.03.


KPR#:1543  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 SP1

Keyword: SEQ

One-line Description:

sequence data set shown as "in progress" although the sequence is longer acquiring

Problem:

In rare cases especially after aborting a sequence it may happen that the sequence data is still displayed as "in progress" in Data Analysis view, even after acquisition for the sequence has stopped.

Temporary Solution:

The "in progress" flag is caused by a file acquiring.txt inside the sequence data folder. To fix the problem, localize the sequence container in Windows Explorer and delete the file acquiring.txt.

Fix Information:

 See Previous Note.


KPR#:1552  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02

Keyword: SEQ

One-line Description:

Sequence is marked as running although it is finished

Problem:

In a scenario where several sequences are acquired into one folder (unique folder off), it can happen that a sequence in Data Analysis View is still reported as in progress.

Temporary Solution:

Close the data path node in the ChemStation Explorer and open it again. This should update the tree view.

Fix Information:

No fix planned. See Previous Note.


KPR#:1554  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: SEQ

One-line Description:

Sequence reprocessing with methods from old revision fails

Problem:

If a method with the old integration settings is used for sequence reprocessing, the process runs into an error: IntegrateObj failed - IntegMode mismatch
There is no migration of the method settings during reprocessing.

Temporary Solution:

Load methods from ChemStation G2070AA before reprocessing. Accept the method migration for integration and spectra options and save the method in the new format.

Fix Information:

No fix planned.


KPR#:1571  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 SP1

Keyword: SEQ

One-line Description:

Missing Vial during sequences causes CTC ChemStation to hang

Problem:

Create a five line sequence using Add-on CTC.
Run sequence, removing vial three.

CTC will report missing vial and it is shown in the log book for CTC.

CTC will reset and go back to ready state, however, ChemStation will hang.

All menu items are grayed out and you cannot touch anything on the system.
The ChemStation must be ended in Task Manager.

Temporary Solution:

n/a

Fix Information:

Issue resolved in OpenLAB C.01.05.


KPR#:1604  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: SEQ

One-line Description:

HeadSpace sequence method will not load when it follows a Valve sequence

Problem:

The valid vial range is determined by the injection source of the loaded method, not the method in the sequence.

As soon as the HS sequence loads, it fails with:
'---------------------------
Error
---------------------------
Invalid location found in FRONT sequence, line 1
There may be others
---------------------------
OK   
---------------------------'
The Valve method has reloaded which is invalid for this HS sequence

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB CDS ChemStation Ed. C.01.06.


KPR#:1636  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: SEQ

One-line Description:

Headspace or PAL Sampler sequence with methods saved to a location other than C:\ does not run

Problem:

Save method on D drive, and start sequence with headspace or PAL Sampler.
    GC stays in "Prep Run".
    headspace and PAL Samplers do not work.

Headspace or PAL Sampler does not start when method is saved to a location other than C:\

This defect affects all headspace hardware models with the B.0x.0x series headspace and PAL Sampler Software.

Temporary Solution:

Workaround is to save and use methods from the C:\ drive.

Fix Information:

Fixed in revision C.01.04


KPR#:1648  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: SEQ

One-line Description:

Sequence parameter subdirectory overwrites sample info parameters during run

Problem:

The sequence parameters (i.e subdirectory, file name, vial location and sample name) overwrites the sample info screen with sequence parameter information during the sequence run only.  

The Sample Info screen goes back to its original settings once the sequence finishes.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1687  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 SP1

Keyword: SEQ

One-line Description:

Missing ‘Reprocessing Only’ in the Dropdown List for Unique Folder Creation OFF

Problem:

1. Start ChemStation (Online)
2. Select View -> Preferences -> Sequence tab
3. Select Unique Folder Creation OFF
4. Select Sequence -> Sequence Parameters
5. Click and open the dropdown list in the part of methods to run field
“Reprocessing Only” is missing.

Temporary Solution:

n/a

Fix Information:

Fixed in revision B.04.03.


KPR#:1708  Product:ChemStation  ChemStation (G2070BA)  B.04.03 GC DSP

Keyword: SEQ

One-line Description:

B.04.03 DSP1 [54] Partial Sequence Acquisition Does Not Work with Headspace with the Back Sequence Table

Problem:

It is not possible to run partial sequence acquisition using the back sequence table with headspace.  

Exception is Found:
                                Exception
                                - Message:    Object reference not set to an instance of an object.
                                - Source:     AgtChemRCNetAdapter
                                - TargetSite: Void .ctor(PartialSequenceLine, SequenceLine, Int32, Int32, System.String, System.String)
                                - HelpLink:   
                                - Data:       System.Collections.ListDictionaryInternal
                                - StackTrace:
                                   at Agilent.ChemStation.RapidControl.DataHelper.RCSequenceEntryHelper..ctor(PartialSequenceLine partialSeqLine, SequenceLine seqLine, Int32 nMethodID, Int32 nPretreatId, String OnSampleIDMismatch, String strInjectorID)
                                   at Agilent.ChemStation.RapidControl.DataHelper.RCSequenceHelper.AddEntry(PartialSequenceLine partialSeqLine, SequenceLine seqLine, Int32 nMethodID, Int32 nPretreatID, String strInjectorID)
                                   at Agilent.ChemStation.RapidControl.AgtChemRCNetAdapter.FullSequence.CreateCollections(PartialSequenceTable partialSeqTable, SequenceContext sequenceContext, MethodCollection MethodCollection, PretreatCollection PretreatCollection, RCSequenceHelper SequenceHelper, Boolean bPreventMethodResolution)
                                   at Agilent.ChemStation.RapidControl.AgtChemRCNetAdapter.FullSequenceDownload.StartFullSequenceMode(String sampleIDMismatch)
                                   at Agilent.ChemStation.RapidControl.AgtChemRCNetAdapter.ControlAdapter.DoPreSequencePhase()


Investigation:  

The exception is the reason why the sequence is not downloaded to the instrument.  It is related to back sequences. Due to a minor mistake, the attempt to resolve a partial Back-sequence line fails, because the ChemStation tries to resolve it from the Front-Sequence Table.

Temporary Solution:

Workaround is to use Front-Sequences only.

Fix Information:

 Fixed in revision B.04.03 SP1.


KPR#:1712  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: SEQ

One-line Description:

Opening the front sequence table, click on a cell, make no changes, and select OK causes sequence to stop with 7697A Headspace Sampler

Problem:

If  the sequence table is opened during a running sequence and the user clicks on a cell and makes no changes, there is a lock placed on the vial so that the instrument doesn't put a vial in the oven. If the user makes no changes and hits OK then the Chemstation never continues past the locked vial.

Temporary Solution:

If the user makes a change to the sequence or hits the cancel button then the sequence works correctly.

Fix Information:

Fixed in Revision B.04.03 SP1.


KPR#:1713  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: SEQ

One-line Description:

Opening the front sequence table, click on a cell, make no changes, and select OK causes sequence to stop with 7697A Headspace Sampler

Problem:

If  the sequence table is opened during a running sequence and the user clicks on a cell and makes no changes, there is a lock placed on the vial so that the instrument doesn't put a vial in the oven. If the user makes no changes and hits OK then the Chemstation never continues past the locked vial.

Temporary Solution:

If the user makes a change to the sequence or hits the cancel button then the sequence works correctly.

Fix Information:

Fixed in revision C.01.04


KPR#:1725  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: SEQ

One-line Description:

Sample statistics are not done, when the first compond in the calibration table is not detected.

Problem:

If the first calibrated compound is not found in the first sample run of a sequence, no sample statistics will be performed.

Temporary Solution:

None

Fix Information:

Fixed in revision  B.04.03 SP1.


KPR#:1728  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: SEQ

One-line Description:

Pausing and resuming a sequence can change the data path.

Problem:

Cases have been observed, where data files from a sequence are stored in a different location after the sequence had been paused and resumed. In those cases data files acquired after a sequence resume have been stored in the folder where sequence templates are stored. Possibly after the sequence template has been saved.

Temporary Solution:

Do not use pause and resume.
Stop the sequence, make modifications and use partial sequence to acquire the remaining runs. Only append sequence lines after the last run that has been injected. Otherwise the file naming scheme will be ambiguous.

Fix Information:

Fixed in revision B.04.03 SP1.


KPR#:1729  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: SEQ

One-line Description:

Headspace or PAL Sampler sequence with methods saved to a location other than C does not run

Problem:

Save method on D drive, and start sequence with headspace or PAL Sampler.
    GC stays in "Prep Run".
    headspace and PAL Samplers do not work.

Headspace or PAL Sampler sequence does not start when method is saved to the D drive

This defect affects all of the B.0x.0x series headspace and PAL Sampler software.

It is possible to run a method.

Temporary Solution:

Workaround is save and use methods from the C drive.

Fix Information:

Point patch available for B.04.03 SP1 [87]

Patch has been tested with:

G2181BA ChemStore rev. B.04.02 SP2
G2183BA Security Pack   rev. B.04.02 SP2
Headspace Control Software B.01.02.1


KPR#:1735  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: SEQ

One-line Description:

Partial sequence run with methods saved to a location other than C:\ doesn’t run with headspace and PAL Samplers

Problem:

A partial sequence run with methods saved to a location other than C:\ simply doesn’t run. The logbook says it is running but the headspace does not run.

Temporary Solution:

Workaround is save and use methods from the C:\ drive.

Fix Information:

Point patch available for B.04.03 SP1 [87]

Patch has been tested with:

G2181BA ChemStore rev. B.04.02 SP2
G2183BA Security Pack   rev. B.04.02 SP2
Headspace Control Software B.01.02.1


KPR#:1736  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: SEQ

One-line Description:

Partial sequence with methods saved to a location other than C: does not run with headspace and PAL Samplers

Problem:

A partial sequence run with methods saved to a location other than C: does not run. The logbook says it is running but the headspace or PAL Samplers do not run.

This defect affects all headspace hardware modules with the B.0x.0x series headspace and PAL Sampler software.

Temporary Solution:

n/a

Fix Information:

Fixed in revision C.01.04


KPR#:1737  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: SEQ

One-line Description:

Partial sequence with methods saved to a location other than on C does not run with headspace and PAL Samplers

Problem:

A partial sequence run with methods saved to a location other than on C does not run. The logbook says it is running but the headspace or PAL Samplers do not run.

Temporary Solution:

n/a

Fix Information:

Point patch available for B.04.03 SP1 [87].


KPR#:1767  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: SEQ

One-line Description:

GC:  sequence does not continue if there is a missing vial, uploads data to ECM and goes to next sequence

Problem:

The GC Chemstation never had the ability to skip a missing vial (the classic GC driver posts a dialog box asking the user what to do, abort or retry).  With RC.NET, if there is no vial on the tray the system stops the sequences, uploads the data to ECM and goes to the next sequence. Need to be able to specify what the system should do if there is a missing vial, skip or wait.

Temporary Solution:

n/a

Fix Information:

Issue resolved in OpenLAB C.01.05.


KPR#:1787  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: SEQ

One-line Description:

78xx: Sequence stops when pushing the GC STOP key during sequence

Problem:

Pushing the stop key at the GC keyboard will stop the current run and generate a report but does not continue to the next line in the sequence.  

This occurs beginning on ChemStation version B.04.01 SP1.

In revisions B.03.01 - B.04.01, pushing the stop key at the GC Keyboard would stop the current run and generate a report and continue to the next line in the sequence.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.06 for 78xx.


KPR#:1788  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: SEQ

One-line Description:

Custom field window not prompted for Back injector on first run of the sequence

Problem:

During the first run of the sequence, the custom field window is only prompted for the Front injector.

Temporary Solution:

n/a

Fix Information:

Issue fixed in OpenLAB C.01.05.


KPR#:1798  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: SEQ

One-line Description:

Sequence Reprocessing: Not possible to stop a sequence that is reprocessing

Problem:

When reprocessing the sequence, it is not possible to stop it.  The stop button is present, but does not to work. The user has to wait till the whole sequence reprocess completes.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1805  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: SEQ

One-line Description:

Sequence Import Mapping is missing LIMS ID columns

Problem:

The dialog box to map a sequence import file (Sequence -> Import Sequence Template) is missing the LimsID columns.  The LimsID 2 and LimsID 3 columns specifically are read-only fields that need to be imported in order to ensure that a unique LimsID can later be used in the result export to the Lims to ensure results are matched to the correct sample.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1822  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.05

Keyword: SEQ

One-line Description:

Sequence parameters: Changes to Fraction start location ignored

Problem:

Sequence parameters: Changes to Fraction start location was ignored.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1824  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.05

Keyword: SEQ

One-line Description:

After upgrade, Sequence modified flag is set even if no changes are made before reprocessing

Problem:

Sequence modified flag is activated even if no changes are made before reprocessing for a dual method sequence

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1830  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03 SP1

Keyword: SEQ

One-line Description:

'Number Format error' in Sequence log file after reprocessing

Problem:

Migrated a sequence from B.04.03 SP1 with 'Unique folder creation OFF' to C.01.04.  

Used the 'Result set migration' tool to migrate the data to a result set.  In online ChemStation, the result set was reprocessed.
In the sequence.log file,  'CP Macro Number format error in: %d' was displayed.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1870  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02

Keyword: SEQ

One-line Description:

Well Plate locations are not accepted in offline sequences

Problem:

In offline sessions of an LC/LCMS ChemStation the sequence table offers by default only the vial range 1-100. It does not accept well plate location like P2-C7.
In order to setup an offline sequence with WPS locations you have to go to the Menu: Instrument>Plate Configuration ....  In the upper left corner you select " Use Wellplate Tray" .

Temporary Solution:

None

Fix Information:

N/A


KPR#:2165  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.07

Keyword: SEQ

One-line Description:

Partial sequence does not support run-to-run relations

Problem:

Some sequence functions are not supported by Partial Sequence. This affects Overlap injections, pre-fetch vial and similar high throughput features..

Temporary Solution:

n/a

Fix Information:

No fix planned.


KPR#:1528  Product:Method Scouting Wizard  Method Scouting Wizard  A.02.01

Keyword: SEQ , ACQ , Method Scouting Wizard

One-line Description:

blank runs are deleted from disk because of a setting in Method Scouting Wizard

Problem:

The Method Scouting Wizard provides an option to delete all blank runs at the end of a campaign. If this option is switched on, blank runs are deleted even if the data is not acquired using Method Scouting Wizard, but by "normal" ChemStation sequences.

Temporary Solution:

Switch off the setting to delete blank runs when not using the Method Scouting Wizard.

Fix Information:

n/a


KPR#:920  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.01

Keyword: Spectra Analysis

One-line Description:

Apex spectra not correctly reported

Problem:

When storing spectra during peaks, the spectrum for the apex may be collected at the wrong point in time, depending on the detector peakwidth parameter.

Temporary Solution:

The instrument hardware’s peak detection algorithm has to identify start, slope, apex, and end of a peak in order to be able to send the correct spectra to the ChemStation software. The results of this process are influenced by the sampling rate / peak width of the detector. The peak detection algorithm of the detector hardware performs best when c. 15 data points per peak are collected. In order to account for broadening of peaks at the end of the chromatogram use the Peakwidth option of the timetable. Peakwidth changes the filters used for peak-controlled spectra acquisition, but not the data rate of a chromatographic signal.

Fix Information:

n/a


KPR#:1036  Product:OpenLAB CDS ChemStation Edition  M83xx  B.03.02

Keyword: Spectra Analysis

One-line Description:

unexpected question whether user wants to save spectra library with the method

Problem:

When there are unsaved changes both in the currently loaded spectra library and the currently loaded method and the user finally saves the method changes, a message comes up asking him whether he wants to save the library with the method.
If the user acknowledges this message with Yes, the library will be stored in the method folder of the currently loaded method.
This even happens, if the method does not reference the library in the library search options or does not use library search at all.

Temporary Solution:

N/A

Fix Information:

No change planned.


KPR#:1323  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02

Keyword: Spectra Analysis

One-line Description:

Issues with scaling of online plot for DAD G4212A

Problem:

Independently of the wavelenth range in the method, the online plot will only display the wavelength range 190 - 400 nm for the DAD G4212A.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1790  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Spectra Analysis

One-line Description:

Mass spectra drawn with strange shapes

Problem:

Mass spectra are drawn with strange shapes instead of distinct vertical lines. When zooming in, then vertical line will be visible.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB CDS ChemStation Edition C.01.04 Hotfix03.


KPR#:13  Product:ChemStation  ChemStation (G2070BA)  A.10.01

Keyword: User Interface

One-line Description:

Fraction collection module  error message given at different  Fraction Collector

Problem:

Using a system with two  or three  Fraction collection modules (but no recovery Fraction collector set)
the first Fraction collector  will show the message  "Could not find a valid next position"  even if that event  occured on one of the other Fraction collectors.

If a system is used as recovery fraction collection,
the message is shown at the right module.

Temporary Solution:

The graphical userinferface will give the right information
about the event occured. 

In case this error message pops up, the user should 
watch the  Fraction collector modules, to see which one is causing the problem.

Fix Information:

 No fix Planned


KPR#:44  Product:ChemStation  ChemStation (G2070BA)  A.09.03

Keyword: User Interface

One-line Description:

Signal Task: Time reference lines only visible with 256 colors on Compaq bundle

Problem:

Compaq chemstation bundle PC EVO 510 only:
Prod ID: X1155AV with Windows 2000, SR 2 or SR3

In Data Analysis signal Tasks the Time Reference Lines are not shown
when selected.

Temporary Solution:

The solution will be to change the hardware acceleration for the video card from the desktop properties.
Beware system settings True color 32-bit and a resolution 1024x768. 

To perform the change: 
1. right click on desktop --> select Properties tab
2. set colors to true color and resolution to 1024x768 à go to advanced button
3. select  Troubleshoot tab à access to slid bar  for  hardware acceleration on the video card. 
4. Set the hardware acceleration to NONE.

Fix Information:

No fix is planned for this problem.


KPR#:62  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: User Interface

One-line Description:

Operator field in the Change access level dialog box retains only 29 characters

Problem:

When the operator name is changed in the Change Access Level dialog, only 29 characters are saved, but no warning is displayed, if more than 29 characters are typed in the dialog.

Temporary Solution:

Specify operator name with less than 30 characters.

Fix Information:

No fix is planned for this problem.


KPR#:68  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: User Interface

One-line Description:

instrument remains in running status after a diagnose test

Problem:

Sporadically it was observed that the ALS remains in running state in the Diagnosis view after running Column Thermostat test.

Temporary Solution:

In case the problem occurs, terminate the test manually by Abort.

Fix Information:

No fix is planned for this issue.


KPR#:69  Product:ChemStation  ChemStation (G2070BA)  A.10.02

Keyword: User Interface

One-line Description:

Fraction time table entries corrected by the biggest delay time

Problem:

A run with one more signals was acquired using fraction collection. In the chromatograms the fraction times are correctly displayed with the correctin by delay time of the detector applied.

In the fraction table below the chromatograms the uncorrected fraction times should be listed because the table is not linked to a particular signal. However, the actually displayed times are corrected by the largest delay time in the signals.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:74  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: User Interface

One-line Description:

"EMPV Cleaning Procedure" displays misleading status

Problem:

When EMPV Cleaning Procedure for a preparative pump is performed, at step 6 (Test EMPV leak) the status remains "Waiting" even when "Continue" button is already available.

Temporary Solution:

Press "Continue" button when it is available to continue with the EMPV Cleaning Procedure.

Fix Information:

No fix is planned for this problem.


KPR#:76  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: User Interface

One-line Description:

Inconsistent update of peak detectors table in Set Up Fraction Collector

Problem:

If the "Peak Detectors" table in the "Set up Fraction Collector" is edited while the fraction trigger mode is "Off", only the changes for upper threshold are retained, while the edits for up slope, down slope, and threshold are discarded upon leaving the "Set up Fraction Collector" dialog.

Temporary Solution:

Only edit the "Peak Detectors" table when the fraction trigger mode is "Peak-based".

Fix Information:

No fix is planned.


KPR#:95  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: User Interface

One-line Description:

Fraction Preview: Shifted marker in Preview

Problem:

There is no fix planned for this problem.

Temporary Solution:

This is only a graphica interface issue. The correct value is taken where the mouse was clicked, and this value is taken into the table and used for fraction collection.

Fix Information:

 No fix Planned


KPR#:121  Product:ChemStation  ChemStation (G2070BA)  (None)

Keyword: User Interface

One-line Description:

Datafile name in File information not updated for overlayed datafiles

Problem:

For ChemStation Rev. A.10.x following issue as been reported for graphical information.
Using the overlay function for datafiles within the Data Analysis-  Integration task section, the datafile information in the lower left corner are displayed when a signal channel is selected. 
Using this option the file information box will be updated with the correct information except the file name, which will refer to the first loaded datafile of the overlaid set.

Temporary Solution:

The correct file name is displayed in the header of the loaded chromatogram.

Fix Information:

 No fix Planned


KPR#:126  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: User Interface

One-line Description:

Hardware related zooming issue

Problem:

Under the following issue has been reported using a Matrox Millenium G450 or G400 graphic card :
Zooming several times into a Datafile by using only a small area of the y-axis strange veritacal lines might be seen in the Datafile.

Temporary Solution:

This is a hardware issue and cannot be solved within the ChemStation.

Fix Information:

n/a


KPR#:232  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: User Interface

One-line Description:

Using "Print window" is not clear what is the active window to print

Problem:

Within different views of the ChemStation it is possible to have multiple items on the screen, now using the "print window" option not always the expected window is printed:

E.g.:

1. load multiple signal, switch to single single view -- ALL signals will be printed, not only the selected one

2.With the chromatogram and integration items on the  screen, it is not possible to define an active window

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:234  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: User Interface

One-line Description:

Minimize Logbook option is not available

Problem:

When you open the Current Logbook in Method and Run Control view, the corresponding window shows the three standard boxes to handle the size. While the other two items are available, the Minimize box cannot be used.

Temporary Solution:

Resize the window manually by using the mouse and drag the corner of the window.

Fix Information:

 No fix Planned


KPR#:239  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: User Interface

One-line Description:

"switch off pump and lamp" question is coming even if alreay switched off

Problem:

When an online session of ChemStation is closed, the user always gets asked whether he wants to switch the column thermostat, pump and detector lamps - even when the user already switched off the instrument manually before.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:286  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: User Interface

One-line Description:

Switching between OQ/PV to  Data Anaylsis view impact the fraction table GUI

Problem:

Using B.01.03 and fraction colletion system, the fraction table in the Data Analysis view "select Fraction Task" can disappear when switching to OQPV context and back again to Data Analysi.

Temporary Solution:

n/a

Fix Information:

No fix is intended for this issue.


KPR#:322  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: User Interface

One-line Description:

in online plot it is difficult to differentiate between two DAD detectors

Problem:

In a system with two DAD detectors, e.g a DAD G1315B and a G1315C, in the online it is very difficult to differentiate which of the signals belong to DAD1 or to DAD2, respectively.

Temporary Solution:

Display the Instrument Actualsto differentiate between the two DADs.

Fix Information:

No fix is planned for this issue.


KPR#:446  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: User Interface

One-line Description:

MultiMode Source is not listed as a possible detector for the OQ/PV

Problem:

In Verification view select either Injector Precision/Carry Over or Response Linearity test. In the setup for these tests only the following detector types are available: EI, APCI and APPI,  while MultiMode Source is missing.

Temporary Solution:

Execute the OQ/PV with customized verifications using the MultiMode Source.

Fix Information:

n/a


KPR#:455  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: User Interface

One-line Description:

Recovery Mode not updated in Instrument Actuals GUI

Problem:

Using ChemStation Rev. B.02.01 running a fraction collector in recovery mode, the Instrument Actual panel will still say "Recovery No".

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:462  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: User Interface

One-line Description:

Incorrect status message of Chip Cube in case of lost synchronization

Problem:

Following issue has been reported using ChemStation B.02.01 running a Chip Cube module. If the LC tower is going into an error state and the sychronization of each module is lost, all LC modules will show an error , the Chip Cube instrument actuals will still indicate the Chip Cube in a working state.

Temporary Solution:

n/a

Fix Information:

No fix is planned for this issue.


KPR#:466  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: User Interface

One-line Description:

Manual injector use is not  present in User Interface

Problem:

Following issue has been reported using ChemStation B.02.01 and manual injectors: In the graphical user interface, the injector is not labeled  as  "Manual" when in manual mode.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:486  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: User Interface

One-line Description:

Sample info desciption for OQPV Fraction Collector tests contains symbols

Problem:

Following issue has been reported using ChemSTation B.02.01 OQ/PV in Verification, running enhanced Tests on the Fraction Collector:
Creating a new verification file perfom the tests and 
define in Instrument Verification some Required vials
the "Sample Info" of the Fraction collector tests mentioned tests show four bars (||||) instead of spaces in the description.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:496  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: User Interface

One-line Description:

System.NullReferenceException error may occur in ChemStation

Problem:

The Windows error System.NullReferenceException may occur during repetitive ChemStation operations (doing something > 1000 times).

Temporary Solution:

This is a Microsoft Window XP SP2 bug. For more details please go to
http://support.microsoft.com/default.aspx?scid=kb;en-us;319740

User will not use msprofile as much as testcase does (1000+ times). 

There also have a workaround available:
WORKAROUND
If this GDI leak is seriously degrading performance, you can turn off the new user interface theming in Windows XP. To do this, follow these steps: 
1. Click Start, and then click Control Panel. 
2. Click Appearance and Themes, and then click Display. 
3. On the Themes tab of the Display Properties dialog box, click Windows Classic in the Theme list. 
4. Click Apply, and then click OK 

Fix Information:

 No fix Planned


KPR#:501  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: User Interface

One-line Description:

toolbars can only be configured when displayed in separate window

Problem:

Toolbars can only be configured (e.g. selection which icons are displayed) when they are dragged to be displayed in separate window.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:517  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: User Interface

One-line Description:

option to create folders from Preferences/Paths only available in Windows 2000

Problem:

Select View -> Preferences. On the Paths tab, press one of the Add buttons. In the upcoming dialog it is possible to create a new folder only with Windows 2000.

Temporary Solution:

n/a

Fix Information:

No fix is planned for this problem.


KPR#:693  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: User Interface

One-line Description:

wrong error message: pressure sensor not connected

Problem:

Using  Firmware A.05.11 running with ChemStation B.01.03  the following has been reported: in case of a defective pressure sensor a wrong error message is displayed:
Instead of the error message "invalid pressure signal", the message " pressure sensor not connected," is coming up.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:723  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: User Interface

One-line Description:

pressure signal plot for the PrepPump is missing in Online Plot

Problem:

Using ChemStation B.02.0x the pressure signal plot for the PrepPump is missing in Online Plot as soon as the user switches to the Diagnosis view and back again.

1. In Method and Run Control online plot view 
   the pressure signal.
2. Switch view to View - Diagnosis: pressure signal 
   plot present
3. Switch back to View - Method&Run Control: no 
   pressure signal plot present

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:731  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR2

Keyword: User Interface

One-line Description:

Error Fetching CP Variable observed sometimes when starting the chemstation

Problem:

Occasionally, when the ChemStation is started or shutdown, these errors are observed:

 “FetchCPVar  Cannot talk to Application”

 “Error Fetching CP Variable   _GUIFInjUpperVial”

 “Error Fetching CP Variable   _GUIBInjLowerVial”

  “Error Fetching CP Variable  _GUIBInjUpperVial”

Temporary Solution:

It seems to be some timing issue during startup when the Sampling and Instrument diagrams are enabled in the software.

The workaround is to turn off all sampling diagrams and instrument diagrams in all 3 views of Method&Run Control, and then shut down and re-start the Chemstation.

It should be possible to turn the views back on after startup.

Fix Information:

n/a


KPR#:809  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: User Interface

One-line Description:

Splitless Mode - GC displays 'Not Ready" - Chemstation displays "Ready"

Problem:

There is a disagreement between GC and Chemstation Status for Splitless Methods with regard to Prep Run

The chemstation status should say:
   "Splitless Inlet Waiting for Prep Run"  and
      the "Not Ready" should be in Yellow for example

The status is confusing.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1045  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: User Interface

One-line Description:

Function keys not displayed on screen

Problem:

Since revision B.02.01, the available function keys are no longer displayed in Method and Run Control view.
This affects both the default function keys and additional function keys introduced via the SetFKey command.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1046  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: User Interface

One-line Description:

Initial Screen Window Size is not kept

Problem:

The Initial Screen Window Size can be changed randomly if the Initial Screen Window Size is set to Normal and then the Chemstation is closed when iconized in the taskbar.  The Chemstation.ini will have the WinSize value = Min and when the Chemstation is started, the error occurs.  If the Chemstation is closed when the ChemStation window is maximized, then the WinSize key will be set to Normal and no errors occur on start up.

So, if the ChemStation is iconized to the taskbar and the user right-clicks on the icon to close the chemstation, the autostart Macro Failed, error# 999 will occur when starting the Chemstation.

Temporary Solution:

In Chemstation.ini, if the WinSize value = Min, change the value to Normal before starting the Chemstation.

[PCS,1]
InstType=gcoffline
InstName=Instrument 1
WinSize=Normal

Use the Configuration Editor to set the Initial Screen Window Size to Normal.

Fix Information:

n/a


KPR#:1125  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.01

Keyword: User Interface

One-line Description:

network datapath not available in ChemStation Explorer after network outage

Problem:

When a network path has been configured as data path in the Preferences dialog, this data path will be available in Data Analysis view in the ChemStation Explorer.
After a network interruption and following restoration, the network path is displayed in ChemStation Explorer again, but it can not expanded anymore to access the data.

Temporary Solution:

Close and restart ChemStation.

Fix Information:

No fix is planned for this problem.


KPR#:1194  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: User Interface

One-line Description:

HeadSpace software error box is not shown until the HeadSpace SW has the focus

Problem:

Headspace software error box is not displayed until the HeadSpace application has the focus.  The user is not shown the error box until selecting the Headspace software user interface.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1230  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: User Interface

One-line Description:

ChemStation displays -1.0 during CTC cycle process pre and post run

Problem:

ChemStation B.04.01
CTC A.01.04

During the CTC process cycles of pre and post run, ChemStation Elapsed time is showing -1.0

This is not while the GC is in a run. 

Set up a CTC run with pre and post run cycle activities.  (Prewashes and Postwashes)

Run Method.

During the pre run before the GC starts, the elapsed time will show -1.0
After the GC run is completed, and the Post run CTC events begin (post washes), the GC elapsed time will show -1.0

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1319  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: User Interface

One-line Description:

ChemStation uses mixture of color settings from Windows and Configuration Editor

Problem:

The chemStation UI uses mixture of color settings from Windows and Configuration Editor.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1336  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: User Interface

One-line Description:

signal description in the online plot does not provide wavelength information

Problem:

When using RC.NET drivers for the LC instrument signal description in the online plot does not provide wavelength information for the detector signals.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1342  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: User Interface

One-line Description:

Verification view does not initialize properly

Problem:

When switching to the Verification view for the first time, this view does not initialize correctly. The icons below the menues are not displayed.

Temporary Solution:

Switching to Method and Run Control and then back to Verifation resolves the situation.

Fix Information:

 No fix Planned


KPR#:1376  Product:ChemStation  ChemStation (G2070BA)  B.04.02

Keyword: User Interface

One-line Description:

not possible to move Edit Integration Events dialog

Problem:

It is not possible to move Edit Integration Events dialog. If you try to move the dialog, only the upper part of the window moves.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:1380  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02

Keyword: User Interface

One-line Description:

wording mismatch between menu and dialog for saving sequence templates

Problem:

The entry in the menu reads: "Save Sequence Template".
The header of the corresponding dialog reads: "Save Sequence".
It should read "Save Sequence Template" in both cases.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1524  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 Aladdin

Keyword: User Interface

One-line Description:

Dropdown box for signal selection does not have a scrollbar

Problem:

In data analysis the user can select the display of individual signals via a dropdown box. This box can display 9 entries at a time. 
With a DAD and several instrument curves more than 9 signals can be available. There is no scrollbar to select more signals.

Temporary Solution:

Open the dropdown menu and use the DOWN arrow on the keyboard to select more signals.

Fix Information:

No fix currently available.


KPR#:1574  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: User Interface

One-line Description:

OpenLAB Control Panel does not prevent connecting to an instrument already  in use

Problem:

With "Fast User Switching" enabled in Microsoft Windows, a user B can connect to an instrument started by user A in a different user session.

Temporary Solution:

n/a

Fix Information:

Fast User Switching must be disabled in Microsoft Windows.


KPR#:1577  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: User Interface

One-line Description:

Slow performances observed while loading datafiles when a networked  printer is not accessible

Problem:

When a networked printer set a default printer is not responding, the ChemStation user interface may become slow when loading datafilles or when changing the scaling mode.

Temporary Solution:

N/A

Fix Information:

N/A


KPR#:1583  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: User Interface

One-line Description:

Additional information about the [ChemStation:Sequence][Save Sequence] privilege

Problem:

When the "Save Sequence" privilege is granted to a user, this user will be able to create and save Sequence Template(s) or Easy Sequence Template(s). 

Without this privilege, the "Save Sequence" and "Save Sequence As..." menus will be greyed out. In addition, the "Easy Sequence Setup" tab will not be visible. 
It is also intended that users will be able to use the "Save Easy Sequence", "Save Easy Sequence as..." and  "Save and Add to Queue" buttons from within the Easy Sequence tab.

Temporary Solution:

N/A

Fix Information:

Fixed in OpenLAB CDS C.01.06.


KPR#:1584  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02 SP1

Keyword: User Interface

One-line Description:

ALS Sampling diagram shows vial usage when using HeadSpace

Problem:

A Headspace method with the injection source set for Headspace with a back inlet connection was run on a 7890A with an ALS mounted on the front. The ALS Sampling diagram is indicating vial usage. The Headspace uses the assigned Headspace vial.

Temporary Solution:

The ALS turret has no effect on the ability of the Headspace to use all vials.

Fix Information:

 See Previous Note.


KPR#:1605  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.03

Keyword: User Interface

One-line Description:

The UI for the CTC sampler does not display the current location

Problem:

The UI for the CTC autosampler does not display the location of the current sample. Instead a vial number 18 is displayed.

Temporary Solution:

Take the current location from the sequence queue line or from the logbook.

Fix Information:

Fix planned.


KPR#:1631  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: User Interface

One-line Description:

Users have restiricted menus, although their privileges should allow all options

Problem:

There are situations where a user is missing some options, e.g the option to save a method. This option is greyed out, although the user hase the ECM-privilege to save methods.
This occurs when the first user that logs on to ChemStation does not have the privilege to see the command line. The first user that logs on to the ChemStation must have administrator privileges. Tnat is also true for a situation where the instrument configuration files are deleted (config.reg and conf_off.reg).

Temporary Solution:

Delete the instrument configuration files and log on with administrator privileges.

Fix Information:

Fixed in C.01.02 Fixed in revision B.04.03 SP1.


KPR#:1644  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: User Interface

One-line Description:

PAL Sampling diagram always shows 'Vial Status: Normal' for every vial

Problem:

The tooltip over any circle (vial) in the CTC PAL Sampler Sampling Diagram reads ‘Vial Status: Normal’, no matter what its real status is.

Temporary Solution:

n/a

Fix Information:

 See Previous Note.


KPR#:1674  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.02

Keyword: User Interface

One-line Description:

Graphic artifacts/software crashes caused by graphic settings

Problem:

Several problems have been observed on various ChemStation revisions that were the result of one common cause: Video settings that were adjusted for best appearance can cause graphic artifacts and can even cause the ChemStation software to crash.

Temporary Solution:

Adjust the display settings for best performance rather than for best appearance. This will free resources and prevent video artifacts from anti-aliasing or filtering attempts by the video driver.

The procedures below refer to Windows Classic screens and Control Panel in Classic View.


The following steps are recommended depending on the operating system:

Windows XP
-----------------

1) Go to Start > Control Panel > System. On the Advanced Tab press the settings button in the Performance section. Under Visual Effects select "Adjust for best performance". Press the OK button.
2) Go to Start > Control Panel > Display. On the Settings tab press the Advanced button. On the Troubleshooting tab set the slider for Hardware acceleration to None. Press the OK button.

Windows Vista and Windows 7
-------------------------------------------

1) Go to Start > Control Panel > System. Select Advanced system settings. On the Advanced Tab ress the Settings... button in the Performance section. Under Visual Effects select "Adjust for best performance". Press the OK button.
2) Go to Start > Control Panel > Administrative Tools. Select Services. Right click on the Desktop Window Manager Session Manager. In the context menu select Properties. Stop the service. Set the startup type to Disabled. This disables the Aero-Glass look.

Fix Information:

n/a


KPR#:1682  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: User Interface

One-line Description:

7890A, 7820A missing sampling diagram when only a rear injector is installed

Problem:

7890A, 7820A has an injector on the back.  The sampling diagram is not shown when there is only a back injector (no front injector or tray).

Temporary Solution:

n/a

Fix Information:

 Fixed in revision B.04.03 SP1.


KPR#:1695  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.03

Keyword: User Interface

One-line Description:

Change Access Level... is not available from menu bar

Problem:

Change Access Level... is grayed out and not available from the menu bar.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1740  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.01

Keyword: User Interface

One-line Description:

The logged-on user is not displayed in lock dialog

Problem:

Set authentication to Internal and create an 'everything' user
Log into Control Panel and open a 6890 session, which fails to initialize
This eventually times out and the lock dialog appears.  The logged-on user is not displayed in lock dialog. Do not know who is logged into Control Panel.

Temporary Solution:

The root cause is that the full name is not set.

Fix Information:

Fixed in revision C.01.04.


KPR#:1745  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: User Interface

One-line Description:

Finish Queue Sequence Help is hidden behind dialog

Problem:

Create a native sequence table; select RunControl\Queue Sequence... and walk through the sequence dialogs to get to the 'Finish Queue Sequence' dialog. Select Help from the button. Help opens but is hidden behind all the windows and will not come in front of the Finish Queue Sequence dialog.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1746  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: User Interface

One-line Description:

Sampling diagram for 68xx Classic driver appears, but if selected and reselected, disappears; also when switching betw Meth&Run and DA

Problem:

Connect to a 68xx Classic driver instrument. The old sampling diagram is correctly displayed in the top to bottom column. Select View\Sampling diagram to make it go away, then select again to make it reappear. Only the column reappears. The turret or tray display is not shown again in that session.

If the user starts the ChemStation in online the sample turret and run start button are shown in Method and run control. But after the user goes to Data Analysis and back to Method and Run Control the turret picture and start button disappear.

Temporary Solution:

Close and reopen the chemstation session and the turret displays again.

The RC.NET driver does not have this problem.  The user can use the RC.NET driver.  Record the IP address, delete the 68xx GC and then add it back again using the RC.NET driver.

Fix Information:

n/a


KPR#:1764  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: User Interface

One-line Description:

Run-time error ‘282’: No foreign application responded to a DDE initiate

Problem:

Run-time error ‘282’: No foreign application responded to a DDE initiate

Temporary Solution:

It seems to be some timing issue during startup when the Sampling and Instrument diagrams are enabled in the software.

The workaround is to turn off all sampling diagrams and instrument diagrams in all 3 views of Method&Run Control, and then shut down and re-start the Chemstation.

It should be possible to turn the views back on after startup.

Fix Information:

n/a


KPR#:1808  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: User Interface

One-line Description:

Values under Overview Table are not aligned with the parameter heading

Problem:

C.01.04 Chinese version, the values under Overview Table are not aligned with the parameter heading of : RT, Sig, Lvl, Amt[wt%], Area, Rsp.Factor ,Ref, ISTD, #, Compound.

Temporary Solution:

n/a

Fix Information:

Fixed in OpenLAB C.01.05.


KPR#:1855  Product:ChemStation  ChemStation (G2070BA)  B.04.03 SP1

Keyword: User Interface

One-line Description:

No sampling diagram displayed with a 7890B on B.04.03 SP1 [87] or B.04.03 SP2 [105]

Problem:

The sampling diagram is not displayed with a 7890B on B.04.03 SP1 [87] or B.04.03 SP2 [105].

Temporary Solution:

n/a

Fix Information:

Fixed in B.04.03 SP2 [106], July 2013.
(which is Hotfix01 for ChemStation B.04.03 SP2 [105]; see service note.)


KPR#:1688  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.01

Keyword: User Interface , Data Analysis

One-line Description:

Preferences set in offline overwrite online Preferences and vice versa

Problem:

Preferences set in offline overwrite online Preferences and vice versa

Temporary Solution:

n/a

Fix Information:

 See Previous Note.


KPR#:1717  Product:ChemStation  ChemStation (G2070BA)  B.04.03

Keyword: User Interface , Data Analysis

One-line Description:

Buttons for manual integration are not accessible, while they can be selected through the menu

Problem:

Under certain conditions it has been observed that users cannot use the buttons to perform the manual integration functions, but the functions are available via the integration menu.
The conditions were that a snapshot from a running datafile was taken and the system was connected to ECM.

Temporary Solution:

No temporary solution available. Upgrade to revision B.04.03 SP1.

Fix Information:

Fixed in revision B.04.03 SP1.


KPR#:1763  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: User Interface , SEQ , ACQ , ECM Integration/OpenLab Option

One-line Description:

In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.

Problem:

By OpenLAB ChemStation Edition C.01.04 the requirement for ECM's remote data path (LCDF) became more strict than before. If for editing the Sequence Parameters dialog is opened, and a backslash at the LCDF's begin is missing, the builtin check of this dialog-box will demand a manual correction by the user. This can happen especially once loading a sequence template file from an earlier version of OpenLAB ChemStation Edition.

The later release of OpenLAB ChemStation Edition (like e.g. C.01.05) will tolerate and correct the missing leading backslash at ECM's remote datat path (LCDF) in Sequence Parameters dialog automatically.

Execution of sequences by OpenLAB ChemStaation Edition is not affected. I.e. sequence templates do NOT have to be corrected, neither for sequence execution nor for sequence reprocessing.

Temporary Solution:

Leaving the opened Sequence Parameters Dialog by OK is interdicted until all parameters are in valid ranges. This now for OpenLAB ChemStation C.01.04 requires the remote data path (LCDF) for ECM to start with a leading backslash. If at some sequence template (e.g. from an earlier version of OpenLab ChemStation Edition) this character was missing, it now has to be inserted. Only then the Sequnece Parameters dialog can be successfully closed and left with OK-button.

Fix Information:

n/a


KPR#:77  Product:ChemStation  ChemStation (G2070BA)  (None)

Keyword: Utilities

One-line Description:

Hyperseq continues running between multiple sequences for one plate after stop

Problem:

If a hypersequence that contains of several sequences for one plate, "Stop Hypersequence" does not stop the hypersequence after the current sequence, but only after all sequences on the plate are completed.

Temporary Solution:

Use "Stop Sequence" to manually stop each single sequence running on the plate or use "Abort" directly after completion of the current sequence.

Fix Information:

There is no fix planned for this problem.


KPR#:199  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Utilities

One-line Description:

B.01.01 and PC Anywhere is giving conflict

Problem:

Installing ChemStation Rev. B.01.01 ChemStation on a system using Symantec's PC Anywhere version 10  following error messga come up during starting the ChemStation:

 "Entry point not found in util.dll."

PC Anywhere and the 32bit ChemStation software have a dll with the same name (util.dll).

Temporary Solution:

Symantec PC Anywhere is not supported using ChemStation Rev. B.01.01 

Renaming the PC Anywhere's dll (or deinstalling the PC Anywhere software)  allowed ChemStation to load.

Path information:
 C:\ProgramFiles\Symantec\PC Anywhere

Fix Information:

n/a


KPR#:209  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Utilities

One-line Description:

Excel workbook issue with "special links" and chemstation.xls reports

Problem:

With ChemStation Rev. B.01.01 and Excel following issue has been reported:

The "paste special link" in Excel is not working immedeatly for ChemStation created linked .xls files:

Conditions:
Creating a ChemStation report using the 
"print to file" option in specify report a report01.xls file including all result information is created.

Opening this file and 
1. create a "new" excel workbook
2. copy the line of interest in the chemstation
   report01.xls and paste that information into the   
   new workbook using the "paste special", "paste link" function in Excel/Edit. Save the new workbook under a specific name.
3. Close Excel
4. Open the new created workbook: a message appears and informs the user that this Excel workbook is linked with the report01.xls file and if this link should be executed: Answer YES
5. The system comes back telling: Unable to read data.

Temporary Solution:

The following workaround can be used:

Open the ChemStation  report01.xls file, than save it and close the file. 
Now the link will be executed.

For audit reasons it would be  a possibility to 

1.print the excel report before and after saving the file in excel, to ensure that no changes have been taken.

2. save the excel file under a different name. The link issue will be solved due to saving the file.
In addition the original file will still be available for comparison.

Fix Information:

No fix is planned for this problem.


KPR#:246  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Utilities

One-line Description:

wrong pressure description using Japanese fonts

Problem:

Using english ChemStation B.01.01 and 
English Operating system, but changing the fonts to 
Japanese the values in the e.g. pressure fields 
will be set to a strange value.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:247  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Utilities

One-line Description:

Would like to be able to view LC/MS fraction and spectra in same View

Problem:

Some customers have asked for the ability to view fraction information in the ChemStation’s Data Analysis Spectral View.  This way they can see both the start and stop times of their fractions, and select spectra from the peaks.

Temporary Solution:

n/a

Fix Information:

 No fix Planned


KPR#:264  Product:ChemStation  ChemStation (G2070BA)  B.01.01

Keyword: Utilities

One-line Description:

XML Import: Import of LimsId fails if sample amount has been set in import XML f

Problem:

Using ChemStation B.01.01x following issue has been reported using the XML functionality: 
As soon as a Sample Amount is predefined in the XML import file, the import of LimsId fails during "import worklist". 

Temporary Solution:

The problem is only seen using a comma instead of a point in the number section Sample Amoung:
Because the XML file gets converted into CSV-file (literally COMMA SEPARATED...) the values for Sample Amount which contain "," (e.g.: "274,42") will break the structure of the file and confuse the parser engine. So the values have to be setted using a point.

or

After the import of the XML worklist the sequence table need to be checked and the Sample Amount need to be added manually.

Fix Information:

No fix is planned for this problem.


KPR#:389  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Utilities

One-line Description:

no annotation or move annotation in 3D overlay window possible

Problem:

Following issue has been reported using the  3D overlay window in ChemStation B.01.03.:
In the 3D overlayed moded It is not possible to
1. add annotations
2. move annotations


Temporary Solution:

Following workaround does solve the issue:

Both issues can be solved, when the chromatogram is ins “overlayed mode”,”display each in full scale”. Only than go for the 3D overlay mode.

If this is not working, select the “draw line” tool (3rd tool in the annotation list on the right site, to be enabled by Grpahics tool bar, Annotations) and click into the 3D overlay screen, than change to the 
e.g. the annotation item and the annotations can be written .

Fix Information:

 No fix Planned


KPR#:477  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Utilities

One-line Description:

OQ/PV test Injector Precision test show pass instead of failed

Problem:

Following issue has been seen executed the OQ/PV test Injector Precision on a micro system with a 20ul/min sensor, using ChemStation B.02.01:
If the measure and calculation of Carry Over Area and Height precision result in negative values, the test is shown as passed.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:484  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Utilities

One-line Description:

Holmium test passed although hardware fails

Problem:

When an hardware error occurs during execution of the Holmium test, the Diagnosis view will nevertheless display the results from the last time the Holmium test has been successfully.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:488  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Utilities

One-line Description:

Cap. Pump Flow Sensor Calibration  required pump switched on

Problem:

Following issue has been reported using ChemStation B.02.01 performing the flow sensor calibration test in Diagnosis: 

If the pump is not switched on while starting the the Capillary Flow Sensor

Temporary Solution:

Ensure to switch on the pump prior running the 
Cap. Pump Flow Sensor Calibration test.

Fix Information:

n/a


KPR#:489  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Utilities

One-line Description:

Setup window for modules grey out after performing OQPV tests

Problem:

Following issue has been reported for some OQPV tests running on ChemStation B.02.01. After closing the a performed OQPV test, the setup dialog box of the used module might greyed out.

Temporary Solution:

If the problem occurs, please load another method, e.g. default method and all set up dialog boxes are available again.

Fix Information:

 No fix Planned


KPR#:541  Product:ChemStation  ChemStation (G2070BA)  B.01.03

Keyword: Utilities

One-line Description:

not possible to open a Purify study when the Purify Data Base Path is set to C:\

Problem:

If the Purity Data Base Path is set to C:\ (Options -> Set Data Base Path), the following error will occur when trying to open a study: "Critical error 52 in frmStudyOpen.Form_load Bad file name or number". Afterwards Purify crashes.

Temporary Solution:

Do not set the Data Base Path to C:\, but use any folder on C:\ instead.

Fix Information:

No fix is planned for this problem.


KPR#:588  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Utilities

One-line Description:

Verification: unit for RID noise values

Problem:

When the Noise and Temperature Stability test for an RID detector is set up and the user enters the limits for noise values, the unit is given as mAU, but it should be nRIU. The same problem is also observed in the report.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:589  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Utilities

One-line Description:

Extracted Chromatograph changes when setting Bandwitdh and exporting

Problem:

Following problem has been reported using the isoplot in ChemStation B.02.01/B.02.01 SR1 to extract signal to Data Analysis.
The selected extracted signal in the isoplot is reduced by one nm as soon as a Bandwith is set. This reduced signal is than exported to Data Analysis.

Temporary Solution:

n/a

Fix Information:

This behaviour is correct. Because not every wavelength is inside the library the closest avaible one is choosen. Depending on the acquired spectras and on the choosen bandwidth not all wavelength are possible.


KPR#:594  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Utilities

One-line Description:

1100 Diagnosis: Draw action performs red message

Problem:

Following item has been reported on ChemStation B.02.01/B.02.01 SR1 using the Autosampler tests in diagnosis view:

Performing the action DRAW or PLUNGER HOME in the Injector steps, the action icon turns red under the following circumstances:

Entering directly a volume 100µL the action icon turns red.

Temporary Solution:

The problem is only graphical nature. Incrementing the draw volume by steps of 10µL till getting 100µL, there is no problem in the User interface. 

Fix Information:

n/a


KPR#:602  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Utilities

One-line Description:

K Prime may not be calculated using void volume

Problem:

Perform a run in ChemStation Rev. B with the column information set, and create a performance report:

If a Void Volume is specified in the column, you ahve to take care to specify a void volume that corresponds to a void time that is within the actual run time of the method. Otherwise, K Prime is not calculated.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:605  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Utilities

One-line Description:

zoomed peaks may be printed incompletely on reports

Problem:

For some reports printed with various Printer drivers using ChemStation Rev. B following problem has been reported, printing a zoomed chromatogram:

If one peak is very big and the user is zooming into the smaller peaks, the big peak might be printed incompletely.

Temporary Solution:

The problem is printer driver related, using the printer driver LJ 4200 PCL 5e or the Postscript driver, the problem is not present.

Fix Information:

It is currently not planned to fix this problem.


KPR#:648  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Utilities

One-line Description:

Annotation Menu disables some of the Graphic, Signal options

Problem:

Following item using the Annotation Menu in ChemStation B.02.01 SR1 has been reported:

As soon as the annotation menu is opened once, the annotation menu settings are the master of the graphics settings used as well in  Graphic, Signal options. Settings defined in the annotation tool set are used.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:673  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Utilities

One-line Description:

Wrong compressibility setting in ChemStation when using Handheld Controler

Problem:

Following item has been reported using ChemStation B.02.01/B.02.01 SR together with a Instand Pilot Handheld Controler:

Defining a setting for the solvent compressibility calibration in the ChemStation and then change the value using the Handheld Controler "Instand Pilot" the  Auxiliary dialog checkbox "use enhanced comp. calibration" is still checked afterwards in ChemStation. The correct value for the composition, set using the Instand Pilot, is shown (grayed out) in the edit field.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:675  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Utilities

One-line Description:

Use of external contacts board on 1367C injector no binary output possible

Problem:

Following problem has been reported using  the external contacs board in a  G1367C injector. Selecting the "contacts" option in "more injector"  and select "use BDC port for" "Binary Output" .

After pressing "OK" the following  error mesage occured: "The Parameters could not be transmitted correctly! Please re-check your settings. "

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:678  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Utilities

One-line Description:

Invalid fraction start location

Problem:

Following issue has been reported using fraction collection on ChemStation B.02.01 : If the first sequence table line is set up a reserved location for fraction collection, there error message "invalid start location" is coming up.

Temporary Solution:

n/a

Fix Information:

No fix is planned for this issue.


KPR#:679  Product:ChemStation  ChemStation (G2070BA)  B.02.01

Keyword: Utilities

One-line Description:

Print report to XLS file strange characters are present

Problem:

Generating XLS reports in ChemStation B.02.01 SR1, the generated file REPORT02.CSV contains  two strange charatcers whem loaded with Excel.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:680  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Utilities

One-line Description:

Purify: sample description only once by wizard

Problem:

Following issue has been reported using ChemStation and Purify Add-On software:
When creating a new studies it is possible to add an information in the "Description" field. Setting up multiples studies based on the same original study with various descriptions, the sample in general get the description of the first study assinged. The samples in unused vial positions will get the new sample description .

Temporary Solution:

Please specify or change the sample description in Study List for each sample and not in the general setup.

Fix Information:

n/a


KPR#:681  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Utilities

One-line Description:

Started sequence with wrong vial position after changing trays

Problem:

Following issue is been reported using the Micro ALS on ChemStation B.02.01/B.02.01SR1:
After exchange a 100 vial tray with a 15 vial tray by inserting the 15 vial tray at right or left position, than then running the sequence using the tray on the left position. An  error message "location: vial must be  in.... range " occurs, if the vial position is set to higher than 15 in sequence line.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:720  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Utilities

One-line Description:

zooming reset when copying to clipboard

Problem:

Following problem with zooming has been reported using ChemStation: 

If the settings in Data Analysis, Graphics, Signal Options, Multi-Chrom., are set to "seperate" and "each in full scale" , the zoomed file is set back and the full chromatogram is exported.

Temporary Solution:

Ensure to set the settings in Data Analysis, Graphics, Signal Options, Multi-Chrom., Scale to "all in same scale", the zoomed file is exported.

Fix Information:

n/a


KPR#:737  Product:ChemStation  ChemStation (G2070BA)  B.02.01 SR1

Keyword: Utilities

One-line Description:

No > 10 is not accepted as vial number for Diagnosis single step tests for AFC

Problem:

The following problem has been reproted for ChemStation B.02.01 using in Diagnosis the Fraction Collector Single Step:

If move to Location is set for Vial 1 - 9, the test is executed.  If Move to Location is set for higher vials the error  ‘Location not Valid’ error occurred, entering number above 10.

Temporary Solution:

Please use the required syntax:

Vial 1, Vial 2,...., Vial 10, ....

Fix Information:

n/a


KPR#:791  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: Utilities

One-line Description:

Excel Seqence summary macro is failing

Problem:

The folliwng problem has been seen in ChemStation B.03.01 printing to the specific excel sequence summary template HPSEQ00  Template using a post-sequence macro. The macro is not executed correctly, no report is created.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:854  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: Utilities

One-line Description:

Prep-Pumps online plot signals need to set again after changing views

Problem:

All Prep pump signals disappear from online plot when switching from Method and Run Control View to Diagnosis and back to Method and  Run Control.

G1361A   A.06.04 x2
G2260A   A.06.04
G1365B   A.06.04
G1364C   A.06.05

This symptom is seen only when switching to Diagnosis view. Switching to other views (Data Analysis, Report Layout, OQ/PV) does not result in this problem.

Temporary Solution:

Reselect the Prep Pump Online Signals again.

Fix Information:

n/a


KPR#:884  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: Utilities

One-line Description:

cannot load batch method, if method (on another drive) is specified

Problem:

Following issue has been reported using batch menu in ChemStation B.02.01 and higher:

Loading a batch  . b , it is possible to select a method to be used with batch. The user can pick a method, and the method is than loaded together with the batch file.

If in Data Analysis a method is currently loaded on the  e.g. D drive (methodA.m) , and the user  select a method on the C drive (methodB.m), the system is NOT able to load the selected method.
The system is loading the method from the sequence container. 

Temporary Solution:

In the Data Analysis view, load a method 
which is in the same drive and directory like the method you want to pick for the batch view, and than it is possible to load a method from another drive.

Fix Information:

No fix is intended for this problem.


KPR#:958  Product:ChemStation  ChemStation (G2070BA)  B.03.01

Keyword: Utilities

One-line Description:

ChemStation crash during MS data AIA export to network drive

Problem:

Export of MS data to AIA format directly onto network drive crashes ChemStation. Afterwards ChemStation cannot be restarted, and the ChemMain.exe process has to be terminated in Windows Task Manager.

Temporary Solution:

Export the AIA data to a local drive and transfer them to the desired server location later on.

Fix Information:

n/a


KPR#:1012  Product:ChemStation  ChemStation (G2070BA)  B.03.02

Keyword: Utilities

One-line Description:

English Acrobat Reader install package on ChemStation CD 2 not working

Problem:

Installation packages for Acrobat Reader are provided on ChemStation CD 2 in English, Chinese, and Japanese.
However, it is impossible to install Acrobat Reader with the English package. Installation fails with error message: "This program is used internally by PackageFromThe Web. It should not be executed directly."

Temporary Solution:

Download Acrobat Reader from Acrobat webpage.

Fix Information:

This problem does no longer exist on the current B.04.03 ChemStation DVD.


KPR#:1134  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Utilities

One-line Description:

Modified or corrupt base reference file IQTRef.xml causes crash of IQT Report

Problem:

A modified or corrupt base reference file IQTRef.xml causes a crash of IQT Report tool.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1135  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Utilities

One-line Description:

with missing base reference file IQTXref.xml the IQT is still reported as passed

Problem:

If previously a sucessfull IQT was performed, a further IQT Report will still show the result as "Passed" in case the base reference file IQTXref.xml was deleted. The previously generated report is shown instead of a new report.

Temporary Solution:

By the date and time stamp the user is able to detect that the report originates from a previous run of the IQT Report tool.

Fix Information:

n/a


KPR#:1233  Product:ChemStation  ChemStation (G2070BA)  B.04.01 SP1

Keyword: Utilities

One-line Description:

Chinese/Japanese ChemStation: cannot import .csv file to sequence

Problem:

Japanese or Chinese Excel will by default save .csv files in ASCII format. ChemStation Sequence Import requires the .csv files to be in Unicode format.

Temporary Solution:

Open the .csv file from Excel with Notepad and save it as .csv file (coding: Unicode).

Fix Information:

No fix is planned for this problem.


KPR#:1263  Product:ChemStation  ChemStation (G2070BA)  B.04.01

Keyword: Utilities

One-line Description:

IQT Report may fail after switching to Diagnosis with invalid files

Problem:

After switching to Diagnosis view, IQT Report may fail with the following files reported as invalid because of checksum mismatch:
\chem32\lc\800\dgdetect.ini
\chem32\lc\800\dgdetec5.ini
This problem occurs with LC/MS systems. The change in the two .ini files causing the checksum mismatch is correct behavior.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1419  Product:OpenLAB CDS ChemStation Edition  M83xx  B.04.02

Keyword: Utilities

One-line Description:

compound custom fields are only exported to result.xml when the compound was found

Problem:

The values of compound custom fields are only exported to the result.xml file for a run, if the compound is actually found in this run.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#:1772  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Utilities

One-line Description:

FPD Checkout Macros no longer work since there is no "autoint.mac" with OL CDS ChemStation C.01.04

Problem:

FPD checkout macro used for FPD sample test.

This FPD macro uses "autoint.mac" .
OL CDS ChemStation C.01.04  no longer includes "autoint.mac" in the CORE directory.

The following message occurs when loading the FPD Checkout macro:

 "Could not open file: C:\CHEM32\CORE\AUTOINT.MAC"

Temporary Solution:

Copy "autoint.mac" from OL CDS ChemStation C.01.03 and add it to the Core directory of OL CDS ChemStation C.01.04 .

Fix Information:

The macro code has been updated and does the following:

•Checks to see if the “AutoIntegrate” macro has already been loaded.  If so, it does nothing. 
•If “Autointegrate” macro has not been loaded, it looks for the file autoint.mac.  If the file is found, it is loaded.
•If autoint.mac is not found, it looks for the file sig_integ.mcx.  If that file is found, it is loaded.

The new macros work with or without autoint.mac.  Autoint.mac is found in the CORE directory up until C.01.03.  

Beginning in C.01.04, autoint.mac is no longer included in Core.  Autoint.mac is now part of sig_integ.mcx.

The eLibrary User Contributed Software has been updated with the updated macros.  They are also attached.


KPR#:1616  Product:OpenLAB CDS ChemStation Edition  M83xx  C.01.04

Keyword: Utilities , Spectra Analysis , MSD Analysis

One-line Description:

Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group

Problem:

A shortcut to the ChemStation Real-Time Display is not included in the OpenLAB CDS ChemStation Edition program group.  However, the Real-Time Display can still be accessed at:

C:\Chem32\CORE\ChemRTD.exe

Temporary Solution:

Manually create a shortcut pointing to

C:\Chem32\CORE\ChemRTD.exe

to access the ChemStation Real-Time Display

Fix Information:

To be resolved in a subsequent OpenLAB CDS ChemStation Edition release